Sie sind auf Seite 1von 113

Waters 432 Conductivity Detector

Operators Guide

34 Maple Street Milford, MA 01757 71500043202, Revision A

NOTICE
The information in this document is subject to change without notice and should not be construed as a commitment by Waters Corporation. Waters Corporation assumes no responsibility for any errors that may appear in this document. This document is believed to be complete and accurate at the time of publication. In no event shall Waters Corporation be liable for incidental or consequential damages in connection with or arising from the use of this document.

19942003 WATERS CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA. ALL RIGHTS RESERVED. THIS DOCUMEBNT OR PARTS THEREOF MAY NOT BE REPRODUCED IN ANY FORM WITHOUT THE WRITTEN PERMISSION OF THE PUBLISHER.

Alliance, Millennium, and Waters are registered trademarks, and Empower, LAC/E, PowerLine, SAT/IN, Sep-Pak, UltraWISP and WISP are trademarks of Waters , Corporation. All other trademarks or registered trademarks are the sole property of their respective owners.

Note: When you use the instrument, follow generally accepted procedures for quality control and methods development. If you observe a change in the retention of a particular compound, in the resolution between two compounds, or in peak shape, immediately determine the reason for the changes. Until you determine the cause of a change, do not rely on the separation results. Note: The Installation Category (Overvoltage Category) for this instrument is Level II. The Level II Category pertains to equipment that receives its electrical power from a local level, such as an electrical wall outlet. Atencin: Changes or modifications to this unit not expressly approved by the party

STOP responsible for compliance could void the users authority to operate the equipment.
Important : Toute modification sur cette unit nayant pas t expressment approuve par lautorit responsable de la conformit la rglementation peut annuler le droit de lutilisateur exploiter lquipement. Achtung: Jedwede nderungen oder Modifikationen an dem Gert ohne die ausdrckliche Genehmigung der fr die ordnungsgeme Funktionstchtigkeit verantwortlichen Personen kann zum Entzug der Bedienungsbefugnis des Systems fhren. Avvertenza: eventuali modifiche o alterazioni apportate a questa unit e non espressamente approvate da un ente responsabile per la conformit annulleranno lautorit dellutente ad operare lapparecchiatura. Atencin: cualquier cambio o modificacin efectuado en esta unidad que no haya sido expresamente aprobado por la parte responsable del cumplimiento puede anular la autorizacin del usuario para utilizar el equipo.

Caution: Use caution when working with any polymer tubing under pressure: Always wear eye protection when near pressurized polymer tubing. Extinguish all nearby flames. Do not use Tefzel tubing that has been severely stressed or kinked. Do not use Tefzel tubing with tetrahydrofuran (THF) or concentrated nitric or sulfuric acids. Be aware that methylene chloride and dimethyl sulfoxide cause Tefzel tubing to swell, which greatly reduces the rupture pressure of the tubing. Attention : Soyez trs prudent en travaillant avec des tuyaux de polymres sous pression : Portez toujours des lunettes de protection quand vous vous trouvez proximit de tuyaux de polymres. Eteignez toutes les flammes se trouvant proximit. N'utilisez pas de tuyau de Tefzel fortement abm ou dform. N'utilisez pas de tuyau de Tefzel avec de l'acide sulfurique ou nitrique, ou du ttrahydrofurane (THF). Sachez que le chlorure de mthylne et le sulfoxyde de dimthyle peuvent provoquer le gonflement des tuyaux de Tefzel, diminuant ainsi fortement leur pression de rupture. Vorsicht: Bei der Arbeit mit Polymerschluchen unter Druck ist besondere Vorsicht angebracht: In der Nhe von unter Druck stehenden Polymerschluchen stets Schutzbrille tragen. Alle offenen Flammen in der Nhe lschen. Keine Tefzel-Schluche verwenden, die stark geknickt oder berbeansprucht sind. Tefzel-Schluche nicht fr Tetrahydrofuran (THF) oder konzentrierte Salpeter- oder Schwefelsure verwenden. Durch Methylenchlorid und Dimethylsulfoxid knnen Tefzel-Schluche quellen; dadurch wird der Berstdruck des Schlauches erheblich reduziert.

Precauzione: prestare attenzione durante le operazioni con i tubi di polimero sotto pressione: Indossare sempre occhiali da lavoro protettivi nei pressi di tubi di polimero pressurizzati. Estinguere ogni fonte di ignizione circostante. Non utilizzare tubi Tefzel soggetti a sollecitazioni eccessive o incurvati. Non utilizzare tubi Tefzel contenenti tetraidrofurano (THF) o acido solforico o nitrico concentrato. Tenere presente che il cloruro di metilene e il dimetilsolfossido provocano rigonfiamento nei tubi Tefzel, che riducono notevolmente il limite di pressione di rottura dei tubi stessi. Advertencia: manipular con precaucin los tubos de polmero bajo presin: Protegerse siempre los ojos en las proximidades de tubos de polmero bajo presin. Apagar todas las llamas que estn a proximidad. No utilizar tubos Tefzel que hayan sufrido tensiones extremas o hayan sido doblados. No utilizar tubos Tefzel con tetrahidrofurano (THF) o cidos ntrico o sulfrico concentrados. No olvidar que el cloruro de metileno y el xido de azufre dimetilo dilatan los tubos Tefzel, lo que reduce en gran medida la presin de ruptura de los tubos.

Caution: The user shall be made aware that if the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. Attention : Lutilisateur doit tre inform que si le matriel est utilis dune faon non spcifie par le fabricant, la protection assure par le matriel risque dtre dfectueuses. Vorsicht: Der Benutzer wird darauf aufmerksam gemacht, dass bei unsachgemer Verwenddung des Gertes unter Umstnden nicht ordnungsgem funktionieren. Precauzione: lutente deve essere al corrente del fatto che, se lapparecchiatura viene usta in un modo specificato dal produttore, la protezione fornita dallapparecchiatura potr essere invalidata. Advertencia: el usuario deber saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas de proteccin del equipo podran ser insuficientes.

Caution: To protect against fire hazard, replace fuses with those of the same type and rating. Attention : Remplacez toujours les fusibles par dautres du mme type et de la mme puissance afin dviter tout risque dincendie. Vorsicht: Zum Schutz gegen Feuergefahr die Sicherungen nur mit Sicherungen des gleichen Typs und Nennwertes ersetzen. Precauzione: per una buona protezione contro i rischi di incendio, sostituire i fusibili con altri dello stesso tipo e amperaggio. Advertencia: sustituya los fusibles por otros del mismo tipo y caractersticas para evitar el riesgo de incendio.

Caution: To avoid possible electrical shock, disconnect the power cord before servicing the instrument. Attention : Afin dviter toute possibilit de commotion lectrique, dbranchez le cordon dalimentation de la prise avant deffectuer la maintenance de linstrument. Vorsicht: Zur Vermeidung von Stromschlgen sollte das Gert vor der Wartung vom Netz getrennt werden. Precauzione: per evitare il rischio di scossa elettrica, scollegare il cavo di alimentazione prima di svolgere la manutenzione dello strumento. Precaucin: para evitar descargas elctricas, desenchufe el cable de alimentacin del instrumento antes de realizar cualquier reparacin.

Commonly Used Symbols Direct current Courant continu Gleichstrom Corrente continua Corriente continua

Alternating current Courant alternatif Wechselstrom Corrente alternata Corriente alterna

Protective conductor terminal Borne du conducteur de protection Schutzleiteranschluss Terminale di conduttore con protezione Borne del conductor de tierra

Commonly Used Symbols (Continued) Frame or chassis terminal Borne du cadre ou du chssis Rahmen- oder Chassisanschluss Terminale di struttura o telaio Borne de la estructura o del chasis

Caution or refer to manual Attention ou reportez-vous au guide Vorsicht, oder lesen Sie das Handbuch Prestare attenzione o fare riferimento alla guida Acte con precaucin o consulte la gua

Caution, hot surface or high temperature Attention, surface chaude ou temprature leve Vorsicht, heie Oberflche oder hohe Temperatur Precauzione, superficie calda o elevata temperatura Precaucin, superficie caliente o temperatura elevada

Commonly Used Symbols (Continued) Caution, risk of electric shock (high voltage) Attention, risque de commotion lectrique (haute tension) Vorsicht, Elektroschockgefahr (Hochspannung) Precauzione, rischio di scossa elettrica (alta tensione) Precaucin, peligro de descarga elctrica (alta tensin)

Caution, risk of needle-stick puncture Attention, risques de perforation de la taille dune aiguille Vorsicht, Gefahr einer Spritzenpunktierung Precauzione, rischio di puntura con ago Precaucin, riesgo de puncin con aguja

UV

Caution, ultraviolet light Attention, rayonnement ultrviolet Vorsicht, Ultraviolettes Licht Precauzione, luce ultravioletta Precaucin, emisiones de luz ultravioleta

Commonly Used Symbols (Continued) Fuse Fusible Sicherung Fusibile Fusible

Electrical power on Sous tension Netzschalter ein Alimentazione elettrica attivata Alimentacin elctrica conectada

Electrical power off Hors tension Netzschalter aus Alimentazione elettrica disattivata Alimentacin elctrica desconectada

432 Conductivity Detector Information


Intended Use
The Waters 432 Conductivity Detector can be used for in-vitro diagnostic testing to analyze many compounds, including diagnostic indicators and therapeutically monitored compounds. When you develop methods, follow the Protocol for the Adoption of Analytical Methods in the Clinical Chemistry Laboratory, American Journal of Medical Technology, 44, 1, pages 3037 (1978). This protocol covers good operating procedures and techniques necessary to validate system and method performance.

Biological Hazard
When you analyze physiological fluids, take all necessary precautions and treat all specimens as potentially infectious. Precautions are outlined in CDC Guidelines on Specimen Handling, CDC NIH Manual, 1984.

Calibration
Follow acceptable methods of calibration with pure standards to calibrate methods. Use a minimum of five standards to generate a standard curve. The concentration range should cover the entire range of quality-control samples, typical specimens, and atypical specimens.

Quality Control
Routinely run three quality-control samples. Quality-control samples should represent subnormal, normal, and above-normal levels of a compound. Ensure that quality-control sample results are within an acceptable range, and evaluate precision from day to day and run to run. Data collected when quality-control samples are out of range may not be valid. Do not report this data until you ensure that chromatographic system performance is acceptable.

Table of Contents
Preface ....................................................................................... 23

Chapter 1 Introduction ...................................................................................... 26 Chapter 2 Installing the Detector ...................................................................... 30 2.1 Selecting the Installation Site................................................ 30 2.2 Unpacking and Inspection..................................................... 31 2.3 AC Power Connection ........................................................... 32 2.4 I/O Signal Connections ......................................................... 35 2.4.1 I/O Signal Descriptions .............................................. 35 2.4.2 PowerLine Controller Connections ............................ 37 2.4.3 Empower and Millennium32 Connections .................. 39 2.4.4 Data Module Connections ......................................... 44 2.4.5 Chart Recorder Connections ..................................... 45 2.4.6 Chart Marker Input Connections................................ 45 2.4.7 Auto Zero Input Connections ..................................... 46 2.4.8 Alliance Separations Module Connections ................ 46 2.5 Making Fluidic Connections .................................................. 48 2.6 Passivating the System ......................................................... 53 2.7 Verifying the Detector............................................................ 54

Table of Contents

15

Chapter 3 Operating the Detector .................................................................... 56 3.1 Controls and Indicators ......................................................... 56 3.2 Startup and Shutdown .......................................................... 61 3.3 Operating Recommendations ............................................... 62 Chapter 4 Performing Ion Analysis ................................................................... 64 4.1 Fundamental Considerations ................................................ 64 4.2 Configuring the System......................................................... 68 4.3 Eluents for Ion Analysis......................................................... 69 4.3.1 Preparing Anion Eluent.............................................. 70 4.3.2 Preparing Cation Eluent ............................................ 70 4.4 Standards for Ion Analysis .................................................... 70 4.4.1 Preparing Anion Standards ....................................... 71 4.4.2 Injecting Anion Standards.......................................... 72 4.4.3 Preparing Cation Standards ...................................... 74 4.4.4 Injecting Cation Standards......................................... 76 Chapter 5 Maintenance .................................................................................... 78 5.1 Routine Maintenance ............................................................ 78 5.1.1 Replacing the Fuse.................................................... 78 5.1.2 Maintaining the Flow Cell........................................... 79 5.2 Cleaning the Detector Exterior.............................................. 82 5.3 Troubleshooting..................................................................... 82

Table of Contents

16

Appendix A Specifications ................................................................................... 87 Appendix B Spare Parts....................................................................................... 90 Appendix C Ion Chromatography Methods .......................................................... 91 C.1 General-Purpose Anion Analysis Using Conductivity and UV Detection ................................................................ 91 C.1.1 Preparing Eluent ....................................................... 93 C.1.2 Preparing Standards ................................................. 93 C.1.3 Preparing a Sample .................................................. 93 C.1.4 Empower Data Processing Method........................... 94 C.1.5 Method Validation ..................................................... 95 C.1.6 Method Linearity........................................................ 95 C.1.7 Quantitation Precision ............................................... 97 C.1.8 Method Detection Limits............................................ 97 C.1.9 Quantitation Accuracy ............................................... 98 C.1.10 Analyte Recovery .................................................... 99 C.1.11 Example of Use..................................................... 100 C.1.12 Using Direct UV Detection .................................... 100 C.1.13 Preparing Lithium Borate/Gluconate 50X Stock Concentrate ................................................... 102 C.1.14 Preparing Lithium Borate/Gluconate Eluent.......... 102

Table of Contents

17

C.2 Alkali and Alkaline Earth Cations, Ammonium, and Amines........................................................................ 103 C.2.1 Preparing Eluent .................................................... 104 C.2.2 Preparing Standards ............................................... 104 C.2.3 Preparing a Sample ................................................ 105 C.2.4 Empower Data Processing Method......................... 105 C.2.5 Method Detection Limits.......................................... 106 C.2.6 Examples of Use..................................................... 107 C.2.7 Preparing Stock Reagent ........................................ 108 Appendix D Validation Support .......................................................................... 109 Index ..................................................................................... 111

