Sie sind auf Seite 1von 57

652 = 662

GOS-600G Service Manual

SAFETY TERMS AND SYMBOLS


These terms may appear in this manual or on the product: WARNING: Warning statements identify condition or practice that could results in injury or loss of lift. CAUTION: Caution statements identify condition or

practices that could result in damage to this product or other property.

The following symbols may appear in this manual or on the product:

DANGERG High Voltage

ATTENTIONG refer to manual

Protective Conductor Terminal

Frame or Chassis Terminal

Earth Terminal (Ground)

GOS-600G Service Manual

FOR UNITED KINGDOM ONLY


NOTE: This lead/appliance must only be wired by competent persons WARNING: THIS APPLIANCE MUST BE EARTHED
IMPORTANT: The wires in this lead are coloured in accordance with the following code:

Green/ Yellow: Earth Blue: Brown:

Neutral Live(Phase)

As the colours of the wires in main leads may not correspond with the colours marking identified in your plug/appliance, proceed as follows: The wire which is coloured Green & Yellow must be connected to the Earth terminal marked with the letter E or by the earth symbol or coloured Green or Green & Yellow. The wire which is coloured Blue must be connected to the terminal which is marked with the letter N or coloured Blue or Black. The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red. If in doubt, consult the instructions provided with the equipment or contact the supplier. This cable/appliance should be protected by a suitably rated and approved HBC mains fuse: refer to the rating information on the equipment and/or user instructions for details. As a guide, cable of 0.75mm2 should be protected by a 3A or 5A fuse. Larger conductors would normally require 13A types, depending on the connection method used. Any moulded mains connector that requires removal /replacement must be destroyed by removal of any fuse & fuse carrier and disposed of immediately, as a plug with bared wires is hazardous if a engaged in live socket. Any re-wiring must be carried out in accordance with the information detailed on this label.

ii

GOS-600G Service Manual

EC Declaration of Conformity
We GOOD WILL INSTRUMENT CO.,LTD. (1) NO. 95 - 11, Pao Chung Rd., Hsin-Tien City, Taipei Hsien, Taiwan (2) Plot 522, Lorong Perusahaan Baru 3, Prai Industrial Estate, 13600 Prai, Penang, Malaysia declare, that the below mentioned product GOS-622G, GOS-626G, GOS-635G, GOS-652G, GOS-653G, GOS-658G are herewith confirmed to comply with the requirements set out in the Council Directive on the Approximation of the Law of Member States relating to Electromagnetic Compatibility (89/336/EEC,92/31/EEC,93/68/EEC) and Low Voltage Equipment Directive(73/23/EEC). For the evaluation regarding the Electromagnetic Compatibility and Low Voltage Equipment Directive, the following standards were applied:
EN50081-1: Electromagnetic compatibility - (1992) EN50082-1: Electromagnetic compatibility -(1992) Generic emission standard Generic immunity standard Part 1: Residential, commercial and light industry Part 1: Residential, commercial and light industry Conducted Emission Radiated Emission Current Harmonics Voltage Fluctuations -----------------EN 61000-3-2 EN 61000-3-3 ---------EN 55022 class B (1994) (1995) (1995) ---------Electrostatic Discharge Radiated Immunity Electrical Fast Transients Surge Immunity IEC 1000-4-2 IEC 1000-4-3 IEC 1000-4-4 IEC 1000-4-5 (1995) (1995) (1995) (1995) (1994)

Voltage Dip/Interruption EN 61000-4-11

EN50081-2: Electromagnetic compatibility - (1993) Generic emission standard Part 2: Industrial Environment Conducted Emission Radiated Emission EN 55011 class B (1991)

Low Voltage Equipment Directive 73/23/EEC Low Voltage Directive


EN 61010-1:(1993)+A2:(1995)

iii

GOS-600G Service Manual

1. GENERAL
1.1 Description
The 6xxG family oscilloscopes are dual-channel oscilloscopes with maximum sensitivity of 1 mV/DIV, and maximum sweep time of 10 nSec/DIV. Each of these oscilloscopes employs a 6-inch rectangular type cathode-ray tube with red internal graticule. 653G and 658G each has a sweep magnification feature with B sweep whereas 626G and 658G provide the read-out function which enables an easy read out for settings and cursor measured values. These oscilloscopes are sturdy, easy to operate and exhibits high operational reliability.

1.2 Features
1) High intensity CRT with high acceleration voltage: The CRT is a high beam transmission, high intensity type with a high acceleration voltage of 2kV for models 622G ,635G and 626G ,and 12KV for models 652G, 653G and 658G. It displays clear readable traces even at high sweep speeds. 2) High stability with less drift: The oscilloscope employs a temperature compensation circuit which is newly developed to reduce the drift of base lines and DC balance disturbance caused by temperature change. 3) A trigger level lock function which makes the triggering adjustment unnecessary: A new trigger level lock circuit is incorporated. This circuit eliminates the procedures of the troublesome triggering adjustment not only for displaying signals but also for that of video signals and large duty-cycle signals. 4) TV sync triggering: The oscilloscope has a sync separator circuit incorporated within the TIME/DIV switch for automatic triggering of TV-V and TV-H signals. 5) Linear focus: Once the beam focus is adjusted to the optimum position, it is automatically maintained regardless to the intensity change.

GOS-600G Service Manual

6) Cursor readout measurement: The unique easy-to-use cursor and numerical readouts make waveform observations and measurement faster and accurate. The on- screen cursors provide seven functions(V, V%, VdB, T, 1/T, DUTY, PHASE).(626G, 658G only)

GOS-600G Service Manual

2. TECHNICAL SPECIFICATIONS
Model Specifications 20MHz OSCILLOSCOPE 622G 626G 1mV 5mV 35MHz 635G 50MHz OSCILLOSCOPE 652G 653G 658G

Sensitivity Sensitivity accuracy Vernier vertical sensitivity Frequency bandwidth ( 3dB)


5mV 5mV 1mV

5V/DIV, 12 steps in 1-2-5 sequence

5V/DIV:3%, 1mV 2mV/DIV: 5% ( 10J to 35J (50K to 95K ) ) << 5 DIV at the center of display>> To 1/2.5 or less of panel-indicated value.

5V/DIV:DC 20MHz,(622G,626G) 5V/DIV:DC 35MHz,(635G) 2mV/DIV:DC 10MHz

5mV 1mV

5V/DIV : DC 50MHz, 2mV/DIV:DC 15MHz

AC coupling: Low limit frequency 10Hz. ( with reference to 100kHz,8DIV. Frequency response with -3dB.) 5mV 5mV 1mV 5V/DIV: 17.5ns(622G,626G) 5V/DIV: 100ns(635G) 2mV/DIV: 35ns 1M ohm 5mV 5V/DIV: 7ns, 1mV 2mV/DIV: 23ns 2% // Approx. 25pF

Rise time Input impedance Square wave characteristics DC balance shift Linearity VERTI-CA L AXIS Vertical modes

Overshoot : 5% ( At 10mV/DIV range ) << 5 DIV at the center of display>> Other distortions and other ranges: 5% added to the above value. ( 10J to 35J (50K to 95K ) ) 5mV < 5V/DIV: 0.5DIV, 1mV 2mV/DIV: 2.0DIV

0.1 DIV of amplitude change when waveform of 2 DIV at graticule center is moved vertically.

CH1 : CH1 single channel. CH2 : CH2 single channel. DUAL : CHOP/ALT are auto-set by TIME/DIV switch.(CHOP:0.5s 5ms/DIV, ALT:2ms 0.1g s/DIV) When CHOP switch is pushed in, the two traces are displayed in the CHOP mode at all range. ADD : CH1 + CH2 algebraic addition. Approx. 250kHz

Chopping repetition frequency Input coupling

AC, DC, GND

Maximum input 400V ( DC+AC peak ), AC: frequency 1kHz or lower. When set probe switch at 1:1, the maximum effective readout is 40Vpp (14Vrms at sine wave), voltage or set probe switch at 1:10, the maximum effective readout is 400Vpp (140Vrms at sine wave). Common mode rejection ratio Isolation between channels CH1 signal output CH2 INV BAL. Dynamic range Signal delay
50:1 or better at 50kHz sinusoidal wave. (When sensitivities of CH1 and CH2 are set equally) > 1000:1 at 50kHz , >30:1 at 20MHz (622G,626G) > 1000:1 at 50kHz , >30:1 at 35MHz(635G) (At 5mV/DIV range) > 1000:1 at 50kHz > 30:1 at 50MHz (At 5mV/DIV range)

Approx. 100mV/DIV without termination, 50mV/DIV with 50 ohm termination. Bandwidth(-3dB) : 622G/626G/635G : 20MHz 652G/653G/658G : 40MHz Balanced point variation : 1 DIV ( Reference at center graticule.) >8DIV at 20MHz >5DIV at 35MHz (635G only) w w w >8DIV at 50MHz w Leading edge can be monitored.

GOS-600G Service Manual


Model Specifications 20MHz OSCILLOSCOPE 622G 626G 35MHz 635G 50MHz OSCILLOSCOPE 652G 653G 658G

Triggering source Coupling Polarity

CH1, CH2, LINE, EXT ( CH1 and CH2 can be selected only when the vertical mode is DUAL or ADD. In ALT mode, if the TRIG. ALT switch is pushed in, it can be use for alternate triggering of two different source. ) AC, HF-REJ, TV, DC ( TV-V/TV-H can be auto-set by TIME/DIV range. TV-V: 0.5s-0.1ms/DIV; TV-H: 50g s-0.1g s/DIV) +/ DC 5MHz : 0.5 DIV ( EXT: 0.1V ) DC 10MHz : 0.5 DIV ( EXT: 0.1V ) 5 20MHz : 1.5 DIV ( EXT: 0.2V ) (622G,626G) 10 50MHz : 1.5 DIV ( EXT: 0.2V ) 5 35MHz : 1.5 DIV ( EXT: 0.2V ) (635G) TV(video signal): 2.0 DIV ( EXT: 0.2V ) AC coupling : Attenuate signal components of lower than 10Hz. HF-REJ: Attenuate signal components of higher than 50kHz.

