Signa EXCITE HD 1.5T TwinSpeed System Installation


R

Copyrighte 2004 by General Electric Company, Inc. All Rights Reserved.

Operating Documentation

DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent, have notation damage in shipment written on all copies of the freight or express bill before delivery is accepted or signed for by a General Electric representative or a hospital receiving agent. Whether noted or concealed, damage MUST be reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and containers held for inspection by the carrier. A transportation company will not pay a claim for damage if an inspection is not requested within this 14 day period. File a report with D Call 18005483366 and use option 8. D Contact your local service coordinator for more information on this process. Rev. 08/15/2003

GE Medical Systems
GE Medical Systems: Telex 3797371 P Box 414, Milwaukee, Wisconsin 53201 U.S.A. .O. (Asia, Pacific, Latin America, North America) GE Medical Systems Europe: Telex 261794 Shortlands, Hammersmith, London W6 8BX U.K.

GE MEDICAL SYSTEMS
REV 1

LANGUAGE POLICY FOR SERVICE DOCUMENTATION DIRECTION 2128126


D THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.

WARNING

D IF A CUSTOMERS SERVICE PROVIDER REQUIRES A LANGUAGE OTHER THAN ENGLISH, IT IS THE CUSTOMERS RESPONSIBILITY TO PROVIDE TRANSLATION SERVICES. D DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD. D FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK, MECHANICAL OR OTHER HAZARDS.

D CE MANUEL DE MAINTENANCE NEST DISPONIBLE QUEN ANGLAIS.

AVERTISSEMENT

D SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRE LANGUE QUE LANGLAIS, CEST AU CLIENT QUIL INCOMBE DE LE FAIRE TRADUIRE. D NE PAS TENTER DINTERVENTION SUR LES QUIPEMENTS TANT QUE LE MANUEL SERVICE NA PAS T CONSULT ET COMPRIS. D LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRANER CHEZ LE TECHNICIEN, LOPRATEUR OU LE PATIENT DES BLESSURES DUES DES DANGERS LECTRIQUES, MCANIQUES OU AUTRES.

WARNUNG

D DIESES KUNDENDIENSTHANDBUCH EXISTIERT NUR IN ENGLISCHER SPRACHE. D FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENTIGT, IST ES AUFGABE DES KUNDEN FR EINE ENTSPRECHENDE BERSETZUNG ZU SORGEN. D VERSUCHEN SIE NICHT, DAS GERT ZU REPARIEREN, BEVOR DIESES KUNDENDIENSTHANDBUCH ZU RATE GEZOGEN UND VERSTANDEN WURDE. D WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE SCHLGE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.

D ESTE MANUAL DE SERVICIO SLO EXISTE EN INGLS.

AVISO

D SI ALGN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA QUE NO SEA EL INGLS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN SERVICIO DE TRADUCCIN. D NO SE DEBER DAR SERVICIO TCNICO AL EQUIPO, SIN HABER CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO. D LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN LESIONES PROVOCADAS POR CAUSAS ELCTRICAS, MECNICAS O DE OTRA NATURALEZA.

1 of 2

GE MEDICAL SYSTEMS
REV 1

LANGUAGE POLICY FOR SERVICE DOCUMENTATION DIRECTION 2128126

ATENO

D ESTE MANUAL DE ASSISTNCIA TCNICA S SE ENCONTRA DISPONVEL EM INGLS. D SE QUALQUER OUTRO SERVIO DE ASSISTNCIA TCNICA, QUE NO A GEMS, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, DA RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIOS DE TRADUO. D NO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E COMPREENDIDO ESTE MANUAL DE ASSISTNCIA TCNICA. D O NO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANA DO TCNICO, OPERADOR OU PACIENTE DEVIDO A CHOQUES ELTRICOS, MECNICOS OU OUTROS.

AVVERTENZA

D IL PRESENTE MANUALE DI MANUTENZIONE DISPONIBILE SOLTANTO IN INGLESE. D SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEMS RICHIEDE IL MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE TENUTO A PROVVEDERE DIRETTAMENTE ALLA TRADUZIONE. D SI PROCEDA ALLA MANUTENZIONE DELLAPPARECCHIATURA SOLO DOPO AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL CONTENUTO. D NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALLADDETTO ALLA MANUTENZIONE, ALLUTILIZZATORE ED AL PAZIENTE PER FOLGORAZIONE ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.

2 of 2

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

REVISION HISTORY
REV DATE PRIMARY REASON FOR CHANGE

1 . . . . . . . Nov 4, 2004 . . . Initial Release.

LIST OF EFFECTIVE PAGES


PAGE REV PAGE REV PAGE REV PAGE REV PAGE REV
Title Page . . . . . . . . . . . 1 Damage in Trans. . . . . Direction 2128126 . . . 1* * A ..................1 i...................1 TAB 1 GETTING STARTED 21649041 . . . . . . . . . . 11 to 113 . . . . . . . . . 1 TAB 2 PENETRATION PANEL 21649042 . . . . . . . . . . 21 to 25 . . . . . . . . . . 1 TAB 3 CABINET POSITIONING 21649043 . . . . . . . . . 31 to 33 . . . . . . . . . . 1 TAB 4 ACGD/PDU SETUP 21649044 . . . . . . . . . 41 to 43 . . . . . . . . . . 1 TAB 5 COOLING SYSTEM INSTALLATION 21649045 . . . . . . . . . 51 to 55 . . . . . . . . . . 1 TAB 6 CABLE & SYSTEM INSTALLATION 21649046 . . . . . . . . . 61 to 672 . . . . . . . . . 1 TAB 7 FINAL ENCLOSURE INSTALLATION 21649047 . . . . . . . . . 71 to 76 . . . . . . . . . . 1 TAB 8 OPERATOR WORKSPACE 21649048 . . . . . . . . . 81 to 812 . . . . . . . . . 1 TAB 9 SYSTEM MECHANICAL CHECKLIST 21649049 . . . . . . . . . 91 to 95 . . . . . . . . . . 1 TAB 10 APPENDIX 216490410 . . . . . . . 101 to 1018 . . . . . . . 1

* This revision number/letter corresponds to the indicated documents revision control system.

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

TABLE OF CONTENTS
SECTION PAGE

SECTION 1 GETTING STARTED . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 SECTION 2 PENETRATION PANEL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 SECTION 3 CABINET POSITIONING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 SECTION 4 HFD CABINET SETUP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 SECTION 5 COOLING SYSTEM INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51 SECTION 6 CABLE & SYSTEM INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61 SECTION 7 ENCLOSURE INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71 SECTION 8 OPERATOR WORKSPACE INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81 SECTION 9 SYSTEM MECHANICAL CHECKLIST COMPLETION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91 SECTION 10 APPENDIX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101

TABLE OF CONTENTS

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

SECTION 1 GETTING STARTED


TABLE OF CONTENTS
SECTION 11 12 13 14 15 16 17 18 19 PAGE SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM CONFIGURATION . . . . . . . . . . . . . . . . . . . . . . . . 12 SITE READY CHECK FOR MECHANICAL INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 REQUIRED TOOLS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 BASIC SYSTEM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 SYSTEM OPTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 PRODUCT DELIVERY INSTRUCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 DAMAGE IN TRANSPORTATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 PRODUCT LOCATOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 INSTALLATION PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 191 192 INSTALLATION FLOW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 INSTALLATION FLOWCHART EXPLANATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

110 INSTALLATION FLOWCHART . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 1101 PREREQUISITES FOR SYSTEM INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 1102 SYSTEM MECHANICAL INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110

11

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

11 SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM CONFIGURATION This manual provides instructions to complete the installation of Singa EXCITE HD 1.5T TwinSpeed System. Additional options are installed according to the installation manual delivered with the option.

12 SITE READY CHECK FOR MECHANICAL INSTALLATION Before equipment is delivered, the following must be complete to avoid delays and confusion. j Preinstallation work must be complete. Refer to Direction 5133302, Signa EXCITE HD 1.5T TwinSpeed Preinstallation, Preinstallation Checklist tab. j Magnet Delivery and Installation complete. Refer to Direction 2226318, GE 1.5T & 1.0T LCC Signa MR/i Magnet Delivery and Installation. j Magnet Subsystem should be installed. Refer to Direction 2192624, GE 1.5T & 1.0T LCC Active Shield Magnet and Cryogen Subsystem, Section 1.

13 REQUIRED TOOLS

WARNING!
FERROUS MATERIAL HAZARD! IF MAGNET IS ENERGIZED, THE CRIMP TOOL, BLOWER BOX, AND OTHER TOOLS AND PARTS REQUIRED FOR THIS INSTALLATION THAT CONTAIN FERROUS MATERIAL AND WILL BE STRONGLY ATTRACTED TO MAGNET AND MAY BECOME DANGEROUS PROJECTILES. KEEP ALL FERROUS TOOLS AT LEAST 10 FEET AWAY FROM THE MAGNET. Tools to install the Signa EXCITE HD 1.5T TwinSpeed system are listed below:

TABLE 12 INSTALLATION EQUIPMENT ITEM 1 2 3 4 5 6 GE PART NUMBER 46271138G1 2319156 DESCRIPTION Ramp for removing cabinets from pallets for International shipments (See Note 1) Wrecking bar Claw hammer, 3/4 lb Restricted Access Control Kit. Contains two plastic warning signs for posting at site during installation and service activity. 4 foot or equivalent carpenter level Aluminum platform ladder, 47.5 inches (1206.5mm) (See Note 1) (Continued)

12

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

13 REQUIRED TOOLS (Continued)


TABLE 12 (Continued) INSTALLATION EQUIPMENT ITEM 7 GE PART NUMBER 2134776 DESCRIPTION Gradient Cable Crimper/Stripper Kit (Note 2) consisting of: S 2134586 Cable stripping tool S 21345862 Stripping tool replacement blade S 2134587 Cable slicer S 46282853P1 Ratcheting crimper S 2135839 1/2 inch terminals S 21358392 3/8 inch terminals NonMagnetic Tool Kit Universal (See Note 2) Both metric and inch NonMagnetic Tool Kits needed. May substitute both of the following kits: S 46320273G1 NonMagnetic Tool Kit Metric S 46320273G2 NonMagnetic Tool Kit Inch 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 Note 1 2 46301450G1 2384858 2352193 5111565 46198094P1 Supplied as part of Signa. Supplied by GE until turnover of system to customer, then available as part of a GE Cryogen and/or Service Contract. Fiber optic connector repair kit (See Note 2) Times Microwave LMR 1200 Stripping Tool (See Note 1) Times Microwave LMR 600 Stripping Tool (See Note 1) Times Microwave cutting Tool (See Note 1) Wrist grounding strap Volt Meter Extension cords, with ground conductor Power strip, grounded type, with minimum of five outlets Plastic or aluminum flashlight Assorted crimp tools. Nonmagnetic level Nonmagnetic tape rule, 12 ft Assorted drill bits Inspection mirror Hobby and utility knives

46320273G3 or G4

13

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

14 BASIC SYSTEM The basic Signa EXCITE 3.0T system, Illustration 11, consists of the following major equipment: D System cooling equipment (See Illustration 12) D 1.5T (15 kilogauss) LCC Active Shield Magnet with Magnet Enclosure D Magnet Accessories D Shield/Cryo Cooler Compressor Cabinet D TwinSpeed Gradient (TRM) and RF body coils and 3.0T General Purpose Head Coil D Main Disconnect Panel (GE supplied or customer supplied) D System electronics cabinets: HFD Cabinet containing HFD Gradient drivers and Power Distribution Unit module with unregulated transformer 200/380/400/415/480 Volt, 50/60 Hz with power filter EXCITE RFS Cabinet containing the Multi Generational Data Acquisition (MGD) chassis with EXCITE 8 or 16 Channel, ReFLex400 or ReFLex800, power supplies for the Magnet Enclosure system components. Twin Accessory Cabinet containing the Gradient Switch and the vacuum pump for Quiet Technology Magnet Enclosure D Operator Workspace equipment: GOC Computer Cabinet with Linux PC, Mouse and Mouse Pad, LCD panel, and chair D Pneumatic Patient Alert System D Patient Transport Table and cradle D Patient accessories such as: a phantom kit, patient log book, head cushion and sponges, chin and forehead straps, body wedges, knee cushions, and security/restraint straps D Gating accessories which include: patient cardiac leads, peripheral gating probe, and respiratory bellows

14

GE MEDICAL SYSTEMS REV 1 14 BASIC SYSTEM (Continued)

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

MAGNET MONITOR (MSM)

BLOWER BOX (MG6) SPT PHANTOM SET STORAGE CART NOTE: STORAGE CART PENETRATION NOT USED WITH TRUE SHIELD COOLER PANEL MOBILE SITE SYSTEM COMPRESSOR CABINET (PP1) (MS5) MAGNET ENCLOSURE (MG2) LIGHTWEIGHT PATIENT TRANSPORT (PT1) GE MDP (M3088TM) Refer to catalog flowchart to determine if required or optional

HFD/PDU CABINET (MR3)

RFS CABINET (MR2)

TWIN ACCESSORY CABINET (TAC)

OPERATOR WORKSPACE (OW1)

Gradient Coil Chiller Option or Coldhead Chiller Option Indoor/Outdoor Air Cooled Chiller (M3088TL) Refer to catalog flowchart to determine if required or optional

Gradient Water Heat Exchanger Option Indoor Air Cooled Chiller (M3088TK)

REMOTE CONTROL PANEL (RCP) MR COMMON CHILLER (MRCC) (GWHX) 1.5T MNS AMP CABINET (MR10)

SIGNA EXCITE 1.5T SYSTEM ILLUSTRATION 11

15

MAIN DISCONNECT PANEL (MDP) MNS Option

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

15 SYSTEM OPTIONS This section lists options for the Signa EXCITE 1.5T TwinSpeed system which have site preparation impact. D 700 Volt Amp UPS for MR Magnet Monitor Equipment (E4504AG) Note The GE MDP (M3088TH) must be ordered and installed when the UPS for MR Magnet Monitor (E4504AG) is installed. The GE MDP provides Emergency Off control of the output of the UPS for system safety. D Advantage Workstation (AW) Note LCD monitor is recommended for a AW to be located near the Signa system Operator Workspace due to gauss field proximity limits. D Bar Code Reader (M1090PM) for use in conjunction with HIS/RIS software option. D Additional Patient Transport table with cradle for WideOpen Enclosure (M1090TL). D MultiNuclear Spectroscopy (MNS) options. D Other hard copy devices and patient accessories.

16 PRODUCT DELIVERY INSTRUCTIONS The Shipping Document lists catalog numbers delivered. Review and confirm that order is delivered complete. Check impact on installation schedule if Catalogs and/or packing boxes are missing and/or noted as shipped short. Labels that are attached to the outside of packing boxes summarize box contents. The labels are color coded for different installation areas as follows: D Green Equipment Room D Purple Operators Room D Orange Magnet Room D Yellow Postinstallation (Phantoms, Coils). Product Delivery Instructions (PDI) specify box contents, part numbers, and shipping procedure. The PDI is numbered according to the catalog number (for example, PDI M3333TA is for the Fixed Site Installation Kit). Lists of items included with each box are detailed by separate checklists, or a separate sheet that provides a summery of that box contents. Refer to PDI and packing lists for information specific to your shipment. A set of service and operator manuals is delivered with the Fixed Site Kit. Refer to checklists packed with Technical Publication boxes for a list of delivered documents.

16

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

17 DAMAGE IN TRANSPORTATION All packages should be closely examined at time of delivery. If damage is apparent, have notation damage in shipment written on all copies of the freight or express bill before delivery is accepted or signed for by a General Electric representative or a hospital receiving agent. Whether noted or concealed, damage MUST be reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and containers held for inspection by the carrier. A transportation company will not pay a claim for damage if an inspection is not requested within this 14 day period. File a report with D Call 18005483366 and use option 8. D Contact your local service coordinator for more information on this process. Rev. 08/15/2003

18 PRODUCT LOCATOR The Global Install Base Database (also known as Product Locator System) tracks shipment, transshipment, and field location of the serialized models. There are now two methods for submitting Product Locator information. 1. At this time, for U.S. ONLY, the preferred method of submitting information is the FE Site Verification Web Site at: http://gein2.med.ge.com/gib The FE Site Verification consists of three components that are available on the web from the main menu. They are: D Install/deinstall product locator model and serial numbers D Add/modify ship to address information D Update CARES FE data for primary/secondary FEs To obtain a copy of the FE training tutorial for using this website, a downloadable copy is available at the following: Product Locator Support Central Page http://supportcentral.ge.com/15563 2. One Shipping Card is filled out and submitted when shipped (extra cards are supplied for transshipments between storage and distribution points), and the Installation Card and extra shipment cards are attached.

Verify that serial and model number on each rating plate matches installation card numbers before removing installation card. Note that there may be one or more shipment cards and bar code labels with the installation card. These shipment cards are used to trace the transfer of serialized units between various inventory storage and distribution points until the product reaches its final installation destination. Process just the installation card and discard any extra shipment cards and labels.

17

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

19 INSTALLATION PROCEDURE 191 Installation Flow Note All onsite construction must be completed before equipment is delivered and installation starts. Attempting to install the system while construction is being completed will impact installation efficiency and further delay site completion. Making sure that all preinstallation and construction work is completed before equipment is delivered will usually result in an earlier turnover date. The flow chart on the next page should be followed for an orderly and efficient installation of the Signa EXCITE HD 1.5T TwinSpeed system. Note that many procedures may be performed in parallel and may be performed in any order according to the specific situation of each site. Note The flow chart references the Tab number and/or Section number for the respective installation procedures. These procedures are similar to cable maps in that they are to be removed from this binder and used at the location where hardware installation is being performed. The Procedures must be returned to the binder for future reference upon completion of the installation. This installation flowchart has been developed assuming that all Signa EXCITE HD 1.5T TwinSpeed equipment has been delivered together. Make sure that every part required is available before starting. 192 Installation Flowchart Explanation Shown below are three examples of the flow chart Installation Procedures and an explanation of the areas within the procedures. Use the correct Magnet Manual that applies to the System being installed to locate the installation procedure listed below DIR 2325852 GE LCC300 Active Shield Magnet & Cryogen Subsystem Complete Section 1

Section within manual to be completed.

Tab within this manual where the Section listed below is located Section within the Tab where the Installation Procedure is located

TAB 6 V Section 61 UNPACKING CABLES

LEGEND FOR FOLLOWING INSTALLATION FLOWCHART

MX
OR

INDICATES MILESTONE COMPLETION GOAL IN DAYS INDICATES AND OPTIONAL PATH DEPENDING ON SYSTEM CONFIGURATION

18

GE MEDICAL SYSTEMS REV 1 110 INSTALLATION FLOWCHART

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

1101 PREREQUISITES FOR SYSTEM INSTALLATION Site Installations Required Prior to System Mechanical Installation MARK OFF PERFORMED ITEMS

n
DIR 5133302 Signa EXCITE HD 1.5T TwinSpeed PreInstallation Preinstall Checklist

IMPORTANT : PER ISO 9001 REQUIREMENTS

DIR 2226318 GE 1.5T & 1.0T LCC Signa MR/i Magnet Delivery and Installation Complete Magnet Delivery Install

DIR 2192624 GE 1.5T & 1.0T Active Shield Magnet & Cryogen Subsystem Complete Section 1

COMPLETE SITE ELECTRICAL WIRING AND HVAC

VENDOR DOCUMENT SYSTEM COOLING CABINET INSTALLATION

MAGNET ROOM EXHAUST FAN INSTALLED & OPERATIONAL

MAGNET ROOM CRYOGEN VENT TO BUILDING INSTALLED

MDP (MAIN DISCONNECT PANEL) INSTALLED

19

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

1102 SYSTEM MECHANICAL INSTALLATION

Procedures to be Completed by Mechanical Installers

A
START

MARK OFF PERFORMED ITEMS

IMPORTANT : PER ISO 9001 REQUIREMENTS

TAB 1 V Section 16 PRODUCT DELIVERY INSTRUCTIONS

TAB 1 V Section 17 DAMAGE IN TRANSPORTATION

TAB 1 V Section 18 PRODUCT LOCATOR OPTION DIRECTION 15336 OXYGEN MONITOR INSTALLATION

LOCKOUT & TAGOUT MAIN DISCONNECT TAB 2 V PENETRATION PANEL INSTALLATION

110

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

1102 SYSTEM MECHANICAL INSTALLATION (Continued)

Procedures to be Completed by Mechanical Installers

B
TAB 3 V POSITIONING EQUIPMENT ROOM EQUIPMENT

MARK OFF PERFORMED ITEMS

IMPORTANT : PER ISO 9001 REQUIREMENTS

TAB 4 V HFD CABINET SETUP

TAB 6 V Section 612 UNPACKING CABLES

TAB 5 V COOLING SYSTEM INSTALLATION

TAB 6 V Sections 621 and 622 CABINET POWER AND GROUND CABLE INSTALLATION

Service Methods CDROM V Set Up & Cal GROUNDING & LEAKAGE CURRENT CHECKS 1. Ground Resistance Checks NOTE: Hard wired signal cables are not to be connected to subsystem cabinets prior to this check. Fiber optic cables may be connected and not impact this check.

111

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

1102 SYSTEM MECHANICAL INSTALLATION (Continued)

Procedures to be Completed by Mechanical Installers MARK OFF PERFORMED ITEMS

C
OPERATOR / EQUIPMENT ROOM

IMPORTANT : PER ISO 9001 REQUIREMENTS MAGNET ROOM

TAB 6 V Section 62 EQUIPMENT ROOM CABLE INSTALLATION TAB 8 V OPERATOR WORKSPACE INSTALLATION

M1

TAB 6 V Section 63 MAGNET ENCLOSURES / CABLE INSTALLATION

TAB 6 V Section 64 EQUIPMENT TO MAGNET ROOM CABLE INSTALLATION

TAB 7 V FINAL ENCLOSURE & PATIENT TRANSPORT INSTALLATION

TAB 2 V Section 23 INSTALL PENETRATION PANEL COVERS

TAB 9 V SYSTEM MECHANICAL CHECKLIST COMPLETION

M2

MECHANICAL INSTALLATION COMPLETE

112

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

1102 SYSTEM MECHANICAL INSTALLATION (Continued)

Procedures to be Completed by FE and Applications n MARK OFF PERFORMED ITEMS IMPORTANT : PER ISO 9001 REQUIREMENTS MR Service Methods CDROM Direction 5133315x V REFER TO: Install/Signa EXCITE HD System Installation Callibration Flow

Applications/Training V Turn site over to Applications who will instruct users.

TURN OVER TO CUSTOMER

113

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

SECTION 2 PENETRATION PANEL INSTALLATION


TABLE OF CONTENTS
SECTION 21 22 23 PAGE PREPARE SCREEN ROOM OPENING FOR PENETRATION PANEL . . . . . . . . . . . . . . . . . . . . . . 22 INSTALL PENETRATION PANEL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 INSTALL PENETRATION PANEL COVERS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 231 232 INSTALL PENETRATION PANEL STANDOFF HARDWARE . . . . . . . . . . . . . . . . . . . . 24 INSTALL PENETRATION PANEL COVERS (After Cable Installation Complete) . . 2-5

2-1

PENETRATION PANEL

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

21 PREPARE SCREEN ROOM OPENING FOR PENETRATION PANEL 1. 2. 3. If still in place, remove vendor supplied blank panel. Save mounting hardware of either method shown below for installation of Penetration Panel. To make sure of good conductivity, use a fine sandpaper or scouring pad to clean surface of screen room wall which will be in contact with Penetration Panel. Do not sand surface of Penetration Panel as it is coated. Grade 8 bolts with black oxide coating should be used as mounting hardware. Tighten to torque of 20 lbft (1.5 to 2 turns after lock washer flattens). Hold Down Method (Top View) Brass Sealing Plate Method (Top View) Panel

Panel

Screen Room Wall

2-2


2 PENETRATION PANEL

GE MEDICAL SYSTEMS REV 1 22 INSTALL PENETRATION PANEL

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Install Penetration Panel using same method and hardware for mounting of blank panel removed in previous procedure.

Note Install the Penetration Panel Floor plate that is supplied with the cover kit prior to routing and connecting cables to Penetration Panel. Refer to Installation Drawing furnished with the Cover Kit.
1. Assemble Penetration Panel around helium lines (Runs 621 and 622) using: (7) #1032 Nylock Lock Nuts (46208937P6) (7) 0.2 ID, 0.5 OD Flat Washers (1000904P425) (7) #1032 x 3/8 Screws (46208560P82) Seal all joints with copper foil tape with a conductive adhesive (3M #1181 or 46-258218P4, 5, or 6) for RFtight integrity. Fill inside of J60 and J61 guides with Bronze Wool (46318068P1) that was supplied with magnet.

