Sie sind auf Seite 1von 6

Cyclic loading of implant-supported prostheses: Changes in component fit

over time
Donna M. Hecker, DDS, MS,a and Steven E. Eckert, DDS, MSb
School of Dentistry, University of Minnesota, Minneapolis, Minn.; Mayo Clinic, Mayo Graduate
School of Medicine, Rochester, Minn.

Statement of problem. Dental literature suggests that an implant-supported prosthesis must exhibit a passive
fit to prevent implant fracture, component breakage, and screw loosening. From a practical standpoint, passive fit
is impossible to achieve; instead, minimal misfit may be the clinical goal. To date no specific range of misfit (below
which problems are minimal and above which catastrophic failure occurs) has been established.
Purpose. The purpose of this study was to determine whether the fit of an implant-supported prosthesis changes
through cyclic loading and to quantify the amount of change between the gold cylinder and implant abutment
over time.
Materials and methods. Fifteen implant-supported frameworks were fabricated with conventional casting
techniques and were cyclically loaded under 3 different loading conditions. Five frameworks were loaded on the
anterior portion of the framework, 5 were loaded on the left unilateral posterior cantilever, and 5 were loaded
bilaterally on the posterior cantilevers with a servohydraulic testing machine. A cyclical load of 200 N was applied
to each framework for up to 200,000 cycles. Linear measurements were made in micrometers of the gap between
the prosthetic cylinder and the implant-supported abutment at 4 predetermined reference points. These mea-
surements were recorded before the application of the cyclical load, after 50,000 cycles, and after 200,000 cycles.
A repeated measures of variance model was fit separately to the data for each load location (P⬍.05).
Results. There was a significant (P⫽.024) decrease in gap dimensions at individual reference points and a
significant (P⫽.031) decrease in the average gap when the load was applied to the anterior portion of the
framework. When the load was applied unilaterally or bilaterally on the posterior cantilever, significant gap closure
was not observed (P⫽.33 and P⫽.35, respectively).
Conclusion. Within the limitations of this study, the fit between the prosthetic superstructure and the implant-
supported abutment changed when simulated functional loading of the anterior portion of the prosthesis was
performed. Simulated functional loading applied unilaterally or bilaterally to the posterior cantilever portion of
the prosthesis did not result in changes in the measured gap sizes. (J Prosthet Dent 2003;89:346-51.)

CLINICAL IMPLICATIONS
This in vitro study suggests that force application to an implant-supported prosthesis may alter
the fit of the prosthesis to its respective supporting components. These changes may be cumulative
over time, having clinical implications toward maintenance of screw joint integrity.

T he recognition of a biocompatible union between


bone and alloplastic implant material creates a number
Technical problems continue to frustrate clinicians.
These include inability to achieve intimate fit on fabri-
of new applications to the field of dentistry. Osseointe- cation of the prosthetic framework, inability to correct
gration offers new treatment options for the edentulous the misfits, and difficulty in developing an occlusal
and partially edentulous patient. The use of dental im- scheme that will not overload the assembly.8
plants to support and retain dental prostheses has been The objective of achieving passive fit of the prosthesis
demonstrated to be clinically efficacious.1-4 Although (gold cylinder) to the abutment is addressed by several
the bone-to-implant interface may be reliable,5-7 clinical authors.9-18 Branemark13 suggested that components
complications can and do occur at the prosthetic level. have no more than a 10-␮m misfit, but modern dental
technology is unlikely to consistently achieve this level of
accuracy. Many manufactured components do not dem-
Presented at the Academy of Prosthodontics Meeting, Santa Fe, NM, onstrate this level of accuracy of fit.19,20 Consequently,
May 2001. misfit of prostheses is a clinical reality, but the amount of
a
Director of Maxillofacial Prosthetic Services and Assistant Clinical
Dental Specialist, Department of Restorative Sciences, University
misfit that can be tolerated without adverse mechani-
of Minnesota. cal21-30 or biologic9,31-34 complications has yet to be
b
Associate Professor, Mayo Graduate School of Medicine. determined.

