Sie sind auf Seite 1von 49

Service Manual for Volumetric Infusion Pump ARGUS 707 V

Made in Switzerland

ARGUS Medical AG, CH-3627 Heimberg / Switzerland


(a member of the CODAN group)

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

1 / 48

TABLE OF CONTENTS 1. INTRODUCTION ...........................................................................................................4


1.1. 2.1. 2.2. 2.3. 2.4. 2.5. 2.6. 2.7. 2.8. 2.9. 3.1. 3.2. 4.1. 4.2. 4.3. 4.4. 5.1. 5.2. 5.3. 5.4. 6.1. 6.2. 6.3. 6.4. 6.5. 6.6. 6.7. 7.1. 7.2. 7.3. General ............................................................................................................................. 4 General ............................................................................................................................. 4 Interrogation mode (without ARGUS service) ................................................................... 5 Configuration mode (without ARGUS service).................................................................. 6 First activation of a PIN code (write protection) ................................................................ 7 Changing an existing PIN code......................................................................................... 8 Address list of the pump configuration (without ARGUS service) ..................................... 9 Index list of the pump configuration (with ARGUS service) ............................................ 10 Special configuration options (with ARGUS service) ...................................................... 14 Medication list ................................................................................................................. 15 General ........................................................................................................................... 18 Serial communication protocol ........................................................................................ 18 General ........................................................................................................................... 19 ARGUS service Configuration...................................................................................... 20 ARGUS service - Calibration (volumetric pums only) ..................................................... 21 ARGUS service - Toolbox ............................................................................................... 22 General ........................................................................................................................... 26 Requirements for a software update ............................................................................... 26 Software update procedure............................................................................................. 26 Safety aspects................................................................................................................. 28 General ........................................................................................................................... 29 Recalibration ................................................................................................................... 29 Pressure calibration ........................................................................................................ 30 Pressure control measurement ....................................................................................... 35 Volume calibration .......................................................................................................... 36 Pump specifications ........................................................................................................ 37 Fault codes ..................................................................................................................... 38 General ........................................................................................................................... 39 Disassembling of the ARGUS 707 V............................................................................... 39 Spare parts ..................................................................................................................... 43

2. PUMP CONFIGURATIONS ...........................................................................................4

3. Serial communication of the pump ..........................................................................18

4. ARGUS service...........................................................................................................19

5. SOFTWARE UPDATES ..............................................................................................26

6. MAINTENANCE ..........................................................................................................29

7. REPLACEMENT OF PARTS.......................................................................................39

8. WIRING DIAGRAMM...................................................................................................46 9. BLOC SCHEMATIC ....................................................................................................46 10. SAFETY STANDARD CHECK ....................................................................................47 11. REPAIR ORDER FORM ..............................................................................................48

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

1 / 48

INTRODUCTION

Park area for drop detector

Display

Drop detector Drop chamber

Global alarm

Air bubble detector

Pressure sensor upstream

automatic free-flow-clamp

Pressure sensor downstream

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

2 / 48

INTRODUCTION

Universal clamp

Staff alert and RS-232-connection

Drop detector connection

AC power connection

Identification plate with pump serial number

DC-connection & QUICK100 Docking Station interface

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

3 / 48

INTRODUCTION 1. INTRODUCTION 1.1. General IMPORTANT! This service manual is intended for the exclusive use of authorized persons who have been trained by ARGUS Medical AG in the maintenance and repair of the ARGUS 707 V infusion pump. The service manual is meant to be used together with the user manual. IMPORTANT! ARGUS Medical AG shall not assume any responsibility for any manipulations which have been carried out on the ARGUS 707 V infusion pump by a nonauthorized person. CAUTION! The ARGUS 707 V infusion pump may only be used with spare parts, accessories, consumables and IV-sets with Luer-Lock connections recommended by ARGUS Medical AG. The functional safety of the pump is not guaranteed if non approved materials are used. The safety of the patient may be endangered. This manual contains the latest data available. It is subject to further modifications in accordance with technical improvements.

2. PUMP CONFIGURATIONS 2.1. General CAUTION! The configuration possibilities with the ARGUS service PC utility tool and without PC assistance constitute a modification of the pump and may only be carried out by authorized persons! CAUTION! After changing the configuration a function check and a control measurement has to be performed!

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

4 / 48

PUMP CONFIGURATIONS 2.2. Interrogation mode (without ARGUS service) With the interrogation mode you can read the present keypad configuration of the pump without the possibility to modify any configurations. For a complete overview, please take the ARGUS service PC-tool!
1. To enter into the interrogation mode, switch the pump on while keeping the keys MODE and START/STOP pressed.

707 P Rog

...5.

1
2. Press the START/ STOP key.

4. To quit the interrogation mode press the ON/OFF key at least 2 seconds.

3. Enter the requested address (see chapter 2.6), e.g. 5. The programmed value of the address 5 appears in the right hand display.

Flashing decimal points indicate which display is ready to accept an input by the keys 100, 10 & 1. In the interrogation mode the left hand display shows the address and the right hand display shows the according value configured at this address. Please refer to chapter 2.6. where the meanings of the addresses are explained. To modify any configuration data you have to go into the configuration mode.

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

5 / 48

PUMP CONFIGURATIONS 2.3. Configuration mode (without ARGUS service) The configuration mode permit you to modify the pump keypad configuration manually using the keypad. Please refer to chapter 2.6. where the meaning of the adresses are explained. To have access to all configuration options, please use the ARGUS service PC-tool! 2.3.1. Step 1
1. To enter into the configuration mode, switch the pump on while keeping the keys MODE and START/STOP pressed. 2. Press key START

707 p rog

3. Press key MODE

4. Press key START again. The display will show Cod ----

2.3.2. Step 2
5. Enter the PIN Code. If no PIN code was configured, no entry is required

Cod

----

6. Press key START to acknowledge the entered PIN code

If the PIN was not accepted, the display will change back to point 5.

You have now access to all addresses in the list of chapter 2.6. Select therefore any address in the left display (see next page).

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

6 / 48

PUMP CONFIGURATIONS 2.3.3. Step 3


8. Enter now the value on the right hand display. The range of the value is given by the table in chap.2.6 7. After entering an address e.g. 5 press the key MODE. The flashing decimal points will change to the right hand display.

...5 .

0
9. Press key START to acknowledge the value entered. The flashing decimal points change back to the left hand display.

10. To quit the configuration mode press the ON/OFF key at least 2 seconds. Changes in configuration become active, after the pump is switched on normally again.

Important remark: Invalid values entered will be corrected automatically by the pump to the maxima or minima value allowed for the according address!

2.4. First activation of a PIN code (write protection) The activation of a PIN code allows you to protect the configuration from unauthorized access. To activate the PIN code, enter into the configuration mode.
1. To enter into the configuration mode, switch the pump on while keeping the keys MODE and START/STOP pressed. 2. Press key START

707 P Rog
4. Press key START again. The display will show Cod ----.

5. Press MODE key (Code 0 will be set) then START key to acknowledge the entered PIN code 0.

3. Press key MODE

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

7 / 48

PUMP CONFIGURATIONS

0 ...0.

