Sie sind auf Seite 1von 100

Catalog Number 62005-18

APA 6000 Process Analyzer


Installation and Maintenance Manual

Hach Company, 2001, 2002, 2004. All rights reserved. Printed in the U.S.A. ce/dk 04 February 2005 Edition 4

Table of Contents
Safety Precautions .................................................................................................................................................... 4 Specifications ............................................................................................................................................................ 5

Section 1 Introduction ........................................................................................................................................ 7


1.1 How to Use the Manuals ..................................................................................................................................... 7 1.2 How the Analyzer Works ..................................................................................................................................... 7 1.2.1 Components .............................................................................................................................................. 8 1.2.2 Flow Through the Instrument..................................................................................................................... 8

Installation ........................................................................................................................................................... 11 Section 2 Installation......................................................................................................................................... 13 2.1 Selecting a Location for the APA 6000 Analyzer ............................................................................................... 13 2.2 Mounting ........................................................................................................................................................... 13 2.2.1 Bench Mount ........................................................................................................................................... 14 2.2.2 Wall Mount............................................................................................................................................... 14 2.3 Electrical Connections ...................................................................................................................................... 15 2.3.1 Power Connections at the Analyzer......................................................................................................... 16 2.3.2 Alarm Relay Connections ........................................................................................................................ 20 2.3.3 Recorder Output Connections ................................................................................................................. 21 2.4 Plumbing Installation ......................................................................................................................................... 22 2.4.1 Sample Line Installation .......................................................................................................................... 23 2.4.1.1 Sample Conditioning................................................................................................................... 24 2.4.2 Drain Line Connection ............................................................................................................................. 25 2.4.2.1 Waste Disposal ........................................................................................................................... 26 2.5 Air Purge ........................................................................................................................................................... 26 Startup .................................................................................................................................................................. 27 Section 3 Startup ............................................................................................................................................... 29 3.1 Introduction ....................................................................................................................................................... 29 3.2 Preparing Sample Conditioning and Starting Sample Flow .............................................................................. 29 3.3 Installing Reagents and Standards ................................................................................................................... 29 3.3.1 Supplying Power ...................................................................................................................................... 30 3.3.2 Priming the Analyzer ............................................................................................................................... 31 Section 4 Instrument Menus ........................................................................................................................... 33 4.1 The AquaTrend Interface................................................................................................................................... 33 4.1.1 The AquaTrend Interface Keypad ............................................................................................................ 33 4.1.2 The AquaTrend Interface Display ............................................................................................................ 34 4.2 Menu Functions................................................................................................................................................. 37 4.2.1 Alarm Log ................................................................................................................................................ 37 4.2.2 AquaTrend Menu ..................................................................................................................................... 37 4.2.2.1 Display Setup.............................................................................................................................. 37 4.2.2.2 Language .................................................................................................................................... 40 4.2.2.3 Edit Name ................................................................................................................................... 40 4.2.2.4 Set Time/Date ............................................................................................................................. 41 4.2.2.5 Security Setup ............................................................................................................................ 42 4.2.3 Sensor Menu ........................................................................................................................................... 44 4.2.3.1 Calibration................................................................................................................................... 44 4.2.3.2 Alarm Setup ................................................................................................................................ 47 4.2.3.3 Recorder Setup........................................................................................................................... 48 4.2.3.4 Measurement Options ................................................................................................................ 50
Page 2 Table of Contents

6200518TOC.fm

Table of Contents
4.2.3.5 Enabling Solution Level Monitoring ............................................................................................ 51 4.2.3.6 Setting Solution Levels ............................................................................................................... 51 4.2.3.7 Solutions & Cleaning Options ..................................................................................................... 52 4.2.3.8 Sensor Diagnostics ..................................................................................................................... 52 4.3 Other Device Menu and Network Menu ............................................................................................................ 55 4.4 Data Display...................................................................................................................................................... 55 4.4.1 Numerical Display.................................................................................................................................... 55

Maintenance ........................................................................................................................................................ 57 Section 5 Maintenance and Troubleshooting ............................................................................................. 59 5.1 Maintenance ..................................................................................................................................................... 59 5.1.1 Shutting the Analyzer Down (Off or Standby).......................................................................................... 59 5.1.2 Scheduled Maintenance .......................................................................................................................... 59 5.1.2.1 Cleaning the Instrument.............................................................................................................. 60 5.1.2.2 Sample Conditioning................................................................................................................... 60 5.1.2.3 Changing Modules ...................................................................................................................... 62 5.1.2.4 Changing Tubing ........................................................................................................................ 63 5.1.2.5 Rotary Valve Module .................................................................................................................. 65 5.1.2.6 Maintaining the Autoburette Module ........................................................................................... 68 5.1.2.7 Greasing the Lead Screw in the Autoburette Module ................................................................. 70 5.1.3 Unscheduled Maintenance ...................................................................................................................... 71 5.1.3.1 Fuse Replacement...................................................................................................................... 71 5.1.3.2 Cleaning the Mixer Module Chamber ......................................................................................... 74 5.2 Troubleshooting................................................................................................................................................. 75 5.2.1 General Troubleshooting Procedure ........................................................................................................ 75 5.2.1.1 Check Alarm Log ........................................................................................................................ 75 5.2.1.2 Check Solutions .......................................................................................................................... 75 5.2.2 Examine the Rotary Valve ....................................................................................................................... 76 5.2.3 Examine the Autoburette ......................................................................................................................... 76 5.2.4 Examine the Tubing and Fittings ............................................................................................................. 76 5.2.5 Back Pressure Gauge ............................................................................................................................. 76 5.3 Specific Alarm/Warning Messages ................................................................................................................... 77 Appendix A Network Connections....................................................................................................................... 81
Network Connections in the APA 6000 Analyzer.......................................................................................... 81

Appendix B Printer and Computer Formats ........................................................................................................ 85


APA 6000 Printer and Computer Formats........................................................................................................ 85

Appendix C Second Sample Options .................................................................................................................. 89


Sample Sequencing ......................................................................................................................................... 89 Quality Control.................................................................................................................................................. 89 Grab Sample .................................................................................................................................................... 89 Setting Up the Sample Sequence .................................................................................................................... 90

General Information........................................................................................................................................... 93
Replacement Parts and Accessories ...................................................................................................................... 94 How to Order........................................................................................................................................................... 96 Repair Service ........................................................................................................................................................ 97 Limited Warranty ..................................................................................................................................................... 98 Certification ............................................................................................................................................................. 99

6200518TOC.fm

Page 3 Table of Contents

Safety Precautions

Section 2.3 Electrical Connections on page 15 Section 2.3.1 Power Connections at the Analyzer on page 16 Section 2.3.2 Alarm Relay Connections on page 20 Section 2.3.3 Recorder Output Connections on page 21 Section 2.4 Plumbing Installation on page 22 Section 3.3 Installing Reagents and Standards on page 29 Section 5.1.2.3 Changing Modules on page 62 Section 5.1.2.4 Changing Tubing on page 63 Section 5.1.2.6 Maintaining the Autoburette Module on page 68 Section 5.1.3.1 Fuse Replacement on page 71

Page 4 Safety Precautions

6200518 Safety.fm

Specifications
Specifications are subject to change without notice. Specifications for range, accuracy, repeatability, response time, cycle time, detection limit, and other parameter specific items are listed in the parameter-specific manual.

Performance
Recorder Outputs: Two isolated, 420 mA outputs standard. Output span programmable over any portion of the range (500 V isolation from earth ground). Alarms: 14 programmable internal alarms Relays: Two SPDT relays with a maximum contact rating of 5 A resistive load at 240 V ac. Additional relays available through optional Signal Output Module. Power Requirements: 95240 V ac, 50 or 60 2 Hz, 150 VA Installation Category: II Sample Inputs: Two sample streams (with optional hardware) Maximum Operating Temperature Range: 5 to 50 C (41 to 122 F) Relative Humidity: 5 to 95% relative humidity, non-condensing Network Interface: LonWorks FTT-10 twisted-pair free topology Maximum Node-to-node Distance: 400 meters (1320 feet)

Features
Instrument Enclosure: NEMA 4X (Indoor) and IEC 529 (IP 66) with provision for air purge Instrument Installation Category: II Reagent Enclosure: Fully enclosed latch cases with protected ducting for instrument drain; additional plumbing connections for instrument case and reagent enclosure overflow drains.

Sample Requirements
Sample Filtration: 0.5 m or smaller. Sample Pressure: 0.5 to 30 psig (0.035 to 2 bar) maximum at sample inlet block Sample Temperature: 5 to 50 C (41 to 122 F) Sample Flow: 1000 to 2000 mL/min.

6200518 Specifications.fm

Page 5 Specifications

Visit http: //www.hach.com

Section 1

Introduction

1.1 How to Use the Manuals


This instrument requires two manuals for complete operating instructions; the APA 6000 Installation and Maintenance Manual and a parameter-specific operation manual. This APA 6000 Installation and Maintenance Manual (Cat. No. 62005-18) provides instructions for instrument installation, menu setup, operation, maintenance, and troubleshooting. The parameter-specific manual contains information related to parameter specific instrument components and software functions. Descriptions of reagents and standards and their consumption rates, method of analysis, operational sequence, and bench comparison testing are all addressed in the parameter-specific manual.

1.2 How the Analyzer Works


The APA 6000 Process Analyzer is a microprocessor-controlled process analyzer designed to continuously monitor a specific substance in a sample stream (see Figure 1). The analyzer consists of five basic components: Figure 1 Module Components AquaTrend Interface Autoburette Module Detector Module and Holding Coil Valve Module Mixer Module

6200518 Introduction.fm

Page 7 Introduction

Section 1
1.2.1 Components
The AquaTrend Interface The APA 6000 Process Analyzer has an AquaTrend Interface built into the instrument door. It provides a numerical or graphical display and a keypad for user input. The AquaTrend Interface displays, graphs, and logs sensor readings, and is used for operator control and sensor setup. Autoburette Module The analyzer uses a positive-displacement autoburette to measure fluids through the various analytical components in the instrument. The autoburette module consists of a clear acrylic block which houses a ceramic piston, a double seal for the piston, a linear actuator, a two-way solenoid valve, and control circuitry. Detector Module Samples are measured in a temperature controlled flow cell detector. All detector modules contain a holding coil. The holding coil acts as a storage area for fluids as they are drawn into the instrument before being directed to the appropriate module for mixing or sensor readings. The detector module for some instruments also contains a reaction coil. The reaction coil optimizes the dispersion utilized in some analysis methods. Valve Module The valve module directs fluids to various components in the instrument. Mixer Module The mixer module mixes reagents with the sample or standards. The mixing action also removes excess air from the sample.

1.2.2 Flow Through the Instrument


The following is a simplified version of how fluids are cycled through the analyzer: 1. In a normal measuring cycle, the instrument uses the autoburette module to draw sample in through the sample conditioning block into the holding coil. 2. The pump then draws reagent(s) into the holding coil. 3. Next, the pump moves the reagent(s) and sample from the holding coil into the mixing module with the sample. 4. When the sample and reagent are adequately mixed, the instrument draws the mixture into the holding coil. 5. The next pump actuation moves the mixture from the holding coil into the reaction coil (not used in all chemistries). As the mixture flows past the reaction coil and through the detector, a measurement is taken and the fluid continues to flow into the drain. The cycle then starts over again at Step 1.

Page 8 How the Analyzer Works

6200518 Introduction.fm

Section 1
Step One The burette draws sample into the Holding Coil. 10-Port Rotary Valve Reaction Coil

Detector
Holding Coil

Mixer Drain

Filtered Sample IN Reagent IN 10-Port Rotary Valve Mixing Chamber Reaction Coil Waste

Step Two The burette draws reagent into the Holding Coil.

Detector
Holding Coil

Mixer Drain

Filtered Sample IN Reagent IN Step Three The burette moves sample and reagent into the Mixing Chamber. 10-Port Rotary Valve Mixing Chamber Reaction Coil Waste

Detector
Holding Coil

Mixer Drain

Filtered Sample IN Reagent IN Step Four The burette draws the sample/reagent mixture into the Holding Coil. 10-Port Rotary Valve Mixing Chamber Reaction Coil Waste

Detector
Holding Coil

Mixer Drain

Filtered Sample IN Reagent IN Step Five The burette moves the sample/reagent mixture through the Reaction Coil and the Detector and measures the response of the Detector. Mixing Chamber Waste

10-Port Rotary Valve

Reaction Coil

Detector
Holding Coil

Mixer Drain

Filtered Sample IN Reagent IN Mixing Chamber Waste

6200518 Introduction.fm

Page 9 How the Analyzer Works

Visit http://www.hach.com

Installation

DANGER
Some of the following manual sections contain information in the form of warnings, cautions and notes that require special attention. Read and follow these instructions carefully to avoid personal injury and damage to the instrument. Only personnel qualified to do so, should conduct the installation/maintenance tasks described in this portion of the manual.

DANGER
Certains des chapitres suivants de ce mode demploi contiennent des informations sous la forme davertissements, messages de prudence et notes qui demandent une attention particulire. Lire et suivre ces instructions attentivement pour viter les risques de blessures des personnes et de dtrioration de lappareil. Les tches dinstallation et dentretien dcrites dans cette partie du mode demploi doivent tre seulement effectues par le personnel qualifi pour le faire.

PELIGRO
Algunos de los captulos del manual que presentamos contienen informacin muy importante en forma de alertas, notas y precauciones a tomar. Lea y siga cuidadosamente estas instrucciones a fin de evitar accidentes personales y daos al instrumento. Las tareas de instalacin y mantenimiento descritas en la presente seccin debern ser efectuadas nicamente por personas debidamente cualificadas.

GEFAHR
Einige der folgenden Abschnitte dieses Handbuchs enthalten Informationen in Form von Warnungen, Vorsichtsmanahmen oder Anmerkungen, die besonders beachtet werden mssen. Lesen und befolgen Sie diese Instruktionen aufmerksam, um Verletzungen von Personen oder Schden am Gert zu vermeiden. In diesem Abschnitt beschriebene Installations- und Wartungsaufgaben drfen nur von qualifiziertem Personal durchgefhrt werden.

PERICOLO
Alcune parti di questo manuale contengono informazioni sotto forma davvertimenti, di precauzioni e di osservazioni le quali richiedono una particolare attenzione. La preghiamo di leggere attentivamente e di rispettare quelle istruzioni per evitare ogni ferita corporale e danneggiamento della macchina. Solo gli operatori qualificati per luso di questa macchina sono autorizzati ad effettuare le operazioni di manutenzione descritte in questa parte del manuale.

6200518 Installation Stopper.fm

Page 11 Installation

Visit http: //www.hach.com

Section 2

Installation

2.1 Selecting a Location for the APA 6000 Analyzer


Locate the analyzer as close to the sampling point as possible. Shorter distances between the process stream and the analyzer mean faster analyzer response to changes in analyte concentration. Long sample lines are also more vulnerable to sediment accumulation. Do not place the sampling point near a point of chemical addition. The process stream at that location may be inadequately mixed, yielding erroneous readings. The instrument enclosure is designed for general-duty, indoor installation. Do not mount the analyzer in direct sunlight. Ambient temperatures may range from 550 C (41122 F), but best performance will result if ambient temperature is between 5 and 35 C (41 and 95 F). Use Figure 2 to determine the space needed to install the instrument. To control the analyzer from a location other than the instrument location, connect a remote AquaTrend Interface. It can be mounted on a bench, pole, or panel as far away as 400 m (1320 feet). Figure 2 Overall Dimensions of the APA 6000 Analyzer

2.2 Mounting
An installation kit (Cat. No. 51040-00) is supplied with the instrument. The kit includes: Grab sample kit (Cat. No. 62037-00) Four 0.5 micron filters (Cat. No. 62011-00) 20 ft. of -in. poly tubing (Cat. No. 47438-00)
Page 13 Installation

6200518 Installation.fm

Section 2
2.2.1 Bench Mount
Important Note: The analyzer must be bolted to the bench.

Secure the rear of the instrument to the bench or table with two -20 UNC screws through the holes in the metal frame on the back of the instrument (see Figure 3). This will prevent the instrument from tipping while the module panel is open.

2.2.2 Wall Mount


The back plate of the instrument has four key-shaped holes for wall mounting. The analyzer can be mounted directly to a wall or to rails capable of holding 157.5 kg (350 lb.). Leave at least 4656 cm (1822 in.) of space above the top of the bolt pattern (see Figure 3) to allow access to the reagent enclosures. 1. Make sure no power is supplied to the analyzer. 2. Install the top two 3/8-16 UNC hex bolts into the wall or fixture according to the dimensions in Figure 3, leaving a -inch gap under each bolt head. 3. Slide the bracket keyholes onto the bolts and lower the frame so the bolts settle in the narrow part of the keyholes. 4. Install two bolts through the lower holes in the frame, into the wall or fixture. 5. Tighten all four bolts securely. Figure 3 Mounting Dimensions

Page 14 Mounting

6200518 Installation.fm

Section 2
2.3 Electrical Connections
DANGER All electrical connections should be made by a qualified technician to assure compliance to all applicable electrical codes. DANGER Tous les branchements lectriques doivent tre effectus par un technicien qualifi pour assurer la conformit toutes les normes lectriques applicables. PERIGO Todas las conexiones elctricas deben ser ubstaladas por personal tcbico cualificado a fin de asegurar su conformidad con las normas elctricas vigentes. GEFAHR Alle elektischen Anschlsse mssen van qualifizierten Techniekem durchgefhrt werden, damit die bereinstimmung mit allen elektrischen Anschlussdaten gewrhrleistet ist. PERICOLO Tutte le connessioni elettriche devono essere fatte da un tecnico qualificato.