Sensitivity

TRIGGER -ING

Triggering modes

AUTO : Sweeps run in the free mode when no triggering input signal is applied. ( Applicable for repetitive signals of frequency 50Hz or over. ) NORM : When no triggering signal is applied, the trace is in the READY state and not displayed. SINGLE : One-shot sweep with triggering signal. Can be reset to the READY state by means of the RESET switch. The READY lamp(LED) turns on when in the READY state or in the sweep operation. (626G,652G,653G and 658G only)

LEVEL LOCK and ALT triggering

Satisfies the value of the above trigger sensitivity plus 0.5 DIV ( EXT: 0.05V ) for signal of duty cycle 20 : 80. Repetition frequency : 50Hz Repetition frequency : 50Hz 20MHz (622G,626G) 35MHz (635G) Repetition frequency : 50Hz 40MHz

EXT triggering signal input

EXT HOR input terminal is used in common.

Input impedance 1M ohm 2% // approx. 35pF Max. input voltage B triggering signal. Horizontal axis display A sweep(main sweep ) time Sweep time accuracy
100V ( DC+AC peak ), AC: Frequency not higher than 1kHz

The A triggering signal of main sweep is used as the B triggering signal.

A, A INT, B, B TRIG'D

0.1g s

0.5s/DIV, 21 steps in 1-2-5 sequence

3%, (10J

to 35J

( 50K to 95F ) )

Vernier sweep HORIZON time control TAL Hold off time AXIS B sweep delay system B sweep(delay sweep) time Sweep time accuracy Delay time

1/2.5 of panel-indicated value Continuous variable >= twice sweep length (time) at 0.1Sec~1mSec/DIV ranges. w w w w w w w w Continuous delay and triggered Delay 0.1g s steps 0.5ms/DIV, 12 ( 50K to

w w

w w

w w

w w

3%, (10J to 35J 95K ) ) 1g s 5msSec

GOS-600G Service Manual


Model Specifications 20MHz OSCILLOSCOPE 622G 626G 35MHz 635G 50MHz OSCILLOSCOPE 652G 653G 658G

Sweep magnification 10MAG sweep time HORIZON accuracy TAL AXIS Linearity Position shift caused by 10MAG Sensitivity Sensitivity accuracy X-Y MODE Frequency bandwidth X-Y phase difference Sensitivity Frequency EXT HOR bandwidth MODE Phase difference between vertical axis Sensitivity Frequency bandwidth Z AXIS Input resistance Maximum input voltage Waveform Frequency CALIBRA- Duty ratio TION VOLTAGE Output voltage Output impedance Type Phosphor Acceleration voltage Effective screen size Graticule
DC

10 times ( maximum sweep time 10ns/DIV )

0.1g s

50ms/DIV

5%, 10ns

50ns/DIV

8% (10J

to 35J (50K to 95K ))

NORM:

3%,

10MAG:

5% (

8% for 10ns

50ns/DIV)

Within 2 div. at CRT screen center

Same as vertical axis.(X-axis:CH1 input signal; Y-axis:CH2 input signal.) NORM: 4%, 10MAG: 6% (10J to 35J (50K to 95K ))

1MHz (-3dB) 3 at DC

DC

2MHz (-3dB) 3 at DC

50kHz

100kHz

Approx. 0.1V/DIV (Trace swept by an external horizontal signal applied to the EXT TRIG IN terminal. Vertical axis modes are CH1, CH2, DUAL and ADD modes in the CHOP mode.) DC 1MHz (-3dB) DC 2MHz (-3dB)

3 at DC

50kHz

3 at DC

100kHz

3 Vp-p ( Trace becomes brighter with negative input.) DC 5MHz

Approx. 5k ohm 50 V ( DC+AC peak, AC frequency 1kHz ) Positive-going square wave 1 kHz 5%

Within 48:52 2 Vp-p 2%

Approx. 2 k ohm.

6-inch rectangular type, internal graticule. P 31 Approx. 2 kV Approx. 12 kV

CRT

10 DIV ( 1 DIV = 10mm(0.39in))

Internal; continuous adjustable illumination (626G,652G,653G,658G only)

GOS-600G Service Manual


Model Specifications 20MHz OSCILLOSCOPE 626G 50MHz OSCILLOSCOPE 658G V , V% , VdB, T ,1/ T ,DUTY ,PHASE

Cursor measurement functions

Cursor display format

(DELTA),

(REF)

CURSOR Cursor resolution READOUT Effective cursor range from center graticule Panel setting display

1/25 DIV Vertical: 3 DIV Horizontal: 4 DIV V/DIV,V-MODE, INV, ALT/CHOP, UNCAL, ADD(SUB), PROBE( 1/ 10), X-Y, A T/D, TV-V/H, B T/D(for 658G only) 10MAG,

Line Power Requirements


Voltage: AC 100V, 120V, 220V, 230V 10% selectable Frequency: 50Hz or 60Hz Power consumption: Approx. 70VA, 60W(max.)

Operating Environment
Indoor use Altitude up to 2000 m Ambient temperature : To satisfy specifications : 5 to 35J ( 41 to 95 F ) Maximum operating ranges: 0 to 40J ( 32 to 104 F ) Relative humidity:85% RH(max.) non condensing Installation Category Pollution degree 2 II

Accessories
Power cord 1; Instruction manual. 1; Probes 2

Mechanical Specifications
Dimensions: 310 W x 150 H x 455 D (mm) Weight: Approx.8.2Kg (18 lbs.)

Storage Temperature & Humidity


-10 to 70J ,70%RH (maximum)

GOS-600G Service Manual

3. PRECAUTIONS BEFORE OPERATING THE OSCILLOSCOPE


3.1 Unpacking the Oscilloscope
The oscilloscope is shipped from the factory after being fully inspected and tested. Upon receiving the instrument, immediately unpack and inspect it for any damages that might have been sustained during transportation. If any sign of damage is found, immediately notify the bearer and/or the dealer.

3.2 Checking the Line Voltage


These oscilloscopes will operate on any one of the line voltage shown in the table below, by inserting the line voltage selector plug in the corresponding position on the rear panel. Before connecting the power plug to an AC line outlet, make sure the voltage selector is set to the correct position corresponding to the line voltage. Note the oscilloscope may be damaged if it is connected to the wrong AC line voltage. WARNING. To avoid electrical shock the power cord protective grounding conductor must be connected to ground. When line voltages are changed, replace the required fuses shown below. Line voltage 100V 120V Range 90-110V 108-132V Fuse T 0.63A 250V Line voltage 220V 230V Range 198-242V 207-250V Fuse T 0.315A 250V

WARNING. To avoid personal injury, disconnect the power cord before removing the fuse holder.

3.3 Environment
The normal ambient temperature range of this instrument is 0 to 40C (32 to 104F). Operation of the instrument above this temperature range may cause damage to the circuits.

GOS-600G Service Manual

Do not use the instrument in a place where strong magnetic or electric field exists. Such fields may disturb the measurement.

3.4 Equipment Installation, and Operation


Ensure there is proper ventilation for the hole vents in the oscilloscope case. If this equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.

3.5 CRT Intensity


To prevent permanent damage to the CRT phosphor, do not make the CRT trace excessively bright or leave the spot stationary for an unreasonably long time.

3.6 Withstanding Voltages of Input Terminals


The withstanding voltages of the instrument input terminals and probe Input terminals are as shown in the following table. Do not apply voltages higher than these limits. Input terminal CH1, CH2, inputs EXT TRIG input Probe inputs Z AXIS input Maximum input voltage 400V (DC + AC peak) 100V (DC + AC peak) 600V (DC + AC peak) 50V (DC + AC peak)

CAUTION. To avoid instrument damage, do not exceed maximum input voltages. Maximum input voltages must have frequencies less than 1 kHz. If an AC voltage which is superimposed on a DC voltage is applied, the maximum peak of CH1 and CH2 input voltage must exceed + or 400V. So for voltages with a mean value of zero volt the maximum peak to peak value is 800Vpp.

GOS-600G Service Manual

Figure 4-1(a) Model 622G & Model 635G

Figure 4-1(b) Model 652G

GOS-600G Service Manual

Figure 4-1(c) Model 653G

Figure 4-1(d) Model 626G

10

GOS-600G Service Manual

Figure 4-1(e) Model 658G

11

GOS-600G Service Manual

4. OPERATION METHOD
4.1 Introduction of Front Panel
CRT : POWER-----(9) Main power switch of the instrument. When this switch is turned on, the LED (8) is also turned on. INTEN-----(2) Controls the brightness of the spot or trace. B INTEN-----(3) (653G & 658G only) Semi-fixed potentiometer for adjusting trace intensity when in B sweep mode. READOUT INTEN-----(7)(626G & 658G only) Semi-fixed potentiometer for adjusting intensity of the readout and cursors. FOCUS-----(4) For focusing the trace to the sharpest image. ILLUM-----(6)(Except 622G & 635G) Graticule illumination adjustment. TRACE ROTATION-----(5) Semi-fixed potentiometer for aligning the horizontal trace in parallel with graticule lines. FILTER-----(42) Filter for ease of waveform viewing. Vertical Axis: CH 1 (X) input.............(12) Vertical input terminal of CH 1. When in X-Y operation, X-axis input terminal. CH 2 (Y) input.............(16) Vertical input terminal of CH 2. AC-DC-GND...............(11)(15) Switch for selecting connection mode between input signal and vertical amplifier. AC DC GND : AC coupling : DC coupling : Vertical amplifier input is grounded and input terminals are disconnected. When in X-Y operation, Y-axis input terminal.