6. Press the MODE key. The flashing decimal points will change to the right hand display

8. Press key START to acknowledge to new PIN.

7. Enter now the new PIN Code (max. 4 digit number). Please remember this code, it will never be visible again! e.g.

CAUTION! After you switch the pump OFF and ON again you can enter into the configuration mode only, if you enter the correct PIN code. e.g. Please note: The interrogation mode can always be accessed without the PIN.

2.5. Changing an existing PIN code Enter the configuration mode using present PIN, select add. 0 and set the new code. Enter actual PIN code and confirm with START key. Press MODE key (#0). The flashing decimal points will change to the right hand display.

Enter the new PIN code and press the START/STOP key to acknowledge the entered code.

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

8 / 48

PUMP CONFIGURATIONS 2.6. Address list of the pump configuration (without ARGUS service) The following list declares the possible configuration options which can be performed on the pump keypad without using the PC. All these options can also be configured by the PC-Software ARGUS service.
Address left display 1 2 3 4 5 6 Index PC 2 10 11 19 44 49 Default Function Unit Range right display 0=No / 1=Yes 0=No / 1=Yes 0=No / 1=Yes 0=No / 1=Yes 0=No / 1=Yes 0=No / 1=Yes

No No No No Yes No

Key ON/OFF only at stop valid Operation without drop detector Recall of the last used ml/h rate Buzzer at start Automatic pressure release Alarm acknowledge with key MODE

100 101 102 103 104

361 362 363 316 318

5 8 7 9 20

Key ON/OFF delay time Display brightness Buzzer volume Default pressure limit (IV-set 1) Drop rate window centre (IV-set 1)

0.1 s level level 100 mbar Drops/ml

0 - 31 2 - 15 5 - 10 1 - 10 10 - 65

200

317

250

Air bubble size (IV-set 1)

50 - 1000

399

707 707

Enter the calibration menu (volume) Enter the calibration menu (pressure)

123 1234

Note! The address does not correspond with the index used by the ARGUS service tool.

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

9 / 48

PUMP CONFIGURATIONS 2.7. Index list of the pump configuration (with ARGUS service)
Index PC 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 Add. Default Yes No No No No No No No No Yes No No No No No Yes No No No No Yes Yes No No No No No No No No No Function Run indication by running decimal point Key ON/OFF only at stop valid Rate change allowed only at STOP Key STOP delayed Second entry of rate Static alarm (staff alerting system) Display elapsed time in run mode Display remaining time Unit Range No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes

# 361 # 3=Yes # 8=No # 7=Yes

2 3

Operation without drop detector Recall of ml/h rate at next power on Recall of ml total at next power on Recall of ml inf. at next power on SBS (step by step function) Display VTBI (volume to be infused) Display Set-X if only one set is enabled KVO (KOR) enabled # 60 Drop alarm only if bottle empty # 10=No, trA Buzzer at start Menu Clr (clear ml inf.) # 15=No Menu trA (transport) Menu PrL (pressure alarm limit) Menu CAP (battery capacity) Menu SEt Fill (prime IV set) Menu InF (ml inf. since last power on) Menu dLo (data lock) Menu Stb (stand by) Menu Med (medication name) Menu tM (timer alarm) Menu boL (bolus release) Menu boLr (bolus rate) Menu tot (bolus total) # 10, 18=No

# 32=Yes # 32=Yes

No No No No No Yes Yes No No No No

Automatic bolus application # 32, 34=Yes Neonatology mode accessible in PrL menu Clear ml/h after infusion completed Clear ml total after infusion completed # 41=Yes Air volume accumulated Automatic pressure release after occlusion Pressure display ON (LED bar graph - 20/40/60/80/100%) Pressure display with indicator # 45=Yes Stand by- and battery pre alarm low volume Flashing numeric display at alarm Alarm acknowledge only with key MODE
ARGUS Medical AG 28.03.06 / PJ

No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes
10 / 48

14.203.B_A707 en.SM.V4.3X

PUMP CONFIGURATIONS
50 51 52 53 54 55 56 60 65 101 102 103 104 105 201 202 203 204 205 206 207 208 209 210 211 212 213 214 215 216 217 221 222 223 224 225 226 227 228 229 230 231 232 233 234 235 236 237 No Yes Start with 1 bar pressure allowed Enable upstream occlusion detection No / Yes No / Yes

No Yes No No Yes No No No No 1065 1054 1044 1035 1027 1020 1014 1008 999 988 976 967 952 942 925 910 1000 1065 1054 1044 1035 1027 1020 1014 1008 999 988 976 967 952 942 925 910 1000

Select RJ-45 connector for serial interface Medication display alternate # 29=Yes KVO only after infusion completed Clear and continue # 15=No

ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio

No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 -1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 1000-1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 -1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 1000-1100
11 / 48

Set 1 enabled Set 2 enabled Set 3 enabled Set 4 enabled Automatic return on default set 1 after power up Correction at 1 ml/h Correction at 3 ml/h Correction at 5 ml/h Correction at 10 ml/h Correction at 25 ml/h Correction at 50 ml/h Correction at 75 ml/h Correction at 100 ml/h Correction at 150 ml/h Correction at 200 ml/h Correction at 250 ml/h Correction at 300 ml/h Correction at 400 ml/h Correction at 500 ml/h Correction at 750 ml/h Correction at 999 ml/h Correction at 1200 ml/h Correction at 1 ml/h Correction at 3 ml/h Correction at 5 ml/h Correction at 10 ml/h Correction at 25 ml/h Correction at 50 ml/h Correction at 75 ml/h Correction at 100 ml/h Correction at 150 ml/h Correction at 200 ml/h Correction at 250 ml/h Correction at 300 ml/h Correction at 400 ml/h Correction at 500 ml/h Correction at 750 ml/h Correction at 999 ml/h Correction at 1200 ml/h

# 101=Yes

SET 1 Do not change these values unless explicitly trained

SET 2 Do not change these values unless explicitly trained

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

PUMP CONFIGURATIONS
238 239 240 241 242 243 244 245 246 247 248 249 250 251 252 253 254 255 256 257 258 259 260 261 262 263 264 265 266 267 268 269 270 271 272 273 274 275 276 277

1065 1054 1044 1035 1027 1020 1014 1008 999 988 976 967 952 942 925 910 1000

Correction at 1 ml/h Correction at 3 ml/h Correction at 5 ml/h Correction at 10 ml/h Correction at 25 ml/h Correction at 50 ml/h Correction at 75 ml/h Correction at 100 ml/h Correction at 150 ml/h Correction at 200 ml/h Correction at 250 ml/h Correction at 300 ml/h Correction at 400 ml/h Correction at 500 ml/h Correction at 750 ml/h Correction at 999 ml/h Correction at 1200 ml/h

SET 3 Do not change these values unless explicitly trained

ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio

900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 1000-1100

1065 1054 1044 1035 1027 1020 1014 1008 999 988 976 967 952 942 925 910 1000

Correction at 1 ml/h Correction at 3 ml/h Correction at 5 ml/h Correction at 10 ml/h Correction at 25 ml/h Correction at 50 ml/h Correction at 75 ml/h Correction at 100 ml/h Correction at 150 ml/h Correction at 200 ml/h Correction at 250 ml/h Correction at 300 ml/h Correction at 400 ml/h Correction at 500 ml/h Correction at 750 ml/h Correction at 999 ml/h Correction at 1200 ml/h

