GUIDANCE NOTE
THE MICROBIOLOGICAL
EXAMINATION OF WATER
SAMPLES
QSOP 57
National Standard Methods are developed, reviewed and updated through an open and wide
consultation process where the views of all participants are considered and the resulting documents
reflect the majority agreement of contributors.
Representatives of several professional organisations, including those whose logos appear on the
front cover, are members of the working groups which develop National Standard Methods. Inclusion
of an organisation’s logo on the front cover implies support for the objectives and process of preparing
standard methods. The representatives participate in the development of the National Standard
Methods but their views are not necessarily those of the entire organisation of which they are a
member. The current list of participating organisations can be obtained by emailing
standards@hpa.org.uk.
The performance of standard methods depends on the quality of reagents, equipment, commercial
and in-house test procedures. Laboratories should ensure that these have been validated and shown
to be fit for purpose. Internal and external quality assurance procedures should also be in place.
Whereas every care has been taken in the preparation of this publication, the Health Protection
Agency or any supporting organisation cannot be responsible for the accuracy of any statement or
representation made or the consequences arising from the use of or alteration to any information
contained in it. These procedures are intended solely as a general resource for practising
professionals in the field, operating in the UK, and specialist advice should be obtained where
necessary. If you make any changes to this publication, it must be made clear where changes have
been made to the original document. The Health Protection Agency (HPA) should at all times be
acknowledged.
The HPA is an independent organisation dedicated to protecting people’s health. It brings together the
expertise formerly in a number of official organisations. More information about the HPA can be found
at www.hpa.org.uk.
The HPA aims to be a fully Caldicott compliant organisation. It seeks to take every possible precaution
to prevent unauthorised disclosure of patient details and to ensure that patient-related records are
kept under secure conditions1.
Please note the references are now formatted using Reference Manager software. If you alter or delete text
without Reference Manager installed on your computer, the references will not be updated automatically.
Health Protection Agency (2005). The microbiological examination of water samples. National
Standard Method QSOP 57 Issue 2. http://www.hpa-standardmethods.org.uk/pdf_sops.asp.
Each National Standard Method has an individual record of amendments. The current amendments
are listed on this page. The amendment history is available from standards@hpa.org.uk.
On issue of revised or new pages each controlled document should be updated by the copyholder in
the laboratory.
Amendment Issue no. Insert Page Section(s) involved Amendment
Number/ Discarded Issue
Date no.
1/ 1 2 7 Section 2.7 Aircraft Text amended. Reference to
06/12/05 waters clostridium species removed
1 INTRODUCTION
The aim of this document is to provide guidance for the Microbiological testing of water
samples across the Health Protection Agency (HPA). The document covers the testing of
water samples for human consumption, pool waters and cooling towers.
The results of a laboratory examination of a single water sample are representative only of
the water at that particular point at which the sample is taken. Contamination is often
intermittent and may not be revealed by the examination of a single sample. The value of
microbiological tests is dependent upon their frequent and regular use. Information gained
over time through monitoring will provide a comprehensive picture of the overall quality of
any particular source of water, any deterioration from which should at once arouse
suspicion.
Sampling techniques and sample transportation can influence sample results and good
practice is essential. Samples collected from treated waters should have any residual
biocide neutralised, wherever possible, at the time of sampling. For oxidising biocides such
as chlorine and bromine this is achieved by using sample bottles containing sodium
thiosulphate as stated in the MDW 20022 Part 2 section 3.1.
3 POOL WATERS
3.1.1 General
Microbiological contamination of the pool water can result in pathogenic (disease producing)
micro-organisms causing infections to bathers. These contaminants can be introduced into
Microbiological samples for indicator organisms should be taken at least once a month as a
routine and quarterly for Legionella species. More frequent sampling may be required
depending on the Risk Assessment, eg if the spa pool is being intensively used and
certainly if there are any adverse health effects reported by the bathers. If adverse health
effects are suspected the enforcing authority (HSE or the Local Authority) and the
microbiologist in the testing laboratory should be informed; as required they will then notify
the CCDC within the Local Health Protection Unit. Microbiological sampling should also be
done
If there are health problems associated with the use of the spa pool, it may be necessary to
test for other organisms, based on epidemiological evidence, such as Staphylococcus
aureus, Cryptosporidium, Giardia and perhaps viruses. In these circumstances advice
should first be sought from the local Health Protection Unit and the microbiologist.
Routine sampling should be done when the spa pool is in use, preferably when heavily
loaded or immediately thereafter.
If results are still unsatisfactory after the repeat samples and investigation, immediate
remedial action is required that may necessitate the spa pool being closed. Note: the
investigation may require the help of the laboratory that does the tests, the local council
Environmental Health Department, or an independent consultant.
3.2.7 Legionella
Well-operated spa pools should not normally contain Legionella species. The
microbiological results should not be considered in isolation but in the context of the
management records for the spa pool.
Where there is evidence of gross contamination the spa pool should be closed to prevent
illness in users and those working near the spa pool. The appropriate enforcing authority
should be contacted.
2. greater than 50 Ps. aeruginosa per 100mL in combination with a high aerobic
colony count (>100 per mL)
The above guidelines for interpretation of Legionella results have been agreed by a number
of environmental health departments but note that some departments have a zero tolerance
approach on the basis that negative Legionella results are achievable.
4 COOLING TOWERS
Cooling towers and evaporative condensers should be routinely monitored to ensure that
control of Legionella growth is in place, according to the Health and Safety Commission
2000 Approved Code of Practice and Guidance L8. To achieve compliance with L8, it is
recommended that systems should be tested on a weekly basis for ACC at 30°C for 49 ±
1hrs (minimum) and quarterly testing for the presence of legionellae (or monthly when
commissioning and establishing a treatment programme for new plant). The sampling
method should be in accordance with Standing Committee of Analysts (2005)14.