12

GOS-600G Service Manual

VOLTS/DIV...............(10)(14) Select the vertical axis sensitivity, from 1mV/DIV to 5V/DIV in 12 ranges. VARIABLE................(13)(17) Fine adjustment of sensitivity, with a factor of 1/2.5 of the indicated value. When in the CAL position, sensitivity is calibrated to indicated value. POSITION..................(40)(37) Vertical positioning control of trace or spot. VERT MODE.............(39) Select operation modes of CH 1 and CH 2 amplifiers. CH 1 CH 2 : The oscilloscope operates as a single-channel instrument with CH 1 alone. : The oscilloscope operates as a single-channel instruments with CH 2 alone.

DUAL : The oscilloscope operates as a dual-channel instrument both CH 1 and CH 2. CHOP/ALT are automatic changed by TIME/DIV switch (18). When CHOP (41) button is pushed in, the two traces are displayed in the CHOP mode at all ranges. ADD : The oscilloscope displays the algebraic sum (CH 1 + CH 2) or difference (CH 1 - CH 2) of the two signals. The pushed in state of CH 2 INV(36) button is for the difference (CH 1 - CH 2).

Triggering: EXT TRIG(EXT HOR) input terminal-----(23) Input terminal is used in common for external triggering signal and external horizontal signal. To use this terminal, set SOURCE switch (26) to the EXT position. SOURCE-----(26) Select the internal triggering source signal, and the EXT HOR input signal. CH 1 (X-Y) : When the VERT MODE switch (39) is set in the DUAL or ADD state, select CH 1 for the internal triggering source signal. When in the X-Y mode, select CH 1 for the X-axis signal. : When the VERT MODE switch (39) is set in the DUAL or ADD state, select CH 2 for the internal triggering source signal.

CH 2

TRIG.ALT(24) : When the VERT MODE switch(39) is set in the DUAL or ADD state, and the SOURCE switch(26) is selected at CH 1 or CH 2, with the engagement of the TRIG.ALT switch(24), it will alternately select CH 1 & CH 2 for the internal triggering source signal. LINE : To select the AC power line frequency signal as the triggering signal.

13

GOS-600G Service Manual

EXT

: The external signal applied through EXT TRIG(EXT HOR) input terminal(23) is used for the external triggering source signal. When in the X-Y, EXT HOR mode, the X-axis operates with the external sweep signal

COUPLING-----(25) Select COUPLING mode (25) between triggering source signal and trigger circuit; select connection of TV sync trigger circuit. AC DC HF REJ TV : AC coupling : DC coupling : Removes signal components above 50kHz(-3dB) : The trigger circuit is connected to the TV sync separator circuit and the triggered sweeps synchronize with TV-V or TV-H signal at a rate selected by the TIME/DIV switch (18) TV-V: 0.5 s/DIV - 0.1ms/DIV TV-H: 50s/DIV - 0.1s/DIV SLOPE-----(22) Select the triggering slope. + : Triggering occurs when the triggering signal crosses the triggering level in positive-going direction. : Triggering occurs when the triggering signal crosses the triggering level in negative-going direction LEVEL-----(30) To display a synchronized stationary waveform and set a start point for the waveform. Toward + : The triggering level moves upward on the display waveform. Toward : The triggering level moves downward on the display waveform. LOCK-----(29) Triggering level is automatically maintained at optimum value irrespective of the signal amplitude (from very small to large amplitudes), requiring no manual adjustment of triggering level. HOLDOFF-----(31) Used when the signal waveform is complex and stable triggering cannot be attained with the LEVEL knob alone.

14

GOS-600G Service Manual

TRIGGER MODE-----(28) Select the desired trigger mode. AUTO NORM SINGLE : When no triggering signal is applied or when triggering signal frequency is less than 50 Hz, sweep runs in the free run mode. : When no triggering signal is applied, sweep is in a ready state and the trace is blanked out. Used primarily for observation of signal 50 Hz. : Use for single sweep.(Except 622G & 635G)

Push to RESET: Operation (one-short triggering operation), and in common as the reset switch. When these three buttons are disengaged, the circuit is in the single trigger mode. The circuit is reset as this button is pressed. When the circuit is reset, the READY lamp turns on. The lamp goes out when the single sweep operation is over. Time Base (A) TIME/DIV-----(18) Select the sweep time of the A sweep. (A and B sweep in common for 658G only, B TIME/DIV < A TIME/DIV) B TIME/DIV-----(43) (653G only)

Select the sweep time of delayed sweep (B sweep). SWP.VAR-----(21) Vernier control of sweep time. When SWP.UNCAL(19) button is pushed in, the sweep time can be made slower by a factor 2.5 of the indicated value. The indicated values are calibrated when this button is not pushed in. POSITION-----(34) Horizontal positioning control of the trace or spot. 10 MAG-----(33) When the button is pushed in, a magnification of 10 occurs. DELAY TIME-----(44)(653G only) Vernier control of the delay time selected by the A TIME/DIV(18) and B TIME/DIV(43) switch to finely select the portion of the A sweep waveform to be magnified. HORIZ. DISPLAY MODE-----(38)(653G & 658G only) Select A and B sweep modes as follows: A : Main sweep(A sweep) mode for general waveform observation.

15

GOS-600G Service Manual

A INT

: This sweep mode is used when selecting the section to be magnified of A sweep, in preparation for delayed sweep. The B sweep section (delayed sweep) corresponding to the A sweep is displayed with a high intensity beam. : Display the delayed sweep(B sweep) alone. : Select between continuous delay and triggered delay. Disengaged : For continuous delay. The B sweep starts immediately after the sweep delay time set by A TIME/DIV and B TIME/DIV switch and DELAY. TIME knob has elapsed. Engaged : For triggered delay. The B sweep starts when the triggering pulse is applied after the sweep delay time set by A TIME/DIV and B TIME/DIV switch and DELAY TIME knob has elapsed. (The triggering signal is used in common for both A sweep and B sweep).

B B TRIGD

X-Y-----(27) Press the X-Y button to enable X-Y operation. Others CAL-----(1) This terminal delivers the calibration voltage of 2 Vp-p, 1kHz, positive square wave. The output resistance is 2k ohm. See page 3 technical specification for tolerance. GND-----(20) Ground terminal of oscilloscope mainframe. Readout Function(Only available in 626G & 658G) CURSOR measurement-----(32) CURSOR ON/OFF : Press the button to turn on/off the cursor measurement. CURSOR Function : Press the button to select the measurement functions. V V% VdB T 1/T DUTY : Voltage difference measurement. : Voltage difference measurement in percentage (5 div =100% ref.) : Voltage gain measurement(5 div = 0dB ref. , VdB = 20 log div/5div) : Time difference measurement. : Frequency measurement. : Duty cycle or time difference percent (T%) measure-ment (5 div =100% ref.)

PHASE : Phase measurement (5 div = 360 ref.)

16

GOS-600G Service Manual

TRACK- (REF)

: Press the button to select the cursor(s) to be moved. The selected cursor is indicated by either a or symbol. When both symbols are displayed, the two cursors can be shifted simultaneously.

POSITION-----(35) : Rotate the cursor POSITION control to position the selected cursor(s). When HORZ.DISPLAY mode is set A INT or B state and CURSOR ON/OFF is set off, the POSITION control is used in common as the DELAY TIME (For 658G only). READOUT ON/OFF PROBE 1/ 10 To turn on/off the readout status on the CRT, press the CURSOR ON/OFF and CURSOR FUNCTION buttons at the same time. To indicate the voltage readout is scaled for 1 or 10 probe, press the TRACK- - (REF) button and rotate the cursor POSITION(35) at the same time

4.2 Introduction of Rear Panel

Figure 4-2 Z AXIS INPUT-----(45) Input terminal for external intensity modulation signal. CH 1 SIGNAL OUTPUT.......(46) Delivers the CH 1 signal with a voltage of approximately 100mV per 1 DIV of graticule. When terminated with 50 ohms, the signal is attenuated to about one half. Suitable for frequency counting, etc.

17

GOS-600G Service Manual

AC POWER Input Circuit AC Power input connector-----(47) AC Power input socket. Connect the AC power cord(supplied) to this connector. FUSE & line voltage selector----- (48) Fuse rating is shown in Page 7. Line voltage selector: to select power sources. Studs-----(49) Studs for laying the oscilloscope on its back to operate it in the upward posture. Also used to take up the power cord.

4.3 Basic Operation


Before connecting the power cord to an AC line outlet, make sure that the AC line voltage input switch on the rear panel of the instrument is correctly set for the AC line voltage. After ensuring the voltage setting, set the switches and controls of the instrument as shown below: Item POWER INTEN FOCUS ILLUM VERT MODE CHOP CH 2 INV POSITION VOLTS/DIV VARIABLE AC-DC-GND SOURCE COUPLING SLOPE TRIG ALT LEVEL LOCK HOLDOFF TRIGGER MODE No (9) (2) (4) (6) (39) (41) (36) Setting Disengage position(OFF) Clockwise (3-oclock position) Mid-position Full anti-clockwise position (Except 622G & 635G) CH 1 Released Released

(40)(37) Mid-position (10)(14) 0.5V/DIV (13)(17) CAL(clockwise position) (11)(15) GND (26) (25) (22) (24) (29) (31) (28) Set to CH 1 AC + Released Pushed in MIN(anti-clockwise) AUTO

18

GOS-600G Service Manual

Item HORZ. DISPLAY mode TIME/DIV SWP.UNCAL POSITION 10 MAG X-Y

No (38) (18) (19) (34) (33) (27) A (653G &658Gonly) 0.5mSec/DIV Released Mid-position Released Released

Setting

After setting the switches and controls as mentioned, connect the power cord to the AC line outlet, and then, continue as follows: 1) Engage the POWER switch and make sure that the power LED is turned on. In about 20 seconds, a trace will appear on the CRT screen. If no trace appears in about 60 seconds, counter checks the switch and control setting. 2) Adjust the trace to an appropriate brightness and image with the INTEN controls and FOCUS control respectively. 3) Align the trace with the horizontal center line of the graticule by adjusting the CH 1 POSITION control and TRACE ROTATION control(adjustable by screwdriver). 4) Connect the probe to the CH 1 INPUT terminal and apply the 2Vp-p CALIBRATOR signal to the probe tip. 5) Set the AC-DC-GND switch to the AC state. A waveform as shown in the figure 4-3 will be displayed on the CRT screen 6) Adjust the FOCUS control so that the trace image appears sharply. 7) For signal viewing, set the VOLTS/DIV switch and TIME/DIV switch in appropriate positions so that signal waveform is displayed clearly. 8) Adjust the POSITION and <> POSITION controls in appropriate positions so that the displayed waveform is aligned with the graticule and voltage(Vp-p) and period(T) can be read conveniently. The above are the basic operating procedures of the oscilloscope. The above procedures are for single-channel operation with CH 1. Single-channel operation with CH 2 can also be achieved in a similar manner. Further operation methods are explained in the subsequent paragraph.

