Beruflich Dokumente
Kultur Dokumente
Page 1 of 16
Service Program for imc Measurement Devices Calibration, Adjustment, Maintenance, Repair and Miscellaneous Service
Page 2 of 16
Service Program for imc Measurement Devices Calibration, Adjustment, Maintenance, Repair and Miscellaneous Service
7.
12. 13.
14.
15.
All test equipment used is subject to our test equipment monitoring; the calibrators (KNICK S 252, Wavetek 9100, FLUKE 5500 A, Burster Kelvimat, HBM bridge calibrator et al) regularly undergo external calibration (generally including a manufacturer's calibration certificate). At all times we keep calibrators having a DKD calibration certificate at the ready.
Page 3 of 16
Service Program for imc Measurement Devices Calibration, Adjustment, Maintenance, Repair and Miscellaneous Service
Page 4 of 16
Service Program for imc Measurement Devices Calibration, Adjustment, Maintenance, Repair and Miscellaneous Service
Service Program for imc Measurement Devices Calibration, Adjustment, Maintenance, Repair and Miscellaneous Service
http://www.ptb.de). The foundation supporting the national standards are the seven basic SI-units and the SIunits derived from these. In Germany the SI-units possess legal status. Thus, liability protection exists for calibration results having a solid traceability chain. The link to the PTB is generally provided by the German Calibration Service (DKD; http://dkd.ptb.de), founded August 25, 1977. International and European cooperation in the filed of accreditation is pursued by a variety of organizations. The main goal of the cooperation is mutual recognition of accreditation in areas not regulated by law. This recognition is encoded in so-called MLAs (Multilateral Agreements). By becoming signatories to the MLA, the accrediting institutions recognize each others accreditations. The DKD is one of currently 34 members within the European cooperation for Accreditation of Laboratories (EAL; http://www.european-accreditation.org), formerly WECC. The MLA ensures a uniform accreditation standard for all accrediting institutions. This reduces or completely obviates the need for multiple appraisals, which means that the vendor only needs one certificate or one report. At the international level, the ILA (International Laboratory Accreditation) was founded already in 1977, with its main goals being international recognition of inspections and close cooperation by experts in the field. In 1996, the ILA became ILACooperation (ILAC http://www.ilac.org). International cooperation in the area of accreditation of certifying institutions for quality management systems, products and personnel is promoted by the IAF (International Accreditation Forum), founded in 1995.
Danach knnen die entsprechenden Forderungen der DIN EN ISO/IEC 17025 als erfllt gelten, wenn [...] (3.1.3) ein Kalibrier- oder Prfschein vom Hersteller der Mess- oder Prfeinrichtung vorliegt, welcher die Kalibrierung und die Rckfhrung auf SI-Einheiten mittels dokumentiert rckgefhrter Bezugsnormale im Sinne des Punktes 3.1.5 und unter Nutzung allgemein anerkannter Verfahren darlegt bzw. die Kalibrierung und die Rckfhrung auf SI-Einheiten durch ein kompetentes Kalibrierlaboratorium darlegt;
Page 6 of 16
Service Program for imc Measurement Devices Calibration, Adjustment, Maintenance, Repair and Miscellaneous Service
Page 7 of 16
Service Program for imc Measurement Devices Calibration, Adjustment, Maintenance, Repair and Miscellaneous Service
Page 8 of 16
Service Program for imc Measurement Devices Calibration, Adjustment, Maintenance, Repair and Miscellaneous Service
Cost Overview
example: -MUSYCS per device -
Cost Comparison
Industry standard: System ships with manufacturers certificate (Explicit protocols upon request at extra charge) Calibration costs: 5 to 20 /measurement point for multimultichannel systems
-MUSYCS without amplifier = 1502 test points SPARTAN-L 635 test points SPARTAN-
Calibration System inspection (flat charge) System revision (flat charge) Repair (basic) Repair incl. system inspection
only in conjunction with SI/SR 890.(bi)annual 890.bi)annual 1290.(tri-,bi-)annual 1290.tri- bi 80/hr. not recommended 80/hr. plus 890.- flat fee 890.-
System expansion 1710.1710.incl. remodeling/system revision plus cost of parts Warrantee extension Premium Support incl. SW upd. upd. 510.- per year 510. 1500.- per year 1500.-
Its worth asking for the calibration instructions. Are really all ranges and measurement points checked?