To access electrical connections in the APA 6000 instrument, remove the two hold-down fasteners in the upper corners of the module panel. Slowly pull the panel forward until it is fully supported by the steel restraining cable (see Figure 4 and Figure 5). Four access holes for electrical connections are provided along the back of the instrument enclosure (see Figure 6). Remove the cover seals by loosening the wing nuts from inside the instrument. This product is designed to meet NEMA 4X (INDOOR) and IEC 529 IP 66 enclosure requirement when the front door is securely latched. Achieving these protection levels, however, depends on the use of proper hardware in the installation of this product. Standard off-the-shelf NEMA-approved conduit hardware is required for proper installation of all electrical connections. For process or industrial applications, the national electrical codes of most countries require that ac service feeds be hard-wired and contained in conduit systems. The APA 6000 Analyzer has been designed to conform to this requirement. Hach recommends conduit for two reasons: 1. It is generally required by most local electrical codes, and 2. use of metal conduit can improve immunity to lightning surges and ac power transients. In hard-wired electrical applications, the power and safety ground service drops for the analyzer should be no longer than 6 meters (20 feet) unless metal conduit is used to shield the ac power wiring. In applications where power cords are allowed by local electrical codes and power surges and transients are not a great concern, an 18-gauge, 3-conductor power cord (including a safety ground wire) can be used, but its length must not exceed 3 meters (10 feet). The power cord or power cable is not supplied with the APA 6000 Analyzer and must be selected to meet applicable electrical codes for the installation site.

6200518 Installation.fm

Page 15 Electrical Connections

Section 2
Hach recommends that care be taken with hook-up configurations other than shown in these illustrations. First, be sure your configuration satisfies local electrical codes and second, ensure that your configuration does not cause susceptibility to lightning, power surges, or transients upsets.

2.3.1

Power Connections at the Analyzer


If power cords are allowed by local electrical code, a 125 V UL/CSA-approved power cord with an approved NEMA-style strain relief and a standard 115 V North American-style plug (Hach Cat. No. 46306-00) or a 230 V VDE-approved power cord with an approved NEMA-style strain relief and a Continental European-style plug (Hach Cat. No. 46308-00) can be ordered. The APA 6000 Analyzer operates on any ac line voltage from 95240 V ac, 50 or 60 Hz. A 12- to 18-gauge power lead supply wire (power cord), with insulation rated at 300 V (600 V peak), is required. The analyzer is designed for hard-wire connection to ensure conformation to local electrical codes.

Note: Be sure the analyzer power switch is turned off. Dry operation may damage the analyzer.

Route the power leads through the access hole on the back of the enclosure. Line voltage connections are wired directly to the terminal strip J10 (see Figure 6 and Figure 7). Make the ground connection to the adjacent ground stud using an appropriate closed loop connector (see Table 1). Table 1 Wiring Color Codes

Connector hot line conductor 01 neutral conductor 02 ground conductor (#8 stud)

North American black white green/green w/yellow stripe

IEC brown blue green with yellow stripe

Page 16 Electrical Connections

6200518 Installation.fm

Section 2
Figure 4 Looking Inside the APA 6000 Analyzer, with Hinged Module Panel Down

Figure 5

Looking Inside the APA 6000 Analyzer, Removing the Customer Access Cover

6200518 Installation.fm

Page 17 Electrical Connections

Section 2
Figure 6 Looking Inside the APA 6000 Analyzer, Locating and Identifying the Customer Connections

Back View of Instrument

Air Purge

Power Connections Alarm Connections Drain

Network Connections Recorder Connections

Page 18 Electrical Connections

6200518 Installation.fm

Section 2
Figure 7 Wiring Options

6200518 Installation.fm

Page 19 Electrical Connections

Section 2
2.3.2 Alarm Relay Connections
DANGER La zone de raccordement des relais est conue pour des connexions soit tout en basse tension (<30 V) soit tout en haute tension (>30 V). Un risque dlectrocution existe si des connexions basse et haute tension sont faites dans la zone de raccordement des relais. PELIGRO El rea de conexin de los rels esta diseada para conexiones que sean todas de baja tensin (inferior a 30V o bien todas de alta tensin (superior a 30V). Existe riesgo de descargas si se combinan conexiones de alto y bajo voltaje en el rea de conexin de los rels. GEFAHR Die Ralaisanschlsse sind entweder fr Niederspannung (< 30V) oder fr Hochspannung (>30 V) vorgesehen. Wenn sowohl Nieder-als auch Hochspannung an die Relais angeschlossen wird, kann es zu einem Stromschlag kommen. PERICOLO La zona riservata ai contatti di allarme disegnata per bassi voltaggi (<30 V) e alti voltaggi (> 30V). Assolutamente non confondersi nell'utilizzo dei bassi e degli alti voltaggi. Note: Limit current to the relay contacts to 5 amps. This can be accomplished with an external switch and a 5-amp fuse or with a switched 5-amp circuit breaker. You must be able to remove power from the relays locally in case of an emergency or for servicing of the product.

DANGER The relay connection area is designed for either all low voltage (<30 V) or all high voltage (>30 V) connections. A shock hazard exists if low and high voltage connections occur simultaneously in the relay connection area.

The analyzer contains two alarm relays designed for use with either high voltage (greater than 30 V-RMS and 42.2 V-PEAK or 60 V dc) or low voltage (less than 30 V-RMS and 42.2 V-PEAK, or less than 60 V dc), but not a combination of both high and low voltage. Wire alarm relays to the connectors labeled Alarm A or Alarm B on the back electrical panel (see Figure 8). For terminal strip and strain relief locations see Figure 6. SPDT relay contacts are unpowered and rated to carry up to 5 amperes at 240 V ac with a resistive load. If high-current or inductive-load devices are to be controlled by analyzer alarms, use a heavy-duty slave relay actuated by the alarm relay. Both normally open and normally closed contacts are available. The connector will accept 22- to 12-gauge wire, insulated as appropriate for the voltage. Size the wires according to the load and local electrical code regulations. If more relays are desired, supplement your system with the AquaTrend Signal Output Module (SOM) (Cat. No. 51250-00). Each SOM provides two additional relays.

Page 20 Electrical Connections

6200518 Installation.fm

Section 2
Figure 8 Alarm Relay Connections

2.3.3

Recorder Output Connections


DANGER Les installations lectriques dans les usines dveloppent des tensions de masse si de mauvaises techniques de mise la terre sont utilises. Pour rduire les risques dlectrocution, prendre des prcautions pour le branchement dquipements raccords la terre des emplacements diffrents. Suivre attentivement les instructions donnes dans ce chapitre. PELIGRO Si no se emplean tcnicas adecuadas de conexin a tierra, los sistemas elctricos de la instalacin pueden generar tensiones de masa. Con el fin de disminuir el riesgo de descarga elctrica, haga la conexin a tierra en lugares diferentes. Respecte las instrucciones que se indican en esta seccin. GEFAHR Spannungsversorgungsanlagen fhren evlt. Erdpotential, wenn eine schlechte Systemerdung vorgenommen wurde. Vosicht ist geboten, wenn das angeschlossene Gert an verschiedenen Orten geerdet ist. Die Anweisungen in diesem Abschnitt sind genau zu befolgen. PERICOLO Assicurarsi che la messa a terra sia attiva, altrimenti esiste la possibilit di tensioni latenti nella carcassa dello strumento.

DANGER Facility power systems may have ground potentials if poor grounding techniques have been used. To reduce the chance of shock hazards, care must be taken when connecting equipment grounded at different locations (potentials). Carefully follow the instructions given in this section.

The analyzer contains two isolated 420 mA analog outputs which can drive a recorder or control a process (see Figure 9). Recommended output impedance is 3.6 to 500 ohms. Route 22-to 12-gauge recorder wires through the access hole below the terminal strip and connect them to the terminal
Page 21 Electrical Connections

6200518 Installation.fm

Section 2
strip J8, labelled Recorder (see Figure 6). Use shielded, twisted-pair cable to connect the analyzer to the recorder. Tie the shield to earth ground only at the analyzer end to avoid ground loops. To ensure the analyzer meets NEMA 4X and IP 66 enclosure ratings, use only NEMA-style enclosure strain relief fittings. If more analog outputs are desired, you can supplement your system with the AquaTrend Signal Output Module (SOM). Each SOM provides two additional analog outputs. Figure 9 APA to Recorder Connections

2.4

Plumbing Installation
The nature of tasks described in this section requires individuals to understand the associated dangers. Please read all caution statements in the left-hand column of this section before installing the instrument. Burns, shock, eye injury, fire, and chemical exposure may occur. Hach Company assumes individuals performing these tasks are qualified and aware of proper safety procedures.

CAUTION This instrument is not intended for use with flammable samples or those containing hydrocarbons.

PRUDENCE Cet instrument nest pas conu pour une utilisation avec des chantillons inflammables ou des chantillons contenant des hydrocarbures. CUIDADO Este instrumento no est destinado para uso con muestras inflamables o que contengan hidrocarburos. VORSICHT Dieses Gert darf nicht fr Tests mit brennbaren Proben oder Proben, die Kohlenwasserstoffe enthalten, benutzt werden. ATTENZIONE Questo strumento non utilizzabile con campioni infiammabili o contenenti idrocarburi.

Page 22 Plumbing Installation

6200518 Installation.fm

Section 2
DANGER This instrument should be installed by qualified technical personnel to ensure adherence to all applicable electrical and plumbing codes. DANGER Cet appareil doit tre install par du personnel technique qualifi, afin dassurer le respect de toutes les normes applicables dlectricit et de plomberie. PELIGRO Este instrumento debe ser instalado por personal tcnico capacitado para assegurar el cumplimiento con todos los cdigos elctricos y de plomera aplicables. GEFAHR Um zu gewhrleisten, da alle elektrischen und sanitrinstallationstechnischen VDE-Vorschriften und gegebenefalls die Zusatzvorschriften der zustndigen Elektrizitts- und Wasserwerke erfllt werden, darf dieses Gert nur von geschultem Fachpersonal installiert werden. PERICOLO La strumentazione deve essere installata da personale qualificato, sia per le connessioni elettriche sia per quelle idrauliche.

2.4.1 Sample Line Installation


Selecting a good, representative sampling point is important for optimum performance from the analyzer. The sample analyzed by the analyzer must be representative of the condition of the entire system. If a sample is drawn from a location too close to points of chemical additions to the process stream, inadequate mixing or an incomplete reaction can result in erratic readings from the analyzer.
Note: Plumbing connections should be made only by qualified personnel to assure conformance to all applicable plumbing codes.

Run sample lines with small diameter tubing to the analyzer to minimize lag time (the time required for sample to flow from sample point to the analyzer). Route the tubing as directly as possible between the sampling point and the analyzer. Long sample lines and low flow rates result in significant lag time between actual process line conditions and the sampling point of the instrument. Install sample line taps into larger process pipes to minimize the chances of ingesting sediment from the pipe line bottom or air bubbles from the top. A tap projecting into the center of the pipe is ideal (see Figure 10).

Figure 10 Sample Line Location in Process Stream

Air (Typical) Sediment (Typical)

Poor

Poor

Good

Best

6200518 Installation.fm

Page 23 Plumbing Installation

Section 2
New tubing, valves and other sample conditioning equipment may be contaminated with substances (oils, dust, etc.) which may contribute to low or high readings on initial startup. Flush the sample line with sample for 12 hours before calibration. For faster cleaning, plumb the sample line to a container of 14 liters of a dilute caustic solution such as 1 N (5%) sodium hydroxide solution and cycle it through to clean sample system components. When the solution is depleted, connect the sample line to the sample source and run the analyzer for an hour to purge the sodium hydroxide solution.

2.4.1.1 Sample Conditioning


The analyzer is equipped with a sample conditioning block that includes a 0.5 m filter. The unit is located on the left side of the instrument. See Figure 11. Use -in OD poly tubing from the sample source to the sample conditioning block. The tubing connects to the block through the lower push-in fitting. From the upper push-in fitting of the sample conditioning block, excess sample flows to drain. A vent assembly is required to avoid excess vacuum at the sample conditioning block. The vent should be located above the sample conditioning block. For wastewater applications or other samples containing high solids, an APA 6000 Micro Filter System is recommended (Cat. No. 51339-00, 115 V model or Cat No. 51339-01, 230 V model). A 40-mesh y-strainer (Cat. No. 46618-00) can be used to remove large solids alone or with the Micro Filter System. When testing highly aerated samples, the manufacturer recommends installing a Process Sample Heater such as Cat. No. 48685-00, prior to the sample conditioning unit. Warming the samples removes some of the excess air and stabilizes the analyzer readings.
Important Note: Sample conditioning requires regular maintenance for proper operation. Important Note: If the sample conditioning system supplied with the instrument is removed, it must be replaced with an alternative filtration system of 0.5 m or smaller. Failure to provide the required filtration will void the instrument warranty.

Page 24 Plumbing Installation

6200518 Installation.fm

Section 2
Figure 11 Location of the Sample Conditioning Block

Sample Conditioning Block

Vent

Unfiltered Sample Out

Drain

Unfiltered Sample In Remove these two screws to take off the Sample Conditioning Block. Filtered sample into instrument.

2.4.2 Drain Line Connection


Note: It is imperative to mount the drain point below the instrument drain block. Failure to do so may result in improper functioning of the instrument or back flow of waste into the instrument cabinet.

The analyzer is connected to a drain system through a -inch hose barb fitting on a drain block on the right side of the analyzer case (see Figure 12). After analysis, the waste exits the instrument through the manifold at the bottom of the instrument and into the drain block. This block connects to a -inch ID hose and provides a vented vacuum break at this point. A vent is located on the top of the drain block. Also see section 2.4.2.1. Route instrument waste through a -in. ID hose to a PVC pipe assembly, which must be located one foot below the level of the instrument drain block. The catastrophic drain is located in the middle of the tubing manifold and is sized to accommodate a 1-in. ID hose. This drain is used if an upset inside the instrument results in a significant leak. The supplied hose from this drain port must be connected and run to a drain which is one foot below the bottom of the instrument. The reagent enclosures each have an overflow fitting in the bottom of the case. Plumb these to the drain by punching out the knock-out in the fitting, attaching a 3/8-inch ID hose to each fitting and running it to a drain which is one foot lower in elevation than the bottom of the instrument.

6200518 Installation.fm

Page 25 Plumbing Installation

Section 2
Figure 12 Drain Line Connection

Drain Block

All drains must be at least one foot below the bottom of the instrument.

Drains from Reagent Enclosures

Drain from Instrument Drain Block (Customer supplied 3/4 inch tubing.) From Catastrophic Drain

2.4.2.1 Waste Disposal


The effluent from the APA analyzer is non-hazardous per federal RCRA regulations and may be safely discharged to a sanitary sewer based solely on federal rules. Other state and local regulations or permitting requirements may apply and it is the responsibility of the end-user to know and apply state or local requirements to their specific applications. Hach Company does not advocate the pouring, dumping, or discharging of any chemicals, hazardous or nonhazardous, to the ground or to septic systems.

2.5 Air Purge


Air purge may be necessary if the analyzer is located in an environment with high humidity and/or caustic vapors. The goal is to maintain a slight positive pressure in the instrument with dry instrument air. The air purge connection is located in the middle of the back of the instrument. To connect an air supply, remove the plug in the quick connect fitting. Connect in. poly tubing by pushing the tubing into the fitting. You should feel two distinct stops as you push. If the tubing is not seated, it will pop out when air pressure is applied. Use only dry instrument air at 15 scfh (7 L/min.)

Page 26 Air Purge

6200518 Installation.fm

Startup

DANGER
Handling chemical samples, standards, and reagents can be dangerous. Review the necessary Material Safety Data Sheets and become familiar with all safety procedures before handling any chemicals.

DANGER
La manipulation des chantillons chimiques, talons et ractifs peut tre dangereuse. Lire les Fiches de Donnes de Scurit des Produits (FDSP) et se familiariser avec toutes les procdures de scurit avant de manipuler tous les produits chimiques.

PELIGRO
La manipulacin de muestras qumicas, estndares y reactivos puede ser peligrosa. Revise las fichas de seguridad de materiales y familiarcese con los procedimientos de seguridad antes de manipular productos qumicos.

GEFAHR
Das Arbeiten mit chemischen Proben, Standards und Reagenzien ist mit Gefahren verbunden. Es wird dem Benutzer dieser Produkte empfohlen, sich vor der Arbeit mit sicheren Verfahrensweisen und dem richtigen Gebrauch der Chemikalien vertraut zu machen und alle entsprechenden Materialsicherheitsdatenbltter aufmerksam zu lesen.

PERICOLO
La manipolazione di campioni, standard e reattivi chimici pu essere pericolosa. La preghiamo di prendere conoscenza delle Schede Techniche necessarie legate alla Sicurezza dei Materiali e di abituarsi con tutte le procedure di sicurezza prima di manipolare ogni prodotto chimico.