19

GOS-600G Service Manual

Figure 4-3

4.4 Dual-channel Operation


Change the VERT MODE switch to the DUAL states so that trace(CH 2) is also displayed(The explanation in the proceeding section is of CH 1). At this state of procedure, the CH 1 trace is the square wave of the calibrator signal and the CH 2 trace is a straight line since no signal is applied to this channel yet. Now, apply the calibrator signal to the vertical input terminal of CH 2 with the probe as is the case for CH 1. Set the AC-DC-GND switch to the AC state. Adjust vertical POSITION knobs (40) and (37) so that both channel signals are displayed as shown in Figure4-4. When in the dual channel operation(DUAL or ADD mode), the CH 1 or CH 2 signal must be selected for the triggering source signal by means of the SOURCE switch. If both CH 1 and CH 2 signals are in a synchronized relationship, both waveforms can be displayed stationary; if not, only the signal selected by the SOURCE switch can be stationary. If the TRIG. ALT push switch is engaged, both waveforms can be stationary(Do not use CHOP and ALT triggering source switch at the same time). Selection between CHOP mode and ALT mode is automatically made by the TIME/DIV switch shown in Figure 4-5. The 5mSec/DIV and lower ranges are used in the CHOP mode and the 2ms/DIV and higher ranges are used in the ALT mode. When the CHOP push switch is engaged, the two traces are displayed in the CHOP operation at all ranges. The CHOP operation has priority over the ALT operation.

20

GOS-600G Service Manual

Signal of CH1

Signal of CH2

Figure 4-4
4.5 ADD Operation
An algebraic sum of the CH 1 and CH 2 signals can be displayed on the screen by setting the VERT MODE switch to the ADD state. The displayed signal is the difference between CH 1 and CH 2 signals if the CH 2 INV push switch is engaged. For accurate addition or subtraction, it is a prerequisite that the sensitivities of the two channels are adjusted accurately at the same value by means of the VARIABLE knobs. Vertical positioning can be made with the POSITION knob of either channel. In view of the linearity of the vertical amplifiers, it is most advantage to set both knobs in their mid-positions.

4.6 X-Y Operation and EXT HOR Operation


When the TIME/DIV switch is set in the X-Y/EXT HOR state, the internal sweep circuit is disconnected and the trace in the horizontal direction is driven by the signal selected by the SOURCE switch. When the SOURCE switch is set to the CH 1 X-Y position, the oscilloscope operates as an X-Y scope with the CH 1 signal for the X-axis; when it is set to the EXT position, the oscilloscope operates in the EXT HOR(external sweep) mode.

21

GOS-600G Service Manual

X-Y Operation The X-Y operation is with CH 1 as X-axis and CH 2 as Y-axis. The bandwidth of the X-axis becomes DC to 1MHz(-3dB)(or DC to 2MHz for 652G, 653G, 658G) and the horizontal POSITION control is directly used as the X-axis POSITION control. For the Y-axis, the CH 2(X-Y) should be selected by the VERT MODE switch.

Y axis (CH2)

X axis (CH1)

Figure 4-6

Dual-Channel X-Y Operation

Figure 4-7
Note: When high frequency signals are displayed in the X-Y operation, pay attention to the frequency bandwidths and phase difference between X and Y-axis.

22

GOS-600G Service Manual

EXT HOR(external sweep) Operation The external signal applied through the EXT HOR terminal(23) drives the X-axis. The Y-axis is with any channels as selected by the VERT MODE switch. When the DUAL mode is selected by the switch, both CH 1 and CH 2 signals are displayed in the CHOP mode.

4.7 Triggering
Proper triggering is essential for efficient operation of an oscilloscope. The user must be thoroughly familiar with the triggering functions and procedures. (1) Functions of SOURCE switch: The displayed signal itself or a trigger signal which has a time relationship with the displayed signal is required to be applied to the trigger circuit to display a stationary signal on the CRT screen. source. CH 1 CH 2 : The internal trigger method which is used most commonly. : The signal applied to the vertical input terminal is branched off from the preamplifier and is fed to the trigger circuit through the VERT MODE switch. Since the triggering signal is the measured signal itself, a stable waveform can be readily displayed on the CRT screen. When in the DUAL or ADD operation, the signal selected by the SOURCE switch is used as the triggering source signal. : The AC power line frequency signal is used as the triggering signal. This method is effective when the measured signal has a relationship with the AC line frequency, especially for measurements of low level AC noise of audio equipment, thyristor circuits, etc. : The sweep is triggered with an external signal applied to the external trigger input terminal. An external signal which has a periodic relationship with respect to the measured signal is used. Since the measured signal is not used as the triggering signal, the waveforms can be displayed more independent than the measured signal. The SOURCE switch is used for selecting such a triggering

LINE

EXT

23

GOS-600G Service Manual

The above triggering source signal selection functions are shown collectively in the following table. VERT.MODE CH1 SOURCE CH1 CH2 ALT LINE EXT (2) Functions of COUPLING switch: This switch is used for selecting the coupling of the triggering signal to the trigger circuit in accordance with the characteristics of the measured signal. AC : This coupling is for AC triggering which is used most commonly. As the triggering signal is applied to the trigger circuit through an AC coupling circuit, stable triggering can be attained without being affected by the DC component of the input signal. The low-range cut off frequency is 10Hz (-3dB). When the ALT trigger mode is used and the sweep speed is slow, jitter may be produced. In such a case, use the DC mode. CH2 DUAL ADD

Triggered by CH1 signal Triggered by CH2 signal Alternately triggered by CH1 & CH2 Trigged by LINE signal Trigged by EXT TRIG input signal

HF REJ : The triggering signal is fed to the trigger circuit through an AC coupling circuit and a low pass filter (approximately 50kHz, -3dB). The higher components of the trigger signal are rejected and only the lower components are applied to the trigger circuit. TV : This coupling is for TV triggering, for observation of TV video signals. The triggering signal is AC-coupled and fed through the triggering circuits (level circuit) to the TV sync separator circuit. The separator circuit picks off the sync signal, which is used to trigger the sweep. Thus, the video signal can be displayed stably. Being linked to the TIME/DIV switch, the sweep speed is switched for TV-V and TV-H as follows: TV-V :0.5 s - 0.1 ms TV-H : 50 s - 0.1 s The SLOPE switch should be set to conform to the video signal as shown in Figure 4-8. : The triggering signal is DC-coupled to the trigger circuit. This mode is used when triggering is desired with the DC component of the triggering signal or when a signal with very low frequency or a signal with large duty cycle ratio is needed to be displayed

DC

24

GOS-600G Service Manual

Figure 4-8 (3) Function of SLOPE switchG This switch selects the slope (polarity) of the triggering signal as shown in Figure 4-9. + - : When set in the + state, triggering occurs as the triggering signal crosses the triggering lever in the positive-going direction. : When set in the - state, triggering occurs as the triggering signal crosses the triggering lever in the negative-going direction.

Slope "-" Range Slope "+" Range

Figure 4-9 Level

(4) Function of Level (LOCK) controlG The function of this control is to adjust the triggering level and display a stationary image. At the instant, the triggering signal has crossed the triggering level set by the control, the sweep is triggered and a waveform is displayed on the screen. The trigger level changes in the positive direction(upward) as this control knob is turned clockwise, and it changes in the negative direction(downward) as the knob is turned counter clockwise. The characteristic changes are as shown in Figure 4-10.

25

GOS-600G Service Manual DIV


4 3 2 1 120
O

60
O

30
O

Trigger level
4 3

30
O

60
O

120

Graticule scale 2 equivalent (DIV)


1

Level knob rotation angle Left Center Right

Figure 4-10
LEVEL LOCK: When LEVEL LOCK push switch is engaged, the triggering level is automatically maintained within the amplitude of the triggering signal, and stable triggering is made without requiring level adjustment (although jitter may not be suppressed when in the ALT mode). This automatic level lock function is effective when the signal amplitude on the screen or the input voltage of the external triggering signal is within the following range: 622G / 626G / 635G: 50 Hz-5MHz 5MHz- 35MHz : 1.0DIV (0.15V) or less : 2.0DIV (0.25V) or less5MHz- 20MHz

652G / 653G / 658G: 50Hz- 10MHz : 1.0DIV (0.15V) or less

10 MHz- 40MHz : 2.0DIV (0.25V) or less (5) Functions of HOLD OFF control: When the measured signal is a complex waveform with two or more repetition frequencies (periods), triggering with the above mentioned LEVEL control alone may not be sufficient to attain a stable waveform display. In such a case, the sweep can be stable synchronized to the measured signal waveform by adjusting the HOLD OFF time (sweep pause time) of the sweep waveform. The control covers at least one full sweep time for sweeps faster than 0.2s/DIV.

26

GOS-600G Service Manual


Complex waveform (digital signal) cycle

Highlighted parts are displayed

Figure 4-11(a)

Sweep waveform

Figure 4-11(a)

Adjusting the HOLD OFF time Figure 4-11(b)

Figure 4-11(b)
Figure 4-11(a) shows several different waveforms which overlapped on the screen, making the signal observation unsuccessful when the HOLD OFF knob is in the MIN state. Figure 4-11(b) shows the undesirable portion of the signal is held off. The same waveforms are displayed on the screen without overlapping.