www.additive-net.de www.imc-berlin.de
www.additive-net.de www.imc-berlin.de
Sample inspection plan for SPARTAN (635 test points: 0.93 EUR / test point)
Competitor I
5 / Measurement point for additional measurement points over and above those specified by manufacturer Adjustment / repair separate billing 97 / h Basic charge Setup time / measurement quantity Measurement point 80 80 23
Competitor II
(16 channels, voltage x 4 input ranges x 3 points, PT 100 x 2 input ranges at 2 points Example: SPARTAN measuremeapiece temperature) 272 nt 1,360 (Basis 5 ) points ,256 (Basis 23 ) 6 Price incl.minor maintenance: 620.00 (403.00 ) These are the prices to expect from a calibration lab. Adjustment, protocols cost extra
Example: SPARTAN-L 2 x 2 current sources / bay 16 digital inputs 4 alarm outputs (closing relay) Function test instructions UPS test Display Long-term test run with data recording Data recording Stand-alone mode
Measurement points Measurement points Measurement points Test steps Function tests Function test final inspection Function test final inspection Function test final inspection Function test final inspection Test points function test Adjustment Calibration => approx. 0.98 / Test point
Service Program for imc Measurement Devices Calibration, Adjustment, Maintenance, Repair and Miscellaneous Service
Sample inspection plan for -MUSYCS (1502 test points: 0.72 EUR / test point)
Example: calibration, adjustment/calibration, inspection -MUSYCS-S (RJ45, AC, C32, list price 8310 ) Device safety test upon receipt Supply voltages Vcc, +/-15 V, +/-5 V, +12 V Function test Calibration: TE Type K Calibration: 16 x diff 4 input ranges at 3 points apiece Calibration: 3 x ICP 2 input ranges at 3 points apiece Calibration: 2 reference channels at 12 points apiece Calibration: current source Calibration: 32 x single end 4 input ranges at 3 points apiece Adjustment / calibration: 32 x single end 4 input ranges at 3 points apiece Adjustment / calibration: 16 x diff 4 input ranges at 3 points apiece Adjustment / calibration: 3 x ICP 2 input ranges at 3 points apiece Adjustment / calibration: 2 reference channels at 12 points apiece Adjustment / calibration: current source Adjustment / calibration: TE Type K Calibration: 8 x DAC at 3 points apiece Test: 32 x digital input Test: 32 x digital output Test: incremental encoder interface 4 channels dual-signal, trigger bit Device safety test: Final test Supply voltages Vcc, +/-15 V, +/-5 V, +12 V Sync-Signal Long-term test run with data recording Final test instructions, All channels are meassured 1 function test 6 1 function test 16 192 18 24 1 384 384 192 18 24 1 16 24 64 64 12 3 6 2 1 48 232 635 635 1502 => points points points points points points points points points points points points points points points points Test points function test Test points function test Test points function test Function test Test steps Test points function test Adjustment/calibration Calibration upon receipt points
Cost of Ownership
Cost of Ownership
5.9 % of purchase price* 8.6 % of purchase price * 2.7 % of purchase price * per user
Responsibility
Updates ensure long life span Device software for all Windows versions Old, new, different devices can run in concert
all Windows versions unlimited due to unchanged measurement methods for years to come (innovation leader) 10 years 10 years
www.additive-net.de www.imc-berlin.de
Page 10 of 16
Service Program for imc Measurement Devices Calibration, Adjustment, Maintenance, Repair and Miscellaneous Service
Example
Verification of imc CRONOS-PL (CRPL/BR-4) CRPL-BR-4 is a plug-in measurement module for the imc CRONOS-PL family of measurement devices. The user can set the following basic functions (for each channel): + Bridge CF: 9 x 3 input ranges @ 5V, 2,5 V, 1 V + Bridge DC: 9 x 3 input ranges @ 5V, 2,5 V, 1 V + 4 mode options (full bridge, half bridge, quarter bridge: 120 Ohm, 350 Ohm) (= 216 bridge measurement ranges) + DC-amplifier: 13 DC-voltage measurement input ranges 5 mV .. 