Each electrophysiology laboratory setup is different, reflecting the requirements of the experiment or the foibles of the experimenter. This unit describes components and considerations that are common to all setups dedicated to measurement of electrical activity in cells. First, the basic concerns of setting up the equipment in a functional and accessible manner are described (see Equipment Placement). Then two typical setups are briefly described: one for in vitro extracellular recording, the other for single-channel patch-clamping. An electrophysiological setup has four main requirements: (1) environment: the means of keeping the preparation healthy; (2) optics: the means of visualizing the preparation; (3) mechanics: the means of stably positioning the microelectrode; and (4) electronics: the means of amplifying and recording the signal. This unit addresses primarily the mechanics (see Vibration Isolation Methods) and electronics (see Electrical Isolation Methods) of the electrophysiological laboratory setup. at hand while one is looking through the microscope. Choice and placement of electronics is again a matter of personal preference. There are minimalists who make do with just an amplifier and a computer; others insist on a loaded instrument rack. An oscilloscope is important, because the computer is often insufficiently flexible. Furthermore, an oscilloscope often reveals unexpected subtleties in the signal that are not apparent on the computer screen if the sample interval is not set exactly right. The oscilloscope should be at eye level. The microelectrode amplifier should be directly above or below so that adjustments can be easily made and monitored. Last, the computer should be placed as far as possible but still within a long arms reach from the microscope. This reduces the radiative noise from the monitor. It also ensures that the experimenter does not bump the microscope while typing during a critical recording. Finally, resist the temptation to try to perfect the setup before starting. As soon as possible, do an experiment. Not only will this provide personal satisfaction, it may also highlight specific problems with the setup that need to be corrected or, better yet, indicate that an anticipated problem is not so pressing after all.
UNIT 6.1
EQUIPMENT PLACEMENT
While the placement of equipment is directed by personal preferences, a survey of electrophysiologists common preferences may be instructive. Electrophysiologists tend to work alone in small rooms. This is partly because their work often involves bursts of intense, intricate activity during which distractions are inadmissible. Furthermore, small rooms are often physically quieter as vibrations and air currents are reduced. It is usually sensible to first set up the microscope and its intimate attachments, such as the chamber, the manipulators, and the temperature control system (if installed). The rationale here is that the first priority is to keep the cells happy in their quiescent state, and the second priority is to ensure that recording from them is not consistently fatal. The former is assisted by a good environment, the latter by good optics and mechanics. Working outward from the microscope, it is clearly prudent to keep such things as perfusion stopcocks and micromanipulator controls off of the vibration isolation table. Ideally, these should be placed on small shelves that extend over but do not touch the table. This ensures that they can be accessed without causing damaging vibrations and are conveniently
6.1.1
Table 6.1.1
Item
Suggested manufacturersa
Extra/intracellular microelectrode setup Vibration isolation table Newport Technical Manufacturing Corp. (Micro-g) Microscope Carl Zeiss Nikon Olympus Micromanipulators Mechanical Narashige Prior Hydraulic Narashige Pneumatic Technical Products International Motorized Newport Piezoelectric Burleigh Microelectrode amplifiers Axon Instruments Warner Instrument Dagan Tape recorders (VCR-based) Instrutech Neurodata Bio-Logic Oscilloscopes Tektronix Gould Instrument Systems Electrode fabrication supplies Glass See UNIT 6.3 Pullers David Kopf Instruments Sutter Instrument Microelectrode holders Axon Instruments E.W. Wright Warner Instrument Chamber, temperature control Medical Systems Narishige Traditional patch-clamp setup Vibration isolation table Microscope, inverted Newport Technical Manufacturing Corp. (Micro-g) Zeiss (Axiovert) Nikon Olympus Narashige Technical Products International Newport Burleigh Sutter Instrument Axon Instruments Warner Instrument Dagan Heka (available through ALA Scientific) Instrutech Neurodata Bio-Logic
continued
6.1.2
Current Protocols in Neuroscience
Table 6.1.1
Item
Suggested manufacturersa
Traditional patch-clamp setup (continued) Oscilloscopes Tektronix Gould Instrument Systems Pipet fabrication supplies Glass See UNIT 6.3 Pullers Narashige Sutter Instrument Microforges Narashige Homemade, based on Zeiss metallurgical microscope or Olympus CH microscope Coaters Narashige Hydrophobic coating Dow Corning (Sylgard 184) GC Electronics (Q-dope) Microelectrode holders Axon Instruments E.W. Wright Chamber, temperature control Medical Systems, Narashige, Warner Instrument Patched slice setupb Microscope, low power Vibratome Optical recording setup Photomultipliers Imaging systems Carl Zeiss Technical Products International Hamamatsu Thorn Axon Instruments ETM Systems Inovision Universal Imaging
aSee SUPPLIERS APPENDIX for contact information. bOther requirements are as for a traditional patch-clamp setup.