4.8 Single-sweep Operation


Non-repetitive signals and one-shot transient signals can hardly be observed on the screen with the regular repetitive sweep operation. Such signals can be measured by displaying them in the single-sweep mode on the screen and photographing them. Measurement of non-repetitive signal: (1) Set the TRIGGER MODE to the NORM state. (2) Apply the measured signal to the vertical input terminal and adjust the triggering level. (3) Set the TRIGGER MODE to the SINGLE state (the three push-button switches are pushed out). (4) Press the SINGLE button. The sweep will run only for one cycle and the measured signal will be displayed only once on the screen.

27

GOS-600G Service Manual

Measurement of single-shot signal: (Except 622G & 635G) (1) Set the TRIGGER MODE to the NORM state. (2) Apple the calibration output signal to the vertical input terminal, and adjust the triggering level at a value corresponding to the predicted amplitude of the measured signal. (3) Set the TRIGGER MODE to the SINGLE state. instead of the calibration signal to the vertical input. (4) Depress the SINGLE button. The sweep circuit is now in the ready state and the READY indicator lamp will be turned on . (5) As the one-shot signal occurs in the input circuit, the sweep runs only for one cycle and the one-shot signal is displayed on the CRT screen. However, this cannot be done when the dual-channel ALT mode is in operation. For the dual-channel one-sweep operation, use the CHOP mode instead. Apply the measured signal

4.9 Sweep Magnification


When a certain part of the displayed waveform is needed to be expanded timewise, a faster sweep speed may be used. However, if the required portion is apart from the

starting point of the sweep, the required portion may run off the CRT screen. In such a case, push in the 10MAG button.

When this has been done, the displayed waveform will be expanded 10 times to the right and left with the center of screen as the center of expansion. The sweep time during the magnification operation is as followsG (Value indicated by TIME/DIV switch) 1/10 Thus, the unmagnified maximum sweep speed (0.1 s/DIV) can be increased with the magnification as followsG 0.1 s/DIV 1/10 = 10 s/DIV When the sweep is magnified and the sweep speed is above 0.1 s/DIV, the trace may become darker. In such a case, the displayed waveform should be expanded in the B sweep mode as explained in the subsequent paragraphs.

28

GOS-600G Service Manual

Figure 4-12

10 x magnification

Any part can be covered by means of POSITION control

4.10 Waveform Magnification with Delayed Sweep (653G / 658G only)


With sweep magnification of the preceding paragraph, although the magnification method is simple, the magnification ratio is limited to 10. With the delayed sweep method of this paragraph, the sweep can be expanded for a wider range from several times to several thousand times according to the ratio between A sweep time and B sweep time. As the measured signal frequency increases, the A sweep range for the non-expanded signal becomes higher whereas the available expansion ratio becomes smaller. Furthermore, as the magnification ratio becomes larger, the trace intensity becomes lower and the delay jitter increases. To cope with these situations, a continuously variable delay circuit and a triggering delay circuit are incorporated into the oscilloscope. (1) Continuous variable delay Set the HORIZ. DISPLAY MODE switch to A and display the signal waveform with the A sweep in the regular operation mode. Next, set the B TIME/DIV switch to a position several steps faster than that of the A TIME/DIV switch. After ensuring the B TRIGD button of the HORIZ. DISPLAY MODE switch is disengaged, engage the HORIZ. DISPLAY MODE switch to the A INTEN position. A portion of the displayed waveform will be accentuated as shown in Figure 4-14, indicating the state ready for delayed sweep. The portion of the accentuated brightness indicates the section corresponding to the B sweep time (DELAYED SWEEP). This portion is expanded on the B sweep.

29

GOS-600G Service Manual

The period from the start of the A sweep to that of the B sweep (the period to the start of trade accentuation) is called SWEEP DELAY TIME. This period is continuously variable by means of the DELAY TIME POSITION knob. Next, change the HORIZ.DISPLAY MODE switch to the B position. The B sweep time will be expanded for the full span of the CRT screen as shown in Figure 4-15. The B sweep time is set by the B TIME/DIV switch, the magnification ratio becomesG Magnification = (A TIME/DIV indication) / (B TIME/DIV indication) (2) Triggering delay When the display waveform is magnified by 100 or higher in the above-mentioned continuous delay method, delay jitter is produced. To suppress the jitter, the triggering delay method may be used. With the triggering delay, delay jitter is reduced by triggering the B sweep again, after a sweep delay time as effected by the continuous delay method has elapsed. For this operation, the A trigger circuit continues to operate even after the B TRIGD button is engaged and the B sweep is triggered by the triggering pulse. Therefore, even when the delay time is continuously varied by turning the TIME DELAY POSITION knob, the starting point of the sweep moves discretely, not continuously. In the A INTEN mode, this operation is characterized by the discrete shifts of the brightness-accentuated section of sweep across the CRT screen; while in the B mode this section remains stationary.

30

GOS-600G Service Manual

Figure 4-13

A Sweep
Sweep delay time

Figure 4-14 HORIZ DISPLAY B INTEN

B Sweep

This start point moves discretely B trigger level A INTEN

Figure 4-15

4.11 Readout Function (626G and 658G)


The selected sensitivity, input, sweep time, etc.. are displayed in the positions as shown in Figure 4-16. NOTEG The CRT will not show any trace or spot when the TRIGGER MODE is in NORM state. To observe the signals, depress the AUTO button.

31

GOS-600G Service Manual

Figure 4-16
(7)

(a)
100 90

(b)

(c)

10 0%

(a)(b) (c) (a) (b) (c) (d) (a)(b) (a)(b) (c) (d)

(6)

(1)

(3)

(2)

(4)

(5)

CH1 Display When the VERT MODE switch is at CH1, DUAL or ADD state, the set values of CH1 are displayed at (1). However, these values are not shown when the VERT MODE is at CH2. (a) (b) (c) (d) : P10 sign is shown when the probe 10 is setted

: > sign is shown when the V/DIV VAR. is at UNCAL position. : Display the selected sensitivity from 1mV to 5V. (Probe x10 from 10mV to 50V) x sign is displayed when the X-Y button is set and the VERT MODE is at CH2. At DUAL X-Y mode, y1 is displayed.

CH2 Displayed Set values of CH2 signal are displayed at (2) when the VERT MODE is at CH2, DUAL or ADD. They are not displayed in the CH1 mode. (a) (b) (c) (d) : P10 sign is shown when the probe 10 is setted.

: > sign is shown when the V/DIV is at UNCAL position. : Display the selected sensitivity from 1mV to 5V. (Probe x10 from 10mV to 50V) : y sign is shown at X-Y mode.y2 sign is shown at DUAL X-Y mode.

32

GOS-600G Service Manual

ADD (SUB) & CH2 INV Display The ADD, SUB and INV functions are displayed at (3). (a) (b) : + is shown when the VERT MODE is at ADD position, then the inputs CH1 & CH2 are algebraically summed. : A sign is displayed when the VERT MODE is at CH2 or DUAL and the CH2 INV button is engaged. When the CH2 INV button is pushed in, the subtraction of CH2 from CH1 is in function.

TIME Display The sweep time is displayed at (4). The A sweep time is shown at the under row, the B sweep time is shown at the upper row. (B sweep for 658G only) (a) (b) (c) : A and B are shown at A and B sweep time. : = is shown normally. * sign is displayed when the x10 MAG button is pushed in. > sign is displayed when the SWP. UNCAL button is engaged. : shows the selected sweep time from 10ns to 0.5s. An X-Y is displayed when the X-Y button is pushed in.

CHOP / ALT Display The CHOP or ALT is displayed at (5) when the VERT MODE is set to DUAL. When X-Y button is engaged, an XEXT is displayed. TV-V / TV-H Display The TV-V or TV-H are displayed at (6) when the TRIG. COUPLING is set to TV position.

Cursor Measured Value Display


The relative measured values of the seven functions are displayed at (7). (a) : Shows each of seven functions ( V, V%, VdB, T, 1/ T, DUTY, PHASE ) which may be selected by the CURSOR FUNCTION button. The V function provide different V ( V1, V2, V12, Vy, Vy1 ) according to the following tableG

33

GOS-600G Service Manual

VERT. MODE CH1 CH1 TRIG. SOURCE CH2 LINE EXT X-Y *1 Vy Vy1 *1 V1 V2 V2 CH2 DUAL V1 V12 ADD

NOTEG *1G When X-Y mode is not set at correct position, the error message X-Y mode error is shown. (b) : In the V function, a + or - polarity is shown. + when the (delta) cursor is above the (REF.) cursor; - when the (delta) cursor is below the (REF.) cursor. : Display the measured value and units of the seven cursor measurement functions.

(c)

V G

0.0V~40.0V ( 400V for PROBE x10) NOTEG When the V/DIV VAR. is set to uncalibrated position, or when the VERT MODE is at ADD but the CH1 & CH2 sensitivities on V/DIV are not the same, the measuring unit will be value is displayed in division(0.00 to 8.00 div.) instead. V% G 0.0%~160% (5 div. = 100% reference) VdBG -41.9dB~+4.08dB (5 div. = 0dB reference) VdB = 20 log V(div.) / 5 div. V(div.)G measured difference division value. 0.0nS~5.00S NOTEG When the SWP UNCAL button is pushed in, the measured value is T G displayed in divisions (0.00 to 10.00 div.). 200.0mHz~2.500GHz 1/ T G NOTEG When the SWP UNCAL button is pushed in or two cursors are overlap, the unknow value displays ???? . DUTY G 0.0%~200.0% (5 div. = 100% reference) PHASEG 0.0~720 (5 div. = 360 reference) NOTEG Except V(%,dB), the other functions ( T, 1/ T, DUTY, PHASE) are selected and then the X-Y button is engaged, the unknown value displays ???? .