6200518 Startup Stopper.fm

Page 27 Startup

Visit http: //www.hach.com

Section 3
3.1 Introduction

Startup

It is assumed that the analyzer has already been plumbed to the sample source and a drain, as well as wired to a power source and any other electrical wiring associated with 420 mA connections, relay connections, and network connections by the installation technician as outlined in Section 2. This section presents the steps, in sequence, an operator should take to begin operation. This section will: describe how to prepare the sample conditioning describe the steps to install the reagents and standards list the steps required to prime the sample lines and reagents.

3.2 Preparing Sample Conditioning and Starting Sample Flow


The analyzer is supplied with a 0.5 m ceramic filter (already installed in the sample conditioning block) and four replacement filters. 1. Set the sample flow rate through the sample conditioning block in the following range: 10002000 mL/min. 2. Check the flow at the air break of the drain (where the tubing from the sample inlet block meets the drain tee). If desired, measure the flow rate by directing the sample into a large measuring container (flask or cylinder) over a set amount of time. Keep the end of the tube at the same height as the push-in fitting. 3. Ensure that the filter is tightly threaded into the sample conditioning block to prevent unfiltered sample from entering the analyzer.

3.3

Installing Reagents and Standards


The solution bottles are labeled (i.e., Reagent 1, Standard 1, Cleaning Solution). The tubing into each of these bottles is also labeled with the solution name to prevent reagent line mix-ups. Some analyzer solutions require special accessories and handling. Please see the parameter specific manual for more information. To change solutions, follow these instructions: 1. Use a permanent marker to put a mark on the tubing where it exits the bottle. 2. Remove the bottle vent cap assembly and tubing from the top of the existing solution bottle. 3. Rinse the straw surrounding the tubing with deionized water and shake off any residual water before inserting it into the new solution. This avoids possible contamination. Be sure the tubing extends to the bottom of the straw. If not, adjust it and re-mark the tubing. 4. Insert the straw into the new bottle and screw the vent cap on. Be sure the mark on the tubing is still located where the tubing exits the bottle. This ensures the inner tube has remained at the bottom of the bottle.

6200518 Startup.fm

Page 29 Startup

Section 3
5. Place the new bottle in the reagent compartment. Hach recommends changing all the solutions at the same time and does not recommend adding fresh reagent to fill a partially full bottle.

3.3.1 Supplying Power


After all appropriate reagents and standards are installed and sample conditioning is prepared, the APA 6000 is ready to start up. Turn on the analyzer by setting the power switch to ON (I). The power switch is located at the center top of the bulkhead that contains the instrument modules (see Figure 13). Close the main door of the analyzer and fasten the latches. The main door should remain closed during operation to maintain correct temperature. Figure 13 Location of the Instrument Power Switch

Page 30 Installing Reagents and Standards

6200518 Startup.fm

Section 3
3.3.2 Priming the Analyzer
Note: Do not allow the instrument to run before installing reagents and standards and providing sample.

After internal checks, the analyzer will begin operation. It is important to prime the analyzer because operating without fluid may damage components. Priming the analyzer ensures that the standards and/or reagents fill the tubing going to the valve module. To prime the analyzer, complete the following steps: 1. Press the MENU key to start from the MAIN MENU. 2. Select SENSOR MENU and press ENTER. 3. Select the sensor by name (default is #1 APA6000) and press ENTER. 4. Select SOLUTIONS & CLEANING and press ENTER. 5. Select PRIMING and press ENTER. 6. Select PRIME ENTIRE SYSTEM NOW. This will prime all the lines from the solution bottles and sample lines. The system prime will take approximately 715 minutes. The analyzer will display various messages about the progress of the cycle. 7. A pop-up window will indicate that priming will begin after the current operation cycle. When priming starts, a pop-up window will indicate the time remaining to complete the prime cycle. 8. When priming is complete, the analyzer will automatically begin measuring sample. The analyzer will require 36 hours for readings to stabilize. During this time, the analyzer will establish temperature control. After readings stabilize, a calibration is required to obtain accurate values. See section 4.2.3.1 on page 44.

6200518 Startup.fm

Page 31 Installing Reagents and Standards

Visit http: //www.hach.com

Section 4

Instrument Menus

4.1 The AquaTrend Interface


The AquaTrend Interface integrated into the door of the APA 6000 Analyzer provides the interface necessary to communicate with the analyzer (sensor) and the Hach network, if present. The interface provides a numerical or graphical display and a keypad for user input. A remote AquaTrend can be connected for access to the analyzer from another location. The AquaTrend interface can be used with a number of Hach sensors and therefore, not all of the functions available are used when interfacing with the APA 6000. Functions that are not appropriate for the APA 6000 will not be explained fully in the following text. Data can be viewed in numeric or graphical format and reviewed on trend graphs for details of measurement history. With the addition of the Hach Serial I/O Module (see Replacement Parts and Accessories on page 94), data can be downloaded to a computer or printer.

4.1.1 The AquaTrend Interface Keypad


The AquaTrend Interface keypad is designed for simple navigation through the AquaTrend Interface menus (see Figure 14). The keypad contains four arrow keys, a DISPLAY key, a MENU key, an EXIT key, and an ENTER key. Figure 14 and Table 2 describe how these keys work. Figure 14 AquaTrend Interface Keypad
1 2 4
DISPLAY EXIT

MENU

ENTER

Table 2 AquaTrend Interface Key Functions


Number 1 Key Function of Keys Allows navigation through the various menus, to change settings, and to move the cursor when observing graph data. Table 3 summarizes the functions of the arrow keys in each menu view. Displays the data display screen. Once the data is displayed, pressing the DISPLAY key toggles between data presented in numerical form and graph form. Displays the Main Menu from any menu level or from the data display. From the Main Menu, press the EXIT key to access the data display screen. From any other menu level, press the EXIT key to display the previous menu screen. The EXIT key can also be used to abandon any alphanumeric entry or selection process. Press the ENTER key while in any menu to select menu options and confirm alphanumeric entries.

6200518 Instrument Menus.fm

Page 33 Instrument Menus

Section 4
The arrow icons and their positions in the arrow field in the top right corner of the display (item 7 on Figure 16) indicates the active arrow keys. For example, if three arrow icons are present in the arrow field, pointing up, down, and right, then the up, down, and right arrow keys can be used while viewing that screen. Figure 15 shows examples of the screen categories and Table 3 contains summaries of the arrow key functions for each screen category. Figure 15 AquaTrend Interface Screen Categories

1. Menu Screen

2. Numeric Data Display Screen

3. Graph Display Screen

4. Pop-up Entry Screen

Table 3 Arrow Key Functions


Screen Category Function Use the arrow keys to move the pointer to the next (down) or previous (up) item. For menus no longer than one page, the pointer will loop around from the bottom to the top or from the top to the bottom. For multiple-page menus, the pointer will move to the next (down) or previous (up) page. Left and Right Arrow: Changes the number of channels shown. Press the Right Arrow to increase the number of channels or press the Left Arrow to decrease the number of channels. Up and Down Arrow: Changes the active channel designation. Up Arrow: Displays preceding channel graph. Down Arrow: Displays next channel graph. Left and Right Arrow: Moves curser if it is enabled; if not enabled, changes the graph time base. Up and Down Arrows: Increment (up) and decrement (down) currently highlighted character. Left and Right Arrows: Moves highlight to next (right) or preceding (left) character. Loops around at either end.

1. Menu Screen

2. Numeric Data Display Screen (1 to 8 numeric readings) 3. Graph Display Screen

4. Pop-up Entry Screen

4.1.2 The AquaTrend Interface Display


AquaTrend Interface screens appear in several different formats (see Figure 15). See Figure 16 and Figure 17 for descriptions of each field. When the Main Menu is displayed on the AquaTrend Interface screen, press the EXIT or DISPLAY key to return to the data display mode. Consecutive presses of the EXIT key from any Menu takes you back, one screen at a time, to the data display screen. Pressing DISPLAY from any menu immediately returns you to the measurement display screen. When changing settings or performing diagnostics on a device on the network, temporarily exit the data display mode and enter the Main Menu by pressing the MENU key. The AquaTrend Interface will display data from the analyzer except when the operator uses it for menu setup. When viewing data, press the DISPLAY key to alternate between displaying the data in numeric and graphical mode. When viewing data in numeric mode, you can display a single measurement or up to sixteen measurement channels at the same time. Pressing the right arrow key causes more measurement channels to be displayed; pressing the left arrow key causes fewer measurement channels to be displayed.
Page 34 The AquaTrend Interface

6200518 Instrument Menus.fm

Section 4
When viewing data in graph mode, you can see a history of readings over the last 2 hours, 8 hours, 30 hours, 7 days, or 30 days. The AquaTrend Interface, however, displays data for only one measurement channel at a time in the graphic mode. For details on selecting the preferred graph time and scale, see section 4.2.2.1 on page 37. Figure 16 AquaTrend Interface Menu and Numeric Data Display Fields

Table 4 Display Fields


Number 1 2 3 4 Field Status Bar Text/Data Date Alarms Security Icon 5 6 7 Hourglass Icon Time Navigation Arrow Information Shown in Display Field Numeric Data display: shows DISPLAY DATA, Main menu: shows MAIN MENU Lower menus: shows top level menu (main menu item) Numeric data (116 channels), graphical data, menus, and pop up boxes appear here. Current date Numbers that appear here, 1 through 8, correspond to the sensor with active alarms. If N appears, a Network alarm is active. If no indicators appear in this field, no system alarm is active. See section 5.2.1.1 on page 75 for information on reviewing alarms. When displayed, indicates a security level of Partial or Total When displayed, indicates the datalog is completing a change. The keyboard buffer is cleared when the icon disappears. Current time displayed as military time (24 hour clock). Direction of these arrow icons correspond to active arrow keys for each screen.

6200518 Instrument Menus.fm

Page 35 The AquaTrend Interface

Section 4
Figure 17 AquaTrend Interface Graph Fields
1 16

AquaTrend

15 14 13

5 12 6 11 7
Alarms

10

Table 5 Graph Fields


Number Field 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 Status Bar Channel assignment Graph Data Y-axis upper limit Cursor Y-data Units Species Graph data Y-axis lower limit Graph time X-axis lower limit Cursor Time/Date Graph time X-axis upper limit Graph Cursor Graph Area Sensor name Units and chemical form Navigation Arrows Current Data Information Shown in Graph Field Shows REVIEW MEASUREMENTS Channel number of the measurement shown in the graph Upper numeric limit of trend graph data. (Ymax) Y value of data at cursor position Units of the measurement on the displayed channel Chemical form of the measurement on the displayed channel. Lower numeric limit of trend graph data (Y-min) Lower time limit (oldest data). If data is being reviewed in the 24 hour or 30 day mode, this field is replaced with the date. If data is being reviewed in the 24 hour or 30 day mode, this field is replaced with the date.

Time or date value of data at cursor position Current time (latest data).

Retrieves value of data at cursor position Graph of 60 minutes, 8 hours, 24 hours, 7 days, or 30 days of data. These settings are selectable in the DISPLAY SETUP menu (see section 4.2.2.1). Depending on the sensor type and configuration, the display shows measurement name or the sensor location, or a combination of both. Units and chemical form of the measurement on the channel Direction of these arrow icons correspond to enabled arrow keys for each screens. Most recent measurement (same as numeric display)

Page 36 The AquaTrend Interface

6200518 Instrument Menus.fm

Section 4
4.2 Menu Functions
The first level of menu functions is comprised of Alarm Log, Aquatrend Menu, Sensor Menu, Other Device Menu, and Network Menu. Each level of menus will be discussed in detail below.

4.2.1 Alarm Log


If an alarm warning is active, a number will appear at the bottom of the display in the alarms area and if the FLASH ON ALARM option has been enabled, the entire display will flash (to disable the Flash on Alarm see section 4.2.2.1). The alarms may suggest an over- or under-range condition, or a Sensor Warning. Proceed to the Alarm Log to find out why the alarm was triggered: 1. With the cursor at Alarm Log, press ENTER. 2. Select device for review of current alarms. 3. Select either Warning/Alarm Log or Setpoint Alarm Log. The Warning/Alarm Log provides information on the operation of the analyzer. The SET POINT ALARM LOG provides information about sample measurements that exceed the setpoint values.

The most recent alarm/warning is listed first. A bell icon to the left of the warning indicates the alarm/warning is active. Selecting a specific alarm/warning and pressing ENTER provides more detailed information about that alarm/warning (i.e., time, date, value).

4.2.2 AquaTrend Menu


The following options are provided in the AquaTrend Menu: Display Setup: Controls specific screen functions. Language: Changes the language of the displayed information. Edit Name: Allows the user to change the name of the AquaTrend. Set Time/Date: Sets the time and date for the AquaTrend. Security Setup: Controls access to the analyzer functions. Print Data Log: If a com-link is set up on the network, data can be output to a printer or computer. AT Diagnostics: Allows the user to test various AquaTrend functions and clear the AquaTrend alarm log.

4.2.2.1 Display Setup


The following options are controlled in the Display Setup Menu: Measurement To Channel: Allows assignment of the analyzer measurement (using the user-selected chemical form and units) to a channel. Channel Graph Setting: Controls the range displayed in the graph.

6200518 Instrument Menus.fm

Page 37 Menu Functions

Section 4
Flash On Alarm: Controls whether or not the display will flash when an alarm has been detected. Contrast Level: Controls the level of contrast in the display.

Measurement to Channel
APA 6000 Analyzers can measure from 1 or 2 sample streams, requiring assignment of a measurement channel for each parameter on a stream. For example, parameter options could be called 1Hardness and 2Hardness. In addition, each sample stream may report up to five different parameters. To assign a measurement to a channel, use the MEASUREMENT TO CHANNEL function.

Attaching a Measurement to a Channel


1. While in the Display Set Up Menu, move the cursor to MEASUREMENT TO CHANNEL and press ENTER. 2. Move the cursor to ATTACH and press ENTER. 3. Move the cursor to the sensor whose measurement is to be attached and press ENTER. 4. Select chemistry measurement and press ENTER. 5. Move the cursor to the channel where the measurement is to be displayed and press ENTER. 6. Move the cursor to ATTACH and press ENTER. The message OPERATION COMPLETE will appear. The screen will return to the display setup menu upon completion of the measurement to channel setup.

Detaching a Measurement From a Channel


1. While in the Display Set Up Menu, move the cursor to MEASUREMENT TO CHANNEL and press ENTER. 2. Move the cursor to DETACH and press ENTER. 3. Move the cursor to the sensor to be detached and press ENTER. 4. Move the cursor to the measurement to be detached and press ENTER. The message OPERATION COMPLETE will appear. The screen will return to the display setup menu upon completion of the measurement to channel setup.

Page 38 Menu Functions

6200518 Instrument Menus.fm

Section 4
Channel Graph Settings
To optimize the graphical display, establish the minimum and maximum values to be displayed. For example, if the expected range is 07 ppm, assign 0 as the measurement minimum and 7 as the measurement maximum. 1. Press the MENU key to access the Main Menu. 2. Select AquaTrend Menu and press ENTER. 3. Select DISPLAY SETUP and press ENTER. 4. Select CHANNEL GRAPH SETTINGS from the Display Setup Menu and press ENTER. A list of available channels will appear. 5. Select the channel of interest and press ENTER. 6. Select YMAX and press ENTER. 7. Use the arrow keys to change the displayed values. Press ENTER to accept the value or EXIT to cancel. 8. Repeat step 6 and step 7 to change the YMIN: displayed value. Enter negative numbers by changing the first character to a negative sign with the arrow keys. Setting the Graph Time (x-axis): 1. Press the DISPLAY key until a graph is displayed. 2. Press the up or down arrow keys to select the channel of interest. 3. Press the right or left arrow key to change the time base. The time base will cycle through 2 hours, 8 hours, 30 hours, 7 days, and 30 days. The time scale is shown below the X-axis.

Flash on Alarm
When an alarm is active, the display can flash to bring attention to the alarm condition. Enable/Disable the Flash on Alarm as follows: 1. Press the MENU key to access the Main Menu. 2. Select AquaTrend Menu and press ENTER. 3. Select Display Setup and press ENTER. 4. Move the pointer to Flash On Alarms. YES or NO will be displayed to the right of that line. If the displayed option is correct, press the EXIT key to return to the previous menu level. If the displayed option is not correct, press ENTER. A pop-up window will appear. Use the arrow keys to move the pointer to the desired option and press the ENTER key to accept it. Press the EXIT key to return to the previous menu level.

6200518 Instrument Menus.fm

Page 39 Menu Functions

Section 4
Adjusting the Display Contrast
Adjusting the contrast level setting of the AquaTrend Interface allows the operator to choose the most readable display for the lighting conditions. 1. Press MENU to start from the Main Menu. 2. Select AquaTrend Menu and press ENTER. 3. Select Display Setup and press ENTER. 4. Select Contrast Level. 5. A pop-up window provides a choice of settings between 0 and 24 (0 being the lowest setting). Use the up and down arrow keys to adjust the contrast. Press ENTER to accept the displayed contrast level.

4.2.2.2 Language
English is the default language of the APA 6000 Analyzer. To display another language follow the instructions below: 1. Press the MENU key to start from the Main Menu. 2. Select AquaTrend Menu and press ENTER. 3. Select Language and press ENTER. 4. Use the up or down arrow key to select the appropriate language and press ENTER. All screens will be displayed in this language.