2. 3.

2-3

PENETRATION PANEL

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

23 INSTALL PENETRATION PANEL COVERS

Note The Penetration Panel Cover can be installed only after all connections and System installation has been completed.
231 1. 2. INSTALL PENETRATION PANEL STANDOFF HARDWARE

Verify that cover will enclose finished wall opening. The distance from the top of Penetration Panel to wall opening must be less than 41/2 inches (114mm). If not, new mounting stud holes in Penetration Panel may be required. Place (8) Reinforcing Plates (46301211P1) over each hole pair on both sides of Penetration Panel and attach using: (16) 5/1618 x 9/16 Brass Nut (46208935P9) (8) 5/1618 x 18 inch Threaded Rod (46301213P2) (8) 5/1618 x 1 inch Barrel Nut (46301212P1) Measure distance from face of Pen Panel to equipment room wall surface. Add six inches (152mm) to obtain length threaded rod should extend from panel. Repeat step 3 procedure for exam room cover. Adjust threaded rods with attached barrel nuts to the previously calculated dimensions from steps 3 and 4. Secure to panel by tightening brass nuts on both sides of Penetration Panel.

3. 4. 5.


2-4

PENETRATION PANEL

GE MEDICAL SYSTEMS REV 1 232 1. 2. 3. 4. 5. 6. 7.

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

INSTALL PENETRATION PANEL COVERS (After Cable Installation Complete)

Remove all barrel nuts from both sides of Penetration Panel. In the Magnet room, temporarily install Top (46306936P1) and Bottom (46306964P1) Cover assembly on threaded rods and secure with four barrel nuts. Adjust position of bottom cover section and mark two lower fastener holes. Remove covers and drill hole in lower cover at each mark. Attach top and bottom covers with furnished 1032 nuts and screws. Repeat procedure for Equipment room cover. After all installation procedures are completed in Magnet and Equipment rooms, install assembled covers on threaded rods and secure with four barrel nuts.


6 2-5

PENETRATION PANEL

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

SECTION 3 CABINET POSITIONING


TABLE OF CONTENTS
SECTION 31 32 33 34 PAGE CABINET INSTALLATION REQUIREMENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 POSITION HFD CABINET . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 POSITION AND SECURE RFSYSTEM CABINET . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 POSITION TAC CABINET . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33

3-1

CABINET POSITIONING

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

31 CABINET INSTALLATION REQUIREMENTS Positioning of cabinets is critical for installing cables and postinstallation system operation. Refer to architectural site layouts for: D Location of all cabinets and related equipment. D For seismic anchoring (where required)

32 POSITION HFD CABINET

WARNING!
DUE TO WEIGHT OF CABINET, AT LEAST TWO PEOPLE ARE REQUIRED TO MOVE CABINET
1. 2. 3. 4. 5. Move cabinet to final position as indicated by site layout drawings. Antitip bars for the HFD Cabinet are shipped in the RFS Cabinet. Remove from RFS Cabinet for this procedure. Attach bars to cabinet. Adjust leveling pads to level cabinet. From the Cabinet ID Label Sheet (5126718), locate the GRADIENT/PDU cabinet magnet and attach to front of cabinet as shown.

1 4

3-2

CABINET POSITIONING

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

33 POSITION AND SECURE RFS CABINET Comply with UL requirements or local codes for securing the cabinet to the floor with locally supplied brackets and floor anchors. 1. 2. Position RFS cabinet to final position as indicated by site layout drawings. From the Cabinet ID Label Sheet (5126718), locate the RF/SYSTEMS CABINET magnet and attach to front of cabinet as shown.

2 1

34 POSITION TAC CABINET Comply with UL requirements or local codes for securing the cabinet to the floor with locally supplied brackets and floor anchors. Note

If supplied in cabinet, Run 975 is a service tool. It will not be part of the installation.
1. 2. Position TAC cabinet as indicated by site layout drawings. From the Cabinet ID Label Sheet (5126718), locate the TAC CABINET magnet and attach to front of cabinet as shown. 1

3-3

CABINET POSITIONING

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

SECTION 4 HFD CABINET SETUP


TABLE OF CONTENTS
SECTION 41 42 43 PAGE INSTALL FAN ASSEMBLY AND ATTACH INSITE MAGNET . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 INPUT VOLTAGE SELECTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 CIRCUIT BREAKER DIP SWITCH SETTINGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43

4-1

HFD CABINET SETUP

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

41 INSTALL FAN ASSEMBLY AND ATTACH INSITE MAGNET 1. 2. 3. 4. Remove Fan Assembly from shipping container. Remove two side trim panels (two screws from each). Requiring two people, place the Fan Assembly on top of cabinet with the cable connector facing the rear of the cabinet. Secure Fan Assembly in place with (4) four screws and washers, two on each side of cabinet. The fasteners are provided in a plastic bag located in PDU module. Rear PDU covers will need to be removed for access to plastic bags with fasteners. Install two side trim panels with previously removed screws. Connect cable from SGA Power Supply to the fan connector. Make sure the GRADIENT/PDU cabinet magnet, part of the Cabinet ID Label Sheet (5126718), is attached to the front of cabinet as shown. 3

5. 6. 7.

2 5 6

4-2

HFD CABINET SETUP

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

DANGER!!
LETHAL VOLTAGES ARE PRESENT WITHIN THE PDU. MAKE SURE THAT POWER AT THE MAIN DISCONNECT IS OFF, LOCKED, AND TAGGED BEFORE PROCEEDING.
42 INPUT VOLTAGE SELECTION 1 2 3 Using a voltmeter or other voltage indicating device, check to be sure that no voltage is being applied to the input of the PDU. Remove the plate from the front of the PDU marked INPUT CIRCUIT BREAKER CURRENT SELECT / INPUT VOLTAGE SELECT. For each of the three phases, be sure the wire is connected to the proper terminal for the actual input voltage at the site and that the terminal screw is tight to make a good connection.
PHASE A (SHOWN WIRED FOR 480V) PHASE B PHASE C

200V

208V

380V

400V

415V

480V

200V

208V

380V

400V

415V

480V

200V

208V

380V

400V

415V

43 CIRCUIT BREAKER DIP SWITCH SETTINGS 1 2 Remove small cover plate under the main circuit breaker to access DIP switches controlling the circuit breaker operation. Table below shows the position of the DIP switches for each input voltage selection. The switches immediately under the letter L and I are the only ones that should be changed.

INPUT DIP SWITCH VOLTAGE L SETTING

480V

DIP SWITCH I SETTING

INPUT VOLTAGE

200V 208V 380V 400V 415V 480V If these switches are present, set both of them in the DOWN position as shown. Set these switches according to the input voltage shown at the left. Set these switches according to the input voltage shown at the right.

200V 208V 380V 400V 415V 480V

4-3

HFD CABINET SETUP

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

SECTION 5 COOLING SYSTEM INSTALLATION


TABLE OF CONTENTS
SECTION 51 52 53 54 55 56 57 PAGE ROUTE/CONNECT COOLANT HOSES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52 INSTALL GROUND CLAMPS ON TEES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52 CONNECT COOLANT HOSES TO TEES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53 INSTALL CONNECTOR WRAPS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53 ROUTE/CONNECT GROUND WIRES TO MAGNET . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54 COOLANT VALVE INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55 COOLANT CABINET OPERATING INSTRUCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55

5-1

COOLING SYSTEM

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

51 CONNECT MAGNET COOLANT HOSES TO TEES The Cooling Cabinet must be installed and electrically powered prior to the start of this procedure. Magnet is delivered to site with four 1/2 inch hoses connected to the Front End Bell. The Twin Gradient Hose Assembly (2295225) is delivered to site and also has four duplicate 1/2 hoses. Discard the 1/2 inch black (22936102) and blue (22936103) hoses that are part of 2295225. Connect the IN and OUT coolant hoses from the Cooling Cabinet to the 1/2 inch hoses attached to the Front End Bell using the Tee Connectors and clamps shown below. 2 Locate Twin Gradient Hose Asm (2295225). Remove and discard the1/2 inch hoses connected to each Tee Connector, that are part of 2295225. Reuse clamps for connecting Tee Connectors to existing 1/2 inch hoses on enclosure that are connected to Front End Bell.

TO COOLING CABINET

NOTE: Dock assembly is not shown for clarity.

Coolant OUT (Black)

Coolant IN (Blue)

1 Route the IN and OUT Coolant hoses from the Cooling Cabinet underneath the magnet to the front of the enclosure. Note: It is important that the IN and OUT hoses are connected to the respective IN and OUT ports on the Cooling Cabinet.

52 INSTALL GROUND CLAMPS ON TEES 1 Verify that one end of each Jumper Wire Asm (5126241) is stripped bare of insulation for 1/2 inch (12.7mm) at one end and insert stripped end into hole of clamp. Tighten the wire securing screw.

Place Ground Clamp (5126242) over hex portion of 3/4 inch hose barb and tighten securely. Repeat steps 1 and 2 for other coolant tee.

5-2

COOLING SYSTEM

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

53 CONNECT COOLANT HOSES TO TEES

NOTE: Dock assembly is not shown for clarity.

TO COOLING CABINET

Coolant OUT Coolant IN (Black) (Blue) 1 Connect the IN and OUT coolant hoses to the correct Tee Connectors at the front of the enclosure. Secure with hose clamps.

54 INSTALL CONNECTOR WRAPS 1 Wrap each Tee Connector with 177 x 200 Connector Wrap (22361302)

1 Coolant OUT (Black) Coolant IN (Blue)

5-3

COOLING SYSTEM

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

55 ROUTE/CONNECT GROUND WIRES TO MAGNET

IMPORTANT !!! Make sure Jumper Wire Asm 5126241 is secure from motion along its entire length.

Note: If Run 040, magnet ground cable to RF Room, is connected to one of the front legs, the attaching nut can not be used for attaching any of the gound wires routed in this procedure. If needed, remove nut from one of the unused ground studs on another leg and use to attach ground wire to above stud.

1 Route one of the ground wires around the left leg to ground stud on inside of leg. Secure wire to leg with adhesive backed Tie Wrap Base (46208747P1) and Tie Wraps (5112522). Securely fasten the ring terminal to ground stud. 2 Route other ground wire to right leg ground stud. Secure and attach ground wire as shown.

5-4

COOLING SYSTEM

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

56 COOLANT VALVE INSTALLATION 1. The Inline Drainage Valves shown above are installed in coolant hoses near the Penetration Panel on the Equipment Room side. The valve end must be positioned towards the Cooling Cabinet and the Tee end towards the TRM coil. Connect IN and OUT coolant hoses to respective piping hose barbs, on outdoor or indoor unit. Typical coolant Supply (IN) and Return (OUT) piping for outdoor cooling unit that is provided by customer in equipment room.

2.

INLINE DRAINAGE VALVES

57 COOLANT CABINET OPERATING INSTRUCTIONS Consult vendor supplied manual that was delivered with Cooling Cabinet for installation, startup, and operating instructions and procedures.

5-5

COOLING SYSTEM

This end towards Cooling Cabinet

This end towards TRM Coil

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

SECTION 6 CABLE & SYSTEM INSTALLATION


TABLE OF CONTENTS
SECTION 61 PAGE CABLE INTERCONNECT DOCUMENTATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62 611 612 62 621 622 623 624 625 626 627 63 631 632 633 634 635 636 637 638 639 6311 64 641 642 643 644 CABLE INTERCONNECT MAPS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62 SORTING AND ROUTING CABLES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 611 HFD CABINET CABLE CONNECTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 612 EQUIPMENT ROOM CABINET POWER CABLE CONNECTIONS . . . . . . . . . . . . . . . . 615 GRADIENT CABLE INSTALLATION IN EQUIPMENT ROOM . . . . . . . . . . . . . . . . . . . . 616 INSTALLATION OF MAGNET MONITORING CABLES . . . . . . . . . . . . . . . . . . . . . . . . . 625 INSTALL RUN 701 RF DOOR SWITCH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 625 INSTALLATION OF FIBER OPTIC CABLES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 626 INSTALLATION OF REMAINING EQUIPMENT ROOM CABLES . . . . . . . . . . . . . . . . . 631 CONNECTION OF MAGNET GROUND, RUN 040 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 634 INSTALLATION OF BORE LAMP ASSEMBLY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 634 MAGNET ROOM FIBER OPTIC PAC/SRI INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . 635 GRADIENT CABLE INSTALLATION IN MAGNET ROOM . . . . . . . . . . . . . . . . . . . . . . . 637 INSTALLATION OF REAR PEDESTAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 645 LEVELING OF BRIDGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 648 INSTALLATION OF DRIVE BELT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 649 ROUTE AND CONNECT CABLES AT REAR PEDESTAL . . . . . . . . . . . . . . . . . . . . . . . 652 INSTALLATION OF AIR HOSE FROM REAR OF MAGNET TO BLOWER BOX . . . . 655 DOCK ADJUSTMENT FOR PATIENT TRANSPORT GAP . . . . . . . . . . . . . . . . . . . . . . . 657 TIMES MICROWAVE RF CABLES INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 658 VACUUM HOSE INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 662 PNEUMATIC PATIENT ALERT INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 664 EMERGENCY OFF CONNECTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 672

EQUIPMENT ROOM CABLE INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 612

MAGNET ROOM CABLE INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 633

6310 ROUTE AND CONNECT CABLES AT FRONT OF MAGNET ENCLOSURE . . . . . . . . 656 INSTALLATION OF CABLES BETWEEN EQUIPMENT ROOM AND MAGNET ROOM . . . . . . 658

6-1

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

61 CABLE INTERCONNECT DOCUMENTATION The cable maps illustrate the Signa EXCITE HD 1.5T TwinSpeed system interconnect for all supplied cables and interface with customer supplied wiring. D Interconnects for additional cables supplied with other options are illustrated in the installation manual shipped with the option. Instructions for connections for optional laser cameras are supplied with option. Connection procedures for other options are covered in the installation manual shipped with the option. Spectroscopy option cabling can be found on page 69. D Procedures for connections made at the Magnet Enclosure, Operator Workspace, cabinets, etc. are detailed in the respective installation section within this manual. D Interconnect details for magnet subsystem are covered in the applicable magnet subsystem manual. 611 CABLE INTERCONNECT MAPS For the convenience of the mechanical installers, a duplicate System Interconnect is provided. This will allow the install team to post one interconnect map in the Magnet Room and the other map in the Equipment/Operator Room. This will improve the efficiency of the installation when marking off cables when their connections are complete. Foldout EXCITE HD 1.5T TwinSpeed System Interconnect Cable Map is provided on the following pages: Description Page

Signa EXCITE HD 1.5T TwinSpeed System Interconnect (Fixed Site) . . . . . . . . . . . . . . . . . . . . . . . . 63 Signa EXCITE HD 1.5T TwinSpeed System Interconnect (Fixed Site) (DUPLICATE) . . . . . . . . . . . 65 Signa EXCITE HD 1.5T TwinSpeed System Cooling Cabinet Interconnect Options . . . . . . . . . . . . . 67 Signa EXCITE HD 1.5T TwinSpeed System Interconnect for 2KW MNS Option . . . . . . . . . . . . . . . . 69 Tape maps in a convenient location in the Magnet Room and the Equipment/Operator Room to check off routed cables. Additional information on each cable is located in the Appedndix. Upon completion of connecting cables, return foldout maps to the binder for future reference.

6-2

CABLE & SYSTEM INSTALLATION

FACILITY EMERGENCY POWER 1

1 5 1 1

A12 (SERVICE LAPTOP) OW1 A15 (HOST COMPUTER) USB

1056
10 A18 (SERVICE KEY) OW1 A13 (MODEM) DB9

PORT 6 PORT 8 OW1 A17 (LAN HUB) PORT 2

13 TO SITE ETHERNET

FACILITY POWER

OM1
(OXYGEN MONITOR)

FILTERS FOR FACILITY POWER, EMERGENCY POWER, AND AUXILIARY WARNING DEVICE EXHAUST FAN 1

MS4
1 1 (ERU)

EXAM ROOM LIGHTS

Signa EXCITE HD 1.5T TwinSpeed, Release 12.x SYSTEM INTERCONNECT (FIXED SITE)
NOTES:
1 CUSTOMER FURNISHED Magnet and superconducting shim power supplies are not included with system, but may remain in place if purchased as an option. Adaptors selected according to coil used (head, surface) and magnet polarity. Typical connection for normally closed emergency off switch Optional: May not be present on all systems. Magnet ramping cables pass through ramp port on PP1 plate. Customer supplied telephone line to be connected from UPS to Modem (MSM3) or optional MUX . Additional telephone line would be needed if optional MUX is installed. Optional: Used only if High Order Shim is installed. PP1 PHPS J155 terminator relocated to High Order Shim in Twin Accessory Cabinet (TAC). Body Hybrid To Body Coil Connections depend on magnet field polarity. Customer supplied telephone line for Insite Modem. Refer to next page for cooling system thats designed for your site. Source of water supply for Cold Head Compressor depends on system used. Refer to Tab 1, Direction 2308287, Gradient Cable Installation, for instructions on tywrapping ground cables to each gradient cable between TAC and PP1. Ethernet cable is customer supplied. RUN 1066 and modem may not be required if connection to InSite is provided thru site ethermet connection at OW1 A7 PORT 8. J5 J1 2 3

606
(MAGNET)

MG3
(REAR PED)

MS1

1 OW1 A6 (HOST MONITOR) R,B.G IN (COLOR) 1f FACILITY POWER 1 3f OW1 A3 (LCD DISPLAY)

1254
TO PDU (S/C SHIM P.S.)

1063 1065 1068 1069 1082 1064 1070


J2

MONITOR 1

RS232 USB SCSI 2 Power

1066 1059 1067

13

MS7

603 604 601

1 5

RF SHIELD ROOM COMMON GROUND 5 J17 6

040
5

MONITOR 2 ON PCI BIT3 SCSI 1

457
RED (2) BLK (2)

458 603 604 F2 F4 601 602 624 621 622 MS1 A5 A2


(PRESSURE TRANSDUCER)

(REMOTE O2 SENSOR)

OM3

4 5 6 7

OW1 A5 (DASM)

T PORT 7

(MAGNET P.S.)

MS6

602

(TERMINAL BOX) P403

MS1 A3 A1 MS1

T 1 OW1 A20 (SCIM)

1083
OW1 A7 (KEYBRD) OW1 A8 (MOUSE)

J1 OW1 A11 J12 J8 (POWER J3 J7 J11 STRIP) J7

1071
FAN

WATER

OW1 A16 (SCSI TOWER)

F1
11 (COLD HEAD CMPRSSR) A1JR

MS5A1

623 621 622 MS5A5


(WATER FLOW)

F3

(MAGNET SERV TURRET) (SHIELD COOLER COLD HEAD) SLEEVE COLDHEAD PREAMP J2 J1

MS1 A2

8 9 10 11

1075 1074 1077


J5 J2 J4 OW1 A19 (CABINET PDU)

1256
J6 J15 J18 OW1 A21 (GOCAA) J8

833
FJ3 FJ4

FEED THRU (J60 & J61) (PENETRATION PANEL) (YELLOW) J48 J10 PP1 A17 (REPEATER PANEL)

OW1 OW1 A1 A10 (OW TABLE) 1 (SPKR) (GREEN)

OW1 A2
(GOC COMPUTER CABINET) (BROWN) J5 J10

PP1

826
J9 (MAGNET MONITOR) J1

827 MSM1
J8 J7

A1 J1 FJ1

832

12 13

1085
10 7

J7 J18 J10

1061 1060

J16 J17

J3

J1

825 824
5 F/O

829 828
FJ2

830

MG3 TB1
(REED SWITCH)

831

TO UPS CRYO COOLER COMPRESSOR TO MAG MON MODEM

MSM4 939
(UPS)

5 STEREO MUSIC RACK IN BORE

1081

823
5

940
(MODEM) 7 MUX 7 PATIENT ALERT CONTROL BOX

3 F/O

848 711 712 716


(J9,10) J4 (J11,12,13) (PAC I/F) BULB P. GATING RESP ECG PPG

862

J1 (RGHT OPER CONTROL)

MSM3

1045 1196
J1

PP1 A17 A1 J91 F/O REPEATER BOARD

711 712
2 F/O

MG2 A4
(OPER. DISPLAY)

MG2 A5

MG2 A33
(SRI3)

(PAC 3) J2 POWER

MG2 A11

ALERT AIR LINE FEED THRU COOLANT FEED THRU COOLANT FEED THRU SHIM FILTER VACUUM FEED THRU
(TWIN GRAD FILTER)

J4 J5 J2 J10 J20 J1 J3

897 1261 422 1260 421

J3 J4 J4

J2

863

(LEFT OPER. CONTROL) J2

MG2 A5

(MAIN DISCONNECT PANEL)

MDP

J5 (HFD CABINET) (ORN/YEL)

974 1179 1180 1181 976 977 978 989 1081 1254
4 F/O 8 T

J12

MR3

J2 J7 J8

TAC
(TWIN ACCESSORY CABINET) (YEL/GRN) J22 J11 GND X

COOLING CABINET 11 VACUUM LINE

COOLANT LINES 8

MG2 A12 A4
(FRNT ENDBELL)

MG3 A1

(MAGNET I/F #1)

611

1152

MG2 A12 A1
(TWIN GRAD COIL)

989

MG2 A36
J1 J2

(BORE TEMP SENSOR)

J8

XWB, X+, Xzm YWB, Y+, Yzm ZWB, Z+, Zzm J5 8

J3 (X) J4 (Y) J5 (Z) 16,17, GRN/YEL 18,19,GRN/YEL,N GRN/YEL

X+/, GND Y Y+/, GND Z+/, GND Z GRADIENT SW J4

979 988 12 980 988 981 988

X Y Z GND

1233
FLEXIBLE EXTENSION CABLES

PP1 A7 Xzm+/, GND Yzm+/, GND Zzm+/, GND Xwb+/, GND Ywb+/, GND Zwb+/, GND J151

MG2 A12 A3 983 982 987 986 985 1198 1199 1202 1197
J1 J2 J3 J4

MG2 A31
(DOCK LIMIT SWITCH)

984

10/12 6/8 (VACUUM 1/2 REAR 3/4 ENDBELL) 5/7 9/11

715 1215 387 MG2 A29


P1 (DOCK ASM)

823 1179 1234


J17 J56 TS1 RF AMP J24 J13 J3 J4 J202 PP1 RF TX SW

1228 1205 1153 1206


J25 J36 J35 J15 J24 J19

MG2 A16
(LOW PROFILE J13 CARRIAGE ASM) J1 (MNS XMT) J20 J1 A8 (QUICK THRU (MC DISCONNECT) J8 ASM) J9 J6 (MULTICOIL) 3

J153 J150 J152

1 1 A B C

10,11,12, GRN/YEL, N

1180

1045 1249 1259 1187 1185 1186 1189 1188 1190 1191 1192 1194 1193 1195 235
J78 J77

MG3 A17
(CABLE I/F)

1214 1207 1208 1228 1209 1209 1210 1212 1211 1213 1206

1238 235

J4

f IN
1255
2 F/O J26 J22 J63

MR2 1083 1082 A11

J202 J200 J201 PP1 A15 (MC SW FLTR)

MG3 A8 J1 (UTNS ANTENNA)

1204

GRN/YEL

(RFS CABINET) (ORANGE) (18)J200 (916)J201 J5 J3 J10

MR3 PD1
1 1 1 (POWER DISTRIBUTION UNIT) (RED) 4,5,6,GRN/YEL,N GRN/YEL SPLICE J2

1200 1201 1217 1216 1203 778 779 782 257 781 780

MG3 A36 J20 (16 CHANNEL SWITCH) J23 J1


J2 J10 J5 J11 J22 J17 J21 J18

A6 A1

PP1 A16 (MC BIAS FLTR) PP1 J203 MC SW FLTR J92 J93 J94 J95 J96 J72 J73 J75 J76 J74

J4 (HEAD J2 (MNS RCVE) J21 B (QUICK J22 CONN DISCONNECT) J11 A (QUICK J12 CONN DISCONNECT) MG2 A14 A1 (MIC) J11 RCV FLTR OUTPUT INPUT

A3/A5

(QUICK DISCONNECT)

A4

(ADAPTOR) (HEAD COIL) (MULTICOIL) (MULTICOIL)

968 037 971


RF DOOR SWITCH 1 2

J1 AC IN (A12) J23 DRIVER J21 MODULE J22 J23 J24 J25

MG3 A3 A5
(BODY PREAMP COMBINER)

PP1 A11 (FILTER)

EO2
(EMERGENCY OFF SWITCH) 1

FRONT J3 FRONT J4 MG2 A12 A2 (RF BODY COIL) REAR J5 J1 REAR J6 J2

893
9

J7 J4 J3 J6 J1

744 743 780 257 1250

J44 J35 J32 PP1 PHPS J89 J14 J47 J20 J15 J11 F5/F6 GND J12

701

(BODY HYBRID SPLITTER)