SET 4 Do not change these values unless explicitly trained

ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio ratio

900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 900 - 1100 1000-1100

310 311 312 313 314 315 316 317 318 319

103 200 104

999.9 999 1200 10 10 20 7 250 20

Max. flow rate Max. prime rate Max. bolus rate Max. bolus total Upstream occlusion sensitivity Air summation volume Downstream default pressure limit PrL Air bubble size Drop rate window center

Set 1 parameters

ml/h ml/h ml/h ml factor 50 l 100 mbar l drops

1 - 999.9 1 - 999 1 - 1200 1 - 999 0 - 60 2 - 40 1 - 10 50 - 1000 10 - 65

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

12 / 48

PUMP CONFIGURATIONS
320 321 322 323 324 325 326 327 328 329 330 331 332 333 334 335 336 337 338 339 340 341 342 343 344 345 346 347 348 349 350 351 361 362 363 364 365 366 367 368 369 390 391 392 393 394 395 396 397 398 100 101 102 103 999.9 999 1200 10 10 20 7 250 20 999.9 999 1200 10 10 20 7 250 20 999.9 999 1200 10 10 20 7 250 20 Max. flow rate Max. prime rate Max. bolus rate Max. bolus total Upstream occlusion sensitivity Air summation volume Downstream default pressure limit PrL Air bubble size Drop rate window center Max. flow rate Max. prime rate Max. bolus rate Max. bolus total Upstream occlusion sensitivity Air summation volume Downstream default pressure limit PrL Air bubble size Drop rate window center Max. flow rate Max. prime rate Max. bolus rate Max. bolus total Upstream occlusion sensitivity Air summation volume Downstream default pressure limit PrL Air bubble size Drop rate window center ml/h ml/h ml/h ml factor 50 l 100 mbar l drops ml/h ml/h ml/h ml factor 50 l 100 mbar l drops ml/h ml/h ml/h ml factor 50 l 100 mbar l drops 1 - 999.9 1 - 999 1 - 1200 1 - 999 0 - 60 2 - 40 1 - 10 50 - 1000 10 - 65 1 - 999.9 1 - 999 1 - 1200 1 - 999 0 - 60 2 - 40 1 - 10 50 - 1000 10 - 65 1 - 999.9 1 - 999 1 - 1200 1 - 999 0 - 60 2 - 40 1 - 10 50 - 1000 10 - 65

Set 2 parameters

Set 3 parameters

Set 4 parameters

4 5 8 10 150 1 300 5 0 0 0 0 0 0

Air volume summation time Key ON/OFF delay time Display brightness Buzzer alarm volume #4

8 min 1/10 s level level mbar Enum min sec year month day month hour xxx 8 yyy

1-8 0 - 31 2 - 15 5 - 10 60 - 250 1-5 45 - 300 5 - 30 0 - 99 0 - 12 0 - 31 0 - 24 0 -10000 xxx 8 yyy

Pressure step for neonatology mode # 39=Yes Pressure display unit (mbar / mmHg / kPa / cmH2O / Psi) Battery capacity (discharge time) Automatic menu fall back delay time Last service date in year Last service date in month Last service date in day Service interval in months Service interval in hours of operation Pump serial number

200

(0 = disabled) (0 = disabled)

399

707

Enter the calibration menu / clears protection key


ARGUS Medical AG 28.03.06 / PJ

123/4
13 / 48

14.203.B_A707 en.SM.V4.3X

PUMP CONFIGURATIONS
Permissions for serial communication 0 = none, 1 = query only, 2 = parametrising, 3 = remote control Allow change of ml total while infusing # 65=No Alarm lamp brightness Display a department info text (after power up)

518 522 523 524

2 No 15

Enum level char

0-3 No / Yes 2 - 15
1-16 ASCII

Using the ARGUS service tool, the complete and detailed pump configuration can be done.

2.8. Special configuration options (with ARGUS service) Enabling a new IV set Follow the steps below to enable and configure a new IV set: a) write Yes in the corresponding index 101, 102, 103, 104 b) make sure the set definitions (index 201 to 277) correspond to the IV sets used. In case of doubt, contact ARGUS Medical AG or your local dealer. c) configure the set parameters (index 310 to 349 and 351) d) perform a complete set calibration as per chapters 6.3. to 6.5.

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

14 / 48

PUMP CONFIGURATIONS 2.9. Medication list 2.9.1. General To display medication names, index 29 (menu "MEd") must be set to "Yes". The selected medication name can be displayed also in alternate mode (rate, ml inf./ med. name) during infusion, for that set index 56 to Yes. After enabled special function Med the following medication names can be selected via pump keypad (see user manual). 2.9.2. User medication 32 user medication names can be custom defined. Choose between capital and small letters for a better displayed medication name. Because of the 7-segment pump display some characters maybe difficult to read.
Index PC 561 562 563 564 565 566 567 568 569 570 571 572 573 574 575 576 577 578 579 580 581 582 583 584 585 586 587 588 589 590 591 592 Default UserM 1 UserM 2 UserM 3 UserM 4 UserM 5 UserM 6 UserM 7 UserM 8 UserM 9 UserM 10 UserM 11 UserM 12 UserM 13 UserM 14 UserM 15 UserM 16 UserM 17 UserM 18 UserM 19 UserM 20 UserM 21 UserM 22 UserM 23 UserM 24 UserM 25 UserM 26 UserM 27 UserM 28 UserM 29 UserM 30 UserM 31 UserM 32 Function User med. name 1 User med. name 2 User med. name 3 User med. name 4 User med. name 5 User med. name 6 User med. name 7 User med. name 8 User med. name 9 User med. name 10 User med. name 11 User med. name 12 User med. name 13 User med. name 14 User med. name 15 User med. name 16 User med. name 17 User med. name 18 User med. name 19 User med. name 20 User med. name 21 User med. name 22 User med. name 23 User med. name 24 User med. name 25 User med. name 26 User med. name 27 User med. name 28 User med. name 29 User med. name 30 User med. name 31 User med. name 32 Unit char char char char char char char char char char char char char char char char char char char char char char char char char char char char char char char char Range 1-8 1-8 1-8 1-8 1-8 1-8 1-8 1-8 1-8 1-8 1-8 1-8 1-8 1-8 1-8 1-8 1-8 1-8 1-8 1-8 1-8 1-8 1-8 1-8 1-8 1-8 1-8 1-8 1-8 1-8 1-8 1-8 ASCII ASCII ASCII ASCII ASCII ASCII ASCII ASCII ASCII ASCII ASCII ASCII ASCII ASCII ASCII ASCII ASCII ASCII ASCII ASCII ASCII ASCII ASCII ASCII ASCII ASCII ASCII ASCII ASCII ASCII ASCII ASCII