4.2.2.3 Edit Name


The EDIT NAME function may be used to assign a unique name to a device connected to the AquaTrend network. (For example: FAB 3, Boiler_Room, etc.) Refer to the AquaTrend Interface Manual for more information. 1. Press the MENU key to start from the Main Menu. 2. Select AquaTrend Menu and press ENTER. 3. Use the up or down arrow key to select Edit Name and press ENTER. 4. A pop-up window will appear instructing the user to Edit AquaTrend Name. 5. Use the left and right arrow keys to choose the character to change and use the up and down arrow keys to cycle through letters, numbers, and symbols until you reach the character you want.
Note: In order to distinguish between devices on the network, use a unique and descriptive name. The name can be up to 12 characters long in any combination of alphanumeric characters. The first character must be a letter or the # sign.

6. Press ENTER to accept the displayed name.

Page 40 Menu Functions

6200518 Instrument Menus.fm

Section 4
4.2.2.4 Set Time/Date
The current time and date have been factory set on the AquaTrend Interface, but the time may need to be adjusted to your time zone. If it is necessary to reset the time and date, proceed as follows:

Setting the Time


1. Press the MENU key to start from the Main Menu. 2. Use the up/down arrow keys to select AquaTrend Menu and press ENTER. 3. Select Set Time/Date and press ENTER. 4. Select Set Time and press ENTER. 5. A pop-up window will prompt for the new time. The numbers representing the hour will be highlighted. Press the up arrow key to increase the hour setting and press the down arrow key to decrease the setting. When the proper hour is displayed, press the right arrow key and set the minutes in the same manner. 6. The AquaTrend Interface uses a 24-hour clock (military time), so there is no selection for a.m. or p.m. If the current time is 4:00 p.m., for example, you would enter the 24-hour format equivalent of 16:00. 7. Press ENTER to save the change or EXIT to cancel.

Setting the Date


Use this procedure to change the current date: 1. Press the MENU key to start from the Main Menu. 2. Use the up/down arrow keys to select AquaTrend Menu and press ENTER. 3. Select Set Time/Date and press ENTER. 4. Select Set Date and press ENTER. 5. A pop-up window will prompt for the new date. The numbers representing the month will be highlighted. Press the up arrow key to increase the month setting and press the down arrow key to decrease the setting. When the proper month is displayed, press the right arrow key and set the day. Repeat to set the last two digits of the current year. 6. Press ENTER to save the change or press EXIT to cancel.

6200518 Instrument Menus.fm

Page 41 Menu Functions

Section 4
Setting the Correct Day of the Week
Use this procedure to change the current day of the week: 1. Press the MENU key to start from the Main Menu. 2. Select AquaTrend Menu and press ENTER. 3. Select Set Time/Date and press ENTER. 4. Select Day and press ENTER. 5. A pop-up window will prompt for the correct day. Use the up or down arrow key to highlight the desired day. 6. Press ENTER to save the change or press EXIT to cancel. The day will change automatically. Check to be sure the clock is set at the correct time.

Changing the Date Format


Use the following procedure to change the format of the date displayed by the AquaTrend Interface: 1. Press the MENU key to start from the Main Menu. 2. Select AquaTrend Menu and press ENTER. 3. Select Set Time/Date and press ENTER. 4. Select Date Format: and press ENTER. 5. A pop-up window will appear with SELECT DATE FORMAT. Use the up and down arrow keys to select a date format: MM/DD/YY is month/day/year, and DD/MM/YY is day/month/year. 6. Press ENTER to save the change or press EXIT to cancel.

4.2.2.5 Security Setup


To limit access and avoid unauthorized changes to the AquaTrend Interface Network setup features, set security options by: setting a password; setting the security level; or accessing a temporary unlock option. This temporarily suspends the full or partial security levels, allowing access to the system.

The default password is HACH. The default security level is NONE. This security level allows access to all functions. If the password is changed and then is lost or forgotten, contact your Hach representative for further instructions.

Page 42 Menu Functions

6200518 Instrument Menus.fm

Section 4
Changing the System Password
1. Press MENU to start from the Main Menu. 2. Select AquaTrend Menu and press ENTER. 3. Select Security Setup and press ENTER. 4. Select Set Password and press ENTER. 5. Enter the current password for access to the AquaTrend Interface menus. Pressing the right and left arrow keys moves the cursor right and left; pressing the up and down arrow keys scrolls through the available characters. The default password is HACH. 6. Press ENTER to confirm or press EXIT to cancel. 7. Enter the new password by using the arrow keys as in step 5. 8. Re-enter the new password.

Setting the Security Level


The analyzers security level determines which menu parameters can be modified by the operator. The security levels are NONE, PARTIAL, and TOTAL. TOTAL lockout means the operator is unable to change any of the analyzers settings or cause the analyzer to perform any operations (priming, cleaning, setting fluid levels, calibrating, etc.) from the keyboard. Any user will be able to review all settings, but will be unable to modify them if they are locked out. PARTIAL lockout allows user access to the following menus and functions: display setup language alarm log execute calibration (see section 4.2.3.1) all priming operations setting all fluid levels clean system now

Setting the security level to NONE allows access, without a password, to any of the menus except SECURITY SETUP.

6200518 Instrument Menus.fm

Page 43 Menu Functions

Section 4
Set the security level as follows:
1. Press the MENU key to start from the Main Menu. 2. Select AquaTrend Menu and press ENTER. 3. Select Security Setup and press ENTER. 4. Select Security Level: and press ENTER. 5. Enter the password (default is HACH). Use the left and right arrow keys to choose the character to change and use the up and down arrow keys to cycle through letters, numbers and symbols until you reach the character you want. 6. Use the arrow keys to select NONE, PARTIAL, or TOTAL. 7. Press ENTER to confirm or press EXIT to cancel.

Using Temporary Unlock and Unlock Duration


Selecting temporary unlock allows the operator to temporarily override the security level setting and have full access to all menu parameters. Unlock duration allows the operator to select the length of time that the temporary unlock is in effect. 1. Press the MENU key to start from the Main Menu. 2. Select AquaTrend Menu and press ENTER. 3. Select Security Setup and press ENTER. 4. Select Temporary Unlock and press ENTER. 5. Enter the password (this is necessary only if the security level is set to Partial or Total). 6. Choose the duration of time you want the keypad to remain unlocked: 30, 60, 120, or 360 minutes. 7. Press ENTER to confirm or press EXIT to cancel.

4.2.3 Sensor Menu


4.2.3.1 Calibration
Calibration prepares the APA analyzer to calculate concentrations in unknown samples based on a calibration generated using two known standards. These two standards are the high and low standard solutions. For immediate trend analysis, an electronic factory default calibration is already established and allows the analyzer to provide approximate measurements.The once-a-day (at 4:00 a.m.)* auto-calibration interval may be changed to every 12 hours, a specific day of the week, or may be turned off. It can also be calibrated on demand as necessary.

* APA6000 Alkalinity and HR Hardness instruments defaults are each Monday at 4:00 a.m. APA6000 HR Hardness instrument default is each Monday at 4:00 a.m. Page 44 Menu Functions

6200518 Instrument Menus.fm

Section 4
Reducing the frequency of the auto-calibration will reduce standard usage, but it may also reduce measurement accuracy. Hach recommends that the analyzer be calibrated when any reagent or standard is changed or as often as necessary to achieve acceptable results. Calibration should only be performed after the analyzer has been turned on for more than five hours. This allows the temperature inside the analyzer to stabilize.

Setting Up the Auto-Calibration Schedule


The analyzer accuracy specifications are based on the factory-set default interval. The APA 6000 Analyzer can calibrate at any time. To change the calibration time, use the auto-cal setup procedure: 1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER. 4. Select Calibration and press ENTER. 5. Select Auto-Cal Setup and press ENTER. 6. The analyzer calibration is a factory set default interval. In order to change the settings, select one of the following: Select AUTO-CAL to turn the automatic calibration option off, or to set the day or hourly interval you want the instrument to calibrate. Choose any day of the week, every day of the week, or every 12 hours (as shown below). Select Auto-Cal Time to specify the clock hour you want the instrument to calibrate (use a 24-hour clock). Use the up and down arrow keys to enter an hour between 1 and 24. For example, if you want the instrument to calibrate at 2:00 p.m., enter 14.

Note: When auto-calibrate is set to 12 hours, the analyzer calibrates at the set auto-cal time and again 12 hours later.

7. Press ENTER to confirm or press EXIT to cancel.

Execute Calibration
To start a calibration manually, proceed as follows. The instrument will finish its current cycle, then perform the calibration. 1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER. 4. Select Calibration and press ENTER. 5. Select Execute Calibration and press ENTER. 6. The CALIBRATION PENDING message will appear.

6200518 Instrument Menus.fm

Page 45 Menu Functions

Section 4
The analyzer will wait until any current measurement or priming cycle is complete before calibrating. Calibration typically takes 1 hours to complete.

Viewing the Calibration History


This option allows you to review calibration data of past calibrations. 1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER. 4. Select Calibration and press ENTER. 5. Select Cal History and press ENTER. 6. A pop-up window will appear with the calibration date and time of the seven most recent calibrations (see the parameter specific operation manual for specific example). Select Review Next Cal to step through the pages of the calibration history. After the last calibration, press ENTER to return to the previous menu. To exit the Cal History Menu, press EXIT. Data displayed for the seven most recent calibrations will include the standard concentrations, and the date and time of the calibration. Please see the parameter-specific operation manual for specific information. Executing the Default Calibration If the analyzer is giving incorrect readings, it can be reset to use the default calibration. The default calibration allows the analyzer to give good approximations of the measurement value. To ensure that measurements meet analyzer accuracy specifications, follow the default calibration with a manual calibration as described above. To set the analyzer to its default calibration, follow these steps: 1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER. 4. Select Calibration and press ENTER. 5. Select Default Calibration and press ENTER. 6. Select Yes and press ENTER to set the analyzer to its default calibration.
Note: To exit without restoring the default calibration, select No and press ENTER or press EXIT.

Page 46 Menu Functions

6200518 Instrument Menus.fm

Section 4
4.2.3.2 Alarm Setup
Each APA 6000 Analyzer provides 14 alarm signals for your system. Two of these, ALA and ALB, are permanently attached to the analyzers internal alarm outputs, labeled Relay A and Relay B. The remaining alarm signals, AL1 through AL12, can be attached to relays in optional Signal Output Modules (SOMs). SOMs can be purchased separately, and provide two relays and two analog outputs each. For information see the SOM Manual (Cat. No. 51250-18). Each of the alarm setup options provides a setting for measurement, enabling/disabling alarms, alarm type, and alarm set point. Alarms can be configured as one of six types: high, low, rate (in mg/L/hour units), sensor warning, sensor alarm, or unconfigured. Table 6 describes the parameters available on each alarm type. Parameters in bold print indicate default settings. Table 6 Parameters for Setting Alarms
Options TYPE MEASUREMENT SET POINT ENABLED MEAS FAILURE Set point Alarms High, Low, Rate, Unconfigured [Sensor measurement] [Legal value] Yes, No Off, On, Hold Sensor Warnings and Alarms Sensor Warning, Sensor Alarm, Unconfigured ----Yes, No ---

To set up the alarms:


1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER. 4. Select Alarm Setup and press ENTER. 5. Select the alarm to set up and press ENTER. ALA and ALB are internal analog outputs connected to relays in the APA analyzer. AL1 through AL12 may be connected to relays in an SOM 6. Select Type and press ENTER. 7. Select one of the following to choose the type of alarm: To disable an alarm, select Unconfigured and press ENTER. Select High, Low, or Rate to configure a set point alarm within those ranges and press ENTER. Go to step 8. Select Sensor Warning or Sensor Alarm to configure an alarm that will be triggered by a sensor alarm or warning. Go to step 9.

8. If High, Low or Rate was selected in step 7, the following setup options are available:

6200518 Instrument Menus.fm

Page 47 Menu Functions

Section 4
Select Measurement to change the sample measurement that is monitored for alarm generation. Select Set Point to specify a numeric value for the alarm. Select Enabled to turn the alarm on or off. Select Meas Failure to set the alarms response when sample is not flowing to the analyzer or the analyzer experiences a fault that prevents it from providing the selected measurement. On and Off turn the alarm on or off; Hold leaves the alarm in the state it was in prior to the failure.

9. If you selected Sensor Warning or Sensor Alarm in step 7, select Enabled. Select Yes or No to turn the alarm on or off, respectively. 10. Press ENTER to save the alarm setting, or press EXIT to cancel.

4.2.3.3 Recorder Setup


Each APA 6000 analyzer provides up to 14 analog output signals. Two of these signals, RCA and RCB, are permanently attached to the analyzers internal analog outputs, labeled Output A and Output B. RCA and RCB may also be configured as recorder outputs. The remaining analog signals, numbered RC1 through RC12, can be attached to analog outputs in SOMs. SOMs can be purchased separately and provide two relays and two analog outputs each (see the AquaTrend Interface Manual for networking information). Table 7 describes the parameters available for recorders and controllers. Parameters in bold print indicate default settings. Table 7 Parameters for Setting Recorders
Options TYPE MEAS MIN MAX ENABLED MEAS FAILURE OUTPUT SETUP Recorder Recorder, Unconfigured [Sensor measurement] [value] (As a recorder output, this value is always shown in the default units of mg/L regardless of what is selected and displayed on the AquaTrend) [value] (As a recorder output, this value is always shown in the default units of mg/L regardless of what is selected and displayed on the AquaTrend) Yes, No Go Max, Go Min., Hold Only RCA and RCB

Configuring Analog Outputs as Recorder Signals


To set up an analog output to a recorder, perform the following steps: 1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER. 4. Select Recorder Setup and press ENTER. 5. Select the appropriate analog output to use and press ENTER. Select RC [1 -12] to configure a SOM output to a recorder. Select RCA or RCB to configure one of the APA analog outputs as a recorder output.
Page 48 Menu Functions

6200518 Instrument Menus.fm

Section 4
6. To change the type of output, select TYPE and press ENTER. 7. Select Recorder to send the analog output to a recorder 8. Press ENTER to confirm the setting or press EXIT to cancel. 9. To change the measurement that the recorder output will track, select Measurement: XXXX and press ENTER. 10. Select the desired measurement and press ENTER. 11. To change the minimum measurement value that corresponds to the minimum recorder output, select Min: XXXX and press ENTER. 12. Modify the minimum value by using the arrow keys and press ENTER when the value entry is complete. 13. To change the maximum measurement value that corresponds to the maximum recorder output, select Max: XXXX and press ENTER. 14. Modify the maximum value by using the arrow keys and press ENTER when the value entry is complete. 15. To enable or disable the recorder output, select Enabled and press ENTER. 16. Select Yes or No to indicate that the recorder output is enabled or disabled, respectively, and press ENTER. 17. To change the recorder output action when the analyzer is unable to perform measurements in the selected measurement range, select MEAS Failure: XXXX and press ENTER. 18. Select Hold, Go Max, or GO Min to indicate what the output action should be when the analyzer cannot perform a selected measurement. Go Max sets the output to 20 mA, Go Min sets the output to 4 mA, and HOLD keeps the output at the current value. Press ENTER. 19. To complete the recorder setup for RCA or RCB see the following section. To complete the recorder setup for RC112 see Analog Output Settings in the SOM Manual.

Adjusting Output for RCA and RCB (Trim)


Use the trim adjustments to set zero or full scale output levels. The trim adjustments allow correction for recorder response without the need to adjust the recorder itself. The range of adjustment is 0.011 mA for zero scale and 13.222.5 mA for full scale. The current values are nominal values and are accurate to 1.5%. The recorder full scale (20 mA) may not exactly match the APA full scale (20 mA). You can adjust the outputs so that full scale on the APA is equal to full scale on the recorder. The same holds true for the zero value.

6200518 Instrument Menus.fm

Page 49 Menu Functions

Section 4
To adjust (trim) the milliamp output range for the internal recorder/controller outputs (RCA or RCB) or to change the range: 1. Select Output A Setup or Output B Setup. Press ENTER. 2. To adjust the mA output at the bottom of the scale: a. Select Zero Scale Trim and press ENTER. b. Use the up/down arrow to select Fine or Coarse Adjustment. Press ENTER. The new trim mA value will appear. c. Select Accept New Value when done and press ENTER. 3. To adjust the mA output at the top of the scale: a. Select Full Scale Trim and press ENTER. b. Use the up/down arrow to select Fine or Coarse Adjustment. Press ENTER. The new trim mA value will appear. c. Select Accept New Value when done and press ENTER 4. To select either 020 mA or 420 mA as the output range, select Output Range and press ENTER.

4.2.3.4 Measurement Options


Selecting Measurement Units
The selection of units that each sensor can display is parameter specific. To view the selection and change the display, follow the directions below.
Important Note: Changing units causes the data log for the measurement to be erased.

1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER. 4. Select Measurement Options and press ENTER. a. Select Chemical Form and press ENTER. Select the desired chemical form and press ENTER. b. Select Sample Sequence and press ENTER. Assign a sample sequence and press ENTER. c. Select Display and assign the reading to the appropriate DDM if desired. Press ENTER. d. Select Units and press ENTER. Select Hold Outputs and press ENTER. Select the units you want to use and press ENTER, or press EXIT to cancel.
Note: When units are changed, the display shows new units with the next measurement.

