Medical, Laboratory, Industrial

Air Systems
Vacuum Systems
Medgas Pipeline

Ohi o Medi c al Cor por at i on

Desk Ref er enc e


Ohio Medical Corporation
has been the leading supplier of Healthcair medical air and vacuum pumping systems, Ohio
Medical Corporation and Amvex medical suction regulators, Ohio Medical and Amvex air and
oxygen flow meters and Aeros portable suction equipment since 1958.
Ohio Medical Corporation is an ISO 9001 and ISO 13485 registered and FDA regulated medical
device manufacturing facility. All Ohio Medical medgas products are produced to conform to NFPA
99 standards.
With a leading research and development team, dedicated service department, and knowledgeable
customer support teams, we will continue to provide the medical industry with leading technology
and quality products as we have for the past five decades.
Ohio Medical is the only name you need to know for your medical suction and oxygen therapy and
medgas equipment needs.
Ohio Medical has maintained a presence in the medical industry for well over 45 years under the
Ohio Medical, Amvex, and Aeros-Instruments & Squire-Cogswell names. We are a leading supplier
of medical pumping systems, with a sound reputation in hospital environments around the world.
In the 1950 the first Ohio Medical Vacuum regulator was introduced under the name of Ohio
Chemical, followed by the Intermittent Vacuum regulator in 1951 and the HH Model in 1953. These
regulators were manufactured by Squire-Cogswell (now Ohio Medical) and private labeled for Ohio
Medical, Ohmeda and successive companies since their introduction over 45 years ago.
In July of 2005, Ohio Medical Corporation was formed, combining the Ohmeda Suction & Oxygen
Therapy products, Aeros portable suction equipment, the Healthcair and Selectorr pumping systems
and medgas pipeline products. Finally in November 2007 Ohio Medical acquired Amvex Corporation
a leading supplier of medical gas devices. This union created a synergistic blend of products and
technology with strong customer and patient safety focus.
www.ohiomedical.com
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
1
PIPELINE
Outlets ....................................................................................................................2
Integrated Flowmeter ..............................................................................................3
Retrot Wall Outlets ................................................................................................4
Zone Valve Boxes ...................................................................................................5
Valves with Extensions ...........................................................................................5
Alarms: Area, Master and Combo .........................................................................6
Refrot Alarm ..........................................................................................................7
Hose Retractor .......................................................................................................7
Hose and Assemblies: Bulk, Spiral Hoses and Hose Assemblies .........................8
Gas Control Panel ..................................................................................................9
Ceiling Columns/Pedastals .....................................................................................9
Manifolds ..............................................................................................................10
Emergency Oxygen Inlet Station ..........................................................................10
SOT PRODUCTS
Vacuum Regulators .............................................................................................. 11
Flowmeters ........................................................................................................... 11
Adapters ............................................................................................................... 11
Portable Suction Equipment ................................................................................. 11
VACUUM SYSTEMS
Rotary Vane Vacuum Systems .............................................................................12
Simplex Rotary Vane Vacuum System .................................................................15
Rotary Claw Vacuum Systems .............................................................................17
Oasis Vacuum Systems ........................................................................................20
Water Mister Liquid Ring Vacuum Systems .......................................................24
AIR SYSTEMS
Scroll Compressed Air Systems ...........................................................................27
Oil-Less Recipriocating Piston Air Compressor Systems .....................................30
Simplex Air System ...............................................................................................33
Instrument Air System ..........................................................................................35
Air Treatment Modules (For Oil-less Compressors with Desiccant Dryers) .........36
TECHNICAL INFORMATION
Calculated Peak Demand for a Medical Vacuum System ....................................38
Calculated Peak Demand for an Medical Air System ...........................................39
Manifold Sizing .....................................................................................................40
Medical Gas Equipment Mounting Height Detail ..................................................41
Suggetsed Medical Gas Master Alarm Points for Level 1 Facilities .....................42
Medical Alarm Panel Wiring ..................................................................................44
Tabl e of Cont ent s
For ordering information, please call 1-800-448-0770 or 847-855-0500
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
2
Out l et s
Features:
Accepts Chemetron/Allied

, Ohio Diamond, Puritan Bennett and DISS specic adapters


Available, Wall/Celing, Console and Ceiling Column
Pin indexed to prevent interchangeability of gas services
Cleaned for medical gas service
NFPA 99 Compliant
Inlet pipe can be rotated 360 degrees for ease of installation
Gas specic back bodies can accept either Quick Connect or DISS front identication bodies
Outlet can accommodate various nished wall thicknesses from 3/8 (10 mm) to 1-1/4 (32 mm).
DESCRIPTION/
PRODUCT
Diamond
Compatible Wall
Chemetron
Compatible Wall
Puritan-Bennett
Compatible Wall
DISS
Wall
DISS
Ceiling
Oxygen 261010-1 261020-1 261040-1 261000-1 261000-3
Vacuum 261010-5 261020-5 261040-5 261000-5 261000-7
Air 261010-13 261020-13 N/A 261000-13 261000-15
Nitrous Oxide 261010-9 261020-9 261040-9 261000-9 261000-11
Evacuation 261010-17 261020-17 261040-13 261000-21 261000-23
Nitrogen N/A N/A N/A 261000-17 261000-19
Carbon Dioxide N/A N/A N/A 261000-25 261000-27
Slide 261690 261690 261290 261690 N/A
Slides, blanks and electrical receptacle are also available.
Rebuild kits available
Chemetron

is a registered trademark of Allied Healthcare Products, Inc.


Ohio Diamond Compatible
Wall Outlet
Chemetron Compatible
Wall Outlet
DISS Wall Outlet
DISS Ceiling Outlet
Puritan-Bennett Compatible
Wall Outlet
Adapters are available from Ohio Medical Corporation
DISS
Console or
Column Outlet
Ohio Diamond Compatible
Console or Celing Column
Outlet
Chemetron Compatible
Console or Ceiling Column
Outlet
Puritan-Bennett Compatible
Console or Ceiling Column
Outlet
www.ohiomedical.com
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
3
I nt egr at ed
Fl ow met er
The Integrated Flowmeter outlet combines a Dial Flowmeter and a Medical Gas Outlet
into a single compact design. The right side port connection has a control knob which
allows the user to adjust the ow setting, while the left side port is a direct connection to
the gas supply. A Flow Indicator displays whether or not ow has been activated by the
Control Knob.
Flow Ranges available: 0-8 LPM, 0-15 LPM, 0-25 LPM.
A large color coded front plate is used for ease of gas identication and aesthetic appeal.
The latch-valve assembly is gas specic.
All outlets are cleaned and degreased for medical gas service, factory assembled and
tested.
All Ohio Medical

products comply with NFPA-99.


FEATURES:
Flow rate meets the requirements of
NFPA and CSA
Each outlet is 100 % tested and cleaned
for medical gas service
Indexed to eliminate interchangeability
of gas services
Flow Indicator
Five year warranty
l
2
3
4
5
6
7
l
2
3
4
5
6
7
Plow Control Knob
Plowmeter Port
Outlet Port
Tubing Nipple
8rass 8arrel
Pastening Screws
|ndex Pin
261XXX-X-XXX
Part Confgurator
Connection Style
000................DISS
010.............Diamond
020........... Chemetron
Gas / Outlet Type
1 .......................Oxy / Wall
13 .......................Air / Wall
2 ....................Oxy / Console
14 ....................Air / Console
Flowrate
08................ 8 LPM
15............... 15 LPM
25............... 25 LPM
Flowmeter Style
C ..................Click
T...............Thorpe**
NOTE: Not all objects appear as shown
** Thorpe style not yet available
as of the date of this publication.
Please contact Ohio Medical
Corporation for updates.
For ordering information, please call 1-800-448-0770 or 847-855-0500
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
4
Quick Connect Diamond 2&3 to
Diamond or DISS Compatible
FEATURES
Accepts Diamond or DISS Gas Specic Adapters.
Pin indexed to prevent interchangeability of gas services.
Each outlet is cleaned for medical gas service and pressure
tested.
Ret r ot Wal l
Out l et s
Retrot kit for diamond 2/3 rough-in part numbers
261220 Rough-In RetroFit Kit Pressure Gas 261221 Vacuum Rough-In RetroFit Kit
Retrot kit for diamond 2/3 latch valve part numbers
261170-1 Latch Valve Oxy 261160-1 Latch Valve DISS Oxy
261170-2 Latch Valve Vac 261160-2 Latch Valve DISS Vac
261170-3 Latch Valve N
2
O 261160-3 Latch Valve DISS N
2
O
261170-4 Latch Valve Air 261160-4 Latch valve DISS Air
261170-6 Latch Valve WAGD 261160-5 Latch Valve DISS N
2
261160-6 Latch Valve DISS WAGD
261160-7 Latch Valve DISS CO
2
www.ohiomedical.com
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
5
Zone Val ve
Boxes
Single Zone Valve Box
For 1/2, 2, 2-1/2 and 3 Valves
Features:
Full port design
Blow-out proof stem
Cleaned for medical gas service
Teon seats
Dual gauge port on pipe extension
Part Numbers:
261901-05 1/2 Single Zone Valve Box
261901-07 3/4 Single Zone Valve Box
261901-10 1 Single Zone Valve Box
261901-12 1-1/4 Single Zone Valve Box
261901-15 1-1/2 Single Zone Valve Box
The following require a 6-1/2 deep box
261901-20 2 Single Zone Valve Box
261901-25 2-1/2 Single Zone Valve Box
261901-30 3 Single Zone Valve Box
Multiple Zone Valve Box
4 & 6 Deep Box
Features:
Full port design
Blow-out proof stem
Cleaned for medical gas service
Teon seats
Dual gauge port on pipe extension
Part Numbers:
261902-XXXX Double Zone Valve Box
261903-XXXXXX Triplex Zone Valve Box
261904-XXXXXXXX Quadruplex Zone Valve Box
261905-XXXXXXXXXX Quintuplex Zone Valve Box
Replace XX with: 05 = 1/2
07 = 3/4
10 = 1
12 = 1 1/4
15 = 1 1/2
20 = 2 (Requires a 6 deep box)
Val ves w i t h Ex t ensi ons
Features:
Full port design
Blow-out proof stem
Sizes range from 1/2 (13mm) to 4 (76.2mm)
Cleaned for medical gas service
Teon seats
Valve with Extensions
1/2 through 4 - Dual gauge ports on pipe extensions
4 - Single gauge port on pipe extension
Non-Locking Valves
Part Number
261630-05
261630-07
261630-10
261630-12
261630-15
261630-20
261630-25
261630-30
261630-40
Locking Valves
Part Number
261600-05
261600-07
261600-10
261600-12
261600-15
261600-20
261600-25
261600-30
261600-40
For ordering information, please call 1-800-448-0770 or 847-855-0500
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
6
Al ar ms
Ar ea, Mast er and Combo
Local
Combo Alarm
(Local or Remote)
Features:
Complies with NFPA 99 & FCC Part 15
Adjustable audible alarm repeat
Field adjustable pressure settings
Equipped with local or remote sensors (Area)
High visibility LED readouts
Optional BMS modules available
UL Listed
Normally Open or Closed Contacts (Master)
261861-X
261862-XX
261863-XXX
261864-XXXX
261865-XXXXX
261866-XXXXXX
261867-XXXXXXX
261868-XXXXXXXX
1 Gas Area Alarm
2 Gas Area Alarm
3 Gas Area Alarm
4 Gas Area Alarm
5 Gas Area Alarm
6 Gas Area Alarm
7 Gas Area Alarm
8 Gas Area Alarm
261851-X
261852-XX
261853-XXX
261854-XXXX
261855-XXXXX
261856-XXXXXX
261857-XXXXXXX
261858-XXXXXXXX
1 Gas Area Alarm
2 Gas Area Alarm
3 Gas Area Alarm
4 Gas Area Alarm
5 Gas Area Alarm
6 Gas Area Alarm
7 Gas Area Alarm
8 Gas Area Alarm
Remote
X = gas 1 = Oxygen
2 = Vacuum
3 = Nitrous Oxide
4 = Air
5 = Nitrogen
6 = Evacuation
7 = Carbon Dioxide
Area Alarm
26189X- XXX
Local 6
Remote 5
Master Alarm Modules
10 = 1, 10 point
20 = 2, 10 point
30 = 3, 10 point
40 = 4, 10 point
Replace X with gas
1 = Oxygen
2 = Vacuum
3 = Nitrous Oxide
4 = Air
5 = Nitrogen
6 = Evacuation
7 = Carbon Dioxide