346 THE JOURNAL OF PROSTHETIC DENTISTRY VOLUME 89 NUMBER 4


HECKER AND ECKERT THE JOURNAL OF PROSTHETIC DENTISTRY

Fig. 1. Model testing system included 5 implants arranged in


arc, with uniform (10 mm) inter-implant spacing and fulcrum Fig. 2. Cylindrical sleeve custom-machined to ensure uni-
line-anterior abutment distance of 9 mm. form identification of reference points scribed on abutments
at every 60 degrees.

This study evaluated the changes that take place in


the gap between the implant and the prosthetic compo-
nents after cyclic loading of prostheses. The goal was to
evaluate changes qualitatively and quantitatively.

MATERIAL AND METHODS


A model system was designed to allow placement of
five 3.75- ⫻ 10-mm implants (Branemark System; No-
bel Biocare, Yorba Linda, Calif.) into a rigid base. The
system used 5 implants arranged in an arc, with uniform
(10 mm) interimplant spacing, a fulcrum line-anterior
Fig. 3. Graphic representation of reference points labeled a,
abutment distance of 9 mm, and a distal extension can-
b, c, and e. Reference points d and f eliminated because of
tilever of 18 mm (Fig. 1). The model system included lack of access with microscope.
load cells (Model MB-250; Interface, Scottsdale, Ariz.)
attached to each implant for the purpose of measuring
applied loads and detecting any changes in load caused 4 different positions, indicating preselected reference
by change in fit of the prosthesis. points. A cylindrical sleeve with a slot every 60 degrees
Polysiloxane (Reprosil; Caulk/Dentsply, York, Pa.) was custom-machined to ensure uniformity between
impressions of the implants were made by use of an open placement of the engraved lines on the abutments (Fig.
custom tray to engage direct implant transfer abutments 2). Three engraved lines were made on the facial aspect
(Nobel Biocare) that had been rigidly connected with of the abutment and 1 on the lingual aspect. These
autopolymerizing resin (GC pattern resin; GC America, positions were consistent on all abutments and were
Chicago, Ill.). After the fabrication of master casts, new labeled a, b, c, and e, respectively (Fig. 3).
standard 4-mm abutments were placed on each of the The abutments were attached to the implants in the
implant analogues. The abutments were labeled and predetermined position and tightened to a torque of
marked to ensure the exact position on each hex could 20 Ncm with a torque driver (Machine, standard 21
be duplicated on the model testing system. mm, Nobel Biocare). The frameworks were attached to
Fifteen frameworks were waxed, cast, finished, and the abutments with the gold retaining screws supplied
polished with conventional lost-wax laboratory tech- by the manufacturer (Nobel Biocare) and were tight-
niques. Frameworks incorporated standard 4-mm gold ened manually with the slotted driver (Slot, 22/37; No-
cylinders (Nobel Biocare). Framework design used an bel Biocare). An electronic torque wrench (Torque mea-
L-shaped configuration in cross-section, with a lingual surement device, custom; NK Biotechnical Corp,
vertical rise of 6 mm and horizontal dimension of Minneapolis, Minn.) with digital read-out capability was
8 mm.35-37 No attempts were made to improve the fit of used to tighten the gold screws to the recommended
the frameworks by sectioning and soldering. A manual preload force of 10 Ncm. These values were recorded
engraving pen was used to mark lines on the abutment at before load application.

APRIL 2003 347


THE JOURNAL OF PROSTHETIC DENTISTRY HECKER AND ECKERT

Fig. 5. View through microscope demonstrates 10-␮m verti-


cal discrepancy (gap).

Fig. 4. Frameworks were evaluated for vertical discrepancies Fig. 6. View through microscope demonstrates 147-␮m ver-
between components with micron microscope. tical discrepancy (gap).