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

15 / 48

PUMP CONFIGURATIONS 2.9.3. Defined medication


Index PC 600 601 602 603 604 605 606 607 608 609 610 611 612 613 614 615 616 617 618 619 620 621 622 623 624 625 626 627 628 629 630 631 632 633 634 635 636 637 638 639 640 641 642 643 644 645 646 647 648 649 650 651 652 653 654 655 656 657 658 659 Def. Yes No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No Function Enable medication Actilyse Actrapid Adrenaline 0.1 Adrenaline 0.2 Adriamycine Ajmalin Alfentanil Alimta Alizapride Alteplase Alupent Ambroxol Aminophylline Amiodaron Amphotericine Amrinone lactate Antibiotic Aprotinin Arterial line Atracurium Bleomycine Blood Bretylium Bupivacne Capecitabine Carboplatin Carmustine Cefazolin sodium Central line Ceruletid Chlormethine Cisplatine Cladribine Clomethiazol Clonidin Cyclophosphamide Cytarabine Dacarbazine Dactinomycine essai cliniq. Daunorubicine liposomale Diltiazem Dobutamin Docetaxel Dopamine Dopexamine Doxapram hcl Doxorubicin Droperidol Ecteinascidine Edrecolomab Epidural Epirubicine Esmolol Etoposide Fentanyl Flecainide Fludarabine Fluimucil Fluorouracil Range No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes Index PC 660 661 662 663 664 665 666 667 668 669 670 671 672 673 674 675 676 677 678 679 680 681 682 683 684 685 686 687 688 689 690 691 692 693 694 695 696 697 698 699 700 701 702 703 704 705 706 707 708 709 710 711 712 713 714 715 716 717 718 719 Def. No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No Function Fotemustine Furosemid Gemcitabine Glucose 5% Glucose 10% Glucose 30% Glucose - n/saline Glycerin Granisetron Hartmann`s Heparin Hydroxycarbamide Ifosfamide Immunoglobulin Insulin Irinotecan Isoproterenol HCL KCL Ketamin Labetalol L-asparaginase Lidocain Liothyronin Lipids Lomustine Magnesium Maintenance line Mannitol Metaraminol bitartrate Methohexital sodium Methotrexate Methyldopate hcl Metoclopramide Metopimazine Midazolam Milrinone Mitoguazone Mitomycine Mitoxantrone Mivacurium chloride Morphin Nacl 0.9% Nacl 0.45% Nalbuphin Naloxone Nicardipine Nifedipin Nimodipin Nitroclycerin Nitroprussiate Noradrenalin Norepinephrine bitartrate Nutriflex Omeprazole Omipressin Ondansetron Oxaliplatine Oxytocin Paclitaxel Pentoxityllin Range No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No /Yes No /Yes No /Yes No /Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes
16 / 48

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

PUMP CONFIGURATIONS
720 721 722 723 724 725 726 727 728 729 730 731 732 733 734 735 736 737 738 739 740 741 742 743 744 745 746 747 748 749 750 751 752 753 754 755 756 757 758 759 760 No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No Phentolamine Phenylephrin Pirarubicine Plicamycin Procainamide Procarbazine Propafenon Propofol Prostaglandin e1 Raltitrexed Rapilysin Remifentanyl Risordan Ritodrine hcl Rituximab Rocuronium bromide Ropivacane Salbutamol Somatostatin Streptokinase 36000 ie Streptokinase 1000 ie Streptokinase 3000 ie Streptokinase 6000 ie Streptokinase Streptozocine Succinylcholine chloride Sufentanil T.P.N Tegafur Temozolomide Terbutaline Theopyllin Thiotepa Tirofiban Tolazolin hcl Topotecan Trastuzumab Trimethaphan camsylate Umbilical arterial catheter Umbilical venous catheter Urapidil No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes 761 762 763 764 765 766 767 768 769 770 771 772 773 774 775 776 777 778 779 780 781 782 783 784 785 786 787 788 789 790 791 792 793 794 795 796 797 798 799 800 No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No No Urokinase Vasopressine Vecuronium Verapamil Vinblastine Vincristine Vindesine Vinorelbine User defined med. 1 User defined med. 2 User defined med. 3 User defined med. 4 User defined med. 5 User defined med. 6 User defined med. 7 User defined med. 8 User defined med. 9 User defined med. 10 User defined med. 11 User defined med. 12 User defined med. 13 User defined med. 14 User defined med. 15 User defined med. 16 User defined med. 17 User defined med. 18 User defined med. 19 User defined med. 20 User defined med. 21 User defined med. 22 User defined med. 23 User defined med. 24 User defined med. 25 User defined med. 26 User defined med. 27 User defined med. 28 User defined med. 29 User defined med. 30 User defined med. 31 User defined med. 32 No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes No / Yes

REMARK : Via barcode reader all medication names can be selected, even if they are not released in the configuration.

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

17 / 48

Serial communication of the pump 3. SERIAL COMMUNICATION OF THE PUMP 3.1. General The ARGUS 708 Volumetric pump has two serial interfaces on board. One is wired to the docking interface connector and one is connected to the RS232 connector. Important remark! Only the RS232 connector is galvanic separated. The docking interface on the pump is a non galvanic isolated interface! Do not use the docking interface on the pump together with the interface cable (part 10.093) on a patient! If the pump is intended to be monitored as a standalone pump on a patient, use the RS232 connector (RJ45, beside the drop detector connector). To use it, configure this connector as the default serial interface (refer index #55 in the configuration). If the pump is docked into a docking station ARGUS 60 M or ARGUS 100 M, the software switches automatically to the docking interface and the docking station builds the separation device (galvanic isolation) then. 3.2. Serial communication protocol The following characteristics are basics for all the ARGUS devices (volumetric pumps, syringe pumps, docking stations with V4.xx and PCs) which are intended to communicate with the device mentioned in this service manual. Full-duplex RS232, currently 4800Baud for single pumps, 9600 Baud for docking stations (also on master/slave-link). Simple master (host/PC) slave (device) communication (host does polling). The host has to repeat the request if there is no valid response. Uses a checksum (CRC-8). Binary data transmission, thus no ASCII/text parsing. Fast & direct communication with pumps on ARGUS docking station. Specified timeouts during remote mode. Basic framing technique used as in the Serial Infrared Link Access Protocol (IrLAP) Version 1.1. Please contact your local distributor or ARGUS Medical AG for the complete serial communication protocol description.

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

18 / 48

ARGUS service 4. ARGUS SERVICE 4.1. General The new ARGUS service utility is a high and user friendly PC software which can configure and upgrade pumps over PC serial COM port. With this Windows based software you can also set pump clock, change PIN code, read and print out history and easily replicate pump configurations, and so on. The modern and clearly structured design of this self-describing PC-tool allows a very easy and rapid modification of the ARGUS 600 Syringe pump, the ARGUS 707 & 708 Volumetric pump and the ARGUS docking station. This software may be available from your local distributor or directly from ARGUS. REMARK: ARGUS service may only be used with software versions greater or equal to 4.00. CAUTION! The infusion pump has to be disconnected from the patient before and while the serial interface cable is connected to the pump. The connection of the A707 over the serial interface RS-232 can be done by connecting the interface cable (REF 10.093) to the serial interface outlet of the serial PC-COM port. Start the ARGUS service

Press the button which confirm to your previously connected device (ARGUS pump or docking station).

Hold the 10 key while switching-ON the pump. Select Start Configuration (see next chapter) or Start Update-Center (see chapter 5.3.2.).

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

19 / 48

ARGUS service

Select the next step by pressing one of the buttons (configuration, calibration or toolbox).

