Sie sind auf Seite 1von 3

1050 LCGC VOLUME 19 NUMBER 10 OCTOBER 2001 www.chromatographyonline.

com

Attacking Carryover
Problems
LC
Troubleshooting C arryover is recognized as the pres-
ence of a small peak that appears
when a blank is injected after a
sample that produces a large peak. Carry-
over can be one of the most frustrating
tion, so a trace of sample remaining from
the previous injection will go unnoticed.
The other common type of carryover
occurs when a small peak appears at a
nearly constant size in all blanks, for
problems in liquid chromatography (LC) example at 0.5% of a high-level sample.
practice. Some methods seem to be This constant carryover probably is a result
immune, whereas other methods fight a of contamination and not true carryover,
constant battle with carryover. This month’s but the general approach to correcting the
“LC Troubleshooting” column focuses on a problem is the same as for classic carry-
systematic approach to solving carryover over, so I will include it in the current
A systematic approach is problems. Readers can use the accompany- discussion.
the best way to solve ing sidebar, “Steps to Correct Carryover
carryover problems. Problems,” as a quick reference guide while Constant Carryover
reading this column and when solving If the consecutive injection of blanks results
future problems. in a small peak of similar size in each run,
the problem is classified as constant carry-
Classify the Problem over. Although most of the troubleshooting
The first step in solving a problem is to process is the same as for classic carryover,
determine what kind of carryover you’re you need to conduct a couple more experi-
experiencing. You can make this determina- ments first.
tion by injecting a high-level sample fol- Because contamination is the most likely
lowed by two or more blank injections of cause of constant carryover, you first should
sample matrix without the analyte (step 1). eliminate the most likely cause of contami-
What I call classic carryover can be recog- nation — the sample blank (step 2). First,
nized by a regular reduction of peak size as replace the blank with a nominally identical
blanks are injected consecutively. For exam-
ple, the first blank might have a peak that
is 1% of the size of the original, and then Steps to Correct
the next drops by another factor of 100. Carryover Problems
Although the washout doesn’t always result
in exactly the same percentage reduction 1. Classify carryover
with each consecutive blank, the peak size 2. Replace blank
will drop significantly with each additional 3. Change injection size
injection, and it should be insignificant by 4. Check fitting assembly
the third or fourth injection. 5. Check wash solvent
The source of classic carryover often is a a. Use fresh wash solvent
small cavity in a system that acts as a reser- b. Increase wash volume
voir of sample that becomes diluted with c. Use more organic solvent in wash
successive injections. Classic carryover is d. Adjust wash pH
especially problematic when large and small 6. Check washing mechanism
analyte values are likely to be found in the 7. Change injection solvent
same sample set such as when an LC 8. Check another sample
method is used to determine the pharmaco- 9. Change hardware
kinetic behavior of a drug in plasma. For a. Change needle seal
applications such as content uniformity or b. Change injection loop
product assay, classic carryover may be of c. Rebuild or replace valve
John W. Dolan little importance because all samples have d. Substitute autosamplers
LC Troubleshooting Editor approximately the same analyte concentra-
1052 LCGC VOLUME 19 NUMBER 10 OCTOBER 2001 www.chromatographyonline.com