A micron microscope (Model # FS-60; Mitutoyo implants within each of the 5 frameworks). A repeated
Corp, Kawasaki, Japan) with a traveling stage and measures analysis of variance model was fit separately to
digital measurement read-out system (VRZ-720-B the data for each of the 3 load locations. In this analysis,
X-Y; Heidenhain, Schaumberg, Ill.) was used to visu- the gap changes that occur with cycling were different
ally aid in quantifying the vertical space between the between the 4 reference points (P⬍.05).
abutment and gold cylinder at the mating surface
(Fig. 4). The vertical discrepancies were measured and
recorded before cycling, after 50,000 cycles, and after RESULTS
200,000 cycles. Measurement values were recorded in There was a significant decrease in gap measurements
micrometers to a resolution of 0.0005 in/0.01 mm considering each reference point separately when the
(Figs. 5 and 6). load was applied to the anterior portion of the frame-
The measurements were made at 4 different locations work (P⫽.024), but not when the load was applied uni-
around the abutment, to quantify the change in misfit at laterally on the posterior cantilever (P⫽.33) or when the
specific locations. These individual points could also be load was applied bilaterally on the posterior cantilever
averaged to quantify change in a planar dimension. The (P⫽.35). When the load was applied to the anterior
values were recorded in a laboratory notebook and then portion of the framework, there was a significant change
entered into a spreadsheet for subsequent analysis. with cycling in the gap measurements at reference point
The summary of the distributions of the gap values e, which corresponds to the lingual aspect of the abut-
(averaged over reference points, implant positions, and ment (P⫽.023). The gap measurements did not change
frames by load location and cycling interval) is shown in significantly with cycling at any of the other 3 reference
Table I. The summary statistics are based on 25 values (5 points.

348 VOLUME 89 NUMBER 4


HECKER AND ECKERT THE JOURNAL OF PROSTHETIC DENTISTRY

Table I. Summary of the distributions of gap measurements (averaged over reference points, implant positions and frames)
by load location and cycling interval

Geometric
Load location Cycle mean Median Mean (std) Minimum Maximum 25th percentile 75th percentile

Unilateral Posterior Pre-load 0.067 0.076 0.108 (0.092) 0.009 0.334 0.038 0.142
Unilateral Posterior 50,000 0.062 0.068 0.105 (0.095) 0.007 0.334 0.033 0.150
Unilateral Posterior 200,000 0.066 0.069 0.104 (0.091) 0.010 0.325 0.038 0.131
Anterior Pre-load 0.054 0.047 0.082 (0.077) 0.014 0.346 0.039 0.115
Anterior 50,000 0.052 0.049 0.082 (0.082) 0.013 0.378 0.035 0.110
Anterior 200,000 0.049 0.048 0.075 (0.070) 0.014 0.299 0.025 0.123
Bilateral Posterior Pre-load 0.047 0.052 0.078 (0.060) 0.005 0.199 0.034 0.124
Bilateral Posterior 50,000 0.042 0.050 0.075 (0.059) 0.007 0.212 0.034 0.122
Bilateral Posterior 200,000 0.041 0.055 0.071 (0.058) 0.006 0.209 0.027 0.116