Master Alarm
261890-XX
10 Point 10
20 Point 20
30 Point 30
40 Point 40
50 Point 50
If Dry Contacts are
required, add C to
the part number
Example: 261890-20C
www.ohiomedical.com
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
7




Ret r ot Al ar m
The Ohio Medical Healthcair

Series hose assemblies are manufactured


with FDA approved medical grade anti-static PVC tubing. The tubing is
reinforced with polyester braiding. The maximum working pressure is 200
psi @ 70
o
F. The ttings have either the coupling or the grips color coded
per CGA C-9 standards. The ttings are permanently crimped on each end
to the tubing with ferrules.
Part Numbers

DISS Female to Male DISS Male - 5 Hose Length
263400-1-05 Oxygen Hose Assembly
263400A-2-05 Vacuum Hose Assembly
263400-3-05 Nitrous Oxide Hose Assembly
263400-4-05 Air Hose Assembly
263400-5-05 Nitrogen Hose Assembly
263400A-6-05 WAGD Hose Assembly
263400-7-05 Carbon Dioxide Hose Assembly
DISS Female to Ohmeda Compatible Quick Connect - 5 Hose
Length
263401-1-05 Oxygen Hose Assembly
263401A-2-05 Vacuum Hose Assembly
263401-3-05 Nitrous Oxide Hose Assembly
263401-4-05 Air Hose Assembly
263401A-6-05 WAGD Hose Assely

DISS Female to Chemetron Compatible Quick Connect - 5
Hose Length
263402-1-05 Oxygen Hose Assembly
263402A-2-05 Vacuum Hose Assembly
263402-3-05 Nitrous Oxide Hose Assembly
263402-4-05 Air Hose Assembly
263402A-6-05 WAGD Hose Assembly
Retractor
261746Retractor
Hose Ret r ac t or
FEATURES AND BENEFITS
High visibility LED readouts
Field adjustable pressure settings for each module
Local or Remote sensors
Adjustable audible repeat cycle
Universal trim plate sized for any brand alarm
The Ohio Medical Retrot Alarm is designed to replace and/or update most manufac-
turers master alarms, area alarms and combination alarms without disturbing the wall.
The retrot alarm kit includes a front panel display, a frame to cover the existing box, a
power supply assembly, and area sensors if required.
The retrot alarm is available in two sizes, four and six bay. The four bay box can ac-
commodate up to 6 gases, or a combination of 4 gases and a 10 point master alarm
with overall frame size of 19L x 11.5H. The six bay can accommodate up to 8 gases
and a 10 point master with an overall frame size of 26L x 11.5H. The retrot box can
be tted in any combination with each vertical slot capable of housing one master mod-
ule or two area gas modules.

For ordering information, please call 1-800-448-0770 or 847-855-0500
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
8
Hose and Assembl i es
Bul k , Spi r al Hoses and Hose Assembl i es
Product Features:
Spiral Hose:
Spiral hose is available in many combination and colors
Single, double, triple, and quadruple
3 ft coiled hose expands to 10 ft working length
Standard working lengths: 10 ft, 15 ft and 20 ft.
Bulk Hose:
Available in a variety of gases
UL Listed, oxygen cleaned and tested
Standard 300 ft. rolls for 1/4 hose
Standard 250 ft. rolls for 5/16 hose
Amvex hose is made of high quality PVC material
Conductive and Non-Conductive available
Available in USA and ISO colors
Hose Assemblies:
CE marked & CSA approved
Available in custom lengths and congurations to suit your requirements
Available in a variety of gases
Available in USA and ISO colors
Part Congurator
Color Code:
USA ......... U
ISO ............I
Gas Type:
Oxygen ...................O
Med Air ................... A
Vacuum ................... V
N2O ........................ 2
Nitrogen ..................N
WAGD ................... W
CO2 ........................C
Instrument Air ..........I
Length in feet
Hose Type:
Conductive 1/4 .................. C4
Conductive 5/16 ................ C5
Non-Conductive 1/4........... N5
Non-Conductive 5/16......... N5
General Purpose 1/4 ......... G4
Ohmeda Female ........................ OF
Ohmeda Male ........................... OM
Chemetron Male ....................... CM
Chemetron Female .................... CF
1/8 FNPT ...................................F2
1/4 FNPT ...................................F4
1/8 MNPT ................................. M2
1/4 MNPT ................................. M4
DISS Male ................................ DM
DISS Male w/demand Check.....DC
DISS Handtight ..........................DH
DISS Hex Nut ............................ DF
Schrader Female ....................... SF
Schrader Male ...........................SM
Puritan Female .......................... PF
Puritan Male ..............................PM
Oxequip Male ............................XM
Medstar Male ............................ MM
Medstar Female.........................MF
British Male ................................BM
British Female............................ BF
French male...............................FM
French Female ...........................FF
NIST Male ................................. NM
NIST Female ............................. NF
German Male ............................ GM
German Female......................... GF
Blank End .................................. XX
NOTE: Not all congurations are possible.
HS-XXYZ-XXYYXX
Inlet and Outlet Connection:
For correct hose matrix follow these rules
when selecting inlet and outlet connections:
1. Female ttings rst
2. Numbers before letters
3. In alphabetical order
www.ohiomedical.com
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
9
Nitrogen, Air, CO2
Pressure Range: 0-300 PSIG
Pull to Lock
Push to Adjust
One DISS Outlet on Unit
Part Numbers:
261760 Nitrogen Control Panel
261911 Air Control Panel
263280 CO
2
Control Panel
261911-SH Air Control Panel with Schrader Fitting
Cei l i ng Col umns
Pedast al s
Available Stationary or Retractable
(Electrical Powered or Manual)
Complies with NFPA 99
Gas Cont r ol
Panel
For ordering information, please call 1-800-448-0770 or 847-855-0500
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
10
Mani f ol ds
Digital LCD Pressure Display
NEMA 4x Enclosure
Automatic Leak Detection
Automatic Transducer Diagnostic Testing
Emer gency Ox ygen I nl et St at i on
Part Numbers:
261825 Emergency Oxygen Inlet Station
261701 Check Valve, 3/4
261702 Check Valve, 1
261703 Check Valve, 1-1/4
261704 Check Valve, 1-1/2
261705 Check Valve, 2
232600 Relief Valve, 1/2 NPT, Set at 75 PSIG
232602 Relief Valve Pipeaway Adapter, 1/2 NPT
Part Number Matrix
2634XX-XX-X*
Control Unit Only ....00
1 x 1 .....................01
2 x 2 .....................02
3 x 3 .....................03
4 x 4 .....................04
5 x 5 .....................05
6 x 6 .....................06
7 x 7 .....................07
8 x 8 .....................08
9 x 9 .....................09
10 x 10 ...................10
120V
Oxygen ...................32
N
2
O .........................34
N
2
.........................36
Air .........................38
CO
2
.........................40
All Other Voltages
Oxygen ...................33
N
2
O .........................35
N
2
.........................37
Air .........................39
CO
2
.........................41
Heater Kit................. H
*Add to the end of the part number
if a heater kit is required
www.ohiomedical.com
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
11
Vac uum Regul at or s:
Di gi t al , Anal og, Pedi at r i c and Neonat al Model s
Fl ow met er s
Por t abl es
Digital:
Easy to read display
No moving parts means no calibration
required
Accuracy is +/- 1% of full scale
Analog:
Analog gauge designed for clarity
Accuracy of +/- 3% of full scale
Impact resistant ow tube and
shroud
D-Shape knob prevents striping
Pressure compensated, accurate
ow readings
White background improves
visibility
Color coded to help prevent cross
connections
5 Year Warranty (1 year on the 0-1
LPM & 0-200 CCM Flowmeter)
Moblvac

III is a multipurpose portable suction unit.


It produces both constant & intermittent suction with
vacuum & ow levels equaling that of wall suction.

IDEAL FOR: Hospitals and extended care facilities
requiring wall vacuum performance for Nasogastric,
Thoracic, Oral, and Tracheal suctioning procedures.
Care-e-vac

3 is a high vacuum & high


ow AC/DC powered suction unit. It
provides constant suction when wall
outlets are unavailable. The unit is
compact and very lightweight. An
optional oor stand & crash cart mount
are available.
IDEAL FOR: Code carts, patient transportation or wherever
a powerful suction unit with battery power is required.
Patented Push-To-Set
TM
Improves patient
safety
Prevents unregulated suction
Simple, one handed operation
Quick-to-Max
2 turns to full wall vacuum for
emergency situations
Intermittent Mode: No-Delay ON cycle
Large, Easy to Read Gauge, Color coded
ranges
Analog: Improved glow-in-the-
dark gauge
Digital: Accuracy is +/- 1% of
full scale
Analog: Accuracy of +/- 5% of
full scale
Modular Components
Digital units (cases) available
in many colors.

Features:
Can easily upgrade from
Continuous to Cont/
Intermittent, or Analog to
Digital
Durable
Easily service and clean the
unit with a cold sterilant
Color cases available for
inventory management
Amvex Corporation manufactures a wide variety of quick
connect medical gas adapters. Amvex offers Ohmeda,
Chemetron, Puritan Bennett, Medstar, Oxequip and Schrader
styles to suit your needs. Available options include 1/8, 1/4
MNPT and FNPT, 1/4 and 5/16 Hose Barb and much more.
Adapt er s
For ordering information, please call 1-800-448-0770 or 847-855-0500
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
12
Rot ar y Vane Vac uum
Syst ems
Continuous on demand
Lubricated oil recirculated or dry running
NFPA 99 compliant
U.L. Listed Electrical Control Panel
Available in Simplex, Duplex, Triplex or Quad
Congurations:
Tank Mounted: Up to 10 Hp
Stack Mounted: Up to 25 Hp
Warranty: 30/24 Months
Vibration Free
Oil sealed systems feature synthetic oil, reducing maintenance
and oil changes.
Design capacity values should correspond with the Calculated
Peak Demand. This will allow for the reserve pump as is re-
quired per NFPA for level 1 installations.
Lubricated
Model
No. HP
SCFM @ 19 HG
(each pump)
S200 2 7.7
S300L 3 11.3
S300 3 16.1
S500L 5 25.6
S500 5 43.2
S750 7.5 60
S1000 10 71.1
S1500 15 103
S2000 20 112
S2500 25 172
Maximum Vacuum 29.1 Hg
Dry
Model
No. HP
SCFM @ 19 Hg
(each pump)
D100 1 3.25
D200 2 8,27
D300 3 12.49
D500L 5 15.37
D500 5 22.1
D750 7.5 30.42
D1000 10 55.4
Maximum Vacuum
1, 2, & 10 Hp................ 25 Hg
3 & 5 Hp....................... 27 Hg
7.5 Hp .......................... 22 Hg
Discharge
Vanes
Rotor
Inlet
Casing
www.ohiomedical.com
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
13
Tot al Rec i r c ul at ed Rot ar y Vane
Cont i nuous On Demand Medi c al Vac uum Syst em
Oil Sealed Rotary Vane Medical Vacuum System
The Ohio Medical Corporation Lubricated Rotary Vane Medical Vacuum System shall be fully NFPA 99 compliant for
use in medical vacuum and dual Medical/Surgical applications. The unit will consist of electric motor driven pumps,
vacuum receiver, electrical control system, and interconnection piping and wiring. The components shall be modularly
assembled to accommodate most existing doorways and designed for serviceability. The packaged unit shall be
factory tested prior to shipment and warranted for a period of (30) months from date of shipment or (24) months from
date of start-up.
Vacuum Pump Module
The total recirculating oil sealed rotary vane vacuum pump shall be single stage air cooled, and capable of producing
a maximum vacuum level of 29.1 Hg. The pump assembly shall include an integral anti-suck back valve, exhaust oil
separator delivering 99.9% oil-free air, oil level sight glass, and an exhaust pressure gauge. Each pump is protected
by a temperature switch, check valve, pump isolation valve, source isolation valve and exible connector.
Vacuum Receiver
The vacuum receiver shall be constructed to ASME standards, rated for full vacuum, and include a valved by-pass,
manual drain valve, vacuum gauge and the National Board label.
System Controls
The Underwriters Laboratories Listed electrical motor control system shall be of a fuseless design in a NEMA 12
enclosure. The Continuous On Demand feature shall stop the operation of the motors during periods of low or no
demand. The controls include individual self protected combination motor controls with short circuit, single phase
and thermal overload protection, individual 120 volt control circuit transformers with fuseless primary and secondary
protection, pressure sensors, and an electronic controller to automatically switch the operating sequence of the
vacuum pumps. The cabinet door shall have an HMI (Human Machine Interface) system status display to include
system pressure, pump operation, accumulated time, maintenance interval, fault conditions and silence button; lighted
Hand-Off-Automatic selector switches and safety disconnect operating handles.
All required local alarm functions shall be integrated into the packaged system. The circuitry shall be designed so the
audible signal can be silenced and the visual indicator will remain until the fault has been cleared and the reset button
actuated. Local alarm functions shall be provided for reserve pump in use (lag alarm). Dry contacts are furnished for
remote (master) alarm signals.
Accessories
Accessories included for job site installation are inlet and discharge exible connectors, source isolation valve, vibration
mounting pads and touch-up paint.
For ordering information, please call 1-800-448-0770 or 847-855-0500
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
14
Tot al Rec i r c ul at ed Rot ar y Vane Cont i nuous
On Demand Vac uum Syst em Sel ec t i on Char t
D
e
s
i
g
n