4.2. ARGUS service Configuration

Important remark: After configuration change, a function check and control measurement has to be done! 4.2.1. Configuration tree structure The configuration is split into 4 parts: Configuration (part 1) All configuration possibilities (indexes) mentioned in chapter 2.7 can be modified herein in its own tree structure as shown below. All indexes which are different from the pump firmware default are high lighted. Calibration (part 2) Details of the pump calibration can be read out of the pump. The calibration cannot be modified herein. Statistic (part 3) Details about last used infusion parameters, total of infused volume and infusion time and so on are shown. Also the last technical failure numbers are listed in this part. ARGUS (part 4) This part contains ex-works settings (e.g. pump serial number)

# 20, 22-30, 32-34, 369 # 1, 7, 8, 15, 16, 45, 46, 48, 56, 366 # 2-5, 14, 39, 49, 361, 365 # 11-13, 17, 38, 40-42, 60 14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 20 / 48

ARGUS service
# 6, 10, 18, 19, 43, 44, 47, 50, 51, 351, 362, 363, 368, 390-394 # 101, 310-318; 102, 320-328; 103, 330-338; 104, 340348 # 201-217; 221-237; 241-257; 261-277 # 55, 514-517 # 600-800, 561-592

4.2.2. How to edit a configuration The following procedure describes how to edit a pump configuration: 1. Press the green Edit button. 2. The software will ask for the pump PIN code as next. The button Edit changes its colour and will be renamed into Download. 3. If you want to import a configuration from a file press the Import button, otherwise skip this point. 4. Select Configuration in the structure tree in the left upper frame. 5. Select the index group you want to modify by selecting the according structure tree and the according index. 6. Modify the according index (within the given restrictions shown). Each value (number) must be acknowledged by the green Enter button. Go through point 5 & 6 for all further indexes you want to modify. 7. Press the Download button if you want to save the modified configuration on the pump. Otherwise you can save the modified configuration into a file by pressing the Save button. 8. Make a functional check on each pump you have configured. Important remark! If a config. has been edited (performed point 1 and 2) once do not switch off the pump! Otherwise the pump will change always into the PC-configuration mode automatically. 4.3. ARGUS service - Calibration (volumetric pums only) Use this feature to easily calibrate the volume delivered by the volumetric pump.

To do this, it is possible to enter the measured rate (of an infusion device tester e.g.) or from a measured volume (of a balance) to define a correction factor according chapter 6.5.2.

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

21 / 48

ARGUS service The calculated correction factor can be applied to IV-sets which are released in the config. (index #101 - 104) only. Please be sure which IV-set you want to calibrate before you press the apply button. Per ex-works settings, only IV-Set number 1 is calibrated. Important remark! It is mandatory to perform also a pressure calibration according chapter 6.3. if a new IV-set will be used. We recommend to do a control measurement after changing the correction factor! 4.4. ARGUS service - Toolbox With the ARGUS service PC-tool you can set the pump clock, change PIN code, read and print out history, etc.

Select the next step by pressing one of the buttons (set clock, change PIN code, view history, service interval or replicate).

4.4.1. ARGUS service - Toolbox - Pump clock Use this feature to synchronize to pump internal clock with your PC time.

Please note: The pump internal clock will set to the central European time zone (Bern, CET, GMT +1.00h) as ex-works settings, the pump internal clock will not switch automatically between summer and winter time. All history logs (refer to chapter 4.4.5) will base on this time.

4.4.2. ARGUS service - Toolbox - PIN code Use this feature to set the pump PIN code.

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

22 / 48

ARGUS service The setting of a PIN code prevents access to the pump configuration of third persons. The default PIN code is 0 by ex-works settings. Please note: The PIN code corresponds with the PIN code mentioned in chapter 2.3.2. If a PIN code greater than 9999 is entered, the pump configuration can only be accessed using the ARGUS service PC tool. 4.4.3. ARGUS service - Toolbox - Service interval Use this feature to set a reminder alarm on the pump for the next service interval. A pending reminder alarm will be shown on the pump display after power up by a flashing CtrL text accompanied by an acoustic sound. The point in time when an active reminder alarm occurs, is given by the settings of the configuration (#393 and #394) and the pump internal clock. Any value higher than 0 on those indexes will release the reminder alarm after the service interval has elapsed. Please check those settings first, before you set the reminder alarm! Please note: By the ex-works settings, the reminder alarm is disabled. 4.4.4. ARGUS service - Toolbox - Replicate Use this feature to replicate fast and easily pump configuration from a saved configuration file or from a pump to another. A configuration can only be replicated if the saved configuration (and pump type) corresponds with the firmware of the connected pump in the first 2 digits (for e.g. 4.30 to 4.31 is possible).

Please note: The pump internal clock and remainder alarm settings must be done individually on each pump!

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

23 / 48

ARGUS service

4.4.5. ARGUS service - Toolbox - Pump history Each registered event has his own date & time stamp. An event is registered on each pump status change. Please refer to the complete list mentioned in chapter 4.4.6 below.

4.4.6. History messages Possible messages appearing in the description of each history event:
No information available Battery defective Battery low prealarm Battery low, pump stop Airbubble, pump stop Door open, pump stop No drops, pump stop Not enough drops, pump stop Too many drops, pump stop Neonatology mode set in PrL PrLimit change Bolus total reached Pump has detected failure Total volume reached, pump stop Total (VTBI) change Pump start Pump stop (KVO) Pump off Pump on Transport off Transport on Data lock off Data lock on Rate change Inf-Set change Infusm cleared Bolus start Bolus stop External power on External power off Pump start in remote mode Rate change during remote mode Pump off in remote mode Enter setup mode Enter PC configuration mode Exit setup or PC configuration mode Downstream occlusion, pump stop Upstream occlusion, pump stop Timer alarm, pump stop (KVO) PC communication timeout reached Logon in PC configuration mode Logoff in PC configuration mode Pump start, ext. changed parameters Any defaults written in EEPROM area CRC error in PC configuration module

4.4.7. History printout example

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

24 / 48

ARGUS service

All pre-alarms, alarms and technical failures are high lighted in a different colour.

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

25 / 48

SOFTWARE UPDATES 5. SOFTWARE UPDATES 5.1. General This chapter describes the procedure to perform a software update on the ARGUS 707 infusion pump. To check the installed software release in your ARGUS 707 V press the "MODE" key while switching on the pump. Please refer to your local distributor or ARGUS Medical AG to determine the latest software release able to run on your device hardware. 5.2. Requirements for a software update To update an ARGUS Medical device, the following items are needed: PC with Microsoft Windows 2000 or newer, .NET Framework must be installed! RS-232 serial interface cable (part no. 10.093) PC configuration tool ARGUS service Latest firmware included in a text file named A707_xxx.txt. (xxx is the placeholder for the firmware version). Those items are available from your local distributor or from ARGUS Medical AG. 5.3. Software update procedure 5.3.1. General Please carefully check the software present installed on the pump. If you have a firmware < version 4.xx please follow chapter 5.3.3 to upgrade the firmware. 5.3.2. Update of a pump with firmware > V4.xx Important remark! The actual calibration (and configuration) will be stored in a file on the PC, please be sure you will restore the correct file into the pump after the firmware update. Otherwise invalid calibration values will be stored on the pump. 1. Connect the pump to the serial interface of your PC. Please remember the COM port number where you have connected the pump. 2. Switch the pump ON while keeping key [10] pressed. 3. Start the PC configuration tool ARGUS service and select the according COM port. 4. Go into the configuration part and save the present pump configuration (incl. calibration) to a file. 5. Close the ARGUS service and switch the pump OFF. 6. Perform point 3 again, go into the Update center.

