Sie sind auf Seite 1von 7

Mass spectrometry (MS) is an analytical technique that measures the massto-charge ratio of charged particles.

It is used for determining masses of particles, for determining the elemental composition of a sample or molecule, and for elucidating the chemical structures of molecules, such as peptidesand other chemical compounds. The MS principle consists of ionizing chemical compounds to generate charged molecules or molecule fragments and measuring their mass-to-charge ratios. In a typical MS procedure: 1. A sample is loaded onto the MS instrument, and undergoes vaporization 2. The components of the sample are ionized by one of a variety of methods (e.g., by impacting them with an electron beam), which results in the formation of charged particles (ions)
[1] [1]

3. The ions are separated according to their mass-to-charge ratio in an analyzer by electromagnetic fields 4. The ions are detected, usually by a quantitative method 5. The ion signal is processed into mass spectra MS instruments consist of three modules: An ion source, which can convert gas phase sample molecules into ions (or, in the case of electrospray ionization, move ions that exist in solution into the gas phase) A mass analyzer, which sorts the ions by their masses by applying electromagnetic fields A detector, which measures the value of an indicator quantity and thus provides data for calculating the abundances of each ion present The technique has both qualitative and quantitative uses. These include identifying unknown compounds, determining the isotopic composition of elements in a molecule, and determining the structure of a compound by observing its fragmentation. Other uses include quantifying the amount of a compound in a sample or studying the fundamentals of gas phase ion chemistry (the chemistry of ions and neutrals in a vacuum). MS is now in very common use in analytical laboratories that study physical, chemical, or biological properties of a great variety of compounds.

Applications
Isotope ratio MS: isotope dating and tracking

Mass spectrometer to determine the

16

O/ O and C/ C isotope ratio on biogenous carbonate

18

12

13

Main article: Isotope ratio mass spectrometry Mass spectrometry is also used to determine the isotopic composition of elements within a sample. Differences in mass among isotopes of an element are very small, and the less abundant isotopes of an element are typically very rare, so a very sensitive instrument is required. These instruments, sometimes referred to as isotope ratio mass spectrometers (IR-MS), usually use a single magnet to bend a beam of ionized particles towards a series ofFaraday cups which convert particle impacts to electric current. A fast online analysis ofdeuterium content of water can be done using Flowing afterglow mass spectrometry, FA-MS. Probably the most sensitive and accurate mass spectrometer for this purpose is theaccelerator mass spectrometer (AMS). Isotope ratios are important markers of a variety of processes. Some isotope ratios are used to determine the age of materials for example as incarbon dating. Labeling with stable isotopes is also used for protein quantification. (see protein characterization below)
Trace gas analysis

Several techniques use ions created in a dedicated ion source injected into a flow tube or a drift tube: selected ion flow tube (SIFT-MS), and proton transfer reaction (PTR-MS), are variants of chemical ionization dedicated for trace gas analysis of air, breath or liquid headspace using well defined reaction time allowing calculations of analyte concentrations from the known reaction kinetics without the need for internal standard or calibration.
Atom probe

Main article: Atom probe An atom probe is an instrument that combines time-of-flight mass spectrometry and field ion microscopy (FIM) to map the location of individual atoms.
Pharmacokinetics

Main article: Pharmacokinetics Pharmacokinetics is often studied using mass spectrometry because of the complex nature of the matrix (often blood or urine) and the need for high sensitivity to observe low dose and long time point data. The most common instrumentation used in this application is LC-MS with a triple quadrupole mass spectrometer. Tandem mass spectrometry is usually employed for added specificity. Standard curves and internal standards are used for quantitation of usually a single pharmaceutical in the samples. The samples represent different time points as a pharmaceutical is administered and then metabolized or cleared from the body. Blank or t=0 samples taken before administration are important in determining background and ensuring data integrity with such complex sample matrices. Much attention is paid to the linearity of the standard curve; however it is not uncommon to use curve fitting with more complex functions such as quadratics since the response of most mass spectrometers is less than linear across large concentration ranges. There is currently considerable interest in the use of very high sensitivity mass spectrometry for microdosing studies, which are seen as a promising alternative to animal experimentation.
[36][37][38]

