910712 Appendix A: Command Menus Page A-15
4.2.3 Calibrate Programmable Goniometer Angles
For channels assembled on a programmable goniometer, the exact two-theta angle can be obtained using the “Cali-
brate Programmable Goniometer Angle” option. A step scan is made over the peak and the results are displayed as a
‘bar graph. In the calibration procedure, the zero offset must be calibrated first using one goniometer channel as aroter-
ence, then the angle is calibrated for each of the other goniometer channels.
CALIBRATE PROGRAMMABLE GONIOMETER ANGLES 26-may 88 10:25:05
Channel Nil
x Energy 27.47
Stepsize 0.01
Time/step =: 1.0
sample ident.
steel
Peak (kCPS) : 30.392
Old peak 48.68
New peak 48.68
4--4- Difference : 0.00
48.47 48.57 48.67 48.77 48.87
(Edit, Start, abort, /] : start
‘The upper right-hand side of the action form shows the input fields, these are:
1. Channel name and X-ray line energy.
2. Scan step size and the measured time per step.
All these input fields can be changed using the Edit mode.
{", after measurement, the new peak angleis different from the previous angle, the commandline willbe replaced by the
question:
“Save new peak angle [YIN]"
‘The results of the calibration are stored in the lower right-hand side of the action form. This shows the peak count, the
‘ld peak angle, the new peak angle and the difference between them.