50 V = 229 basic input ranges (also possible: current measurement, ICP-measurement) DKD 2622 Part 20 describes the calibration of DC-voltage measuring amplifiers. Among other things, this guideline recommends calibration of preferably 10 points per polarity and input range, and repeated preferably 6 times3 = 6 x 4.580 input ranges per channel. By complying with this recommendation to the extent of measuring 5 points per input range, (no repeat measurements), then calibrating the entire measurement functionality of the 4-channel measurement amplifier would entail calibrating 4.580 measurement points. Important note! Even then, that would be calibrating only at one sampling rate and one filter setting! For a complete and manual inspection, assuming a test duration of 1 min per measured point (transient subsidence, data recording in Excel, for instance), the overall test duration comes to more than 70 h or approx. 10 days labor. Estimating an hourly wage rate of 95 EUR, this would come to more than 7000 EUR calculating on a T&L basis; some calibration labs calculate the price per measurement point... This manner of proceeding is uneconomical, not suitable for serial inspection and therefore employed only in exceptional cases as verification of calibration. An alternative for obtaining the conformity rating Usable measurement functionality was calibrated is calibration according to imc manufacturers specifications: Ee the example "T5311353.prt" (text file - excerpt of adjustment protocol for a CRPL/BR-4) In the Internet at www.imc-berlin.de or www.additive-net.de => Instrumentation => Search for 99999 (registration for download area required) In designing the measuring amplifier, the procedures for its adjustment and calibration are also developed at the same time and verified in the type inspection. One part of the procedure is to adjust various parameters due to their dependence on properties of components and on the running temperature; additional parameters such as the sampling rate selected or the selected filtering may not affect the measurement precision according to design targets. The resulting acceptance tests include procedures for the serial production and servicing, which reflect an optimum balance of test precision, standards conformity and economic viability In accordance with imc manufacturers specifications, the following measurement points are adjusted and calibrated for the BR-4 (on each of 4 channels): + DC 5 mV ... 50 V: Gain: 13 input ranges * 3 measurement points, + Offset: 13 measurement points (208 measurement points) + internal reference quantities (not externally accessible): 18 measurement points + CF-mode full bridge, 5 V power supply, low-pass 10 Hz in the input ranges 1 mV/V, 10 mV/V, 50 mV/V, 200 mV/V: 16 measurement points + DC mode full bridge 9 input ranges: 108 measurement points = in total 350 measurement points
3 DKD2622, Part 20 (Draft 09/2003) quote from Section 3.2.3: To characterize the response of measuring amplifiers for a defined input signal range, measurement of the output signal of the measuring amplifier at equidistant input values (preferably at 10 points per polarity) is repeated preferably six times... ... Service_and_calibration_of_imc_devices_2006.doc; MS, 25.01.2006 Page 11 of 16
Service Program for imc Measurement Devices Calibration, Adjustment, Maintenance, Repair and Miscellaneous Service
In order to be able to test according to these specifications, the condition for calibrating must be met; this is checked in a function test prior to calibration, in which additional measurement points are tested according to imc inspection plans with a pass/fail rating:
-
Offset measurement and noise test shorted directly at the device, with a high sampling rate and no filter Background: If either the offset or the noise is too high, there is a defect which may violate the calibration conditions. Too much noise, in conjunction with averaging of the signal, can lead to an apparent offset or apparent non-linearities. Testing of reference voltages, supply voltages Calibration signals emitted by the test object (e.g. shunt calibration of bridges) Test of basic functions such as auto bridge balancing Test of power consumption
These tests are generally conducted during the warm-up phase. The inspection plan for confirming the calibration conditions is modified by the Service department in adaptation to emerging experience. If the function test indicating fulfillment of the calibration conditions according to the imc inspection plan is passed, and the subsequent calibration according to imc manufacturers specifications carried out successfully, the conformity declaration "Calibrated according to manufacturers specifications traceable to national standards". Any additional measurement points specially ordered by the customer must be calibrated subsequently.