Table of Contents

18

List of Figures
1-1 1-2 2-1 2-2 2-3 2-4 2-5 2-6 2-7 2-8 2-9 2-10 2-11 2-12 2-13 3-1 4-1 4-2 4-3 4-4 5-1 5-2 C-1 Waters 432 Conductivity Detector ................................................. 26 Flow Cell Schematic ...................................................................... 28 Rear Panel ..................................................................................... 33 Changing the Voltage Setting ........................................................ 34 I/O Terminal Strip ........................................................................... 36 IEEE-488 Address Switch.............................................................. 38 Bus SAT/IN Module (Front Panel) .................................................. 40 Bus SAT/IN to Bus LAC/E Connections ......................................... 42 Bus SAT/IN to 432 Detector Connections...................................... 43 Alliance Separations Module Connections to the 432 Detector Auto-Zero on Inject.......................................................... 47 Alliance Separations Module Connections to the 432 Detector Chart Mark on Inject........................................................ 48 Fluid Connections .......................................................................... 49 Cutting Polymeric Tubing ............................................................... 50 Ferrule and Compression Screw Assembly................................... 51 Pulse Dampener ............................................................................ 53 Front Panel .................................................................................... 56 Soda Lime Tube............................................................................. 67 System Configuration for Ion Analysis........................................... 68 Chromatogram of a 7-Anion Standard........................................... 74 Chromatogram of an 8-Cation Standard........................................ 77 Installing Operating Voltage Fuses ................................................ 79 Flow Cell Assembly ....................................................................... 81 Common Anion Standards ............................................................ 92

List of Figures

19

C-2 C-3 C-4 C-5 C-6 C-7 C-8 C-9 C-10 C-11 C-12

Calibration Curves for Chloride, Fluoride, and Bromide ............... 95 Calibration Curves for Nitrite and Nitrate ...................................... 96 Calibration Curves for Sulfate and Phosphate .............................. 96 100-mL Injection ........................................................................... 97 Typical Drinking Water, No Dilution Required ............................ 100 Direct UV Detection .................................................................... 101 100-ppb Anion Standard ............................................................. 101 1-ppm Standard .......................................................................... 103 25-ppb Cation Standard .............................................................. 106 Typical Drinking Water, No Dilution Required ............................ 107 Typical Municipal Wastewater, Diluted 1:50, Overlay of Duplicate Injections.................................................................. 107

List of Figures

20

List of Tables
1-1 2-1 2-2 2-3 2-4 2-5 2-6 2-7 2-8 2-9 2-10 2-11 2-12 3-1 3-2 4-1 4-2 4-3 4-4 4-5 5-1 A-1 A-2 A-4 A-3 Limiting Equivalent Conductance of Ions in Water at 25 C ......... 29 Power Cord Wire Identification ..................................................... 33 Nominal Operating Voltage...................................................... 34 I/O Signal Descriptions ........................................................... 36 IEEE-488 DIP Switch Setting .................................................. 38 Bus SAT/IN Cable Connections ............................................... 44 Data Module Signal Cable Connections .................................. 44 Data Module Chart Mark Cable Connections........................... 45 Chart Recorder Cable Connections......................................... 45 Autosampler Chart Mark Cable Connections ........................... 46 Autosampler Auto Zero Cable Connections ............................. 46 Connections for Generating Auto-Zero on Inject ...................... 47 Connections for Generating Chart Mark on Inject .................... 48 Key Descriptions ........................................................................... 58 Setting the Beep Function ....................................................... 60 Shelf-Life of Standards ................................................................. 71 Salts for Anion Standard Concentrates.................................... 72 Anion Concentrate Dilutions .................................................... 72 Salts for Cation Standard Concentrates .................................. 75 Cation Concentrate Dilutions................................................... 75 Troubleshooting Guide .................................................................. 85 Operational Specifications ............................................................ 87 Mechanical Specifications ............................................................ 87 Electrical Specifications ................................................................ 88 Environmental Specifications........................................................ 88

List of Tables

21

A-5 B-1 C-1 C-1 C-2 C-3 C-4 C-5 C-6 C-1 C-2 C-3

Communications ........................................................................... 89 Spare Parts ................................................................................... 90 Required Instrumentation ............................................................. 91 Analysis Conditions ...................................................................... 92 IC Processing Method Using Peak Apex for Retention Time ....... 94 Method Validation ......................................................................... 95 Quantitation Precision................................................................... 97 Quantitation Accuracy................................................................... 98 Analyte Recovery.......................................................................... 99 Required Instrumentation ........................................................... 103 Analysis Conditions .................................................................... 104 IC Processing Method Using Peak Apex for Retention Time ..... 105

List of Tables

22

Preface
The Waters 432 Conductivity Detector Operators Guide details the procedures for unpacking, installing, operating, maintaining, and troubleshooting the 432 Conductivity Detector. It also includes appendixes listing specifications and spare parts and describing validation support. This guide is intended for use by personnel who need to install, operate, maintain, or troubleshoot the 432 Detector. This guide assumes an understanding of the principles of chromatography.

Organization
This guide contains the following: Chapter 1 describes the features and method of operation of the 432 Detector. Chapter 2 describes the procedures for installing the 432 Detector and making fluid and electrical connections. Chapter 3 describes the controls and indicators of the 432 Detector, and provides general operating instructions. Chapter 4 describes the system configuration, eluents, and standards recommended for anion and cation analysis. Chapter 5 describes simple maintenance procedures and provides troubleshooting tables to aid in problem diagnosis. Appendix A describes the operational specifications and requirements of the 432 Detector. Appendix B lists the recommended spare parts for the 432 Detector. Appendix C describes ion chromatography methods. Appendix D describes the recommended validation protocols and Waters validation support.

Related Documentation
Waters Licenses, Warranties, and Support: Provides software license and warranty information, describes training and extended support, and tells how Waters handles shipments, damages, claims, and returns.

23

Documentation on the Web Related product information and documentation can be found on the World Wide Web. Our address is http://www.waters.com.

Related Adobe Acrobat Reader Documentation


For detailed information about using Adobe Acrobat Reader, see the Adobe Acrobat Reader Online Guide. This guide covers procedures such as viewing, navigating, and printing electronic documentation from Adobe Acrobat Reader.

Printing This Electronic Document


Adobe Acrobat Reader lets you easily print pages, page ranges, or the entire document by selecting File > Print. For optimum print quantity, Waters recommends that you specify a PostScript printer driver for your printer. Ideally, use a printer that supports 600 dpi print resolution.

Documentation Conventions
The following conventions can be used in this guide: Convention Purple Usage Purple text indicates user action such as keys to press, menu selections, and commands. For example, Click Next to go to the next page. Italic indicates information that you supply such as variables. It also indicates emphasis and document titles. For example, Replace file_name with the actual name of your file. Courier indicates examples of source code and system output. For example, The SVRMGR> prompt appears. Courier bold indicates characters that you type or keys you press in examples of source code. For example, At the LSNRCTL> prompt, enter set password oracle to access Oracle. Indicates hypertext cross-references to a specific chapter, section, subsection, or sidehead. Clicking this topic using the hand symbol brings you to this topic within the document. Right-clicking and selecting Go Back from the shortcut menu returns you to the originating topic. For example, The detectors I/O signals are described in Section 2.4, I/O Signal Connections. The word key refers to a computer key on the keypad or keyboard. Screen keys refer to the keys on the instrument located immediately below the screen. For example, The A/B screen key on the 2414 Detector displays the selected channel.

Italic

Courier Courier Bold

Underlined Blue

Keys

24

Convention

Usage Three periods indicate that more of the same type of item can optionally follow. For example, You can store filename1, filename2, in each folder. A right arrow between menu options indicates you should choose each option in sequence. For example, Select File > Exit means you should select File from the menu bar, then select Exit from the File menu.

>

Notes Notes call out information that is helpful to the operator. For example: Note: Record your result before you proceed to the next step. Attentions Attentions provide information about preventing damage to the system or equipment. For example: Attention: To avoid damaging the detector flow cell, do not touch the flow cell

STOP window.
Cautions Cautions provide information essential to the safety of the operator. For example: Caution: To avoid burns, turn off the lamp at least 30 minutes before removing it for replacement or adjustment.

Caution: To avoid electrical shock and injury, unplug the power cord before performing maintenance procedures.

Caution: To avoid chemical or electrical hazards, observe safe laboratory practices when operating the system.

25

Chapter 1 Introduction
Features
The Waters 432 Conductivity Detector (Figure 1-1) is specifically designed to be integrated into chromatographic systems. The following features contribute to its performance in measuring the conductivity of column eluents: Unique 5-electrode flow cell design Heat exchanger and a built-in automatic temperature control system for stable operation Auto baseline/auto zero External recorder/integrator and chart mark connections Three time constant selections Leak-detected alarm signal

Conductivity

Waters 432
Detector

IN

OUT

TP01268

Figure 1-1 Waters 432 Conductivity Detector

26

Method of Operation
This section discusses the method of operation of the 432 Detector. Additional descriptive information appears in these sections: Section 2.4.1, I/O Signal Descriptions Section 3.1, Controls and Indicators Appendix A, Specifications Measurement Technique The 432 Detector responds to all ions present in the flow cell, since all ions in solution conduct electricity. This allows the 432 Detector to detect a wide variety of sample ions. The 432 Detector eliminates the eluents contribution to conductivity with an electronic technique called baseline suppression. The detector measures the eluent conductivity and assigns it a value of zero. Thus, any sample ions appear as positive or negative measurements, relative to the baseline. The temperature of an ionic solution affects the conductivity of the ions. Generally, a solutions conductivity rises about 2% for every degree Celsius of temperature increase. The special flow cell heater in the 432 Detector minimizes the effect of ambient temperature fluctuations on measurement accuracy. Flow Cell Design The flow cell in the 432 Detector contains five electrodes connected in a measuring circuit: two reference electrodes, two detection electrodes, and a guard electrode that provides a local electrical ground (Figure 1-2). Column eluent flows through the heater to attain the set temperature, and then flows through the cell, directly contacting the electrodes. The 5-electrode design permits measurement of conductivity to be made with a very low current at the detection electrodes. The low current employed eliminates impedance and other problems associated with simpler designs, and results in a stable baseline and an extended range of linearity.

Introduction

27

1= Reference Electrodes 2= Detection Electrodes 3= Guard Electrode Flow Cell Block (heated)

Fluid Outlet

TP01271

Figure 1-2 Flow Cell Schematic

Ion Detection Theory


The conductance of a solution of known concentration can be calculated using the following equation:

C G = ---------10 3
G = measured conductance of the solution, in Siemens (1 S = ohm1) C = concentration in equivalents per 1000 cm3 K = length/area of cell (the cell constant) = equivalent conductance in S cm2 equiv1 Table 1-1 lists the equivalent conductances of some common ions.1 Concentrations above 105 to 103 N, generally exhibit decreased equivalent conductance due to interionic effects.

28

Table 1-1 Limiting Equivalent Conductance of Ions in Water at 25 C Cations H


+

l+
349.8 38.6

Anions OH F Cl Br I NO 3 ClO3 ClO4 IO


4

198.6 55.4 76.4 78.1 76.8 71.5 64.6 67.4 54.5 54.6 40.9 32.4 80.0 69.3 111.0

Li + Na K
+ + + + + +

50.1 73.5 77.8 61.9 73.3 51.8 53.0 53.1 59.5 63.6 53.6 52.8 69.7 69.8

Rb Ag

NH4 Hg Mg Ca Ba Cu Zn La Ce

(CH3)2NH2
2+ 2+ 2+ 2+ 2+ 2+ 3+ 3+

Formate Acetate Benzoate SO4 2 CO3


2

Fe(CN)6
4

1. Henry H. Bauer et al., eds. Instrumental Analysis, Allyn and Bacon, Boston (1978), p. 115. Reprinted with permission from the publisher.

Introduction

29

Chapter 2 Installing the Detector


This chapter guides you through the following steps in preparing the 432 Detector for operation in a chromatographic system: Selecting an installation site that satisfies the detectors power and environmental requirements Unpacking and inspecting the 432 Detector and accompanying items Connecting the detector to your AC power supply Connecting the detector electrically to the other components of your chromatographic system Connecting the detector inlet to the column and the detector outlet to a waste receptacle (and, if required, installing the pulse dampener) Passivating the detector and other post-column fluid path components After you have successfully completed this chapter, familiarize yourself with the information in Section 3.1, Controls and Indicators. When you are ready to operate the detector, perform the startup procedure described in Section 3.2, Startup and Shutdown.

2.1 Selecting the Installation Site


Operating Environment
The 432 Detector operates in any standard laboratory environment that provides suitable electrical power and remains within the following ranges: Temperature: 5 to 35 C (40 to 95 F) Humidity: 20 to 80%, noncondensing Install the instrument in a clean area that is free from exposure to: Temperature or humidity extremes, which can be found near direct sunlight, heat registers, and air conditioning vents Strong electromagnetic radiation, such as from large motors or arcing contacts Appreciable shock or vibration

Selecting the Installation Site

30

Required Space
The 432 Detector requires bench space that measures approximately: 10 inches (25 cm) high 14 inches (34 cm) wide 24 inches (60 cm) deep Attention: Make sure that air can circulate freely through the ventilation slots on both

STOP side panels.

Power Requirements
The 432 Detector requires: One properly grounded AC voltage outlet. Correct voltage and fuse selections as shown in Table 2-2.

2.2 Unpacking and Inspection


Unpacking
The 432 Detector is shipped in one carton that contains the following items: Waters 432 Conductivity Detector Startup Kit Validation certificate Waters 432 Conductivity Detector Operators Guide Packing list Declaration of conformity Note: If you purchased the 432 Detector as part of an ion/liquid chromatograph system, a Waters representative will perform the installation and startup. To unpack the 432 Detector: 1. Locate the packing list. 2. Unpack the contents of the shipping carton and check the contents against the packing list to make sure that you received all items. 3. Check the contents of the Startup Kit against the Waters 432 Conductivity Detector Startup Kit List. 4. Save the shipping carton for future transport or shipment.

Installing the Detector

31

Inspection
Inspect all items. If you find any damage or discrepancy, immediately contact the shipping agent and Waters. For more information about the instrument warranty, refer to Waters Licenses, Warranties, and Support. If the shipment is complete and undamaged, record the installation date and serial number of the 432 Detector in the spaces provided in Appendix B, Spare Parts.

2.3 AC Power Connection


Caution: To avoid a potential fire hazard and damage to the 432 Detector, make sure that the voltage selector in the power connector is set correctly to match the available AC power source, and that the correct fuses are installed before you apply AC power.

Power Cord
The power connector is located on the lower-right corner of the rear panel, as shown in Figure 2-1. If a power plug other than the one supplied is needed for your location, consult Table 2-1 and observe the existing applicable regulations.

AC Power Connection

32

IEEE DIP Switch Cover

Figure 2-1 Rear Panel

Table 2-1 Power Cord Wire Identification Wire (USA) Black White Green Wire (International) Brown Blue Green/Yellow Connection Hot Neutral Ground (Earth)

The 432 Detector can be adapted to operate within two voltage ranges at 50 or 60 Hz. Table 2-2 describes these voltage ranges and the fuse value that is appropriate to each.