C
a
p
a
c
i
t
y

S
C
F
M

@

1
9
"

H
g

/


c
a
p
a
c
i
t
y

o
f

e
a
c
h

p
u
m
p
D
e
s
i
g
n

C
a
p
a
c
i
t
y

A
C
F
M

@

2
5
"

H
g

/


c
a
p
a
c
i
t
y

o
f

e
a
c
h

p
u
m
p
S
y
s
t
e
m
C
o
n


g
u
r
a
t
i
o
n
H
P
T
a
n
k

S
i
z
e

(
g
a
l
l
o
n
s
)
L
e
n
g
t
h

(
A
)
W
i
d
t
h

(
B
)
H
e
i
g
h
t

(

C
)
W
e
i
g
h
t
S
y
s
t
e
m

M
o
d
e
l

N
u
m
b
e
r
7.7 / 7.7 19.7 / 19.7
Vertical Tank
Mounted Duplex
2 80 60 40 75 1000 S200B-T2V
11.3 / 11.3 30.8 / 30.8
Vertical Tank
Mounted Duplex
3 80 60 40 75 1100 S300LB-T2V
16.1 / 16.1 44 / 44
Vertical Tank
Mounted Duplex
3 80 60 40 80 110 S300B-T2V
25.6 / 25.6 70 / 70
Vertical Tank
Mounted Duplex
5 80 60 43 80 1500 S500LB-T2V
43.2 / 43.2 114.7 / 114.7
Horizontal Tank
Mounted Duplex
5 120 80 36 66 1300 S500B-T2
60 / 60 163.8 / 163.8
Horizontal Tank
Mounted Duplex
7.5 120 90 42 77 1700 S750B-T2
71.1 / 71.1 194 / 194
Horizontal Tank
Mounted Duplex
10 120 90 46 77 2000 S1000B-T2
101 / 101 271 / 271
Stack Mounted
Duplex
15 240 83 80 90 3900 S1500B-ST2
112 / 112 306 / 306
Stack Mounted
Duplex
20 240 80 85 90 4400 S2000B-ST2
120 / 60 327.6 / 163.8
Stack Mounted
Triplex
7.5 240 107 66 69 3300 S750B-ST3
142.2 / 71.1 388 / 194
Stack Mounted
Triplex
10 240 107 66 69 3800 S1000B-ST3
172 / 172 460 / 460
Stack Mounted
Duplex
25 240 80 94 90 4900 S2500B-ST2
180 / 60 491.4 / 163.8
Stack Mounted
Quarduplex
7.5 240 107 66 80 4000 S750B-ST4
202 / 101 542 / 271
Stack Mounted
Triplex
15 240 137 72 90 5800 S1500B-ST3
213.3 / 71.1 582 / 194
Stack Mounted
Quarduplex
10 240 107 66 80 4800 S1000B-ST4
224 / 112 612 / 306
Stack Mounted
Triplex
20 240 131 74 90 5900 S2000B-ST3
303 / 101 813 / 271
Stack Mounted
Quarduplex
15 240 137 72 90 7300 S1500B-ST4
336 / 112 918 / 306
Stack Mounted
Quarduplex
20 240 131 74 90 7500 S2000B-ST4
344 / 172 920 / 460
Stack Mounted
Triplex
25 240 131 82 90 7000 S2500B-ST3
516 / 172 1380 / 460
Stack Mounted
Quarduplex
25 240 131 82 90 8800 S2500B-ST4
www.ohiomedical.com
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
15
Si mpl ex Rot ar y Vane Vac uum Syst em
Spec i c at i on/Sel ec t i on Char t
A
C
F
M

@

1
9
"

H
g

(
2
7
7

T
O
R
R
)
M
a
x
i
m
u
m

V
a
c
u
u
m

(
"
H
g
/
T
O
R
R
)
S
y
s
t
e
m

C
o
n


g
u
r
a
t
i
o
n
H
P
T
a
n
k

S
i
z
e

(
g
a
l
l
o
n
s
)
L
e
n
g
t
h

(
A
)
W
i
d
t
h

(
B
)
H
e
i
g
h
t

(

C
)
W
e
i
g
h
t
S
y
s
t
e
m

m
o
d
e
l

N
u
m
b
e
r
21 29.1/20
Vertical Tank
Mounted
2 60 30 43 74 290 S200B-T1V
30.8 29.1/20
Vertical Tank
Mounted
3 60 30 45 74 310 S300LB-T1V
44 29.1/20
Vertical Tank
Mounted
3 60 30 46 75 350 S300B-T1V
70 29.1/20
Vertical Tank
Mounted
5 60 30 46 75 450 S500LB-T1V
118 29.1/20
Vertical Tank
Mounted
5 90 31 50 80 550 S500B-T1V
163.8 29.1/20
Vertical Tank
Mounted
7.5 90 31 50 81 600 S750B-T1V
194 29.1/20
Vertical Tank
Mounted
10 90 31 54 81 1050 S1000B-T1V
Oil Sealed Rotary Vane Simplex Vacuum System
The Ohio Medical Lubricated Rotary Vane Vacuum System shall consist of an electric motor driven pump, vacuum receiver,
electrical control system, and interconnection piping and wiring. The components shall be modularly assembled to accommodate
most existing doorways and designed for serviceability. The packaged unit shall be factory tested prior to shipment and warranted
for a period of (30) months from date of shipment or (24) months from date of start-up.
Vacuum Pump Module
The total recirculating oil sealed rotary vane vacuum pump shall be single stage air cooled, and capable of producing a maximum
vacuum level of 29.1 Hg. The pump assembly shall include an integral anti-suck back valve, exhaust oil separator delivering 99.9%
oil-free air, oil level sight glass, and an exhaust pressure gauge. Each pump is protected by a temperature switch, check valve,
pump isolation valve, source isolation valve and exible connector.
Vacuum Receiver
The vacuum receiver shall be constructed to ASME standards, rated for full vacuum, and include a valved by-pass, manual drain
valve, vacuum gauge and the National Board label.
System Controls
The Underwriters Laboratories Listed simplex electrical motor control system shall be of a fuseless design in a NEMA 12 (dustproof)
enclosure. The Continuous On Demand feature shall stop the operation of the motor during periods of low or no demand. The
controls include a self protected combination motor control with short circuit, single phase and thermal overload protection, 120 volt
control circuit transformer with fuseless primary and secondary protection, pressure sensor and a temperature switch with reset
button. The cabinet door shall have an HMI (Human Machine Interface) system status display to include accumulated time and fault
conditions; lighted Hand-Off-Automatic selector switch and safety disconnect operating handle.
All required local alarm functions shall be integrated into the packaged system. The circuitry shall be designed so the visual
indicator will remain until the fault has been cleared and the reset button actuated. Local alarm functions shall be provided for high
air discharge temperature and thermal overload.
Accessories
Accessories included for job site installation are inlet and discharge exible connectors, source isolation valve, vibration mounting
pads and touch-up paint.
For ordering information, please call 1-800-448-0770 or 847-855-0500
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
16
Dr y Runni ng Rot ar y Vane Vac uum Syst em
Spec i c at i ons/Sel ec t i on Char t
D
e
s
i
g
n

C
a
p
a
c
i
t
y














S
C
F
M

@

1
9
"

H
g
/

c
a
p
a
c
i
t
y

o
f

e
a
c
h

p
u
m
p
D
e
s
i
g
n

C
a
p
a
c
i
t
y














A
C
F
M

@

2
5
"

H
g
/

c
a
p
a
c
i
t
y

o
f

e
a
c
h

p
u
m
p
S
y
s
t
e
m

C
o
n


g
u
r
a
t
i
o
n
H
P
T
a
n
k

S
i
z
e

(
g
a
l
l
o
n
s
)
L
e
n
g
t
h

(
A
)
W
i
d
t
h

(
B
)
H
e
i
g
h
t

(
C
)
W
e
i
g
h
t
S
y
s
t
e
m

m
o
d
e
l

N
u
m
b
e
r
3.25 / 3.25 5.8 / 5.8
Vertical Tank
Mounted Duplex
1 80 63 30 78 740 D100-T2V
8.27 / 8.27 13.8 / 13.8
Vertical Tank
Mounted Duplex
2 80 63 30 80 810 D200B-T2V
12.49 / 12.49 18 / 18
Vertical Tank
Mounted Duplex
3 80 63 32 83 940 D300B-T2V
15.37 / 15.37 20 / 20
Vertical Tank
Mounted Duplex
5 80 63 32 83 980 D500LB-T2V
22.1 / 22.1 30 / 30
Vertical Tank
Mounted Duplex
5 80 63 34 83 1000 D500B-T2V
30.42 / 30.42 42 / 42
Horizontal Tank
Mounted Duplex
7.5 120 90 42 77 1670 D750B-T2
55.4 / 55.4 114 / 114
Horizontal Tank
Mounted Duplex
10 120 102 60 77 2820 D1000B-T2
Oil-less Rotary Vane Vacuum System
Ohio Medical Corporation Oil-less Rotary Vane, NFPA 99 compliant vacuum system shall consist of electric motor driven pumps,
vacuum receiver, and an electrical control system. The components shall be modularly assembled to accommodate most existing
doorways. The system shall include interconnecting piping and wiring to provide a functional operating package with applicable
electrical and plumbing connections at the installation site. The packaged unit shall be factory tested prior to shipment and warranted
for a period of (30) months from date of shipment or (24) months from date of start-up.
Vacuum Pump Module
The dry running (oil-less) rotary vane vacuum pumps shall be single stage, air cooled, with self lubricating graphite vanes. The
vacuum pump assembly shall include an integral inlet lter, vacuum regulating valve, check valve, pump isolation valve, a source
isolation valve and exible connector.
Vacuum Receiver
The vacuum receiver shall be constructed to ASME standards, rated for full vacuum, and include a valved by-pass, manual drain
valve, vacuum gauge and the National Board label.
System Controls
The Underwriters Laboratories Listed electrical motor control system shall be of a fuseless design in a NEMA 12 enclosure. The
Continuous On Demand feature shall stop the operation of the motors during periods of low or no demand. The controls include
individual self protected combination motor controls with short circuit, single phase and thermal overload protection, individual 120
volt control circuit transformers with fuseless primary and secondary protection, pressure sensors, and an electronic controller to
automatically switch the operating sequence of the vacuum pumps. The cabinet door shall have an HMI (Human Machine Interface)
system status display to include system pressure, pump operation, accumulated time, maintenance interval, fault conditions and
silence button; lighted Hand-Off-Automatic selector switches and safety disconnect operating handles.
All required local alarm functions shall be integrated into the packaged system. The circuitry shall be designed so the audible signal
can be silenced and the visual indicator will remain until the fault has been cleared and the reset button actuated. Local alarm
functions shall be provided for reserve pump in use (lag alarm). Dry contacts are furnished for remote (master) alarm signals.
Accessories
Included for job site installation are inlet and discharge exible connectors, source isolation valve, vibration mounting pads, and
touch-up paint.
www.ohiomedical.com
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
17
Rot ar y Cl aw Vac uum
Syst ems
Dry, non-contacting design
No friction (or wearing parts) in pumping
chamber
Low maintenance
Long life
Reliability
High efciency (cfm/hp)
Duplex Tank Mounted
Model Number
H.P.
Each
Maximum
Vacuum Level
("Hg)
SCFM @ 19"
Hg Each
Design Capacity
SCFM @ 19"
Hg
C300-T2 3 25.5 23.8 23.8
C500-T2 5 24 38.7 38.7
C750L-T2 7.5 24 52.6 52.6
C750-T2 7.5 22.5 68.2 68.2
Triplex Stack Mounted
Model Number
H.P.
Each
Maximum
Vacuum Level
("Hg)
SCFM @ 19"
Hg Each
Design Capacity
SCFM @ 19"
Hg
C500-ST3 5 24 38.7 77.4
C750L-T2 7.5 24 52.6 105.2
C750-T2 7.5 22.5 68.2 136.4
C1500-ST3 15 22.5 120.1 222.6
Quadruplex Stack Mounted
Model Number H.P. Each
Maximum Vacuum
Level ("Hg)
SCFM @ 19" Hg
Each
Design Capacity
SCFM @ 19" Hg
C500-ST4 5 24 38.7 116.1
C750L-T2 7.5 24 52.6 157.8
C750-T2 7.5 22.5 68.2 204.6
C1500-ST4 15 22.5 120.1 333.9
Low cost of ownership
Clean air
NFPA 99 compliant
Warranty: 30/24
For ordering information, please call 1-800-448-0770 or 847-855-0500
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
18
Dr y Runni ng Cont ac t l ess Rot ar y Cl aw
Medi c al Vac uum Syst em
Dry Running Contactless Rotary Claw Medical Vacuum System
The Ohio Medical Corporation