34

GOS-600G Service Manual

4.12 Calibration of Probe


As explained previously, the probe makes up a wide range attenuator. Unless phase compensation is properly done, the displayed waveform is distorted causing measurement errors. Therefore, the probe must be properly compensated before use. Connect the probe BNC to the INPUT terminal of CH1 or CH2 and set VOLTS/DIV switch at 50mV. Connect the probe tip to the calibration voltage output terminal and adjust the compensation trimmer on probe for optimum square wave (minimum overshoot, rounding off and tilt). Refer to 4-18 G

(a) Correct compensation

(b) Over compensation Figure 4-18

(c) Insufficient compensation

35

GOS-600G Service Manual

5. MAINTENANCE
WARNING
The following instructions are for use by qualified personnel only. To avoid electrical shock, do not perform any servicing other than in the operating instructions unless you are qualified to do so.

5.1 Fuse Replacement


If the fuse blows, the power lamp indicators will not light and the oscilloscope will not operate. The fuse should not normally open unless a problem has developed in the unit. Try to determine and correct the cause of the blown fuse. The replace only with a fuse of the correct rating and type (see page 7 ) The fuse is located on the rear panel (see fig. 4-2).

WARNING. For continued fire protection. Replace fuse only with 250V fuse of the specified type and rating, and disconnect power cord before replacing fuse.

5.2 Line Voltage Conversion


The primary winding of the power transformer is tapped to permit operation from 100,120,220,or 230VAC 50/60Hz line voltage. Conversion from one line voltage to another is done by changing the line voltage selector switch as shown in Fig. 4.2. The rear panel identifies the line voltage to which the unit was factory set. To convert to a different line voltage, perform the following procedure: (1) Make sure the power cord is unplugged. (2) Change the line voltage selector switch to the desired line voltage position. (3) A change in line voltage may also require a corresponding change of fuse value. Install the correct fuse value as listed on rear panel.

36

GOS-600G Service Manual

5.3 Cleaning
To clean the oscilloscope, use a soft cloth dampened in a solution of mild detergent and water. Do not spray cleaner directly onto the oscilloscope because it may leak into the cabinet and cause damage. Do not use chemicals containing benzine, benzene, toluene, xylene, acetone, or similar solvents. Do not use abrasive cleaners on any portion of the oscilloscope.

37

GOS-600G Service Manual

6.BLOCK DIAGRAM
CH1 SIGNAL OUTPUT 652G/ 653G/ 65G/ 8

+12KV CH1 (X) INPUT


CH1 ATT CH1 PREAMP CHARACTER GENERATOR

626G/6G 85
(MICROPROCESSOR) READOUT CONTROL CIRCUIT CONTROL PANEL

CH1 TRIG (X-AXIS) PICKUP AMP

VERT SIGNAL

VERT MODE SWITCH

VERTICAL OUTPUT AMP

CH2 (Y) INPUT


CH2 ATT CH2 PREAMP SWITCHING LOGIC Z-AXIS AMP TRIG SIGNAL TRIG SWITCH Z-AXIS INPUT CRT CIRCUIT

- HV

CH2 TRIG PICKUP AMP

H.V. SUPPLY X-AXIS SIGNAL

CH1 CH2

TRIG INPUT AMP

TRIG GENERATOR

(A) SWEEP GENERATOR

HORIZONTAL SWITCHING

HORIZONTAL OUTPUT AMP

EXT TRIG & EXT HOR INPUT

REE RUN F SIGNAL

EXT

AUTO CIRCUIT

(B) SWEEP GENERATOR 653G/6G 85

TO EACH BLOCK

POWER SUPPLY

AC LINE 060 Hz 5 /

LINE
LINE TRIG

2Vp-1KHz /p SQUARE-WAVE

CALIBRATOR

BLOCK DIAGRAM

38

GOS-65XG ADJUSTMENT PROCEDURES

GOS-65XG ADJUSTMENT PROCEDURES WARNING


The following instructions are for qualified service personnel only. To avoid electrical shock, do not perform servicing unless you are qualified to do so. High voltage appears when covers are removed and the unit is in operation. Remember that high voltage may be retained indefinitely on high voltage capacitors. In addition, the AC line voltage is present on line voltage input circuit (including the fuse holder, line cord receptacle, and POWER switch) whenever the instrument is plugged into an AC outlet, even when it is turned off. Unplug the oscilloscope and discharge high voltage capacitors before performing service procedures unless it is necessary to have its power turned on.

GENERAL
This unit was carefully checked and calibrated at the factory before shipment. After the oscilloscope has been used for a period of time, it should be calibrated to maintain instrument accuracy. Calibration is also required if repairs have been made in a circuit which may affect the former calibration. Keep in mind that some calibration adjustments require high precision test instruments. Those adjustments should be attempted only if the proper test equipment is available and you are experienced in using it. The unit should be calibrated at an ambient temperature of 23 degrees C ( 5 degrees C). Connect the unit to AC line voltage, 50/60 Hz line source and allow it to warm up for at least 30 minutes before any adjustment. The following test equipment required complete calibration:: GOODWILL GDM-8045G Multi-meter or equivalent. TEKTRONIX CG-5001 & SG-503 Oscilloscope Calibrator or equivalent. FLUKE High Voltage Probe or equivalent with capability of measuring at least 12kV. NOTE All voltage measurements are to be taken with respect to earth (chassis) ground unless otherwise noted. The adjusting procedures MUST be done in sequence. Unit Under Test position must direct to South-North direction.

---- 1 ----

GOS-65XG ADJUSTMENT PROCEDURES

GOS-65XG ADJUSTMENT PROCEDURES


ITEM 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 Page

SUPPLY VOLTAGE ADJUSTMENT & CHECK 3 INTENSITY ADJUSTMENT. 3 B INTEN CHECK (EXCEPT GOS-652G).3 FOCUS ADJUSTMENT. 4 ASTIG ADJUSTMENT... ILLUM CHECK. 4 TRACE ROTATION CHECK... 4 STEP DC BAL. ADJUSTMENT... 4 MAG DC BAL. ADJUSTMENT... 5 VAR. DC BAL. ADJUSTMENT.. 5 CH1 & CH2 VERT POS CENTRE & CH2 INV. ADJUSTMENT.. 5 ADD BAL. ADJUSTMENT.. 6 X10 MAG. CENTRE ADJUSTMENT.. 6 A-START ADJUSTMENT. 6 B-START SWEEP ADJUSTMENT (EXCEPT GOS-652G). 6 X-Y POSITION ADJUSTMENT... 7 TIME/DIV ACCURACY ADJUSTMENT. 7 B TIME ACCURACY ADJUSTMENT (EXCEPT GOS-652G)... 7 SENSITIVITY ADJUSTMENT. 8 MAG. SENSITIVITY ADJUSTMENT.. 8 PHASE COMPENSATION ADJUSTMENT 9 X-GAIN ADJUSTMENT... 9 CAPACITANCE ADJUSTMENT. 9 1MHZ OVERSHOOT ADJUSTMENT. 10 FREQUENCY RESPONSE CHECK. 10 TRIG. DC OFFSET ADJUSTMENT. 11 CH1 OUTPUT DC OFFSET ADJUSTMENT... 11 TRIG-SLOPE BAL. ADJUSTMENT. 12 TRIGGERING CENTRE ADJUSTMENT.12 CURSOR READOUT ACCURACY ADJUSTMENT.. 13 CAL FREQUENCY ADJUSTMENT. 13

---- 2 ----

GOS-65XG ADJUSTMENT PROCEDURES

SUPPLY VOLTAGE ADJUSTMENT & CHECK USE GDM-8045G DIGITAL MULTIMETER & 149-10A H.V.DIGITAL METER INSTRUMENT a). Connect the METER between test point on 03 PCB and chassis. Check the voltage accordingly as shown in the following table. Test Point Pin 1 of J603 Pin 2 of J603 Pin 3 of J603 Pin 5 of J603 Voltage + 12 V - 12 V + 5V + 145 V Adjust & Check VR601 CHECK CHECK CHECK Tolerance +11.95V ~ +12.05V -11.80V ~ -12.20V +4.75V ~ +5.25V +140V ~ +150V

b). Connect the DC HIGH VOLTAGE METER to -H.V point. Spec : - 1725V ~ - 1875V WARNING ! Extreme caution must be taken while working in the high voltage section so as to prevent personal injury.

INTENSITY ADJUSTMENT a). Set HORIZ DISPLAY MODE to A, A TIME/DIV to 1ms/Div, Rotate HOLD OFF VR fully anti clockwise, rotate A INTENSITY VR until write indicator drop at center of L, AC-GND-DC switch to GND. (B INTEN VR fully anti clockwise. Except GOS-652G). b). Adjust VR603 (SUB INTEN) until a dim trace appears on CRT display. c). Check INTEN VR when rotate clockwise sweep will become brighten & rotate anti clockwise sweep will disappear.

B INTEN CHECK(EXCEPT GOS-652G) a). Set: A TIME/DIV to 1ms/Div, HORIZ DISPLAY MODE to A INT, B TIME/DIV to 0.5ms/DIV & DELAY TIME POSITION VR rotate fully anti clockwise. b). Check when B INTEN VR rotate clockwise, sweep INTEN A will become dim but sweep B INTEN look like no change. (B sweep length is half of A sweep).

---- 3 ----

GOS-65XG ADJUSTMENT PROCEDURES

FOCUS ADJUSTMENT a). Rotate FOCUS VR to center position, set X-Y, rotate HOR POS VR until a spot at center. (CURSOR READOUT ON. GOS-658 only) b). Adjust VR602 (SUB FOCUS) until optimum spot is obtained.

ASTIG ADJUSTMENT a). Adjust VR604 (ASTIG) until when rotating FOCUS VR the best round spot obtained. (CURSOR READOUT must become a good focus. GOS-658 only)

ILLUM CHECK a). Rotate ILLUM VR clockwise will illuminate on CRT screen, when rotate anti clock light will disappear.

TRACE ROTATION CHECK. a). Set TIME/DIV to 1ms/Div, AC-GND-DC to GND, rotate HOR POS VR to left until sweep start from first graticule & rotate VERT POS VR until sweep at center. b). When rotate TRACE ROTATION VR fully clockwise & anti clockwise, sweep will move UP & DOWN. Spec : +/- 0.2 Div above.