Installing the Detector

33

Table 2-2 Nominal Operating Voltage Nominal Voltage (VAC) 100/120 220/240 Fuse T 2A T 1A

Required Material
You need a flat-blade screwdriver to perform this procedure.

Procedure
Caution: To avoid the possibility of electrical shock, turn off the front panel power switch and unplug the power cord.

To change the operating voltage setting: 1. Remove the power cord from its connector on the rear panel of the controller and pry open the power connector cover with a flat-blade screwdriver. 2. Remove the voltage selection barrel and locate the correct voltage setting (Figure 2-2). 3. Reinstall the voltage selection barrel so the desired voltage setting appears through the window when you close the power connector cover (Figure 2-2).
Voltage Settings

Figure 2-2 Changing the Voltage Setting 4. Determine if you need to change the fuses (see Table 2-2). All units are supplied with two 2-A fuses installed for 100/120 volt operation. If you operate the unit on 220/240 volt power, change the fuse as outlined in Section 5.1.1, Replacing Fuses. 5. Reinstall the power connector cover and the power cord.

AC Power Connection

34

2.4 I/O Signal Connections


The 432 Detector is usually installed as an integral part of a data collection system. You can control the 432 Detector either locally from the keypad on the front panel or remotely from a PowerLine controller, such as the Waters 600S. This section describes the detectors I/O signals and how they connect to the following devices: PowerLine controller Empower or Millennium Data module SAT/IN module Chart recorder Device signalling the Chart Marker input Device signalling the Auto Zero input Attention: To meet the regulatory requirements of immunity from external electrical
32

software

STOP disturbances that may affect the performance of this instrument, do not use cables longer
than 9.8 feet (3 meters) when connecting to the screw-type barrier terminal strips. In addition, ensure you always connect the shield of the cable to chassis ground at one instrument only.

2.4.1 I/O Signal Descriptions


The 432 Detector rear panel has an IEEE-488 connector for communication with a PowerLine controller, and a terminal strip (Figure 2-3) for the input/output signals. These signals are described in Table 2-3.

Installing the Detector

35

+ +


+ +

INT Int

REC Rec

LEAK Leak

+ +

Marker In IN
MARKER Out Marker OUT

MARKER

+ +

AUTO ZERO Auto Zero

Figure 2-3 I/O Terminal Strip

Table 2-3 I/O Signal Descriptions Terminal Pairs Rec (+ and ) Function Recorder output A 10-mV full-scale analog output signal appears on these terminals. The measurement range is determined by the product of the Base Range and Sensitivity settings: for example, 500 S (base range) x 0.005 (sensitivity) = 2.5 S full scale. Integrator output A 1-V full-scale analog output signal appears on these terminals. The measurement range is selectable: 10, 50, or 100 S full scale. Marker output A 1-second contact closure signal appears on these terminals when either of the following events occurs: The Chart Mark key on the keypad is pressed A contact closure signal occurs between the Marker In terminals Leak Alert output A contact closure signal appears on these terminals if a leak is detected inside the detector. Auto Zero input The voltage at the Recorder and Integrator output terminals is set to the user-selected balance offset level when a contact closure occurs between these terminals. Marker input A chart mark (~0.5 mV for 3 seconds) is added to the Recorder output signal when a contact closure signal appears between these terminals.

Int (+ and )

Marker Out

Leak Auto Zero (+ and ) Marker In (+ and )

I/O Signal Connections

36

Required Material
To connect cables to the I/O terminals, use a small flat-blade screwdriver.

Other Rear Panel Connections and DIP Switch


In addition to the I/O terminal strip, the rear panel also contains the following items: IEEE-488 connector Communication bus for use with a Waters PowerLine system controller, such as the Waters 600S. DIP switch Sets the IEEE-488 address seen by the system controller. Ground lugs Used to connect the 432 Detector to an earth ground connection and also used as a chassis ground connection to other system instruments.

2.4.2 PowerLine Controller Connections


The 432 Detector can be programmed remotely by a PowerLine controller (such as the Waters 600S) via the IEEE-488 data communications bus.

Required Material
You need a 2.5-mm Allen wrench to connect to the 432 Detector.

Procedure
To connect the 432 Detector to a PowerLine controller: 1. Turn off the PowerLine controller and the 432 Detector. 2. Plug one end of the IEEE-488 cable (included in the Startup Kit) into the bus connector on the rear panel of the 432 Detector (Figure 2-1) and the other end of the cable into the bus connector on the PowerLine controller. 3. Remove the DIP switch cover (Figure 2-1) using a 2.5-mm Allen wrench. 4. Refer to Table 2-4 to set the DIP switches on the rear panel of the 432 Detector (Figure 2-4) to a unique IEEE-488 address between 2 and 29. 5. After you set the DIP switches, reinstall the DIP switch cover. Note: To operate the 432 Detector in local mode, press the front panel Remote key. The illuminated light above the key will go out.

Installing the Detector

37

4 1 2
1 2 3 4 5

16

Switch 1
OFF

Switch 5 (Address 8 Shown)

Figure 2-4 IEEE-488 Address Switch The IEEE-488 address DIP switch employs positive logic to determine the address of the 432 Detector from the switch settings. Table 2-4 shows the settings for valid addresses. Table 2-4 IEEE-488 DIP Switch Setting IEEE-488 Address 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 DIP Switch Settings 1 OFF ON OFF ON OFF ON OFF ON OFF ON OFF ON OFF ON OFF ON OFF ON 2 ON ON OFF OFF ON ON OFF OFF ON ON OFF OFF ON ON OFF OFF ON ON 3 OFF OFF ON ON ON ON OFF OFF OFF OFF ON ON ON ON OFF OFF OFF OFF 4 OFF OFF OFF OFF OFF OFF ON ON ON ON ON ON ON ON OFF OFF OFF OFF 5 OFF OFF OFF OFF OFF OFF OFF OFF OFF OFF OFF OFF OFF OFF ON ON ON ON

I/O Signal Connections

38

Table 2-4 IEEE-488 DIP Switch Setting (Continued) IEEE-488 Address 20 21 22 23 24 25 26 27 28 29 DIP Switch Settings 1 OFF ON OFF ON OFF ON OFF ON OFF ON 2 OFF OFF ON ON OFF OFF ON ON OFF OFF 3 ON ON ON ON OFF OFF OFF OFF ON ON 4 OFF OFF OFF OFF ON ON ON ON ON ON 5 ON ON ON ON ON ON ON ON ON ON

PowerLine Operation
Under PowerLine control, the 432 Detector is recognized as a 431 Detector and it retains the functionality of the 431 Detector with the following differences: The Balance field on the detector setup page of the PowerLine controller affects the Integrator Balance and the Integrator Output only. When you press the Setup key on the controller, the selected Balance value is sent to the 432 Detector from the PowerLine controller. However, the 432 Detector output does not change to the selected balance until the detector is autozeroed by a contact closure at the Auto Zero input terminals on the rear panel (remote or local mode) or when you press the Auto Zero key on the front panel (local mode only). Under PowerLine control, the 432 Detector retains the full functionality of local mode operation, except for the following differences: The Recorder Sensitivity ranges of 0.0002 and 0.0001 are not accessible. The Integrator Sensitivity ranges are not accessible. The 432 Detector does not automatically perform an Auto Zero after an Auto Base routine has occurred.

2.4.3 Empower and Millennium32 Connections


Empower and Millennium software perform data acquisition, processing, and management of chromatographic information. This software requires the detectors analog signal to be converted to a digital form.
32

Installing the Detector

39

Empower and Millennium are menu-driven applications specifically designed by Waters for chromatographers. Use the software to: Acquire data Process data Generate and print reports Store information (or data) in a central area and share this information with users who have proper security access To connect the 432 Detector to an Empower or Millennium
32

32

computer, be sure to:

Connect the Bus Satellite Interface (SAT/IN) module to the Bus Laboratory Acquisition and Control/Environment (LAC/E) card in the Empower computer, Millennium32 computer, acquisition client, or LAC/E32. Connect the 432 Detector to the Bus SAT/IN module (Channel 1 or 2). Remove the IEEE-488 cable from the rear panel of the 432 Detector, if it is connected. The 432 Detector is in local mode when it is connected to an Empower and Millennium computer.
32

Bus SAT/IN Module


The Waters Bus SAT/IN module, shown in Figure 2-5, translates analog signals into digital form. It then transmits these digital signals to the Bus LAC/E card inside the workstation, 32 acquisition client, or LAC/E .

Waters SAT/IN Module


IN

EVENTS CH1 CH2

CHANNEL 1 CHANNEL 2

OUT

IN

OUT

CH 1

CH 2

OK

1 2 3 4

5 6 7 8

Figure 2-5 Bus SAT/IN Module (Front Panel)

I/O Signal Connections

40

Note: To prevent damage to the unit, always disconnect the power cord at either the wall outlet or the power supply before you attach or remove the power connection to the Bus SAT/IN module. The Bus SAT/IN module does not have a power switch.

Connecting the Bus SAT/IN Module to the Bus LAC/E Card


The Bus SAT/IN module connects to the Bus LAC/E through an I/O distribution box, as shown in Figure 2-6. To connect the Bus SAT/IN module to the Bus LAC/E card: 1. Use the I/O distribution cable to connect the I/O distribution box to the 9-pin I/O 32 distribution port on the Bus LAC/E card at the back of the Millennium computer. 2. Use a serial cable to connect the data terminal on the back of the Bus SAT/IN to a port of the I/O distribution box. 3. Configure the serial port for the Bus SAT/IN module as described in the Empower or Millennium32 installation and configuration guides.

Installing the Detector

41

I/O Distribution Por t (9-pin) of Bus LAC/E Card

I/O Distribution Cable

Modified Modular Jack Connections

I/O Distribution Box Serial Cable

SAT/IN Module Rear Panel

Connect SAT/IN to Port 1 on the I/O Distribution Box

BCD DATA PWR

AC to DC Converter

Figure 2-6 Bus SAT/IN to Bus LAC/E Connections

I/O Signal Connections

42

Connecting the Bus SAT/IN Module to the 432 Detector


The Bus SAT/IN module connects to the 432 Detector as shown in Figure 2-7. Refer to the procedure following the figure and Table 2-5 for complete details. Attention: To prevent damage to the unit, do not plug in the power cord of the Bus

STOP SAT/IN module until you perform all of the procedures described in the Waters Bus
SAT/IN Module Installation Guide.

Waters 432 Detector

White Red Black Black

+ +

INT

REC

LEAK

+
Waters SA T/IN Module
IN

MARKER IN

EVENTS CH1 CH2

CHANNEL 1 CHANNEL 2

OUT

IN

OU T

CH 1

CH 2

MARKER OUT
OK

1 2 3

4 5 6 7 8

AUTO ZERO

TP01264

Figure 2-7 Bus SAT/IN to 432 Detector Connections To connect the 432 Detector to the Bus SAT/IN module: 1. Connect the white wire of the analog cable (included with the Bus SAT/IN module) to the Int + terminal on the rear panel of the 432 Detector. Connect the black wire to the Int terminal. 2. Connect the other end of the cable to either the Channel 1 or Channel 2 connector on the front panel of the Bus SAT/IN module.

Installing the Detector

43

3. Connect the Event In terminals of the channel you chose in the previous step to the Inject Start output signal of the Waters Alliance solvent delivery system or the Waters 717plus (or equivalent) Autosampler. 4. Remove the IEEE-488 cable from the rear panel of the 432 Detector, if it is connected. The connections from the 432 Detector to the Bus SAT/IN are summarized in Table 2-5. Table 2-5 Bus SAT/IN Cable Connections 432 Detector I/O Connector Terminal Int (+) Int () Bus SAT/IN Cable White wire Black wire Channel 1 or 2 Bus SAT/IN Connector

2.4.4 Data Module Connections


This section describes how to connect the analog output signal from the 432 Detector to the Waters 746 Data Module. Attention: Remember to meet the regulatory requirements of immunity from external

STOP electrical disturbances that may affect the performance of this instrument, do not use
cables longer than 9.8 feet (3 meters) when connecting to the screw-type barrier terminal strips. In addition, ensure you always connect the shield of the cable to chassis ground.

Analog Signal
To send the analog output signal from the 432 Detector to a Waters data module, connect the signal cable in the 432 Detector Startup Kit as described in Table 2-6. Table 2-6 Data Module Signal Cable Connections Wire Red Black Shield 432 Detector I/O Connector Terminal Int (+) Int () Ground lug 746 Terminal (+) () None

Marker Out Signal


The Marker Out terminals of the 432 Detector provide a contact closure output signal when either of the following events occurs: Chart Mark key is pressed Marker In terminals are shorted together I/O Signal Connections 44

Use the signal to start a Waters 746 Data Module by connecting a signal cable to the modules data cable (Table 2-7). Table 2-7 Data Module Chart Mark Cable Connections 432 Detector I/O Connector Terminal Marker Out Marker Out 746 Cable Join to both Remote Start wires (white and red) Green wire

Wire Either wire Other wire

2.4.5 Chart Recorder Connections


To connect the 432 Detector to a chart recorder: 1. Attach the Recorder cable (see Appendix B, Spare Parts) to the 432 Detector REC output terminals, as indicated in Table 2-8. 2. Connect the cable shield to the ground lug on the 432 Detector rear panel. 3. Connect the other end of the cable to the 10-mV input terminals on the chart recorder, as indicated in Table 2-8. Table 2-8 Chart Recorder Cable Connections Wire Red Black 432 Detector I/O Connector Terminal Rec (+) Rec () Chart Recorder Terminal Pen (+) Pen ()

2.4.6 Chart Marker Input Connections


The 432 Detector accepts a chart mark (start inject) signal from the following devices: Waters 717plus Autosampler Any other device that provides a compatible switch closure

Waters 717plus Autosampler


To connect the 432 Detector to a Waters 717/717plus Autosampler, connect a signal cable as indicated in Table 2-9.