Page 12 of 16
Service Program for imc Measurement Devices Calibration, Adjustment, Maintenance, Repair and Miscellaneous Service
Sample inspection plan for CANSAS-L-UNI8 (840 test points included in calibration)
only calibration Item Remarks
Calibration / Adjustment
1 2 3 4 5 6 6.1 6.2 6.3 6.4 6.5 6.6 6.7 6.8 7 7.1 7.2 7.3 7.4 7.5 7.6 7.7 7.8 8 9 9.1 9.2 9.3 9.4 9.5 9.6 9.7 9.8 10 11 12 13 x x x x x x x x
x x x x
x x
Option
Admin
Designation
Description
Goods Received Administration Service standard Function test (IN) Temperature calibration (IN)
intake, unpacking, recording, device intake printout records, oder clarification, record service order, check history, assign to P, order confirmation to customer determine SW, in case of repairs: replicate error (see 4) manually with ISO-thermal box, Type K thermocouple 1 MP * 1 input range * 8 channels (room temperature) verify faults, clarify automated (no input needed), 840 MP 1 MP * 13 input ranges * 8 channels = 104 MP 1 MP * 13 input ranges * 8 channels = 104 measurement points 3 MP * 13 input ranges * 8 channels = 312 MP 11 MP * 8 channels = 88 MP 4 MP 4 MP * 8 channels = 32 MP 4 MP * input ranges * 8 channels = 192 MP 4 MP automated (no input needed), 840 MP 1 MP * 13 input ranges * 8 channels = 104 MP 1 MP * 13 input ranges * 8 channels = 104 MP 3 MP * 13 input ranges * 8 channels = 312 MP 11 MP * 8 channels = 88 MP 4 MP 4 MP * 8 channels = 32 MP 4 MP * input ranges * 8 channels = 192 MP 4 MP plausibility, check validity on server manually upon instruction
x x
x Administration Service Reparatur (IN) Calibration (IN) Noise test 1 (RMS) Noise test 2 (peak-peak) Voltage measurement PT100 measurement internal reference PT1000 Bridge Sense Current measurement Bridge Supply Adjustment / calibration (OUT) Noise test 1 (RMS) Noise test 2 (peak-peak) Voltage measurement PT100 measurement internal reference PT1000 Bridge Sense Current measurement Bridge Supply Check protocol, generate certificate electronically
Final inspection (OUT) Module measurement Test PT100 @ 50 C Sensorflash Robustness test 1 Sensor info Test Divider test Robustness test 2 Sensor info Test Test shunt calibration Temperature calibration (OUT)
manually with ISO-thermal box, Type K thermocouple 1 MP * 1 input range * 8 channels (room temperature) File service documents, print certificate, inspection markrt on the module optional
Document results, label module x Print protocol and certificate per ISO17025 x Recording of time and materials, file and edit device intake document
14 15 16 17 18
Inspect documents, file in inforunder Service, edit device intake document Inspect plausibility, adaptation to customer wishes Shipping documents, packaging, shipping Conclude procedures, statistics Statistics, reports, archiving
x x x
Page 13 of 16
Service Program for imc Measurement Devices Calibration, Adjustment, Maintenance, Repair and Miscellaneous Service
Maintenance Work
Definitions 1.) Calibration: 2.) Inspection 3.) Revision 4.) Repair, Remodeling, Update To date: IMC calibration = 1 + 2 +(sometimes) 3 Now: System-inspection or Revision System1 can be done by anybody, 2 4 only the manufacturer imc
System Calibration
Comparison with a reference having the same units Calibration means certificate of traceability Only valid at the time performed Re-calibration appointment the users responsibility ReUser must monitor interval between calibrations We recommend: Calibration interval of 1-2 years 1Who can perform calibration? Any calibration laboratory
