Lecture 1

The Principles of Microscopy


BMS 524 - Introduction to Confocal Microscopy and Image Analysis Purdue University Department of Basic Medical Sciences, School of Veterinary Medicine

J.Paul Robinson, Ph.D. Professor of Immunopharmacology Director, Purdue University Cytometry Laboratories
These slides are intended for use in a lecture series. Copies of the graphics are distributed and students encouraged to take their notes on these graphics. The intent is to have the student NOT try to reproduce the figures, but to LISTEN and UNDERSTAND the material. All material copyright J.Paul Robinson unless stated. Textbook for this lecture series in Jim Pawleys Handbook of Confocal Microscopy Plenum Press which has been used extensively for material and ideas to support the class.
UPDATED December 27, 1998 J.Paul Robinson - Purdue University Cytometry Laboratories

Slide 1 t:/PowerPoint/confoc/524lect1.ppt

Evaluation
End of term quiz - 100% grade

J.Paul Robinson - Purdue University Cytometry Laboratories

Slide 2 t:/PowerPoint/confoc/524lect1.ppt

Introduction to the Course


Microscopy Fluorescence Basic Optics Confocal Microscopes Basic Image Analysis 3D image analysis Live Cell Studies Advanced Applications
Slide 3 t:/PowerPoint/confoc/524lect1.ppt

J.Paul Robinson - Purdue University Cytometry Laboratories

Introduction to Lecture 1
Early Microscopes Modern Microscopes Magnification Nature of Light Optical Designs

J.Paul Robinson - Purdue University Cytometry Laboratories

Slide 4 t:/PowerPoint/confoc/524lect1.ppt

Microscopes
Upright Inverted Khler Illumination Fluorescence Illumination
"Microscope" was first coined by members of the first "Academia dei Lincei" a scientific society which included Galileo

J.Paul Robinson - Purdue University Cytometry Laboratories

Slide 5 t:/PowerPoint/confoc/524lect1.ppt

Earliest Microscopes
1590 - Hans & Zacharias Janssen of Middleburg, Holland manufactured the first compound microscopes 1660 - Marcello Malpighi circa 1660, was one of the first great microscopists, considered the father embryology and early histology - observed capillaries in 1660 1665 - Robert Hooke (1635-1703)- book Micrographia, published in 1665, devised the compound microscope most famous microscopical observation was his study of thin slices of cork. He wrote:

. . . I could exceedingly plainly perceive it to be all perforated and porous. . . these pores, or cells, . . . were indeed the first microscopical pores I ever saw, and perhaps, that were ever seen, for I had not met with any Writer or Person, that had made any mention of them before this.

J.Paul Robinson - Purdue University Cytometry Laboratories

Slide 6 t:/PowerPoint/confoc/524lect1.ppt

Earliest Microscopes
1673 - Antioni van Leeuwenhoek (1632-1723) Delft, Holland, worked as a draper (a fabric merchant); he is also known to have worked as a surveyor, a wine assayer, and as a minor city official. Leeuwenhoek is incorrectly called "the inventor of the microscope" Created a simple microscope that could magnify to about 275x, and published drawings of microorganisms in 1683 Could reach magnifications of over 200x with simple ground lenses - however compound microscopes were mostly of poor quality and could only magnify up to 20-30 times. Hooke claimed they were too difficult to use - his eyesight was poor. Discovered bacteria, free-living and parasitic microscopic protists, sperm cells, blood cells, microscopic nematodes In 1673, Leeuwenhoek began writing letters to the Royal Society of London - published in Philosophical Transactions of the Royal Society In 1680 he was elected a full member of the Royal Society, joining Robert Hooke, Henry Oldenburg, Robert Boyle, Christopher Wren