blank that you feel is free of contamination. tubing is to shut off the pump, loosen each for removing contaminants than methanol
If the blank is water or a simple solvent fitting, push the tubing firmly to the bot- or acetonitrile.
mixture, just make a fresh batch, preferably tom of the tube port, and retighten the fit- Most mobile phases contain a buffer,
from another source of reagents, and rinse ting. Don’t try to tighten PEEK fittings acid, or base to control the pH. For wash
the injection vial before use to ensure its with pressure in the system, because the solvents, I generally like to avoid using
cleanliness. If the sample is the result of tubing can slip in the fitting when the nut buffers or salts that leave a residue when
extraction of the analyte from a biological is rotated. Even if you didn’t notice obvious evaporated. In my experience, these addi-
or environmental matrix, you may find it problems, it is a good idea to repeat the tives leave white crusty deposits around the
more difficult to obtain a blank unrelated carryover test (high sample followed by two needle wash port and injection port. Thus,
to the suspect sample. If all else fails, blanks) to see if the problem has been phosphate buffer is banned for use in wash
replace the blank with another solution that resolved. solvents in my laboratory. However, creat-
is compatible with the chromatographic ing acidic or basic conditions is all that is
system, such as a few milliliters of mobile Wash Solvent Problems necessary for use as a wash solvent, so you
phase. If the carryover peak disappears with The next several steps are related to the can use a volatile acid or base. Workers in
a fresh blank, the solution to your problem injector wash solvent. You may want to fol- my laboratory often use formic, acetic, tri-
may be as simple as preparing fresh blanks low these steps in order or if one seems fluoroacetic, or heptafluorobutyric acid or
to use with the analysis. On the other more appropriate for your sample, go ammonium hydroxide to create acidic or
hand, you may need to spend considerable directly to it. First, replace the wash solvent basic wash solvent conditions. These addi-
effort tracking down the source of contami- with a fresh batch (step 5a). When trouble- tives are freely soluble in all concentrations
nation, which may be a reagent or glass- shooting carryover problems, I like to err of water, acetonitrile, and methanol. Typi-
ware. on the side of caution, so I generally replace cally a 0.1–1% solution by volume is suffi-
As a cautionary measure, if your labora- the wash solvent with a freshly prepared cient. Try adding acid or base to the wash
tory is involved in both trace and prepara- batch and replace the reservoir with a clean solvent (step 5d) to see if the carryover is
tive chromatography, you should segregate one. The autosamplers my co-workers and improved. If you plan to use a basic wash
the glassware. I have seen contamination I use in our laboratory work best if the solvent routinely, be sure that system com-
problems arise when traces of analyte wash solvent is degassed, so we helium- ponents such as the injection valve seals are
remain on the glassware after use of that sparge the solvents before putting them on compatible.
glassware for synthesis or preparative the autosampler. Degassing, however, is a
chromatography. Washing highly contami- potential source of contamination, so you Washing Mechanism
nated glassware in the same wash cycle as may want to try using a fresh solvent with If your efforts at correcting the problem by
analytical glassware may produce cross- and without degassing to see if you notice changing the wash solvent (step 5) were
contamination. any difference. After the wash solvent is unsuccessful, it is time you look into the
If replacing the blank did not solve the replaced, cycle the autosampler through the mechanical aspects of the autosampler wash
problem, try adjusting the injection volume purge or wash cycle several times to ensure (step 6). Each autosampler design uses a
by a factor of two or more (step 3). If the that all traces of the previous solvent have slightly different wash mechanism, so you
carryover peak increases or decreases in pro- been removed. will find it helpful to consult your opera-
portion to the injection volume change, it If changing the wash solvent solves the tor’s manual for specific help. Many auto-
is highly probable that the blank is con- problem only temporarily, perhaps you are samplers use an overflow cup assembly to
taminated, so go back to step 2 and spend using an insufficient wash volume between wash the outside of the autosampler needle.
more time trying to identify the source. If samples. Increase the wash volume (step For example, the autosampler will insert the
the peak remains at constant size, continue 5b) or the number of wash cycles to see if needle into a small wash station, which
with the steps described below. either of these fixes will correct the prob- may look like a microvial. The autosampler
lem. pushes wash solvent through the needle,
Fittings Problems If the problem persists, the solvent prop- and the solvent flows up the sides of the
If steps 2 or 3 did not identify the source of erties of the wash solvent might be insuffi- needle from the needle tip and overflows to
the constant carryover problem or if the cient to remove all of the analyte from the waste.
problem is classic carryover, move to step 4. previous injection. To my knowledge, none This process usually is effective at rinsing
You should check all the compression fit- of the currently available autosamplers any contamination from the outside of the
tings and tube connections that the sample inject wash solvent, so as long as the wash needle. In some cases, however, the wash
contacts for proper assembly. This proce- solvent is miscible with the mobile phase, station can become a source of contamina-
dure generally is limited to the fittings on it doesn’t need to be of similar solvent tion. If nonsoluble sample components
the injection valve and all fittings down- strength. My next step usually is to increase collect in the cup or on the outside of the
stream to the detector. Stainless steel fit- the strength of the wash solvent (step 5c). needle, each cycle through the wash station
tings rarely are a problem, but the popular Often I’ll use a wash solvent that is roughly contaminates the needle. Some auto-
polyetheretherketone (PEEK) fittings and equivalent in strength to the mobile phase, samplers use a septum or other wiping
tubing can slip under high pressures; for so the first step when using a stronger sol- mechanism to physically wipe the outside
example, greater than approximately 4000 vent would be to use more organic solvent of the needle as it enters and leaves the
psi or 275 bar. or even 100% of the strong solvent (usually wash station. This septum can become
You can check stainless steel fittings by acetonitrile or methanol). Isopropanol worn out, damaged, or contaminated. If
slightly tightening them with a wrench. makes an excellent wash solvent for many the wash station drain line is positioned
The best approach for PEEK fittings and applications and proves to be more effective improperly, contaminated wash solvent can
1054 LCGC VOLUME 19 NUMBER 10 OCTOBER 2001 www.chromatographyonline.com

