Beruflich Dokumente
Kultur Dokumente
2. Identification of substances
IR spectroscopy is used to establish whether a given sample of an organic substance is
identical with another or not. This is because large number of absorption bands is observed
in the IR spectra of organic molecules and the probability that any two compounds will
produce identical spectra is almost zero. So if two compounds have identical IR spectra then
both of them must be samples of the same substances.
IR spectra of two enatiomeric compound are identical. So IR spectroscopy fails to
distinguish between enantiomers.
For example, an IR spectrum of benzaldehyde is observed as follows.
C-H stretching of aromatic ring- 3080 cm-1
C-H stretching of aldehyde- 2860 cm-1 and 2775 cm-1
C=O stretching of an aromatic aldehyde- 1700 cm-1
C=C stretching of an aromatic ring- 1595 cm-1
C-H bending- 745 cm-1 and 685 cm-1
No other compound then benzaldehyde produces same IR spectra as shown above.
It is possible to obtain an IR spectrum from samples in many different forms, such as liquid,
solid, and
gas. However, many materials are opaque to IR radiation and must be dissolved or diluted in a
transparent matrix in order to obtain spectra. Alternatively, it is possible to obtain reflectance or
emission
spectra directly from opaque samples. Some popular sampling techniques and accessories are
discussed
here.
Liquid cells are used for dilute solutions of solid and liquid samples that are dissolved in
relatively
IR-transparent solvents. Sampling in solution results in enhanced reproducibility and is often the
preferred choice. Unfortunately, no single solvent is transparent through the entire mid IR region.
The analyst usually chooses solvents that have transparent windows in the region of interest. The
conventional
popular solvents are carbon tetrachloride for the region between 4000 and 1330 cm
–1
and carbon disulfide for the region between 1330 and 625 cm
–1
Both solvents are quite toxic, and thus must be handled .
carefully. One may replace carbon tetrachloride with the less-toxic tetrachloroethylene or
methylene
chloride and substitute carbon disulfide with n-hexane or n-heptane. Polar solvents such as water
and
alcohols are seldom used because they absorb strongly in the mid IR range and react with alkali-
metal
halides, such as NaCl, commonly used for cell windows.
Acquiring acceptable IR spectra of aqueous samples requires use of special types of liquid cells
such as thin cells of BaF2
, AgCl, or KRS-5(a mixed thallium bromide–thallium iodide). Aqueous solution measurements
can also be accomplished with attenuated total reflectance (ATR) accessories,
which are discussed later in this chapter.
Typically, solutions of 0.05 to 10% in concentration are handled in IR cells of 0.1 to 1 mm in
thickness. Concentration of 10% and cell path length of 0.1 mm represent one practical
combination. In a
double-beam spectrometer, a compensating cell is filled with pure solvent and placed in the
reference
beam. In the single-beam FT instrument, the solvent bands are mostly removed by obtaining the
difference spectra through subtraction of solvent spectra from sample spectra. Both fixed-
thickness and variable-thickness liquid cells are available commercially. They normally consist
of metal frame plates, IRtransmitting windows, and gaskets that determine the path length of the
cells.
Salt plates of IR-transmitting materials can be used for semivolatile and nonvolatile liquid
samples. Sodium chloride disks are the most popular and economical choice for nonaqueous
liquids. Silver chloride or barium fluoride plates may be used for samples that dissolve or react
with NaCl plates.
A drop of the neat sample is squeezed between two salt plates to form a film of approximately
0.01
mm in thickness. The plates can be held together by capillary attraction, or they may be clamped
in a
screw-tightened holder or pressed to form a good contact in a press fit O-ring supported holder.
It is
also possible to place a film of samples on salt plates by melting a relatively low-melting solid
and
squeezing it between two plates. Sodium chloride salt plates can usually be cleaned with dry
methylene chloride or acetone. This smear technique is one of the simplest ways to obtain IR
spectra.
Thin films of nonvolatile liquids or solids can be deposited on an IR-transmitting salt plate by
solvent evaporation. The sample is first dissolved in a reasonably volatile solvent. A few drops
of the resulting solution are placed on the plate. After evaporating off the solvent, a thin film of
sample is
obtained for subsequent spectra acquisition.
Water absorbs strongly at 3710cm-1 due to O-H stretch and 1630 cm-1 due to O-H bending.
These absorptions may overlap with the absorption bands due to the sample under examination.
Hence water cannot be used as solvent in IR spectra and also the sample has to dry.
Beer–Lambert law,
the law states that there is a logarithmic dependence between the transmission (or transmissivity), T, of
light through a substance and the product of the absorption coefficient of the substance, α, and the
distance the light travels through the material (i.e. the path length), ℓ. Theabsorption coefficient can, in
turn, be written as a product of either a molar absorptivity (extinction coefficient) of the absorber, ε, and
theconcentration c of absorbing species in the material, or an absorption cross section, σ, and the
(number) density N' of absorbers.
Transmittance, T = P / P0
% Transmittance, %T = 100 T
Absorbance,
A = log10 P0 / P
A = log10 1 / T
A = log10 100 / %T
A = 2 - log10 %T
the last equation, A = 2 - log %T , is worth remembering because it allows you
10
A=ε bc
%T = 100 P/P0 = e -ε bc
where:
A=Absorbance
c= Concentration
e= extinction coefficient
l = optical path length (typically 1 cm)
The most common of these are (a) and (b), which both result in a concave-down curvature of the
Beer's Law plot; (c) and (d) are easily avoided by proper experiment and instrument design
(square cuvettes, well-mixed solutions); (e) is only a problem at very high concentrations; (f) is
pretty common in real-world applications to complicated samples, but can be minimized by
special measurement techniques and instrument designs; (g) and (h) can be avoided by buffering
the solutions to constant pH and adjusting the concentration of reagents; (i) and (j) occur rarely
with some particular absorber molecules and must be treated on a case-by-case basis; (k) never
occurs in standard laboratory instruments with conventional light sources..