Protein characterization

Main article: Protein mass spectrometry Mass spectrometry is an important emerging method for the characterization and sequencing of proteins. The two primary methods for ionization of whole proteins are electrospray ionization (ESI) and matrix-assisted laser desorption/ionization (MALDI). In keeping with the performance and mass range of available mass spectrometers, two approaches are used for characterizing proteins. In the first, intact proteins are ionized by either of the two techniques described above, and then introduced to a mass analyzer. This approach is referred to as "top-down" strategy of protein analysis. In the second, proteins are enzymatically digested into smaller peptides using proteases such as trypsin or pepsin, either in solution or in gel afterelectrophoretic separation. Other proteolytic agents are also used. The collection of peptide products are then introduced to the mass analyzer. When the characteristic pattern of peptides is used for the identification of the protein the method is called peptide mass fingerprinting (PMF), if the identification is performed using the sequence data determined in tandem MS analysis it is called de novo sequencing. These

procedures of protein analysis are also referred to as the "bottom-up" approach.


Glycan analysis

Mass spectrometry (MS), with its low sample requirement and high sensitivity, has been the predominantly used in glycobiology for characterization and elucidation of glycan structures. Mass spectrometry provides a complementary method to HPLC for the analysis of glycans. Intact glycans may be detected directly as singly charged ions by matrix-assisted laser desorption/ionization mass spectrometry (MALDI-MS) or, following permethylation or peracetylation, by fast atom bombardment mass spectrometry (FAB-MS). Electrospray ionization mass spectrometry (ESI-MS)
[39] [40]

also gives good signals for the smaller glycans. Various free and commercial software are now available which interpret MS data and aid in Glycan structure characterization.
[41]

Space exploration

As a standard method for analysis, mass spectrometers have reached other planets and moons. Two were taken to Mars by the Viking program. In early 2005 the Cassini-Huygens mission delivered a specialized GC-MS instrument aboard the Huygens probe through the atmosphere of Titan, the largest moon of the planet Saturn. This instrument analyzed atmospheric samples along its descent trajectory and was able to vaporize and analyze samples of Titan's frozen, hydrocarbon covered surface once the probe had landed. These measurements compare the abundance of isotope(s) of each particle comparatively to earth's natural abundance. Also onboard the CassiniHuygens spacecraft is an ion and neutral mass spectrometer which has been taking measurements of Titan's atmospheric composition as well as the composition of Enceladus' plumes. A Thermal and Evolved Gas Analyzer mass spectrometer was carried by the Mars Phoenix Lander launched in 2007. Mass spectrometers are also widely used in space missions to measure the composition of plasmas. For example, the Cassini spacecraft carries the
[42] [43]

Cassini Plasma Spectrometer (CAPS), which measures the mass of ions in Saturn's magnetosphere.
[44]

Respired gas monitor

Mass spectrometers were used in hospitals for respiratory gas analysis beginning around 1975 through the end of the century. Some are probably still in use but none are currently being manufactured.
[45]

Found mostly in the operating room, they were a part of a complex system, in which respired gas samples from patients undergoing anesthesia were drawn into the instrument through a valve mechanism designed to sequentially connect up to 32 rooms to the mass spectrometer. A computer directed all operations of the system. The data collected from the mass spectrometer was delivered to the individual rooms for the anesthesiologist to use. The uniqueness of this magnetic sector mass spectrometer may have been the fact that a plane of detectors, each purposely positioned to collect all of the ion species expected to be in the samples, allowed the instrument to simultaneously report all of the gases respired by the patient. Although the mass range was limited to slightly over 120 u, fragmentation of some of the heavier molecules negated the need for a higher detection limit. Electron spectrometer: In an Electron spectrometer, an incoming beam of electrons is bent with electric or magnetic fields. As higher energy electrons will be bent less by the beam, this produces a spatially distributed range of energies. Electron spectrometers are used on a range of scientific equipment, including particle accelerators, transmission electron microscopes, andastronomical satellites. An electrostatic electron spectrometer uses the electric field, which cause electrons to move along field gradients, whereas magnetic devices cause electrons to move at right angles to the field. Magnetic fields will act in a direction perpendicular to the electron propagation, thereby conserving velocity, whereas electrostatic fields will cause electrons to move along the field gradient, which may change electron energies if the component of the direction of propagation and field gradients are not perpendicular. Owing to these effects, sector based designs are commonly used in electron spectrometers. The effective potential in the solution of motion in a magnetic or electric system with rotational symmetry leads to radial focusing onto a mean
[1] [46]