MG3 A3

J15 J125

715 385 318 784 045 282 297 EO1 (EMERGENCY OFF SWITCH) 1
(BLOWER BOX) AIR HOSE GND J1

1184 1233 1181


8 T

387 300 MG6


(LONGITUDINAL MOTOR DRIVE)

J155(R) J154(P)

FLEXIBLE EXTENSION CABLE

385

(PATIENT ALIGNMENT LIGHTS) JP1

MG2 A30

MG3 A7

(+,)

7, N, GRN/YEL

1182 1085

(BORE LIGHT)

MG2 A15

FIBEROPTIC BUNDLE AIR HOSE

(PATIENT COMFORT MODULE)

MG2 A14

296 REV 1

296

J18

J7 (PATIENT J8 COMMUNICATION BOX) J9 J6

MG3 A2

MG2 A14 A2 414 MG2 A35 (FRONT


(SPEAKER) ENDBELL MICROPHONE)

6-3

DIRECTION 5133312

THIS PAGE IS LEFT INTENTIONALLY BLANK

REV 1

6-4

DIRECTION 5133312

FACILITY EMERGENCY POWER 1

1 5 1 1

A12 (SERVICE LAPTOP) OW1 A15 (HOST COMPUTER) USB

1056
10 A18 (SERVICE KEY) OW1 A13 (MODEM) DB9

PORT 6 PORT 8 OW1 A17 (LAN HUB) PORT 2

13 TO SITE ETHERNET

FACILITY POWER

OM1
(OXYGEN MONITOR)

FILTERS FOR FACILITY POWER, EMERGENCY POWER, AND AUXILIARY WARNING DEVICE EXHAUST FAN 1

MS4
1 1 (ERU)

EXAM ROOM LIGHTS

606
(MAGNET)

Signa EXCITE HD 1.5T TwinSpeed, Release 12.x SYSTEM INTERCONNECT (FIXED SITE) (DUPLICATE OF PAGE 63)
NOTES:
1 2 CUSTOMER FURNISHED Magnet and superconducting shim power supplies are not included with system, but may remain in place if purchased as an option. Adaptors selected according to coil used (head, surface) and magnet polarity. Typical connection for normally closed emergency off switch Optional: May not be present on all systems. Magnet ramping cables pass through ramp port on PP1 plate. Customer supplied telephone line to be connected from UPS to Modem (MSM3) or optional MUX . Additional telephone line would be needed if optional MUX is installed. Optional: Used only if High Order Shim is installed. PP1 PHPS J155 terminator relocated to High Order Shim in Twin Accessory Cabinet (TAC). Body Hybrid To Body Coil Connections depend on magnet field polarity. Customer supplied telephone line for Insite Modem. Refer to next page for cooling system thats designed for your site. Source of water supply for Cold Head Compressor depends on system used. Refer to Tab 1, Direction 2308287, Gradient Cable Installation, for instructions on tywrapping ground cables to each gradient cable between TAC and PP1. Ethernet cable is customer supplied. RUN 1066 and modem may not be required if connection to InSite is provided thru site ethermet connection at OW1 A7 PORT 8. J5 J1

MG3
(REAR PED)

MS1

1 OW1 A6 (HOST MONITOR) R,B.G IN (COLOR) 1f FACILITY POWER 1 3f OW1 A3 (LCD DISPLAY)

1254
TO PDU (S/C SHIM P.S.)

1063 1065 1068 1069 1082 1064 1070


J2 J8

MONITOR 1

RS232 USB SCSI 2 Power

1066 1059 1067

13

MS7

603 604 601

1 5

RF SHIELD ROOM COMMON GROUND 5 J17 6

040
5 3

MONITOR 2 ON PCI BIT3 SCSI 1

457
RED (2) BLK (2)

458 603 604 F2 F4 601 602 624 621 622 MS1 A5 A2


(PRESSURE TRANSDUCER)

(REMOTE O2 SENSOR)

OM3

4 5 6 7

OW1 A5 (DASM)

T PORT 7

(MAGNET P.S.)

MS6

602

(TERMINAL BOX) P403

MS1 A3 A1 MS1

T 1 OW1 A20 (SCIM)

1083
OW1 A7 (KEYBRD) OW1 A8 (MOUSE)

J1 OW1 A11 J12 (POWER J3 J7 J11 STRIP) J7

1071
FAN

WATER

OW1 A16 (SCSI TOWER)

F1
11 (COLD HEAD CMPRSSR) A1JR

MS5A1

623 621 622 MS5A5


(WATER FLOW)

F3

(MAGNET SERV TURRET) (SHIELD COOLER COLD HEAD) SLEEVE COLDHEAD PREAMP J2 J1

MS1 A2

8 9 10

1075 1074 1077


J5 J2 J4 OW1 A19 (CABINET PDU)

1256
J6 J15 J18 OW1 A21 (GOCAA) J8 J7 J18 J10 J16 J17

833
FJ3 FJ4

FEED THRU (J60 & J61) (PENETRATION PANEL) (YELLOW) J48 J10 PP1 A17 (REPEATER PANEL)

OW1 A10 (OW TABLE) 1 (SPKR) (GREEN)

OW1 A1

OW1 A2
(GOC COMPUTER CABINET) (BROWN) J5 J10

PP1

826
J9 (MAGNET MONITOR) J1

827 MSM1
J8 J7

A1 J1 FJ1

832

11 12 13

1085
10 7

1061 1060

J3

J1

825 824
5 F/O

829 828
FJ2

830

MG3 TB1
(REED SWITCH)

831

TO UPS CRYO COOLER COMPRESSOR TO MAG MON MODEM

MSM4 939
(UPS)

5 STEREO MUSIC RACK IN BORE

1081

823
5

940
(MODEM) 7 MUX 7 PATIENT ALERT CONTROL BOX

3 F/O

848 711 712 716


(J9,10) J4 (J11,12,13) (PAC I/F) BULB P. GATING RESP ECG PPG

862

J1 (RGHT OPER CONTROL)

MSM3

1045 1196
J1

PP1 A17 A1 J91 F/O REPEATER BOARD

711 712
2 F/O

MG2 A4
(OPER. DISPLAY)

MG2 A5

MG2 A33
(SRI3)

(PAC 3) J2 POWER

MG2 A11

ALERT AIR LINE FEED THRU COOLANT FEED THRU COOLANT FEED THRU SHIM FILTER VACUUM FEED THRU
(TWIN GRAD FILTER)

J4 J5 J2 J10 J20 J1 J3

897 1261 422 1260 421

J3 J4 J4

J2

863

(LEFT OPER. CONTROL) J2

MG2 A5

(MAIN DISCONNECT PANEL)

MDP

J5 (HFD CABINET) (ORN/YEL)

974 1179 1180 1181 976 977 978 989 1081 1254
4 F/O 8 T

J12

MR3

J2 J7 J8

TAC
(TWIN ACCESSORY CABINET) (YEL/GRN) J22 J11 GND X

COOLING CABINET 11 VACUUM LINE

COOLANT LINES 8

MG2 A12 A4
(FRNT ENDBELL)

MG3 A1

(MAGNET I/F #1)

611

1152

MG2 A12 A1
(TWIN GRAD COIL)

989

MG2 A36
J1 J2

(BORE TEMP SENSOR)

J8

MG2 A12 A3 983 982 987 986 985 1198 1199 1202 1197
J1 J2 J3 J4

XWB, X+, Xzm YWB, Y+, Yzm ZWB, Z+, Zzm J5 8

J3 (X) J4 (Y) J5 (Z) 16,17, GRN/YEL 18,19,GRN/YEL,N GRN/YEL

X+/, GND Y Y+/, GND Z+/, GND Z GRADIENT SW J4

979 988 12 980 988 981 988

X Y Z GND

1233
FLEXIBLE EXTENSION CABLES

PP1 A7 Xzm+/, GND Yzm+/, GND Zzm+/, GND Xwb+/, GND Ywb+/, GND Zwb+/, GND J151

MG2 A31
(DOCK LIMIT SWITCH)

984

10/12 6/8 (VACUUM 1/2 REAR 3/4 ENDBELL) 5/7 9/11

715 1215 387 MG2 A29


P1 (DOCK ASM)

823 1179 1234


J17 J56 TS1 RF AMP J24 J13 J3 J4 J202 PP1 RF TX SW

1228 1205 1153 1206


J25 J36 J35 J15 J24 J19

MG2 A16
(LOW PROFILE J13 CARRIAGE ASM) J1 (MNS XMT) J20 J1 A8 (QUICK THRU (MC DISCONNECT) J8 ASM) J9 J6 (MULTICOIL) 3

J153 J150 J152

1 1 A B C

10,11,12, GRN/YEL, N

1180

1045 1249 1259 1187 1185 1186 1189 1188 1190 1191 1192 1194 1193 1195 235
J78 J77

MG3 A17
(CABLE I/F)

1214 1207 1208 1228 1209 1209 1210 1212 1211 1213 1206

1238 235

J4

f IN
1255
2 F/O J26 J22 J63

MR2 1083 1082 A11

J202 J200 J201 PP1 A15 (MC SW FLTR)

MG3 A8 J1 (UTNS ANTENNA)

1204

GRN/YEL

(RFS CABINET) (ORANGE) (18)J200 (916)J201 J5 J3 J10

MR3 PD1
1 1 1 (POWER DISTRIBUTION UNIT) (RED) 4,5,6,GRN/YEL,N GRN/YEL SPLICE J2

1200 1201 1217 1216 1203 778 779 782 257 781 780

MG3 A36 J20 (16 CHANNEL SWITCH) J23 J1


J2 J10 J5 J11 J22 J17 J21 J18

A6 A1

PP1 A16 (MC BIAS FLTR) PP1 J203 MC SW FLTR J92 J93 J94 J95 J96 J72 J73 J75 J76 J74

J4 (HEAD J2 (MNS RCVE) J21 B (QUICK J22 CONN DISCONNECT) J11 A (QUICK J12 CONN DISCONNECT) MG2 A14 A1 (MIC) J11 RCV FLTR OUTPUT INPUT

A3/A5

(QUICK DISCONNECT)

A4

(ADAPTOR) (HEAD COIL) (MULTICOIL) (MULTICOIL)

968 037 971


RF DOOR SWITCH 1 2

J1 AC IN (A12) J23 DRIVER J21 MODULE J22 J23 J24 J25

MG3 A3 A5
(BODY PREAMP COMBINER)

PP1 A11 (FILTER)

EO2
(EMERGENCY OFF SWITCH) 1

FRONT J3 FRONT J4 MG2 A12 A2 (RF BODY COIL) REAR J5 J1 REAR J6 J2

893
9

J7 J4 J3 J6 J1

744 743 780 257 1250

J44 J35 J32 PP1 PHPS J89 J14 J47 J20 J15 J11 F5/F6 GND J12

701

(BODY HYBRID SPLITTER)

MG3 A3

J15 J125

715 385 318 784 045 282 297 EO1 (EMERGENCY OFF SWITCH) 1
(BLOWER BOX) AIR HOSE GND J1

1184 1233 1181


8 T

387 300 MG6


(LONGITUDINAL MOTOR DRIVE)

J155(R) J154(P)

FLEXIBLE EXTENSION CABLE

385

(PATIENT ALIGNMENT LIGHTS) JP1

MG2 A30

MG3 A7

(+,)

7, N, GRN/YEL

1182 1085

(BORE LIGHT)

MG2 A15

FIBEROPTIC BUNDLE AIR HOSE

(PATIENT COMFORT MODULE)

MG2 A14

296 REV 1

296

J18

J7 (PATIENT J8 COMMUNICATION BOX) J9 J6

MG3 A2

MG2 A14 A2 414 MG2 A35 (FRONT


(SPEAKER) ENDBELL MICROPHONE)

6-5

DIRECTION 5133312

THIS PAGE IS LEFT INTENTIONALLY BLANK

REV 1

6-6

DIRECTION 5133312

NOTES:
1 1 2 2

CUSTOMER FURNISHED. THIS GROUP CONTAINS WATER LINES WHICH SHALL BE ROUTED SEPARATE FROM ELECTRICAL LINES (I.E. POWER & SIGNAL). NOTE: D ONLY INTERCONNECTS SPECIFIC TO MRCC SUBSYSTEM EQUIPMENT SHOWN HERE.

EXTERNAL TO BUILDING
1 2

INTERNAL TO BUILDING
1 2

NOTES:
1 2

CUSTOMER FURNISHED. THIS GROUP CONTAINS WATER LINES WHICH SHALL BE ROUTED SEPARATE FROM ELECTRICAL LINES (I.E. POWER & SIGNAL). NOTE: D ONLY INTERCONNECTS SPECIFIC TO MRCC SUBSYSTEM EQUIPMENT SHOWN HERE.

INDOOR MRCC

1 1 1 1 1

TO GRADIENT COIL RCP

OUTDOOR MRCC

1 1 1 1 1

TO GRADIENT COIL RCP

MAIN DISCONNECT PANEL RCP OUTDOOR MRCC MS5

MAIN DISCONNECT PANEL RCP


1

INDOOR MRCC

MS5
1 2 1 2

DUAL INDOOR MRCC SUBSYSTEM GROUP INTERCONNECT DIAGRAM NOTES:


1 2 1 2

DUAL OUTDOOR MRCC SUBSYSTEM GROUP INTERCONNECT DIAGRAM NOTES: EXTERNAL TO BUILDING
1 2

CUSTOMER FURNISHED. THIS GROUP CONTAINS WATER LINES WHICH SHALL BE ROUTED SEPARATE FROM ELECTRICAL LINES (I.E. POWER & SIGNAL). NOTE: D ONLY INTERCONNECTS SPECIFIC TO MRCC SUBSYSTEM EQUIPMENT SHOWN HERE.

INTERNAL TO BUILDING
1 2

1 2

CUSTOMER FURNISHED. THIS GROUP CONTAINS WATER LINES WHICH SHALL BE ROUTED SEPARATE FROM ELECTRICAL LINES (I.E. POWER & SIGNAL). NOTE: D ONLY INTERCONNECTS SPECIFIC TO MRCC SUBSYSTEM EQUIPMENT SHOWN HERE.

INDOOR MRCC

1 1 1

TO GRADIENT COIL RCP

OUTDOOR MRCC

1 1 1

TO GRADIENT COIL RCP

MAIN DISCONNECT PANEL

MAIN DISCONNECT PANEL

FACILITY WATER SUPPLY


2

MS5

FACILITY WATER SUPPLY


2

MS5

INDOOR MRCC SUBSYSTEM GROUP INTERCONNECT DIAGRAM

OUTDOOR MRCC SUBSYSTEM GROUP INTERCONNECT DIAGRAM

MR3 PD1
4 2

INDOOR WATER COOLED GWHX FACILITY WATER SUPPLY


3 1 2

TO GRADIENT COIL NOTES: 1 CUSTOMER FURNISHED.


2

THIS GROUP CONTAINS WATER LINES WHICH SHALL BE ROUTED SEPARATE FROM ELECTRICAL LINES (I.E. POWER & SIGNAL). FOR INDOOR WATER COOLED GWHX ONLY. GWHX PROVIDED CABLE.

3 4

FACILITY WATER SUPPLY


2

MS5

NOTE: D ONLY INTERCONNECTS SPECIFIC TO GWHX SUBSYSTEM EQUIPMENT SHOWN HERE.

INDOOR GWHX SUBSYSTEM GROUP INTERCONNECT DIAGRAM

REV 1

6-7

DIRECTION 5133312

THIS PAGE IS LEFT INTENTIONALLY BLANK

REV 1

6-8

DIRECTION 5133312

Signa EXCITE 1.5T TwinSpeed, Release 12.0: 2KW MNS OPTION SYSTEM INTERCONNECT (FIXED SITE)

MR2 A11
(RFS CABINET) (ORANGE)

(PENETRATION PANEL) (YELLOW) J200

PP1
J200

Driver Module J8

1185

1200

MG3 A36
J1 (16 CHANNEL SWITCH) J23

(LOW PROFILE CARRIAGE ASSEMBLY)

MG2 A16
J2 (RX) J1 (TX)

(MNS T/R ASM)

1209 1153

MR3 PD1
(POWER DISTRIBUTION UNIT) (RED)

J52 J123 J102 J103 J104 J105

NOTCH FILTER
J1

1221 1224
J3 J9

1222 1223 1227 1225 1226


J7 J1 J11 J12 J13 FLEXIBLE EXTENSION CABLE

J3

J3

MG3 A17
(CABLE I/F)

13,14 GRN/YEL

1237

J8

J2

1248

1146

J83

1147

MR3

MR10
(1.5T MNS AMP CABINET) (YEL/BLU)

(HFD CABINET) (ORN/YEL)

REV 1

6-9

DIRECTION 5133312

THIS PAGE IS LEFT INTENTIONALLY BLANK

REV 1

6-10

DIRECTION 5133312

GE MEDICAL SYSTEMS REV 1 612 SORTING AND ROUTING CABLES

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

CAUTION
Tywraps must be cut flush with no protruding sharp edges or points. Failure to do so can result in numerous laceration hazards when servicing the equipment. System power, ground, and data cables are color coded at each end by destination per colors shown on Cable Map. 1. Sort cables by destination. Normally, it is best to sort cables originating from the farthest cabinet or component from the PDU first. This will make it easier to pull cables through the troughs.

Note: Carefully review architectural drawings to insure that all ducts are properly installed and that all signal and power cables are accounted for before starting to route cable runs. 2. 3. Unroll coiled cables along the intended route to insure that they will lay freely without twists or kinks. Route cables in accordance with the applicable System Interconnect Diagram. Plan for storage location of coiled cable excess length if cables are to remain at delivered length. See Illustration 21. Some cables are usually cut to length such as gradient cables, power cables, Head and Body RF, and Fiberoptic cables. 1 Excess cable to be stored in cable trough must be routed in a Figure Eight pattern as shown. DO NOT COIL IN CIRCLES!

2A Dress and tywrap ends of looped cable.

OR
2B Dress and tywrap middle of looped cable. 3 Repeat as necessary for cables in Magnet or Equipment room.

PROPER STORAGE OF EXCESS CABLES ILLUSTRATION 2-1

4.

Place cables in provided troughs, conduit, or ducts. Leave sufficient slack for later connection when cabinets are in final position according to site architectural plans. Route cables inside exam room to Rear Pedestal area with sufficient length remaining to complete routing and connecting within Magnet Enclosure.

6-11

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

62 EQUIPMENT ROOM CABLE INSTALLATION The magnet enclosure and magnet room cables must be installed simultaneously. Cable locations depend on how enclosures are installed. 621 HFD CABINET CABLE CONNECTIONS

DANGER!!
MAKE SURE ALL POWER TO CABINET IS OFF, LOCKED, & TAGGED BEFORE CONNECTING WIRES.

WARNING!
IF 3 PHASE WYE WITH NEUTRAL AND GROUND (5 WIRE SYSTEM) INPUT IS USED THE NEUTRAL MUST BE TERMINATED INSIDE THE MAIN DISCONNECT CONTROL AND NOT BROUGHT TO THE HFD CABINET.
621A 1. 2. 3. 4. 5. HFD Cabinet Power and Ground Cable Installation

Remove rear PDU module covers. Route facility input power cable and ground wire thru interface panel. Connect L1L3 conductors to terminal block. Connect facility input dedicated ground conductor to Ground (GRN/YEL) Bus Bar. Connect RF screen room ground conductor to Ground Bus Bar. 5 1 4 3

6-12

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 621B 1.

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Route Power Cables Thru Interface Panel

Route Runs thru interface panel as marked.

Note: The GRN/YEL ground wire, Run 037, MUST be routed with Run 968 thru the same port. 2. Route any option Runs as marked. Run 989 Run 037 Run 968 Run 1081 Run 976

Run 1179

Run 1182

Run 977 Facility Power Run 978 Run 1254 (RF Shield Room GND)

6-13

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 621C 1. 2. 3.

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Connect Power Cables to PDU Module

Route/connect power cables to pressure connectors and studs on back of PDU module as shown in Table below. Relable PDU connection panel as shown below. Trim and connect ground wire cables (Runs 037 and 1254) to ground bus connectors. Tighten clamps of interface panel to secure cables in place.
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 30
GRADIENT 420V POWER

WARNING!
LT. BLU

RF/SYSTEM CABINET

PHPS/SSM

ALL POWER CABLES LISTED BELOW MUST BE CONNECTED TO PDU MODULE LOCATED WITHIN HFD CABINET.

RF/AMP

MNS/ SPARE 2 TAC/BRM GWC PC (BLK) HOST (BRN) SPARE #1 SPARE 3/TRM GWC UPS IN UPS OUT PATIENT BLOWER

GRN/YEL

3 2

CONNECT TO
4,5,6 7 10,11,12 16,17 18,19 (MDP)* or 30 (PDU) GRN/YEL GRN/YEL

RUN
968 1182 1113 989 1081

PHASE
f A, B, C f A f A, B, C f A,C f C,A f A or f A

DESTINATION
RFS Cabinet PHPS Module NB RF Amp Cabinet TAC Cabinet Operator Workspace AirSys Water Cooled Cabinet or AirSys Heat Exchanger 3f , Neutral, GND Single f Neutral, GND 3f , Neutral, GND Single f , GND Single f , Neutral, GND Single f , Neutral, GND Single f , Neutral, GND

037 1254

System Cabinet Ground Wire RF Shield Room Ground (Customer Supplied)

COLOR CODE: for remainder of cables: Neutral = Light blue 3f = Black, Red, Orange Single f = Brown, Black

Gnd = Grn/Yel

* Connection to Main Disconnect Panel to be completed by customer electrician.

6-14

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

622

EQUIPMENT ROOM CABINET POWER CABLE CONNECTIONS

WARNING!
CABINET POWER AND GROUND CABLES MUST BE CONNECTED TO PDU MODULE LOCATED WITHIN HFD CABINET.
622A 1. 2. RF/System Cabinet Power and Ground Cable Connections

Connect Run 968 cable plug to J1. Be sure to push and turn plug for secure connection. Connect Run 037 GND cable to ground stud.

622B 1.

TAC Cabinet Power Cable Connection

Connect Run 989 cable plug to J8.

J22

J11

J27 J12 J10

J9

1
J8

6-15

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 623

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

GRADIENT CABLE INSTALLATION IN EQUIPMENT ROOM

Nine cables have been delivered to site. All have three lead wires with terminals and attached cable labels at one end. The leads on this cable end are connected to the Twin Gradient Filter on the Penetration Panel. The other end is cut flush with no labels or terminals. The nine cables with supplied labels are as follows: RUN PART # 979 980 981 982 983 984 985 986 987 234540010 234540011 234540012 234540018 234540017 234540016 234540024 234540023 234540022 LABEL PP1A7X PP1A7Y PP1A7Z PP1A7ZZM PP1A7YZM PP1A7XZM PP1A7ZWB PP1A7YWB PP1A7XWB LENGTH SUPPLIED LABELS 59.5 ft. 59.5 ft. 59.5 ft. 26.5 ft. 26.5 ft. 26.5 ft. 26.5 ft. 26.5 ft. 26.5 ft. (1) RUN 976, (1) RUN 979, (2) X+, (1) X, (1) GND, (1) XZM, (1) XWB (1) RUN 977, (1) RUN 980, (2) Y+, (1) Y, (1) GND, (1) YZM (1) YWB (1) RUN 978, (1) RUN 981, (2) Z+, (1) Z, (1) GND, (1) ZZM (1) ZWB (1) RUN 982, (1) ZZM (1) ZZM+ (1) RUN 983, (1) YZM (1) YZM+ (1) RUN 984, (1) XZM (1) XZM+ (1) RUN 985, (1) ZWB (1) ZWB+ (1) RUN 986, (1) YWB (1) YWB+ (1) RUN 987, (1) XWB (1) XWB+

Refer to Section 631 for installation of Gradient cables in Magnet Room. 623A 1. 2. 3. 4. 5. 6. 7. 8. 9. General Steps for Cutting Cables to Length and Routing in Equipment Room

Move Run 979, 980, and 981 cables to equipment room. The end of cable with terminals will be connected to Penetration Panel. Route cables from Gradient Filter of Penetration Panel to right side of interface panel at rear of Twin Accessory Cabinet (TAC). Verify sufficient slack of at least two feet (610mm) is provided for termination and connection at TAC Cabinet. Mark location for cutting. Cut cables to length and label cut ends with applicable run number. Refer to Procedure 623E to terminate and label TAC end cable Runs 979, 980, and 981 Refer to Procedure 623F for connection of Runs 979, 980, and 981 to TAC cabinet. Route remainder of cables from rear of TAC to rear of HFD cabinet. Make sure enough slack is provided for termination and connection to cabinet. Mark location for cutting. Cut cables to length. Refer to Procedure 623H for connection of Runs 976, 977, and 978 to TAC cabinet. Verify sufficient slack of at least two feet (610mm) is provided for termination and connection at HFD Cabinet. Mark location for cutting. Cut cables to length and label cut ends with applicable run number. Refer to Procedure 623K for connection of Runs 976, 977, and 978 to HFD cabinet.