STEP DC BAL ADJUSTMENT a). Set AC-GND-DC (both) to GND, VOLTS/DIV VAR. (both) to CAL.VERT MODE select CH1 (CH2), rotate VOLT/DIV switch to 10mv. (both) b). Lets sweep at center of graticule line, rotate VOLT/DIV to 5mV & adjust CH1: VR101 (CH2: VR201) until the sweep at center. Repeat above step until in spec. Spec : +/- 0.05 Div c). Check VOLT/DIV from 5mV~5V range. Spec : +/- 0.05 Div

---- 4 ----

GOS-65XG ADJUSTMENT PROCEDURES

MAG DC BAL ADJUSTMENT a). Rotate VOLT/DIV switch to 5mV(both). VERT MODE select CH1 (CH2). b). Lets sweep at center of graticule line. Rotate VOLT/DIV switch to 2mV & adjust CH1: VR104 (CH2: VR220) until the sweep at center. Repeat above step until in spec. Spec : +/- 0.2 Div c). Check VOLT/DIV from 2mV~1mV range.

10

VAR DC BAL ADJUSTMENT a). Rotate VOLT/DIV switch to 10mV (both). VERT MODE select CH1 (CH2). b). Rotate the VAR VR fully anti clockwise, rotate VERT POS VR until the sweep at center graticule line & rotate VAR VR fully clockwise, then adjust CH1: VR106 (CH2: VR206) until the sweep at center. Repeat above step until in spec. Spec : +/- 0.1 Div c). Rotate VOLT/DIV switch to 2mV & check when rotate CH1:VAR VR(CH2) fully anti clockwise, the deflection must be in spec. Spec : +/- 0.2 Div

11

CH1 & CH2 VERT POS CENTRE & CH2 INV ADJUSTMENT *(UUT must put horizontally) a). Set AC-GND-DC to GND.VOLT/ DIV (both) switch to 10mV & VERT POS knob (both) to mid position. Set VERT MODE to CH2. b). Press INV switch & adjust VR213 until the trace stays stationary irrespective to pressing or releasing. Spec : +/- 0.2 Div c). Adjust VR303 until CH2 sweep at center of graticule line, check again INV. d). VERT MODE selects to CH1 & adjusts VR113 until sweep at center. Spec : +/- 0.2 Div

---- 5 ----

GOS-65XG ADJUSTMENT PROCEDURES

12

ADD BAL ADJUSTMENT *(UUT must put horizontally) a). Set AC-GND-DC to GND (BOTH), VOLT/DIV to 10mV (both), VERT MODE to CH1 (CH2). b). Rotate VERT POS VR until sweep at center of graticule line (both) c). VERT MODE select to ADD. d). Adjust VR301 until sweep at center of graticule line. e). Check VERT MODE CH1, CH2.DUAL, ADD. Sweep must at center of graticule line. Spec : +/- 0.1 Div

13

X10 MAG CENTRE ADJUSTMENT *(UUT must put horizontally) a). Set T/D to 1ms/Div, press X10 MAG sw. & rotate HOR POS VR until sweep start point at center. b). Release X10MAG sw. & adjust VR502 until sweep start point at center. Spec : +/- 0.3 Div

14

A-START ADJUSTMENT *(UUT must put horizontally) a). Set A TIME/DIV to 1ms/Div, rotate HOR POS VR knob point to center. b). Adjust VR408 to obtain a start sweep at first vertical graticule line. Spec : +/- 0.3 Div

15

B-START SWEEP ADJUSTMENT(EXCEPT GOS-652G) *(UUT must put horizontally) a). Set HOR POS to A & A T/D to 0.5ms/div . b) Rotate HOR POS VR until a start sweep at first vertical graticule line. c). Set HOR DISPLAY to B (release B TRIGD), B T/D to 0.2ms/div, DELAY TIME POSITION VR fully anti clockwise. d). Adjust VRC01 to obtain a start sweep at first vertical graticule line.

---- 6 ----

GOS-65XG ADJUSTMENT PROCEDURES

e). HOR DISPLAY set to A INT, DELAY TIME VR fully clockwise. f). Adjust VRA06 until B SWEEP inten dim trace appear on CRT SCREEN. 16 X-Y POSITION ADJUSTMENT *(UUT must put horizontally) a). Set T/D to 1ms/Div & rotate HOR POS VR until sweep start from first vertical graticule line. b). Press X-Y sw. & adjust VR401 until the spot at center on graticule line. Spec : +/- 0.2 Div 17 A TIME/DIV ACCURACY ADJUSTMENT USE SC-340 SCOPE CALIBRATION INSTRUMENT a). Set VERT MODE to CH1, Release SWP UNCAL & TIME/DIV to 1ms/Div. b). Input Time Marker with TIME/DIV according to scope setting below. c). Use 2nd marker & 9th marker for reference. Spec : +/- 0.2 Div SC-340 SETTING 1ms/Div 1us/Div .1us/Div 18 T/D SETTING 1ms/Div 1us/Div .1us/Div ADJUST VR VR405 VR404 VR802

B TIME/DIV ACCURACY ADJUSTMENT (EXCEPT GOS-652G) USE SC-340 SCOPE CALIBRATION INSTRUMENT a). HOR DISPLAY A, A T/D 2us/Div & HOR DISPLAY set to B, T/D set 1us/div, DELAY TIME POS. fully anti clockwise, input 1us/div pulse train to CH1. b). Adjust VRC02 until pulse train align with CRT vertical line. c). Set HOR DISPLAY A, A T/D set 0.2us/div & HOR DISPLAY set to B.B T/D set 0.1us/div, input 0.1us/div pulse train. d). Adjust VRC03 until pulse train align with CRT vertical line. Spec : +/- 0.2 Div

---- 7 ----

GOS-65XG ADJUSTMENT PROCEDURES

19

SENSITIVITY ADJUSTMENT USE SC-340 SCOPE CALIBRATION INSTRUMENT a). Set VOLT/DIV (both) to 10mv,VOLT/DIV (both) to CAL, AC-GND-DC to DC, TIME/DIV to 0.2ms/Div, VERT MODE to CH1. b). Input 1KHz, 60mv Vp-p square wave to CH1 & CH2 terminal. c). Adjust VR108 until amplitude at 6.06 Div. (or adjust 6.00 div by setting calibrator 60mV1%). (Adjust to exactly 6.00 div if adjust with cover) (See fig 1) d). Set VERT MODE to DUAL, AC-GND-DC to GND (both), rotate CH1 & CH2 VERTICAL POSITION VR until two sweep at vertical center graticule line. e). Set VERT MODE to ADD, AC-GND-DC to DC (both), Push in CH2 INV

switch. f). Adjust VR208 until get one sweep. (See fig 2) Spec : +/- 0.05Div

Figure 1 20 MAG SENSITIVITY ADJUSTMENT

Figure 2

USE SC-340 SCOPE CALIBRATION INSTRUMENT a). Set VOLT/DIV (both) at 2mV,VOLT/DIV VAR at CAL, AC-GND-DC at DC & VERT MODE set CH1 (CH2). b). Input square wave 1KHz, 12mvVp-p to CH1 (CH2) input terminal. c). Adjust VR102 (VR202) until 6.06 Div (or adjust 6.00 div by setting calibrator 12mV1%). (Adjust to exactly 6.00 div if adjust with cover) Spec : +/- 0.1 Div

---- 8 ----

GOS-65XG ADJUSTMENT PROCEDURES

21

PHASE COMPENSATION ADJUSTMENT USE SC-340 SCOPE CALIBRATION INSTRUMENT a). Set VOLT/DIV (both) set at 0.1V,VOLT/DIV at CAL, AC-GND-DC to DC & VERT MODE CH1 (CH2). b). Input square wave 1KHz, 0.6Vp-p to CH1 (CH2) input terminal. c). Adjust CH1, 1/10 VC102 (CH2, 1/10 VC202) until obtain an ideal waveform. d). VOLT/DIV (both) set at 1V. Input 6Vp-p to CH1 (CH2) input terminal. e). Adjust CH1, 1/100 VC104 (CH2, 1/100 VC204) until obtain an ideal waveform. Spec : +/- 0.1 Div

22

X-GAIN ADJUSTMENT USE SC-340 SCOPE CALIBRATION INSTRUMENT a). Set VOLT/DIV (both) to 10mV, set X-Y mode, TRIG-COUPLING to AC, ACGND-DC to AC, TRIG-SOURCE to CH1, VERT MODE to CH2, VOLT/DIV VAR (both) to CAL. b). Input 1KHz, 60mVp-p square wave to CH1. c). Adjust VR109 until 6 Div. d). Input 1KHz 12mVp-p square wave e). VOLT/DIV set to 2mV & check sweep length becomes 6 Div. Spec : +/- 0.18 Div

23

CAPACITANCE ADJUSTMENT USE CAPACITANCE METER INSTRUMENT a). Set V/D (both) to 10mv,VERT MODE to CH1 (CH2) b). Input CAPACITANCE METER BNC to CH1 (CH2)& read capacitance value for reference. c). Set V/D to 0.1V (both), adjust CH1=VC101 (CH2=VC201) until get same reading with reference value. Spec : 25pf +/- 3pf d). Set V/D to 1V (both), adjust CH1=VC103 (CH2=VC203) until get same reading with reference value. Spec : 25pf +/- 3pf

---- 9 ----

GOS-65XG ADJUSTMENT PROCEDURES

24

1MHz OVERSHOOT ADJUSTMENT(USE 50 LOAD) USE PG506 CALIBRATION GENERATOR INSTRUMENT a). Set VOLT/DIV to 10mV(both), TIME/DIV 0.2s/Div, AC-GND-DC to DC (both), VERT MODE to CH1 (CH2), TRIG-SOURE to CH1(CH2). b). Input 1MHz square wave to CH1 (CH2) input terminal. Set the frequency amplitude until 5 div on screen. c). Adjust 1ST COMP.VC105 (CH2=VC205) & 2ND COMP. VC106

