Airway Management
AIRWAY MANAGEMENT
OBJECTIVES:
EQUIPMENT:
PROCEDURE:
2. Assess patient’s need for suctioning. Since endotracheal suctioning can be hazardous and
causes discomfort, it is not recommended in the absence of apparent need.
4. Turn on suction apparatus and set vacuum regulator to appropriate negative pressure.
Recommend 80-120 mmHg; adjust lower for children and the elderly. Significant hypoxia and
damage to tracheal mucosa can result from excessive negative pressure.
5. Prepares suction apparatus. Secure one end of connecting tube to suction machine, and place
other end in a convenient location within reach.
6. Use in-line suction catheter or open sterile package (catheter size not exceeding one-half the
inner diameter of the airway) on a clean surface, using the inside of the wrapping as a sterile
field.
8. Prepare catheter flush solution.With in-line catheter use sterile saline bullets to flush catheter.
With regular suctioning set up sterile solution container and being careful not to touch the inside
of the container, fill with enough sterile saline or water to flush catheter.
9. With in-line suction catheter use clean gloves. With regular suctioning, done sterile gloves.
Maintain sterility. Universal precautions. In regular suctioning the dominant hand must remain
sterile throughout the procedure.
10. Pick up suction catheter, being careful to avoid touching nonsterile surfaces. With
nondominant hand, pick up connecting tubing. Secure suction catheter to connecting tubing.
Maintains catheter sterility. Connects suction catheter and connecting tubing
11. Ensures equipment function. Check equipment for proper functioning by suctioning a small
amount of sterile saline from the container. (skip this step in in-line suctioning)
12. Remove or open oxygen or humidity device to the patient with nondominant hand. (skip this
step with in-line suctioning). Opens artificial airway for catheter entrance. Have second person
assist when indicated to avoid unintentional extubation.
13. Replace O2 delivery device or reconnect patient to the ventilator. Hyperoxygenate and
hyperventilate via 3 breaths by giving patient additional manual breaths on the ventilator before
suctioning. Hyperoxygenation with 100% O2 is used to offset hypoxemia during interrupted
oxygenation and ventilation. Preoxygenation offsets volume and O2 loss with suctioning.
Patients with PEEP should be suctioned through an adapter on the closed suction system.
14. Without applying suction, gently but quickly insert catheter with dominant hand during
inspiration until resistance is met; then pull back 1-2 cm. Catheter is now in tracheobronchial
tree. Application of suction pressure upon insertion increases hypoxia and results in damage to
the tracheal mucosa.
15. Apply intermittent suction by placing and releasing dominant thumb over the control vent of
the catheter. Rotate the catheter between the dominant thumb and forefinger as you slowly
withdraw the catheter. With in-line suction, apply continuous suction by depressing suction valve
and pull catheter straight back. Time should not exceed 10-15 seconds. Intermittent suction and
catheter rotation prevent tracheal mucosa when using regular suctioning methods. Unable to
rotate with closed- suction method.
16. Replace oxygen delivery device. Hyperoxygenate between passes of catheter and following
suctioning procedure. Replenishes O2. Recovery to base PaO2 takes 1 to 5 minutes. Reduces
incidence of hypoxemia and atelectasis.
17. Rinse catheter and connecting tubing with normal saline until clear. Removes catheter
secretions.
18. Monitor patient’s cardiopulmonary status during and between suction passes. Observe for
signs of hypoxemia, e.g. dysrhythmias, cyanosis, anxiety, bronchospasms, and changes in mental
status.
19. Once the lower airway has been adequately cleared of secretions, perform nasal and oral
pharyngeal or upper airway suctioning. Removes upper airway secretions. The catheter is
contaminated after nasal and oral pharyngeal suctioning and should not be reinserted into the
endotracheal or tracheostomy tube.
20. Upon completion of upper airway suctioning, wrap catheter around dominant hand. Pull
glove off inside out. Catheter will remain in glove. Pull off other glove in same fashion and
discard. Turn off suction device. Reduces transmission of microorganisms.
21. Reposition patient. Supports ventilatory effort; promotes comfort; communicates caring
attitude.
23. Dispose of suction liners and connecting tubing, sterile saline solution every 24 hours and set
up new system. Decreases incidence of organism colonization and subsequent pulmonary
contamination. Universal precautions.
PRECAUTIONS:
a. Avoid using catheters larger than one-half the diameter of the airway.
b. Administer one or more postsuctioning hyperinflations, using manual or sigh breaths on the
ventilator or ambu bag if not ventilated.
2. Maintain rigorous sterile technique when suctioning the intubated patient. Impaired pulmonary
defense systems and invasive instrumentation of the pulmonary tract predisposes these patients
to colonization and infection. Never use same catheter to suction the trachea after it has been
used in the nose or the mouth.
3. Limit the frequency of suctioning and avoid, as much as possible, catheter impaction in the
bronchial tree when the patient is anticoagulated or when hemorrhage from suction-induced
trauma is evident.
4. Minimize the frequency and duration of suctioning when patient is on positive end-expiratory
pressure (PEEP) greater than 5 cm or continuous positive airway pressure (CPAP). Small
suctioning-induced changes may have profound effects on these marginally oxygenated patients.
6. The use of saline installations for loosening secretions has been controversial and recent
research shows that in fact it is detrimental and poses a greater risk of pneumonia for the patient.
RELATED CARE:
1. Include strategies to move secretions through peripheral airways. These measures are:
appropriate hydration and adequate humidification of inspired gases (to keep secretions thin);
coughing and deep breathing; frequent position changes (may need rotation bed); chest
physiotherapy; and bronchodilating agents as ordered.
2. Monitor the patient carefully during ET/tracheal suctioning for ectopic dysrhythmias
aggravated by suction-induced hypoxemia and other dysrhythmias, particularly conduction
disturbances, related to catheter irritation of vagal receptors within the respiratory tract (requires
immediate cessation of suctioning and hyperoxygenation).
POTENTIAL COMPLICATIONS
Hypoxemia
Atelectasis
Dysrhythmias
Sepsis
Cardiac arrest