6-16

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 623B 1. 2.

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Connect Runs 979, 980, and 981 to Penetration Panel

Remove gradient filter terminal covers. Route Run 981 (Z) to left side and Runs 979 (X) and 980 (Y) to right side of gradient filter on Penetration Panel.

Note: The three X and three Y terminals are routed to the right side and three Z terminals to the left side. 3. Connect +, ZM (cable marked as TWO), & WB (cable marked as THREE) wires to gradient filter terminals as marked on each side of gadient filter. Note: Red tape indicates positive (+) wires. Align cables to obtain maximum spacing from other cables and conductive metal objects. Note: When tightening nuts, it is important to hold screw body. Tighten nuts to a torque of 35 ftlbs (6774Nm) or after tightening, verify that nothing is loose. Note: Final check is by hand. Make sure screw is fixed, nut is tight, and cable is not detached from terminal lug. Route and secure each cable Run to strainrelief bracket with Ubolts. EQUIPMENT ROOM VIEW 1 2 3 4 5 EQUIPMENT ROOM RIGHT SIDE
Y+

4. 5. 6.

EQUIPMENT ROOM LEFT SIDE ZWB 3 4 5

YZM

YWB XWB

XZM

Z+

X+
GND

ZZM

6
RUN

980

RUN

RUN

979

981

6-17

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 623C

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Attach Ground Cables to Runs 979, 980, and 981

For each gradient cable routed between the Penetration Panel and TAC Cabinet, a ground wire, Run 988 (2283297), has to be tywrapped to each cable. 1. 2. One Ground Wire will be routed along each of the Gradient Cables, Runs 979, 980, and 981. Each Run 988 must be tywrapped about every 12 inches along entire length of each gradient cable. Note: Run 988 must be routed parallel to the gradient cable. It MUST NOT be routed spirally around the gradient cable. End of Run 988 at TAC cabinet should extend about 12 inches (305mm) past end of outer jacket after trimming. Note: Refer to Procedure 623E for length of gradient cable jacket material to be removed.

3.

Excess ground cable can not be coiled. It must be cut to length and reterminated.
3
12 in (305mm) 12 in (305mm)

2 GND WIRE 1 PEN PANEL CABLE ENDS

TAC CABINET CABLE ENDS

RUN 988 GND WIRE

6-18

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 623D 1. 2.

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Connect Run 988 Cables Ground Wires to Gradient Filter

Connect the three Run 988 Gound Wires to GND on left side of Gradient Filter. Each Run 988 must be tywrapped about every 12 inches along entire length of each gradient cable. Note: Run 988 must be routed parallel to the gradient cable. It MUST NOT be routed spirally around the gradient cable. Reinstall gradient filter terminal covers. Cut each Run 988 to length, reterminate with 3/8 ring terminal for connection to GND stud on TAC Cabinet. EQUIPMENT ROOM LEFT SIDE ZWB

3. 4.

EQUIPMENT ROOM RIGHT SIDE


Y+

YZM

YWB XWB

XZM

Z+

X+
GND

ZZM

1 2

Note: For clarity, only Run 981 is shown.

RUN

980

RUN

RUN

979

981
12 in (305mm) 12 in (305mm)

RUN 988

TAC CABINET CABLE ENDS

6-19

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 623E

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Terminate Runs 979, 980, and 981 for TAC Cabinet Connection

The following steps are required to terminate the end of the cable that was cut in Procedure 623A. The other end of the cable is connected to the Penetration Panel. 1. 2. 3. 4. Remove 18 inches of jacket from end of cable Runs 979 (X), 980 (Y), and 981 (Z). Trim rubber cable filler strands flush with end of jacket. Strip 0.75 in. (19mm) of insulation from end of wires ONE, TWO and THREE and crimp a furnished T & B 1/2 (G2312) terminal to each using Ratcheting Crimper with #2 die. Place furnished X+ label on wire ONE. Place furnished XZM label on wire TWO. Place furnished XWB label on wire THREE. Wrap red tape around wire ONE. Place furnished RUN 979 label on cable. Steps above show completion of Run 979. Repeat same process for Runs 980 (Y+, YZM, YWB) and 981 (Z+, ZZM, ZWB). 1
1.5 ft. (457mm)
ONE TWO THREE

5. 6. 7.

2 GRADIENT CABLE (RUN 9__) TAC END

0.75 IN (19mm)
ONE ONE TWO THREE

GRADIENT CABLE (RUN 9__) TAC END 4

ONE ONE TWO

X+ XZM

5
RUN 979 TACCABX 210094819 GE COMPANY

THREE XWB

6 7

6-20

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 623F

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Connect Runs 979, 980, and 981 to TAC Cabinet

To make installation easier, route and connect Run 981 (Z terminals) first, then Run 980 (Y), and last Run 979 (X). 1. 2. 3. 4. Connect X+, Y+, and Z+ cables as marked. Cables should have red tape applied towards end of cable. Connect XZM, YZM, and ZZM cables as marked. Connect XWB, YWB, and ZWB cables as marked. Route and connect the three Run 988 Ground wires to GND stud on interface panel.

X+ XWB Y+ YWB Z+ ZWB ZZM YZM XZM

3 4

To Pen Panel 623G Preterminated cable ends for Runs 976, 977, and 978

Cables delivered to site will be preterminated for the TAC Cabinet connection of Runs 976, 977, and 978. Refer to
Procedure 623A for cutting cables to length for connection to HFD Cabinet. A typical preterminated end for each cable is shown below.
ONE ONE TWO X+ X RUN 976 TACCABX (FROM MR3) 210094834 GE COMPANY

THREE GND

6-21

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 623H

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Connect Runs 976, 977, and 978 to TAC Cabinet

To make installation easier, route and connect Run 978 (Z terminals) first, then Run 977 (Y), and last Run 976 (X). 1. 2. 3. Connect X+, Y+, and Z+ cables as marked. Cables should have red tape applied towards end of cable. Connect X, Y, and Z cables as marked. Connect Ground leads from each cable to M8 GND stud located at bottom of cabinet. Note: Connect cables to all studs with terminals facing outward as shown. Do not install with terminals facing toward the module or facing one another. 1 2

X+ Y+ Z+

X Y Z

M6 M8 To ACGD/PDU Cabinet GND Studs 3

6-22

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 623I

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Terminate Runs 976, 977, and 978 for Connection to HFD Cabinet

The following steps are required to terminate the end of the cable that is to be connected to the HFD Cabinet. The other end of the cable is connected to the TAC Cabinet. Run 976 Jacket Length to be Removed 1. 2. 3. 4. 5. 8.0 IN (203mm) Run 977 11.0 IN (280mm) Run 978 14.0 IN (356mm)

Refer to above table to determine length of jacket and foil to be removed from end of cable Runs 976 (X), 977 (Y), and 978 (Z). Trim rubber cable filler strands flush with end of jacket. Strip 0.75 in. (19mm) of insulation from end of wires ONE and TWO. Cut wire THREE to 6.0 inches (152mm). Place furnished X+ label on wire ONE. Place furnished X label on wire TWO. Place furnished GND label on wire THREE. Wrap red tape around wire ONE. Crimp a furnished T&B #G238 terminal on GND wire end using a Ratcheting Crimper with #2 die. Place furnished RUN 976 label on cable.

6. 7. 8.

Steps in illustration below show completion of Run 976. Repeat same process for Runs 977 (Y+, Y) and 978 (Z+, Z). Red tape, Y+ (RUN 977) and Z+ (RUN 978) labels are attached to wire ONE. 1
See Table
ONE TWO THREE

2 OUTPUT CABLE (RUN 97_) ACGD END

0.75 IN (19mm)
ONE ONE TWO

3 OUTPUT CABLE (RUN 97_) ACGD END


THREE

0.68 IN (17mm) 6.0 IN (152mm)

6
ONE ONE TWO X+ X GNDGND

RUN 976 MR3 CAB I/F J3 210094819 GE COMPANY

6-23

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 623J 1. 2.

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Gradient Output Cable Interface Panel Access at HFD Cabinet

Remove Rear Cover Plate by removing seven screws that hold it in place. Loosen screws on clamps for gradient output cables. Insert ends of gradient output cables into Cabinet Interface as marked. (Runs 976 into J3, Run 977 into J4, and Run 978 into J5)

2 ACGD CABINET (REAR VIEW) 623K 1. 2. Gradient Output Cable Routing and Connection in HFD Cabinet

Connect gradient output cables with redtaped (+) #1 wires to + output terminals and black () #2 wires to output terminals. Tighten six screws to 5265 inchlbs (5,887,34 NM). This is about what can be applied using only your thumb and little finger on wrench. Loosen 1/2 turn and re-tighten to the same torque. Repeat approximately 6 times or until wrench returns to the same spot after torquing. Connect the GND leads of each run to the Ground Bus Bar studs. Tighten clamps on Interface Panel to secure gradient output cables. GRADIENT CABLE CONNECTIONS

3. 4.

CAUTION
Over tightening of screws in Step 2 can damage terminal block. 2 1 2 3 4 5 6 X+ X Y+ Y Z+ Z
TWO TWO TWO ONE ONE ONE ONE

RED TAPE BLK

RUN 976

RUN 977

RUN 978

6-24

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 624

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

INSTALLATION OF MAGNET MONITORING CABLES

Refer to Direction 2230681, Magnet Monitor Hardware Installation Manual, for instructions on connecting cables in the System Cabinet. This manual should be found in the box containing the parts for Magnet Monitoring.

Connect Run 823 to J24

625

INSTALL RUN 701 RF DOOR SWITCH

Run 701 is a 100 ft cable with a 9pin subminature D plug on one end. The other end is cut off with the black and red leads from pair #1 dressed for connecting to a switch. Should it be desired to cut off excess length, make sure that the black and red leads have continuity to pins 1 and 6 on the 9 pin subminature D plug. 1. 2. 3. 4. 5. Connect Run #701 to J15 on System Cabinet and route to RF Door Switch. Connect black lead on Run #701 to RF Door Switch COM (common) contact. If RF door switch is normally open, proceed to step 4. If normally closed, proceed to step 5. Connect red lead on Run #701 to RF Door Switch N.O. (normally open) contact. If RF door switch is not normally open, connect red lead on Run #701 to RF Door Switch N.C. (normally closed) contact.

6-25

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 626

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

INSTALLATION OF FIBER OPTIC CABLES

Plan for storage of excess Run 1034 conduit of as it cannot be reterminated. If storage of Run 709 and 1033 excess conduit is a problem in the equipment room, excess may be cutoff and ends terminated with the Fiber Optic Connector Repair Kit. Instructions for terminating fiber optic cables are in Appendix. 626A 1. 2. Routing of Fiber Optic Cables

Unroll coiled fiber optic conduits for Runs 711/712, 1045,1083, and 1179 along intended routes. Connection of Runs is shown in diagram below.
ST SIMPLEX CONNECTOR AND GLASS CABLE HP SIMPLEX CONNECTOR AND PLASTIC CABLE HP DUPLEX CONNECTOR AND PLASTIC CABLE REPEATER PANEL BOARD (PP1 A17 A1) J7 (BL) J2 (BL) J8 (GY) J1 (GY) MAGNET ENCL (MG2) PAC (MG2A11) J10 J9 J11

RF/SYSTEM CABINET (MR2) HFD CABINET (MR3) MR3 GP X J2 Y J3 Z J4 CLK J1 MGD CHASSIS STIF BOARD (SLOT 13R) J14 J13 J12 J11 J18 J19 J15 J16 J17 GOC COMPUTER CAB (OW1 A2) LINUX PC X1 (OW1 A15) R1 APS BOARD (SLOT 1FPCI)

7121 7111 7112

1179

1045

J9 (BL) J3 (GY) J10 (GY) J4 (BL) J11 (BL) J5 (GY)

J12 J13

1083

R1 X1

SRI (MG2A33)

626B 1. 2. 3. 4.

Connect Run 1045 to Penetration Panel

Move Run 1045 to the Equipment Room side and attach PP1A17 end to access bracket on Penetration Panel. Refer to Direction 2123149 (packed with bag of FO Conduit connectors) for connector installation instructions. Adjust route of individual fiber optic runs to maintain large radius and connect Run 1045 as shown. Adjust position of protective spiral wrap and secure with tiewrap. Route other end of Run 1045 to System Cabinet. 3

2 2338146 (J11) 23381463 (J9, J10) 23381462 (J7, J8)

1 Run 1045

6-26

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 626C

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Route Fiber Optic Runs 1045, 1083, and 1179 in RF/System Cabinet

CAUTION
Handle fiber optic cables carefully. Do not bend fiber optic cables to radius smaller than two inches (50mm). Avoid scratching connector ends. Keep connectors protected until ready to connect.

SYSTEM CABINET (REAR VIEW)

Route spiral wrapped fiber optic cables behind rails along side of cabinet to MGD Chassis and tiewrap to veritcal rail as shown.

1 2

Attach Run 1179 to J17*. (use 1 inch connector) Attach Run 1045 to J13*. (use 1 inch connector)

Attach Run 1083 to J22*. (use 3/4 inch connector)

* Instructions provided with connectors.

6-27

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 626D

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Route/Connect Fiber Optic Run 1083 to APS Board

CAUTION
Handle fiber optic cables carefully. Do not bend fiber optic cables to radius smaller than two inches (50mm). Avoid scratching connector ends. Keep connectors protected until ready to connect. 1. 2. 3. Route Run 1083 fiber optic cables from cabinet I/F J22 to front of APS module. Secure cable to side of cabinet. Do not bend fiber optic cables to radius smaller than two inches (50mm). Connect the Orange and Blue fiber optic cables to the TIMGD Daughter Board as indicated. Do not connect the Green and Brown fiber optic cables to the APS Module (these are spares). Lay them on top of MGD Chassis. 1 3

SYSTEM CABINET (FRONT VIEW)

6-28

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 626E

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Route/Connect Fiber Optic Runs1079 and 1045 to STIF Board

CAUTION
Handle fiber optic cables carefully. Do not bend fiber optic cables to radius smaller than two inches (50mm). Avoid scratching connector ends. Keep connectors protected until ready to connect. 1. 2. Route/Connect Run 1045 fiber optic cables to J15, J16, J17, J18 and J19 on the STIF Board. Route/Connect Run 1034 fiber optic cables to J11, J12, J13, and J14 on the STIF Board.

J19 J18 J17 J16 J15 J14 J13 J12 J11

1 Note: Use tie-wraps to secure fiber optic cables to MGD Chassis cables. 2

SYSTEM CABINET (REAR VIEW)

6-29

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 626F

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Route/Connect Fiber Optic Runs 709 and 1034 to GP3 Module in HFD

CAUTION
Handle fiber optic cables carefully. Do not bend fiber optic cables to radius smaller than two inches (50mm). Avoid scratching connector ends. Keep connectors protected until ready to connect. Routing of fiber optic cables must be done with care to prevent damage to optical fibers. 1. 2. 3. Attach Run 1179 to I/F J2. Route fiber optic cables to GP3 Module. Leave enough slack for extending GP module forward. Connect fiber optics to GP3 module according to table.
GP INTERFACE PANEL CLK J1 X J2 Y J3 Z J4 MDS IN OUT J5 J6

RUN -4

RUN -1

RUN -2

RUN -3

6-30

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 627

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

INSTALLATION OF REMAINING EQUIPMENT ROOM CABLES

Connect remaining Runs within equipment room. Dress and secure all cables after connections are made. Reference interconnect map on page 63 and Cable Table in Appendix for to/from designator information and Run number reference. 627A 1. 2. RF/System Cabinet Connections

Connect routed cables to Interface Panels at bottom and middle of cabinet and to the Driver Module as marked. Refer to cable map. Dress and secure cables. Adjust as necessary.

Driver Module

Middle Interface Panel

Bottom Interface Panel

627B 1. 2. 3.

HFD Cabinet PDU Module Conenctions

Attach rear PDU module covers removed earlier. Connect Run 1179 to J2. Connect Runs 974, 1180, and 1181 to J5, J7, and J8.

2 3

6-31

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 627C 1. 2. 3. TAC Cabinet Connections

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Connect Run 974 to J12 on Interface panel. Connect Run1233 to J5 on Shim Supply Module. Install Can Terminator on J4 of Shim Supply Module.

1 2

6-32

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

63 MAGNET ROOM CABLE INSTALLATION The magnet enclosure and magnet room cables must be installed simultaneously. Cable locations depend on how enclosures are installed.

WARNING!
FERROUS MATERIAL HAZARD! PARTS REQUIRED FOR THIS UPGRADE INSTALLATION CONTAIN FERROUS MATERIAL AND WILL BE STRONGLY ATTRACTED TO MAGNET AND MAY BECOME DANGEROUS PROJECTILES. IF MAGNET IS AT FULL FIELD KEEP ALL FERROUS TOOLS AND MATERIALS AT LEAST 10 FEET AWAY FROM THE MAGNET.

If not already done at time of Magnet delivery and installation, carefully unroll all cables and hoses from bore at rear of magnet Straighten and dress cables.

6-33

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 631

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

CONNECTION OF MAGNET GROUND, RUN 040

The PAC/SRI Platform is located on right side of magnet. The Magnet ground cable , RUN 040, has to be connected to the opposite side rear left leg as shown above. The other end of Run 040 is connected to the RF Screen Room common ground.

Magnet Front View

Magnet Rear View

PAC/SRI Platform

Ground Stud

632 1. 2. 3. 4. 5. 6.

INSTALLATION OF BORE LAMP ASSEMBLY

Move Bore light Assembly (2222547) to area of left rear magnet leg. Connect the (+) and () wires of Run 300 to the terminal strip as shown. Route Run 300 thru opening in rear leg, under Magnet and Rear Pedestal and connect to J15 on Penetration Panel. Route fiber bundle from Rear End Bell thru leg opening and insert into bushing. Fasten to bushing with set screws. Insert into clip on Bore light Assembly. Bushing flanges to be inside clip. Peel off protective paper on velcro and install assembly in opening in rear leg. 4 5

(+) ()
1 6 2 3

6-34

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 633

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

MAGNET ROOM FIBER OPTIC PAC/SRI INSTALLATION

Fiber Optic Run 711/712 is delivered to site connected to the SRI and PAC modules on the platform assembly mounted to the right side of the magnet between the legs. The conduit is stored for shipment in the bore of the magnet. Unroll conduit along designated path in cable trough to Penetration Panel. Shown below are the connections for the fiber optics cables routed in the magnet room from the Penetration Panel, thru the cable trough to the Rear Pedestal, and under the magnet to the PAC/SRI platform.
MAGNET ENCLOSURE (MG2) SRI (MG2A33) J11

REPEATER PANEL BOARD (PP1 A17 A1) J9 (BL) J10 (GY) J11 (BL) J7 (BL) J8 (GY) J3 (GY) J4 (BL) J5 (GY) J2 (BL) J1 (GY)

7111 7112 7121

J12 J13 J10 J9 PAC (MG2A11)

ST SIMPLEX CONNECTOR AND GLASS CABLE HP SIMPLEX CONNECTOR AND PLASTIC CABLE HP DUPLEX CONNECTOR AND PLASTIC CABLE

633A

Repeater Panel Feedthru Bracket Removal

Unfasten two #1032 screws and two #1032 nuts securing feedthru access bracket to J91 feedthru block.

2 Remove feedthru access bracket and fastening hardware from panel.

6-35

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 633B

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Routing Run 711/712 in Magnet Room

Note: If Run 711/712 excess conduit is a problem in the magnet room, excess may be cut and ends terminated with the Fiber Optic Connector Repair Kit. Refer to Appendix for instructions. If cables need to be shortened, mark each cable prior to cutting. This will aid in identifying the cables when reterminating. 1 Move Run 711/712 (46328079G9) to the Magnet Room side of the Penetration Panel. The PP1J91 end will be attached to interface bracket. 2 Attach Run 711/712 to access bracket. Refer to Direction 2123149 (packed with bag of FO Conduit connectors) for connection installation.

3 Route ends of individual fiberoptic Runs through J91 opening to the Repeater Board (PP1 A17 A1)

Route Run 711/712 to Magnet Enclosure Rear Pedestal and underneath magnet.

633C

Connecting Fiber Optic Run 711/712 to Repeater Panel

Note: Take care to maintain the minimum 2 inch bend radius.

1 2

Adjust route of individual fiber optic runs to maintain large radius. Connect Runs 7121, 7111, and 7112 as shown. Run 7112 (J5) Run 7111 (J3, J4) Run 7121 (J1, J2)

3 4

Gather fiber optic runs and route through wave guide notch in feedthru block.

Replace feedthru access bracket and fasten with two #1032 screws and two #1032 nuts removed in step A1.

6-36

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 634

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

GRADIENT CABLE INSTALLATION IN MAGNET ROOM

Nine cables have been delivered to site. All have three lead wires with terminals and attached cable labels at one end. The leads on this cable end are connected to the Twin Gradient Filter on the Penetration Panel. The other end is cut flush with no labels or terminals. The nine cables with supplied labels are as follows: RUN PART # 979 980 981 982 983 984 985 986 987 234540010 234540011 234540012 234540018 234540017 234540016 234540024 234540023 234540022 LABEL PP1A7X PP1A7Y PP1A7Z PP1A7ZZM PP1A7YZM PP1A7XZM PP1A7ZWB PP1A7YWB PP1A7XWB LENGTH SUPPLIED LABELS 59.5 ft. 59.5 ft. 59.5 ft. 26.5 ft. 26.5 ft. 26.5 ft. 26.5 ft. 26.5 ft. 26.5 ft. (1) RUN 976, (1) RUN 979, (2) X+, (1) X, (1) GND, (1) XZM, (1) XWB (1) RUN 977, (1) RUN 980, (2) Y+, (1) Y, (1) GND, (1) YZM (1) YWB (1) RUN 978, (1) RUN 981, (2) Z+, (1) Z, (1) GND, (1) ZZM (1) ZWB (1) RUN 982, (1) ZZM (1) ZZM+ (1) RUN 983, (1) YZM (1) YZM+ (1) RUN 984, (1) XZM (1) XZM+ (1) RUN 985, (1) ZWB (1) ZWB+ (1) RUN 986, (1) YWB (1) YWB+ (1) RUN 987, (1) XWB (1) XWB+

Refer to Section 623 for installation of Gradient cables in Equipment Room.

6-37

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 634A 1. 2. 3. 4. 5. 6. 7.

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Connect ZM (Zoom) Cables to Pentration Panel

Remove gradient filter terminal covers. Route Run 982, 983, and 984 to right side of gradient filter on Penetration Panel. Note: The end of the cable with terminals will be connected to Penetration Panel. Connect six ZM and ZM+ wires to gradient filter terminals as marked. Note: Red tape indicates positive (+) wires. Align cables to obtain maximum spacing from other cables and conductive metal objects. Note: When tightening nuts, it is important to hold screw body. Tighten nuts to a torque of 35 ftlbs (6774Nm) or after tightening, verify that nothing is loose. Note: Final check is by hand. Make sure screw is fixed, nut is tight, and cable is not detached from terminal lug. Route and secure each cable Run to strainrelief bracket with Ubolts. Secure GND lead wire on Ground Stud for each cable Run. MAGNET ROOM VIEW 1 2 3 4 5 YZM+ Z+ MAGNET ROOM ZZM ZZM+ EQUIPMENT ROOM ZWB

YZM XZM

XZM+

ZZM

7 RUN RUN RUN 982 983 984

6-38

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 634B 1. 2. 3. 4. 5. 6. 7.

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Connect WB (Whole Body) Cables to Pentration Panel

Route Run 985, 986, and 987 to left side of gradient filter on Penetration Panel. Note: The end of the cable with terminals will be connected to Penetration Panel. Connect six WB and WB+ wires to gradient filter terminals as marked. Note: Red tape indicates positive (+) wires. Align cables to obtain maximum spacing from other cables and conductive metal objects. Note: When tightening nuts, it is important to hold screw body. Tighten nuts to a torque of 35 ftlbs (6774Nm) or after tightening, verify that nothing is loose. Note: Final check is by hand. Make sure screw is fixed, nut is tight, and cable is not detached from terminal lug. Secure GND lead wire on Ground Stud for each cable Run. Route and secure each cable Run to strainrelief bracket with Ubolts. Replace gradient filter terminal covers. MAGNET ROOM VIEW 2 3 4 EQUIPMENT ROOM Y+ YZM MAGNET ROOM ZWB+ ZWB

YWB

XZM

YWB+ YWB

XWB

X+

XWB

XWB+

5 6 RUN RUN RUN 987 986 985

6-39

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 634C

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Route Gradient Coil Output Cables Thru Rear Pedestal

Gradient cables will arrive bundled with tiewraps installed.