Installing the Detector

45

Table 2-9 Autosampler Chart Mark Cable Connections 432 Detector I/O Connector Terminal Marker In (+) Marker In () Autosampler Terminal Either Inject Start terminal of a pair Other Inject Start terminal of the same pair

2.4.7 Auto Zero Input Connections


The voltage at the Recorder and Integrator outputs is set to the user-selected balance offset level when a contact closure occurs between the Auto Zero terminals. This section describes how to connect the 432 Detector to the following devices (so that an auto zero occurs at the injection point): Waters 717plus Autosampler Any other device that provides a compatible switch closure

Waters 717plus Autosampler


To connect the 432 Detector to a Waters 717plus Autosampler, connect a signal cable as indicated in Table 2-10. Table 2-10 Autosampler Auto Zero Cable Connections 432 Detector I/O Connector Terminal Auto Zero (+) Auto Zero () Autosampler Terminal Either Inject Start terminal of a pair Other Inject Start terminal of the same pair

2.4.8 Alliance Separations Module Connections


Connect the detector to Waters Alliance Separations Modules, when it is not under the 32 control of the Millennium software, to perform the following tasks: Auto-Zero on inject Chart mark on inject Method start

Generating Auto-Zero on Inject


To generate the Auto-Zero function on the 432 Detector at the start of an injection, make the connections summarized in Table 2-11 and illustrated in Figure 2-8.

I/O Signal Connections

46

Table 2-11 Connections for Generating Auto-Zero on Inject Alliance Separations Modules (B Inputs and Outputs) Pin 1 Inject Start Pin 2 Inject Start 432 Detector (A Inputs) Auto-Zero (+) Auto-Zero ()

Before you can generate an Auto-Zero from an Alliance Separations Module, you must configure the Auto-Zero signal at the 432 Detector front panel. The default Auto-Zero signal is Low.
Waters Alliance B (Inputs and Outputs) Inject Start + Inject Start Ground Stop Flow + Stop Flow Waters 432 Detector A (Inputs)

Hold Inject 1+ Hold Inject 1 Hold Inject 2 + Hold Inject 2 Ground Chart Out + Chart Out

INT Int 1 Inject Start + 2 Inject Start + + REC Rec 3 Ground 4 Lamp On/Off + 5 Lamp On/Off Leak 6 Chart Mark LEAK +

+ +

7 Chart Mark + 8 Ground + MARKER Marker In IN 9 Auto-Zero + 10 Auto-Zero MARKER Out Marker
OUT

+ +

AUTO Zero Auto ZERO

Figure 2-8 Alliance Separations Module Connections to the 432 Detector Auto-Zero on Inject

Generating Chart Mark on Inject


To generate the chart mark function at the start of an injection, make the connections summarized in Table 2-12 and illustrated in Figure 2-9.

Installing the Detector

47

Table 2-12 Connections for Generating Chart Mark on Inject Alliance Separations Modules (B Inputs and Outputs) Pin 1 Inject Start Pin 2 Inject Start 432 Detector (A Inputs) Marker In (+) Marker In ()

Before you can generate a chart mark from an Alliance Separations Module, you must configure the chart mark signal at the front panel. The default chart mark signal is Low.
Waters Alliance B (Inputs and Outputs) Inject Start + Inject Start Ground Stop Flow + Stop Flow Waters 432 Detector A (Inputs)

Hold Inject 1+ Hold Inject 1 Hold Inject 2 + Hold Inject 2 Ground Chart Out + Chart Out

INT Int 1 Inject Start + 2 Inject Start + + Rec 3 Ground REC 4 Lamp On/Off + 5 Lamp On/Off Leak 6 Chart Mark LEAK + 7 Chart Mark + 8 Ground + MARKERIn Marker IN 9 Auto-Zero + 10 Auto-Zero MARKEROut Marker OUT

+ +

+ +

AUTO Zero Auto ZERO

Figure 2-9 Alliance Separations Module Connections to the 432 Detector Chart Mark on Inject

2.5 Making Fluidic Connections


Fluid lines to a column and waste container connect to the front of the 432 Detector, as shown in Figure 2-10. To make these connections: Cut the tubing. Assemble compression fittings and ferrules. Connect the tubing to the detector.

Making Fluidic Connections

48

This section will guide you through each of these procedures. Attention: Conductivity detection is sensitive to flow rate fluctuations. If you use a

STOP non-Waters pump or a Waters pump without the SILK microflow compensation algorithm,
you must install the pulse dampener kit supplied in the Startup Kit for optimum performance. Refer to the installation procedure in this section.

Conductivity

Waters 432

Detector

InIn from Column From Column

IN OU T

Out to Waste inches. Out To Waste (18(18 inches, 0.009-inch ID) 0.009-inch I.D.)

Figure 2-10 Fluid Connections

Cutting Stainless Steel Tubing


You need the following tools to cut stainless steel tubing: A file with cutting edge Two cloth- or plastic-covered pliers To cut the tubing: 1. Measure the length of 1/16-inch OD, 0.009-inch ID, stainless steel tubing you need to make the following connections: Column to the detector inlet Detector outlet to a suitable waste container 2. Use a file with a cutting edge to scribe the circumference of the tubing at the desired length. 3. Grasp the tubing on both sides of the scribe mark with cloth-covered pliers. Gently work the tubing back and forth until it separates. 4. File the ends smooth. Installing the Detector 49

Cutting Polymeric Tubing


Waters chromatography systems are supplied with a tubing cutter (similar to the one in Figure 2-11) to facilitate cutting polymeric tubing. This section presents the recommended procedure for using the tubing cutter. Note: To avoid bandspreading caused by angled cuts, always use a tubing cutter. Angled cuts leave unswept dead volumes at the connection junction due to the poor fit of the tubing against the connector or port. To cut a length of polymeric tubing: 1. Estimate the length of tubing required to connect the components. Allow slack so that the tubing is not pulled tightly around sharp corners. 2. Insert the tubing into the cutter so that the tubing extending from the metal side is the length required. Use the proper hole to have a snug enough fit so that the tubing is not flexed by the blade when you cut it.

Figure 2-11 Cutting Polymeric Tubing 3. Press down on the razor blade to cut the tubing (Figure 2-11). Discard the excess tubing that extends from the clear side of the cutter. 4. Inspect the cut for burrs or scratches and for the perpendicularity of the cut.

Assembling Compression Fittings


To assemble each compression fitting: 1. Slide the compression screw over the tubing end, followed by the ferrule (Figure 2-12). 2. Mount the ferrule with its taper end facing the end of the tubing (Figure 2-12).

Making Fluidic Connections

50

Compression Compression Screw Screw

Ferrule Ferrule

0.009-inch I.D. Tubing (0.23 mm) (0.23 mm)


Figure 2-12 Ferrule and Compression Screw Assembly

0.009-inch ID Tubing

Connecting to the 432 Detector


To make connections at the column outlet and detector inlet, and at the detector outlet: 1. Install a compression screw and then a ferrule on the length of 0.009-inch tubing from the column outlet. Use stainless steel fittings on stainless steel tubing and PEEK fittings on PEEK tubing. Note: If you are using a column with 1/428 end fittings and there is a length of tubing with 1/428 fittings on each end, use the 1/428 to Z-detail adapter (included in the Startup Kit) to connect this tubing to the tubing that leads to the detector inlet. The Waters IC-Pak C column comes supplied with a length of tubing that has a 1/428 fitting on one end (column outlet) and a Waters compression screw and ferrule on the other end (detector inlet). 2. Push the free end of the tubing as far as it will go into the IN fitting on the 432 Detector. While you hold it there, use a 5/16-inch open-end wrench to tighten the compression screw 3/4-turn past finger-tight. Note: The 432 Detector and IC-Pak series of columns have very deep ferrules. 3. Remove the compression screw and tubing from the connection and verify that fluid can flow freely. 4. Reconnect the tubing to the IN fitting, making sure to push the tubing all the way into the fitting. 5. Install a ferrule on an 18-inch length of 0.009-inch tubing and connect it to the OUT connection on the 432 Detector. Use stainless steel fittings on stainless steel tubing and PEEK fittings on PEEK tubing. 6. Place the other end of the tube in a waste container. If you are using any Teflon tubing, attach it after the stainless steel or PEEK tubing.

Installing the Detector

51

Installing the Pulse Dampener


To achieve the best performance from the 432 Detector in a chromatographic system with a non-Waters pump, Breeze software, or Waters HPLC 515 Pump, you must install the pulse dampener kit supplied in the Startup Kit. The pulse dampener is not required if you are using a Waters 2695 Separations Module. To install the pulse dampener between the pump and the injector: 1. Assemble the pulse dampener (Figure 2-13) using the instructions in the pulse dampener kit. 2. Connect the large-ID (0.020-inch) tubing to the pump outlet using a stainless steel compression screw and ferrule. 3. Connect the small-ID (0.009-inch) tubing to the injector inlet using a stainless steel compression screw and ferrule. 4. Disconnect the tubing from the injector inlet. 5. Pump ASTM Type I reagent water at 2 mL/min through the pulse dampener assembly until you see a constant stream exiting from the restrictor assembly outlet line. 6. Reconnect the tubing to the injector inlet.

Making Fluidic Connections

52

0.020- IDI.D. i nch 0.020-inch Tubing Tubi ng

0.00 i IDI.D. 9- nch 0.009-inch Tubi Tubingng

Fro Pu m m From Pump p

Union U ni on

To Inect j To Injector or

Low Pressure Low Pressure Filter A ssembly Filt Assembly er

Restrictor or R est i r ct Assembly A ssembly

Figure 2-13 Pulse Dampener

2.6 Passivating the System


Passivating the system removes potential contamination from the wetted surfaces of all system components. Perform passivation on a new system, and subsequently, whenever you suspect that contamination may have occurred. See Section 5.3, Troubleshooting, for help diagnosing performance problems. Use this procedure for Waters hardware only. For other equipment, check with the manufacturer before you continue with this procedure. Attention: If you are installing the 432 Detector into an existing Waters system, replace

STOP the pump seals before you passivate. Use the new pump seals supplied in the Startup Kit
and refer to the replacement procedure in the pump manual.

Installing the Detector

53

To passivate the system: 1. Replace the column with a union fitting. 2. If the system is not new, flush it thoroughly with ASTM Type I reagent water to remove any residual solvents or salts. 3. Connect the power cord to the 432 Detector and plug the other end into an AC power outlet. Push the 432 Detector power switch to turn on the instrument. Caution: To avoid chemical hazards, always wear safety glasses and gloves when you are using solvents.

4. Prime the pump with 6 N nitric acid (HNO3) and run it at a flow rate of 1.2 mL/min for 20 minutes to passivate all the wetted parts of the detector. Press the Clear key to stop the overrange alarm. 5. Stop the pump. 6. Remove the inlet line from the nitric acid and place it in ASTM Type I reagent water. 7. Flush the system using one of the following methods: Prime and start the pump, then flush it with ASTM Type I reagent water at 1.2 mL/min until you observe a consistent reading of less than 20 S (base range set to 50 S). Flush the system overnight with 100% methanol at a reduced flow rate. By the next morning the system will be passivated and ready for use. Note: If you are using a pump with seal-wash capability, skip step 8. 8. Use a syringe to flush the back of the pump seals and pistons by slowly running about 5 mL of water into the top hole in the baseplate of the pump heads. Place a tissue under the baseplates to absorb the water. 9. Set the pump flow rate to 0.0 mL/min. It is not necessary to turn off the 432 Detector unless it will be idle for an extended period (14 days). For best results, always leave the power on to maintain cell temperature; it takes a minimum of 2 to 3 hours once the detector is turned on to equilibrate the flow cell at the selected operating temperature.

2.7 Verifying the Detector


This procedure is a guideline for verifying that the detector works correctly within its expected operational range. The detector is calibrated before shipping, and recalibration is not normally required.

Verifying the Detector

54

Verify the detector when any of these conditions apply: When you replace the flow cell To verify accuracy When you make adjustments

Calibration Procedure
Note: You need solution of 1 mM potassium chloride (KCl) to calibrate the detector. Note: Waters suggests one of its Technical Service Representatives perform this procedure. 1. Turn on the 432 Detector and set the temperature control to 35 C. Allow 2 to 3 hours for the temperature in the flow cell to equilibrate. 2. Set the base range to 200 S. 3. Set the Filter Time Response to Fast. 4. Pump 1 mM KCl solution through the detector (without a column in place). 5. Verify that the front panel output is 147 S 5 S.

Installing the Detector

55

Chapter 3 Operating the Detector


This chapter contains: A description of front panel controls and displays Procedures for starting up, shutting down, and long-term storage Recommended operating practices

3.1 Controls and Indicators


Figure 3-1 illustrates the controls and indicators on the front panel of the 432 Detector.

Waters 432
Conductivity Detector

3
274
CONDUCT ( S/cm)

500
BASE ( S/cm/FS)

0.0005
SENS

Remote

Temp.

Pol.

Chart Mark

7
Auto Base Auto Zero

8 5 2

9 6 3
Clear

4
Base Range Sens. Range

1 0
Shift

Resp.

Bal.

Enter

ON

IN

OUT
OFF

Figure 3-1 Front Panel

Controls and Indicators

56

Power Switch
The power switch (located in the lower-right corner of the front panel) controls power to the 432 Detector. Upon startup, an initialization routine verifies the data in ROM memory, tests RAM memory function, and checks for any internal leakage or an eluent conductivity over-range condition.

Display
The display shows instrument status and parameter values in two 20-character lines of text. Upon startup, Waters 432 Self Check appears briefly. If any error conditions are detected during startup or normal operation, the appropriate error message is displayed. The main screen shows the measured conductivity, as well as the base range and sensitivity settings. When you set an operating parameter, the display shows the selected or entered value. Error Messages A corresponding error message is displayed if one of the following conditions occurs: ROM/RAM error (checked during startup only) Error: ROM/RAM Leakage detected Error: Leak Temperature control failure Error: Temp Over-range (above base range setting) Error: Over Range Overflow (above 10,000 S) Error: Over Flow Press the Clear key to clear an error alarm and message. For a continuing error condition, the error message remains after the audio alarm is cleared.

Keypad
Use the keypad to control the operation of the 432 Detector. Table 3-1 describes the function of each key.