Dry Running Rotary Claw Medical Vacuum System shall be fully NFPA 99 compliant for use in
medical vacuum and WAGD applications. The unit will consist of electric motor driven pumps, vacuum receiver, electrical control
system, and interconnection piping and wiring. The components shall be modularly assembled to accommodate most existing
doorways and designed for serviceability. The packaged unit shall be factory tested prior to shipment and warranted for a period of
(30) months from date of shipment or (24) months from date of start-up.
Vacuum Pump Module
The dry running contactless rotary claw vacuum pump shall be single stage, and air cooled. The pump design utilizes two ductile
iron claw type rotors rotating in opposite direction inside a cast iron housing. The timing of the rotors shall be maintained by a
precision gear set which is separated from the oil free pumping chamber by a combination of lip seals. The vacuum pump assembly
shall also include a 5 micron particulate inlet vacuum lter, vacuum regulator/relief valve, and exhaust silencer. The assembly shall
also include check valves, pump isolation valves, a source isolation valve and exible connectors.
Vacuum Receiver
The vacuum receiver shall be constructed to ASME standards, rated for full vacuum, and include a valved by-pass, manual drain
valve, vacuum gauge and the National Board label.
System Controls
The Underwriters Laboratories Listed electrical motor control system shall be of a fuseless design in a NEMA 12 enclosure. The
Continuous On Demand feature shall stop the operation of the motors during periods of low or no demand. The controls include
individual self protected combination motor controls with short circuit, single phase and thermal overload protection, individual 120
volt control circuit transformers with fuseless primary and secondary protection, pressure sensors, and an electronic controller to
automatically switch the operating sequence of the vacuum pumps. The cabinet door shall have an HMI (Human Machine Interface)
system status display to include system pressure, pump operation, accumulated time, maintenance interval, fault conditions and
silence button; lighted Hand-Off-Automatic selector switches and safety disconnect operating handles.
All required local alarm functions shall be integrated into the packaged system. The circuitry shall be designed so the audible signal
can be silenced and the visual indicator will remain until the fault has been cleared and the reset button actuated. Local alarm
functions shall be provided for reserve pump in use (lag alarm). Dry contacts are furnished for remote (master) alarm signals.
Accessories
Included for job site installation are inlet and discharge exible connectors, source isolation valve, vibration mounting pads, and
touch-up paint.
www.ohiomedical.com
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
19
Dr y Runni ng Cont ac t l ess Rot ar y Cl aw Medi c al
Vac uum Syst em Sel ec t i on Char t
D
e
s
i
g
n

C
a
p
a
c
i
t
y














S
C
F
M

@

1
9
"

H
g

/
c
a
p
a
c
i
t
y

o
f

e
a
c
h

p
u
m
p
S
y
s
t
e
m

C
o
n


g
u
r
a
t
i
o
n
H
P
T
a
n
k

S
i
z
e

(
g
a
l
l
o
n
s
)
L
e
n
g
t
h
(
A
)
W
i
d
t
h
(
B
)
H
e
i
g
h
t
(
C
)
W
e
i
g
h
t
S
y
s
t
e
m

m
o
d
e
l

N
u
m
b
e
r
23 / 23
Tank Mounted
Duplex
3 120 105 30 68 710 C300B-T2
38.7 / 38.7
Tank Mounted
Duplex
5 120 106 43 76 1930 C500B-T2
52.6 / 52.6
Tank Mounted
Duplex
7.5 120 106 43 76 2060 C750B-T2
77.4 / 38.7
Stack Mounted
Triplex
5 240 114 62 96 3920 C500B-ST3
68.2 / 68.2
Tank Mounted
Duplex
10 120 120 48 80 2600 C1000B-T2
116.1 / 38.7
Stack Mounted
Quadruplex
5 204 114 62 96 4990 C500B-ST4
105.2 / 52.6
Stack Mounted
Triplex
7.5 240 114 62 96 4110 C750B-ST3
136.4 / 52.6
Stack Mounted
Triplex
10 240 114 62 96 4600 C1000B-ST3
157.8 / 52.6
Stack Mounted
Quadruplex
7.5 240 114 62 96 5250 C750B-ST4
222.6 / 111.3
Stack Mounted
Triplex
15 240 114 62 96 4640 C1500B-ST3
204.6 / 68.2
Stack Mounted
Quadruplex
10 240 114 62 96 5800 C1000B-ST4
333.9 / 111.3
Stack Mounted
Quadruplex
15 240 114 62 96 5930 C1500B-ST4
For ordering information, please call 1-800-448-0770 or 847-855-0500
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
20
Oasi s Vac uum Pumps
Pr i nc i pl e of Oper at i on
Figure 1
Trapped
Air
Sealant Ring
(oil or water)
Casing
Impeller
Air Inlet Air and Sealant
Discharge
The Oasis is a total recirculation liquid ring vacuum pump module. All components are contained within a 304 Stainless
Steel enclosure.
The above gure shows a cross section of a liquid ring vacuum pump. The impeller is cylindrical in shape with several
ns and is located eccentrically in the impeller housing, which is partially lled with water. A port plate with suction and
discharge openings is located on each side of the impeller cell.
As the impeller rotates, the centrifugal force pushes the sealing liquid outward where it forms a liquid ring which follows
the contour of the impeller housing. At the start, the impeller cell is completely lled with the sealing liquid.
As the impeller advances, the liquid in the cell recedes and creates a vacuum in the empty space, which draws in the
air or gas through the suction port. Continued rotation increases the gas volume in the impeller cell until the end of
the suction cycle.
From that , the liquid is gradually forced back into the impeller cell, compressing the gas to near atmosphere, until the
compressed gas is expelled through the discharge port together with a portion of the sealing liquid.
The complete cycle can be compared with a piston in a cylinder where the piston is the liquid ring and the impeller
cell is the cylinder. The heat generated by the pump during the compression cycle is dissipated into the sealing liquid.
The discharge is piped to a separation and recirculation tank. The air is discharged and the seal water is piped to an
on board, high efciency, fan driven heat exchanger where the seal water is cooled and recirculated via a return line
back to use in the pump. Any water lost due to evaporation is automatically added.
Figure 2
Oasis Model H.P. SCFM @ 19Hg Capacity (ACFM) Make-Up Water Per Hour (gal)
OL150 1.5 6.6 18 0.04
OL300 3 12.1 33 0.08
OL400 4 17.6 48 0.15
OL500 5 26.5 72 0.20
OL750 7.5 31.6 86 0.22
OL1000 10 53 144 0.38
OL1500 15 69 187 0.53 Estimated water
usage in a typical
medical application
www.ohiomedical.com
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
21
Oasi s Vac uum
Syst ems
NFPA Complaint
U.L. Listed Electrical Control Panel
Water Sealed/Air cooled
Available in simplex, duplex, triplex or quad
Modular Stack Mounted
Medical or Laboratory Applications
Minimal Maintenance, Maximum Durability
No Contacting Parts
Quiet Operation
Low Operating Temperature
Warranty 30/24 Months
Up to 48 hours of operation without make up water.*
* System can run up to 48 hours without need for make up water in emergency
situations based on typical medical vacuum application.
For ordering information, please call 1-800-448-0770 or 847-855-0500
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
22
Oasi s Vac uum
Syst em
Oasis Liquid Ring Water Sealed/Air Cooled Total Recirculating Vacuum System
Ohio Medical Corporation Oasis Liquid Ring, NFPA 99 compliant vacuum system shall consist of electric motor driven
positive displacement non-pulsating liquid ring vacuum pumps, ASME vacuum receiver, electrical control system,
and interconnection piping and wiring. The components shall be modularly assembled to accommodate most existing
doorways and designed for serviceability. The packaged unit shall be factory tested prior to shipment and warranted
for a period of (30) months from date of shipment or (24) months from date of start-up.
Vacuum Pump Module
Each liquid ring vacuum pump shall have a cast iron body, stainless shaft, bronze impeller, and mechanical seals to
be direct driven by a TEFC motor. Pump suction accessories shall include an inlet check valve, isolation valve and
inlet exible connector. Pump discharge accessories shall include a water reservoir constructed of stainless steel. The
recirculation reservoir shall have a level switch and sight glass to maintain proper liquid level.
The recirculation lines shall include a water/air radiator. The total recirculation service liquid system, after initial
priming, the pump shall require no more than 3 gallons of water per hour.
Vacuum Receiver
The vacuum receiver shall be constructed to ASME standards, rated for full vacuum, and include a valved by-pass,
manual drain valve, sight glass, vacuum gauge and the National Board label.
System Controls
The Underwriters Laboratories Listed electrical motor control system shall be of a fuseless design in a NEMA 12
enclosure. The Continuous On Demand feature shall stop the operation of the motors during periods of low or no
demand. The controls include individual self protected combination motor controls with short circuit, single phase
and thermal overload protection, individual 120 volt control circuit transformers with fuseless primary and secondary
protection, pressure sensors, and an electronic controller to automatically switch the operating sequence of the
vacuum pumps. The cabinet door shall have an HMI (Human Machine Interface) system status display to include
system pressure, pump operation, accumulated time, maintenance interval, fault conditions and silence button; lighted
Hand-Off-Automatic selector switches and safety disconnect operating handles.
All required local alarm functions shall be integrated into the packaged system. The circuitry shall be designed so the
audible signal can be silenced and the visual indicator will remain until the fault has been cleared and the reset button
actuated. Local alarm functions shall be provided for reserve pump in use (lag alarm). Dry contacts are furnished for
remote (master) alarm signals.
Accessories
Accessories included for job site installation are inlet and discharge exible connectors, source isolation valve, vibration
mounting pads and touch-up paint.
www.ohiomedical.com
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
23
Oasi s Vac uum Syst ems
Sel ec t i on Char t
D
e
s
i
g
n

C
a
p
a
c
i
t
y

S
C
F
M

@

1
9
"