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

26 / 48

SOFTWARE UPDATES 7. Select the requested pump firmware file by pressing the button . 8. Press Update the pump firmware. Follow the instructions displayed on the PC. The firmware will be installed and the pump will be switched off automatically. 9. Go into the configuration part again (refer to point 2-3). Press the Edit button and enter the pump PIN code (default PIN after firmware update is 0). 10. Restore the old configuration (incl. calibration) from the previous created file. 11. Restore the configuration by pressing the Download button. 12. Perform a standard safety check (see chapter 10), normally the calibration will not be destroyed if the procedure is carefully performed step by step. 5.3.3. Upgrade of a pump with firmware < V4.xx With the Update center it is also possible to upgrade pump firmware older than V4.xx. Important remark! The actual calibration (and configuration) will be stored temporary on the PC, please perform the upgrade procedure pump by pump. Otherwise invalid calibration values will be stored on the pump. It is urgent necessary to perform a standard safety check (see chapter 10)! 1. Go into the Update center (see point 1-4 of chapter 5.3.2.): 2. Switch the pump ON by keeping the key [10] pressed. 3. Backup the legacy configuration (present configuration before the firmware update). This may take several seconds. 4. Switch the pump OFF. 5. Press Upgrade to bootloader 4. Follow the instructions displayed on the PC. The bootloader will be upgraded then. 6. Select the requested pump firmware file by pressing the button . 7. Press Update the pump firmware. Follow the instructions displayed on the PC. The firmware will be installed and the pump will be switched off automatically.
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 27 / 48

SOFTWARE UPDATES 8. Switch the pump ON while keeping key [10] pressed. Start the ARGUS service tool and select the according COM port. 9. Import configuration from backup. The calibration values and configuration of last connected pump will be imported. 10. Download it to the pump by pressing the Download button. 11. Important: Perform a standard safety check (see chap.10), the calibration values maybe lost during the upgrade procedure!

5.4. Safety aspects Be aware of the following points:

! !

For medical device traceability your local distributor or ARGUS Medical AG needs to be informed about every device updates (serial number) you performed! Do not make any software updates when the device is used and/or connected to a patient!

CAUTION! A standard safety check (see chap.10) has to be performed after every software update!

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

28 / 48

MAINTENANCE 6. MAINTENANCE 6.1. General CAUTION! Only authorized persons who have been trained by ARGUS Medical AG or by the local distributor are allowed to service the ARGUS 707 V infusion pump. In case of repair request, send the unit with the filled out repair order form (see chapter 11) to the local distributor. Further information is available from: ARGUS Medical AG CH-3627 Heimberg / Switzerland E-mail: info@argusmedical.com CAUTION! The safety standard check (SSC) has to be performed at least every 24 month or after 10'000 hrs of operation. The check has to be done in accordance to the chapter 10. No special maintenance of the ARGUS 707 V infusion pump is necessary. There are no wear and tear parts. 6.2. Recalibration 6.2.1. General Unless otherwise specified by the customer, the ARGUS 707 V has been calibrated by the manufacturer with the CODAN L86 PVC (REF 43.4304) infusion set. If a different infusion set is used (see recommended list in the appendix of the user manual), a recalibration is required. CAUTION! A new set calibration always requires pressure sensors (up-/ downstream) and a volume calibration (see chapter 6.3 & 6.5)! It is mandatory to execute first the calibration procedure of the pressure sensors and afterwards the volume calibration.

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

29 / 48

MAINTENANCE 6.3. Pressure calibration 6.3.1. General The volumetric infusion pump ARGUS 707 V contains two pressure sensors: - One upstream sensor, bottle side (left input) - One downstream sensor, patient side (right output). A pressure calibration becomes necessary if: - the pressure control measurement is not accurate enough - a pressure sensor is replaced - the main board is replaced or a new set is configured Needed equipment: - a manometer with a resolution of 0.1 bar - a 3-way stop cock and a clamping shears - the chosen IV set (refer to the list including the recommended IV set in the user manual)

CAUTION!

The pressure calibration procedure always includes all of the following steps, including a control measurement.

Key functions during pressure calibration: By the "MODE" key you can switch to the next calibration step. By the "START/STOP" key you can store the value of the involved pressure sensor, this will be acknowledged by a sound (beep). By pressing the "1" key you can switch off and on the actual value of the pressure sensor in the display. The value is displayed in mV.
Upstream sensor Downstream sensor

854

727

By pressing the "10" key you can switch off and on the motor. Don't use it by the standard calibration. By pressing the "100" key, the pump rate can be changed from 100 to 200 ml/h and back (toggle action). It is recommended to reduce the rate when the pressure is near 1.2 (1.0) bar, this allows a more precise calibration.
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 30 / 48

MAINTENANCE

6.3.2. Enter into the pressure calibration mode

.399 1234

1. Enter into the configuration mode (see chapter 2.3.).

2. Select address 399 and press MODE.

3. Enter value 1234.

4. Press START/STOP. Display shows Set - 1-.

6.3.3. Set selection before calibrating


Before you select an IV set, check chap. 2.8.

1. Press key "1" to select the set number you want to calibrate (1 to 4).

2. Press START/STOP. Display shows CAL. door.

6.3.4. Calibrating the offset of both sensors


CAL. door

1. Remove the IV set and open the door.

2. Press START/STOP to store the values for the upand downstream sensor.

3. Press MODE. Display shows CAL 1.0b.

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

31 / 48

MAINTENANCE 6.3.5. Calibrating the downstream sensor


CAL. 1.0b

1. Take a new IV set filled with water. 2. Simulate a downstream occlusion by the 3-way stop cock before you insert the set. 3. Open the roller clamp, insert the new IV set and close the door. 4. The pump generates a pressure build-up automatically. When the pressure 1.0 bar reaches press START/STOP. 5. Release the pressure and press "MODE". Display shows uPSt rEAM.

6.3.6. Calibrating the upstream sensor part 1


uPSt rEAM 1. Install a clamping shears near the pump (bottle side) and wait 15 sec, then press START/STOP. * 2. Release the clamping shears and press MODE. Display shows CAL 0.2b.

* Remark:

By pressing key 1 the pressure signal in (mV) of the upstream sensor will be displayed, after approximately 15 seconds the value does not decrease that fast any more; at this point press START/STOP to store the value (refer to step 1). The software will not allow the value to drop lower than the previous saved door open offset value +50mV.

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

32 / 48

MAINTENANCE 6.3.7. Calibrating the downstream sensor


CAL 0.2b 1. Simulate a downstream occlusion by the 3-way stop cock. 2. The pump generates a pressure build-up automatically. When the pressure 0.2 bar reaches press START/STOP.

3. Press "MODE". Display shows CAL 1.2b.

CAL 1.2b

4. The pump continues on with the pressure build-up. When the pressure 1.2 bar reaches press START/STOP. 5. Release the pressure and press "MODE". Display shows CAL -1.0b.

6. Close the roller clamp and open the door. The pump pumps reverse.

Remark: The pump must reach a pressure of at least 1.2 bar. Otherwise there might be a mechanical problem (Check the pump unit and the door).

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

33 / 48

MAINTENANCE

6.3.8. Calibrating the upstream sensor part 2

CAL -1.0b

1. Simulate an upstream occlusion by the 3-way stop cock. 2. Open the roller clamp, insert the IV set in the reverse direction and close the door. 3. The pump generate a pressure buildup automatically. When the pressure 1.0 bar reaches press START/STOP. 4. Release the pressure and press "MODE". Display shows CAL -0.2b.