Page 50 Menu Functions

6200518 Instrument Menus.fm

Section 4
4.2.3.5 Enabling Solution Level Monitoring
Solution level monitoring by default is set to Off. The display will show Level Monitor: On/Off. When On is chosen (solution monitoring enabled), solution levels will be displayed as 0 to 100% full. To turn Solution Level Monitoring on or off proceed as follows: 1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER. 4. Select Solutions and Cleaning and press ENTER. 5. Select Level Monitor: On/Off and press ENTER. a. If monitoring is off the display will show: Turn on Level Monitor Levels Will Start @ 100% b. Select Yes and press ENTER to enable monitoring or select No to exit. c. If monitoring is on, the display will show: Turn off Level Monitor. Select Yes and press ENTER to disable or select No to exit.

4.2.3.6 Setting Solution Levels


1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER. 4. Select Solutions and Cleaning and press ENTER. 5. Select Solution Levels and press ENTER. 6. The current % Full is displayed for each reagent. 7. If all reagents have been cleaned and refilled, select Reset All to 100% and select ENTER. 8. If all reagents have not been cleaned and refilled to 100%, select the solution(s) value to be changed and press ENTER. 9. A pop-up box will show the default setting of 100%. Use the up and down ARROW keys to change the setting to reflect the amount of solution in the bottle. 10. Press the ENTER key to accept the value or press Exit to Cancel.

6200518 Instrument Menus.fm

Page 51 Menu Functions

Section 4
4.2.3.7 Solutions & Cleaning Options
There are two cleaning options, Clean System Now and Auto-Clean. Clean System Now will start the clean cycle as soon as the current operation finishes. The Auto-Clean option allows the user to change the frequency of the cleaning cycle. 1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER. 4. Select Solutions & Cleaning and press ENTER. 5. Select Instrument Clean and press ENTER. 6. Select Auto-Clean and press ENTER. Select the Automatic Clean Frequency and press ENTER.
Note: Using the default settings is strongly recommended by the manufacturer and should only be changed if sample components cause measurement problems. If cleaning frequency is decreased, verify that the cleanliness and operation of the analyzer is not degraded. Changing the cleaning frequency may interfere with the autocalibration.

4.2.3.8 Sensor Diagnostics


The diagnostic menu provides options to help the operator diagnose and solve potential problems.

Hold Outputs
Holds all outputs at the current level for 30 minutes, until this function is manually cancelled or until power is cycled off/on. 1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER. 4. Select Sensor Diagnostics and press ENTER. 5. Select Hold Outputs and press ENTER. 6. Select Hold Outputs for 30 Min or Abort Output Hold and press ENTER.

Internal Relay Test


Turns the relays on and off. 1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER.

Page 52 Menu Functions

6200518 Instrument Menus.fm

Section 4
4. Select Sensor Diagnostics and press ENTER. 5. Select Internal Relay Test and press ENTER. 6. Select A or B and press ENTER. 7. Select Test Open Relay or Test Closed Relay and press ENTER.

Internal Output Test


Sends the 420 signal either high, low, or mid-scale and the result can be monitored on an attached recorder. 1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER. 4. Select Sensor Diagnostics and press ENTER. 5. Select Internal Output Test and press ENTER. 6. Select Analog Out A or Analog Out B and press ENTER. 7. Select Zero, Half, or Full Level test and press ENTER.

Sensor Warning Options


This option allows you to enable/disable warnings generated by power failures. The default setting for power failure warning is OFF. 1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER. 4. Select Sensor Diagnostics and press ENTER. 5. Select Sensor Warning Options and press ENTER. 6. Select Power Failure and press ENTER. A pop-up menu will appear. 7. Select ON or OFF to enable or disable the warning. Press ENTER.

Default Calibration
Restores the factory default calibration curve. 1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER. 4. Select Sensor Diagnostics and press ENTER. 5. Select Default Calibration and press ENTER.
Page 53 Menu Functions

6200518 Instrument Menus.fm

Section 4
6. A pop-up window will appear. 7. Select Yes or No and press ENTER.

Sample Streams
This option allows the user to add or delete a second sample stream to the analyzer. 1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER. 4. Select Sensor Diagnostics and press ENTER. 5. Select Sample Streams and press ENTER. 6. A pop-up window will appear. 7. Select One or Two and Press ENTER.

Clear Alarm Log


Selecting the Clear Alarm Log option lets the operator delete past alarms and warnings from the sensor alarm log. To clear the alarm log: 1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER. 4. Select Sensor Diagnostics and press ENTER. 5. Select Clear Alarm Log and press ENTER. 6. Select Yes and press ENTER to clear the alarm log.

Show Version Number


Displays the version of the currently loaded software.

SIO1 and SIO2 Setup


The APA 6000 sends out measurement readings, alarms, and warnings through the optional SIO module. This section instructs how to select a particular format. 1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name. 4. Scroll to and select Sensor Diagnostics and press ENTER. 5. Scroll to and select SIO1 SETUP or SIO2 SETUP and press ENTER.
Page 54 Menu Functions

6200518 Instrument Menus.fm

Section 4
Note: Note the selections under SIO1 SETUP and SIO2 SETUP are the same. Up to two SIOs can be connected to the APA 6000. If installing a second SIO or the analyzer is connected to the second SIO on the network, select SIO2 SETUP.

If Computer format is selected and Output is set to Auto, the readings, alarms, and warnings will be sent out as they occur. If Output is set to Queried, the information is sent only when a query command is sent from the computer. The Time/Date will be sent with readings if Yes is selected and can be turned off by setting to No. If Printer Format is selected, the Output and Time/Date options are ignored and not shown in the menu. See Appendix B on page 85 for more information on printer and computer formats.

Sensor Status
The sensor status function indicates if the APA 6000 is in Standby or Online. The analyzer will go into Standby on a system Alarm condition. This menu also contains a Restart function. When this function is used, the sensor status reverts to Online when the restart procedure is complete.

4.3 Other Device Menu and Network Menu


These two menus are used when employing a network of multiple instruments. If your APA 6000 is not operating on a network, these selections will not apply. If building an AquaTrend network of multiple analyzers, refer to the AquaTrend Interface Manual for these menus.

4.4 Data Display


4.4.1 Numerical Display
While the AquaTrend Interface can display data from up to 16 sensor channels the most it can display at one time is eight. 1. Access the display mode by pressing the DISPLAY key. 2. Use the right arrow key to increase the number of channels displayed on the screen. 3. Press the left arrow key to decrease the number of channels displayed. Use the up and down arrow keys to move up or down through the displayed channels. The most current measurement is displayed for each channel. If a current measurement for a channel is not available, the display will show - - - -. When viewing more than one channel, the display will not identify the sensor that is sending measurement data. If the APA is not networked, the measurement channel is associated only with that specific APA. If the APA is networked, the display may show the same channel numbers displayed, even though the channels show data from different analyzers. For example, two 1HARDNESS channels may be displayed. To identify the instrument that is sending the displayed data, highlight the measurement channel and press ENTER. The display will show all the information for that measurement. Press ENTER again to return to the multiple display screen.
Page 55 Other Device Menu and Network Menu

6200518 Instrument Menus.fm

Visit http: //www.hach.com

Maintenance

DANGER
Handling chemical samples, standards, and reagents can be dangerous. Review the necessary Material Safety Data Sheets and become familiar with all safety procedures before handling any chemicals.

DANGER
La manipulation des chantillons chimiques, talons et ractifs peut tre dangereuse. Lire les Fiches de Donnes de Scurit des Produits (FDSP) et se familiariser avec toutes les procdures de scurit avant de manipuler tous les produits chimiques.

PELIGRO
La manipulacin de muestras qumicas, estndares y reactivos puede ser peligrosa. Revise las fichas de seguridad de materiales y familiarcese con los procedimientos de seguridad antes de manipular productos qumicos.

GEFAHR
Das Arbeiten mit chemischen Proben, Standards und Reagenzien ist mit Gefahren verbunden. Es wird dem Benutzer dieser Produkte empfohlen, sich vor der Arbeit mit sicheren Verfahrensweisen und dem richtigen Gebrauch der Chemikalien vertraut zu machen und alle entsprechenden Materialsicherheitsdatenbltter aufmerksam zu lesen.

PERICOLO
La manipolazione di campioni, standard e reattivi chimici pu essere pericolosa. La preghiamo di prendere conoscenza delle Schede Techniche necessarie legate alla Sicurezza dei Materiali e di abituarsi con tutte le procedure di sicurezza prima di manipolare ogni prodotto chimico.

6200518 Operation Stopper.fm

Page 57 Maintenance

Visit http: //www.hach.com

Section 5
5.1 Maintenance

Maintenance and Troubleshooting

5.1.1 Shutting the Analyzer Down (Off or Standby)


Before changing solutions or inspecting/replacing parts, place the analyzer in standby then turn the instrument off. To turn the instrument off, open the front door and locate the on/off switch near the top of the analyzer. Move the switch to the off position. Placing the instrument in standby leaves the analyzer in a known condition so restarting begins at a consistent place. If the analyzer will be shut down for several days, it is best to run a clean cycle just before shutdown to fill all the wetted parts with cleaning solution and help prevent fouling/biogrowth in the system. See section 5.1.2.1 on page 60 for cleaning instructions. To place the analyzer in standby or re-start it:
Note: Putting the analyzer into Standby Mode terminates all operations. When the analyzer is restarted, it will automatically start performing measurements.

1. Press the MENU key. 2. Select Sensor Menu, then press ENTER. 3. In the Select Sensor Menu, select the sensor name and press ENTER. 4. Select Sensor Status and press ENTER. A pop-up window will appear. Use the arrow keys to select Standby or Restart. Press ENTER If the analyzer will be shut down for an extended period of time, all reagents, standards, and samples should be flushed out of the system. This is accomplished by supplying deionized water in place of all reagents, standards, and samples and performing a Prime Entire System function.

5.1.2 Scheduled Maintenance


The following tables specify requirements for one, three, and six month maintenance. If any of the items are in need of repair at the time of the scheduled maintenance check, follow the direction under the Maintenance Required column.
Important Note: Failure to perform monthly maintenance on the autoburette module will void the warranty.

Table 8 Monthly Maintenance


Item Reagents and standards Cleaning solution* Autoburette module Valve module Sample conditioning filter* Mixer module Maintenance Required Replace reagents and standards. Make sure the reagent compartment and tubing is clean. Make sure cleaning solution container is full. Inspect for seal or fitting leaks. Inspect for particulate build up. Lubricate lead screw and ceramic piston guide, see Figure 26 on page 70 and Figure 27 on page 71. Inspect module and associated tubing for leaks. Inspect the filter and check sample flow. Clean or replace the filter. Inspect for particulate build up.

* Check/replace filter and fill cleaning solution every two weeks in typical municipal wastewater applications.

6200518 Operation.fm

Page 59 Maintenance and Troubleshooting

Section 5
Table 9 Three-Month Maintenance
Item Autoburette module Valve module Maintenance Required Replace piston seals and O-rings. Replace valve rotor and inspect condition of stator (dry the stator before inspection); if scored, replace stator. Check for leaks. Replace as needed.

Table 10 Six-Month Maintenance


Item Autoburette module Tubing and fittings Maintenance Required Check for need to replace piston seals. Inspect for signs of leakage. Inspect for leaks or damage. Replace as needed.

5.1.2.1 Cleaning the Instrument


To clean the outside of the instrument, wipe with a damp cloth as needed. Be sure to clean up all reagent spills as soon as possible with a dry cloth, followed by a damp cloth. To clean the wetted parts of the analyzer, the system default is set to run a normal clean cycle every day. However, this option can also be used in other circumstances. During operation air bubbles may develop inside the autoburette, tubing, or the colorimeter. This is especially true of new instruments where system components are not thoroughly wetted. In addition, clean samples may require more cleaning cycles because clean water does not contain enough surfactants to keep the system thoroughly wetted. To clear bubbles from the system, use the Clean function. If bubbles persist in a new analyzer, use the Cleaning Solution (Cat. No. 26974-53) as described in section 4.2.3.7 on page 52. The cleaning option dispenses the cleaning solution through all the wetted parts in the analyzer. The cleaning operation has several functions: It is used to prime the system (cleaning solution is pulled directly into the autoburette). It is used routinely to ensure the wetted parts stay free of contaminant buildup. It is used to aid in the removal of air bubbles retained in the system.

5.1.2.2 Sample Conditioning


All samples are conditioned using the sample conditioning filter. Sample conditioning eliminates particles greater than 0.5 m. Regular maintenance of the sample conditioning unit is required for proper operation. For samples that tend to clog the analyzer filter or tubing, the following sample conditioning options are available: Increase the filter replacement frequency. Increase the flow to the sample conditioning block. This will improve the self-cleaning of the filter.

Page 60 Maintenance

6200518 Operation.fm

Section 5
Filtered sample is aspirated through the filter as needed. The instrument is not affected by how fast the sample is flowing through the block as long as the flow is between 1000 and 2000 mL per minute. If the filter becomes plugged, replace it with one of the spare filters supplied with the unit (Cat. No. 62011-00). To remove the filter from the sample conditioning block: 1. Turn off the sample flow and place the analyzer in standby mode.
Note: Sample will drain from the block when the filter holder is removed. Be prepared to dispose of this appropriately.

2. Remove the filter holder from the block by unscrewing the large knurled nut on the bottom of the block (see Figure 18). 3. Remove the quad-ring at the top of the filter assembly (see Figure 19). 4. Slide the filter off the top of the assembly. 5. Install a new filter. 6. After the new filter is in place, replace the quad-ring and screw the filter assembly into the sample conditioning block. 7. Turn sample flow on and restart the analyzer.
Note: The manufacturer does not recommend cleaning and reusing clogged filters.

Depending on your sample, the filter may quickly become plugged. The APA 6000 Micro Filter System (Cat. No. 51339-00, 115 V or Cat. No. 51339-01, 230 V) may be purchased to provide longer periods of operation between maintenance intervals. Figure 18 Removing the Sample Conditioning Block

Sample Conditioning Block

Unfiltered sample OUT

Unfiltered sample IN

Knurled Filter Holder

6200518 Operation.fm

Page 61 Maintenance

Section 5
Figure 19 Removing the Filter

Remove the Quad-Ring from the top of the Filter Assembly.

Slide the filter off the top of the assembly.

Quad-Ring (Cat. No. 62012-00)

Filter (Cat. No. 62011-00)

Filter Holder (Cat. No. 62019-00)

5.1.2.3

Changing Modules
Refer to Figure 20 and follow the instructions below to replace a module:
Note: Chemicals are present in the pump, valve, and reagent tubing. Before doing any maintenance, review all MSDSs, and use recommended safety measures to prevent exposure to potential chemical hazards.

1. Put the analyzer in standby (see section 5.1.1 on page 59). This will extend the piston in the autoburette body and put the valve in a known position. Turn off power to the instrument by toggling the on/off switch at the top center of the module panel to off (O). 2. Break the siphon of any reagent/cleaning tubing that is connected to the module by removing the tubing from the bottles (see section 3.3 on page 29). Shut off the sample stream to the analyzer. 3. Remove all fittings from the module. 4. Loosen the two captive screws that secure the module to the instrument panel with a Phillips screwdriver. 5. Pull the module straight out to remove. 6. Reverse the process to reinstall the module.

Page 62 Maintenance

6200518 Operation.fm

Section 5
Figure 20 Part Identification for Module Removal

5.1.2.4 Changing Tubing


CAUTION To familiarize yourself with handling precautions, dangers and emergency procedures, always review the Material Safety Data Sheets prior to handling containers, reservoirs, and delivery systems that contain chemical reagents and standards. Protective eye wear is always recommended when contact with chemicals is possible. PRUDENCE Pour se familiariser avec les prcautions prendre lors de la manipulation, les dangers et les procdures d'urgence, toujours lire les Fiches de Donnes de Scurit des Produits avant de manipuler les rcipients, les rservoirs et les systmes de distribution contenant les ractifs chimiques et les solutions talons. Il est toujours recommand de porter des lunettes de protection lorsqu'un contact avec les produits chimiques est possible. CUIDADO Para familiarizarse con las precauciones de manipulacin, los peligros y los procedimientos de emergencia, siempre estudie las Hojas de Datos de Seguridad de los Materiales antes de manipular recipientes, depsitos y sistemas de entrega que contengan reactivos y patrones qumicos. Siempre se recomienda el uso de protectores oculares cuando sea posible el contacto con productos qumicos. VORISCHT Es wird dringend empfohlen, die Sicherheitsdatenbltter vor der Handhabung von Behltern, Tanks und Zufuhrsystemen, die chemische Reagenzien und Standardsubstanzen enthalten, aufmerksam durchzulesen, damit Sie sich mit den beim Umgang mit diesen Chemikalien notwendigen Vorsichtsmanahmen, Risiken und Notfallschutzmanahmen vertraut machen, Es wird empfohlen, in allen Situationen, in denen mit einem Kontakt mit Chemikalien zu rechnen ist, eine Schutzbrille zu tragen. ATTENZIONE Per familiarizzare con misure di precauzione di controllo manuale, con procedure pericolose e di emergenza, rileggere sempre le schede di sicurezza del materiale prima di maneggiare contenitori, imballaggi per spedizione che contengono reagenti chimici e standards. Si consiglia di indossare sempre gli occhiali protettivi quando possibile un contatto con agenti chimici.

6200518 Operation.fm

Page 63 Maintenance

Section 5
Tubing replacement will be required if a tube becomes crimped or kinked and is interfering with fluid flow or the tubing becomes fouled due to sample conditioning upset (or improper sample conditioning).