H
g
P
e
r

P
u
m
p

S
C
F
M

@

1
9
"
H
g
S
y
s
t
e
m

C
o
n


g
u
r
a
t
i
o
n
H
P
T
a
n
k

S
i
z
e

(
g
a
l
l
o
n
s
)
L
e
n
g
t
h

"
A
"
W
i
d
t
h

"
B
"
H
e
i
g
h
t

"
C
"
G
P
H

o
f

W
a
t
e
r

P
e
r

P
u
m
p
*
S
y
s
t
e
m

M
o
d
e
l

N
u
m
b
e
r
12.1 12.1
Stack Mounted
Duplex
3 120 79" 70" 92" 0.08 LO300B-ST2
17.6 17.6
Stack Mounted
Duplex
5 120 79" 70" 92" 0.2 LO500B-ST2
31.6 31.6
Stack Mounted
Duplex
7.5 120 79" 70" 92" 0.22 LO750B-ST2
35.2 17.6
Stack Mounted
Triplex
5 240 116" 70" 92" 0.2 LO500B-ST3
52.8 17.6
Stack Mounted
Quarduplex
5 240 154" 70" 92" 0.2 LO500B-ST4
53 53
Stack Mounted
Duplex
10 120 79" 70" 92" 0.38 LO1000B-ST2
63.2 31.6
Stack Mounted
Triplex
7.5 240 116" 70" 92" 0.22 LO750B-ST3
69 69
Stack Mounted
Duplex
15 120 79" 70" 92" 0.53 LO1500B-ST2
94.8 31.6
Stack Mounted
Quarduplex
7.5 240 154" 70" 92" 0.22 LO750B-ST4
106 53
Stack Mounted
Triplex
10 240 116" 70" 92" 0.38 LO1000B-ST3
138 69
Stack Mounted
Triplex
15 240 116" 70" 92" 0.53 LO1500B-ST3
159 53
Stack Mounted
Quarduplex
10 240 154" 70" 92" 0.38 LO1000B-ST4
207 69
Stack Mounted
Quarduplex
15 240 154" 70" 92" 0.53 LO1500B-ST4
For ordering information, please call 1-800-448-0770 or 847-855-0500
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
24
Wat er Mi ser
Li qui d Ri ng Vac uum Syst ems
NFPA 99 compliant
U.L. Listed Electrical Control panel
Water Sealed
Available in Simplex, Duplex, Triplex or Quad
Congurations: Tank, Base or Stack Mounted
Medical or Laboratory Applications
Minimal maintenance, Maximum Durability
No contacting parts in pumping chamber
Cool running and long wear
Optional Heat Exchanger shown below for optimal seal uid
conservation
Warranty: 30/24 Months
s
Model No. HP
SCFM @ 19" Hg
(each pump)
LR-3-1 3 11.7
LR-5-1 5 21.6
LR-75-1 7.5 33.7
LR-10-1 10 47.5
LR-15-1 15 75.5
LR-20-1 20 100.7
LR-30-1 30 137.4
LR-40-1 40 199.6
Design capacity values should correspond with the
Calculated Peak Demand. This will allow for the
reserve pump as is required per NFPA for level 1
installations.
www.ohiomedical.com
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
25
Li qui d Ri ng Cont i nuous On Demand
Vac uum Syst em, Par t i al Rec i r c ul at i on
Liquid Ring Water Sealed Partial Recirculating Vacuum System
Ohio Medical Corporation

partial recirculating, NFPA 99 compliant vacuum system shall consist of electric motor driven positive
displacement non-pulsating liquid ring vacuum pumps, ASME vacuum receiver, electrical control system, and interconnection
piping and wiring. The components shall be modularly assembled to accommodate most existing doorways and designed for
serviceability. The packaged unit shall be factory tested prior to shipment and warranted for a period of (30) months from date of
shipment or (24) months from date of start-up.
Vacuum Pump Module
Each liquid ring vacuum pump shall have cast iron body, stainless steel impeller, and mechanical seals to be direct driven by
a TEFC motor. Pump suction accessories shall include an inlet check valve, isolation valve and inlet exible connector. Pump
discharge accessories shall include a discharge air/water separator constructed of high density polyethylene that also serves as
the partial recirculation reservoir tank. The recirculation reservoir shall have a two station, combination level sensor/sight glass to
maintain proper liquid level. The pump discharge separator vents shall be manifolded, and a exible connector shall be provided
for job site installation to facilitate connection of vent piping to the exterior of the building.
Recirculation liquid control will be interactive through a temperature sensor, with a range of 75F to 90F. Each recirculation liquid
line shall include a strainer and isolation/ow regulation valve.
The make-up water lines shall include a strainer, regulating valve, solenoid valve with manual priming valve and exible connector.
Each pump will require 59
o
F cooling water.
Vacuum Receiver
The vacuum receiver shall be constructed to ASME standards, rated for full vacuum, and include a valved by-pass, manual drain
valve, sight glass, vacuum gauge and the National Board label.
System Controls
The Underwriters Laboratories Listed electrical motor control system shall be of a fuseless design in a NEMA 12 enclosure. The
Continuous On Demand feature shall stop the operation of the motors during periods of low or no demand. The controls include
individual self protected combination motor controls with short circuit, single phase and thermal overload protection, individual 120
volt control circuit transformers with fuseless primary and secondary protection, pressure sensors, and an electronic controller to
automatically switch the operating sequence of the vacuum pumps. The cabinet door shall have an HMI (Human Machine Interface)
system status display to include system pressure, pump operation, accumulated time, maintenance interval, fault conditions and
silence button; lighted Hand-Off-Automatic selector switches and safety disconnect operating handles.
All required local alarm functions shall be integrated into the packaged system. The circuitry shall be designed so the audible signal
can be silenced and the visual indicator will remain until the fault has been cleared and the reset button actuated. Local alarm
functions shall be provided for reserve pump in use (lag alarm). Dry contacts are furnished for remote (master) alarm signals.
Accessories
Accessories included for job site installation are inlet and discharge exible connectors, source isolation valve, vibration mounting
pads and touch-up paint.
For ordering information, please call 1-800-448-0770 or 847-855-0500
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
26
Li qui d Ri ng Cont i nuous On Demand Vac uum
Syst ems Sel ec t i on Char t , Par t i al Rec i r c ul at i on
D
e
s
i
g
n

C
a
p
a
c
i
t
y













S
C
F
M

@

1
9
"

H
g
/








c
a
p
a
c
i
t
y

o
f

e
a
c
h

p
u
m
p
D
e
s
i
g
n

C
a
p
a
c
i
t
y














A
C
F
M

@

2
5
"

H
g
/








c
a
p
a
c
i
t
y

o
f

e
a
c
h

p
u
m
p
S
y
s
t
e
m

C
o
n


g
u
r
a
t
i
o
n
H
P
T
a
n
k

S
i
z
e















(
g
a
l
l
o
n
s
)
A
v
e
r
a
g
e

m
a
k
e
-
u
p

w
a
t
e
r

r
e
q
u
i
r
e
d

@

1
9
"

H
g

(
G
P
M
)
L
e
n
g
t
h

(
A
)
W
i
d
t
h

(
B
)
H
e
i
g
h
t

(

C
)
W
e
i
g
h
t
S
y
s
t
e
m

M
o
d
e
l

N
u
m
b
e
r
11.7 / 11.7 27.5 / 27.5
Tank Mounted
Duplex
3 120 0.7 88 40 86 1370 LP3B-T2
21.6 / 21.6 69.5 / 69.5
Tank Mounted
Duplex
5 120 1.1 92 46 70 1600 LP5B-T2
33.7 / 33.7 85 / 85
Tank Mounted
Duplex
7.5 200 1.6 92 46 76 2050 LP75B-T2
47.5 / 47.5 122 / 122
Tank Mounted
Duplex
10 200 2.0 92 56 76 2340 LP10B-T2
75.5 / 75.5 200 / 200
Tank Mounted
Duplex
15 200 2 92 56 76 2630 LP15B-T2
95 / 47.5 244 / 122
Stack Mounted
Triplex
10 240 2.0 114 64 94 3940 LP10B-ST3
100.7 / 100.7 248 / 248
Base Mounted
Duplex
20 240 3.3 114 64 86 3300 LP20B-B2
101.1 / 33.7 255 / 85
Stack Mounted
Quadruplex
7.5 240 1.6 114 64 92 4200 LP75B-ST4
137.4 / 137.4 360 / 360
Base Mounted
Duplex
30 240 4.9 130 70 100 4300 LP30B-B2
142.6 / 47.5 366 / 122
Stack Mounted
Quadruplex
10 240 2.0 114 64 92 4900 LP10B-ST4
151 / 75.5 400 / 200
Stack Mounted
Triplex
15 240 2.9 114 64 94 4370 LP15B-ST3
201.4 / 100.7 496 / 248
Base Mounted
Triplex
20 240 3.3 150 64 86 4600 LP20B-B3
226.5 / 75.5 600 / 200
Stack Mounted
Quadruplex
15 240 2.9 114 64 94 5500 LP15B-ST4
274.8 / 137.4 720 / 360
Base Mounted
Triplex
30 240 4.9 176 70 100 6260 LP30B-B3
www.ohiomedical.com
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
27
Sc r ol l Compr essed Ai r
Syst ems
Ohio Medical Scroll Systems features a dedicated pump to motor design. This ensures optimal design safety
and energy efciency by cycling on fewer motors during low demand in lieu of starting multi pump modules.
Quiet
Low vibration
10,000 hour maintenance interval for tip seals
5,000 hour maintenance interval for shaft bearing
Warranty: 30/24
Base Mounted Modules (with motor)
Part
Number
Description H.P. RPM
SCFM @
100 PSIG
SCFM @
140 PSIG
Max
Pressure
Voltage
230072 Scroll Module Low Pressure 3 2200 8.5 - 115 208/230/460/3/60
230073 Scroll Module Low Pressure 5 3150 14.5 - 115 208/230/460/3/60
230082 Scroll Module High Pressure 3 2000 - 7.2 145 208/230/460/3/60
230083 Scroll Module High Pressure 5 2900 - 12 145 208/230/460/3/60
Bare Compressor Air Ends (without motor)
Part
Number
Description H.P.
Max
Pressure
230074 Scroll Air End Low Pressure 3 or 5 115
230075 Scroll Air End High Pressure 3 or 5 145
For ordering information, please call 1-800-448-0770 or 847-855-0500
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
28
Oi l -Less Rot ar y Sc r ol l Ai r Syst ems w i t h
Desi c c ant Dr yer s
Oil-Less Rotary Scroll Compressed Air System with Desiccant Dryers
Ohio Medical Corporation

Oil-less Rotary Scroll Compressed Air System shall be fully NFPA 99 compliant. The
unit will consist of electric motor driven single stage oil-less rotary scroll air compressor modules, electrical control
system, ASME air receiver, air cooled aftercoolers with individual drains, inline intake air lters, air treatment module,
interconnection copper/brass piping, and wiring. The components shall be modularly assembled to accommodate
most existing doorways and designed for serviceability. The packaged unit shall be factory tested prior to shipment and
warranted for a period of (30) months from date of shipment or (24) months from date of start-up.
Air Compressor Modules
The compressors shall have PTFE composite tip seals and be rated for 10,000 hours of operation. Compressors
bearings shall be external to the compression chamber and shall be serviceable. Compressors with bearings that
are not accessible for service and have limited life span shall not be accepted. Each belt driven compressor module
inlet shall include an inline lter with particle retention of 10 microns, inlet isolation valve and ex connector. The
compressor discharge piping includes air cooled after cooler with moisture separator and automatic drain, isolation
valves, ASME safety pressure relief valve, thermal malfunction shutdown devise.
Air Receiver
The corrosion resistant coated receiver is equipped with an ASME safety pressure relief valve, sight glass pressure
gauge, automatic drain, three-valve by-pass and source isolation valve.
Air Treatment Module
The air treatment module shall include dual dryers, dual ltration system, and a CO and Dew Point monitoring devise
with local audible and visual signals and dry contacts for remote monitoring. The components shall be mounted on a
common base with interconnecting copper/brass piping and upstream and downstream isolation valves. The isolation
valves shall allow either set of components to be serviced without shutting down the system.
The dryer system shall be duplexed and provide a pressure dew point of 10oF. Dryers shall be heatless desiccant
design selected and sized to provide for the peak-calculated demand. The desiccant dryer shall be equipped with a
Dewpoint Dependent Switching feature to minimize the need for purge air.
The dual ltration system is designed to remove liquids and particulate matter, and consists of one micron coalescing
lters with differential pressure indicators and automatic drain, air line pressure regulators with gauges, nal pressure
relief valve, and sampling valve.
System Controls
The Underwriters Laboratories listed electrical motor control system shall be of a fuseless design in a NEMA 12
enclosure. The Continuous On Demand feature shall stop the operation of the motors during periods of low or no
demand. The controls include individual self protected combination motor controls with short circuit, single phase
and thermal overload protection, individual 120 volt control circuit transformers with fuseless primary and secondary
protection, pressure sensors, temperature switches with reset button, and an electronic controller to automatically
change the operating sequence of the compressors. The cabinet door shall have an HMI (Human Machine Interface)
system status display to include system pressure, dewpoint, pump operation, accumulated time, maintenance interval,
fault conditions, and silence button; lighted Hand-Off-Automatic selector switches and safety disconnect operating
handles. All required local alarm functions shall be integrated into the packaged system.
Accessories
Accessories included for job site installation are inlet and discharge exible connectors, vibration mounting pads, and
source isolation valve.
www.ohiomedical.com
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
29
Oi l -Less Rot ar y Sc r ol l Ai r Syst ems w i t h
Desi c c ant Dr yer s Sel ec t i on Char t
D
e
s
i
g
n