siphon into the wash station rather than to plumbing. If the carryover is specific to the repair options, the only option left is to
waste when the needle is removed. Disas- one compound, you should search for a use another autosampler (step 9d). If you
semble the wash station, clean it, and chemical solution to the problem, such as take this route, be sure you are not just
check for any of the problems noted above changing the injection or wash solvents. making a substitution that delays the
or additional mechanical problems that are recurrence of carryover problems. For
specific to your brand and model of Hardware Changes example, my laboratory has two models of
autosampler. If you’re still trying to solve the carryover the same brand of autosampler; one model
problem, it is highly likely that the prob- is much less susceptible to carryover than
Injection Solvent lem is related to the autosampler hardware. the other. So if we just replace a problem
The injection solvent serves as a carrier to At this point, you should substitute parts unit with another clean unit of the same
facilitate transfer of the sample onto the until you locate the problem. Often the model, the carryover problem likely will
column. In general, and especially if small easiest first step is to try a manual injection recur before long.
injection volumes (less than 20 mL) are valve or to replace the autosampler with
used, the choice of injection solvent is one that has not been exposed to the com- Conclusion
unimportant. As long as the injection sol- pound of interest (step 9d). If either of Carryover problems can be some of the
vent is no stronger than the mobile phase these approaches correct the problem, most vexing quandries in LC. It is tempt-
and the injection volume is no larger than it confirms that the problem is in the ing to try a quick fix such as changing the
approximately 15% of the resulting peak autosampler and further troubleshooting wash solvent and seals and thoroughly
volume, the injection solvent composition efforts along this line are worthwhile. cleaning the system. Sometimes this route
will have little effect on the chromatogra- If you haven’t done it already, replace the will work, but it rarely identifies the cause
phy. For this reason, chromatographers needle seal on the injection valve (step 9a). of the problem. Thus, when the problem
should choose their injection solvent for Many autosamplers use a polymeric seal to shows up again, you have little experience
ease of sample preparation. For example, if prevent sample loss during sample loop fill- to help you solve it more quickly the next
the sample is highly water soluble, the ing. These seals can become worn and time. In my experience, carryover problems
injection solvent could be water, even require occasional adjustment or replace- of one sample type often show up as carry-
though the mobile phase contains 50% ment. Autosampler designs in which the over problems for another sample. By tak-
organic solvent. With some samples, the needle and loop are combined may use a ing the time to systematically find the
selected injection solvent may facilitate graphite or hard polymer seal to provide cause of the problem, you will be able to
adsorption of the sample on system com- high-pressure sealing. If this seal is worn or locate and correct the problem much more
ponents such as the autosampler loop. This damaged by improper needle alignment, quickly in the future.
practice most commonly becomes prob- you may observe increased carryover even
lematic when the injection solvent is 100% if the fitting doesn’t leak. When the needle
water or buffer. Often, adding a small seal is replaced, carefully examine the tip John W. Dolan
“LC Troubleshoot-
amount of organic solvent such as 5% and outside of the injection needle. If any ing” editor John W.
methanol or acetonitrile to the injection roughness or corrosion is observed, replace Dolan is president of
solvent will eliminate any sample adsorp- the needle to avoid rapid wear of the new LC Resources Inc. of
tion. Check for problems related to the seal. Walnut Creek, Cali-
injection solvent by changing to another Sometimes the sample adsorbs on the fornia, and a mem-
ber of LCGC’s editor-
solvent or by adding more organic solvent sample loop because of the characteristics ial advisory board.
to the injection solvent (step 7). of sample and injection solvent. Replacing Direct correspon-
the loop (step 9b) with one of different dence about this col-
Sample-Specific Carryover composition can help solve this problem. umn to “LC Troubleshooting,” LCGC, 859
Injection loops are available in stainless Willamette Street, Eugene, OR 97401, e-mail
By now you’ve tried all the easy fixes to
John.Dolan@LCResources.com.
carryover problems, so it is worth the trou- steel, PEEK, and titanium. A combination For an ongoing discussion of LC trouble-
ble to see if the problem is specific to your of a new loop material and a different shooting with John Dolan and other chro-
sample (step 8). Sometimes you can check injection solvent (step 7) should eliminate matographers, visit the Chromatography
by examining another peak in the sample. sample adsorption on the loop. Forum discussion group at http://www.
chromforum.com.
For example, if your method uses an inter- The internal components of the injec-
nal standard, try injecting a large amount tion valve will become worn over time;
of internal standard and then injecting repairing or replacing the injection valve
blanks without an internal standard. Per- (step 9c) will be necessary from time to
haps you can inject another chemically time. A worn injection seal can distort the
related compound that will be eluted flow of fluids through the valve and may
under the current method conditions. If all create conditions that facilitate carryover.
else fails, you can change the mobile phase In other cases, sample can adsorb on the
or column and run another method. polymeric seals inside the valve. Seal mate-
If you find that the carryover problem rials of different composition help address
is common to different compounds or this problem.
method conditions, most likely the carry- Other components specific to your
over is a physical problem associated with brand and model of autosampler can be
the autosampler hardware or system cleaned or replaced. Once you’ve exhausted

Das könnte Ihnen auch gefallen