WARNING!
FERROUS MATERIAL HAZARD! If magnet is ramped, the wheel assemblies attached to the Rear Pedestal for shipping and delivery MUST be removed before moving Rear Pedestal into Magnet Room.

CAUTION
The weight of the Rear Pedestal requires at least THREE people to move Rear Pedestal into position at rear of magnet. 1. 2. Move Rear Pedestal close to its final position at rear of magnet. Route the bundled Gradient Cables above lower cross brace at front of Rear Pedestal.

Note: Routing the Gradient Cables above rather than below the cross brace will reduce white pixels.

Note: Cables and hoses can be routed to exit out of rear or side of Rear Pedestal.

6-40

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 634D

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Install Gradient Cable Restraint Blocks

The gradient cables leads have to be secured as they exit from the Pang Block at rear of End Bell. Twelve Gradient Cable Clamp Blocks (5126956) and Six Covers (5126957) with screws are to be installed as shown below. 1 Remove any TieWraps are in place around cables in area where blocks are to be installed.

2 Install a 2 Blocks and a Cover around each set of 6 cables. Note: Orientate the Restraint Blocks to cables exiting the Pang Block as shown. 3 Install remaining 4 sets of Blocks and Covers at positions shown about 2 inches (50mm) apart.

634E 1.

Cut Cables to Length and Route in Magnet Room

Route Gradient Output cables from Gradient Filter of Penetration Panel thru designated cable trough to rear of magnet. The TRM Body Coil gradient cables should have been routed to area in cable trough at rear of magnet.

CAUTION
2.

Do not cut TRM Body Coil gradient cables. Only Gradient Output cables from Penetration Panel are to be cut to length.

Verify sufficient slack of at least two feet (610mm) is allowed for termination and connection. Mark location for cutting. Cut cables to length and label cut ends with applicable run number.

6-41

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 634F

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Terminate Runs 982, 983, And 984 for Magnet Room

CAUTION
Rear End Bell MUST be installed on magnet with gradient coil cables routed thru block before connecting gradient coil cables to gradient cables routed from Penetration Panel with splices in cable trough. Failure to do so will result in inablility to install Rear End Bell. Note: The following steps are required to terminate the end of the cable that was cut in Step 634E. The other end of the cable is connected to the Penetration Panel. 1. 2. 3. 4. 5. 6. 7. 8. Remove 24 in. (288 mm) of jacket from end of cable Run 982. Trim rubber cable filler strands flush with end of jacket. Cut and remove GROUND wire. Strip 0.75 in. (19 mm) of insulation from end of each wire. Wrap red tape (from Pen Panel end of cable) around wire ONE. Place furnished ZZM+ label on wire ONE. Place furnished ZZM label on wire TWO. Place furnished Run 982 label on cable. Repeat Steps 1 through 7 for Runs 983 and 984. Note: Red tape, Yzm+ (RUN 983) and Xzm+ (RUN 984) labels are attached to wire ONE.
24 IN (288mm) ONE TWO GND

GRADIENT CABLE (MAGNET ROOM) (RUN 982)

2
ONE TWO GND GRADIENT CABLE (MAGNET ROOM) (RUN 982)

3 4 5
RUN 982 MG3A32TS1ZZM

ZZM

TWO

6 8

6-42

ZZM+

ONE

2xxxxxxxx GE COMAPNY

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 634G

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Terminate Runs 985, 986, And 987 for Magnet Room

Note: The following steps are required to terminate the end of the cable that was cut in Step 634E. The other end of the cable is connected to the Penetration Panel. 1. 2. 3. 4. 5. 6. 7. 8. Remove 24 in. (288 mm) of jacket from end of cable Run 982. Trim rubber cable filler strands flush with end of jacket. Cut and remove GROUND wire. Strip 0.75 in. (19 mm) of insulation from end of each wire. Wrap red tape (from Pen Panel end of cable) around wire ONE. Place furnished ZWB+ label on wire ONE. Place furnished ZWB label on wire TWO. Place furnished Run 985 label on cable. Repeat above process for Runs 986 and 987. Note: Red tape, YWB+ (RUN 986) and XWB+ (RUN 987) labels are attached to wire ONE.
24 IN (288mm) ONE TWO GND

GRADIENT CABLE (MAGNET ROOM) (RUN 982)

2
ONE TWO GND GRADIENT CABLE (MAGNET ROOM) (RUN 982)

3 4 5
ZWB+ ZWB
ONE TWO

RUN 985 MG3A32TS1ZWB

6 8

634H

Connection of Body Coil Gradient Cables

WARNING!
THE RATCHETING CRIMP TOOL REQUIRED FOR THIS PROCEDURE CONTAINS FERROUS MATERIAL AND WILL BE STRONGLY ATTRACTED TO MAGNET WHICH COULD CAUSE INJURY OR DAMAGE TO THE EQUIPMENT.

6-43

2xxxxxxxx GE COMAPNY

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 634H 1.

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Connection of Body Coil Gradient Cables (Continued)

Cable connection process shown below is for Run 987. Repeat process for Runs 982, 983, 984, 985 and 986. Note: Red tape, XWB+ (RUN 987), YWB+ (RUN 986), ZWB+ (RUN 985), XZM+ (RUN 984), YZM+ (RUN 983), and ZZM+ (RUN 982) labels are attached to wire ONE. Position Cold Shrink tubing (2218148) on output cables prior to connecting. Insert ends of Run 987 into correct butt splice (2218149) on Body Coil gradient cable. Crimp butt splices using a T&B WT 3185 Gradient Cable Crimper (2144091) with #2 die. Locate Cold Shrink tubing at desired starting point. Position tubing 11/2 to 2 inches (38 to 50mm) beyond end of connector. Pull loose tab gently while unwinding in a counterclockwise direction. Repeat process for other cable. Connection of cables and installation of Cold Shrink tubing is now complete. Note: Illustration below shows a typical cable routing for connection of gradient output cables within an enclosed cable trough. Note: Cold Shrink Tubing (2218148) can be found in the Magnet Field Spares Kit delivered with the Magnet

2. 3. 4. 5. 6. 7.

Gradient Cables from Body Coil RED TAPE

Gradient Output Cables from Penetration Panel 2 RED TAPE

3 4

6-44

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 635

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

INSTALLATION OF REAR PEDESTAL Remove Rear Bridge Support Bracket

635A 1.

Remove Rear Bridge Support bracket and discard. It is held in place with four screws.

635B 1.

Install Rear Pedestal Support Studs

Remove the Rear Pedestal Support Studs (2365032) from the Rear Pedestal Assembly and install into two lower holes on Rear End Bell.

6-45

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 635C Position Rear Pedestal

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

CAUTION
The weight of the Rear Pedestal requires at least THREE people to move Rear Pedestal into position at rear of magnet. 1. 2. Move Rear Pedestal into position at rear on magnet as shown below. Remove four 10M hex nuts (21098773) and washers (22287172) from end of Bridge in magnet. These will be used for attaching the Rear Split Bridge.

Note: Dress and align all cables and hoses routed thru Rear Pedestal while moving Rear Pedestal closer to magnet. 3. 4. 5. Move Rear Pedestal forward and engage front split bridge studs. Adjust the Rear Pedestal on each side to level the Rear Split Bridge with the Front Split Bridge. Fasten Front and Rear Split Bridge with four nuts and washers that were removed from end of Front Split Bridge. make sure Rear Bridge Support Bracket has been removed, refer to Procedure 635A. 3 1 2 5

6-46

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 635D

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Connect Front and Rear Split Bridge

Note: Tongue to Bridge alignment is critical to smooth belt tracking and operation. Make sure Tongue is square to Bridge and Rear Pedestal. 1. 2. 3. Center Rear Pedestal on Rear Endbell and attach to Support Studs using M10 Nut (21098754) and Flat Washer (21098755) on each stud. Adjustment of Rear Pedestal location can be made at this slotted hole location when conecting the Split Bridge. Route Ground cable from Rear Pedestal under magnet and connect to same Ground Stud as Magnet ground.

6-47

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

636

LEVELING OF BRIDGE Rear End Bell notch.

NOTE: Make sure that surfaces between front and rear Bridge surfaces are even. This will provide smooth travel for the Carriage wheels. NOTE: Rear Pedestal must be adjusted to maintain full contact between bottom of Bridge and top of Rear Pedestal.

NOTE: LPCA and Cable Track not shown for clarity.

1 Level Bridge and make sure that Bridge surface is no less than 2mm and no more than 4mm above the notches on the front and rear End Bells. Tighten Vertical Adjustment Fasteners.

Adjust leg screws on both sides as needed to level Rear Pedetal.

Front End Bell notch.

Put a level across front of Bridge.

Level front of Bridge by adjusting right or left slider screws to center the bubble.

6-48

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 637 INSTALLATION OF DRIVE BELT Remove LPCA Cover 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

637A

Unfasten the four screws that attach the cover to the base plate. Unfasten 6 screws and remove Rear Cover (2386346) from LPCA

Carefully remove cover from base plate assembly.

6-49

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 637B Route and Attach Drive Belt 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Route one end of Drive Belt delivered with Front Split Bridge thru opening in LPCA Base Plate and attach with clamp. 6 Position Belt under Clamp Plate and attach.

Route other end of drive belt from front of Bridge towards Tension Arm Pulley. 3 Release Tension Arm.

637C

Final Adjustment of Drive Belt 3 Tighten Clamps and trim excess Belt length. 4 Fasten Carriage Cover to Carriage with four screws removed in Procedure 637A.

2 (IF NECESSARY): Loosen Clamp and adjust position of Belt one tooth at a time until desired tension is obtained.

Check Belt tension by closing Tension Arm. (See Note)

NOTE: Tension Arm should lock crisply into position without excessive force (which could cause later problems). If adjustment is necessary go to step 2. If not, go to step 3.


6-50

Route Belt under and through opening to top of Carriage. Rear Pedestal Side View

4 Feed back end of Belt underneath Bridge, around Tension Arm Pulley and rear pulleys into Bridge.

NOTE: Check to verify that belt is not caught around spacers near rear pulley.

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 637D Installation of Sensor Flag

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

NOTE: The following Step is needed if flag was removed during routing of drive belt.

Photo Sensor

1 Install Flag Assembly on Drive Belt in line with Bridge mounted Photo Sensor.

Front of Magnet

637E

Installation of LPCA Cover

3 Attach the cover to the base plate using four screws previously removed.

2 Attach Rear Cover using 6 screws previously removed. 1 Carefully place cover on base plate assembly.
A

6-51

CABLE & SYSTEM INSTALLATION

NOTE: Function of the Flag is to interrupt the reflected limit switch sensor beam when the Carriage is in home position all the way forward in the bore.

GE MEDICAL SYSTEMS REV 1 638

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

ROUTE AND CONNECT CABLES AT REAR PEDESTAL Route/Connect Runs 282, 414, 422, 781, 782, 1057, Multicoil, & Front End Bell Speaker Cable

638A 1. 2. 3. 4. 5. 6. 7. 8. 9.

Route and connect RF Bias cables from RF Coil to bracket as marked. Connect Run 781 to Bias cable J5 on bracket and route to J75 on the Penetration Panel. Connect Run 782 to Bias cable J6 on bracket and route to J76 on the Penetration Panel. Wrap both Bias connections with 177 x 200 Connector Wrap (22361302). Route/connect Front End Bell microphone cable Run 414 to J9. Route/connect speaker cable from Patient Comfort Module to J8. Route/connect Run 282 from J12 on Pen Panel to J6. Route/connect Run 1152 from Shim Filter on Penetration Panel to Shim Filter cable thru Rear End Bell gasket. Route/connect Run 422 from J2 on SRI to J4 on Magnet Interface #1.

2 3 1

4 5 9 6 7 8

6-52

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 638B 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

CONNECT REMAINING CABLES TO REAR PEDESTAL

Route and connect Runs to 16 Channel Switch and Bulkhead Interface as indicated.

Run 1201 to J2 Run 1200 to J1 Run 1215 to J20 Run 1216 to J5 Run 1197 to J25 Run 1217 to J10 Run 1203 to J11 Run 1202 to J4 Run 1198 to J1 Run 1199 to J2

638C 1. 2. 3.

Route/Connect Run 318 to Longitudinal Drive Motor

Route Run 318 from J11 on Penetration Panel tand connect to Longitudinal Drive Motor cable. Wrap both connectors with 177 x 200 Connector Wrap (22361302). Tywrap cable to motor to prevent movement. 3 1

Run 318

6-53

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

638D

Connect Cables to Hybrid Splitter

CAUTION
The RF Drive Input cables utilize a corregated solid RF shield. If mishandled, the cable can be kinked and subsequently broken.

Also, use care when connecting these cables to the RF Splitter connectors J3 and J4. Cross threading or inadequate engagement of the center pin can cause arcing and damage the cables. 1. Route and connect I and Q RF Drive input cables from RF Coil and connect as marked to J1 and J2 on bracket.

Note: Runs 743 and 744, Body Coil to Hybrid Splitter cable connections, may be reversed depending on magnet field polarity. Consult local Field Engineer for proper routing. 2. 3. 4. 5. Connect Run 743 to J2 and Run 744 to J1 on bracket. Connect RF Flexible cable Run 1160 to J1. Refer to Procedure 641B for instructions on installing the Times Microwave cables. Other end is connected to Run 257. Route and connect Run 780 from J74 on Penetration panel to J6 at front of Hybrid Splitter. If present, remove the two Ground Isolation Filter Boxes (5118952) that are installed on end of Hybrid Splitter and connected to J3 and J4. Reconnect Runs 743 and 744 to J3 an dJ4 on Hybrid Splitter.

2 1

5 4

6-54

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 639

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

INSTALLATION OF AIR HOSE FROM REAR OF MAGNET TO BLOWER BOX

Install Blower Cabinet (2241999) according to Site Design (architectural) specifications.

WARNING!
RF SHIELD INTEGRITY MUST BE MAINTAINED FOR MOUNTING BLOWER BOX WITHIN THE MAGNET ROOM 1. Location of cabinet and type of fastener used for securing Blower Cabinet outside of 200 Gauss range should be specified in the Site Design document.

The following items are part of Blower Hose Kit, 2129412-5 2. 3. 4. Connect 2 inch air hose from Rear End Bell to Reducer (2175143) with 2 inch clamp (46208765P28). Do not cut or reduce the length of the 2 inch hose. Connect 4 inch air hose (46282666P11) to reducer with 4 inch clamp (46208765P34). If necessary, the 4 inch hose may need to be cut to reduce length. Route 4 inch hose to Blower Cabinet. Trim to appropriate length and attach to vent opening with 4 inch Hose Clamp (46208765P34).

1 2 3 4 NOTE: If needed, Blower Hose Cover can be cut and modified.

6-55

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

6310 ROUTE AND CONNECT CABLES AT FRONT OF MAGNET ENCLOSURE

CAUTION No metal to metal contact is allowed for connectors to other areas. i.e. connector touching the magnet.
1. 2. 3. 4. 5. 6. 7. Route Runs 778 and 779 from J72 and J73 on Penetration Panel and connect to front RF Bias cables J3 & J4. Wrap the RF Bias cable connections with one 177 x 200 Connector Wrap (22361302). Route Run 1260, Bore Temperature Sensor, from Front End Bell and connect ot J10 on SRI. Route Front End Bell Microphone and Speaker cables underneath manet to Rear Pedestal and connect to Patient Communication Box as marked. Connect Run 387 to Dock Cable (MG2 A29 P1). If needed to prevent contact with other metal parts, cover connectors with provided velcro jacket or tape. Connect Dock Cable to J2 on Dock Limit Box Route/Connect Run 421 from SRI J3 and connect to J1 on Dock Limit Box.

Note: Dock Assembly not shown for clarity.

3 4

5 6 7

6-56

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

6311 DOCK ADJUSTMENT FOR PATIENT TRANSPORT GAP

WARNING!
DOCK CONTAINS FERROUS MATERIAL. HANDLE WITH CARE DURING INSTALLATION. 1. 2. Start with two (2) 11mm washers on each side between Dock and Dock Support. If adjustment is required, add or remove washers until proper gap is acheived.

Note: Gap between Patient Transport front rubber inlay and front of End Bell must be 0.75 Inches  0.25 (19.06.4mm). See Step 1 below.

Note: Floor stud installed by RF room vendor/mechanical contractor.

1 2

6-57

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

64 INSTALLATION OF CABLES THAT RUN BETWEEN EQUIPMENT ROOM AND MAGNET ROOM 641 TIMES MICROWAVE RF CABLES INSTALLATION Cable Information

641A

Five cables, three extension cables, and other associated parts to be installed are provided for this procedure. Also provided are two Times Microwave stripping tools for LMR 1200 (2384858) and the LMR 600 (2352193), and the Times Microwave cutting tool (5111565).

CAUTION
Do not subject LMR600FR cable (Head) to a bend radius less than 3 inches (75mm). Do not subject LMR1200FR cable (Body) to a bend radius less than 6.5 inches (165mm). The cables will be permanently damaged if it is kinked by a bend radius less than the specified minimum. Two extension cables, 51162132, Run 1249 (Head) and 5121269, Run 1159 (Body) are connected to the RF Amplifier module in the RF/System Cabinet. The other extension cable, Run 1250, 5121245, is connected to the Body Hybrid Splitter in the Rear Pedestal.

Run 1249 and Run1250

Run 1159 Cable 23165842 (Body, Runs 235 & 257), will be cut to length to make two cables and terminated for routing in the equipment and scan rooms.

Run 235 and Run 257 Cable 5115915 (Head, Runs 1198 & 1199), will be cut to length to make two cables and terminated for routing in the scan room.

Run 1198 and Run 1199 Cable 5115916 (Head, Run 1238), will be cut to length and terminated for routing in the equipment room.

Run 1238 The remaining connectors, fasteners, and labels will be installed as directed in the following steps.

6-58

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

641B

Install Runs 235, 257, 1238, 1249, 1250, and 1259

RF cables will be routed and connected as indicated in the following steps. 1. 2. 3. Connect extension cable Run 1259 to J4 Body Output at rear of RF Amp in RFS cabinet. Connect end of Times RF Body cable marked (RUN 235) FJ5 to Run 1259, Body extension cable. Attach label RUN 235, Connect to Run 1259 (5121244) over existing label. Route cable to J44 at Penetration Panel. Determine if any takeup slack is needed and cut to length using provided cutting tool. Refer to Times Connector Installation Procedures provided with cables for cutting information. Refer to Procedure 641E for termination procedure of cut end. Connect Run 1250 to J1 on Hybrid Splitter. Move remainder of RF Body cable to Scan Room and connect other end of cable marked (RUN 257) MG3A3J1 to Run 1250 at Rear Pedestal as shown below. Attach label RUN 257, Connect to Run 1250 (51212442) over existing label. Route remainder of cable to J44 at Penetration Panel. Determine if any takeup slack is needed and cut to length. Refer to Times Connector Installation Procedures provided with cables for cutting information. Refer to Procedure 641E for termination instructions of cut end. Connect extension cable Run 1249 to J3 Head Output at rear of RF Amp IN RFS cabinet. Connect end of Times RF Head cable marked (RUN 1238) PP1J4 to J4 at Penetration Panel. Route to RF Head extrension cable at rear of SRF cabinet. Determine if any takeup slack is needed and cut to length using provided cutting tool. Refer to Times Connector Installation Procedures provided with cables for cutting information. Refer to Procedure 641D for termination procedure of cut end.

4. 5.

6.

7. 8.

3 1 8 2

5 Note: If cables and hoses are to exit this side of Rear Pedestal, move Coil Cable Takeup Interface to opposite side of Rear Pedestal. If Interface bracket is moved to other side, connectors on bracket need to be removed, turned end for end, and reinstalled.


6-59

Coil Cable Takeup Interface (MG3 A17)

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 641C Install Runs 1198 and 1199 1 2

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Connect Run 1198 end of cable to J151 on the RF Tranmitter Switch on Penetration Panel.

Route to Bulkhead Interface on Rear Pedestal. Determine if any takeup slack is needed and cut to length using provided cutting tool. Refer to Times Connector Installation Procedures provided with cables for cutting information.

3 Refer to Times Connector Installation Procedures provided with cables for information on installing an N Connector (2352195) on end of the RF cable. 4 Attach label RUN 1198, MG3PED IFJ1 (51162142) to end of cable . Connect cable to J1. 5 Connect other end of cable marked Run 1199 to J153 on the RF Tranmitter Switch. 7

Repeat steps 2 and 3 above for installing connector on end of cable

Attach label RUN 1199, MG3PED IFJ2 (51162143) to end of cable . Connect cable to J2.

641D

Terminate and Connect RF Head Coil Cable

Perform the following steps for terminating and connecting the RF Head cables in the Equipment room. 1. Refer to Times Connector Installation Procedures provided with cables for information on installing a Straight Male N Connector (2352195) on each end of the RF cable. Use provided cutting tool (5111565) and the provided LMR600 Stripping Tool (2352193). Attach label RUN 1238 CONNECT TO RUN 1249 (5116214) to cable as shown in Equipment room. Connect cable to Run 1249.

2. 3.

1 3

6-60

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 641E

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Terminate and Connect RF Body Cables at Penetration Panel

Perform the following steps for terminating and connecting the RF Body cables at the Penetration Panel J44. 1. Refer to Times Connector Installation Procedures provided with cables for information on installing an EIA Right Angle Connector (2384859) on each end of the RF cables. Use provided cutting tool (5111565) and the LMR1200 Stripping Tool (2384858). Note Make sure the fastener holes in the connector flanges align with the holes in the Penetration Panel so the cables will route toward floor when installed. 2. 3. 4. Attach label RUN 235, PP1J44 (2357230) to cable as shown on Equipment room side of Penetration Panel. Attach label RUN 257, PP1J44 (23572302) to cable as shown on Scan room side of Penetration Panel. Insert Interface Pin into one of the connectors. At J44, align connectors on each side of Penetration Panel and push together. Fasten together using three of each of the following: D 1/420 x 1.25 Hex Head Bolt (brass) D 1/4 Screw Size Washer (phos bronze) D 1/420, 7/16 Hex x 3/16 thk Nut (brass)

Interface Pin

6-61


CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 642 VACUUM HOSE INSTALLATION

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

The parts for installing the vacuum hose are supplied with the Twin Vacuum Hose Assembly (2295226). Note: Vacuum Grease must be applied to all Centering Rings (2293608) and outside edge of vacuum connector flanges. 642A 1. 2. 3. 4. Connect Vacuum Hose to Magnet

Install Elbow (2293609) to End Bell using centering ring (2293608) and chain clamp (22936182). Tighten chain clamp to finger tight. Route vacuum hose (2293607) underneath Rear Pedestal, under magnet end cover, and thru hole to Elbow on End Bell. At end of vacuum hose insert Hose Connector (2293616), apply vacuum grease to ouside edge of connector and fasten with two (2) Clamps (22940103). Apply vacuum grease to outside edge of Hose Connector flanges and respective connector flanges on End Bell. 1

4 3 2

6-62

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 642B 1. Route Vacuum Hose

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Route Vacuum Hose thru wave guide in Penetration Panel to TAC Cabinet.

642C 1. 2. 3. 4.

Connect Vacuum Hose to TAC Cabinet

Route vacuum tube to J11 at back of cabinet and cut to length. Install 1 Hose Connector (2293616) in the end of hose and attach with wormdrive clamp (22940103). Apply vacuum grease to centering ring (2293608) and connect vacuum tube to J11 using supplied Chain Clamp (22936182). Apply vacuum grease to outside edge of hose connector flanges and respective connector flanges on TAC Cabinet.

J22

J11

J27 J12 J10

J9

1 2 3 4

6-63

CABLE & SYSTEM INSTALLATION


J8

GE MEDICAL SYSTEMS REV 1 643

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

PNEUMATIC PATIENT ALERT INSTALLATION

The customer and users should be involved in the following decisions: D Routing of Pneumatic Tubing D Location of Control Box

CAUTION
Do not substitute other types of tubing for the pneumatic tubing supplied with this Installation Kit. Substitution of other types of tubing may cause a control box malfunction. 643A 1. SQUEEZE BULB INSTALLATION

Insert end of Pneumatic Tubing into Squeeze Bulb approximately 1/2 inch (13mm). For ease of pneumatic tubing insertion, pinch pneumatic tubing in one hand while squeezing squeeze bulb and pushing pneumatic tubing in with the other hand. This will create a positive pressure and expand squeeze bulb opening. Before cutting pneumatic tubing from roll to length required, verify that when the squeeze bulb is held by a patient being scanned, the tubing length can be routed to the CALL connector on the Remote PAC Interface Assembly without getting in the way of operator or patient during use. 1

2.