C
a
p
a
c
i
t
y

S
C
F
M

@

5
0

P
S
I
G
S
C
F
M

p
e
r

m
o
d
u
l
e

@

5
0

P
S
I
G
S
y
s
t
e
m

C
o
n


g
u
r
a
t
i
o
n
H
P

p
e
r

p
u
m
p
N
o
.

o
f

P
u
m
p
l
e
n
g
t
h

(
A
)
W
i
d
t
h

(
B
)
H
e
i
g
h
t
(

C
)

S
y
s
t
e
m

m
o
d
e
l

N
u
m
b
e
r
9.8 9.8
Tank Mounted
Duplex
3 2 74 32 71 AS300-T2-DD
16.3 16.3
Tank Mounted
Duplex
5 2 74 32 71 AS500-T2-DD
32.6 16.3
Tank Mounted
Triplex
5 3 84 64 72 AS500-T3-DD
48.9 16.3
Stack Mounted
Quadruplex
5 4 84 64 72 AS500-ST4-DD
65.2 16.3
Stack Mounted
Pentaplex
5 5 119 64 72 AS500-ST5-DD
81.5 16.3
Stack Mounted
Hexaplex
5 6 116 64 72 AS500-ST6-DD
97.8 16.3
Stack Mounted
Heptaplex
5 7 119 64 72 AS500-ST7-DD
114.1 16.3
Stack Mounted
Octoplex
5 8 119 64 72 AS500-ST8-DD
130.4 16.3
Stack Mounted
Enneaplex
5 9 155 64 72 AS500-ST9-DD
146.7 16.3
Stack Mounted
Decaplex
5 10 155 64 72 AS500-ST10-DD
163 16.3
Stack Mounted
Hendecaplex
5 11 155 64 72 AS500-ST11-DD
179.3 16.3
Stack Mounted
Dodecaplex
5 12 155 64 72 AS500-ST12-DD
For ordering information, please call 1-800-448-0770 or 847-855-0500
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
30
Oi l -Less Rec i pr oc at i ng Pi st on
Ai r Compr essor Syst ems
Oil-Less Air Compressors
NFPA 99 compliant
U.L. Listed Electrical Control panel
Available in Simplex, Duplex, Triplex and Quadruplex
Congurations: Tank, Base or Stack Mounted
Medical and Laboratory Congurations
Warranty: 30/24 months
Model # HP
SCFM @
50 PSIG
SCFM @
100 PSIG
A200 2 8.2 5.1
A300 3 12 10.1
A500 5 20.9 17
A750 7.5 32 26.8
A1000 10 41.3 36
A1500 15 58 51
A2000 20 71.5 62
A2500 25 109 94
A3000 30 118 102
www.ohiomedical.com
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
31
Rec i pr oc at i ng Oi l -Less Cont i nuous On Demand
Compr essed Ai r Syst em w i t h Desi c c ant Dr yer s
Oil-Less Reciprocating Continuous on Demand Compressed Air System with Desiccant Dryers
Ohio Medical Corporation reciprocating NFPA 99 compliant air compressor system shall consist of electric motor
driven oil-less air compressors, an Underwriters Laboratories listed electrical control mounted in a NEMA 12 enclosure,
ASME air receiver, air cooled aftercoolers with individual drains, inline intake air lters and an air treatment module.
The components shall be modularly assembled to accommodate most existing doorways. The system shall include
interconnecting copper/brass piping and wiring to provide a functional operating package with applicable electrical
and plumbing connections at the installation site. The packaged unit shall be factory tested prior to shipment and
warranted for a period of (30) months from date of shipment or (24) months from date of start-up.
Compressor Modules
The compressors shall consist of a crankcase, connecting rods, integral counterweights for smooth operation, and
cylinders and heads designed for efcient heat dissipation. Piston rings shall be provided to reduce wear and have
a life span of 10,000 hours. Each compressor cylinder is protected by a temperature switch, which will stop the drive
motor and provide an alarm signal in the event of abnormal discharge air temperature. Each belt driven compressor
module shall include an inline lter with a particle retention of 10 microns, inlet isolation valve discharge isolation valve,
and ASME safety pressure relief valve.
Air Receiver
The corrosion resistant coated receiver is equipped with an ASME safety pressure relief valve, sight glass pressure
gauge, automatic drain, three-valve by-pass and source isolation valve.
Air Treatment Module
The air treatment module shall include dual dryers, dual ltration system, and a CO and Dew Point monitoring devise
with local audible and visual signals and dry contacts for remote monitoring. The components shall be mounted on a
common base with interconnecting copper/brass piping and upstream and downstream isolation valves. The isolation
valves shall allow either set of components to be serviced without shutting down the system.
The dryer system shall be duplexed and provide a pressure dew point of 10oF. Dryers shall be heatless desiccant
design selected and sized to provide for the peak-calculated demand. The desiccant dryer shall be equipped with a
Dewpoint Dependent Switching feature to minimize the need for purge air.
The dual ltration system is designed to remove liquids and particulate matter, and consists of one micron coalescing
lters with differential pressure indicators and automatic drain, air line pressure regulators with gauges, nal pressure
relief valve, and sampling valve.
System Controls
The Underwriters Laboratories listed electrical motor control system shall be of a fuseless design in a NEMA 12
enclosure. The Continuous On Demand feature shall stop the operation of the motors during periods of low or no
demand. The controls include individual self protected combination motor controls with short circuit, single phase
and thermal overload protection, individual 120 volt control circuit transformers with fuseless primary and secondary
protection, pressure sensors, temperature switches with reset button, and an electronic controller to automatically
change the operating sequence of the compressors. The cabinet door shall have an HMI (Human Machine Interface)
system status display to include system pressure, dewpoint, pump operation, accumulated time, maintenance interval,
fault conditions, and silence button; lighted Hand-Off-Automatic selector switches and safety disconnect operating
handles. All required local alarm functions shall be integrated into the packaged system.
Accessories
Included for job site installation are inlet and discharge exible connectors, vibration mounting pads, source isolation
valve, and touch-up paint.
For ordering information, please call 1-800-448-0770 or 847-855-0500
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
32
Rec i pr oc at i ng Oi l -Less Cont i nuous On Demand Compr essed
Ai r Syst ems w /Desi c c ant Dr yer s Sel ec t i on Char t
D
e
s
i
g
n

C
a
p
a
c
-
i
t
y


S
C
F
M

@

5
0

P
S
I
G

e
a
c
h
C
a
p
a
c
i
t
y

o
f

e
a
c
h

p
u
m
p

(
o
n
e

i
n

r
e
s
e
r
v
e

o
n

m
u
l
t
i
l
p
e
x

s
y
s
t
e
m
s
)
S
y
s
t
e
m

C
o
n


g
u
r
a
t
i
o
n
H
P
T
a
n
k

S
i
z
e

(
g
a
l
l
o
n
s
)
Q
t
y

o
f

M
o
t
o
r
s
l
e
n
g
t
h
(
A
)
W
i
d
t
h

(
B
)
H
e
i
g
h
t

(

C
)
W
e
i
g
h
t
S
y
s
t
e
m

m
o
d
e
l

N
u
m
b
e
r
8.2 8.2
Tank Mounted
Duplex
2 120 2 78 47 69 1900 A200B-T2-DD
12 12
Tank Mounted
Duplex
3 120 2 78 47 69 2000 A300B-T2-DD
20.9 20.9
Tank Mounted
Duplex
5 120 2 80 50 77 2300 A500B-T2-DD
32 32
Tank Mounted
Duplex
7.5 120 2 81 52 77 3000 A750B-T2-DD
41.3 41.3
Tank Mounted
Duplex
10 120 2 67.5 105 85 3900 A1000B-T2-DD
58 58
Stack Mounted
Duplex
15 120 2 69.5 126 92 5400 A1500B-ST2-DD
64 32
Stack Mounted
Triplex
7.5 120 3 107.5 112 85 5100 A750B-ST3-DD
71.5 71.5
Stack Mounted
Duplex
20 120 2 69.5 126 92 5900 A2000B-ST2-DD
82.6 41.3
Stack Mounted
Triplex
10 120 3 107.5 112 85 5350 A1000B-ST3-DD
96 32
Stack Mounted
Quadruplex
7.5 240 4 116 116 85 7000 A750B-ST4-DD
109 109
Base Mounted
Duplex
25 120 2 139.5 125 80 6300 A2500B-B2-DD
116 58
Stack Mounted
Triplex
15 240 3 120 128 90 7600 A1500B-ST3-DD
118 118
Base Mounted
Duplex
30 120 2 139.5 125 80 6300 A3000B-B2-DD
123.9 41.3
Stack Mounted
Quadruplex
10 240 4 116 116 85 7350 A1000B-ST4-DD
143 71.5
Stack Mounted
Triplex
20 240 3 120 132 90 8200 A2000B-ST3-DD
174 58
Stack Mounted
Quadruplex
15 240 4 120 145 90 10600 A1500B-ST4-DD
214.5 71.5
Stack Mounted
Quadruplex
20 240 4 120 145 90 11500 A2000B-ST4-DD
218 109
Base Mounted
Triplex
25 240 3 204.5 142 95 9000 A2500B-B3-DD
236 118
Base Mounted
Triplex
30 240 3 204.5 142 95 9000 A3000B-B3-DD
www.ohiomedical.com
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
33
Si mpl ex
Ai r Syst em
Oil-Less Reciprocating Simplex Compressed Air System
Ohio Medical Corporation air compressor system shall consist of electric motor driven oil-less air compressors, an
Underwriters Laboratories listed electrical control mounted in a NEMA 12 enclosure, ASME air receiver, air cooled
aftercoolers with individual drains, inline intake air lters and an air treatment module. The components shall be
modularly assembled to accommodate most existing doorways. The system shall include interconnecting copper/
brass piping and wiring to provide a functional operating package with applicable electrical and plumbing connections
at the installation site. The packaged unit shall be factory tested prior to shipment and warranted for a period of (30)
months from date of shipment or (24) months from date of start-up.
Compressor Modules
The compressors shall consist of a crankcase, connecting rods, integral counterweights for smooth operation, and
cylinders and heads designed for efcient heat dissipation. Piston rings shall be provided to reduce wear and have
a life span of 10,000 hours. Each compressor cylinder is protected by a temperature switch, which will stop the drive
motor and provide an alarm signal in the event of abnormal discharge air temperature. Each belt driven compressor
module shall include an inline lter with a particle retention of 10 microns, inlet isolation valve discharge isolation valve,
and ASME safety pressure relief valve.
Air Receiver
The corrosion resistant coated receiver is equipped with an ASME safety pressure relief valve, sight glass pressure
gauge, automatic drain, three-valve by-pass and source isolation valve.
System Controls
The Underwriters Laboratories Listed simplex electrical motor control system shall be of a fuseless design in a NEMA
12 (dustproof) enclosure. The Continuous On Demand feature shall stop the operation of the motor during periods
of low or no demand. The controls include a individual self protected combination motor control with short circuit,
single phase and thermal overload protection, 120 volt control circuit transformer with fuseless primary and secondary
protection, pressure sensor and a temperature switch with reset button. The cabinet door shall have an HMI (Human
Machine Interface) system status display to include accumulated time and fault conditions; lighted Hand-Off-Automatic
selector switch and safety disconnect operating handle.