Replacing a single tube:


1. Isolate the tube to be replaced from the others it is grouped with (see the Plumbing Diagram in the parameter-specific manual). If it is an external tube (to reagent containers, sample inlet, or drain) remove the tubing cover from the back, underside of the instrument case. 2. Remove the fittings on either end of the tube from their ports. Note whether it is a -28 flat bottom fitting or a 10-32 rotary valve fitting. 3. Remove the fittings from the ends of the tubing and discard. 4. Prepare a section of tubing the appropriate length, according to Tubing Replacement Length Table in the parameter-specific manual.
Important Note: Tube length is critical to proper instrument function. Replacement tubing must be cut to the length specified.

5. If the tube is an external tube, slide the rubber grommet onto the tubing and insert into the appropriate position in the manifold on the inside bottom of the instrument case. 6. Route the tubing to its appropriate place on the outside and inside of the instrument (refer to Plumbing Diagram in the parameter-specific manual). 7. Install the fitting onto the tubing, as shown in Figure 21 or Figure 22. 8. Tighten fittings in the ports.
Warning: DO NOT over-tighten the fittings. Fittings should be firmly finger tightened. Loose fittings will result in air or liquid leaks and poor analyzer performance. In some case, a tool may be required to sufficiently tighten a fitting.

Figure 21 #10-32 Fitting Assembly (Valve Stator)


Ferrule Tubing

#10-32 Nut

Page 64 Maintenance

6200518 Operation.fm

Section 5
Figure 22 -28 Fitting Assembly

Ferrule and Locking Ring (see Detail A)

1/4-28 Flangeless Nut

Tubing

Chamfered End

Flat End
Chamfered End of Locking Ring Flattened End of Locking Ring

Detail A

5.1.2.5 Rotary Valve Module


The rotary valve directs fluids to various components in the instrument. All tubing is factory-assembled and labeled for each individual chemistry. The valve head and holding coil are enclosed in a temperature-controlled metal housing.

Maintaining the Valve Rotor and Stator


It may be necessary to clean inside the rotary valve and or inspect the surface of the rotor and stator. If the finish of the these two surfaces become scored from normal wear or scratched by debris, analyzer performance will deteriorate.

Servicing the Rotary Valve


Refer to Figure 23 and these instructions: 1. Place the instrument in standby, then turn it off (see section 5.1.1 on page 59). 2. Shut off the sample flow. 3. Break the siphon to the reagent bottles and standard bottles as follows: Disconnect the reagent line from the reagent vent cap insert. Lift the reagent line so the bottom of the reagent line is above the liquid level in each reagent/standard bottle. Re-collar the reagent line in the vent cap insert to prevent it from moving around.
Note: Avoid contaminating the standards.

4. Before removing the valve head, remove tubing to ports 4, 7, and 10 for wrench clearance (see Figure 24). There is no need to drain sample line 3 as the sample was turned off in step 2. The line to port 7 is the waste line.

6200518 Operation.fm

Page 65 Maintenance

Section 5
5. Using a 7/64-in. Allen wrench, loosen the three recessed socket head cap screws in the valve stator (valve head). Do not completely remove the screws, but back them out far enough to remove the head. These screws will help hold a ring spacer that lies under the valve head or stator. Be careful not to lose this spacer. Pull the valve head away from the valve module and inspect surfaces of the rotor and stator.
Note: Do not the scratch the surfaces of the rotor and stator. Note: A small amount of fluid may drain from the stator. If the volume is large, be sure the siphon to the reagent bottles was broken.

6. Inspect the slot in the rotor for any plastic shavings or sample debris. If these are visible, remove the rotor from its holder for cleaning. Carefully clean this slot with a small stylus, or a soft bottle brush, taking care not to scratch the valve surface. 7. Dry and then inspect the valve surface of the valve head/stator. If this surface is scored, replace the valve head along with the rotor. Figure 23 Valve Rotor Part Identification

7/64" Allen Cap Screw 3 Places


10
5
6
1

Valve Stator
7

8
4

Page 66 Maintenance

Spacer Valve Rotor Valve Body

6200518 Operation.fm

Section 5
Figure 24 Removing Screws From the Valve Stator

Reinstallation
1. Before reinstalling the rotary valve, dry all surfaces with paper toweling. Reposition the valve head with port 1 at 12 oclock (top center) on the valve body. Make sure the locating pins in the valve body engage the locating pin holes on the back side of the valve head/stator. 2. Hold the valve stator in place as you engage the three socket head cap screws. Take care to evenly tighten these three screws. You will feel an even resistance as the valve stator tightens down. This is because a spring washer will be compressed during the rotor/stator mating. Tighten these screws in even increments until tight (100 to 120 inch-ounces of torque or 0,7 to 1,4 Joules of torque). 3. Re-attach lines 4, 7, and 10 to the valve head/stator. If the valve stator is replaced, connect the remaining lines to the new stator.
Note: Do not over-tighten these valve fittings. These fitting are finger-tight fittings. Torque these fitting to no more that 18 inch-ounces (0.13 Joule).

4. Return the raised reagent lines of the reagent/standard bottles to their original positions. 5. Start the analyzer. Go to the Prime Menu and select the Prime Entire System option. Once readings stabilize, recalibrate the instrument.

6200518 Operation.fm

Page 67 Maintenance

Section 5
5.1.2.6 Maintaining the Autoburette Module
CAUTION To familiarize yourself with handling precautions, dangers and emergency procedures, always review the Material Safety Data Sheets prior to handling containers, reservoirs, and delivery systems that contain chemical reagents and standards. Protective eye wear is always recommended when contact with chemicals is possible. PRUDENCE Pour se familiariser avec les prcautions prendre lors de la manipulation, les dangers et les procdures d'urgence, toujours lire les Fiches de Donnes de Scurit des Produits avant de manipuler les rcipients, les rservoirs et les systmes de distribution contenant les ractifs chimiques et les solutions talons. Il est toujours recommand de porter des lunettes de protection lorsqu'un contact avec les produits chimiques est possible. CUIDADO Para familiarizarse con las precauciones de manipulacin, los peligros y los procedimientos de emergencia, siempre estudie las Hojas de Datos de Seguridad de los Materiales antes de manipular recipientes, depsitos y sistemas de entrega que contengan reactivos y patrones qumicos. Siempre se recomienda el uso de protectores oculares cuando sea posible el contacto con productos qumicos. VORISCHT Es wird dringend empfohlen, die Sicherheitsdatenbltter vor der Handhabung von Behltern, Tanks und Zufuhrsystemen, die chemische Reagenzien und Standardsubstanzen enthalten, aufmerksam durchzulesen, damit Sie sich mit den beim Umgang mit diesen Chemikalien notwendigen Vorsichtsmanahmen, Risiken und Notfallschutzmanahmen vertraut machen, Es wird empfohlen, in allen Situationen, in denen mit einem Kontakt mit Chemikalien zu rechnen ist, eine Schutzbrille zu tragen. ATTENZIONE Per familiarizzare con misure di precauzione di controllo manuale, con procedure pericolose e di emergenza, rileggere sempre le schede di sicurezza del materiale prima di maneggiare contenitori, imballaggi per spedizione che contengono reagenti chimici e standards. Si consiglia di indossare sempre gli occhiali protettivi quando possibile un contatto con agenti chimici.

Replace autoburette piston seals if: They have been in use longer than 3 months. Air leakage into the autoburette body through the seals is visible. There are signs of leakage on the back side of the autoburette module.

Important Note: Regular maintenance of the Autoburette Module is critical for correct analyzer function. Failure to perform the required regular maintenance will void the manufacturer warranty.

Use the following procedure, Figure 25, and Figure 26 to replace the autoburette seals: 1. Place the instrument in standby, then turn it off. The piston will be fully forward. 2. Break the siphon to the cleaning solution by disconnecting the reagent tubing from the reagent vent cap insert. Lift the reagent line until the bottom of the reagent line is above the liquid level in the bottle. 3. Remove the two tubes connected to the autoburette body. 4. Remove the knurled lock ring from the autoburette body; set it aside for reuse when reassembling the autoburette.
Page 68 Maintenance

6200518 Operation.fm

Section 5
5. Gently slide the autoburette body assembly off of the piston by pulling it away from the module base. Be careful not to bend the piston and lead screw assembly during removal. 6. Use a -inch open-end wrench to loosen the ceramic piston guide from the Autoburette Body assembly. Unscrew the piston guide and set it aside. Figure 25 Autoburette Module

7. Remove the two piston seals from the seal retaining ring and discard. Take care not to damage the seal retaining ring. 8. Remove the O-ring that is seated within the assembly. Do not use a sharp tool for O-ring removal, damage to the auto-burette body could result. 9. Install a new O-ring; make sure it is fully seated in the groove. 10. Install the two new piston seals onto the seal retaining ring. Place one seal flanged-side up and press the seal retaining ring onto it. Repeat for the opposite side. Ensure both rings are fully engaged and that the resulting edge is flat. 11. Drop the assembled piece into the autoburette body so it is resting on the O-ring. 12. Install the ceramic piston guide and tighten just until resistance is felt. 13. Fully insert the stainless steel alignment tool. 14. Securely hand-tighten the ceramic piece into the Autoburette Body.
Page 69 Maintenance

6200518 Operation.fm

Section 5
Note: The adequacy of sealing is not determined by the tightness of the piston guide. Overtightening of the piston guide will lead to premature failure.

15. Slide the alignment tool out of the assembled Autoburette Body.
Note: If the alignment tool does not slide out easily, the ceramic piece is too tight. Loosen the ceramic piece slightly and remove the alignment tool.

16. Lubricate the outer circumference of the ceramic piston guide with the pump lubricant (Cat. No. 62044-00). Do not allow the lubricant to touch the piston. 17. Slide the assembled unit onto the Autoburette Piston with the weep hole oriented at the bottom. 18. Reinstall the knurled lock ring and hand-tighten it to complete the installation.
Note: There should NOT be any movement of the autoburette body once the lock ring is fully in place.

19. Reinstall the two tubing connections. 20. Replumb the instrument; refer to the plumbing diagram in the parameterspecific manual supplied with your instrument. 21. Restart and prime the instrument. Figure 26 Autoburette Body Assembly

5.1.2.7 Greasing the Lead Screw in the Autoburette Module


The motor lead screw must be lubricated with the supplied grease every 30 days for proper operation. Proceed as follows: 1. Open the front cover of the instrument and watch the Autoburette body assembly. 2. When the piston is in its full forward position, use the power switch to turn the analyzer off. 3. Loosen the two phillips-head screws on the front of the module and slide the module out approximately 3 inches.
Page 70 Maintenance

6200518 Operation.fm

Section 5
4. Locate the lead screw in the recessed area at the top of the module. 5. Apply a thin bead of grease the length of the exposed screw.
Note: Do not apply grease to the portion of the screw protruding from the back of the module. Grease the part of the screw that is exposed in the recessed area on the top of the module only.

6. Slide the module back into the instrument, retighten the screws and turn the power back on.
Important Note: Failure to preform monthly maintenance on the autoburette module will void the warranty.

Figure 27 Greasing the Lead Screw

5.1.3 Unscheduled Maintenance


The following are maintenance procedures that may need to be performed more or less frequently depending on the parameter measured and the condition of sample being introduced to the instrument.

5.1.3.1

Fuse Replacement
DANGER Couper l'alimentation lectrique de l'appareil pour retirer ou installer un fusible. PELIGRO Apagar la electricidad del instrumento al quitar o instalar fusibles. GEFAHR Beim Entfernen oder Einsetzen der Sicherung muss die Stromzufuhr zum Gert unterbrochen werden. PERICOLO Staccare lo strumento dall'alimentazione a rete prima di togliere o inserire un fusibile.

DANGER Remove power from the instrument when removing or installing a fuse.

6200518 Operation.fm

Page 71 Maintenance

Section 5
The analyzer contains two ac line (5mm x 20 mm) fuses (Cat. No. 52083-00). These are located on the power supply board in the back of the instrument. They protrude through the power supply cover on the lower right side and are labeled F1 and F2 (see Figure 28). It is not necessary to remove the power supply cover to replace the fuses. Replace the fuses as follows: 1. Disconnect all ac power to the analyzer. 2. Turn the fuse carrier turn counterclockwise to remove it from the fuse holder. 3. Replace the fuse (T1.6 A. 250 V, Cat. No. 52083-00) inside the carrier. 4. Turn the fuse carrier turn clockwise to reinstall it in the fuse holder. 5. Reconnect ac power to the analyzer.
DANGER For continued protection against fire hazard, replace the fuse only with a fuse of the same type and rating. DANGER Pour assurer la protection contre les risques d'incendies, remplacez les fusibles uniquement par des fusibles du mme type et pour la mme intensit. PELIGRO Para una continua proteccin contra incendios, reemplace los fusibles nicamente por los del tipo y capacidad recomendados. GEFAHR Zur Wahrung des kontinuierlichen Brandschutzes drfen die Sicherungen nur mit Sicherungen des gleichen Typs und mit gleichen Stromkennwerten verwendet werden. PERICOLO Per evitare ogni pericolo di incendio, sostituire ogni fusibile solamente con uno dello stesso tipo ed amperaggio.

Page 72 Maintenance

6200518 Operation.fm

Section 5
Figure 28 APA 6000 Analyzer Electrical Terminals and Fuse Location

6200518 Operation.fm

Page 73 Maintenance

Section 5
5.1.3.2 Cleaning the Mixer Module Chamber
It may become necessary to clean the mixer module chamber and the surface where the stir bar rests. This is usually indicated when the module chamber becomes discolored by biogrowth, staining, or sediment. 1. For safety reasons, run a CLEAN cycle before disassembling the mixer module. 2. Put the instrument in standby. 3. Notice the mixer housing is split. Use a Phillips screwdriver to loosed the two captive screws that secure the module to the instrument panel. 4. Hold the part of the housing next to the block while pulling the top of the housing out from the block. The part of the housing that holds the stir bar should stay attached to the instrument. Let the top part of the housing hang from the attached tubing. 5. Use a cotton swab soaked with a 10% household bleach solution to clean the mixing chamber and the module housing top. Rinse well with deionized water. See Figure 29. Figure 29 Mixer Module Parts

Page 74 Maintenance

6200518 Operation.fm

Section 5
6. If necessary, use a cotton swab soaked with a 10% household bleach solution to clean the surface the stir bar rests on and the stir bar. Rinse with deionized water. 7. Reassemble the housing, making sure the o-ring and stir bar are in place. See Figure 29. 8. Tighten the Phillips screws that hold the top of the housing to the block. 9. Restart the instrument.

5.2 Troubleshooting
5.2.1 General Troubleshooting Procedure
This section provides general procedures to solve problems that affect the analyzers performance. The first sign of a problem may be an alarm or warning, but erroneous or erratic readings may not always cause an alarm or warning. Being familiar with the steps in this section will help to quickly solve operational problems.

5.2.1.1 Check Alarm Log


Occasionally check the bottom line of the AquaTrend for N or 1 through 8. These are indicators for network or sensor alarms/warnings. If any alarm or warnings occur, this will help narrow the search to a specific module or problem. See section 5.3 on page 77 for specific alarm and warning messages and corrective actions. See section 4.2.1 on page 37 to review the alarm log.

5.2.1.2 Check Solutions


Verify Reagent, Standard, and Cleaning Solution bottles are not empty and are connected to the correct tube. Ensure the reagent line in the guide tube goes to the bottom of the bottle. Make sure standards are the correct value. Check tubing at the rotary valve and make sure the tubing is filled with fluid, not air. If ports on the rotary valve have been removed recently, refer to the plumbing diagram and ensure the ports were reconnected properly. If reagent or standard is not delivered to the valve, check for kinks in the tubing. Replace any tubing that is kinked or crushed. If reagents or standards are degrading, check the seal on the bottle caps and the tubing connections. Replace if necessary.

6200518 Operation.fm

Page 75 Troubleshooting

Section 5
5.2.2 Examine the Rotary Valve
Watch the rotary valve movement during a cycle. If fluid is moving from the common port to more than one outlet port, cross-leakage in the rotary valve is occurring. Inspect the rotary valve per section 5.1.2.5 on page 65. It is likely that stator will need to be replaced. Verify fluid is aspirated in when the autoburette is aspirating. If air is pulled into the tubing connected to the center port, and there is fluid at the selected port, the selected port may be clogged. Inspect and clean the rotary valve per section 5.1.2.5 on page 65. If reagent or standard is not delivered, disassemble and inspect the valve rotor and stator for blockage or buildup. Also inspect the reagent and standard port tubing.

5.2.3 Examine the Autoburette


If air is moving through the piston seals during operation, check the vortex filter. If the filter is plugged, clean it as described in section 5.1.2.2 on page 60. If the filter is not plugged, replace the seals per section 5.1.2.6 on page 68. If there is fluid leakage around the autoburette body, replace the seals.

5.2.4 Examine the Tubing and Fittings


Check the tubing for problems which may cause reduced or blocked flow. Reduced or blocked flow may also be caused by fittings that are overtightened. If fluid in the holding coil is not observed, the holding coil may be blocked. Replace the block containing the coil

Important Note: Do not shorten the tubing lengths for the holding coil.

Examine the fittings for leakage and tighten if necessary. The fittings should be firmly finger tight. If tubing was recently replaced, check the tubing lengths against the table presented in the parameter-specific manual.