CAL -0.2b

5. Simulate an upstream occlusion by the 3-way stop cock.

6. The pump generates a pressure build-up automatically. Is the pressure of 0.2 bar reached press START/STOP.

7. Press "MODE". Display shows CAL -1.2b.

CAL -1.2b

8. The pump continues on with the pressure build-up. When the pressure 1.2 bar reaches press START/STOP. 9. Release the pressure and press "MODE". Display shows SEt - 1-. 10. Switch the pump off. Close the roller clamp and remove the IV set. 11. Perform a control measurement according chapter 6.4.

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

34 / 48

MAINTENANCE 6.4. Pressure control measurement 6.4.1. Downstream control measurement

1. Insert the IV set and close the door. Open the roller clamp. 2. Switch on the pump and set the infusion rate at 200 ml/h. Start the infusion. After 10 seconds you have to simulate a downstream occlusion. 3. At 700 mbar the pump must stop and release an alarm. The tolerance is +150/-50 mbar.

Remark: A pressure limit of 700 mbar must not be the default value in menu PrL (configurable), please select 700 mbar as default for this control measurement. If the result of this control measurement does not fulfil the stated requirement, a pressure calibration according to chapter "Pressure calibration" has to be carried out again.

6.4.2. Upstream control measurement Start an infusion at a rate of 200 ml/h according the user manual. Simulate an upstream occlusion by installing a clamping shears (see chapter 6.3.6) or by kinking the tube. To avoid the drop alarm release, simulate falling drops on the drop detector. The pump must stop after several seconds and an upstream alarm (right hand LED in the pressure bar graph) must be activated. If the result of this control measurement does not fulfil the stated requirement, a pressure calibration according to chapter "Pressure calibration" has to be carried out again.

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

35 / 48

MAINTENANCE

6.5. Volume calibration 6.5.1. General On the A707 Volumetric pump it is possible to calibrate up to 4 different IV-Sets. For each IV-set you can define a correction factor, which calibrate the volume delivered. There are two ways to calibrate the volume delivered by the ARGUS 707 Volumetric pump, select one: By entering the correction factor With the internal calibration program of the pump Needed equipment: balance with a resolution of 0,1g at least

6.5.2. Volume calibration of the pump by defining the correction factor Please be sure, which IV-set you want to calibrate. If it is not enabled in the configuration, please enable it first and select the according IV-set in the normal mode of the pump in menu Set -x-. 1. Insert a new IV set (only recommended IV sets may be used, see appendix in the user manual) in the pump and perform a warm up infusion of 40 ml at an infusion rate of 999.9 m/h! 2. Infuse now a volume of 30 ml of water in a measuring cup on a zeroed balance at an infusion rate of 250 ml/h. 3. Determine the weight of the delivered water. 4. Connect the pump to a PC and start the ARGUS service PC tool as described in chapter 4, go into the calibration part of this tool. 5. Refer to chapter 4.3, select by VOLUME on the left side in the calibration window. 6. Enter 30 in the target volume setting field, enter the measured volume (weight) in the actual measurement reading field. The deviation will be calculated by the PC software. 7. Apply this value to the IV-Set you want to calibrate. 8. Perform a control measurement according to the steps 2 & 3 above. Repeat the calibration procedure if necessary.

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

36 / 48

MAINTENANCE 6.5.3. Volume calibration of set 1 with the pump integrated calibration program Needed equipment: balance with a resolution of 0,1g at least 1. Insert a new IV set (only recommended IV sets may be used, see appendix in the user manual) in the pump (filled with water) and connect the infusion line to a measuring cup on a balance. Enter into the configuration mode as described in chapter 2.3. Enter at address 399 the value "123". Press key START to acknowledge the entered value. Now you can select your preferred set (in our case set 1). Press the key START/STOP. The pump delivers a volume of 40 ml at an infusion rate of 999.9 ml/h (tArA is flashing). When tArA stops flashing, reset the balance to 0. Press the key START/STOP. The pump will display tM 432 (infusion delivery time is 432 seconds). It should now deliver a volume of 30 ml at an infusion rate of 250 ml/h. After the delivery time has elapsed, the pump stops and shows bAL. 30.00 in the displays. Enter now the value of the balance, e.g. 29.80g. This value must be within the range of 25.50 34.50. Otherwise switch the pump off and restart the volume calibration. Press the key START/STOP to acknowledge the entered value. The pump displays the new correction factor, e.g. Cor. 993.

2. 3.

4. 5. 6.

7.

8. 9.

10. Press the key START/STOP to store the new correction factor acknowledged by a buzzer sound. The pump display changes back to SEt -1- again. 11. Switch the pump off. Do not remove the infusion set! 12. Perform a control measurement with an infusion rate of 250 ml/h and an infusion total of 30 ml. Repeat the calibration procedure if necessary. For an optimized long time accuracy over 24 hours, we recommend to do a control measurement over 24 hours on a rate of 25 ml/h.

6.6. Pump specifications Please refer to the user manual for the specifications (chapter 9).

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

37 / 48

MAINTENANCE 6.7. Fault codes A technical failure will be signalled by the pump with a continuous alarm display and a continuous sound. During this state, the fault code which causes the pump to fail can be displayed by pressing the key MODE. If the pump was switched OFF after a detected failure, the fault code will be stored in the configuration of the pump (see index 380 - 389 via ARGUS service PC-tool). The possible fault codes registered in the configuration are listed in the table below:
Fault Code F_21 F_22 F_23 F_24 F_25 F_26 F_27 F_28 F_29 F_32 F_33 F_37 F_38 F_39 F_40 F_44 F_45 F_46 F_47 F_48 F_49 F_50 F_51 F_52** F_53 F_54 F_55 F_56 F_57 F_58 Failure ROM test ROM check (Runtime) RAM test/check XRAM test/check CPU test Invalid function menu EEPROM data invalid RTC data invalid, no RTC etc. Stepper motor power test (delayed 5s) 5Volt supply out of range 24Volt supply out of range (delayed 5s) Downstream pressure sensor test failed (always > 4.7V or < 0.2V, delayed 5s) Upstream pressure sensor test failed (always > 4.7V or < 0.2V, delayed 5s) * Downstream pressure sensor test failed (dynamical test failed) Upstream pressure sensor test failed (dynamical test failed) * Address invalid for config-eeprom Address invalid for history-eeprom Frequency from uC or RTC out of range Display-print not present Key(s) too long active Sensor-print not present AIL (Air in line) detector test failed Movement test failed (Home-Pulse < (Hall / Home)) Movement test failed (Home-Pulse > (Hall / Home)) More than one rotation at 'STOP' without 'KVO' Infused sum <> Calculated sum (Rotations) Frequency calculation Invalid volume adjustment over time Rotation (SW overflow) Internal volume control (10/ml)

We recommend replacing the main board in case a fault code is not included in this list above. * ** Failure released only if configured with upstream sensor (index 51=Yes). Fault code F_52 may occur after a software update or if peristaltic is manually turned during service. In this case, turn the pump off then on again and restart the pump. If a failure free start can be performed, omit this fault code.
ARGUS Medical AG 28.03.06 / PJ 38 / 48

14.203.B_A707 en.SM.V4.3X

REPLACEMENT OF PARTS 7. REPLACEMENT OF PARTS 7.1. General CAUTION! The ARGUS 707 V may only be used with accessories and spare parts which have been approved by ARGUS Medical AG for safe technical use. CAUTION! If a door, a housing, a pressure sensor or a main board is replaced, a full calibration (pressure sensors and volume calibration) is required. 7.2. Disassembling of the ARGUS 707 V CAUTION! Disconnect the mains cable from the power outlet before opening the housing! Observe the antistatic protection rules when disassembling the ARGUS 707 V (the use of an antistatic table mat and a grounded clip are recommended). 7.2.1. Disassembly of the case

1. Remove the four screws on the back side.

2. Separate the front and the back side.

3. Disconnect the battery, the supply and the docking interface connection.
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 39 / 48

REPLACEMENT OF PARTS 7.2.2. Removing the pump door

1. Gently press this hinge stopper backwards. 2. Pull the shaft towards the centre. Repeat point 1 and 2 for the other side of the door and then remove the door.