J.Paul Robinson - Purdue University Cytometry Laboratories

Slide 7 t:/PowerPoint/confoc/524lect1.ppt

Secondary Microscopes
George Adams Sr. made many microscopes from about 1740-1772 but he was predominantly just a good manufacturer not inventor (in fact it is thought he was more than a copier!) Simple microscopes could attain around 2 micron resolution, while the best compound microscopes were limited to around 5 microns because of chromatic aberration In the 1730s a barrister names Chester More Hall observed that flint glass (newly made glass) dispersed colors much more than crown glass (older glass). He designed a system that used a concave lens next to a convex lens which could realign all the colors. This was the first achromatic lens. George Bass was the lens-maker that actually made the lenses, but he did not divulge the secret until over 20 years later to John Dolland who copied the idea in 1759 and patented the achromatic lens. In 1827 Giovanni Battista Amici, built high quality microscopes and introduced the first matched achromatic microscope in 1827. He had previously (1813 designed reflecting microscopes using curved mirrors rather than lenses. He recognized the importance of coverslip thickness and developed the concept of water immersion

J.Paul Robinson - Purdue University Cytometry Laboratories

Slide 8 t:/PowerPoint/confoc/524lect1.ppt

Lister, Abbe, Zeiss & Schott


In 1830, by Joseph Jackson Lister (father of Lord Joseph Lister) solved the problem of Spherical Aberration - caused by light passing through different parts of the same lens. He solved it mathematically and published this in the Philosophical Transactions in 1830 Ernst Abbe together with Carl Zeiss published a paper in 1877 defining the physical laws that determined resolving distance of an objective. Known as Abbes Law
minimum resolving distance (d) is related to the wavelength of light (lambda) divided by the Numeric Aperture, which is proportional to the angle of the light cone (theta) formed by a point on the object, to the objective.

Abbe and Zeiss developed oil immersion systems by making oils that matched the refractive index of glass. Thus they were able to make the a Numeric Aperture (N.A.) to the maximum of 1.4 allowing light microscopes to resolve two points distanced only 0.2 microns apart (the theoretical maximum resolution of visible light microscopes). Leitz was also making microscope at this time. Dr Otto Schott formulated glass lenses that color-corrected objectives and produced the first apochromatic objectives in 1886.

J.Paul Robinson - Purdue University Cytometry Laboratories

Slide 9 t:/PowerPoint/confoc/524lect1.ppt

Modern Microscopes
Early 20th Century Professor Khler developed the method of illumination still called Khler Illumination Khler recognized that using shorter wavelength light (UV) could improve resolution

J.Paul Robinson - Purdue University Cytometry Laboratories

Slide 10 t:/PowerPoint/confoc/524lect1.ppt

Khler
Khler illumination creates an evenly illuminated field of view while illuminating the specimen with a very wide cone of light Two conjugate image planes are formed
one contains an image of the specimen and the other the filament from the light

J.Paul Robinson - Purdue University Cytometry Laboratories

Slide 11 t:/PowerPoint/confoc/524lect1.ppt

Khler Illumination
condenser

Field iris

Specimen

eyepiece Field stop retina

Conjugate planes for image-forming rays

Field iris

Specimen

Field stop

Conjugate planes for illuminating rays

J.Paul Robinson - Purdue University Cytometry Laboratories

Slide 12 t:/PowerPoint/confoc/524lect1.ppt

Some Definitions
Absorption
When light passes through an object the intensity is reduced depending upon the color absorbed. Thus the selective absorption of white light produces colored light.

Refraction
Direction change of a ray of light passing from one transparent medium to another with different optical density. A ray from less to more dense medium is bent perpendicular to the surface, with greater deviation for shorter wavelengths

Diffraction
Light rays bend around edges - new wavefronts are generated at sharp edges - the smaller the aperture the lower the definition

Dispersion
Separation of light into its constituent wavelengths when entering a transparent medium - the change of refractive index with wavelength, such as the spectrum produced by a prism or a rainbow

J.Paul Robinson - Purdue University Cytometry Laboratories

Slide 13 t:/PowerPoint/confoc/524lect1.ppt

Refraction
Short wavelengths are bent more than long wavelengths

dispersion

Light is bent and the resultant colors separate (dispersion). Red is least refracted, violet most refracted.
J.Paul Robinson - Purdue University Cytometry Laboratories Slide 14 t:/PowerPoint/confoc/524lect1.ppt

Refraction

He sees the fish here .

