Beruflich Dokumente
Kultur Dokumente
Abstract - In the medical field, medical instruments used in the operations should be measured their purification level
after the washing and sterilization process, because it is necessary to evaluate them to be clear excellent. However, some
problems existed in a current evaluation of purification level. The qualitative and the quantitative evaluation were not
enough, and the residual detergent on them has been the other problem. Then, we examined an application of the
Fourier transform infrared spectroscopy to qualitative and the quantitative evaluation concerning the soil in this research.
This paper aims to propose a new evaluation of purification level.
This measurement system is able to evaluate purification level promptly and accurately. In addition, this measurement
system needs no preprocessing in every measurement. As a characteristic of measurement, anyone could use this system,
and there is not influence of individual difference.
Key Words: medical treatment, the evaluation of purification level, qualitative and the quantitative evaluation, Fourier
transform infrared rays spectroscopy measurement, the diffuse reflectance spectrum measurement
Recently, we human beings could receive a lot of fruit When the molecule in the material receives the infrared
of the medical technology. For example, very clear rays light, the molecule is absorbed specific infrared rays
images of our stomach could be easily obtained and an light, and changes from a low state of energy to a high
adequate diagnosis makes us healthier. In these medical state of energy. A molecular vibration is generated by the
fields, cleanness should be verified for all instruments to change in this state of energy. The qualitative analysis of
minimize the risk of the infection. the material becomes possible by measuring the infrared
Therefore, every used instruments must be evaluated absorption spectrum that attributes to this molecular
the purity after washing and sterilization in order to vibration. At the same time, the intensity of the
verify the adequate cleanness. Until now, evaluation of absorption peak is proportional to the density and the
purification level of medical instruments has been carried optical path length of the material (Law of Lambert
out by the visual inspection, amido Black method, or -Beer). Therefore, the quantitative analysis becomes
adenosine triphosphate (ATP) method [1] [2]. However, possible by measuring the absorption spectrum, too.
the personal error, need of rinsing the residual chemicals
or lack of the reproducibility has not been solved in these 3. EXPERIMENTAL PROCEDURES
methods. Therefore, FT-IR measurement system
equipped with diffuse reflectance spectrum was 3.1 Experimental apparatus
developed to evaluate the real medical instruments. This This research uses a Fourier transform infrared
study used plane sheets having ruggedness on the surface. spectrophotometer (SHIMADZU CORPORATION: IR
Based on this measuring system, qualitative and/or Prestige-2 1). The configuration of it is shown in Fig. 1.
quantitative evaluation of soil was studied. The infrared light is emitted from the light source and is
1458
uneven plane model could be measured. Therefore, we The similar experiment of the using lard as a sample,
could expect the efficient estimation method to verify the using 88 levels of lard-in-hexane solutions, 176 spectra
cleanness of the medical instruments of complicated in total were measured to make calibration model. Fig.4
form. is the scatter diagram when PLS regression is analyzed
by 4000-lOOOcm'1. At this time, the correlation
The spectrum of the 000 and 1% are shown in Fig.3. coefficient 0.890 and measurement accuracy ±3.58mg
0.42mg and 0.78 mg are the reference chemical was obtained. Therefore, it is possible to measurement
remaining. In this Spectrum of 0°, 0.42mg is influence remaining dirt using lard as a sample because it could get
of solvent. The absorption peaks which can't see with a significant correlation.
spectrum of 0O are seen with 3000cm'1-2850cm'1,
1545cm'1, 1455cm'1, 1400cm'1, 1270cm'l- 1190cm' and The spectrum of the 00O and 1% are shown in Fig.5.
1190cm'1 -1140cm-' with a spectrum of I1%. The absorption peaks which can't see with a spectrum of
Therefore, because the absorption peak of the fresh 0O are seen with 2920cm-1, 2850cm-1, 1750cm'1, 1460
horse blood is seen on the spectrum at the arrow of Fig.3, cm'i and 1160cm'1 with a spectrum of 1%. As well as
there could be a detection limit with this method in less fresh horse blood measurement these absorption peaks
than 0.78mg. And, this measurement limit is about the are the peak of the lard to be consistent with the
same as the technique except that it is reported. wavenumber of the main absorption peaks of the lard.
Fresh horse blood 0% 0.42[mgj
1545 Lard 0% * * 0.1 6[mg]
Fresh horse blood 1%* 0.78[mg]
0.6 /1455 Lard 1% .3.86[mgj
1.6
0.5 2850 1.4 2850
O 1.2
1750
0 .4 (. 1
2920/
0.3 C 0.8
70 1460 1160
[00
c, 0.2
0- 1 1270 n
U, 0.4
0.24--
4 * pA(
/XXoo,
_
.0
o
<0.1 1190 0
0
-0. 00 3500 3000 2500 2000 1500 1000
-0.140[00
3500 3000 2500 2000 1500 1 00 -0.4
Wave number[cm-11
Wave number[cm-1
Fig.5: Qualitative evaluation of Lard
Fig.3: Qualitative evaluation of Horse Blood
It makes clear that it is possible in this range to analyze
the lipid dirt on the uneven plane model. At the same
4.2 Lard measurement time, it was proved that it is excellent in the
reproducibility when using other sample.
30
The time to get a spectrum is about two minutes in this
M R=0.89C0 experiment. This time is the shortest time among the
20
usual methods. In the measurement, only put sample on
a)
the sample room. From this, it is clear that measurement
t** *00¢tv
a) 0~~~~~ * a
>* '**
1459
measurement method of remaining dirt would meet a
requirement in the medical field.
5. CONCLUSION
ACKNOWLEDGMENT
REFERENCES
1460