When the load was applied bilaterally to the posterior these methods require visualization of the space be-
cantilevers, there was a significant change with cycling in tween 2 components. Tactile evaluation of fit is per-
the gap measurements at reference point e (P⫽.003). formed by placing the prosthesis in contact with its re-
The gap measurements did not change significantly with spective abutments and feeling for even seating of the
cycling at any of the other 3 reference points. prosthesis. Once the initial seating is confirmed, retain-
In an attempt to summarize the gaps across the 4 ing screws are alternately tightened and loosened to de-
reference points on the abutment, the average of these termine whether movement of the prosthesis occurs
gap measurements per abutment was used. Averaging while the individual screws are secured.
the 4 individual reference points allows inferences to be The visual and tactile methods of evaluation attempt
made about changes occurring in the planar dimension. to bring objectivity to the evaluation process, but both
A repeated measures analysis of variance model was fit retain a level of subjectivity. Conversely, laboratory
separately to the data for each of the 3 load locations. methods for measuring fit 3-dimensionally at the im-
There was evidence of a significant decrease in the aver- plant-prosthodontic interface may eliminate this level of
age gaps due to cycling when the load was applied to the subjectivity. There are essentially 3 methods for labora-
anterior portion of the framework (P⫽.031); the aver- tory assessment: stylus contact techniques,16 laser video-
age gaps were significantly less after 200,000 cycles than graphy,17 and photogrammetry.18 Each method pro-
the average gaps before testing. vides precise measurements to the level of approximately
3.5 ␮m. Of course, any discrepancy between the oral
DISCUSSION cavity and the master cast will continue to create clinical
The purpose of this study was to investigate changes misfit of the prosthetic frame.
that occur in fit at the gold cylinder to implant abutment Although there are many ways to assess 3-dimen-
interface over time. The investigation was initiated in sional fit of prostheses to abutments,16-18 the sophisti-
response to the recognition that there is little likelihood cated testing equipment required for such testing is not
of achieving a perfect fit between these components with readily available. The use of more common measuring
current dental technology. Given the inevitable nature techniques still provides information on the relative fit
of imperfection in fit, it seems prudent to understand the or misfit of a prosthesis while recognizing the fact that
effect of cyclic loading on the fitting surfaces of the this is not as precise as the other methods.
implant-supported prosthesis. Another factor that must be considered in the fabri-
In evaluating the determinants of a passively fitting cation of prostheses is the reproducible fit of the prema-
prosthesis, several factors should be considered. These chined components. These components are machined
include the definition of passive fit and the accuracy of by the manufacturer to ensure a specific tolerance. Ma-
the procedures used to achieve this degree of passivity. chining tolerances define the limit of acceptability when
Other factors include the method by which the fit is assessing the measurement of a component feature in
evaluated and the role of the premachined components. relation to the specified design value target. There can
Numerous methods exist for evaluating framework be no consistently greater accuracy of the final prosthesis
fit. Different methods apply to the clinical and labora- fit to its abutments than the level of machining toler-
tory settings. Both settings involve use of visual or tactile ance. Ma et al21 demonstrated a range of tolerance for
inspection. Visual inspection may include the 1-screw the implant components (abutment, gold cylinder, im-
test,14 use of disclosing medium,15 fabrication of a cast pression coping, and brass abutment analog) in the hor-
verification jig,20 or use of radiographic films.22 All of izontal plane from 22 to 100 ␮m. This study also dem-