radius. By superposition of a quadrupole field axial focusing is possible while weakening the radial focusing, until the astigmatism vanishes. By breaking the rotational symmetry a bit and varying the electrostatic potential along the mean path of the spherical aberration is minimized. The spectrometer can use entrance and exit slits or use a small source, which only emits into specific angle and a small detector. Photoelectronspectra from
[1]

single crystals exhibit a dependency on the emission angle, so that the entrance slit is needed. All the electrons from an isotopic source may be sucked off and focused into a directed beam (much like in an electron gun), which can then be analyzed. A position sensitive detector can detect the energy along one direction and depending on the additional optics lateral resolution or one angle along the other direction. Electrostatic spectrometers preserve the spin, which can be resolved afterwards. Color spectrometer: They are photometer instrument used in spectrophotometry to measure the intensity or transmitted light as function of wavelength or color.There are many applications of spectrophotometers, such as measuring the amount of light absorbed by a medium which the light is passing through. Spectrometers are also used to measure the amount of light which have been reflected by an object, these are called reflectance spectrophotometers. There are two major classes of measurement performed by spectrophotometers, these are called single beam and double beam. single beam spectrophotometers measures the change in intensity of a single light path through a medium. A double beam spectrophotometer compares the light intensity of two separate light paths and determine a ratio. Applications are throughout industry including, aerospace, material manufacturing,pharmaceutical,pulp or paper,building materials,chemical,food,glass,paint or coatings or plastic industries.

Constructing a cereal-box spectrograph


We can make good use of these household diffraction gratings for constructing our own spectrometer. It consists of two important elements: a CD, which separates the light into its different colours, and a narrow slit on the opposite side of the box which produces a narrow beam of light.
Construction of the cereal-

The slit is constructed on one side of the box using some thick paper and duct

tape. A slightly fancier model can be is placed at a 60 angle to made using the two small blades from a disposable razor, which are taped down the bottom of the box with the sharp edges facing each other, as Image courtesy of Mark Tiele shown in the picture. If the slit is too broad, the spectrum will be blurred, and if Westra it is too narrow, the spectrum will be too dim. A width of 0.2 mm seems to work well, but one can experiment. The quality of the spectra obtained is dependent on the quality of the slit, so it should be constructed with care.
box spectrometer. The CD

On the other side of the box, a CD (the author used an empty rewritable CD, cut in half using a pair of scissors) is mounted at a 60 angle to the bottom of the box. On top, a hole is cut, through which the CD can be observed. Stray light is excluded from the box by covering any holes around the CD, as well as the edges of the CD, with dark tape. To observe a spectrum, the slit is pointed towards a source of light (the closer the better), and the investigator looks down through the hole on top. Move the box around a little to get a feel for what to look at. Photos of the observed spectra can be made using a simple digital camera, with macro capability for close-up focus, attached to the box by tape or rubber bands. The camera should preferably have manual focus, as it can be difficult to get sharp spectra with auto-focus. Infrared spectrometer

An instrument used to identify and measure the concentration of chemical compounds (gases, nonaqueous liquids, and solids) with electromagnetic radiation from 800 nanometers to 1 millimeter. Infrared Spectroscopy is a method used to identify substances using spectrum analysis. It is a way used to identify specific substances by analyzing the spectrum it has when it absorbs particular wavelengths of infrared radiation

Das könnte Ihnen auch gefallen