Operating the Detector

57

Note: Three keys (Balance, Sensitivity Range, and the numeral 1) perform an alternate function when they are preceded by the Shift key. Table 3-1 Key Descriptions Key Remote Function Remote key: Toggles between local and remote operating modes. In remote mode, the light above the key is on and all other front panel controls are disabled. Polarity key: Toggles the polarity of the signal to the external chart recorder and integrator. When positive polarity is selected, the light above the key is illuminated. Base Range key: Sets the base sensitivity range of the 432 Detector to the appropriate value for the eluent being used. The base sensitivity is set to one of ten steps, from 10 S (maximum gain) to 10,000 S, using the Up and Down keys or the numeric keypad. Sensitivity Range key: Sets the sensitivity range multiplier of the 432 Detector. The sensitivity range has twelve steps, from 0.0001 (maximum sensitivity) to 1.0 (available only with 100 S multiplier setting), and is set using the Up and Down keys or the numeric keypad. The 10-mV full-scale recorder response is calculated by multiplying the Base Range by the Sensitivity Range to obtain a value of x S / 10 mV FS. The recorder range is 1 to 0.0001 for the 100 S setting and 0.1 to 0.0001 for the two lower settings. Shift key then Sensitivity Range key: Sets the sensitivity range multiplier of the integrator to 100, 50, or 10 S using the Up and Down keys or the numeric keypad; the integrator output is 100, 50, or 10 S/1 V, respectively. Bal. Balance key: Manually sets the offset (%) of the signal to the external chart recorder. (Use the numeric keypad or Up and Down keys.) Shift key then Balance key: Manually sets the offset (%) of the signal to the integrator. (Use the numeric keypad or Up and Down keys.) Shift key after Balance key: Changes the polarity of the offset. Allowable values are 100 to +100%. Temp. Temperature key: Sets the temperature of the detection cell. Use the Up and Down keys or the numeric keypad to turn temperature control off (Setting 0) or select one of the following eight settings: 30, 35, 40, 45, 50, 55, 60, or 65 C. The light above the key is illuminated when the temperature control is on.

Pol.

Base Range

Sens. Range

Controls and Indicators

58

Table 3-1 Key Descriptions (Continued) Key Resp. Function Response key: Sets the response time constant of the 432 Detector to optimize signal-to-noise ratio. Use the Up and Down keys or the numeric keypad to choose Setting 1 (Fast, 0.25 sec) for very narrow peaks, Setting 2 (Standard, 0.5 sec), or Setting 3 (Slow, 1.0 sec) to detect wider peaks. Setting 2 is used for most applications. Auto Zero key: Automatically zeros the Recorder and Integrator signals to the specified Recorder Balance and Integrator Balance offsets, respectively. Auto Base key: Automatically sets the base range of the 432 Detector to the appropriate value for the eluent being used. This is the next highest setting above the actual background conductivity of the eluent. Shift key: Press the Shift key before, not along with, other keys to access additional functions and also to change polarity when you set balance offset values. When the Shift key is pressed, an asterisk (*) appears at the right side of the display; press Shift again to return to normal mode. Shift key then Balance key: Displays the integrator balance offset value. When setting the balance offset, press Shift to change polarity. Shift key then Sensitivity Range key: Displays integrator range value. Shift key then 1 key: Displays the current, actual value of the chart recorder balance offset. Press Enter to return to the main screen. Chart Mark Chart Mark key: When this key is pressed, a 1-second, 1-mV signal is sent to the Recorder terminals and a 1-second contact closure is sent to the Marker Out terminals. Enter key: When you manually set offsets, sensitivity range, or base range, pressing Enter records the displayed value and returns the display to the main screen. The Enter key is also used to access the beep setting function. Clear key: Erases a value input from the keypad. The Clear key is also used to clear an error alarm and message. For a continuing error condition, the error message remains after the audio alarm is cleared.

Auto Zero Auto Base

Shift

Enter

Clear

Operating the Detector

59

Table 3-1 Key Descriptions (Continued) Key Function Up key: Increments the current parameter setting.

Down key: Decrements the current parameter setting.

Beep Function
You can set the 432 Detector to beep when a key is pressed and/or an error condition is detected. Use the Clear key to stop an error alarm. For a continuing error condition, the error message remains after the beep is cleared. To set the beep function, follow the procedure in Table 3-2. Table 3-2 Setting the Beep Function Keystroke 1 2 3 4 5 6 Enter . (decimal point) Enter 1 0 1 0 Enter Turns on the beep-on-error function. Turns off the beep-on-error function. Turns on the beep-on-keystroke function. Turns off the beep-on-keystroke function. Saves settings and returns to the main screen. Key Function Accesses the beep function parameters.

Note: Since the void volume in many separations contains highly conductive counter-ions, the error alarm sounds upon each injection. See Table 3-2 to silence the beep-on-error alarm.

Controls and Indicators

60

3.2 Startup and Shutdown


Startup Procedure
Perform the following procedure to start the 432 Detector. Typically, this procedure is done at the beginning of each workday. Note: This procedure assumes that the flow cell has stabilized at the selected temperature (minimum 2 to 3 hours). Standard practice is to leave the detector powered and with the temperature control on unless the instrument will be unused for several days. Set the temperature at least 5 C above the highest ambient temperature expected for the duration of the application. 1. Prime the pump with properly degassed eluent and set the flow rate to 1.2 mL/min or to the flow rate recommended for your particular column or application. Do not sparge eluents, since sparge gasses may contain ionic contaminants. 2. Set the response (time constant) to the desired setting by pressing the Response key. A standard setting (0.5 seconds) is preferred for most applications. 3. Set the base value by pressing the Auto Base key or by manually entering the base range that is the next highest setting above the eluents background conductivity. 4. Turn on the recorder/integrator and run the system until the baseline stabilizes. 5. Depending on whether you are using a recorder or an integrator, do one of the following actions: If you are using a 10-mV recorder connected to the Recorder terminals on the rear panel, select the desired sensitivity by pressing the Sensitivity Range key, then the appropriate Up or Down arrow key. If you are using an integrator connected to the Integrator terminals on the rear panel, select the desired sensitivity by pressing the Shift and Sensitivity Range keys, then the appropriate Up or Down arrow key. 6. Zero the recorder/integrator by pressing the Auto Zero key or have the Auto Zero terminals of the rear panel I/O terminal strip connected to your manual injector or autosampler. The 432 Detector is now ready for operation.

Standby Setup
To eliminate the need to allow time for the flow cell temperature to equilibrate, leave the 432 Detector turned on at the end of the workday or workweek. Set the temperature control to the operating temperature and the pump flow rate to 0.01 to 0.1 mL/min (depending on the pump).

Operating the Detector

61

Long-Term Storage
If the 432 Detector is to be removed from a system for storage or if the system itself is to be stored for a long time, flush the detector/system with 100% water, then 100% HPLC-grade methanol. Leave the methanol in the system after shutdown. If you are removing the 432 Detector from the system, seal the inlet and outlet bulkheads with dead-end fittings or a loop of tubing.

3.3 Operating Recommendations


Observe the following recommendations for best detector performance.

Temperature Equilibration
The 432 Detector should be powered up and set at the desired operating temperature for two to three hours before use. Select a temperature at least 5 C above the highest ambient temperature expected during the duration of the application. The detector is usually set at 35 C, but it should be set higher if the ambient temperature will exceed 30 C. You may choose to leave the 432 Detector powered up overnight at a flow rate of 0.01 0.1 mL/min (depending on the pump) to minimize the daily reequilibration time. A drifting baseline is one indication that the temperature of the flow cell is not uniform across the flow cell or stable over time.

Base Range
The Base Range is normally set at the next setting above the background conductivity of the eluent. For example, if the conductivity of borate/gluconate eluent is 270 S, set the Base Range to 500 S.

Integrator Output
The Integrator output is not attenuated; signals should be below 1 V. Set the Integrator output to 10 S/V for small signals or to 50 S/V when you expect a signal greater than 10 S. If you are using the 432 Detector with chemical suppression, set the integrator output to 100 S/V.

Recorder Output
The Recorder output is attenuated and the Sensitivity Range should be adjusted to provide the appropriate output level.

Operating Recommendations

62

Polarity
Signal polarity depends on eluent conductivity. If necessary, press the Polarity key to obtain peaks rather than dips.

Eluent Handling
Replace your eluent reservoir filter regularly. When you analyze cations, use an all-plastic eluent reservoir filter. Filter and degas eluents to prolong column life, reduce pressure fluctuations, and decrease baseline noise. When you change eluents, flush the flow cell and associated tubing thoroughly with the new eluent.

Operating the Detector

63

Chapter 4 Performing Ion Analysis


This chapter provides essential information about ion analysis techniques. Two representative columns serve as typical examples: the Waters IC-Pak A for anions and the IC Pak C M/D for cations. The following topics are discussed: Fundamental considerations Configuring the system Selecting and preparing eluents Preparing and injecting standards Also refer to the manufacturers manual for the particular column you are using. A recommended source for more information about ion analysis in general is Ion Chromatography: Principles and Applications by Paul R. Haddad and Peter E. Jackson, Elsevier Science Publishing, New York, 1990.

4.1 Fundamental Considerations


To obtain full performance from the 432 Detector in chromatographic applications, observe the guidelines presented in this section regarding: Water Containers High-pH eluents Sample preparation

Water
Water purity (the absence of conductivity) is the most important consideration in ion analysis. Note: Using water other than ASTM Type I reagent water will compromise the accuracy of your results. Freshly drawn ASTM Type I reagent water, conforming to ASTM specification D1193, with total organic carbon <100 ppb is recommended. Sophisticated laboratory water systems that use a combination of reverse osmosis, mixed bed ion exchange, and carbon

Fundamental Considerations

64

adsorption cartridges produce ASTM Type I reagent water, and are recommended for ion chromatography applications. Do not use HPLC-grade water or distilled water. Attention: To avoid damage to the detector flow cell, do not allow the flow cell to

STOP dry out.

Containers
Use plastic to contain all anion and cation samples, cation standards, and cation eluents. When you analyze trace level ions in water, polystyrene containers such as tissue culture flasks are recommended; polypropylene or polymethylpentene containers suit most other applications. Use polystyrene tissue culture flasks for long-term storage. If your system operates on Breeze software or contains a 2695 Separations Module, use 4-mL polycarbonate vials to hold your samples and standards. Attention: Avoid glass containers (which tend to leach sodium cations) when you are

STOP analyzing for cations.


Preparing Containers for Low-Level Analysis To prepare plastic containers for low-level analysis: 1. Soak all containers for 5 hours with a 1:1 solution of nitric acid (HNO3) and ASTM Type I reagent water. 2. Rinse with plenty of ASTM Type I reagent water. The containers are ready for analysis in the ppm range. 3. For analysis in the ppb range, fill each container completely with ASTM Type I reagent water and let soak overnight. Certain applications that involve ppb level analysis may require container considerations beyond the scope of this manual. For further instructions on trace metal cleaning of plasticware, see Selection and Cleaning of Plastic Containers for Storage of Trace Element Samples, JR Moody and RM Lindstrom, Analytical Chemistry, v. 49, Dec 1977, pp. 2264-67, or contact the Waters Technical Services Department. Cleaning Syringes To avoid contamination, always rinse a syringe two to three times with ASTM Type I reagent water before you draw standards or samples for injection.

High-pH Eluents
High-pH eluents (such as hydroxide eluent) absorb atmospheric CO2, which slowly acidifies the eluent causing baseline drift and retention time changes. To minimize

Performing Ion Analysis

65

carbonate absorption, connect a soda lime (Ascarite ) tube (Figure 4-1) to the eluent bottle as follows: 1. Insert a 3/4-inch (2-cm) piece of glass wool in one end of a polyethylene tube with end fittings. Attach the end fitting. Caution: To avoid chemical burns, wear gloves, lab coat, and eye glasses when you are handling soda lime.

2. Fill the tube with soda lime (Ascarite) until it reaches 3/4 inches (2 cm) from the top. 3. Add another piece of glass wool to the other end of the tube and attach the end fitting. 4. Drill a hole in the cap of the reagent bottle. The hole should be large enough to accommodate the end fitting. Drill a second hole for the pump inlet line. 5. Pass the pump inlet line through the hole. Seal the hole with paraffin film. 6. Change the soda lime in the tube when it is exhausted.

Fundamental Considerations

66

End Fittings

Glass Wool

Soda Lime

Polypropylene Tube

Pump Inlet Line

4
Reagent Bottle

Figure 4-1 Soda Lime Tube

Sample Preparation
Sample preparation is very important in ion analysis. Contact the Waters Technical Services Department, if you need assistance.

Performing Ion Analysis

67

As a general rule, to analyze a sample of completely unknown ionic concentration, initially prepare at least a 1:100 dilution and inject 100 L. For best results, injections should contain a total anion concentration of no more than 300 ppm for the IC-Pak A column or a total cation concentration of no more than 10 ppm per ion for the IC-Pak C M/D column. The sample volume (usually 100 L) generally equilibrates to the pH of the eluent upon injection. However, for samples with pH values that differ greatly from that of the eluent (for example, strong acids and bases), bring the sample pH close to that of the eluent before you inject the sample, if possible. Do not inject concentrated samples directly into the mobile phase. Direct injection may cause precipitation of the salts in the sample. Dissolve (or dilute) samples in an appropriate volume of the mobile phase first. If you must use other solvents, watch for precipitation upon injection into the eluent. Always filter samples before you use them. Cationic samples that contain organic amines may exhibit hydrophobic interaction between the mobile phase and packing. You may use a water-miscible organic mobile phase, such as acetonitrile, as a modifier to reduce this. Pretreat the sample with a Sep-Pak C18 cartridge to remove hydrophobic compounds.

4.2 Configuring the System


Figure 4-2 shows a typical system configuration. Refer to Section 2.5, Making Fluidic Connections, for the procedures to cut tubing and assemble fittings.
Waters In-Line Degasser * Guard Column Holder* Pump Injector Eluent Reservoir Pulse Dampener Column Waters 432 Detector

4
To Waste

*Optional t non-Waters pumps or SILK or non-W with pumps such Required for Waters pumps without Waters pumpsaters Breeze software,TP01269 as the HPLC 515

Figure 4-2 System Configuration for Ion Analysis

Configuring the System

68

Pulse Dampener
If your system uses a non-Waters pump or a Waters pump with Breeze software, such as the HPLC 515, use a pulse dampener (supplied in the Startup Kit) to achieve the best performance from the 432 Detector. Install the pulse dampener between the pump and the injector, as described in Installing the Pulse Dampener on page 52.

4.3 Eluents for Ion Analysis


This section describes how to select, prepare, and use eluents for ion analysis. Attention: Never recirculate eluents. Ions from sample and standard injections

STOP progressively contaminate a recirculating eluent.