643B 1.

CONNECT TUBE TO REMOTE PAC INTERFACE ASSEMBLY

IConnect end of tubing to CALL connection on Remote PAC Interface Assembly.

6-64

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

643C

MAGNET ROOM ROUTING OF TUBING & WAVE GUIDE INSTALLATION

CAUTION
Make sure that pneumatic tubing is not routed where it can be stepped on, pinched, or have an object set on it. The control box alarm will not work if pneumatic tubing is pinched. Also, make sure that the pneumatic tubing can be routed from the PAC connector through the Rear Pedestal, and to the Penetration panel without being pinched along the route by other cables. 1. Push end of pneumatic tubing onto PACII tubing connector.

Note: If the pneumatic tubing is to be pulled thru fiberoptic conduit, coat with baby powder, not grease, to allow easier pulling. 2. 3. 4. Route Pneumatic Tubing from Rear Pedestal to Penetration Panel with other system cables. Insert Wave Guide thru ALERT AIR LINE hole and fasten to Penetration Panel with nut on each side. Route end of pneumatic tubing from Magnet Room side of Penetration Panel thru Wave Guide into the Equipment Room where the Control Box will be installed.

4 1

6-65

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 643D

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

DETERMINING NEED OF EXTENDER BOX

If installation requires greater than 115 feet of pneumatic tubing between the squeeze bulb (hanging on Magnet Enclosure) and control box (near Operators Console), Extender Kit, 46317758P2, must be ordered. D If the pneumatic tubing is long enough to reach control box: The extender is not needed. Go to Procedure 643F, CONTROL BOX INSTALLATION. D If the pneumatic tubing is not long enough to reach control box: The extender is needed. Go to Procedure 643E, INSTALLATION OF EXTENDER KIT. 643E INSTALLATION OF EXTENDER KIT (46317758P2)

CAUTION
Do not mount Extender Box in Magnet Room because the box contains ferrous parts. Install Extender Box 1. 2. 3. 4. 5. 6. Choose the mounting location for extender box on equipment room wall next to Penetration Panel. Extender box will be mounted with pneumatic and electrical jacks on side of extender box. Cut a two inch long piece of Velcro loops and a two inch long piece of Velcro hooks. Both are obtained from the Pneumatic Patient Alert Kit. Clean the back of extender box, remove protective paper from back of Velcro loops, and attach Velcro to center of extender box back. Clean wall in the area where extender box will be mounted, remove protective paper from back of Velcro hooks, and attach velcro to wall in same orientation as Velcro loops on extender box. Push end of pneumatic tubing fully onto Feedthru connector on Extender Box. Push the jack on one end of the 65 foot Extender Wire into the plug on side of Extender Box.

5 6

6-66

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 643E

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Installation of Extender Kit (46317758P2) (Continued)

Connection of Extender Wire to Control Box 1. 2. Route the extender wire from extender box to control box. Control box will be installed near Operator Console, within five feet of an electrical outlet. Push the jack at other end of extender wire into the plug on back of control box.

643F

CONTROL BOX INSTALLATION OPTIONS

Consult with the operator in choosing the mounting orientation for the control box. Some key factors to consider are ease of use by operator, remaining within sight of operator, and remaining within five feet of an electrical outlet. Choose one of the following control box locations: D Mount control box on a wall or other vertical surface per Procedure 643G, CONTROL BOX INSTALLATION ON WALL OR UNDER SHELF. D Mount control box under shelf per Procedure 643G, CONTROL BOX INSTALLATION ON WALL OR UNDER SHELF. D Place Control Box on a counter top, desk top, or other horizontal surface. Set the Control Box near Operator Console and within five feet of electrical outlet. Control Box should be placed within sight of operator. Go to Procedure 643K, FINAL CONNECTIONS TO CONTROL BOX. Note: The installer must provide additional mounting hardware if supplied screws or adhesive backed Velcro are not adequate for installation of the control box.

6-67

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 643G 1. 2. 3.

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

CONTROL BOX INSTALLATION ON WALL OR UNDER SHELF

Clean Control Box surface. Remove adhesive backing from metalized nameplate and apply over silk screened lettering as shown. Select one of the following for installation of Control Box: D If control box will be mounted to WALL or SHELF with Velcro, go to Procedure 643H D If control box will be mounted to WALL with screws, go to Procedure 643I D If control box will be mounted to SHELF with screws, go to Procedure 643J 1 2

643H 1. 2. 3. 4. 5. 6. 7. 8.

Control Box Mounted With Velcro

Clean the bottom surface of the Control Box. From the Pneumatic Patient Alert Kit, cut a three inch long piece of Velcro loops and a three inch piece of Velcro hooks. Remove protective paper from back of Velcro loops, and attach Velcro to center bottom of control box. Choose the mounting location for control box on wall (or under shelf) next to operator console. Control box should be mounted within sight of operator. Clean the area where control box will be mounted. Remove protective paper from back of Velcro hooks and attach to mounting location in same orientation as Velcro on box. jMount the control box by pressing Velcro on control box against Velcro on mounting location and twisting slightly. Proceed to Procedure 643K, FINAL CONNECTIONS TO CONTROL BOX. 1 2 3 6

6-68

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 643I

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Installing Control Box with Mounting Screws on Vertical Surface

Choose the mounting area near the Operator Console and install mounting screws (item 9). Control box should be mounted within sight of operator. 1. 2. 3. 4. 5. 6. 7. 8. Determine which of the shown keyhole slot spacings are applicable to Control Box and select appropriate dimensions to use. Mark top mounting hole position on wall. Mark bottom mounting hole position on wall. Drill a 1/16 in. pilot hole at locations if furnished screws are used. Install screws to depth as shown. Center box keyhole slots over screws, push in on box, and then push downward to seat screws within slots. Tighten or loosen screws as necessary to obtain a snug fit of box against wall. Proceed to Procedure 643K, FINAL CONNECTIONS TO CONTROL BOX.

Wall Ceiling or Shelf Bottom 1 21/8 in. or 15/8 in. 11/2 in. Minimum or 2 in. Minimum 2 4 6 3 4

0.20 in. 5 7 8

6-69

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 643J

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Installing Control Box with Mounting Screws Under Shelf

Choose the mounting area near the Operator Console and install mounting screws (item 9). Control box should be mounted within sight of operator. 1. 2. 3. 4. 5. 6. 7. 8. Determine which of the shown keyhole slot spacings are applicable to Control Box and select appropriate dimensions to use. Mark front mounting hole position on underside of shelf. Mark rear mounting hole position on underside of shelf. Drill a 1/16 in. pilot hole at locations if furnished screws are used. Install screws to depth as shown. Center box keyhole slots over screws, push in on box, and then push inward to seat screws within slots. Tighten or loosen screws as necessary to obtain a snug fit of box against wall. Proceed to Procedure 643K, FINAL CONNECTIONS TO CONTROL BOX. 1

0.20 in. 2 4

21/8 in. or 15/8 in. 3 4

11/2 in. Minimum or 2 in. Minimum

7 8 6

6-70

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 643K 1. 2. 3. 4. 5. 6.

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Final Connections to Control Box

Bring the pneumatic tubing to back of control box and mark length. Leave two inches of extra pneumatic tubing for connection. Cut tubing and push fully onto feedthru connector on back of control box. jPlug Power Supply Jack into Control Box. Set Power Supply to outlet voltage (115VAC or 230VAC). Plug Power Supply into AC outlet. Check that green LED is lighted. 5

1 2

6-71

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1 644 EMERGENCY OFF CONNECTIONS

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

Run 296 is routed to EO2 (Equipment Room Emergency Off) and connected to EO2 and wire from Facility Disconnect. Run 297 is routed from Penetration Panel to EO1 (Magnet Room Emergency Off Button). Refer to Direction 5133302, Signa EXCITE 1.5T TwinSpeed PreInstallation, Section 5, Protective Disconnect Device, for detailed wiring diagram. 1. Complete routing of Run 296 to EO2 (Equipment Room Emergency Off) location, and trim cable to length. Locate red and black pair of wires and prepare ends for applicable terminals.

Note: In Illustration below, black and red wires are used for connections in Runs 295 and 297. Actually any pair of wires on these runs could be used so long as both ends are consistent with one another. Runs 296 and 297 are actually nine wire cables. 2. 3. 4. 5. 6. 7. 8. At equipment room room Emergency Off (EO2) location, terminate red wire at end of Run 296 with local supplied terminal and connect to customer supplied Equipment Room Emergency Off Button (EO2). Terminate customer supplied wire (from fuse in Protective Disconnect Device) with local supplied applicable terminal and connect to customer supplied Emergency Off Button (EO2). Terminate customer supplied wire from R1 in Protective Disconnect Device with local supplied pushon terminal. Terminate black wire at end of Run 296 with local supplied pushon terminal. Connect black wire from end of Run 296 to customer supplied wire from protective disconnect device with local supplied inline splice. Complete routing of Run 297 to EO1 (Magnet Room Emergency Off) location, and trim cable to length. Locate red and black pair of wires and prepare ends for applicable terminals. Terminate red and black wires at end of Run 297 with local supplied terminals and connect to customer supplied Magnet Room Emergency Off Button (EO1).
EO2 EQUIPMENT ROOM EMERGENCY OFF BUTTON R1 24 VAC WIRES FROM FACILITY DISCONNECT IN-LINE SPLICE BLK

EO1 MAGNET ROOM EMERGENCY OFF BUTTON BLK RUN 296 J18 RUN 297 NC RED

NC

RED

PROTECTIVE DISCONNECT DEVICE (CUSTOMER SUPPLIED)

PENETRATION PANEL (PP1)

6-72

CABLE & SYSTEM INSTALLATION

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

SECTION 7 FINAL ENCLOSURE INSTALLATION


TABLE OF CONTENTS
SECTION 71 72 73 74 75 76 77 PAGE INSTALL BASE TRIM COVERS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72 INSTALL ENCLOSURE TOP COVERS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72 INSTALL REAR PEDESTAL COVERS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73 INSTALL RATING PLATES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73 INSTALL LASER ALIGNMENT LIGHT WARNING LABELS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74 PATIENT TRANSPORT INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75 INSTALLATION OF HEARING LOSS WARNING SIGN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76

7-1

FINAL ENCLOSURE INSTALLATION

GE MEDICAL SYSTEMS REV 1 71 INSTALL BASE TRIM COVERS 22259824 End Bell

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

72 INSTALL ENCLOSURE TOP COVERS

Note: All enclosure seams and gaps visible to the operator and patient shall be less than 0.25 inches (6.35mm). This applies to the mating seam of the two Side Top Covers to the Front Top Cover Gap uniformity along any continuous seam shall be +/ 0.0625 inch (1.59mm)
1 Install front (2217457) and rear (2217458) Top Covers as shown using M10 x 25mm Low Profile hex screws and blue Loctite. 2217457 2270873

22173122 End Bell Spacer 22173102 22259823

Foot Cover

Slide front Foot Covers (22173102 & 22173082) from side, above spring clip, and between End Bell Flange and End Bell Spacers. Install remaining side Base Trim Covers as shown.

Note: The gap from the bottom of the Base Trim to floor should be a nominal 0.50 inch (13mm).

22173082

22173112

Install side Top Covers as shown.

2217458 22708732

Place a level against vertical surface of Top Covers. If top of covers are leaning towards magnet, cut a notch at this location and reinstall. Repeat as necessary until vertically level.

7-2

FINAL ENCLOSURE INSTALLATION

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

73 INSTALL REAR PEDESTAL COVERS 4 Pedestal Covers (2217704 & 2217705) require an opening to be cut out depending on routing of cables and hoses. Covers are marked in three places (left side, right side, or rear) for cutting. After cutting out appropriate opening, hang covers on Rear Pedestal. If covers need adjustment, loosen the screws at each end of the support bar and reposition bar as necessary. 1 Remove mating piece of velcro from Rear End Bell Cover Plates (2228970 & 22289702) and attach to Rear Pedestal Covers 2 Remove mating velcro and attach to back side of Arc Covers.

5 6

2228970 (This Side) 22289702 Other Side)

Velcro from Cover Plates 7 Fasten covers together with furnished hardware. 3 Attach Rear End Bell Cover Plates (2228970 & 22289702) to both sides of Rear Pedestal with furnished hardware two M4 washers (2109879) and two M4 x 8mm Pan head screws (210987313) for each plate.

74 INSTALL RATING PLATES

1 If not already applied, place the following Rating Plates on Bridge Rear Plate: 46302200P7 Serial #, Model # 2142857 ETL List for Global Certification 2117390 CE International Approval 2132620 CE Marking Label Class A Group 1

7-3

FINAL ENCLOSURE INSTALLATION

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

75 INSTALL LASER ALIGNMENT LIGHT WARNING LABELS A sheet of Laser Alignment Light Warning Labels has been supplied with the magnet, attached to the magnet bridge. The magnet enclosure is shipped with a label in English attached below each of the three laser lights.

For those sites which require a label other than English, peel off the appropriate label and affix over the English labels at the three laser light locations.

French

German

Portuguese

Spanish

Italian

Swedish

Japanese

Chinese

7-4

FINAL ENCLOSURE INSTALLATION

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

76 PATIENT TRANSPORT INSTALLATION Patient Transport is shipped covered with protective cardboard which should remain in place until ready to be moved into the magnet room. Cover table with plastic previously removed when checks and adjustments are completed. 1. 2. 3. Unlock casters and move Patient Transport to magnet room for adjustments and function checks at the dock. Remove protective plastic sheet from table. Perform mechanial procedures in MR CDROM, Set up and Calibration: Patient Transport, LIGHTWEIGHT PATIENT TRANSPORT.

Note: If hardware fastening procedures for installation of the dock and support brackets were followed, the hardware used to fasten the dock and dock supports was left loose. The loose hardware permits a quick and easy dock centering adjustment with table docked. 4. If additional Patient Transports are supplied, refer to Direction 15477, Second Patient Transport Option Installation, (delivered with M1020BD option).

PATIENT TRANSPORT

CASTER LOCK

7-5

FINAL ENCLOSURE INSTALLATION

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

77

INSTALLATION OF HEARING LOSS WARNING SIGN


2353850

Remove applicable language warning label from HEARING PROTECTION REQUIRED sheet of 8 labels (2353850).

ENGLISH

GERMAN

FRENCH

ITALIAN

WARNING
SPANISH PORTUGUESE JAPANESE CHINESE

Possible Hearing Loss


All persons in the scan room are required to wear hearing protection during a scan.

2 Attach WARNING label to outside of magnet room door.

7-6

FINAL ENCLOSURE INSTALLATION

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

SECTION 8 OPERATOR WORKSPACE


TABLE OF CONTENTS
SECTION 81 82 83 84 85 86 87 88 89 PAGE OPERATOR WORKSPACE TABLE ANCHOR INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . 82 SECURE OPERATOR WORKSPACE TABLE TO FLOOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83 POSITION COMPUTER CABINET . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84 SEISMIC ANCHORING (IF REQUIRED) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84 COMPUTER CABINET CABLE ROUTING AND CONNECTIONS . . . . . . . . . . . . . . . . . . . . . . . . . 85 ROUTE CABLES FOR LOWER OPENING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86

POWER STRIP CONNECTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87 LINUX PC CABLE CONNECTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88 FINAL POSITIONING OF GOC CABINET . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89

810 INSTALL SCIM/KEYBOARD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 810 811 INSTALL MONITOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 811 812 INSTALL SCSI TOWER AND LCD DISPLAY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 812

8-1

OPERATOR WORKSPACE

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

81

OPERATOR WORKSPACE TABLE ANCHOR INSTALLATION

Refer to room layout plan and consult with Site customer for exact location of Operator Workspace Table. Items installed are part of Color Monitor Table Tie Down Kit, 2154928. 1. 2. 3. 4. Remove and discard the two (2) cover plugs. Locate and mark on floor the two hole locations at the front of the table legs. Move table to prepare for installing the floor anchors in Steps 3 and 4. At each hole marked in Step 2, drill a .38 inch hole and insert 1/420 x .375 diameter anchor (2168253). Secure anchor in place with supplied Setting Tool (2168781).


30.55 (776)
TOP VIEW

52.75 (1340)

8-2

OPERATOR WORKSPACE

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

82

SECURE OPERATOR WORKSPACE TABLE TO FLOOR

Items installed are part of Color Monitor Table Tie Down Kit, 2154928. 1. 2. 3. 4. 5. 6. Insert Bushing (2151454) in table leg hole. Insert as many Flat Washers (2168251) as required (3 provided for each leg) for securing table to floor. Insert one Lock Washer (2168252) on top of Flat Washer(s). Depending upon depth of previously installed anchor, install either the 13/4 inch long or the 2 inch long 1/4 x 20 Hex Socket Cap Screw. Install black nylon Plug (2168254). Repeat anchoring process below for other leg. 5 4 3 2 1

8-3

OPERATOR WORKSPACE

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

83

POSITION COMPUTER CABINET

Consult with Site customer for exact location of Global Operator Cabinet (GOC). The location could be underneath the left or right side of the Operator Workspace table, or on the outside left or right side of the table.

GOC Cabinet

84

SEISMIC ANCHORING (IF REQUIRED)

If seismic anchoring is required, refer to site architectural drawings for fastener type details or applicable Pre-installation manual. Illustration below indicates hole location for installing brackets. 4.13 (105) NOTE: D ALL DIMENSIONS ARE IN INCHES. ALL BRACKETED () DIMENSIONS ARE IN MILLIMETERS. CENTERLINE OF FRAME

2.90 (73.5)

SEISMIC ANCHOR HOLE LOCATION

29.82 (757.5)

FRONT

1.67 (42.5)

3.35 (85)

Seismic Brackets

8-4

OPERATOR WORKSPACE

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

85

COMPUTER CABINET CABLE ROUTING AND CONNECTIONS

The GOC (Computer) Cabinet is shipped with seven cables connected to the computer that will be routed and connected to equipment that will be installed on the top of the OW Table. The cables are: D Three power cords D Two video ouput cables D Keyboard output cable D SCSI output cable These cables are already connected to the computer and routed thru the opening located at the top of the rear of the GOC cabinet as shown in the illustration below. The remainder of the cables to be connected to designated modules within the GOC Cabinet will have to be routed thru the lower opening. 1. 2. 3. 4. Runs 1063, 1064, 1065, 1068, 1069, 1070, and 1256 are preconnected to computer and are routed thru top opening. Runs 1077, 1081, 1082, 1083, 1085, 1255, power cord for the Patient Alert system, and any other option cables will be routed thru the lower opening. Route all cables before connecting any cables. Refer to cable map for connection information. If GOC Cabinet is to be located under OW Table, the cable coming out of the top opening of the cabinet will be routed above the table skirt. The cables coming out of the lower opening will be routed below the tabel skirt. 6 5

4 2 3 Table Skirt

The Computer Cabinet covers need to be removed to access the cable connection points. 5. 6. 7. For right side cover, remove three screws at rear of cover, slide toward rear of cabinet and then lift up. For left side cover, remove four screws at rear of cover, slide toward rear of cabinet and then lift up. Loosen four screws holding the back cover. Lift and remove cover.

8-5

OPERATOR WORKSPACE

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

86

ROUTE CABLES FOR LOWER OPENING

Refer to cable map for information on cables to be connected. Route cables thru clamp at lower rear of cabinet and connect cable runs as shown in illustration below. NOTE: The USB to Serial Converter that is included in the GOC Cabinet is NOT USED in this system. 1. 2. 3. 4. 5. 6. 7. 8. Run 1077 to LAN Hub. Run 1085 to Audio Asm. Run 1255 routes from GOC Cabinet thru clamp to J26 on RFS Cabinet. Make sure Modem power cable is connected to J5 on Computer Cabinet PDU. Run 1081 to PDU J1. Patient Alert Power cable to power strip at J12. Refer to Section 87 for other power cable connections to power strip. Runs 1082 and 1083 to Linux PC computer. Refer to Section 88 for all cables connected to Linux PC computer. Tighten clamp to secure cables in place.

Modem (Optional)

1 7 DASM (Optional) 6 5 8

4 Cable Clamp

8-6

OPERATOR WORKSPACE

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

87

POWER STRIP CONNECTIONS

The illustration below show the power cable connection locations to the power strip in the GOC cabinet. Make sure power cables are attached as shown.

J1

J7

POSITION J1 J2 J3 J4 J5 J6 J7 J8 J9 J10 J11 J12 TOTAL

NAME Host Host LCD Fan

MAX AMP CIRCUIT 1 CIRCUIT 2 6.4 0.8 0.2

J2

J8

J3

J9

Hub 2nd LCD DASM SCSI Tower Patient Alert 7.4

0.2 0.8 0.6 7.0 0.2 8.8

J4

J10

J5

J11

J6

J12

8-7

OPERATOR WORKSPACE

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

88

LINUX PC CABLE CONNECTIONS

The illustration below shows the connection points of cables routed to the Linux PC. All cables shown are preconnected except for Runs 1082 and 1083, which have to be routed and connected at site.

Run 1071 Power Cable

Run 1061 Mouse/Keyboard Cable (from WIM J8) Run 1055 USB Service Key Cable

Run 1066 Global Modem RS232 Cable (Option) Run 1063 Host Monitor Cable Run 1068 LCD Monitor Cable Run 1067 DASM Cable (Option) Run 1082 Ethernet Cable (from System Cabinet J63) Run 1064 SCSI Tower Cable Run 1083 Bit3 Cable (from System Cabinet J22) Run 1059 Lan Hub Cable

8-8

OPERATOR WORKSPACE

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

89

FINAL POSITIONING OF GOC CABINET

Complete positioning of GOC Cabinet. 1. 2. 3. 4. Reattach rear cover and tighten screws. Install both side covers by inserting tabs at bottom of covers into slots (as indicated in illustration below) and slide forward. Install screws at rear of cover that were removed in earlier procedure. Route remaining cables from top opening at rear of GOC Cabinet thru opening of table skirt. Place cabinet in final position as determined by customer. 2 3

8-9

OPERATOR WORKSPACE

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

810 INSTALL SCIM/KEYBOARD The ScanControl Intercom Module (SCIM) is the current keyboard on all Signa LX and OpenSpeed systems. Consult with Site customer for exact location of SCIM/Keyboard and Monitor on Operator Workspace (OW) Table. 1. Remove the SCIM, Data Cable, and Keyboard From the box.

SCIM Module 2. 3.

Data Cable

Keyboard

Turn the SCIM unit up side down exposing the cable connections. Route data cable Run 1262, preconnected to the GOC Cabinet and is located on back side of cabinet, from Computer Cabinet to bottom side of SCIM Module. Route thru data cable strain relief and connect as marked. (Run 805, supplied with SCIM Kit (2307175) is not used for EXCITE) Attach Run 1262 cable marked for mouse to the Mouse. Attach Run 1262 cable marked for keyboard and attach to the Keyboard. Data Cable Strain Relief 3 SCIM Module Keyboard

4. 5.

2 5

Data Cable Connector 6. 7.

Center Opening

Place keyboard in front of SCIM, pull the keyboard forward, flip it, and place on top of the SCIM up side down. Remove the plastic strips from the velcro located on SCIM plate, exposing the adhesive surface (see illustration on next page).

8-10

OPERATOR WORKSPACE

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

810 INSTALL SCIM/KEYBOARD (Continued) 8. 9. Guide the cable back into the center notched opening until all the cable is under the SCIM base. Place the keyboard tight against the SCIM, Center it as best as possible, and drop into place. The Velcro pads will now be sticking to the bottom of the keyboard. 8 9

Note The adhesive on the velcro takes 24 hours to completely adhere to the bottom of the keyboard. Allow the adhesive to dry completely before removing keyboard from SCIM base.

811 INSTALL MONITOR 1. 2. Place Host LCD Monitor on table in back of SCIM. Route Run 1063 from GOC Cabinet and connect to monitor. Route power cable Run1065 from Computer Cabinet and connect to monitor. Power cable must be routed and secured thru tywrap located on the under side of table top.

Host LCD Monitor 2 1

SCIM Module Keyboard

8-11


OPERATOR WORKSPACE

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

812 INSTALL SCSI TOWER AND LCD DISPLAY 1. Place the SCSI Tower on the OW table and connect Runs 1064 and 1070 from GOC Cabinet. Install terminator on back of SCSI Tower. Power cable Run 1070 must be routed and secured thru tywrap located on the under side of table top. Place LCD Display (Waveform Monitor) on OW table and connect Runs 1068 and 1069 from GOC Cabinet. Power cable Run 1069 must be routed and secured thru tywrap located on the under side of table top.

2.