APRIL 2003 349


THE JOURNAL OF PROSTHETIC DENTISTRY HECKER AND ECKERT

onstrated a discrepancy between the machining 8. Zervas PJ, Papazoglou E, Beck FM, Carr AB. Distortion of three-unit
implant frameworks during casting, soldering, and simulated porcelain
tolerances of the prosthetic components and the labora- firings. J Prosthodont 1999;8:171-9.
tory analogs. The machining tolerance of brass analog 9. Rangert B, Jemt T, Jorneus L. Forces and moments on Branemark implants.
and gold cylinder was significantly larger compared with Int J Oral Maxillofac Implants 1989;4:241-7.
that of the abutment and gold cylinder. This level of 10. Rangert B, Gunne J, Sullivan DY. Mechanical aspects of a Branemark
implant connected to a natural tooth: an in vitro study. Int J Oral Maxil-
discrepancy may result in a prosthesis that appears to fit lofac Implants 1991;6:177-86.
in the laboratory, whereas the same prosthesis may have 11. Weinberg LA. The biomechanics of force distribution in implant-sup-
a larger misfit in the clinical setting. Rubenstein and ported prostheses. Int J Oral Maxillofac Implants 1993;8:19-31.
12. Zarb GA, Schmitt A. The longitudinal clinical effectiveness of osseointe-
Ma19 found similar results when comparing components grated implants: the Toronto study. Part III: Problems and complications
for titanium laser-welded and standard cast implant encountered. J Prosthet Dent 1990;64:185-94.
components. A range of 23.1 to 51.7 ␮m for machining 13. Branemark PI, Zarb G, Albrektsson T. Tissue-integrated prostheses: os-
tolerances was reported. seointegration in clinical dentistry. Special edition for Nobelpharma. Chi-
cago: Quintessence; 1987. p. 268-71.
This study demonstrated that under specific loading 14. Jemt T. Failures and complications in 391 consecutively inserted fixed
conditions, the gap between the implant abutment and prostheses supported by Branemark implants in edentulous jaws: a study
the prosthesis can undergo significant change. In this of treatment from the time of prosthesis placement to the first annual
checkup. Int J Oral Maxillofac Implants 1991;6:270-6.
study, the change resulted in smaller gaps after cyclic 15. Yanase RT, Binon PP, Jemt T, Gulbransen HJ, Parel SP. Current issues
loading. This change is probably due to wear at the forum: How do you test a cast framework for for a full-arch fixed implant-
component interface, because some point of contact supported prosthesis? Int J Oral Maxillofac Implants 1994;9:469-74
must prevent the components from achieving the man- 16. Tan KB, Rubenstein JE, Nicholls JI, Yuodelis RA. Three-dimensional anal-
ysis of the casting accuracy of one-piece, osseointegrated implant-retained
ufactured fit. Once this wear occurs and prosthetic re- prostheses. Int J Prosthodont 1993;6:346-63.
taining screws are retightened to ideal levels, it is possi- 17. Riedy SJ, Lang BR, Lang BE. Fit of implant frameworks fabricated by
ble that the net effect may be favorable toward future different techniques. J Prosthet Dent 1997;78:596-604.
18. Lie A, Jemt T. Photogrammetric measurements of implant positions. De-
component loosening or breakage. However, if compo- scription of a technique to determine the fit between implants and super-
nents do become dislodged, the orientation of the worn structures. Clin Oral Implants Res 1994;5:30-6.
surfaces could be altered. The consequences of such a 19. Rubenstein JE, Ma T. Comparison of interface relationships between im-
change are currently being investigated in a subsequent plant components for laser-welded titanium frameworks and standard cast
frameworks. Int J Oral Maxillofac Implants 1999;14:491-5.
study. 20. Binon PP. Evaluation of machining accuracy and consistency of selected
implants, standard abutments, and laboratory analogs. Int J Prosthodont
1995;8:162-78.
CONCLUSIONS 21. Ma T, Nicholls JI, Rubenstein JE. Tolerance measurements of various
Within the limitations of this study, cyclic loading of implant components. Int J Oral Maxillofac Implants 1997;12:371-5.
22. Cox JF, Pharoah M. An alternative holder for radiographic evaluation of
implant-supported frames caused changes in the fit of tissue-integrated prostheses. J Prosthet Dent 1986;56:338-41.
the frame to the supporting implant abutments. The fit 23. Jemt T, Lekholm U. Measurements of bone and frame-work deformations
of the superstructure showed significant gap reductions induced by misfit of implant superstructures: a pilot study in rabbits. Clin
Oral Implants Res 1998;9:272-80.
when the prosthesis was loaded on the anterior segment.
24. Piattelli A, Piattelli M, Scarano A, Montesani L. Light and scanning elec-
Fit did not change significantly when the frameworks tron microscopic report of four fractured implants. Int J Oral Maxillofac
were loaded on the unilateral or bilateral posterior can- Implants 1998;13:561-4.
tilever portions of the prostheses. 25. Eckert SE, Wollan PC. Retrospective review of 1170 endosseous implants
placed in partially edentulous jaws. J Prosthet Dent 1998;79:415-21.
26. Taylor TD. Prosthodontic problems and limitations associated with os-
REFERENCES seointegration. J Prosthet Dent 1998;79:74-8.
27. Binon PP, McHugh MJ. The effect of eliminating implant/abutment rota-
1. Adell R, Lekholm U, Rockler B, Branemark PI. A 15-year study of os- tional misfit on screw joint stability. Int J Prosthod 1996;9:511-9.
seointegrated implants in the treatment of the edentulous jaw. Int J Oral 28. Niznick GA. Bending overload and implant fracture: a retrospective clin-
Surg 1981;10:387-416.
ical analysis. Int J Oral Maxillofac Implants 1996;11:431-2.
2. Adell R, Eriksson B, Lekholm U, Branemark PI, Jemt T. Long-term fol-
29. Rangert B, Krogh PH, Langer B, Van Roekel N. Bending overload and
low-up study of osseointegrated implants in the treatment of totally eden-
implant fracture: a retrospective clinical analysis. Int J Oral Maxillofac
tulous jaws. Int J Oral Maxillofac Implants 1990;5:347-59.
Implants 1995;10:326-34.
3. Cox JF, Zarb GA. The longitudinal clinical efficacy of osseointegrated
30. Kallus T, Bessing C. Loose gold screws frequently occur in full-arch fixed
dental implants: a 3-year report. Int J Oral Maxillofac Implants 1987;2:
prostheses supported by osseointegrated implants after 5 years. Int J Oral
91-100.
4. Laney WR, Tolman DE, Keller EE, Desjardins RP, Van Roekel NB, Brane- Maxillofac Implants 1994;9:169-78.
mark PI. Dental implants: tissue-integrated prosthesis utilizing the os- 31. Goodacre CJ, Kan JY, Rungcharassaeng K. Clinical complications of os-
seointegration concept. Mayo Clin Proc 1986;61:91-7. seointegrated implants. J Prosthet Dent 1999;81:537-52.
5. Branemark PI, Adell R, Breine U, Hansson BO, Lindstrom J, Ohlsson A. 32. Eckert SE, Meraw SJ, Cal E, Ow RK. Analysis of incidence and associated
Intra-osseous anchorage of dental prostheses. I. Experimental studies. factors with fractured implants: a retrospective study. Int J Oral Maxillofac
Scand J Plast Reconstr Surg 1969;3:81-100. Implants 2000;15:662-7.
6. Carr AB, Gerard DA, Larsen PE. The response of bone in primates around 33. Carr AB, Gerard DA, Larsen PE. The response of bone in primates around
unloaded dental implants supporting prostheses with different levels of fit. unloaded dental implants supporting prostheses with different levels of fit.
J Prosthet Dent 1996;76:500-9. J Prosthet Dent 1996;76:500-9.
7. Clelland NL, Carr AB, Gilat A. Comparison of strains transferred to a bone 34. Jemt T, Lekholm U, Johansson CB. Bone response to implant-supported
simulant between as-cast and postsoldered implant frameworks for a frameworks with differing degrees of misfit preload: in vivo study in
five-implant-supported fixed prosthesis. J Prosthodont 1996;5:193-200. rabbits. Clin Implant Dent Relat Res 2000;2:129-37.