(CH2=VC206) to maximum overshoot. d). Adjust 1ST COMP.VR103 (CH2=VC203) to specification. ~ Adjust 3RD COMP.VR302 & VC301 for a flat-off & roll-off on the leading edge. ~ Adjust 2ND COMP.SVR for a flat-top & roll-off on the leading edge. ~ Adjust 2ND COMP.SVC for flat-top & roll-off. Spec : +0.17 ~ -0.17 div (+/-3.3% for 10mv), other range: +6.7% / -6.7% (+/-0.3 div) . 0.17 Div 0.17 Div

25

FREQUENCY RESPONSE CHECK (USE 50 LOAD) USE SG-503 LEVELED SINEWAVE GENERATOR INSTRUMENT a). Common mode rejection ration checks (at 10Mhz): b). Connect the sine wave generator to CH1& CH2 input terminal, set the frequency 10MHz & the output for display of 8 Div. Press in CH2 INV, VERT MODE= ADD. Check AC Vp-p. If out of spec, repeat step 24. Spec : 0.4Div c). Set freq 50khz, o/p amplitude 5 div, connect to CH1 (CH2) input terminal. Increase the frequency to 53MHz. Check the amplitude Spec : 3.6~5Div

---- 10 ----

GOS-65XG ADJUSTMENT PROCEDURES

d). Repeat frequency response check for 0.1V/Div & 1V/Div range. (Repeat step 24 if necessary) e). Set VOLT/DIV to 2mV (both), set frequency to 50KHz & the output amplitude for display of 5 Div, check frequency between 5MHz~10MHz. Spec : 3.6~5.3 Div f). Set input frequency to 18MHz. Spec : 3.6~5.3 Div 26 TRIG DC OFFSET ADJUSTMENT USE 33120A 15MHZ FUNCTION GENERATOR INSTRUMENT *(UUT must put horizontally) a). Set VOLT/DIV to 10mV(both), AC-GND-DC to DC (both), VERT MODE set CH1 (CH2), TRIG SOURE CH1(CH2), TRIG COUPLING to AC. b). Move sweep to center of vertical graticule line & input 1KHz sine wave to CH1 (CH2) and amplitude show 8 Div on screen. c). Rotate TRIG-LEVEL until trigger point start at the center. d). TRIG-COUPLING from AC to DC position & Adjust VR110 (CH2=VR210) until trigger point start from same position. e). Check TRIG-COUPLING from AC to DC the start point must be center. Spec : +/- 0.1 Div 27 CH1 OUTPUT DC OFFSET ADJUSTMENT a). Set CH1 & CH2 VOLT/DIV to 10mV, AC-GND-DC to GND, VERT MODE to BOTH, TRIG-SOURCE to CH1. b). Rotate CH1 & CH2 VERT POSITION to center of CRT & connect CH1 OUTPUT to CH2 input by a cable BNC TO BNC. c). Release CH2 GND switch & adjust VR112 until sweep at center of graticule line. (see fig below) Spec : +/- 0.22 Div

---- 11 ----

GOS-65XG ADJUSTMENT PROCEDURES

28

TRIG-SLOPE BAL ADJUSTMENT USE 33120A 15MHZ FUNCTION GENERATOR INSTRUMENT a). Set TIME/DIV to 1ms/Div, VERT MODE to CH1, Press-in LOCK switch, Pressin SLOPE switch & AC-GND-DC to GND. b). Adjust VR403 Use GDM-8045G DIGITAL MULTIMETER to measure voltage at Q414(C) Spec : 1.40+/- 0.02V (Release GND sw, input 1KHz 0.4 Div sinewave, check +/- slope both can sync.)

29

TRIGGERING CENTRE ADJUSTMENT USE 3310A 15MHZ FUNCTION GENERATOR INSTRUMENT a). Set TIME/DIV to 1ms/Div, TRIG-LEVEL unlock, AC-GND-DC to AC, input 1KHz 8 Div sinewave to CH1 input terminal. (make sure AC-GND-DC at GND, sweep at center of graticule line.) b) Set LEVEL VR knob at center & adjust VR407 until when slope +/- trigger point deviation must be same. (see fig below) Spec: LOCK: A, B1.5Div, UNLOCK: A, B0.2Div & A-B0.2Div

A B

---- 12 ----

GOS-65XG ADJUSTMENT PROCEDURES

30

CURSOR READOUT ACCURACY ADJUSTMENT (GOS-658G ONLY) a). Press CURSOR ON/OFF & TRACK switch at same time, CALIBRATION READOUT will appear. b). Adjust VRA01 for X-GAIN, VRA02 for X-CENT, VRA03 for Y-GAIN, VRA04 for Y-CENT, adjust until get 6x8Div. Spec : +/- 0.5% d). VERT MODE select from DUAL to ADD, CHARATOR will move up/down, adjust VRA05 until the CHARATOR cant move. Spec : < 0.02 Div

31

CAL. FREQUENCY ADJUSTMENT USE GFC-8130G INTELLIGENT COUNTER a). Measure the CAL. FREQUENCY by the counter. b). Adjust VR701 until frequency 1KHz. Spec : 0.980KHz~1020KHz.

---- 13 ----

GOS-62XG PRE-ADJUST CHECKING PROCEDURES


1 HIGH-POT TEST a) At 120V : AC 1.2KV,<3.5mV.Time 1minute. b) At 220V : AC 2KV,<3.5mV.Time 1minute. INTENSITY CHECK a) Rotate Intensity clockwise & anti clockwise,sweep will become brightness & disappear. FOCUS CHECK a) Press X-Y Mode,Rotate FOCUS VR untill optimum spot is obtained. READOUT INTEN CHECK(GOS-626G ONLY) a) Rotate READOUT VR fully clockwise READOUT will become brightness & when fully anti clockwise READOUT will disappear. TRACE ROTATION CHECK a) When rotate TRACE ROTATION VR clockwise & anti clockwise,the START point & END point of sweep will move UP & DOWN. ILLUM CHECK a) Rotate ILLUM VR clockwise will illuminated on CRT screen,when rotate anti clockwise light will disappear. CH1 & CH2 POSITION CHECK a) Rotate CH1 & CH2 POSITION VR clockwise & anti clockwise,sweep will move up & down. HORIZONTAL POSITION CHECK a) Rotate HORIZONTAL POSITION VR clockwise & anti clockwise,sweep will move left & right. HOLD OFF CHECK a) Rotate HOLD OFF VR clockwise,sweep intensity will change. CH1 & CH2 VOLT/DIV CHECK a) Input CAL frequency to CH1 & CH2 input terminal. b) Rotate CH1 & CH2 VOLT/DIV switch,amplitude will change.(CH1 & CH2 readout will change at the same time.-GOS-658G only)

10

11

GOS-65XG PRE-ADJUST CHECKING PROCEDURES


11 TIME/DIV CHECK a) Rotate TIME/DIV switch from 0.5s/Div to 0.1us/Div,sweep start from slow to fast.(Readout will change at the same time.-GOS-658G only) CHOP & ALT CHECK a) VERT MODE set DUAL,T/D=0.5s/Div~5ms/Div will appear CHOP sweep(see fig 1),T/D=2ms/Div~0.1us/Div will appear ALT sweep(see fig 2) & Press-in CHOP sw. ALT will change to CHOP sweep.(Readout will change at the same time.-GOS-658G only) (Fig 1) (Fig 2)

12

13

ADD CHECK a) When select from DUAL to ADD mode,CH1 & CH2 sweep will ADD to become one sweep.VERTICAL POS VR of CH1 & CH2 can control the sweep. (Readout will display [ + ].GOS-658G only) CH2 INVERT CHECK a) Input CAL. frequency to CH1 input terminal. b) When press-in INVERT sw.phase will invert 180 degree.(Readout will display [ ] GOS-658G only) X10 MAG CHECK a) When press-in X10MAG sw. sweep will become longer.(Readout will change, example display [A=1ms],when press-in X10MAG sw.will display [A.1ms] GOS-658G only) X-Y MODE CHECK a) Press-in X-Y sw.& the spot will appear.(Readout display X-Y,GOS-658G only) AUTO/ NORM/SGL. MODE CHECK a) Press-in AUTO sw. sweep will appear. b) Set to NORM mode, sweep will disappear but when rotate LEVEL VR sweep will appear. c) Press-in SGL sw. sweep will appear at that moment.

14

15

16

17

12

GOS-65XG PRE-ADJUST CHECKING PROCEDURES


18 COUPLING TV-V & TV-H READOUT CHECK(GOS-658G ONLY) a) Select COUPLING to TV position & rotate T/D sw. from 0.5s/Div~0.1ms/Div (Display TV-V)from 50us/Div ~ 0.1us/Div(Display TV-H) CH1/CH2 VAR.& SWP UNCAL CHECK a) Input CAL frequency to CH1 & CH2 input terminal. b) Rotate CH1 & CH2 VAR VR will reduce the amplitude of waveform. c) Press-in SWP UNCAL sw.& rotate the VAR VR will change the waveform length.(Readout of CH1/CH2 will display [<] & SWP UNCAL display from[=] to [<].-GOS-658G only) READOUT CALIBRATION CHECK(GOS-658G ONLY) a) Adjust all VR for calibration 6X8Div. i) VRA01 for CHR-X GAIN ii) VRA02 for CHR-X POSITION iii) VRA03 for CHR-Y GAIN iv) VRA04 for CHR-Y POSITION CURSOR READOUT CHECK(GOS-658G ONLY) a) Press CURSOR ON/OFF, CURSOR READOUT will Display,Press CURSOR FUNCTION readout function will change & Press TRACK change each other. b)Rotate CORSOR POSITION sw. for move grid line & changing readout at same time. c) Press TRACK sw. & rotate CURSOR POSITION sw. will Display [ Pin ]

19

20

21

13

Das könnte Ihnen auch gefallen