Voltage excursions are measured in the 10- to 10-mV range; thus, a low-noise voltage amplifier with a gain of at least 1000 is required (see UNIT 6.2). Table 6.1.1 lists the equipment necessary for extracellular recording along with suggested manufacturers.
ment, such as Nomarski, phase, or Hoffman. Nomarski (or differential interference contrast, DIC) is best for critical placement of the electrode because it gives a very crisp image with a narrow depth of field. However, it is typically expensive and may be incompatible with applications that also seek to make fluorescence measurements simultaneously with electrophysiological ones, as this requires placing polarizing filters in the light path, which attenuates the intensity considerably. Phase contrast is acceptable for less demanding applications and provides better contrast for fine processes; it is typically best for routine patch-clamp applications. Hoffman presently ranks as a less expensive, slightly degraded version of Nomarski. Regardless of the contrast method selected, an inverted microscope is preferable for two reasons: (1) it usually allows easier top access
Neurophysiology
6.1.3
Current Protocols in Neuroscience
for the electrode, as the objective lens is underneath the chamber, and (2) it usually provides a larger, more solid platform upon which to bolt the micromanipulator. If a top-focusing (upright) microscope is the only option, it is critical that there be a fixed stage, and that the focus mechanism move the objective, not the stage. Upright microscopes have recently become more widely used as brain slice thickness has increased. Superficial neurons in thick slices are not easily viewed with an inverted scope. Recent improvements in optics by many manufacturers have made it possible to record and measure fluorescent signals using water immersion optics. The micromanipulator should permit fine, smooth movement down to a few microns per second or less. The vibration and stability requirements of the micromanipulator depend upon whether one wishes to record from a cell-attached or a cell-free (inside-out or outside-out) patch. In the latter case, the micromanipulator needs to be stable only as long as it takes to form a seal and pull away from the cell. This usually takes less than a minute. For cellattached or whole-cell recording (see UNIT 6.6), stability requirements are more stringent. Finally, the electronic requirements for singlechannel recording are more complex than for extracellular recording. Excellent amplifiers suitable for this purpose are commercially available from a number of suppliers (see Table 6.1.1). The patched slice technique, a recent extension of patch-clamping, requires a setup that borrows features from both in vitro extracellular and conventional patch-clamp configurations. For example, this technique may require a chamber that continuously perfuses and oxygenates the slice. In most other respects, the setup is similar to the conventional patch-clamp setup, except that the optical requirements depend upon whether one is using the thick-slice or thin-slice approach. Whereas a simple dissecting microscope may suffice for some thickslice applications, the thin-slice approach requires a microscope that provides 300- to 400fold magnification, preferably upright with contrast enhancement.
than on a complicated air table that tries to compensate for a micromanipulators inadequacies. A good micromanipulator is solidly constructed and compact, so that the moment arm (through the body of the manipulator and the tip of the electrode, to the cell in the chamber) is as short as possible. Ideally, the micromanipulator should be attached close to the chamber, preferably bolted directly to the microscope stage. The headstage of the recording amplifier should, in turn, be bolted directly to the manipulator (not suspended on a rod), and the electrode should be short. For most fine work, such as patch-clamp recording, it is preferable to use remote-controlled micromanipulators to eliminate hand vibration (although a fine mechanical manipulator, coupled with a steady hand, may sometimes be adequate). Currently, there are three main types of remote controlled micromanipulators available: motorized, hydraulic/pneumatic, and piezoelectric (see Table 6.1.1). Motorized manipulators tend to be solid and compact and have excellent long-term stability. However, they are often slow and clumsy to move into position, and may exhibit backlash when changing direction. Hydraulic drives are fast, convenient, and generally backlash-free, but some models may exhibit slow drift when used in certain configurations. Piezoelectric manipulators have properties similar to motorized drives, except for their stepwise advancement. These components may be combined in order to provide easy short- and long-distance manipulation. The headstage can then be controlled separately to move a great distance (e.g., to change an electrode) or a small distance (e.g., for the final approach to the cell). Antivibration tables are usually composed of a heavy slab on pneumatic supports. Tables of varying cost and complexity are commercially available, but a homemade table consisting of a slab resting on partially inflated smalldiameter inner tubes may be adequate, especially if high-quality micromanipulators are used. Tennis balls and tombstones are among the materials that some investigators have put to work as elements of vibration isolation tables.
6.1.4
Current Protocols in Neuroscience
the area of the loop cut by the flux, or try shielding the magnetic source with mu-metal.
6.1.5
Current Protocols in Neuroscience
Patch-Slice Recording
Blanton, M.G., Lo Turco, J.J., and Kriegstein, A.R. 1989. Whole cell recording from neurons in slices of reptilian and mammalian cerebral cortex. J. Neurosci. Methods 30:203-210. Edwards, F.A., Konnerth, A., Sakmann, B., and Takahashi, T. 1989. A thin slice preparation for patch clamp recordings from neurons of the mammalian central nervous system. Pflgers Arch. Eur. J. Physiol. 414:600-612.
Contributed by Alan Finkel Axon Instruments Foster City, California Richard Bookman University of Miami School of Medicine Miami, Florida