350 VOLUME 89 NUMBER 4


HECKER AND ECKERT THE JOURNAL OF PROSTHETIC DENTISTRY

35. Clelland NL, Papazoglou E, Carr AB, Gilat A. Comparison of strains Reprint requests to:
transferred to a bone simulant among implant overdenture bars with DR DONNA M HECKER
various levels of misfit. J Prosthodont 1995;4:243-50. UNIVERSITY OF MINNESOTA
36. Michaels GC, Carr AB, Larsen PE. Effect of prosthetic superstructure ROOM 6-284 MOOS TOWER
accuracy on the osteointegrated implant bone interface. Oral Surg Oral 515 DELAWARE ST SE
Med Oral Pathol Oral Radiol Endod 1997;83:198-205. MINNEAPOLIS, MN 55455
37. Stewart RB, Desjardins RP, Laney WR, Chao EY. Fatigue strength of FAX: 612-624-2660
cantilevered metal frameworks for tissue-integrated prostheses. J Prosthet E-MAIL: hecke003@tc.umn.edu
Dent 1992;68:83-92.
38. Stewart RB, Staab GH. Cross-sectional design and fatigue durability of Copyright © 2003 by The Editorial Council of The Journal of Prosthetic
cantilevered sections of fixed implant-supported prostheses. J Prosthodont Dentistry.
1995;4:188-94. 0022-3913/2003/$30.00 ⫹ 0
39. Staab GH, Stewart RB. Theoretical assessment of cross sections for canti-
levered implant-supported prostheses. J Prosthodont 1994;3:23-30. doi:10.1067/mpr.2003.71

APRIL 2003 351

Das könnte Ihnen auch gefallen