7.2.3. Disassembling the pump unit

1. Unscrew these screws. Remove the pump unit.

7.2.4. Assembling the pump


1. Use the screw with two washers at these places (torque 1.0 Nm)!

2. Use the screw without a washer at these places (torque 1.0 Nm)!

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

40 / 48

REPLACEMENT OF PARTS

3. Reinstall the door. Make sure that a washer has been installed on each side!

4. Make sure that this part touches nowhere, otherwise noises may be generated!

5. Reconnect the back side with the front side (battery, power & docking interface)!

6. Make sure that both cables are correctly placed in the notch before the main board is inserted!

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

41 / 48

REPLACEMENT OF PARTS

7. Use the short screw at these places (torque 1.1 Nm)!

7. Use the long screw at these places (torque 1.1 Nm)!

Mind the torque for the screws listed below: What: Case Pump unit Combination clamp Pressure sensor Stop flow Air detector Sensor board Display board Transformer Clamp holder Mains plug Edge board Where Backplane into front Into front Into clamp holder Into front Into front Into front Into pump unit Into front Into Backplane Into Backplane Into Backplane Into clamp holder Torque in Nm 1.10 1.00 1.50 0.60 0.60 0.60 0.40 0.50 1.00 0.50 0.50 0.50

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

42 / 48

REPLACEMENT OF PARTS 7.3. Spare parts

10.087 Combination clamp

10.089 External drop detector

10.093 Interface cable docking pumps

10.119 Battery NiMH 12V/1500mAh

10.120 Display board A707

10.121 Pressure sensor

10.122 Air detector

10.124 Main board A707

10.123 Stop flow A707


14.203.B_A707 en.SM.V4.3X

10.125 Sensor board A707


ARGUS Medical AG 28.03.06 / PJ

10.126 Pump unit A707


43 / 48

REPLACEMENT OF PARTS

10.127 Door complete A707

10.128 Casing back plane A707 30VAC complete

10.136 Casing back plane A707

10.129 Casing forepart (without door) complete

10.137 Casing forepart

11.221 Sealing A707

12.042 Cable staff alert 2m

12.011 Interface cable 8pol/2m & 12.012 Interface adapter

11.268 Alarm lights

11.132 Label external drop detector


14.203.B_A707 en.SM.V4.3X

11.237 Identification plate A707

11.238 Label flow direction

ARGUS Medical AG 28.03.06 / PJ

44 / 48

REPLACEMENT OF PARTS

10.135 Edge board complete A707

Part no. 11.222 Front panel A707

12.035 Pressure gauge with stopcock

11.219 + 11.239 11.247 Short instructions


(SW, DE, EN, FR, DK, NL, CZ, PT, SP, IT)

11.005 Bottle holder 45 cm 11.043 Bottle holder 60 cm

12.044 Plug

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

45 / 48

WIRING DIAGRAMM 8. WIRING DIAGRAMM

9. BLOC SCHEMATIC
100 Mode Start/Stop
Buzzer

Nurse-call

ml total

On/Off

ml inf. h. min DOOR

RS-232

KVO

RS232 Display Driver

Motor

Motor Driver

Reset + Watchdog Power Switch Test Switcher

Latch

FLASH

EEPROM AC Main AC DC Ext. DC DC Ni-Cd Akku 12V DC + 5V (Logic) Air Bubble Detector Drop Detector Pressure Sensor (in) Door Control Pressure Sensor (out) Motion Control DC DC + 24V (Motor + Buzzer) 512 Hz I C RTC
2

Li-Batt. 3 Volt

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

Pressure

1 3

Controller Alarm Relay

10

0,1

46 / 48

SAFETY STANDARD CHECK 10. SAFETY STANDARD CHECK

Safety Standard Check (SSC)


Serial-no.: ......... Hospital: .

ARGUS 707 V_en


Inventory-no.: ........ Department: ...

Customer: .

The SSC has to be performed at least every 24 months or after 10'000 hours of operation. The check has to be done in accordance to the user- and service manuals. 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19
Electrical test according to EN60601-1
(all measurements made with a power cable 2.5m)

Check if a software upgrade is required? Visual check for damage, cleanness and completeness Test the function of the stop flow clamp Keep "MODE" pressed while switching on the pump Connect/disconnect the pump to the mains Test the drop detector by simulating drops Test the door switch, open and close the door Open the door and remove any IV set Install a water filled IV set, close the door Set rate to 333.3 ml/h, press "START", disconnect the drop detector Press "MODE" Check the external connector nurse call Calibration of the pressure sensors IV set type used: Codan .... Other . Volume calibration Charge the battery while the pump is running 16 hours, at a rate of 30.0 ml/h Battery check at a rate of 30 ml/h. Run the battery test until the pump switches off automatically Printout the pump history (refer to chapter 4.4.5.) - Housing, labels, accessories, connectors, power cable, etc. - Proper movement of the clamp
-

Display pump type and software release Display of 2, 4, 7, F., in numeric display Display of all operation- and alarm indicators The indicator external supply turns on/off Check the green drop indicator Door open the door indicator lights up Door close the door indicator turns off Indicator air bubble lights up Indicator air bubble turns off The red indicator drop lights up (delayed) The acoustical alarm turns on The acoustical alarm mutes Relay contact switches (see chapter 9) See chapter 6.3. and 6.4. of the service manual See chapter 6.5. of the service manual

- The indicator external supply must light - The green indicator "battery" lights up during this test - Check the pump internal time and data - Check the battery run time by checking the latest history entries, > 4h 30min - Visual check of mains connector - Measurements attached

Charge the battery after this test!

The pump has passed the SSC and is safe for use
Date / Name: ..... Signature: ............

ARGUS Medical AG
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 47 / 48

REPAIR ORDER FORM 11. REPAIR ORDER FORM

ARGUS Medical AG / Heimberg Switzerland REPAIR ORDER FORM


Purchase order / Proforma invoice number:
Customer name and address:

Name of contact person: A414 A400 A404 A200 A300 Accessory: Detailed failure or problem description: ARGUS 100 P ARGUS 100 M ARGUS 600 S ARGUS 707 V ARGUS 708 V

Tel. number:

Device:

Serial Number: Serial Number / Production code:

Expected work / repair to be done: Repair Warranty repair Replacement Other Description:

Date:

Signature:

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

48 / 48