All required local alarm functions shall be integrated into the packaged system. The circuitry shall be designed so the
visual indicator will remain until the fault has been cleared and the reset button actuated. Local alarm functions shall
be provided for high air discharge temperature and thermal overload.
Ship Separate Items & Accessories (Field Installed)
Ship Separate Items: Non-cycling refrigerated type air dryer, coalescing 1 micron lter, line pressure regulator with
gauge.
Accessories: Inlet and discharge exible connectors, vibration mounting pads, source isolation valve and touch-up
paint.
For ordering information, please call 1-800-448-0770 or 847-855-0500
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
34
Si mpl ex
Ai r Sel ec t i on Mat r i x
S
C
F
M

@

5
0

P
S
I
G
S
y
s
t
e
m

C
o
n


g
u
r
a
t
i
o
n
H
P
T
a
n
k

S
i
z
e

(
g
a
l
l
o
n
s
)
l
e
n
g
t
h

(
A
)
W
i
d
t
h

(
B
)
H
e
i
g
h
t

(

C
)
W
e
i
g
h
t
S
y
s
t
e
m

m
o
d
e
l

N
u
m
b
e
r
6.7 Oil-Less Scroll
Tank Mounted
Simplex
2 80 75 32 54 745 AS200-T1
8.2
Oil-Less
Reciprocating
Tank Mounted
Simplex
2 80 75 32 54 745 A200B-T1
9.9 Oil-Less Scroll
Tank Mounted
Simplex
3 80 75 32 54 770 AS300-T1
12
Oil-Less
Reciprocating
Tank Mounted
Simplex
3 80 75 32 54 770 A300B-T1
16.4 Oil-Less Scroll
Tank Mounted
Simplex
5 80 75 32 54 962 AS500-T1
20.9
Oil-Less
Reciprocating
Tank Mounted
Simplex
5 80 75 32 54 962 A500B-T1
32
Oil-Less
Reciprocating
Tank Mounted
Simplex
7.5 80 75 32 54 1045 A750B-T1
41.3
Oil-Less
Reciprocating
Tank Mounted
Simplex
10 80 75 32 54 1156 A1000B-T1
www.ohiomedical.com
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
35
I nst r ument Ai r
Syst em
High Pressure Reciprocating Compressed Instrument Air System
with Desiccant Dryers
Ohio Medical Corporation Instrument Air NFPA 99 compliant compressor system shall consist of electric motor driven
pressure lubricated air compressors, an Underwriters Laboratories listed electrical control mounted in a NEMA 12
enclosure, ASME air receiver, air cooled aftercoolers with individual drains, inline intake air lters and an air treatment
module. The components shall be modularly assembled to accommodate most existing doorways. The system shall
include interconnecting copper/brass piping and wiring to provide a functional operating package with applicable
electrical and plumbing connections at the installation site. The packaged unit shall be factory tested prior to shipment
and warranted for a period of (30) months from date of shipment or (24) months from date of start-up.
Compressor Modules
Each compressor is to be a Quincy Compressor model QR-25 pressure lubricated reciprocating design, manufactured
in the United States of America.
Each pressure lubricated compressor will be v belt driven, single-acting, two-stage, air-cooled reciprocating type.
Cylinders will be cast iron, with deep cooling ns. Valve assemblies can be removed individually without removing the
cast iron head and inlet or discharge piping. The compressor shall have hardened and lapped disc-type valves for
continuous service. There will be two pieces to the connecting rods with cast in tubes for positive oil distribution to the
piston pins. Low pressure pistons will be aluminum with three automotive type compression rings and one automotive
type oil ring. The high pressure pistons will be cast iron with three automotive type compression rings and one
automotive oil ring. Connecting rod bearings shall be steel backed Babbitt lined automotive type. Crankshafts shall
have integral counterweights and be rie drilled for pressure lubrication to ensure positive oil distribution. Adjustable
anti-friction tapered roller bearings shall be mounted on each end of the crankshaft to take both radial and thrust loads.
Compressor ywheels shall be dynamically balanced and of the fan type. Finned intercoolers shall be an integral part
of all two-stage units, and located in the direct blast of air from the fan-type ywheel. Compressors shall be equipped
with a crankshaft driven oil pump located in the bearing carrier for easy removal and inspection. There is also an oil
lter located on the bearing carrier as well as a pre-lter screen in the crankcase to help keep the oil clean. Low oil
protection shall be a standard feature of the compressor, along with loadless starting.
Air Receiver
The ASME corrosion resistant coated air receiver is equipped with sight glass, pressure gauge, manual and electronic
drains, three-valve by-pass and source isolation valve. The receiver shall be rated for 350 psig.
Air Treatment Module
Desiccant air dryers shall be modular adsorption design. Each desiccant dryer shall be capable - 40F pressure
dew point at 100 psig. The desiccant dryer shall be twin tower design with switching controls to change from one
tower to the other during regeneration. The dryer will come standard with energy management feature to control and
delay purge air requirement. Carbon monoxide/dew point monitor shall be combination type and shall be in NEMA
12 enclosure. Desiccant dryers will have integral coalescing and particulate lters with activated charcoal lters with
bypass piping. Coalescing lter will be rated at 98% efciency at 0.01 micron.
System Controls
The air treatment module shall include a microprocessor based, UL/CUL-508 listed and NFPA-99 compliant alternating
control panel with a NEMA type 12 enclosure. The control panel is equipped with through the door, rotary type fused
disconnect switches and a through the door, rotary type, non-fused disconnect switch which controls incoming power
to the transformer that powers the CO/Dew Point Monitor. There are independent controls for each compressor with
primary and secondary fused, 120VAC control circuit transformers. The panel will include an ON/OFF selector switch,
elapsed time meter, two pressure switches (lead and lag), two motor starters. Each display board has a POWER
ON L.E.D., COMPRESSOR RUNNING L.E.D., 2 mode of operation L.E.D.s (Continuous or Start/Stop), 2 starting
sequence L.E.D.s (Lead or Auto). Three pressure sensitive membrane style pushbuttons to select the mode of
operation (Continuous or Start/Stop), starting sequence (Lead or Auto) and ALARM CANCEL that silences the audible
alarm horn and deactivates the alarm contacts. Four ashing fault indicators (Overload, Hi Temp, Lag Comp and
System). One fault code display which indicates the type of system fault that has occurred. Audible alarm horn that
sounds when one of the four fault conditions has occurred.
For ordering information, please call 1-800-448-0770 or 847-855-0500
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
36
D
e
s
i
g
n

C
a
p
a
c
i
t
y












(
o
n
e

p
u
m
p

i
n

r
e
s
e
r
v
e
)






















S
C
F
M

@

2
5
0

P
S
I
G

e
a
c
h
C
a
p
a
c
i
t
y

o
f

e
a
c
h

p
u
m
p







@

2
5
0

P
S
I
G
S
y
s
t
e
m

C
o
n


g
u
r
a
t
i
o
n
H
P
T
a
n
k

S
i
z
e

(
g
a
l
l
o
n
s
)
Q
t
y

o
f

M
o
t
o
r
s
l
e
n
g
t
h

(
A
)
W
i
d
t
h

(
B
)
H
e
i
g
h
t

(

C
)
S
y
s
t
e
m

m
o
d
e
l

N
u
m
b
e
r
23.4 23.4
Tank Mounted
Duplex
7.5 120 2 68" 80" 83" AI750B-T2
30.2 30.2
Tank Mounted
Duplex
10 120 2 68" 80" 83" AI1000B-T2
46.8 23.4
Stack Mounted
Triplex
7.5 120 3 106" 80" 83" AI750B-ST3
60.4 30.2
Stack Mounted
Triplex
10 120 3 106" 80" 83" AI1000B-ST3
70.2 23.4
Stack Mounted
Quadruplex
7.5 120 4 106" 80" 83" AI750B -ST4
90.6 30.2
Stack Mounted
Quadruplex
10 120 4 106" 80" 83" AI1000B-ST4
I nst r ument Ai r
Sel ec t i on Mat r i x
www.ohiomedical.com
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
37
Ai r Tr eat ment Modul es
(For Oi l -Less Compr essor s w i t h Desi c c ant Dr yer s)
Ohio Medical compressor control panel provides highly
efcient purge and switching control
Clean, dry and oil free compressed air
compact and lightweight design
Perfect solution for a eld upgrade on older existing air
compressor systems
Model
Number
SCFM @
100 PSIG
Dimensions
(LxWlH)
Replacement Elements
Pre-Filter After Filter
HCTM-B-DDS-6 6 34x16x36 233722 233777
HCTM-B-DDS-10 10 34x16x36 233722 233777
HCTM-B-DDS-24 24 34x20x40 233722 233777
HCTM-B-DDS-32 32 34x30x49 233722 233777
HCTM-B-DDS-42 42 34x20x56 233722 233777
HCTM-B-DDS-53 53 34x20x61 233773 223778
HCTM-B-DDS-65 65 34x20x69 233774 233779
HCTM-B-DDS-88 88 34x20x79 233774 233779
HCTM-B-DDS-106 106 40x30x66 233774 233779
HCTM-B-DDS-130 130 40x30x72 223775 233780
HCTM-B-DDS-176 176 40x30x82 223775 233780
HCTM-B-DDS-240 240 50x40x72 223776 233781
HCTM-B-DDS-360 360 50x40x72 - -
Combination CO / Dewpont Monitor (-112F to +68F)
Model Number Description
233013 CO/Dewpoint Combination Monitor
233015 CO Calibration Kit
2610-001 Humidier (for CO Monitoring)
261922
Gas Specict DISS Air Check Assembly -
1/8" NPT
For ordering information, please call 1-800-448-0770 or 847-855-0500
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
38
Cal c ul at ed Peak Demand f or a Medi c al Sur gi c al Vac uum
Syst em i n a Typi c al Shor t Ter m Gener al Hospi t al
Qty
Medical Gas Vacuum
Outlet Locations
Design Flow in SCFM
(Free Air)
Simultaneous
Use Factor
%
Total
Per
Room
Per Bed Per Outlet
ANESTHETIZING LOCATIONS:
Specialized Surgeries (Open Heart, Organ
Transplant)
4.0 100
Major / Outpatient OR 3.5 100
Cystoscopy / Endoscopy 2.0 100
Delivery Room 1.0 100
Emergency Operating Room 3.0 100
Cardiac Catheterization 1.0 10
Other Anesthetizing Areas (Minor O.R., Induction
Rooms, etc.)
1.0 50
Waste Anesthetizing Gas Evacuation 2.0 100
ACUTE CARE (NON-ANESTHETIZING LOCA-
TIONS):
Post Operative Recovery Room 3.0 50
O.B. Recovery Room 2.0 50
Intensive Care Units (Except Cardiac) 2.0 75
Emergency Room (Trauma, Cardiac) 1.0 100
Cardiac Intensive Care 2.0 50
Neonatal I.C.U. 1.0 50
Special Procedure (X-Ray,Dialysis, Radiology) 1.5 25
Surgical Excision Room 1.0 10
SUBACUTE PATIENT CARE AREAS:
Normal Nursery 1.0 10
Premature Nursery 1.0 20
Respiratory Care 1.5 10
Labor/Birthing 1.0 10
Patient Room (Surgical) 1.5 50
Patient Room (Medical) 1.0 10
Exam & Treatment Rooms 1.0 10
ER (Cast Room, OB/GYN) 1.0 10
Pre-Op Holding 1.0 10
Stress Test EKG & EEG 1.0 10
OTHER AREAS:
Autopsy / Morgue 2.0 20
Central Supply / Sterile 1.5 10
Equipment Repair,Calibration & Teaching 1.5 10
Medical Lab / Pharmacy 1.0 10
Anesthesia/Workroom 1.5 10
Total Estimated Peak Flow (In SCFM @ 19" Hg)
Total Estimated Peak Flow (in SCFM @ 19 Hg)
For the actual worksheet go to www.ohiomedical.com, go to Technical Assistance, select Calculated Peak Demand for a Typical Hospital
(vacuum). This will take you to an excel spreadsheet, which all that you have to do is ll in the quantity and it will calculate for you.
www.ohiomedical.com
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
39
Cal c ul at ed Peak Demand f or a Medi c al Ai r Syst em i n a
Typi c al Shor t Ter m Gener al Hospi t al
Qty
Medical Air
Outlet Locations
Design Flow in SCFM
(Free Air)
Simultane-
ous
Use Factor
%
Totals
Per Unit Per Room Per Bed Per Outlet
ANESTHETIZING LOCATIONS:
Special Surgery (Open Heart, Organ Trans-
plant, Ortho)
0.5 100
Major / Outpatient Surgery 0.5 100
Minor Surgery 0.5 75
Emergency Surgery 0.5 50
Cardiac Catherization 0.5 50
Endoscopy / Cytoscopy 1.0 10
Delivery Room / C-Section 0.5 100
ACUTE CARE LOCATIONS:
Recovery Room/Surgical 2.0 50
Recovery Room - OB 0.5 25
ICU/CCU & Pediatric I.C.U. 2.0 50
Emergency Room (Trauma, Cardiac) 2.0 50
Anesthesia Workroom 1.5 10
Neonatal ICU 1.5 75
Pre-Op Holding 1.5 10
Dialysis Unit, Radiology 0.5 10
Ventilators Adult* 3.5 50
Ventilators Infants* 3.5 50
SUBACUTE CARE LOCATIONS:
Pediatric Croup Tents** 2.0 50
Nursery Full Term & Isolation 0.5 25
ER (Cast Room, OB/GYN) 1.0 10
Patient Rooms (Medical & Surgical) 0.5 10
EEG & EKG 1.0 50
Exam & Treatment 1.0 10
Respiratory Care 1.0 50
Pulmonary Function Lab 1.0 50
Birthing & LDRP 1.0 50
OTHER:
Respiratory Care Workroom 1.5 10
Autopsy / Morgue 1.5 10
Equipment Repair 1.5 10
Total Estimated Peak Flow (In SCFM @ 50 PSIG)
* Important! You must consult with respiratory therapy personnel concerning the maximum number of ventilators that could be used at one time
and the average ow rates for each ventilator. DO NOT CALCULATE your total peak requirements without this information.
** If powered by Medical Air
For the actual worksheet go to www.ohiomedical.com, go to Technical Assistance, select Calculated Peak Demand for a Typical Hospital (air).
This will take you to an excel spreadsheet, which all that you have to do is ll in the quantity and it will calculate for you.
For ordering information, please call 1-800-448-0770 or 847-855-0500
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
40
Mani f ol d
Si zi ng
Medical Air Manifold:
Cylinders Per Bank = 400 Ft
3
x No. of beds x 12 52 234 2
(cubic ft of gas used
per month per bed)
(months
per year)
(# of weeks
per year)
(Cubic ft per
cylinder)
(Cylinders
per bank)
Surgery Center Oxygen Cylinder Manifolds:
Where usage is not known
Cylinders Per Bank = 400 Ft
3
x No. of beds x 12 52 244 2
(cubic ft of gas used
per month per bed)
(months
per year)
(# of weeks
per year)
(Cubic ft per
cylinder)
(Cylinders
per bank)
Hospital Oxygen Cylinder Manifolds:
Where usage is not known
Cylinders Per Bank = 700 Ft
3
x No. of beds x 12 52 244 2
(cubic ft of gas used
per month per bed)
(months
per year)
(# of weeks
per year)
(Cubic ft per
cylinder)
(Cylinders
per bank)
*With a monthly demand of 30,000 ft, bulk O
2
should be used.
Where usage is known: Duplex Manifold Size
Usage, Ft
3
per Month Divide by 2 for each Bank
5850 6
9750 10
13650 14
17600 18
21500 22
25350 26
29250 30
N
2
O Cylinder Manifold:
Divide total by 2 for each Bank
No. of Total
Operating No. of
Rooms Cylinders
1 to 4 4
5 to 8 8
9 to 12 12
13 to 16 16
17 to 20 20
CO
2
Cylinder Manifold:
Divide total by 2 for each Bank
No. of Total
Operating No. of
Rooms Cylinders
1 to 8 4
9 to 16 8
17 to 32 12
N
2
Cylinder Manifold:
Divide total by 2 for each Bank
No. of Total
Operating No. of
Rooms Cylinders
1 to 2 4
3 to 4 8
5 to 6 12
7 to 8 16
9 to 10 20
11 to 12 24
13 to 14 28
www.ohiomedical.com
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
41
Medi c al Gas Equi pment Mount i ng
Hei ght Det ai l
F
I
N
I
S
H
E
D