General Guidelines
Observe the following guidelines when you prepare eluents for ion analysis: Use only ASTM Type I reagent water with total organic carbon <100 ppb. Use the highest purity salts and reagents available. A pH meter is recommended for checking the pH of eluents; care should be taken to avoid cross contamination. Adjust the pH with potassium hydroxide (KOH) or lithium hydroxide (LiOH). For eluents such as octane sulfonate, test an aliquot of the eluent with pH paper. Never immerse pH paper directly into a batch of eluent. Use the following formula to prepare eluents: Formula Wt of Salt x Molarity = g/L Salt

Eluent Filtering and Degassing


The Waters Solvent Clarification Kit is recommended for eluent filtration and preliminary degassing. Durapore 0.22-m filters can be used for all ion chromatography eluents. Millipore 0.45-m HATF filters may be used for aqueous eluents containing no organic modifier. For eluents containing organic modifier, use the Durapore filters. Attention: To avoid contamination when you analyze for cations, minimize the time that

STOP the eluent is in contact with the glass filtration apparatus and transfer the eluent to a
suitable pre-cleaned plastic container as soon as possible. After you install a new filter, pass 20 to 30 mL of eluent through the filter under vacuum. Turn off the vacuum, swirl the eluent throughout the flask and discard. Reattach the flask to the filter apparatus and filter the remaining eluent. As soon as filtration is complete,

Performing Ion Analysis

69

transfer the eluent to a precleaned plastic container, introducing the least possible amount of bubbles in the process. The Waters In-line Degasser is recommended for continuous online degassing.

4.3.1 Preparing Anion Eluent


This section presents the procedure for the preparation of sodium borate/gluconate concentrate and eluent. Consult the manufacturers manual for your column (IC-Pak Column and Guard Column Care and Use Manual included with Waters columns) for additional instructions on the selection and preparation of eluents. A recommended source for more information about ion analysis in general is Ion Chromatography: Principles and Applications, by Paul R. Haddad and Peter E. Jackson, Elsevier Science Publishing, New York, 1990.

Preparing Lithium Borate/Gluconate Concentrate


To prepare sodium borate/gluconate concentrate, refer to Section C.1.13, Preparing Lithium Borate/Gluconate 50X Stock Concentrate and Section C.1.14, Preparing Lithium Borate/Gluconate Eluent.

4.3.2 Preparing Cation Eluent


To prepare 1 L of cation eluent, refer to Section C.2.1, Preparing Eluent.

4.4 Standards for Ion Analysis


This section describes how to prepare and inject ion standards. Note: It is recommended to purchase certified 1000-ppm anion standards instead of preparing them manually. Certify all manual standards against National Institute of Science and Technology traceable standards. Standard concentrations in this manual are defined in terms of mass. For example, 1 mg of sample per liter of water equals a 1 ppm concentration, since 1 L of water has a nominal mass of 1 kg (0.997 kg at 25 C). 1 part per thousand = 1 mg/mL = 1 g/L = 1000 ppm 1 part per million (ppm) = 1 g/mL = 1 mg/L = 1000 ppb 1 part per billion (ppb) = 1 ng/mL = 1 g/L = 1000 ppt 1 part per trillion (ppt) = 1 pg/mL = 1 ng/L

Standards for Ion Analysis

70

Storing Standards
For accurate quantitative results, do not store standards beyond the approximate periods listed in Table 4-1. Be aware that shelf-life depends on many factors and may be significantly shorter than shown here. Table 4-1 Shelf-Life of Standards Standard Carbonate, ppm Chloride, ppm All, ppb All other anion concentrates Cation standards Cation concentrates 1 day 3 weeks 1 day 1 to 2 months 1 month 6 months Shelf-Life

Nitrite and carbonate concentrates 1 week

Cation standards must be stored in properly prepared plasticware. See Containers on page 65.

4.4.1 Preparing Anion Standards


This section presents the procedure for preparing a 7-anion standard. If a simpler standard suffices, follow the procedure, but select only three or four salts, such as sodium chloride, sodium nitrate, and sodium sulfate. Always use salts of at least reagent-grade purity. If you require quantitative results or you use hygroscopic salts, dry the salts overnight at 80 C before you make solutions. Store the dried salts in a desiccator.

Preparing a 7-Anion Standard


To prepare a 7-anion standard: 1. Weigh out the amounts of dry salts shown in Table 4-2 or use the following formula to calculate the amount for a salt not listed: (Mol. Wt. Salt / Mol. Wt. Cation) x 0.1 = g Salt

Performing Ion Analysis

71

Table 4-2 Salts for Anion Standard Concentrates Salt (Anion) Sodium fluoride (F ) Sodium chloride (Cl ) Sodium nitrite (NO2 ) Potassium bromide (Br ) Sodium nitrate (NO 3 ) Potassium phosphate, monobasic (HPO4 2) Sodium sulfate (SO42 )

Weight (mg) 221.0 164.9 150.0 148.9 137.1 141.8 147.9

2. Place each salt in a separate plastic100-mL volumetric flask and dilute to the mark with ASTM Type I reagent water. Each concentrate contains 1000 ppm of the anion. 3. Combine the amounts listed in Table 4-3 in a 100-mL volumetric flask with ASTM Type I reagent water. Table 4-3 Anion Concentrate Dilutions Anion Fluoride Chloride Nitrite Bromide Nitrate Phosphate Sulfate Amount (L) 100 L 200 L 400 L 400 L 400 L 600 L 400 L Final Concentration (ppm) 1 ppm 2 ppm 4 ppm 4 ppm 4 ppm 6 ppm 4 ppm

4.4.2 Injecting Anion Standards


Required Materials
To inject a standard, obtain the following materials: Borate/gluconate eluent Refer to Section C.1.13, Preparing Lithium Borate/Gluconate 50X Stock Concentrate and Section C.1.14, Preparing Lithium Borate/Gluconate Eluent.

Standards for Ion Analysis

72

1-cc disposable plastic syringe Pharmaseal Stylex disposable syringe with a Luer Slip tip, or equivalent. Autoinjector or manual injector with 100-L loop Ion chromatography commonly uses a 100-L injection volume. When you use a fixed loop, overfill a minimum of three times.

Injecting the Standard


To inject the standard: 1. Set up the 432 Detector as follows: Base Sensitivity = 500 S Integrator Sensitivity = 10 S/V Recorder Sensitivity = 0.01 (strip chart) Response = STD (0.5 seconds) Temperature = 35 C Polarity = + 2. Equilibrate the 432 Detector as described in Startup Procedure on page 61. 3. Rinse a 1-cc disposable plastic syringe two or three times with ASTM Type I reagent water, then load the standard. 4. Place the syringe tip into the sample loading port and overfill the 100 L loop at least three times (that is, 300 L). 5. Inject the sample. Figure 4-3 shows a representative chromatogram of the 7-anion standard run on an IC-Pak A (4.6 mm x 5.0 cm) column with borate/gluconate eluent at 1.2 mL/min flow rate. The separation of the standard typically takes 12 to 15 minutes with this setup.

Performing Ion Analysis

73

Figure 4-3 Chromatogram of a 7-Anion Standard

4.4.3 Preparing Cation Standards


This section presents the procedure for preparing an 8-cation standard. If a simpler standard suffices, follow this procedure selecting only those salts that you want in the standard. For accurate quantitative results, use only properly prepared plasticware and do not store standards beyond the recommended shelf-lives listed in Table 4-1.

Preparing Cation Standard Concentrates


Note: It is recommended that you use certified 1000-ppm cation standards not prepared in acid with this method. To prepare concentrated stock solutions for an 8-cation standard (prepare fewer types of cations, if a simpler standard suffices): 1. Weigh out the amounts of dry salts shown in Table 4-4 or use the following formula to calculate the amount for a salt not listed.

Standards for Ion Analysis

74

(Mol. Wt. Salt / Mol. Wt. Cation) = g Salt If you choose to use other salts, be sure to avoid any combinations that will form a precipitate. Table 4-4 Salts for Cation Standard Concentrates Salt (Cation) Lithium hydroxide monohydrate (Li+) Sodium chloride (Na +) Ammonium chloride (NH4+) Potassium chloride (K+) Magnesium nitrate hexahydrate Calcium nitrate tetrahydrate Barium chloride dihydrate Strontium nitrate tetrahydrate (Mg2+) (Ca2+) (Sr2+) Weight (g) 6.0476 2.5421 2.9640 1.9067 10.5466 5.8919 3.2377 1.7786

(Ba2+)

2. Place each salt in a separate plastic1-L volumetric flask and dilute to the mark with reagent-grade water. Each concentrate contains 1000 ppm of the cation.

Preparing an 8-Cation Standard


To prepare 1 liter of 8-cation standard: 1. Add the volume of stock (concentrate) standard listed in Table 4-5 to a plastic 1-L volumetric flask. Table 4-5 Cation Concentrate Dilutions Cation Lithium Sodium Ammonium Potassium Magnesium Calcium Strontium Barium Amount (mL) 0.25 1.00 1.00 3.00 2.00 3.00 5.00 5.00 Final Concentration (ppm) 0.25 1.00 1.00 3.00 2.00 3.00 5.00 5.00

2. Fill the flask to the mark with ASTM Type I reagent water.

Performing Ion Analysis

75

4.4.4 Injecting Cation Standards


Required Materials
To inject the standard, obtain the following materials: 0.1 mM EDTA/ 3 mM HNO3 cation eluent Refer to Section 4.3.2, Preparing Cation Eluent. 1-cc disposable plastic syringe Pharmaseal Stylex disposable syringe with a Luer Slip tip, or equivalent. Injector or autosampler with a 100-L loop Ion chromatography commonly uses a 100-L injection volume. When you use a fixed loop, overfill a minimum of three times.

Injecting the Standard


Note: You can substitute the method described in Section C.2, Alkali and Alkaline Earth Cations, Ammonium, and Amines, for the following procedure. Use this procedure to inject the standard. 1. Set up the 432 Detector as follows: Base Sensitivity = 2000 S Integrator Sensitivity = 50 S/V Recorder Sensitivity = 0.01 (strip chart) Response = STD (0.5 seconds) Temperature = 35 C Polarity = (negative) 2. Equilibrate the 432 Detector as described in Startup Procedure on page 61. 3. Rinse a 1-cc disposable plastic syringe two or three times with ASTM Type I reagent water, then load the standard. 4. Place the syringe tip into the sample loading port and overfill the 100-L loop at least three times (that is, 300 L). 5. Inject the sample. Figure 4-4 shows a representative chromatogram of an 8-cation standard run on an IC-Pak C M/D column with 0.1 mM EDTA/3 mM HNO3 eluent at 1.0 mL/min flow rate. The separation of the standard typically takes 20 to 25 minutes with this setup.

Standards for Ion Analysis

76

Figure 4-4 Chromatogram of an 8-Cation Standard

Performing Ion Analysis

77

Chapter 5 Maintenance
This chapter contains information about maintaining the 432 Detector and troubleshooting charts to help you isolate and correct problems. Caution: To avoid the possibility of electric shock, power off the detector and disconnect the power cord before you service the instrument.

5.1 Routine Maintenance


This section contains information designed to help you maintain the 432 Detector. Routine maintenance for the 432 Detector includes: Replacing the fuse Calibrating the detector Maintaining the flow cell Waters service specialists provide maintenance for the 432 Detector on a corrective, as required, basis. Contact the Waters Technical Services Department if you have questions regarding the repair or performance of your instrument. Follow these suggestions to help you maintain the 432 Detector: Stock the recommended spare parts listed in Appendix B to reduce downtime. Contact the Waters Service Department for assistance. Record the initial installation date and serial number of your instrument in Appendix B for easy reference. Keep a file of typical chromatograms for comparison when you suspect problems.

5.1.1 Replacing the Fuse


To change the operating voltage fuse: 1. Turn off the front panel power switch and remove the power cord from its connector on the rear panel of the detector. Caution: To avoid the possibility of electric shock, turn off the front panel power switch, and unplug the power cord from the rear panel.

Routine Maintenance

78

2. Pry open the power connector cover with a screwdriver. 3. To change the AC power fuses, pull out each fuse holder as though opening a drawer. Spare fuses are included in the System Startup Kit. For ordering information, see Appendix B, Spare Parts. 4. Table 2-2 on page 34 lists the operating voltage fuses (for use in either North America or Europe). 5. Install the correct fuse in the holder and slide it back into place (Figure 5-1). The arrow on each fuse holder points up when in the correct position.

Figure 5-1 Installing Operating Voltage Fuses 6. Close the power connector cover. Then plug the power cord into its connector on the rear panel of the detector.

5.1.2 Maintaining the Flow Cell


Maintenance for the 432 Detector consists of ensuring the flow cell is free of foreign material. Foreign material in the flow cell may cause baseline drift, cycling, or noise. Attention: To avoid damaging the column, remove it before you flush the system. Do not

STOP reconnect the column until the eluent has returned to approximately pH 7.
To clean the cell: 1. Flush the system with ASTM Type I reagent water. 2. Flush the system with 20 mL of 6 N nitric acid (HNO3). 3. Flush the system again with ASTM Type I reagent water. Do not reconnect the column until the eluent has returned to about pH 7.

Maintenance

79

Refer to Appendix B, Spare Parts, to order a replacement flow cell. The following tools are required to replace the flow cell: Phillips-head screwdriver 5/16-inch open-end wrench Knife or flat-blade screwdriver Caution: To avoid electrical hazards, always unplug the power cord before you perform any of the following replacement procedures.

1. Unplug the 432 Detector from the power source, and completely disconnect all electrical cables and fluid connections. 2. Remove the 432 Detector cover (four Phillips-head screws, two on each side). 3. Remove the two pins and pin holders that hold the cell block cover in place (see Figure 5-2). Use a knife or flat-blade screwdriver to gently pry the pins and holders out. 4. Pull off the cover of the flow cell unit and remove the top layer of insulation. 5. Remove the four Phillips-head screws from the upper plate of the cell block, and remove the plate. Note the orientation of the plate: a notch is cut into the underside to clear one of the components within the cell block. 6. Carefully disconnect the inlet and outlet fittings from the flow cell. 7. Remove the two Phillips-head screws from the flow cell mounting bracket. 8. Unplug the flow cell cable connector from its socket in the cell block. 9. Remove the flow cell assembly. 10. Install the new flow cell by following steps 2 through 9 in reverse order. Be sure to orient the upper plate of the cell block properly before you install the four screws.

Routine Maintenance

80

Cell block cover Pin Pin holder

Pin holder Pin Insulation

Upper plate screws

Upper plate of cell block

Cell mounting bracket screws Cell mounting bracket Flow cell Connector

Cell block

Figure 5-2 Flow Cell Assembly

Maintenance

81

5.2 Cleaning the Detector Exterior


To clean the outside of the 432 Detector, use only a soft lint-free paper or cloth dampened with mild soap and water.

5.3 Troubleshooting
This section contains troubleshooting charts to help you isolate and correct problems with the 432 Detector. Keep in mind that the source of apparent detector problems may lie within the chromatography or hardware of your system. The Waters Guide to Successful Operation of Your LC System contains detailed chromatographic troubleshooting tables. (Contact your nearest Waters office for information on how to get a copy.) If you cannot correct a problem, contact the Waters Technical Services Department for assistance.