THE MICROBIOLOGICAL EXAMINATION OF WATER SAMPLES
Issue no: 2 Issue date: 06.12.05 Issued by: Standards Unit, Evaluations & Standards Laboratory Page 12 of 26
Reference no: QSOP 57i2
This SOP should be used in conjunction with the series of SOPs from the Health Protection Agency
www.evaluations-standards.org.uk
Email standards@hpa.org.uk
An ACC of not more than 10,000 bacteria per mL and less than 100 Legionella per litre
indicate that the system is under control. For levels of 10,000 – 100,000 ACC per mL, or
100 – 1,000 Legionella per litre, a review of control measures and a risk assessment should
be undertaken. Immediate corrective action including disinfection and re-sample should be
implemented if levels exceed 1,000,000 ACC per mL or 1,000 Legionella per litre.
Standards Unit
Evaluations and Standards Laboratory
Centre for Infections
Health Protection Agency
Colindale, London
NW9 5EQ
e-mail standards@hpa.org.uk
MAINS WATER
Colony count 1mL No abnormal For one off samples of little value except W4
22°C for 68 ± change post disinfection e.g. new water system in a
4hrs hospital or to investigate complaints about
taste / odour.
Colony count 1mL No abnormal For one off samples of little value except W4
37°C for 44 ± change post disinfection e.g. new water system in a
4hrs hospital or to investigate complaints about
taste / odour.
Enterococci 100mL 0 >0 Not required by PWS Regs 1991 W3 Will be required for all audit samples when PWS
Regs are revised
THE MICROBIOLOGICAL EXAMINATION OF WATER SAMPLES
Issue no: 2 Issue date: 06.12.05 Issued by: Standards Unit, Evaluations & Standards Laboratory Page 16 of 26
Reference no: QSOP 57i2
This SOP should be used in conjunction with the series of SOPs from the Health Protection Agency
www.evaluations-standards.org.uk
Email standards@hpa.org.uk
BOTTLED WATER / SPRING WATER
Volume Acceptable Action Comments HPA UK Legislation / Guidance
Test to be level level Method
tested
Bottled water Based on Council Directive 98/83/EC W3 The Natural Mineral Water, Spring Water and
/ spring water Bottled Drinking Water Regulations 1999
(Statutory Instrument 1540) as amended by the
Natural Mineral Water, Spring Water and Bottled
Water (Amendment) (England) Regulations 2003
(SI 2003 No. 666 )and the Natural Mineral Water,
Spring Water and Bottled Drinking Water
(amendment) (England) Regulations 2004
(Statutory Instrument 2004 No. 656).
Legionella 1000mL <100 >100 If >1000 close pool W12 / Currently not required but recommend quarterly in
W13 / the new HSE/HPA guidance Management of Spa
W14 Pools:- Controlling The Risks of Infection currently
in press)
Legionella 1000mL <100/L >100 If >1000/L immediately resample and then W12 /
disinfect W13 /
W14
Colony count 1mL < 10,000 >10,00 If > 100,000.mL immediately resample and W4
30°C for 49 ± 0 then disinfect
1hrs
Mineral Water *The microbiological requirements have not changed since the original directive;
tests in unshaded boxes are required on all samples;
tests in shaded boxes are only required for some samples
temperature tolerances are those in Microbiology of Drinking water 2002;
requested by the CCDC; or to support adverse results of inspections by EHO’s subject to local agreement with the laboratory.
1. Department of Health NHS Executive: The Caldicott Committee. Report on the review of patient-
identifiable information. London. December 1997.
2. Standing Committee of Analysts. The Microbiology of Drinking Water 2002. Methods for the
Examination of waters and Associated Materials. Environment Agency.
http://www.environmentagency.gov.uk/science/219094/401849/?lang=_e®ion=&projectstatus=&th
eme=&subject=&searchfor=SCA&topic=&area=&month.
3. Water Supply (Water Quality) Regulations 2000 (SI 2000 No. 3184).
http://www.dwi.gov.uk/regs/si3184/3184.htm.
4. The Private Water Supplies Regulations (Statutory Instrument No. 2790). London: HMSO; 1991.
5. Council Directive 98/83/EC: Relating to the Quality of Water Intended for Human Consumption.
Official Journal of the European Communities; 1998. p. L330-32-L330/53
6. The Natural Mineral Water, Spring Water and Bottled Drinking Water Regulations (Statutory
Instrument No. 666). London: HMSO; 2003.
7. Council Directive 80/777/EC: On the approximation of the laws of the Member States relating to the
exploitation and marketing of natural mineral waters. Official Journal of the European Communities;
1998. p. L229 1-L229 10
8. The Natural Mineral Water, Spring Water and Bottled Drinking Water (amendment) (England)
Regulations 2004 (Statutory Instrument No. 656). 2004.
9. The Natural Mineral Water, Spring Water and Bottled Drinking Water Regulations 1999 (Statutory
Instrument 1540). 1999.
10. Health Protection Agency. Guidelines for water quality on board merchant ships including passenger
vessels. London: Health Protection Agency; 2003.
11. PHLS The Microbiological Quality of Water Onboard Aircraft. London: PHLS; 2003.
12. British Standard BS PAS 39:2003. Management of public swimming pools. Water treatment
systems, water treatment plant and heating and ventilation systems. Code of Practice; 2003.
http://www.pwtag.org/home.html.
13. PHLS. Hygiene for Hydrotherapy Pools. 2nd ed. London: PHLS; 1999.
14. Part I - Rationale of surveying and sampling and ISO 11731:1998. http://www.environment-
agency.gov.uk/commondata/acrobat/book_200_1028650.pdf.