1 2

8-12

OPERATOR WORKSPACE

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

SECTION 9 COMPLETION CHECKLIST


TABLE OF CONTENTS
SECTION 91 911 912 92 921 922 923 PAGE RETURNING SHIPPING MATERIAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92 FINAL MECHANICAL INSTALLATION STEPS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92 PHANTOMS AND SERVICE KITS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92 FINAL GE FIELD ENGINEER INSTALLATION STEPS . . . . . . . . . . . . . . . . . . . . . . . . . . 93 SUPPLIER PERFORMANCE REPORT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95 MECHANICAL INSTALLATION COMPLETION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92

GE FIELD ENGINEER RESPONSIBILITIES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92

9-1

INSTALLATION COMPLETION CHECKLIST

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

91 MECHANICAL INSTALLATION COMPLETION The following steps need to be reviewed and completed by GE Field Engineer, especially if the system was installed by nonGE personel. Check off boxes at the front of each item below when complete. - Check mechanical flowchart to make sure all steps are complete - Resolve shipment shortages - Resolve ommissions by mechanical contractors - Make sure all rating plates are installed - Recheck all wiring connections using cable map - Correct for wiring errors if necessary - Complete Section 911. - Complete Section 912. 911 RETURNING SHIPPING MATERIAL The shipping material must be returned as soon as possible. Prompt return is required so that production shipments can continue with these labor saving dollies and carts. These can be included with the Return pile, Recycling Operation will route them to appropriate location. 912 FINAL MECHANICAL INSTALLATION STEPS 1. 2. 3. Complete assembly of PPG cable and probe head. Make sure all Cabinet, Magnet Enclosure, and Rear Pedestal covers have been installed. Make sure Patient Transport is in position at front of Magnet.

92 GE FIELD ENGINEER RESPONSIBILITIES 921 PHANTOMS AND SERVICE KITS 1. 2. 3. Locate Phantom Kit and SPT Phantom Cart Kit. Unpack the phantoms carefully to reduce risk of damage, breakage, or chemical spill. Locate the multilingual (provided in eight languages) phantom attention labels shipped with each of the phantoms.

CAUTION
Three types of multilingual phantom attention labels are provided and each type must be applied to the correct phantom. The directions at the top of each label sheet describe which phantoms that can receive that label.

9-2

INSTALLATION COMPLETION CHECKLIST

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

9-2-1 PHANTOMS AND SERVICE KITS (Continued) 4. Read the directions provided at the top of each of the multilingual attention label sheets and apply the appropriate language attention label for the site to each of the phantoms that should receive that type of label. a. Each phantom will receive the correct type of attention label written in the appropriate language for that site. b. Note that the LVshim phantom assembly will receive 3 labels, one for each phantom component. 5. 6. Place service phantoms into SPT Phantom Cart or service storage cabinet to be organized and ready to be used during applicable checks and calibrations. The DQA Phantom is to be stored in a customer designated location. Locate TPS RF I/F Kit (if applicable) and Signa Spares Kit. Place kits into service storage cabinet for ready use.

922 FINAL GE FIELD ENGINEER INSTALLATION STEPS 1. 2. Record and enter applicable data into applicable site configuration files and records. Complete Product Locator information for all installed serialized components, new or updated, via either: D (U.S. Only) FE Site Verification Web Site or D Process and return product locater installation cards for all serialized components to: Product Locator File, P Box 414, W-523, Milwaukee, WI 53201-0414 .O. Refer to Section 18 for details on submitting Product Locator information. Note Failure to fill out and return Product Locator Cards may result in failure of your site to receive future FMIs. 3. 4. 5. Process the Supplier Performance Report located in Common Forms and Section 923 (reference copy). Store the delivered sites set of service tools and spares kit in service cabinet at site. Locate surface coil (quick disconnect) adapters and any optional surface coils and positioning accessories and place on Patient Transport. Be sure only the correct polarity (normal or reverse but not both) head coil adapter is left at site. Locate Box To be opened by Applications Specialist and set aside for later visit by Applications Specialist. Leave the sites set of Manuals and CDROM at site. Organize and set up reference cabinet for the Manuals. Locate Material Safety Data Sheets (MSDS) packed with phantoms and gradient coil coolant. They must be retained on site. Customer is to be informed that material with MSDS was brought into site and customer should know/decide where MSDS should be retained at site. Verify that coordination of application support for instruction of site users on the appropriate Software Release has been made before returning system to the users. Turn site over to applications who will instruct users.

6. 7. 8.

9.

9-3

INSTALLATION COMPLETION CHECKLIST

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

This page left blank intentionally

9-4

INSTALLATION COMPLETION CHECKLIST

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

923 SUPPLIER PERFORMANCE REPORT GE Medical Systems Field Service Send completed forms to the 4 addresses at the end of this form. Supplier Name: Work Performed / Evaluated: Commodity / Item Purchased: GE Customer Name: FDO Number: GEMS Evaluator Name: Please rate suppliers as 1 to 5 on each of the 15 questions below: 1 = Unacceptable 2 = Poor 3 = Average 4 = Excellent 5 = Outstanding There are 5 groups with three related questions in each group. Groups can be rated either as a group or 3 separate questions. Any rating below 3 triggers a discrepancy review with the supplier. Corrective actions will be faster and more effective when you provide examples and/or detail information.
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Supplier schedules well and: starts on time? finishes on time? avoids disruptions? Supplier is flexible and: responds to GEMS/customer requests? cooperates with changes in schedule? deals with unexpected problems? Supplier tools and equipment are: appropriate for contracted work? adequate for contracted work? calibrated and in good working order? Supplier meets GEMS quality standards with employees who: are trained, skilled and technically competent? always keep job sites clean and free of debris? ensure both function and appearance are as expected? Supplier employees enhance GEMS image by their: appearance? behavior? attitude? 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 5 5 5 5 5 5 5 5 5 5 5 5 5 5 5

Date:

Test Equipment Barcode No;

Examples / Details:

Any rating below 3 triggers a discrepancy review with the supplier. SEND COMPLETED FORMS TO: 1. 2. 3. 4. Senior Operations Specialist for your LCT The contractor who performed the work ISS (already in default distribution) Sourcing Supplier Quality (already in default distribution)

9-5

INSTALLATION COMPLETION CHECKLIST

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

SECTION 10 APPENDIX
TABLE OF CONTENTS
SECTION PAGE 101 SYSTEM CONFIGURATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102 1011 BASIC SYSTEM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102 1012 SYSTEM COOLING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105 102 CABLE IDENTIFICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107 1021 PDU POWER AND GROUND CABLES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107 1022 MAGNET ROOM CABLES NOT CUT TO LENGTH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108 1023 MAGNET MONITORING CABLES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1010 1024 GE MAGNET CABLES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1011 1025 FIBER OPTIC CABLES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1011 1026 EQUIPMENT ROOM & MAGNET ROOM CABLES & HOSES CUT TO LENGTH . . . 1012 1027 EQUIPMENT AND OPERATOR ROOM CABLES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1013 1028 CABLE SPARES (FRUS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1014 1029 EXCITE 4KW AND 8KW MNS CABLES (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1015 103 FIBER OPTIC TERMINATION INSTRUCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1016 1031 FIBER OPTIC CABLE TERMINATION KIT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1016 1032 STRIPPING FIBER OPTIC OUTER JACKET . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1017 1033 CRIMPING CONNECTOR TO FIBER OPTIC CABLE . . . . . . . . . . . . . . . . . . . . . . . . . . . 1017 1034 TRIMMING EXCESS FIBER FROM CONNECTOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1018 1035 POLISHING FIBER OPTIC END . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1018

10-1

APPENDIX

GE MEDICAL SYSTEMS REV 1 101 SYSTEM CONFIGURATION

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

The tables in this section list the catalogs for the hardware equipment which comprise the Signa EXCITE HD 1.5T TwinSpeed system. 1011 BASIC SYSTEM The basic system is configured from several catalogs. The following tables indicate what catalog(s) has been installed for this site. For those tables which offer more than one catalog, select the catalog which has been delivered to the site. TABLE 101 SIGNA EXCITE HD NEW SYSTEM CATALOGS
8 channel TwinSpeed Reflex 400 (M3000TE) Fixed Site Twin Fixed Site Collector (M3043PF) Ceiling Height Kit (M1060SR) 1.5T LCC Magnet with Twin Enclosure and TRM Body Coil Note For Japan (M3060TC) ONly: Air Cooled Shield/Cryo Cooler Compressor Shield/Cryo [AKA Coldhead Compressor] Cooler (M1060JW) Compressor procured locally. Coldhead Chiller: Customer provided water Indoor/Outdoor Air Cooled Chiller cooling required for (M3088TL) Coldhead Compressor 1.5T LCC Magnet with Twin Enclosure and TRM Body Coil (M3060TC) Shield/Cryo Cooler Compressor [AKA Coldhead Compressor] (M1060JW) Transportable Relocatable Coldhead Chiller: Indoor/Outdoor Air Cooled Chiller (M3088TL) Gradient Coil Chiller: Indoor/Outdoor Air Cooled Chiller (M3088TL) TwinSpeed Main Disconnect Panel 400/480 VAC, 60Hz designed for 2 chillers (M3088TM) Chilled water for Coldhead Compressor and Gradient Coil provided by Mobile equipment provider. Main Disconnect Panel provided by Mobile equipment provider. 1.5T LCC Magnet with Twin Enclosure and TRM Body Coil (M3061TC) Shield/Cryo Cooler Compressor [AKA Coldhead Compressor] (M1060JW) 16 channel TwinSpeed Reflex 800 (M3000Tg) T/R Twin T/R Site Collector (M3063PF) True Mobile Twin Mobile Site Collector (M3063PF)

Gradient Coil Chiller: Indoor/Outdoor Air Cooled Chiller (M3088TL)

Gradient Coil Chiller: H20 Single Heat Exchanger (M3088TK) Requires facility chilled water Customer provided MDP per GE specifications

Chilled water for Coldhead Compressor and Gradient Coil provided by Transporatable equipment provider. Main Disconnect Panel provided by Transporatable equipment provider.

TwinSpeed Main Disconnect Panel 400/480 VAC, 60Hz designed for 2 chillers (M3088TM) Signa Patient Table (M1090TL) Signa Patient Table (M1090TL) 1.5T Fixed System Cables Config A (M3143PE) 1.5T Fixed System Cables Config B (M3144PE) Wide OW Table M1000MW

Signa Patient Table (M1090TL) 1.5T Fixed System Cables Config C (M3145PE) 1.5T T/R System Cables Config B (M3163PE)

Signa Patient Table (M1090TL) 1.5T Mobile System Cables Config B (M3153PE)

10-2

APPENDIX

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

1011 BASIC SYSTEM (Continued)

A
YES 1.5T MNS Amp/Hardware (M3100LA) Carbon TR Switch (M3100LB) Phosphorous TR Switch (M3100LC) MNS? NO

Carbon TR Switch (M3100LB)

Phosphorous TR Switch (M3100LC)

1 English Warning Sign Kit included in Magnet catalog

English Spanish Chinese Italian Portugues French German Swedish Japanese Warning Warning Warning Warning e Warning Warning Warning Warning Warning Sign Kit Sign Kit Sign Kit Sign Kit Sign Kit Sign Kit Sign Kit Sign Kit Sign Kit (M2000XA) (M2000XB) (M2000XC) (M2000XD) (M2000XE) (M2000XF) (M2000XG) (M2000XH) (M2000XJ)

MR Opterator Chair E8803BE LCD Monitor (18 inch) M1000NT

English Keyboard M1000MN

French Keyboard M1000MP

German Keyboard M1000MR

Scandinavian Keyboard M1000MS

Italian Keyboard M1000NH

Portuguese Keyboard M1000NJ

Spanish Keyboard M1000NK

Broadband Connectivity E1000LG

IIP Global Modem M1000NW

Digital DASM M1000MK

DICOM Print will be used exclusively for software filming to DICOM print peripheral devices

Oxygen Monitor M1060KM

END

10-3

APPENDIX

GE MEDICAL SYSTEMS REV 1 1011 BASIC SYSTEM (Continued)

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

MAGNET MONITOR (MSM)

BLOWER BOX (MG6) SPT PHANTOM SET STORAGE CART NOTE: STORAGE CART PENETRATION NOT USED WITH TRUE SHIELD COOLER PANEL MOBILE SITE SYSTEM COMPRESSOR CABINET (PP1) (MS5) MAGNET ENCLOSURE (MG2) LIGHTWEIGHT PATIENT TRANSPORT (PT1) GE MDP (M3088TM) Refer to catalog flowchart to determine if required or optional

HFD/PDU CABINET (MR3)

RFS CABINET (MR2)

TWIN ACCESSORY CABINET (TAC)

OPERATOR WORKSPACE (OW1)

Gradient Coil Chiller Option or Coldhead Chiller Option Indoor/Outdoor Air Cooled Chiller (M3088TL) Refer to catalog flowchart to determine if required or optional

Gradient Water Heat Exchanger Option Indoor Air Cooled Chiller (M3088TK)

REMOTE CONTROL PANEL (RCP) MR COMMON CHILLER (MRCC) (GWHX) 1.5T MNS AMP CABINET (MR10)

SIGNA EXCITE 3.0T SYSTEM ILLUSTRATION 101

10-4

MAIN DISCONNECT PANEL (MDP) MNS Option

APPENDIX

GE MEDICAL SYSTEMS REV 1 1012 SYSTEM COOLING

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

The Signa EXCITE HD 1.5T TwinSpeed system requires water cooling for the Magnet Shield/Cryo Cooler Compressor Cabinet and the TRM Gradient Coil located in the Magnet bore. The following system cooling equipment is available with the MR system: Note See Illustration 101 flowchart for system cooling equipment configurations options and relationship to other system equipment/options selections. For water cooling specifications and requirements for Site Provided Water Cooling, refer to Direction 5133302, Signa EXCITE HD 1.5T TwinSpeed PreInstallation manual. D MR Common Chiller (MRCC) The MRCC is a singleloop 10KW 50/60 Hz water chiller that can be used to cool either Shield/Cryo Cooler Compressor or the TwinSpeed Gradient Coil. The MRCC can be installed either indoors or outdoors. There is a Remote Panel to be installed in the MR system Equipment Room that can stop or restart the chiller and can display supply water temperature, chiller setpoints and alarms. The MRCC power is supplied by the MR system Main Disconnect Panel (MDP). This chiller will be serviced by a third party vendor. D Gradient Water Heat Exchanger (GWHX) The GWHX is a 10KW 50/60 Hz watertowater heat exchanger that is used for cooling the TwinSpeed Gradient Coil. The GWHX is for indoor use only. The GWHX requires facility provided chilled water. The GWHX power is supplied by the MR system Power Distribution Unit (PDU). This heat exchanger will be serviced by GE Service. When the GWHX is used then the customer must supply water cooling for the Shield/Cryo Cooler Compressor. Note Gradient Coil water cooling must be supplied by cooling equipment (MRCC or GWHX) supplied with the MR system to prevent contamination/damage to the coil and for proper image quality. There are several possible scenarios which a customer may utilize to meet the system water cooling requirements for the LCC Magnet Shield/Cryo Cooler Compressor Cabinet and the TRM Gradient Coil. D Two MRCC installed outdoors where one unit provides Shield/Cryo Cooler Compressor cooling and the other provides Gradient Coil cooling. The 2 MRCC Remote Control Panels to be installed in the MR system Equipment Room. D Two MRCC installed indoors where one unit provides Shield/Cryo Cooler Compressor cooling and the other provides Gradient Coil cooling. The 2 MRCC Remote Control Panels to be installed in the MR system Equipment Room. Customer supplied air cooling is required for the MRCC units. D One GWHX installed indoors provides water cooling to the Gradient Coil (GWHX requires facility water cooling) and Customer provided facility water cooling for the LCC Magnet Shield/Cryo Cooler Compressor Cabinet. D One MRCC installed outdoors provides water cooling to the Gradient Coil and Customer provided facility water cooling for the LCC Magnet Shield/Cryo Cooler Compressor Cabinet. The MRCC Remote Control Panel to be installed in the MR system Equipment Room. D One MRCC installed indoors provides water cooling to the Gradient Coil and Customer provided facility water cooling for the LCC Magnet Shield/Cryo Cooler Compressor Cabinet. The MRCC Remote Control Panel to be installed in the MR system Equipment Room. Customer supplied air cooling is required for the MRCC unit.

10-5

APPENDIX

GE MEDICAL SYSTEMS REV 1 1012 SYSTEM COOLING (Continued)

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

TABLE 102 TWIN SYSTEM COOLING CABINET (SEE TABLE NOTE 1)


CATALOG DESCRIPTION

M3088TK M3088TL

Gradient Water Heat Exchanger (GWHX) See Note 2 MR Common Chiller (MRCC) See Note 3

Note 1 See llustration NO TAG to determine system cooling equipment configurations that must order and install the GE MDP. 2 When GWHX is selected, customer supplied water cooling is required for the LCC Magnet Shield/Cryo Cooler Compressor. Refer to Direction 2355548, Signa EXCITE 3.0T PreInstallation for water cooling specifications. 3 Two M3088TL installed outdoors is standard cooling equipment configuration of Transportable/Relocatable Mobile systems. Note True Mobile system water cooling for the Gradient Coil and the Shield/Cryo Cooler Compressor is provided by the Mobile Van equipment.

GWHX (M3088TK)

MRCC (M3088TL)

REMOTE CONTROL PANEL (RCP)

SYSTEM COOLING EQUIPMENT SELECTIONS ILLUSTRATION 102

10-6

APPENDIX

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

102

CABLE IDENTIFICATION

Cables delivered to site for installation have specifications dependent on its usage. The following tables contain information on cable runs for this system as it was originally installed. The table below provides definitions for the following cable tables.
COLUMN HEADING RUN FROM & TO DESCRIPTION CABLE TYPE VOLTAGE RATING ACTUAL VOLTAGE TEMP RATING FLAME RATING REMARKS DESCRIPTION The run number as seen on the delivered cable Labels the destination of the cable Could include information on length, number of conductors, type of cable, and end plug type Type designation for the cable The highest voltage that may be continously applied to a wire Nominal for continuous; maximum for variable voltage Maximum temperature insulating material my be used without loss of basic properties Measure of materials ability to support combustion Describe the cables use in the MR system Specific to the core cable manufacturer given. If voltage varies during use, the maximum is given. Specific to the core cable manufacturer given. Specific to the core cable manufacturer given. Total length of cable delivered to site, type of cable specifies power from data. SPECIAL NOTES The vendor cable may not have run numbers.

1021 PDU POWER AND GROUND CABLES The cables listed in Table 103 are connected at one end to the PDU module in the ACGD/PDU Cabinet. TABLE 103 PDU POWER AND GROUND CABLES
VOLTAGE RATING ACTUAL VOLTAGE TEMP RATING CABLE TYPE FLAME RATING

RUN

FROM

TO

DESCRIPTION

REMARKS

037 040 045 968 988 989 1081

PD1 GND MSI PP1 GND MR1 A12 J1 PP1 A7 GND MR3 PD1 OW1 A19 J1 PP1 PHPS PWR MR2 RF AMP MR3 PD1 PD1 MAG/SHIM OUTLET

MR2 A12 GND RF COM GND MG6 GND MR3 PD1 TAC GND TAC J8 MR3 PD1 MR3 PD1 MR3 PD1 RF COM GND MAG/SHIM POWER SUPPLY

36 ft Ground Wire (#10) 69 ft Ground Wire (#1/0) 41 ft Ground Wire (#10) 35 ft Power (5#8) 60 ft. Ground (#8) 25 ft. Power (3#10) 65.5 ft Power (4#10) 61 ft. Power (3#10) 35.5 ft Power (4#8) 69 ft Ground Wire (#1/0) 50 ft Power (5#8) With 30A Plug and Receptacle

SO SO SO SO SO SO SO AWM SO SO

600 600 600 600 600 600 600 600 600 600

0 208Y 0 208Y 0 120 120 120 208 208Y

105 105 105 75 80 90 75 90 90 105

VW1 VW1 VW1 FT4 VW1 FT4 FT4 FT4 FT4 VW1

System Cabinet ground Magnet Ground Cable Blower Cabinet Separate ground RFS Cabinet Power TAC to PP1 ground (3) TAC Power Operator Workspace Power PHPS Module Power RF Amplifier Power Magnet Ground Cable Cable for PDU to GE Magnet Power Supply or SC Shim Supply Power

1182 1234
1254 N/A

10-7

APPENDIX

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

1022 MAGNET ROOM CABLES NOT CUT TO LENGTH The cables listed in Table 104 are routed and connected in the Magnet Room and are not cut to length. TABLE 104 MAGNET ROOM RUNS (NOT CUT TO LENGTH)
VOLTAGE RATING ACTUAL VOLTAGE TEMP RATING CABLE TYPE FLAME RATING

RUN

FROM

TO

DESCRIPTION

REMARKS

282 297 300 318 385 387 414 421 422 458 715 716 743 744 778 779 780 781 782 784

PP1 J12 PP1 J18 PP1 J15 PP1 J11 PP1 J20 PP1 J47 MG3 A35 MG2 A33 J1 MG2 A33 J2 PP1 J17 PP1 J14 PP1 J89 MG2 A12 A1 J2 MG2 A12 A1 J1 PP1 A11 J72 PP1 A11 J73 PP1 A11 J74 PP1 A11 J75 PP1 A11 J76 PP1 F5, F6

MG3 A2 J6 EO1 MG2 A15 MG3 A7 MG2 A30 MG2 A29 J1 MG3 A2 J9 MG3 A31 J1 MG3 A1 J4 OM3 MG2 A33 J1 MG2 A11 A1 J2 MG3 A3 A1 J3 MG3 A3 A1 J4 MG2 A12 A2 J3 MG2 A12 A2 J4 MG3 A3 A3 J6 MG2 A12 A2 J5 MG2 A12 A2 J6 MG6 J1

45 ft. 25 pin SubD 90 ft. 9 pin SubD 60 ft. 25 pin SubD 45 ft. 9 pin SubD 67 ft. 37 pin SubD 60 ft. 25 pin SubD 30 ft. 9pin SubD 15 ft. 15 pin SubD 25 ft. 37 pin SubD 9 pin SubD 65 ft. 37 pin SubD 65 ft. 15 pin SubD 10 ft Coax w/M&F N 10 ft Coax w/M&F N 70 ft. MHV Coax 70 ft. MHV Coax 60 ft. MHV Coax 60 ft. MHV Coax 60 ft. MHV Coax 40 ft. 2 cond power cable

CI3 CI3 CI3 CI3 CI3 CI3 CI3 CI3 CI3 CI3 CI3 CI3 STO STO CI2 CI2 CI2 CI2 CI2 SO

300 300 300 300 300 300 300 300 300 300 300 300 2500 VDC 2500 VDC 1900 VRMS 1900 VRMS 1900 VRMS 1900 VRMS 1900 VRMS 600

<30 <30 <30 38.5 VDC <30 38.5 VDC <30 <30 <30 <30 <30 <30 950 950 1000 VDC 1000 VDC 1000 VDC 1000 VDC 1000 VDC 120

80 80 80 80 80 80 80 80 80 80 80 80 85 85 80 80 80 80 80 80

FT4 FT4 FT4 FT4 FT4 FT4 FT4 FT4 FT4 FT4 FT4 FT4 VW1 VW1 VW1 VW1 VW1 VW1 VW1 FT4

Patient Intercom Emergency off switch Bore Light power Long Drive Motor power Patient Alignment Light power Table Elevation motor power Front Cover Microphone Dock Limit Long Drive Encoder/Limit Sw Optional Remote O2 Sensor SRI Power PAC Power Body Coil Drive Body Coil Drive Dynamic Disable (body coil frnt) Dynamic Disable (body coil frnt) Inductive Drive Dynamic Disable (body coil rear) Dynamic Disable (body coil rear) Patient Cooling Blower power

10-8

APPENDIX

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

1022 MAGNET ROOM CABLES NOT CUT TO LENGTH (Continued) TABLE 104 (CONTD) MAGNET ROOM RUNS (NOT CUT TO LENGTH)
VOLTAGE RATING ACTUAL VOLTAGE TEMP RATING CABLE TYPE FLAME RATING

RUN

FROM

TO

DESCRIPTION

REMARKS

848 862 863 893 897 1152 1197 1200 1201 1202 1203 1215 1216 1217 1250 1260 1261

MG2 A11 J4 MG2 A4 J1 MG2 A4 J2 MG3 A3 J2 MG2 A4 J3 PP1 A5 J1 PP1 RFTXSW J150 PP1 J200 PP1 J201 PP1 J202 PP1 MC SW FLTR J203 MG3 16CHSW J20 PP1 A16 J77 PP1 A15 J78 MG3 A3 J1 MG2 A33 J10 MG2 A4 J4