When You Call Waters Service


To expedite your request for service, have the following information available when you call Waters regarding a 432 Detector problem: Symptom Type of column Eluent(s) Flow rate Operating pressure Base Range setting Integrator Sensitivity setting Recorder Sensitivity setting Type of injector (automatic or manual) Type of data integrator

Detector Does Not Turn On


If your detector is completely inoperative (for example, the lights do not illuminate and the display remains completely blank when the unit is turned on), the fuse may require replacement. Refer to Section 5.1.1, Replacing the Fuse.

Cleaning the Detector Exterior

82

Startup Diagnostics
The 432 Detector performs startup diagnostics that check the internal memory (both RAM and ROM), and the associated processing circuitry.

Power Supply
The following factors can adversely affect the operation of the 432 Detector: Power surges Line spikes Transient energy sources Be sure that the electrical supply used for the 432 Detector is properly grounded and free from any of these conditions.

Error Messages
The error messages displayed by the 432 Detector are listed below along with the recommended action for each: Error: ROM/RAM ROM/RAM error (checked during startup only) Call Waters service. Error: Leak Leakage detected Check flow cell and associated plumbing connections. Error: Temp Temperature control failure Call Waters service. Error: Over Range Base over-range condition Set Base Range to the next setting above the background conductivity of the eluent. Error: Over Flow Conductivity overflow (above 10,000 S) Dilute eluent to remain within measurable range. Press the Clear key to clear an error alarm and message. For a continuing error condition, the error message remains after the audio alarm is cleared.

Troubleshooting Procedure
As soon as you realize that a problem may exist: 1. Visually examine the integrity of the electrical and fluid connections as you verify proper system configuration and installation.

Maintenance

83

2. If the results of previous runs are available, compare the current system operation with the system operation before you identified the problem. For example, if your system usually runs at a certain pressure with a certain method: Is the system pressure in the same range, or is it drastically higher (possibly caused by a blocked line) or lower (possibly caused by a leak)? Are pressure fluctuations in the same range as during normal operation? 3. Isolate the parameter that varies from normal operation. The parameters to observe include: Baseline noise Peak retention time Peak resolution Qualitative/quantitative chromatographic results System pressure Evaluate the parameters in the order presented above to rule out simple causes of the problem. 4. Use Table 5-1 to determine corrective actions for the problems that you identify.

Removing Bubbles
Bubbles in the flow cell are evident when the noise is equal to or greater than 2 S. Use this method to remove bubbles. 1. Disconnect the tubing from the inlet and outlet of the 432 Detector. 2. Attach a 1-mL tuberculin syringe to a priming syringe cannula which is screwed into the inlet of the detector. 3. Flush four times with 1-mL portions of ASTM Type I reagent water. 4. Flush four times with 1-mL portions of HPLC-grade methanol. 5. Flush four times with 1-mL portions of ASTM Type I reagent water. 6. Reattach the tubing from the 432 Detector outlet to a waste receptacle (18-inch length of 0.009-inch ID stainless steel). 7. Start eluent flowing through the system at a flow rate of at least 1 mL/min. 8. With the eluent flowing, reattach the detector inlet tubing to the column. 9. Allow a few minutes for temperature reequilibration, then check the noise level. If it is not reduced, repeat steps 1 through 4, then proceed to steps 10 through 13. 10. Attach a dead-end fitting to the 432 Detector outlet. 11. Remove the priming syringe cannula and attach a dead-end fitting to the 432 Detector inlet.

Troubleshooting

84

12. Allow the detector to stand overnight (>12 hours) with temperature on and with methanol in the flow cell. 13. Repeat steps 5 through 9. Table 5-1 Troubleshooting Guide Symptom Noisy baseline Possible Cause Pulse dampener not installed Pulsing pump Bubbles in flow cell Voltage fluctuation Spikes on baseline Dirty flow cell Flow cell leak Solution See Installing the Pulse Dampener on page 52. Check the pump; see the pump manual. Remove bubbles and degas the solvent. Use the voltage regulator. Clean the cell. Check flow cell fittings and tighten. If leak continues, replace the flow cell. Remove bubbles and degas the solvent. Control ambient temperature, locate drafts, and insulate tubing and column, if necessary. Set the cell temperature to a minimum of 5 C above ambient. Replace the column. Make sure the temperature control is turned on. Call Waters service. Control ambient temperature, locate drafts, and insulate tubing and column, if necessary. Set the cell temperature to a minimum of 5 C above ambient.

Bubbles in flow cell Irregular noise on baseline Temperature changes in room

Cell temperature set lower than ambient Defective column Excessive baseline drift Unstable temperature control Defective cell heater Temperature changes in room

Cell temperature set lower than ambient

Maintenance

85

Table 5-1 Troubleshooting Guide (Continued) Symptom Possible Cause Bubbles in flow cell Solvent changeover Solution Remove bubbles and degas the solvent. Wait until baseline stabilizes (purge autosampler a few times). Check flow cell fittings and tighten. If leak continues, replace the flow cell. Wait until the baseline stabilizes. Remove bubbles and degas the solvent. Install a high-sensitivity noise filter on the pump.

Flow cell leak

Detector cannot be zeroed

Solvent changeover Bubbles in flow cell

Continuous noise at high sensitivity (<1S)

Pump crossover noise

Troubleshooting

86

Appendix A Specifications
This appendix includes information on: Operational specifications Mechanical specifications Environmental specifications Electrical specifications Communications Table A-1 Operational Specifications Condition Drift Specification Less than 0.05 S/hr/C (ambient) Base: 200 S Sensitivity: 0.005 Eluent: 1 mM KCI Less than 0.005 S/cm Base: 200 S Sensitivity: 0.005 Eluent: 1 mM KCI Front-panel selectable: OFF, 30, 35, 40, 45, 50, 55, 60, 65 C Performance: 0.5 C/hr Fast: 0.25 sec Standard: 0.5 sec Slow: 1.0 sec

Noise

Temperature control

Response times

Table A-2 Mechanical Specifications Condition Cell volume Wetted materials 0.6 L 316 stainless steel, PTFE, and PCTFE Specification

Specifications

87

Table A-2 Mechanical Specifications (Continued) Condition Operating pressure Height Length Width Weight Specification 70 psi maximum 9.4 inches (23.8 cm) 21 inches (53.3 cm) 11.5 inches (29.2 cm) 17.7 pounds (8 kg)

Table A-3 Environmental Specifications Condition Operating temperature range Operating humidity Specification 4 to 35 C (40 to 95 F) 20 to 80% RH, noncondensing

Table A-4 Electrical Specifications Condition Protection class


a b

Specification Class I II 2 Normal (IPXO) 100/120 VAC 220/240 VAC 50 Hz: 2 Hz 60 Hz: 2 Hz T2 A (20 mm) T1 A (20 mm) 0.6 A

Over-voltage category Pollution degree


c d

Moisture protection

Line voltages (grounded AC), nominal Line frequency ranges 100/120 VAC fuse rating 220/240 VAC fuse rating Current (Max)

a. Protection Class I The insulating scheme used in the instrument to protect you from electrical shock. Class I identifies a single level of insulation between live parts (wires) and exposed conductive parts (metal panels), in which the exposed conductive parts are connected to a grounding system. In turn, this grounding system is connected to the third pin (ground pin) on the electrical power cord plug.

Specifications

88

b. Over Voltage Category II Pertains to instruments that receive their electrical power from a local level such as an electrical wall outlet. c. Pollution Degree 2 A measure of pollution on electrical circuits, which may produce a reduction of dielectric strength or surface resistivity. Degree 2 refers to normally only nonconductive pollution. Occasionally, however, a temporary conductivity caused by condensation must be expected. d. Moisture Protection Normal (IPXO) IPXO means that there is NO Ingress Protection against any type of dripping or sprayed water. The X is a placeholder to identify protection against dust, if applicable.

Table A-5 Communications Signal Recorder output Integrator output Marker output Controller bus Specification 0 to 10 mV 10, 50, 100 S/1V FS Isolated contact output IEEE-488, PowerLine

Specifications

89

Appendix B Spare Parts


The parts listed in Table B-1 are spare parts recommended for installation by you, the customer. Any parts that are not listed may require installation by a trained service representative. Order a supply of the parts listed in Table B-1 to keep in stock for use as needed. Table B-1 Spare Parts Item Fuse, Time Delay, 1A, 250V, IEC Fuse, Time Delay, 2A, 5x20 mm T Fitting kit Pump Seal Replacement kit Union Quantity 2 2 1 2 1 Part Number WAT165-11 WAT165-14 WAT025604 WAT022934 WAT097332

Note: The flow cell (part number 043069) is considered a replacement part. Order the flow cell only when it is needed for replacement in the Waters 432 Detector. Fill in the information below for easy reference when you order parts or request service.

Installation Date: _____________ Serial Number: ____________

Spare Parts

90

Appendix C Ion Chromatography Methods


This appendix provides information about: General-purpose anion analysis using conductivity and UV detection Alkali and alkaline earth cations, ammonium, and amines

C.1 General-Purpose Anion Analysis Using Conductivity and UV Detection


Table C-1 Required Instrumentation Instrument Part Number

Alliance, 2695 Separations Module or Breeze N/A (with column heater, seal wash, and degasser) 432 Conductivity Detector busSAT/IN Module Empower/Breeze data processing UV Detector (optional) 043061 200415 Contact Waters Contact Waters

Ion Chromatography Methods

91

Figure C-1 Common Anion Standards

C
Table C-1 Analysis Conditions Condition Column Eluent Back conductivity Degas Flow rate Backpressure Temperature Injection Needle wash Detection Base range Attenuation Polarity Value IC-Pak A/HR (part number 026765) Borate/gluconate containing 12% AcCN 240 20 S Continuous 1 mL/min. 1200 200 psi 30 C for column heater, 35 C for detector 100 L 12% AcCN in DI water Direct conductivity 500 50 S/Volt unattenuated Positive

Ion Chromatography Methods

92

C.1.1 Preparing Eluent


1. Add 20 mL of lithium borate/gluconate concentrate and 120 mL of HPLC-grade acetonitrile (AcCN) into a 1-liter volumetric flask (see Section C.1.13, Preparing Lithium Borate/Gluconate 50X Stock Concentrate, for concentrate preparation). 2. Dilute to volume with DI water. 3. Ensure the natural pH is 8.2 0.1. 4. Vacuum degas through a 0.45-m aqueous and solvent-compatible membrane filter. 5. Store in a glass or plastic container at ambient temperature. Discard after 1 month.

C.1.2 Preparing Standards


It is recommended that you use certified 1000-ppm anion standards with this method. If unavailable, see Section C.2.7, Preparing Stock Reagent, for uncertified standard preparation. Prepare at least three mixed analyte standards within the expected range of sample analyte concentration. This method is linear from 0.1 to 100 ppm. After you validate the multi-point calibration curve, a single-point calibration curve within the expected analyte concentration range is appropriate for recalibration.

C.1.3 Preparing a Sample


1. Determine the expected range of analyte concentration and other anionic component in the sample matrix. The major analyte should be less than 100 ppm for best results. 2. Dilute the sample with DI water, if necessary. 3. If the sample contains high amounts of neutral organics or is highly colored, pass the diluted sample through a C 18 Sep-Pak cartridge. Anions pass through unretained, but there can be a loss of fluoride recovery. 4. Filter samples containing suspended solids through a 0.45-m aqueous-compatible disk before injection. Failure to filter solids can increase column backpressure. Note: For best results, ensure sample pH is between 3 and 11. Samples treated with a sample preparation disk in the H form, used to remove cations and neutralize high pH, will yield chromatograms similar to suppressed conductivity chromatograms.
+

Ion Chromatography Methods

93

C.1.4 Empower Data Processing Method


Table C-2 IC Processing Method Using Peak Apex for Retention Time Process Integration Peak Width = 30.0 Minimum Area = 3000 Inhibit Intg. = 0 to 2 min. Threshold = 10 to 25 Minimum Height = 150 Calibration Averaging = None Update RT = Never Peak Match = Closest Quant By = Peak Area Fit Type = Linear for multi-point calibration, Linear Through Zero for single-point calibration Report Analyte Name Analyte Retention Time Peak Area Amounts Values

Ion Chromatography Methods

94

C.1.5 Method Validation


This validation design is abstracted from ASTM/EPA validation. It has been used to validate all anion analysis methods. Many of the methods using this validation design are linear above 50 ppm. Table C-3 Method Validation Individual Youden Pair Standard, in ppm 1 Analyte Anion Cl Br NO2 SO4 NO3 F PO4 0.7 2.0 3.0 40.0 15.0 2.0 50.0 2.0 3.0 40.0 50.0 20.0 0.7 40.0 2 3.0 15.0 20.0 0.5 40.0 0.5 20.0 3 4 15.0 40.0 15.0 0.7 50.0 3.0 0.5 5 40.0 20.0 50.0 2.0 0.5 10.0 3.0 6 20.0 50.0 0.5 3.0 0.7 7.0 2.0 7 50.0 0.7 2.0 15.0 2.0 20.0 0.7 0.5 0.5 0.7 20.0 3.0 25.0 15.0 8

C.1.6 Method Linearity


Cl F Br r2 = 0.9999 r = 0.9986 r2 = 0.9999
2

Figure C-2 Calibration Curves for Chloride, Fluoride, and Bromide

Ion Chromatography Methods

95

NO2 NO3

r = 0.9999 r2 = 0.9992

Figure C-3 Calibration Curves for Nitrite and Nitrate

SO4 PO4

r2 = 0.9999 r = 0.9992
2

Figure C-4 Calibration Curves for Sulfate and Phosphate

Ion Chromatography Methods

96

C.1.7 Quantitation Precision


Quantitation Precision is the percent RSD of analyte peak area at each concentration. Data is based on seven replicate injections of the validation standards. Table C-4 Quantitation Precision Analyte 0.5 ppm Concentration 0.7 2 3 15 20 40 50 F 0.95 0.67 0.17 0.43 0.44 0.45 CI 1.11 1.64 0.18 0.17 0.05 0.04 0.05 0.13 NO2 3.44 0.78 0.56 0.20 0.28 0.05 0.04 0.02 Br 5.17 1.73 1.14 0.67 0.11 0.30 0.03 0.26 NO3 0.32 0.75 0.91 0.19 0.16 0.06 0.08 0.03 PO4 12.98 9.29 2.91 3.49 0.50 0.52 0.37 1.56 SO4 7.62 3.90 1.07 0.64 0.32 0.25 0.26 0.16

C.1.8 Method Detection Limits

C
100-PPB Standards

Figure C-5 100-L Injection

Ion Chromatography Methods

97

Based on this representative chromatogram using a 100-L injection, the estimated detection limits, as ppb, at three times signal to noise (S/N) are as follows: Fluoride = 50 Chloride = 25 Nitrite = 50 Bromide = 75 Nitrate = 75 Phosphate = 125 Sulfate = 75 Quantitation below these detection limits is not advised. You can obtain lower detection limits using a 250-L injection.