Usually in accordance with DIN EN ISO/IEC 17025
www.additive-net.de www.imc-berlin.de
www.additive-net.de www.imc-berlin.de
Cost of Ownership
Cost of Ownership
System Inspection
Calibration of inputs Maintenance (repair if appropriate) Adjustment Function test Final inspection incl. Calibration of outputs Shipping with manufacturers calibration certificate and service report
Protocol of data for each channel and input range offered for extra charge extra
System Revision
Inspection plus System Update
Bringing systems up to date Benefits of further developments Extends system life span Revision only performed by imc Recommended at 12 - 36 month intervals
www.additive-net.de www.imc-berlin.de
www.additive-net.de www.imc-berlin.de
Cost of Ownership
Cost of Ownership
Repairs
Repair (basic)
Test for reproducibility of error message and repair Function test of overall system (incomplete) Calibration history is lost
Why perform a system inspection (former calibration) with every repair order
The device returns to the user in reliable working condition Calibration history not lost Best, most reliable 100% function test (even finds hidden errors) errors) Fault remedy/ component replacement often necessitates readjustment, anyway readjustment, Example: an adapter defect is reported. Error profile: defective input channels strain the adapter. Also: DAC (analog outputs) Channel 2 defective.
(not recommended)
www.additive-net.de www.imc-berlin.de
Page 14 of 16
Service Program for imc Measurement Devices Calibration, Adjustment, Maintenance, Repair and Miscellaneous Service
Page 15 of 16
Service Program for imc Measurement Devices Calibration, Adjustment, Maintenance, Repair and Miscellaneous Service
Contacts
Youre welcome @ http://www.imc-berlin.de/en imc Mesysteme GmbH integrated measurement & control Quality Assurance Voltastrae 5 D-13355 Berlin Additive GmbH hotline Tel.: +49-(0)30-467090-26 Fax: +49-(0)30-467090-22 email: hotline@imc-berlin.de Internet: http://www.imc-berlin.de Tel.: +49 (0) 641 5905-0 email: hotline@additive-net.de Internet: http://www.additive-net.de
DIN EN ISO/IEC 17025:2005 General requirements for the competence of testing and calibration laboratories (replaces EN 45000, implements ISO Guide 25) Calibration guidelines for electrical measurement equipment DIN ISO 10012-1: Measurement monitoring systems ISO 8402 (1986): Quality terminology also: DIN 55350 T 11-15, 17, 18, 21-24, 31 (Terminology) + DIN Taschenbuch 223 DKD Guidelines, miscellaneous (see also http://www.dkd.ptb.de) VDI/VDE/DGQ/DKD 2622 (draft): Kalibrieren von Messmitteln fr elektrische Gren (describes calibration procedures generally) Blatt 3: Kalibrierrichtlinien fr Digitalmultimeter Blatt 4: Kalibrierrichtlinien fr Oszilloskope Blatt 5: Funktionsgeneratoren Blatt 11: Kalibrierrichtlinien fr Spektrumanalysatoren Blatt 20: Kalibrierung von Gleichspannungsmessverstrkern DKD-R 5-5 regulates calibration of temperature indication devices DKD-R 6-1 regulates calibration of pressure measurement modules DKD-3: Specifies measurement uncertainty in calibrations DKD-4 (1991) Traceability of test equipment to national standards EA Richtlinien (European cooperation for Accreditation of Laboratories): Overview: DAR-INF2, e.g. DAR-EM6: Interne Audits und QM-Bewertung in Laboratorien DAR-EM32: Anleitung zur Akkreditierung nach ISO/IEC 17025 DAR-EM23: Sammlung von Begriffen DIN V ENV 13005: Leitfaden zur Angabe der Messunsicherheit beim Messen (identical to Guide to the expression of uncertainty in measurement = GUM)
Page 16 of 16