F
L
O
O
R
C
E
I
L
I
N
G
C
O
L
U
M
N
S

-
M
A
N
U
A
L

O
R
E
L
E
C
T
R
I
C
A
L
L
Y
R
E
T
R
A
C
T
A
B
L
E
H
O
S
E

D
R
O
P
&

R
E
T
R
A
C
T
O
R
A
R
E
A

A
L
A
R
M
M
A
S
T
E
R

A
L
A
R
M
Z
O
N
E

V
A
L
V
E
B
O
X
G
A
S
P
A
N
E
L
O
U
T
L
E
T
S
F
I
N
I
S
H
E
D

C
E
I
L
I
N
G
6
3

6
6

6
6

6
6

6
2

6
0

6
2


I
N
A
L
L
O
.
R
.

S
For ordering information, please call 1-800-448-0770 or 847-855-0500
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
42
Suggest ed Medi c al Gas Mast er
Mechanical Room
Soiurce/Mainline
Isolation Valve
SUGGESTED ALARM POINTS FOR:
MEDICAL/SURGICAL VACUUM PUMPS
RESERVE VACUUM PUMP RUNNING
5.1.3.6.8 / 5.1.9.2.4(9) / 5.1.9.5.2 / 5.1.9.5.4(4)
(Termination point located in vacuum control panel)
MAIN LINE VACUUM LOW 5.1.9.2.4(8)
(Terminal point from main line vacuum switch)

WASTE ANESTHETIC GAS VACUUM PUMPS
RESERVE VACUUM PUMP RUNNING
5.1.9.5.2 / 5.1.3.7.4.1 / 5.1.9.2.4(9) / 5.1.9.5.4(5)
(Termination point located in vacuum control panel)
MAINLINE EVACUATION LOW
5.1.9.2.4(8)
(Termination point from main line vacuum switch)


SUGGESTED ALARM POINTS FOR:
MEDICAL AIR COMPRESSOR (All Types)
RESERVE COMPRESSOR RUNNING 5.1.9.5.4(1) / 5.1.3.5.14.5
(Termination point located in compressor control panel)
MAINLINE PRESSURE HIGH 5.1.9.2.4(7)
(Termination point from mainline pressure switch)
MAINLINE PRESSURE LOW 5.1.9.2.4(7)
(Termination point from mainline pressure switch)
CARBON MONOXIDE HIGH 5.1.9.4.5(2) / 5.1.3.5.15(2)
(Termination point from CO monitor)
DEW POINT HIGH 5.1.9.2.4(10) / 5.1.3.5.15(1)
(Termination point located in compressor control panel)
Also, For Reciprocating (Piston type) Compressors
HIGH TEMPERATURE 5.1.3.5.14.3 / 5.1.9.5.4(9) / 5.1.3.5.14.4(1)
(Termination point located in compressor control panel)
RECEIVER FLOODED (see note #1) 5.1.3.5.14.1 / 5.1.9.5.4(7)
(Termination point located in compressor control panel)
Also, For Liquid Ring Compressors:
RECEIVER FLOODED 5.1.3.5.14.1 / 5.1.9.5.4(7)
(Termination point located in compressor control panel)
HIGH WATER SEPARATOR 5.1.3.5.14.2 / 5.1.9.5.4(8)
(Termination point located in compressor control panel)
Soiurce/Mainline
Isolation Valve
MASTER ALARM #1
MAY BE LOCATED IN SECORITY
(Requires 24 Hourt/Day Surveillance)
www.ohiomedical.com
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
43
Medical Gas Room
MASTER ALARM #2
MAY BE LOCATED IN ENGINEERING
DEPARTMENT/AREA
SUGGESTED ALARM FUNCTIONS FOR:
MANIFOLDS WITHOUT RESERVE
CYLINDER BANK CHANGEOVER 5.1.9.2.4(1) / 5.1.3.4.10.6
(Termination point located in power supply)
MAINLINE PRESSURE HIGH 5.1.9.2.4(7)
(Termination point from mainline pressure switch)
MAINLINE PRESSURE LOW 5.1.9.2.4(7)
(Termination point from mainline pressure switch)
SUGGESTED ALARM FUNCTIONS FOR BULK
OXYGEN SYSTEM
CRYOGENIC BULK GAS UNITS WITH CRYOGENIC
RESERVE

LIQUID LEVEL LOW 5.1.9.2.4(2) / 5.1.3.4.13.6(1)
RESERVE IN USE 5.1.9.2.4(3) / 5.1.3.4.13.6(2)
MAIN LINE PRESSURE HIGH 5.1.9.2.4(7)
MAIN LINE PRESSURE LOW 5.1.9.2.4(7)
RESERVE LOW 5.1.9.2.4(5) / 5.1.3.4.11.6(3)
CHANGEOVER 5.1.9.2.4(1) / 5.1.3.4.13.6(5)
RESERVE PRESSURE LOW 5.1.9.2.4(6) /
5.1.3.4.11.6(4) (Not Functional)

CRYOGENIC BULK GAS UNITS WITH CYLINDER
RESERVE
LIQUID LEVEL LOW 5.1.3.4.13.6(1) / 5.1.9.2.4(2)
RESERVE IN USE 5.1.9.2.4(3) / 5.1.3.4.13.6(2) /
5.1.3.4.13.5
MAIN LINE PRESSURE HIGH 5.1.9.2.4(7)
MAIN LINE PRESSURE LOW 5.1.9.2.4(7)
CHANGEOVER 5.1.9.2.4(1) / 5.1.3.4.13.6(5)
SUGGESTED ALARM FUNCTIONS FOR:

MANIFOLDS WITH RESERVE
CYLINDER BANK CHANGEOVER 5.1.9.2.4(1) / 5.1.3.4.12.9(1)
(Termination point located in power supply)
RESERVE IN USE 5.1.9.2.4(3) / 5.1.3.4.15.5 / 5.1.3.4.12.9(3)
(Terminal point from actuating pressure switch 261920*)
RESERVE LOW 5.1.3.4.12.9(4) / 5.1.9.2.4(5)
(Terminal point from actuating pressure switch 261921*)
MAINLINE PRESSURE HIGH 5.1.9.2.4(7)
(Termination point from mainline pressure switch)
MAINLINE PRESSURE LOW 5.1.9.2.4(7)
(Termination point from mainline pressure switch)
Typical Nitrogen (N2) Installation
Soiurce/Mainline
Isolation Valve
MASTER ALARM #2
261921 Actuating
Pressure Switch
(Reserve Low,
500 PSIG)
Regulator
SUGGESTED HEIGHT
FOR CYLINDER
Source/Mainline
Isolation Valve
261920* Actuating Pressure
Switch (Reserve In Use, Activate
HP Cylinders set at 85 PSIG)
Typical Oxygen (O2), Medical Air (Air),
Nitrous Oxide (N2O), Carbon Dioxide (CO2)
Installation
Check Valve
Located
in Pigtail
Al ar m Poi nt s f or Level 1 Fac i l i t i es
For ordering information, please call 1-800-448-0770 or 847-855-0500
Information/products in this catalog subject to change without notice.
255322 (Rev.11) 06/2010
44
Ohio Medical Corporation

Your Single Source:


From the Basement to the Bedside
You can Trust our Experience
Factory Direct Solutions
Ohio Medical Corporation

1111 Lake Side Drive


Gurnee, IL 60031-4099 USA
P: 800-448-0770
P: 847-855-0500
F: 847-855-6300
www.ohiomedical.com
www.squire-cogswell.com
Ohio Medical Corporation

Your Single Source:


From the Basement to the Bedside
You can Trust our Experience
Factory Direct Solutions
P/N 255322

Das könnte Ihnen auch gefallen