C.1.9 Quantitation Accuracy


The Certified Performance Evaluation Standards were diluted 1:100 with DI water. Amounts are based on a multi-point calibration curve prepared from certified standards. Table C-5 Quantitation Accuracy Analyte Performance Evaluation Standard Official Anion Methods Wet Chem & IC True Value in ppm Measured Mean Measured Std Dev Ave IC n=3 IC Using Alliance IC Pak A/HR and B/G Eluent IC/Mean IC/True Value F 2.69 CI 43.00 NO2 1.77 NO3 15.37 PO4 6.29 SO4 37.20

2.75 0.26 2.63 0.05 0.956 0.978

43.30 3.09 43.87 0.09 1.013 1.020

1.77 0.07 1.93 0.01 1.090 1.090

15.42 1.15 15.04 0.06 0.975 0.979

6.38 0.21 6.47 0.09 1.014 1.029

37.00 2.24 37.03 0.12 1.001 0.995

Ion Chromatography Methods

98

C.1.10 Analyte Recovery


The Certified Performance Evaluation Standards were diluted 1:100 with typical drinking water. Amounts are based on a multi-point calibration. Table C-6 Analyte Recovery Analyte F CI 25.82 0.04 0.16 43.00 69.64 0.08 0.11% 102.5% 1.77 1.82 0.004 0.21% 102.8% NO2 Not detected NO3 0.23 0.002 0.92 15.37 15.52 0.02 0.10% 99.5% 6.29 5.35 0.05 0.92% 85.1% PO4 Not detected SO4 8.30 0.02 0.27 37.20 46.46 0.17 0.37% 102.8%

Not Milford Drinking Water detected n=3, as ppm %RSD Performance 2.69 Evaluation Std MDW + PES n=3; as ppm %RSD % Recovery 2.46 0.04 1.51% 91.4%

Ion Chromatography Methods

99

C.1.11 Example of Use

Figure C-6 Typical Drinking Water, No Dilution Required

C.1.12 Using Direct UV Detection


Many anions are UV active in the range of 205 to 214 nm, such as NO2, Br, and NO3, and the use of direct UV detection provides a degree of detector selectivity. Figure C-7 shows the chromatogram of the anion standard that demonstrates this selectivity. Generally, the lower the wavelength of detection, the greater is the response, as seen with the difference between 205- and 214-nm chromatograms. However, the borate/gluconate eluent has some UV absorption which causes negatives peaks at the retention time of the UV transparent anion, such as F, Cl, PO4, and SO4, if present.

Ion Chromatography Methods

100

Figure C-7 Direct UV Detection Direct UV detection is five times more responsive for nitrite and nitrate than is conductivity detection and therefore provides lower detection limits. The chromatogram in Figure C-8 of a 100-ppb anion standard demonstrates the improved sensitivity.

Waters 996 Photodiode Array Detector

Figure C-8 100-ppb Anion Standard

Ion Chromatography Methods

101

The borate/gluconate eluent has some UV absorption. Use of eluents that are UV transparent, such as hydroxide and carbonate/bicarbonate, provide lower detection limits.

C.1.13 Preparing Lithium Borate/Gluconate 50X Stock Concentrate


1. Using a 60-mL plastic syringe, add approximately 25 g of BioRad AG-50W-X12 strong cation exchange resin in the hydrogen form, or equivalent. Wash the resin with five 20-mL portions of DI water to remove any ionic impurities from the resin. Discard the washings. 2. Dissolve 9.06 g of sodium gluconate in approximately 20-mL of DI water. After dissolution, transfer this solution into the 60-mL syringe with resin. Slowly pass this solution into a 1-liter volumetric flask. Wash the resin with five 20-mL portions of DI water, then add the washing to the volumetric flask. Discard the syringe and resin. An alternative is to use commercially available 50% gluconic acid. However, it comes as a brown solution that, when diluted, gives a yellow tint to the eluent that can affect long-term performance. You can remove the brown color by passing 5 mL of 50% gluconic acid through a C18 Sep-Pak cartridge. This requires three Sep-Pak cartridges. Use 13.2 mL of 50% gluconic acid for the eluent concentrate. 3. Adjust the volume in the flask to approximately 500 mL with DI water and use a stirring bar. Add 7.2 g of lithium hydroxide monohydrate and 25.5 g of boric acid. With the aid of a magnetic stirrer, mix until all reagents are dissolved. 4. Add 94 mL of 95% glycerol and mix. Remove the stirring bar and fill to the mark with DI water. 5. Store this lithium borate/gluconate concentrate in a plastic container at ambient temperature for up to 6 months. You can store it at 4 C for up to 1 year, but warm it to ambient temperature before use. 6. A white string-like material observed at the bottom of the container indicates significant microbiological growth. If present, discard and prepare again.

C.1.14 Preparing Lithium Borate/Gluconate Eluent


1. In a 1-liter volumetric flask, add 20 mL of the 50X lithium borate/gluconate concentrate and dissolve in 500 mL of DI water. Add 120 mL of HPLC-grade acetonitrile. Mix and fill to the mark with DI water. 2. Vacuum degas through a 0.45-m aqueous / organic membrane. The background conductivity of the lithium borate/gluconate eluent is between 220 and 270 S.

Ion Chromatography Methods

102

C.2 Alkali and Alkaline Earth Cations, Ammonium, and Amines


Table C-1 Required Instrumentation Instrument Part Number

Alliance, 2695 Separations Module or Breeze N/A (with column heater, seal wash, and degasser) 432 Conductivity Detector busSAT/IN Module Empower/Breeze data processing 043061 200415 Contact Waters

Figure C-9 1-ppm Standard

Ion Chromatography Methods

103

Table C-2 Analysis Conditions Condition Column Eluent Back conductivity Degas Flow rate Backpressure Temperature Injection Needle wash Detection Base range Attenuation Polarity IC-Pak C/MD 3 mM HNO3/0.1 mM EDTA 1250 50 S Continuous 1 mL/min. 2100 psi 30 C for column heater, 35 C for detector 100 L 12% AcCN in DI water Indirect conductivity 2000 100 S/Volt unattenuated Negative Value

C.2.1 Preparing Eluent


1. Add 0.029 g of EDTA as the free acid into a 1-liter plastic volumetric flask. 2. Dissolve in 500 mL of DI water with a stirring bar for 30 min. 3. Add 30 mL of 100 mM HNO3 (or 189 L of concentrated HNO3). 4. Dilute to volume with DI water. 5. Vacuum degas through a 0.45-m aqueous-compatible membrane to remove excess EDTA crystals. 6. Store in a plastic container at ambient temperature. Discard after 1 month.

C.2.2 Preparing Standards


It is recommended that you use certified 1000-ppm anion standards with this method. If unavailable, see Section C.2.7, Preparing Stock Reagent, for uncertified standard preparation. Prepare at least three mixed analyte standards, using plastic volumetric flasks, within the expected range of the sample analyte concentration. This method is linear from 0.05 to 20 ppm for lithium, sodium, and ammonium, and 0.05 to 50 ppm for potassium, magnesium, and calcium. Above these concentrations, the response is off scale. After the multi-point

Ion Chromatography Methods

104

calibration curve is validated, a single-point calibration curve within the expected analyte concentration is appropriate for future calibrations. You can use this method for the analysis of Rb, Cs, Sr, and Ba.

C.2.3 Preparing a Sample


1. Determine the expected range of analyte concentration and other anionic component in the sample matrix. Sodium should be less than 20-ppm for best results. 2. Dilute the sample with DI water, if necessary. 3. If the sample contains high amounts of neutral organics or is highly colored, pass the diluted sample through a C 18 Sep-Pak cartridge. Cations pass through unretained. There can be residual Na contamination from the cartridge. 4. Filter samples containing suspended solids through a 0.45-m aqueous-compatible disk before injection. Failure to filter solids can increase column backpressure. Note: For best results, ensure sample pH is between 2 and 7 (especially the alkaline earth cations). Samples with pH less than 10 are appropriate for the alkali cations, ammonium, and amines. 5. For samples with pH less than 2, dilute the sample 1:10 with DI water or treat the sample with an Alltech IC-OH cartridge to remove anions and neutralize pH.

C.2.4 Empower Data Processing Method


Table C-3 IC Processing Method Using Peak Apex for Retention Time Process Integration Peak Width = 30.0 Minimum Area = 3000 Inhibit Intg. = 0 to 2 min. Threshold = 25 to 40 Minimum Height = 500 Calibration Averaging = None Update RT = Never Peak Match = Closest Quant By = Peak Area Fit Type = Linear for multi-point calibration, Linear Through Zero for single-point calibration Values

Ion Chromatography Methods

105

Table C-3 IC Processing Method Using Peak Apex for Retention Time (Continued) Process Report Analyte Name Analyte Retention Time Peak Area Amounts Values

C.2.5 Method Detection Limits

Figure C-10 25-ppb Cation Standard Based on this representative chromatogram using a 100-L injection in a 2695 Separations Module, the estimated detection limits, as ppb, at three times signal to noise (S/N) are as follows: Lithium = 1 Potassium = 15 Sodium = 5 Magnesium = 10 Ammonium = 5

Ion Chromatography Methods

106

Calcium = 15 You can achieve lower detection limits by using a 250-L injection.

C.2.6 Examples of Use

C
Figure C-11 Typical Drinking Water, No Dilution Required

Figure C-12 Typical Municipal Wastewater, Diluted 1:50, Overlay of Duplicate Injections

Ion Chromatography Methods

107

Note: Alkyl and alkanol amine analysis standards are between 1 and 5 ppm, 3 mM HNO3/0.1 mM EDTA/3% AcCN.

C.2.7 Preparing Stock Reagent


Because it is difficult to prepare a stock eluent for this column, it is recommended to prepare fresh working eluent. To prepare stock reagent: 1. In a 1-liter plastic volumetric flask, add 0.029 g of EDTA (as the free acid, not its salts) in 800 mL of DI water. Place on a magnetic stir plate and stir for 10 minutes. 2. While stirring, add 189 L of concentrated nitric acid and mix for 5 minutes. 3. Remove the stirring bar and fill to the mark with DI water. 4. Filter through a 0.45-m aqueous-compatible membrane filter before use. There can be some remaining white crystals on the filter (EDTA). This does not affect the performance of the eluent. Discard the filter.

Ion Chromatography Methods

108

Appendix D Validation Support


This appendix provides information about: Validation regulation overview Waters regulatory compliance support

Validation Regulation Overview


Federal regulatory codes require that instrumentation and automated systems used for generation, measurement, and assessment of data undergo: Operational qualification (performance verification) following repairs, maintenance, and substantial periods of operation Routine maintenance Federal regulatory codes also require that laboratories maintain: Written Standard Operating Procedures (SOPs) that indicate dates of operational qualification and maintenance Logs of results Note: Designate a person to be responsible for maintaining federally required records.

Waters Regulatory Compliance Support


Waters provides a wide range of documentation and services to assist customers in complying with the following areas of Standard Operating Procedure regulatory requirements: Basic operation Instrument maintenance Error messages, diagnostics, and tests Instrument performance qualification (calibration) Basic Operation Regulatory requirements regarding proper instruction for preparing, programming, and operating the 432 Detector are satisfied by the Waters 432 Conductivity Detector Operators Guide.

Validation Support

109

Instrument Maintenance Chapter 5, Maintenance, in the Waters 432 Conductivity Detector Operators Guide satisfies regulatory requirements for routine instrument maintenance. Chapter 5 includes: Maintenance considerations Calibration adjustment Replacing the flow cell Replacing fuses Summary of 432 Detector error messages Troubleshooting tables

Additional Waters Support


For more information about compliance support products and services, contact Waters Technical Service Department.

Validation Support

110

Index
A
Auto Base 59 Auto Zero 36, 39, 46, 59

D
Display 57 Documentation conventions 24 related 23

I N D E X

B
Balance 39 Balance key 58 Base Range 62 Base Range key 58 Baseline drift 85 noise 85 Beep-on-error 60 Beep-on-keystroke 60 Bubbles, removing 84 Bus LAC/E connections 41 Bus SAT/IN connections 40

E
Eluent anion, preparing 70 cation, preparing 70 general guidelines 69 high-pH 65 nitric acid, preparing 70 Eluent handling 63 Error messages 57, 83

F
Flow cell removing bubbles 84 theory 27 Fluid connections assembling fittings 50 making connections 51

C
Carbonate absorption, minimizing 66 Chart Mark key 59 Chart recorder connections 45 Chart recorder offset 58 Clear key 59 Compression screw 50 Configuration, system 68 Connections auto zero input 46 Bus SAT/IN to Bus LAC/E 41 chart recorder 45 marker input 45 32 Millennium 39 PowerLine controller 37 Containers, selecting and preparing 65 Contamination, removal. See passivation Conventions, documentation 24

I
IEEE-488 address 38 connector 37 Initialization self-test 57 Integrator offset 58 output 36, 62 sensitivity 39 sensitivity range multiplier 58 Ion chromotography methods 91 Index 111

K
Key descriptions 58 60

R
Recorder output 36, 45, 62 sensitivity 39 Related documentation 23 Remote key 58 Response key 59

L
LAC/E connections. See Bus LAC/E connections Leak Alert output 36

I N D E X

M
Marker input 36, 45 Marker output 36, 44, 59 32 Millennium connections 39

S
Sample preparation 67 Sensitivity Range key 58 Shift key 59 Soda lime tube, using 66 Spare parts 90 Specifications 87 Standards anion, injecting 72 anion, preparing 71 cation, injecting 76 cation, preparing 74 concentration 70 shelf-life 70 Standby setup 61 Storage, long-term 62 System configuration 68

N
Nitric acid eluent, preparing 70 Noisy baseline, troubleshooting 85

O
Offset chart recorder 58 integrator 58 polarity 58

P
Passivation 53 Polarity key 58, 63 Power requirements 31 Power switch 57 PowerLine operation 39 Pulse dampener 52, 69

T
Temperature equilibration 62 Temperature key 58 Time constant, detector 59 Troubleshooting 82 Tubing cutting polymeric 50 cutting steel 49

Index 112

U
Unpacking the 432 Detector 31

V
Validation support 109 Voltage, operating 34 ranges 34

I N D E X

W
Water purity 64

Index 113

Das könnte Ihnen auch gefallen