5.2.5 Back Pressure Gauge


The APA 6000 is equipped with a back pressure gauge positioned on the right-front of the analyzer, attached to the drain block. The gauge monitors the pressure applied on the detection line of the analyzer. This line is pressurized via a small spring-loaded device located on the back of the drain block. The factory setting of the device is 20 psi. The detector and valve are temperature controlled at an elevated temperature. This may cause entrained air to leave the solution. The reason for pressurizing this line is to minimize out-gassing as the solution passes through the detector and reaction coil. The gauge can serve as a good indication of the proper operation of the analyzer. The back pressure will reach a maximum value (20 psi) as the

Page 76 Troubleshooting

6200518 Operation.fm

Section 5
autoburette dispenses fluid through the detector line. Following dispensation, the pressure may drop 34 psi and then should remain approximately constant. The gauge can indicate possible problems in the following ways: If the back pressure never reaches 20 psi and fluctuates as the autoburette is dispensing fluid, this may indicate leaking seals in the autoburette. If the back pressure reaches 20 psi as the autoburette dispenses, but then drops below 5 psi after the burette has stopped, this may indicate a leak in the valve which may also indicate the beginning of wear that can result in leakage between valve ports and poor performance. Finally, if the gauge reads above 25 psi, it is likely that the back pressure device or detector line has buildup and needs to be replaced.

5.3 Specific Alarm/Warning Messages


There are two types of alarms, system alarms and sample setpoint alarms. System alarms cause the analyzer to go into standby mode and require operator intervention to correct the problem. Sample setpoint alarms are caused by sample concentrations exceeding the user-selected setpoint values. Warnings notify the operator of potential problems which might degrade the performance of the analyzer. When an alarm or warning occurs, it is placed in the alarm log, which stores up to 14 alarm/warning messages. The alarm log can be reviewed from the AquaTrend (see section 4.2.1 on page 37). The following tables describe the possible set point alarms (Table 11), system alarms (Table 12) and system warnings (Table 13), along with causes, analyzer actions, and corrective actions. Use these tables to return the analyzer to proper operation. Table 11 APA 6000 Set Point Alarms
Alarm/Warning Cause Sample is below user setting for concentration setpoints. Analyzer Action Enable assigned relay when alarm condition occurs. Disable assigned relay when condition clears. Alarm is registered in ALARM LOG. Enable assigned relay when alarm condition occurs. Disable assigned relay when condition clears. Alarm is registered in ALARM LOG. Enable assigned relay when alarm condition occurs. Disable assigned relay when condition clears. Alarm is registered in ALARM LOG. Corrective Action

LOW ALARM value units

None.

HIGH ALARM value units

Sample exceeds user setting for concentration setpoints.

None.

RATE ALARM value units

Sample rate change exceeds user setting for concentration setpoints.

None.

6200518 Operation.fm

Page 77 Specific Alarm/Warning Messages

Section 5
Table 12 APA 6000 System Alarms
Alarm/Warning A/D FAILURE Cause Detector module main analog to digital converter has failed. RAM on APA main circuit board has malfunctioned. Chemistry dipswitch is set to an illegal location. Valve module is not communicating with controller. Burette module is not communicating with controller. Mixer module is not communicating with controller. Detector module is not communicating with controller. Internal communication problem. Valve was unable to find its home signal. Valves home signal is always on. Valve was unable to move to the desired port. Analyzer Action Analyzer shuts down. Corrective Action Restart the analyzer. Replace detector module if problem persists. Restart the analyzer. Replace main board if problem persists. Verify dipswitch setting is correct. If the problem persists, replace the main board. Check to ensure all cable connections are secure. Replace module if problem persists. Replace the burette module. Replace the mixer module. Replace the detector module. Replace the main circuit board. Replace module if problem persists. Replace the valve module. Replace module if problem persists. Check for clogged lines or incorrect plumbing in the analyzer. Rebuild the pump. If problem persists, replace autoburette. Check for clogged lines or incorrect plumbing in the analyzer. Rebuild the pump. If problem persists, replace autoburette. Check for clogged lines or incorrect plumbing in the analyzer. Rebuild the pump. If problem persists, replace autoburette. Check for clogged lines or incorrect plumbing in the analyzer. Rebuild the pump. If problem persists, replace autoburette. Service the Sample Conditioning Unit or restore the sample flow.

RAM TEST FAILED

Analyzer shuts down. An alarm is logged and the analyzer shuts down. An alarm is logged and the analyzer shuts down. An alarm is logged and the analyzer shuts down. An alarm is logged and the analyzer shuts down. An alarm is logged and the analyzer shuts down. An alarm is logged and the analyzer shuts down. An alarm is logged and the analyzer shuts down. An alarm is logged and the analyzer shuts down. An alarm is logged and the analyzer shuts down.

INVALID DIPSWITCH

VALVE COMM FAIL

BURETTE1 COMM FAIL MIXER COMM FAIL DETECTOR COMM FAIL INTERNAL COMM FAIL VALVE NO HOME VLV HOME SIG ON VLV MOVE ERR

BURETTE1 HOME ERR

The Home location for Burette 1 has drifted outside of its limit from the last Home check.

An alarm is logged and the analyzer shuts down.

BURETTE1 HOME DRIFT

The cumulative offset drift from Home has exceeded the acceptable limits.

An alarm is logged and the analyzer shuts down.

BURETTE1 NO HOME

Burette1 was unable to ever find its home location.

An alarm is logged and the analyzer shuts down.

BURETTE1 HOME ON

The Home signal for Burette1 is always on.

An alarm is logged and the analyzer shuts down.

SAMPLEx OUT

Sample inlet x has no sample fluid.

An alarm is logged and the analyzer shuts down.

Page 78 Specific Alarm/Warning Messages

6200518 Operation.fm

Section 5
Table 13 System Warnings
Alarm/Warning POWER FAILURE Cause Power was turned off or lost. Analyzer Action A warning message is logged. Analyzer continues to operate. A warning message is logged. Analyzer continues to operate using the average of the cal readings. A warning message is logged. Analyzer continues to operate using the previous calibration. Corrective Action None.

CAL STDx REPEATABILITY

Cal Standard xs readings are varying.

Typically a mechanical failure. Follow trouble shooting guidelines for modules. Typically a mechanical failure, also may be out of reagents and/or standards.

STDx FAIL ACC/REP

Standard xs readings are outside acceptable calibration limits and are not repeatable.

STDx LOW or REAGENTx LOW or CLEANER LOW

Standard, reagent or cleaning solution level is less than 5%.

Check fluid level in each bottle. If fluid level is almost empty, replace the bottle and reset the Analyzer continues to operate fluid level. If the fluid level is not near empty, reset the fluid level to reflect the correct level. A warning message is logged. Analyzer continues to operate using the previous calibration. A warning message is logged. Analyzer continues to operate. Analyzer continues to operate. Analyzer continues to operate. Reagent or standard problem. Replace reagents and/or standards.

STDx FAIL ACCURACY

Standard xs readings are outside acceptable calibration limits.

DETECTOR LIGHT LEAKAGE LED OUTPUT LOW LED OUTPUT HIGH

Light is leaking into the colorimeter. Colorimeter LED light output is too low. Colorimeter LED light output is too high.

Replace detector module. Replace detector module if problem persists. Replace detector module if problem persists.

6200518 Operation.fm

Page 79 Specific Alarm/Warning Messages

Visit http: //www.hach.com

Appendix A

Network Connections
The connections described below use the following parts: Shielded, two-conductor cable (Cat. No. 52157-00) Cable Termination Kit (Cat. No. 52156-00) - includes: Strain Relief (2) (Cat. No. 52025-00) Sealing Washer (2) (Cat. No. 10338-14) Locknut (2) (Cat. No. 10596-12) Insulated Quick Connects (6) (Cat. No. 52143-00) Heat Shrink Tubing (1 foot) (Cat. No. 40825-00)

Network Connections in the APA 6000 Analyzer

The Cable Termination Kit is supplied with the SIO, SOM, or PS1201 Modules. If connecting the APA 6000 to another APA 6000, order this kit and cable along with the additional APA 6000. 1. Make sure there is no power supplied to the APA 6000. 2. Open the APA 6000 door and remove the screws on the hinged panel holding the modules to access the electrical connections at the back of the instrument. Refer to Figure 30. 3. Remove the two lower wing nuts on the lower back panel, then remove the lower terminal strip cover. Refer to Figure 31. Figure 30 Electrical Connections

6200518 Appendix A.fm

Page 81 Network Connections

Appendix A
Figure 31 Removing the Customer Access Cover

4. Remove the wire access plug below the network terminal strip where the network cable will be routed into the APA 6000. Refer to Figure 32. 5. Take one end of the network cable and strip 15 cm (6 in.) of outer insulation from the cable. 6. Put the strain relief and sealing washer (supplied with the cable) over the cable and slide the strain relief and sealing washer 7 inches down from the end of the cable. Tighten the strain relief nut until the cable does not slip through the strain relief.
Note: Connect shield at one end of cable only. If connecting to another analyzer, then connect shield in only one of the analyzers. Insulate the non-connected end with tape or heat-shrink tubing.

7. Remove inch of insulation from the white and green wires. Either insulate the shield wire with appropriate length of heat shrink tubing or remove inch of insulation from the shield wire. Pull the cable through the APA 6000 wire access opening until the strain relief/sealing washer is pulled flush with the opening. From inside the analyzer, screw the locknut onto the strain relief and tighten.

Page 82 Network Connections

6200518 Appendix A.fm

Appendix A
Figure 32 Access to the Network Terminal Strip

Back View of Instrument

Air Purge

Power Connections Alarm Connections Drain

Network Connections Recorder Connections

6200518 Appendix A.fm

Page 83 Network Connections

Appendix A
8. Connect the white wire to the APA 6000 position A on the Network terminal block. Connect the green wire to position B on the Network terminal block. If connecting the shield wire, connect the ground wire of the cable to the terminal stud near the network terminal block. 9. If the AquaTrend Interface in the analyzer will be the master AquaTrend Interface or if the analyzer is not connected to a network, connect network terminator by placing the jumper over pins 2 and 3 on the J11 connector (see Figure 33). If this is not the master AquaTrend Interface, place the jumper over pin 2. 10. Replace the lower terminal strip cover over the electrical connections and install the two wing nuts to secure the cover. 11. Close the hinged panel holding the modules and reinsert and tighten screws holding the module door shut. Figure 33 Network Terminator

J11 Connector 3 1

Jumper

J11 Disabled 3 1 1

J11 Enabled 3

Page 84 Network Connections

6200518 Appendix A.fm

Appendix B

Printer and Computer Formats

APA 6000 Printer and Computer Formats


Printer Format Since the SIO can receive data from up to 8 analyzers, the SIO channel is shown in the first three characters (i.e., [1] specifies information is from the analyzer connected to SIO channel 1).
[1]01/01/97 00:23: 3.234 S1P1 [1]01/01/97 00:23: 0.527 S1P2 [1]01/01/97 11:28: 3.977 S1P1 *W [1]01/01/97 11:28: SMP1 OUT ALARM *A [1]01/01/97 11:28: 0.611 S1P2 *A * W [1]01/01/97 11:29: End SMP1 OUT ALARM *A

First occurrence of any alarm or warning will be sent out. The time/date of the alarm or warning is printed followed by either ALRM or WARN to specify it is an alarm or warning. The specific alarm or warning is then printed on the next line. When the alarm or warning condition clears it will be sent preceded by END: to indicate the condition is cleared. Some alarms and warnings provide information only and do not have an end (i.e. a fail to calibrate message is printed out only when the condition occurs). If a alarm or warning condition is active, then a *A for alarm or a *W for warning is appended to the measurement readings. Computer Format When the SIO is setup to transmit to a computer, the format is similar to the printer format described above. In addition, the year is included in the date and the alarms and warnings are printed out on the same line as the time and date. When using PC format, the time/date stamp can be disabled and the readings can be queried instead of being sent out in real-time. If the SIO is in Queried mode, the readings are sent out only when the analyzer receives RMR command, alarms and warnings are sent only when receive RAR command (refer to Table 14 for more on RMR and RAR commands). Remote RS232 Commands Table 14 lists the commands that can be sent to the APA 6000 Analyzer for remote analyzer operation.The analyzer response to remote commands will be OK if changing instrument settings. Commands which query settings or values will be responded to with the current settings or readings. If the analyzer does not recognize the command then a BAD CMD message is returned. If the analyzer recognizes the command but the settings are invalid then BAD SETTINGS is returned. Messages to the analyzer must be terminated with a carriage return character to signify end of command. Linefeed characters and spaces are ignored. All incoming characters are converted to uppercase before evaluated. When sending commands through the SIO, the channel number must precede the command. For example, if the analyzer is connected to SIO channel 1 and you wish to restart the analyzer, 1SRT should be sent. The analyzer will reply with OK and the SIO inserts the channel assigned for the analyzer (i.e. [1]OK will be the response to the computer).

6200518 Appendix B.fm

Page 85 Printer and Computer Formats

Appendix B
Table 14 APA 6000 Remote RS232 Commands
Description of Remote Command Restart system Put instrument in standby Perform a calibration Perform a clean cycle Prime entire system Prime reagents Prime standards Command SRT HLT CAL CLN PMA PMR PMS Description of Command/Settings Restarts the system. Puts the analyzer in standby. Performs a calibration at end of the current cycle. Performs a clean cycle at end of the current cycle. Primes all the reagent and sample lines now. Primes reagents lines at end of current cycle. Primes standard lines at end of current cycle. Sending just first three characters CTM will result in current auto-cal. settings to be returned. Add the equal sign and one of the characters listed in the Command column to the left to set new auto-calibration interval. Calibrate every 12 hours. x = 0 to 23 indicates time of day to calibrate (using a 24-hour clock). If x=22, the analyzer calibrates at 10:00 p.m. and a.m. Calibrate daily. x = 0 to 23 indicates time of day to calibrate (using a 24Hhour clock). Auto-Calibrate Weekly. x = 6 to 12 which signifies the day of week to calibrate where Monday=6 and Sunday=12. y = 0 to 23 to indicate time of day to calibrate (using 24-hour clock). Sending just first three characters recalls the calibration values from the last calibration. Adding =A will recall all calibrations stored (up to 7). Resets calibration curve to factory default values. Sending just first three characters results in current settings being returned. Adding =x will result in following settings where x =1=6 hours, x=2=12 hours and x=3=24 hours. Sending just first three characters results in current settings being returned. Adding =x will result in following settings: x =1=6 hours, x=2 =12 hours and x=3=24 hours. x=4 to 10 signifies clean weekly where the day of the week to clean is 4 for Monday to 10 for Sunday. Returns SEQUENCE=12...1 which represents the order in which the samples are measured. Can have up to 8 values in sequence. The numbers after the equal sign represent the order in which sample measurements are performed. Nitrate V1.0 C1.0, Copper V1.0C1.0, Ammonium V1.0 C1.0, Amm and Nitr V1.0 C1.0, Hardness V1.0 C1.0, Alkalinity V1.0 C1.0 Output up to last 10 alarms/warning messages since last RAR command. Only enabled if SIOx setup in Query mode. Output last measurement reading for all cycles (grab and normal measurement readings & calibrations). Only enabled if SIOx setup in Query mode. Resets all standards, reagents, and cleaning solutions to 100% Send just first three characters if requesting level The third character is the number of the reagent. Add = followed by reagent level in percent if setting to a new level. Allowed values are between 0 and 100%.

CTM

Set/query Auto-Calibration setting

CTM=4,x

CTM=5,x

CTM=x,y

Recall calibration Reset to default calibration Get/Set auto-refresh settings

RCL=A RDC ELI=x

Get/Set auto-clean settings

CLI=x

SEQ Get/Set sample sequence SEQ=12. Get instrument ID Output latest alarms/warnings RID RAR

Output latest readings Reset all fluid levels to 100%

RMR RFL

Query/Set reagent levels

RG1=xxx

Page 86 Printer and Computer Formats

6200518 Appendix B.fm

Appendix B
Table 14 APA 6000 Remote RS232 Commands (continued)
Description of Remote Command Command Description of Command/Settings Send just first three characters if requesting level The third character is the number of the standard. Add = followed by standard level in percent if setting to a new level. Allowed values are between 0 and 100%. Send just first three characters if requesting level Add = followed by fluid level in percent if setting to new level. Allowed values are between 0 and 100%. The percent sign is optional. Sends out all Instrument settings. Send first three characters to get instrument name. Default name is #1 APA 6000. Add equal sign plus text to set new name. Maximum length of name is 12 characters. Note: all letters will be converted to upper case. Set x = F or H or Z for output recorder full-scale, half-scale or zero respectively. Send just first three characters to stop test. Set x = F or H or Z for output recorder full-scale, half-scale or zero respectively. Send just first three characters to stop test. Set x = 0 or 1 for relay off or on. Send just first three characters to stop test. Set x = 0 or 1 for relay off or on. Send just first three characters to stop test. Get version numbers of instrument and its modules. For Hardness V1.0 C1.0, EXT NEURON: EXT0101, INT NEURON: INT0101, (and following if used:), VALVE MODULE: VLV0103 DETECT MODULE: DET0102, BURETTE1 MODULE: PMP0102, MIXER MODULE: PMP0102 Transmits all messages in current alarm log. Clears alarm logs.

Query/Set standard levels

SD1=xxx

Query/Set pump wash fluid level Request all instrument settings and information

WSH=xxx

ALL

Set/Query instrument name

RIN=xxx

Test internal recorder channel A Test internal recorder channel B. Test internal relay A. Test internal relay B.