MG2 A4 J5 MG2 A5 J1 MG2 A5 J2 MG3 A3 A5 INPUT MG2 A33 J4 MG2 A12 A3 P1 MG3 16CHSW J25 MG3 16CHSW J1 MG3 16CHSW J2 MG3 A17 J3 MG3 16CHSW J11 MG2 SRI J20 MG3 A36 J5 MG3 A36 J10 CONNECT TO RUN 257 MG2 A36 MG2 A33 J5

16 ft. 25 pin SubD/rt. angl 2 ft. 26 pin Rt. angle 2 ft. 26 pin Rt. angle 6.6 Inch BNC 20 ft. 26 pin Amplite to 37 pin SubD 29.5 ft. 37 pin CPC to 16 pin 49 ft. 9 pin SubD 8Coax 8W8 Combo D 8Coax 8W8 Combo D 2Coax 8W8 Combo D 49 ft. 37 pin SubD 37 pin SubD 80 ft. 15 SubD 80 ft. 15 SubD

CI3 CI2 CI2

300 300 300 1900 VRMS

<30 <30 <30 <30 <30

80 80 80 80 80 105

FT4 VW1 VW1 VW1 FT4 FT4 FT4 FT4 FT4 FT4 FT4 FT4 FT4 FT4

Operator Display to PAC II Oper Display to Rt. Op Cntrl Oper Display to Lt. Op Cntrl Spliiter to Preamp Cable Dispay to SRI Shim Filter Cable

CI3 PLTC CI3 CATV R CATV R CATV R CI3 CI3 CI3 CI3

300 300 300 1500 1500 1500 300 300 300 300

<30 <30 <30 <30 <30 <30 <30 <30

80 85 85 85 80 80 80 80

Monitor to Compressor I/F SRI Power

35 ft 3 cond w/ 9Sub D 20 ft. Data Cable (80#28)

CI3 CI2

300 300

<30 <30

105 80

FT4 VW1

Bore Temperature Sensor SRI Splitter to Oper Display

10-9

APPENDIX

GE MEDICAL SYSTEMS REV 1 1023 MAGNET MONITORING CABLES

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

The Magnet Monitoring cables listed in Table 105 are delivered with the Magnet. TABLE 105 MAGNET MONITORING CABLES DELIVERED WITH MAGNET
VOLTAGE RATING ACTUAL VOLTAGE TEMP RATING CABLE TYPE FLAME RATING

RUN

FROM

TO

DESCRIPTION

REMARKS

823 824 825 826 827 828 829 830 831 832 833 939 940 1077

MSM1 A1 J10 PP1 J10 PP1 J48 MSM1 A1 J9 FJ3 PP1 J10 PP1 J48 MS1 FJ2

MR2 A11 J24 FJ1 MSM1 A1 J8 FJ3 MS5 A5 A1 J1 / FJ4 MA1 A3 A1 P403 MS1 FJ2 MS1 A2 J2/ MS1 A5 A2 J1/ MG3 TB1 MSM1 A1 J7 MS1 SLEEVE/ MS1 COLDHEAD MS5 A1 A6 JR MSM4 MSM1 J1 OW1 A2 A4 PRT7 MR2 IPG J9/ MR2 PDU J5

Data Cable (15#22) Data Cable (9#22) Data Cable (25#22) Data Cable (15#22) Data Cable (8#22) & Data Cable (4#22) Data Cable (9#22) Data Cable (25#22) 3cond/3cond/25cond

CI3 CI3 CI3 CI3 CI3 CI3 CI3 CI3

300 300 300 300 300 300 300 300

<30 <30 <30 <30 <30 <30 <30 <30

80 80 80 80 80 80 80 80

FT4 FT4 FT4 FT4 FT4 FT4 FT4 FT4

Monitor to System Cabinet I/F Monitor Sensor I/F to Penetration Panel Magnet Monitor connector to Penetration Panel Monitor to Compressor I/F Compressor I/F Penetration Panel to Sensor Penetration Panel to Magnet I/F Magnet I/F

FJ1 MS1 A2 J1 FJ4 MDP UPS MSM3 RS232 MSM1 J5 MR2 A11 J24

Data Cable (8#24) Data Cable (8#22) Data Cable (6#24) 6 ft. Power (5#15) 6 ft. Data Cable (9#22) 80 ft. Cat 5 Ethernet

CI3 CM CM SO CI3 CI3

300 300 300 600 300 300

<30 <30 <30 120 <30 <30

80 80 80 85 80 80

FT4 FT4 FT4 VW1 FT4 FT4

Monitor to Sensor I/F Buffer Amp to Cooling Sleeves Compressor to Compressor I/F MDP to UPS Monitor to Modem Monitor to OW Ethernet System Cabinet I/F

10-10

APPENDIX

GE MEDICAL SYSTEMS REV 1 1024 GE MAGNET CABLES

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

The Magnet cables listed in Table 106 arrive on site with the magnet and are installed at the time of Magnet delivery. TABLE 106 GE MAGNET CABLES
VOLTAGE RATING ACTUAL VOLTAGE TEMP RATING CABLE TYPE FLAME RATING

RUN

FROM

TO

DESCRIPTION

REMARKS

601 602 603 604 606 609 621 622 623 624

MS8 MS8 MS7 MS7 or MS8 MS4 MS7 OR MS8 MS5 A1 MS5 A1 MS5 A1 MS1 A2

MS1 MS1 MS1 A3 A1 MS1 A3 A1 MS1 A3 A1 PDU MS1 A2 MS1 A2 PP1 F1, F2, F3, F4 PP1 F1, F2, F3, F4

Power Cable (#4/0) Power Cable (#4/0) Power Cable (#4/0) Data Cable (12#22) Data Cable (4#22) Power Cable (5#16) 7/8 Gas Flex Line 7/8 Gas Flex Line Power Cable (4#18) Power Cable (4#18)

SO SO SO CI3 CI3 SO STO STO SO SO

600 600 600 150 300 300 N/A N/A 600 600

208 208 208 60 VDC <30 208Y N/A N/A 200 200

80 80 80 80 105 105 85 85 105 105

VW1 VW1 VW1 FT4 FT4 FT4 VW1 VW1 VW1 VW1

Ramp cables, Power Assembly Positive 4/0 Ramp cables, Power Assembly Negative 4/0 Ramp and Shim Cable Assembly Switch Heaters Emergency Run Down Unit Feeder Cables Supply Flexible Gas Line (PP1 J60 or J61) Return Flexible Gas Line (PP1 J60 or J61) Compressor to Pen Panel Cable (equipment room) Cold Head Control Cable

1025 FIBER OPTIC CABLES The Fiber Optic cables listed in Table 107 are routed and connected in the Magnet Room and the Equipment Room. TABLE 107 FIBER OPTIC CABLES
VOLTAGE RATING ACTUAL VOLTAGE TEMP RATING FLAME RATING CABLE TYPE

RUN

FROM

TO

DESCRIPTION

REMARKS

711/ 712 1045 1083 1179

PP1J91

MG2 A33 and MG2 A11 A1

100 ft. 3/4 in. fiber optic conduit

N/A

N/A

85

N/A

One duplex fiber optic cables and one simplex fiber optic cable and One duplex plastic fiber optic cable for PACII Two duplex fiber optic and one simplex fiber optic cables Two fiber optic cables with two spares Four glass fiber optic cables

MR2 A11 J13 MR2 A11 J22 MR2 A11 J17

PP1 A17 A1 OW1 A15 XMT1 MR3 A11 J2

3/4 in. fiber optic conduit 72.2 ft. 3/8 in. jacketed cable

N/A N/A N/A

N/A N/A N/A

0 0 0

85 85 85

N/A N/A N/A

20 ft. 3/4 in. fiber optic conduit

10-11

APPENDIX

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

1026 EQUIPMENT ROOM AND MAGNET ROOM CABLES AND HOSES CUT TO LENGTH The Gradient cables and the RF Head and Body cables listed in Table 108 will be routed, cut, and terminated at designated cable ends according to instructions found in the CABLE INSTALLATION section of this manual. TABLE 108 EQUIPMENT ROOM AND MAGNET ROOM CABLES & HOSES CUT TO LENGTH
VOLTAGE RATING ACTUAL VOLTAGE CABLE TYPE TEMP RATING FLAME RATING

RUN

FROM

TO

DESCRIPTION

REMARKS

235 257 536 537 976 977 978 979 980 981 982 983 984 985 986 987 1198 1199 1238

CONNECT TO RUN 1259 PP1 J44 WC1 WC1 MR3 CABI/F J3 MR3 CABI/F J4 MR3 CABI/F J5 TAC X+,WB,ZM TAC Y+,WB,ZM TAC Z+,WB,ZM PP1 ZZM+,ZM PP1 YZM+,ZM PP1 XZM+,ZM PP1 ZWB+,WB PP1 YWB+,WB PP1 XWB+,WB PP1 RFTXSW J151 PP1 RX TX SW J153 PP1 J4

PP1 J44 CONNECT TO RUN 1250 MG2 A12 A1 MG2 A12 A1 TAC X TAC Y TAC Z PP1X+,WB,ZM PP1Y+,WB,ZM PP1Z+,WB,ZM MG2 A12 A1 7/8 MG2 A12 A1 9/10 MG2 A12 A1 11/12 MG2 A12 A1 1/2 MG2 A12 A1 3/4 MG2 A12 A1 5/6 MG3 PEDI/F J1 MG3 PEDI/F J2 CONNECT TO RUN 1249

Times Microwave LMR900FR Times Microwave LMR900FR Flexible Tubing Flexible Tubing Conductor (3#2) Conductor (3#2) Conductor (3#2) Conductor (3#2) Conductor (3#2) Conductor (3#2) Conductor (3#2) Conductor (3#2) Conductor (3#2) Conductor (3#2) Conductor (3#2) Conductor (3#2) Times Microwave LMR600FR Times Microwave LMR600FR Times Microwave LMR600FR

CATV R CATV R N/A N/A STO STO STO STO STO STO STO STO STO STO STO STO CATV R CATV R CATV R

5000 VDC 5000 VDC N/A N/A 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 4000 VDC 4000 VDC 6000 VDC N/A N/A 600 600 600 600 600 600 600 600 600 600 600 600

85 85 N/A N/A 90 90 90 90 90 90 90 90 90 90 90 90 85 85 85

FT4 FT4 N/A N/A FT4 FT4 FT4 FT4 FT4 FT4 FT4 FT4 FT4 FT4 FT4 FT4 FT4 FT4 FT4

1.5T Body Coil Run created from kit 1.5T Body Coil Transmit Gradient Coil Cooling Line Gradient Coil Cooling Line Twin X Gradient Power Twin Y Gradient Power Twin Z Gradient Power Twin X Gradient Filter Twin Y Gradient Filter Twin Z Gradient Filter Twin Zzm Gradient Output Twin Yzm Gradient Output Twin Xzm Gradient Output Twin ZWB Gradient Output Twin YWB Gradient Output Twin XWB Gradient Output RF Head Transmit RF Head Transmit RF Head Transmit

10-12

APPENDIX

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

1027 EQUIPMENT AND OPERATOR ROOM CABLES The cables listed in Table 109 are routed and connected between Penetration Pnael, cabinets in equipment room and Operator Workspace. TABLE 109 EQUIPMENT & OPERATOR ROOM RUNS
VOLTAGE RATING ACTUAL VOLTAGE CABLE TYPE TEMP RATING FLAME RATING

RUN

FROM

TO

DESCRIPTION

REMARKS

296 457 611 701 971 974 1056 1059 1060 1061 1063 1064 1065 1066 1067 1068 1069 1070 1071 1074 1075 1082 1085 1180 1181 1184 1185 1186 1187

FACILITY DISCNCT OM1 TAC J22 MR2 A11 J15 MR2 A11 J23 MR3 PD1 J5 OW1 A17 PRT6 OW1 A17 PRT2 OW1 A19 J3 OW1 A15 KYBD/ MS OW1 A15 Video 1 OW1 A15 SCSI 1 OW1 A11 J2 OW1 A15 OW1 A15 SCSI 2 OW1 A15 Video2 OW1 A11 J8 OW1 A11 J11 OW1 A11 J1 OW1 A19 J2 OW1 A19 J4 OW1 A15 ETH SLOT 2 OW1 A21 J7 MR3 J7 MR3 J8 MR2 A11 J125 MR2 A11 J200 MR2 A11 J201 MR2 A11 J202

PP1 J18 PP1 J17 PP1 SHIM RF Door switch MR3 PD1 J2

45 ft 9 pin SubD 9 pin SubD

CI3 CI3

300 300

<30 <30

80 80

FT4 FT4

Emergency off switch Optional O2 Monitor cable Resiitive Shim P.S. Option

70 ft 9 pin SubD

CI3 CI3 CI3 CI3 CI3 CI3 CI3 CI3 CI3 SVT CM CI3 CI3 SVT SVT SVT SVT SVT CI3 CI3 CI3 CI3 CI3 CATV R CATV E CATV R

300 300 300 30 30 300 300 300 300 300 300 30 300 300 300 300 600 600 300 300 300 300 300 1500 1500 1500

<30 <30 <30 <30 <30 <30 <30 <30 <30 120 <30 <30 <30 120 120 120 120 120 <30 <30 <30 <30 <30 <30 <30 <30

80 80 80 80 80 80 90 80 80 80 80 80 80 80 80 80 105 105 80 80 80 80 80 85 85 85

FT4 FT4 FT4 FT4 FT4 FT4 CL2 VW1 VW1 VW1 FT4 VW1 VW1 VW1 VW1 VW1 VW1 VW1 FT4 FT4 FT4 FT4 FT4 FT4 FT4 FT4

RF Door switch PDU Status (ACGD) Twin GIP/GP Laptop Access cable Lan Hub cable GOCAA Power cord Keyboard/Mouse cable Host LCD Display Video cable SCSI Tower Data cable Host LCD Display Power cable Modem I/O cable DASM I/O cable LCD Display Video cable LCD Display Power cable SCSI Tower Power cable Host Computer Power cable Power Strip cord Power Strip cord EXCITE Ethernet Ccable Intercom Can Can

40 ft 15 pin SubD 49 ft. 9 pin SubD 3.3 ft. Ethernet Cat 5 5.9 ft. Ethernet Cat 5 3.9 ft. 9 pin SubD 2.9 ft. 25 pin SubD 2 PS/2s 9.9 ft. 29 pin DVI, 15 pin VGA 9.9 ft. SCSI cable 50 9.9 ft. Power (3#18) 3.9 ft. 25/9 pn SubD 5.9 ft. SCSI cable 50 9.9 ft. Data (15#28) 9.9 ft. Power (3#18) 9.9 ft. Power (3#18) 9.9 ft. Power (3#18) 3.3 ft. Power (3#14) 3.3 ft. Power (3#14) 75 ft. Ethernet Cat 5
65 ft 25 pin SubD 26 ft. 9 pin SubD 65 ft 9 pin SubD 65 ft 25 pin SubD 8Coax 8W8 Combo D 8Coax 8W8 Combo D 2Coax 8W8 Combo D

TAC J12 OW1A12 ETHRNET OW1A15 ETHRNET OW1 A21 J17


OW1 A21 J8

OW1 A6 HOST LCD OW1 A16 SCSI TOWER OW1 A6 Pwr


OW1 A13

OW1 A5 DASM
OW1 A3 LCD

OW1 A3 Pwr OW1 A16 Pwr OW1 A15 Pwr OW1 A11 OW1 A11 MR2 A11 J63
PP1 J12 MR2 A11 J56 PP1 PHPS J154 PP1 PHPS J32 PP1 J200 PP1 J201 PP1 J202

10-13

APPENDIX

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

1027 EQUIPMENT AND OPERATOR ROOM CABLES (Continued) TABLE 109 (CONTD) EQUIPMENT & OPERATOR ROOM RUNS
VOLTAGE RATING ACTUAL VOLTAGE TEMP RATING CABLE TYPE FLAME RATING

RUN

FROM

TO

DESCRIPTION

REMARKS

1188 1189 1190 1191 1192 1193 1194 1195 1196 1233 1249 1255 1256 1259

MR2 DRV J3 MR2 DRV J5 MR2 DRV J10 MR2 DRV J21 MR2 DRV J22 MR2 DRV J23 MR2 DRV J24 MR2 DRV J25 PP1 A17 J1 TAC J116 (J5) MR8 J3 OW1 A21 J18 OW1 A20, A7, A8 MR2 RF AMP J4

PP1 A1

J77

65 ft 15 pin SubD 65 ft 37 pin SubD 65 ft 37 pin SubD 65 ft. RG 58C/U Coas 65 ft. RG 58C/U Coas 65 ft. RG 58C/U Coas 65 ft. RG 58C/U Coas 65 ft. RG 58C/U Coas 2 ft 15 pin SubD 30 ft. 9 pin SubD 5 ft. Times Microwave LMR 60 ft 9 pin SubD 13.2 ft. 50 pin SCSI, 9 pin SubD

CI3 CI3 CI3 CI2 CI2 CI2 CI2 CI2 CI3 CI3 CATV R CI3 CI3

300 300 300 1900 VRMS 1900 VRMS 1900 VRMS 1900 VRMS 1900 VRMS 300 300 6000 VDC 300 300

<30 <30 <30 1000 VDC 1000 VDC 1000 VDC 1000 VDC 1000 VDC <30 <30

80 80 80 80 80 80 80 80 80 80 85

FT4 FT4 FT4 VW1 VW1 VW1 VW1 VW1 FT4 FT4 FT4 FT4 VW1

16 Channel SW Cntrl Power MC Drive MC Drive Dynamic Disable Dynamic Disable Dynamic Disable Dynamic Disable Dynamic Disable Repeater Board Power Can RF Extension cable Estop SCIM, Mouse, Keyboard cable

PP1 A17 J78 PP1 J203 PP1 A11 J92 PP1 A11 J93 PP1 A11 J94 PP1 A11 J95 PP1 A11 J96 PP1 PHPS J35 PP1 PHPS J155 CONNECT TO RUN 1238 MR2 A11 J26

<30 <30

80 80

OW1 A21 J6
CONNECT TO RUN 235

1028 CABLE SPARES (FRUS) TABLE 1010 EQUIPMENT & OPERATOR ROOM RUNS CABLE PART NUMBER 46244710G30 46244712G30 46244713G30 46244714G30 CABLE RUN Run 1227 Run 1236 Run 1188 Run 1181 Run 1180 Run 1181 Run 1233 Run 1189 Run 1190

46258877G30

10-14

APPENDIX

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

1029 EXCITE 4KW AND 8KW MNS CABLES (Optional) TABLE 1011 EXCITE 4KW AND 8KW MNS RUNS (optional)
VOLTAGE RATING ACTUAL VOLTAGE TEMP RATING CABLE TYPE FLAME RATING

RUN

FROM

TO

DESCRIPTION

REMARKS

1129 1130 1136 1137 1138 1139 1140 1141 1142 1143 1144 1145 1146 1147 1158 1248

CPC J8 CPC J3 CPC J9 RUN 1158 MR6 J12 AMT J3 MR1 J123 MR1 J102 MR1 J103 MR1 J104 MR1 J105 MR2 A11 J4 RUN 1159 MG3 A17 J3 RUN 1137 RUN 1146

MR3 PD1 MR2 A11 J52 MR2 A11 J4 PP1 J83 MR3 PD1 Mr2 A11 J52 MR6 J1 AMT J5 AMT J6 AMT J7 AMT J8 AMT J11 PP1 J83 PP! J83 MR1 CPC J2 MR9 J4

35 ft. Power (3#10) 30 ft. RG 223/U Coax 30 ft. RG 223/U Coax 100 ft. Times Microwave LMR 600FR 65 ft. Power (5#8) 30 ft. RG 223/U Coax 30 ft. 25 pin SubD 30 ft. RG 223/U Coax 30 ft. RG 223/U Coax 30 ft. RG 223/U Coax 30 ft. RG 223/U Coax 30 ft. RG 223/U Coax 100 ft. Times Microwave LMR 600FR 100 ft. Times Microwave LMR 600FR 5 ft. Times Microwave LMR 400FR 5 ft. Times Microwave LMR 400FR

STO CMX CMX CATV R SO CMX CI3 CMX CMX CMX CMX CMX CATV R CATV R CATV R CATV R

600 1900V RMS 1900V RMS 4000V DC 600 1900V RMS 300 1900V RMS 1900V RMS 1900V RMS 1900V RMS 1900V RMS 4000V DC 4000V DC 2500V DC 2500V DC

208 <30 <30

90 80 80 85

FT4 VW1 VW1 FT4 FT4 VW1 FT4 VW1 VW1 VW1 VW1 VW1 FT4 FT4 FT4 FT4

4KW CPC Amp Power cable 4KW MNS cable 4KW MNS cable 4KW MNS cable BB RF Amp Power Cable 8KW MNS cable 8KW MNS cable 8KW MNS cable 8KW MNS cable 8KW MNS cable 8KW MNS cable 8KW MNS cable 8KW MNS cable 4KW & 8KW MNS cable 4KW MNS Extension cable 8KW MNS Extension cable

208Y <30 <30 <30 <30 <30 <30 <30

75 80 80 80 80 80 80 80 85 85 85 85

10-15

APPENDIX

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

103

FIBER OPTIC TERMINATION INSTRUCTIONS

WARNING!
FERROUS MATERIAL HAZARD! CRIMP TOOL AND OTHER TOOLS REQUIRED FOR THIS PROCEDURE CONTAIN FERROUS MATERIAL AND WILL BE STRONGLY ATTRACTED TO MAGNET AND MAY BECOME DANGEROUS PROJECTILES. KEEP ALL FERROUS TOOLS AT LEAST 10 FEET AWAY FROM THE MAGNET. 1031 FIBER OPTIC CABLE TERMINATION KIT The connectors on plastic fiber optic cables can be replaced if damaged or cut off to shorten. 1. 2. The Fiber Optic Repair Kit, 46301450G1 shown below, contains all the necessary materials to replace a damaged connector. The Spares Kit, delivered with the Shipping Collector, contains a 46320119P1 Termination Kit for Fiber Optic Cables. This kit consists of consumables such as terminals and polishing sheets used to terminate cables during installation. Remove the end of fiber optic cable to be reterminated from magnet room if possible. f it is not possible to remove the end to be repaired from the magnet room, the cable must be at least 10 feet away from the magnet before it can be worked on. Cut cable to desired length. Note: The connectors on glass fiber optic cables cannot be replaced using the F/O Repair Kit. At this time, there is no kit available to repair glass fiber optic cable connectors.

3. 4. 5.

Crimping Tool Polishing Kit Six bags of Connectors with Crimp Rings

10-16

APPENDIX

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

1032 STRIPPING FIBER OPTIC OUTER JACKET The separated duplex cable must be stripped to equal lengths on each side of the cable. This allows proper seating of the cable into the connector. 1. 2. For Duplex Cables, seperate the two fibers approximately 2 in. (50mm) back from the ends. Strip off approximately 0.3 in. (7 mm) of the outer jacket with the 16 gauge wire strippers. Excess webbing on duplex cable may have to be trimmed to allow the connector to slide over the cable. 1 2 in. 50mm 2

1033 CRIMPING CONNECTOR TO FIBER OPTIC CABLE The Connector Crimp Rings cannot be interchanged between simplex and duplex connectors. 1. 2. 3. Place the Crimp Ring and Connector over the end of the cable. The fiber should extend approximately 0.12 in. (3mm) through the end of the connector. Carefully position the ring so that it is entirely on the connector and then crimp the ring in place with the crimping tool.

The Duplex Crimp Ring is shiny and has a thinner wall 2 The Simplex Crimp Ring has a dull luster.

10-17

APPENDIX

GE MEDICAL SYSTEMS REV 1

SIGNA EXCITE HD 1.5T TwinSpeed SYSTEM INSTALLATION DIRECTION 5133312

1034 TRIMMING EXCESS FIBER FROM CONNECTOR 1. Cut off excess fiber protruding from the connector. The trimmed fiber should extend approximately 0.06 inch (11/2mm).

0.06 in. (1.5mm)

1035 POLISHING FIBER OPTIC END The four dots on the bottom of the polishing fixture are wear indicators. Replace the polishing fixture when any dot is no longer visible. 1. 2. 3. 4. 5. Insert the connector fully into the polishing fixture with the trimmed fiber end protruding from the bottom of the fixture. Place the 600 grit abrasive paper on a flat smooth surface. Pressing down on the indicator, polish the fiber and the connector using a figure eight motion until the connector is flush with the end of the polishing fixture. Wipe the connector and fixture with a clean cloth or tissue. Place the flush connector and polishing fixture on the dull side of the 3 micron pink lapping film and continue to polish the fiber and connector for approximately 25 strokes. The fiber end should be flat, smooth and clean. The cable end can now be connected. 1 2

10-18

APPENDIX