RCA=x RCB=x RLA=x RLB=x

Get software version numbers.

RVN

Query alarm/warning logs Clear alarm/warning logs

TAL WAL

6200518 Appendix B.fm

Page 87 Printer and Computer Formats

Visit http: //www.hach.com

Appendix C

Second Sample Options


The APA 6000 Analyzer is capable of analyzing two sample streams. If a second sample stream is added to the analyzer, the sample must be filtered to 0.5 m.

Sample Sequencing
Hach recommends using the Sample Sequencing Kit (Cat. No. 62009-00) to add a second sample stream to the analyzer in clean water applications. The Sample Sequencing Kit contains a vortex filter sample conditioning block, tubing, fittings, spare filters, and instructions. For wastewater applications Hach recommends using the APA 6000 Micro filter Kit (Cat. No. 51339-00 115 V or 51339-01 230 V).

Quality Control
A second sample stream may be used to verify the operation of the APA 6000 by reading a known sample at regular intervals. Hach recommends using the Quality Control Kit (Cat. No. 62032-00).This kit includes a carboy capable of holding a one month supply of a quality control standard, plus the tubing and fittings necessary to plumb the carboy to the analyzer.

Grab Sample
To occasionally check a second sample or verify operation by reading a known sample, use the Grab Sample Kit (Cat. No. 62037-00) which is supplied with the instrument.

6200518 Appendix C.fm

Page 89 Second Sample Options

Appendix C
Setting Up the Sample Sequence
For analyzers with two sample streams, the measurement sequence option allows the user to customize the sequence of sample stream measurements. The analyzer will measure the sample streams based on the step order the operator selects. The sample streams can be in any order; the analyzer will sequence up to eight sample cycles. For example: 1, 2, or 1, 2, 1, 1, 1, 2.
Note: The sequence will terminate at the first unused step. Subsequent steps will be cleared.

1. Add the new sample stream to the analyzer:


Key Press 1. MENU SENSOR 2. ENTER SELECT SENSOR Select sensor by name 3. ENTER SENSOR DIAGNOSTICS 4. ENTER SAMPLE STREAMS 5. ENTER TWO 6. ENTER Menu Selection

2. Assign a measurement to channel for the new sample stream:


Key Press 1. MENU AQUATREND MENU 2. ENTER DISPLAY SETUP 3. ENTER MEASUREMENT TO CHANNEL Select ATTACH 4. ENTER SELECT SENSOR Select sensor by name 5. ENTER SELECT MEASUREMENT / CHANNEL 6. ENTER Select the channel to which the measurement is to be displayed 7. ENTER A pop-up window will appear Select ATTACH 8. A pop-up window will confirm the attachment Menu Selection

Page 90 Second Sample Options

6200518 Appendix C.fm

Appendix C
3. Use the sample sequencing function to control the frequency of the readings taken from the new sample stream.
Key Press 1. 2. MENU SENSOR ENTER SELECT SENSOR Select sensor by name 3. 4. 5. ENTER MEASUREMENT OPTIONS ENTER SAMPLE SEQUENCE ENTER STEP X Select the first unused step. 6. ENTER SAMPLE NUMBER Assign either sample stream 1 or 2 to the step 7. ENTER Menu Selection

For each STEP X (18) in Step 4, choose one of the measurement streams to sequence. The instrument will analyze in this sequence down to the first step left unused and will then repeat the sequence. For example, if you want the analyzer to alternate readings between the sample and the quality control standard, you would only program 1, 2. If you want the analyzer to read the quality control standard every fourth measurement, program 1, 1, 1, 2.

6200518 Appendix C.fm

Page 91 Second Sample Options

Visit http: //www.hach.com

General Information

At Hach Company, customer service is an important part of every product we make. With that in mind, we have compiled the following information for your convenience.

6200518 General Information Stopper.fm

Page 93 General Information

Replacement Parts and Accessories


See the parameter-specific manual for information on analyzer configurations, reagents, standards, and other specific parts.

Optional Equipment and Supplies


Description Unit Cat. No

AquaTrend Instrument................................................................................................................each.......... 51200-00 AquaTrend Instrument with SOM ...............................................................................................each.......... 51350-00 Cable Termination Kit .................................................................................................................each.......... 52156-00 Digital Display Module (DDM) ....................................................................................................each.......... 52400-00 Power Cord, UL/CSA, 115 V, w/ NEMA-style strain relief ..........................................................each.......... 46306-00 Power Cord, VDE, 230 V, w/ Continental-style plug and NEMA-style strain relief .....................each.......... 46308-00 PS2401 Power Supply, 115 V ....................................................................................................each.......... 54352-00 PS2401 Power Supply, 230 V ....................................................................................................each.......... 54353-00 Serial Input/Output Module (SIO)...............................................................................................each.......... 52074-00 Signal Output Module (SOM) .....................................................................................................each.......... 51250-00

Sample Conditioning
APA 6000 Micro Filter System, 115 V ........................................................................................each.......... 51339-00 APA 6000 Micro Filter System, 230 V ........................................................................................each.......... 51339-01 Sample Conditioning Block, -in fittings....................................................................................each.......... 62008-00 Sample Sequencing Kit..............................................................................................................each.......... 62009-00 Y-strainer Kit...............................................................................................................................each.......... 46618-00

Replacement Parts
AquaTrend Instrument Manual ...................................................................................................each.......... 51350-18 Binder, D-ring, for AquaTrend manual........................................................................................each.......... 52001-77 Autoburette Module ....................................................................................................................each.......... 57150-00 Back Pressure Regulator Assembly...........................................................................................each.......... 55226-00 Bulkhead Union, 0.25 OD tubing to tubing (for air purge bulkhead) ..........................................each.......... 51104-00 Cable, backplane .......................................................................................................................each.......... 51161-00 Cable, AquaTrend ......................................................................................................................each.......... 51164-00 Circuit Board, AquaTrend ...........................................................................................................each.......... 51092-00 Faceplate Assembly, APA 6000 (plastic bezel only, does not include circuit board) ..................each.......... 51022-00 Filter, Ceramic, for Filter assembly.............................................................................................each.......... 62011-00 Fitting, Ferrule, 0.062 OD...........................................................................................................each.......... 51130-00 Fitting, Ferrule, with ring, 0.062 OD ...........................................................................................each.......... 51132-00 Fitting, Union, Tefzel, -28 ........................................................................................................each.......... 51359-00 Flangeless Hex Nut Fitting, 0.062 OD........................................................................................each.......... 51129-00 Flangeless Nut Fitting, 0.062 OD ...............................................................................................each.......... 51131-00 Fuse, 1.6 A, 250 V, 5 mm x 20 mm, UL, CSA, VDE approved...................................................each.......... 52083-00 Grommet, manifold, APA 6000 (for tubing).................................................................................each.......... 51036-00 Plug, air purge, for -in. tubing..................................................................................................each.......... 51140-00 Plug, manifold, APA 6000...........................................................................................................each.......... 51224-00 Plug, flat bottom, Tefzal, -28....................................................................................................each.......... 51293-00 Power Switch, 4A, 250 VAC .......................................................................................................each.......... 51155-00 Reagent Bottle Vent Cap Assembly ...........................................................................................each.......... 51215-00 Reagent Housing Assembly.......................................................................................................each.......... 51048-00 Restraining Cable, steel, w/eyelets (for module panel) ..............................................................each.......... 51417-00 Rotary Valve, 10-port, with actuator...........................................................................................each.......... 51150-00 Rotary Valve, head only .............................................................................................................each.......... 51507-00 Seal, oil tight, 0.88-in. diameter .................................................................................................each.......... 42210-00 Stator, for 10-port valve..............................................................................................................each.......... 51347-00 Stir Bar, micro, 3 mm x 10 mm...................................................................................................each.......... 20549-59 Teflon Tubing, 0.170 ID x 0.250 OD ...................................................................................... per foot.......... 45462-00

Page 94 Replacement Parts and Accessories

6200518 Replacement Parts.fm

Replacement Parts and Accessories


Replacement Parts, continued
Description unit Cat. No.

Teflon Thread Tape, 0.25-in wide ...............................................................................................1 roll.......... 70608-24 Tubing, Tefzel, 0.030 ID x 0.062 OD, 2000 psi ...................................................................... per foot.......... 51178-00 Valve Body Replacement, 10-port valve ....................................................................................each.......... 51507-00 Valve Module Assembly, APA 6000............................................................................................each.......... 62050-00 Valve Plug, 10-32 .......................................................................................................................each.......... 51393-00

Kits
Installation Kit............................................................................................................................................... 51040-00 Filter, 0.5 micron ..................................................................................................................each.......... 62011-00 Grab Sample Kit ..................................................................................................................each.......... 62037-00 Poly Tubing, -in. OD......................................................................................................20 feet........... 47438-00 Tool Kit ......................................................................................................................................................... 51291-00 Alignment tool...................................................................................................................1 each.......... 57161-00 Allen Wrench, 7/64 in., long arm (1) ..................................................................................1 each.......... 51345-00 Fitting, Hex Nut, 0.062-in.OD tubing (4) ...........................................................................4 each.......... 51129-00 Fitting, Ferrule, 0.062 in. OD tubing (4) ............................................................................4 each.......... 51130-00 Fitting, Flangeless Nut, 0.062 in. OD tubing (10) ...........................................................10 each.......... 51131-00 Fitting, Ferrule w/ Ring, 0.062 in. OD tubing ((10) ..........................................................10 each.......... 51132-00 Nut Extender Tool, Flangeless (1) ....................................................................................1 each.......... 51174-00 Plug Fitting, Hex Nut Valve, 10-32 (3)...............................................................................3 each.......... 51393-00 Plug, Flat-bottom Tefzel, -28 (4) ....................................................................................4 each.......... 51293-00 Tefzel tubing, 0.030 ID x 0.062 in. OD, 2000 PSI ............................................................15 feet.......... 51178-00 Tubing cutter, for 1/16-in. Tubing (1) ..................................................................................1 each.......... 51356-00 Wrench, Valco Valve, -inch (1).......................................................................................1 each.......... 51357-00 Maintenance Kit ........................................................................................................................................... 62028-00 Auto burette piston seals .................................................................................................8 each.......... 57163-00 Grease, Lead Screw ......................................................................................................... 1 tube.......... 62044-00 O-ring, fluorocarbon..........................................................................................................4 each.......... 62014-00 Valve Rotor .............................................................................................................................2 each.......... 51348-00

6200518 Replacement Parts.fm

Page 95 Replacement Parts and Accessories

How to Order
U.S.A. Customers
By Telephone: 6:30 a.m. to 5:00 p.m. MST Monday through Friday (800) 227-HACH (800-227-4224) By Fax: (970) 669-2932 By Mail: Hach Company P.O. Box 389 Loveland, Colorado 80539-0389 U.S.A.

Ordering information by e-mail: orders@hach.com

Information Required
Hach account number (if available) Your name and phone number Purchase order number Brief description or model number billing address Shipping address Catalog number Quantity

International Customers
Hach maintains a worldwide network of dealers and distributors. To locate the representative nearest you, send e-mail to intl@hach.com or contact: Hach Company World Headquarters; Loveland, Colorado, U.S.A. Telephone: (970) 669-3050; Fax: (970) 669-2932

Technical and Customer Service (U.S.A. only)


Hach Technical and Customer Service Department personnel are eager to answer questions about our products and their use. Specialists in analytical methods, they are happy to put their talents to work for you.

Call 1-800-227-4224 or e-mail techhelp@hach.com

Page 96 How to Order

6200518 General Information.fm

Repair Service
Authorization must be obtained from Hach Company before sending any items for repair. Please contact the Hach Service Center serving your location. In the United States: Hach Company Ames Service 100 Dayton Avenue Ames, Iowa 50010 (800) 227-4224 (U.S.A. only) FAX: (515) 232-3835

In Canada: Hach Sales & Service Canada Ltd. 1313 Border Street, Unit 34 Winnipeg, Manitoba R3H 0X4 (800) 665-7635 (Canada only) Telephone: (204) 632-5598 FAX: (204) 694-5134 E-mail: canada@hach.com In Latin America, the Caribbean, the Far East, the Indian Subcontinent, Africa, Europe, or the Middle East: Hach Company World Headquarters, P.O. Box 389 Loveland, Colorado, 80539-0389 U.S.A. Telephone: (970) 669-3050 FAX: (970) 669-2932 E-mail: intl@hach.com

6200518 General Information.fm

Page 97 Repair Service

Limited Warranty
Hach Company warrants this product against any defects that are due to faulty material or workmanship for a period of one year from the date of purchase. In the event that a defect is discovered during the warranty period, Hach Company agrees that, at its option, it will repair or replace the defective product or refund the purchase price of the defective product less any charges for shipping and handling. This warranty does not apply to: (i) damage caused by Act of God; (ii) damage caused by misuse, neglect, accident or improper application or installation; (iii) damage caused by any repair or attempted repair not authorized by Hach Company; (iv) any product not used in accordance with the instructions furnished by Hach Company. Any product repaired or replaced under this warranty will itself be warranted only for the remainder of the warranty period of the original product being repair or replaced. This warranty does not apply to short-life or consumable components of a product, such as, without limitation, light bulbs and chemical reagents.

How To Obtain Service


Because Hach Company manufactures and warrants a variety of products, you must receive a Return Authorization and details as to how warranty service may be obtained from Hach Company by calling the appropriate number for your product, see Repair Service on page 97.

Modifications and Limitations


This warranty contains the sole express warranty made by Hach Company in connection with its products. All implied warranties, including without limitation, the warranties of merchantability and fitness for a particular purpose, are expressly disclaimed. Some states do not allow the disclaimer of implied warranties and if this is true in your state the above limitation may not apply to you. This warranty gives you specific rights, and you may also have other rights that vary from state to state. This warranty constitutes the final, complete and exclusive statement of warranty terms and no person is authorized to make any other warranties or representations on behalf of Hach Company.

Limitation of Remedies
The remedies of repair, replacement or refund of purchase price as stated above are the exclusive remedies for the breach of this warranty. In no event shall Hach Company be liable for any incidental or consequential damages of any kind for breach of warranty, negligence, on the basis of strict Liability or under any other legal theory.

Geographical Limitation
This warranty is extended only to products purchased, delivered and used in the 50 United States.

Page 98 Limited Warranty

6200518 General Information.fm

Certification
Hach Company certifies this instrument was tested thoroughly, inspected and found to meet its published specifications when it was shipped from the factory. The APA 6000 Analyzer has been tested and is certified as indicated to the following instrumentation standards:

Product Safety
UL 3101-1 (ETL Listing # H0492805390) CSA C22.2 No. 1010.1 (ETLc Certification # H0492805390) Certified by Hach to EN 61010-1 (IEC1010-1) per 73/23/EEC, supporting test records by Intertek Testing Services.

Immunity
EN 61326 (EMC Requirements for Electrical Equipment for Measurement, Control and Laboratory Use) per 89/336/EEC EMC: Supporting test records by Hach Company, certified compliance by Hach Company. Standards include: IEC 1000-4-2:1995 (EN 61000-4-2:1995) Electro-Static Discharge Immunity
IEC 1000-4-3:1995 (EN 61000-4-3:1996) Radiated RF Electro-Magnetic Field Immunity IEC 1000-4-4:1995 (EN 61000-4-4:1995) Electrical Fast Transients/Burst IEC 1000-4-5:1995 (EN 61000-4-5:1995) Surge IEC 1000-4-6:1996 (EN 61000-4-6:1996) Conducted Disturbances Induced by RF Fields IEC 1000-4-11:1994 (EN 61000-4-11:1994) Voltage Dip/Short Interruptions

Additional immunity Standard/s include: ENV 50204:1996 Radiated Electro-Magnetic Field from Digital Telephones

Emissions
Per 89/336/EEC EMC: EN 61326:1998 (Electrical Equipment for measurement, control and laboratory use-EMC requirements) Class "A" emission limits. Supporting test records by Criterion Technology O.A.T.S. (NVLAP #0369), certified compliance by Hach Company. Standards include: EN 61000-3-2 Harmonic Disturbances Caused by Electrical Equipment
EN 61000-3-3 Voltage Fluctuation (Flicker) Disturbances Caused by Electrical Equipment

Canadian Interference-causing Equipment Regulation, IECS-003, Class A: Supporting test records by Criterion Technology O.A.T.S. (NVLAP #0369), certified compliance by Hach Company.
This Class A digital apparatus meets all requirements of the Canadian Interference- Causing Equipment Regulations. Cet appareil numrique de la classe A respecte toutes les exigences du Rglement sur le matriel brouilleur du Canada.
Page 99 Certification

6200518 Certification.fm

Certification
FCC PART 15, CLASS A LIMITS: Supporting test records by Criterion Technology O.A.T.S. (NVLAP #0369), certified compliance by Hach Company. This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. Changes or modifications to this unit not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference, in which case the user will be required to correct the interference at his own expense.

The following techniques of reducing the interference problems are applied easily. 1. Disconnect the APA 6000 Analyzer from its power source to verify that it is or is not the source of the interference. 2. If the APA 6000 Analyzer is connected into the same outlet as the device with which it is interfering, try another outlet. 3. Move the APA 6000 Analyzer away from the device receiving the interference. 4. Reposition the receiving antenna for the device receiving the interference. 5. Try combinations of the above.

Page 100 Certification

6200518 Certification.fm

Das könnte Ihnen auch gefallen