But it is really here!!


J.Paul Robinson - Purdue University Cytometry Laboratories Slide 15 t:/PowerPoint/confoc/524lect1.ppt

Control

Absorption

No blue/green light
red filter
J.Paul Robinson - Purdue University Cytometry Laboratories Slide 16 t:/PowerPoint/confoc/524lect1.ppt

Light absorption
white light blue light red light green light

J.Paul Robinson - Purdue University Cytometry Laboratories

Slide 17 t:/PowerPoint/confoc/524lect1.ppt

Absorption Chart
Color in white light Color in white light red blue green
yellow magenta

Color of light absorbed Color of light absorbed blue red red blue green red red pink green green blue blue
Slide 18 t:/PowerPoint/confoc/524lect1.ppt

green green blue

cyan black gray

J.Paul Robinson - Purdue University Cytometry Laboratories

The light spectrum


Wavelength ---- Frequency
Blue light 488 nm short wavelength high frequency high energy (2 times the red)
Photon as a wave packet of energy

Red light 650 nm long wavelength low frequency low energy


J.Paul Robinson - Purdue University Cytometry Laboratories Slide 19 t:/PowerPoint/confoc/524lect1.ppt

Magnification
An object can be focussed generally no closer than 250 mm from the eye (depending upon how old you are!) this is considered to be the normal viewing distance for 1x magnification Young people may be able to focus as close as 125 mm so they can magnify as much as 2x because the image covers a larger part of the retina - that is it is magnified at the place where the image is formed
J.Paul Robinson - Purdue University Cytometry Laboratories Slide 20 t:/PowerPoint/confoc/524lect1.ppt

Magnification
35 mm slide 24x35 mm

1000mm

1000 mm M = 35 mm = 28

The projected image is 28 times larger than we would see it at 250 mm from our eyes. If we used a 10x magnifier we would have a magnification of 280x, but we would reduce the field of view by a factor of 10x.
Slide 21 t:/PowerPoint/confoc/524lect1.ppt

J.Paul Robinson - Purdue University Cytometry Laboratories

Some Principles
Rule of thumb is is not to exceed 1,000 times the NA of the objective Modern microscopes magnify both in the objective and the ocular and thus are called compound microscopes - Simple microscopes have only a single lens

J.Paul Robinson - Purdue University Cytometry Laboratories

Slide 22 t:/PowerPoint/confoc/524lect1.ppt

Basic Microscopy
Bright field illumination does not reveal differences in brightness between structural details - i.e. no contrast Structural details emerge via phase differences and by staining of components The edge effects (diffraction, refraction, reflection) produce contrast and detail
Slide 23 t:/PowerPoint/confoc/524lect1.ppt

J.Paul Robinson - Purdue University Cytometry Laboratories

Microscope Basics
Originally conformed to the German DIN standard Standard required the following
real image formed at a tube length of 160mm the parfocal distance set to 45 mm object to image distance set to 195 mm
Object to Image Distance = 195 mm

Mechanical tube length = 160 mm

Focal length of objective = 45 mm

Currently we use the ISO standard

J.Paul Robinson - Purdue University Cytometry Laboratories

Slide 24 t:/PowerPoint/confoc/524lect1.ppt

The Conventional Microscope

Mechanical tube length = 160 mm

Object to Image Distance = 195 mm

Focal length of objective = 45 mm


Modified from Pawley Handbook of Confocal Microscopy, Plenum Press

J.Paul Robinson - Purdue University Cytometry Laboratories

Slide 25 t:/PowerPoint/confoc/524lect1.ppt

Upright Scope
Epiillumination Source

Brightfield Source

J.Paul Robinson - Purdue University Cytometry Laboratories

Slide 26 t:/PowerPoint/confoc/524lect1.ppt

Inverted Microscope
Brightfield Source

Epiillumination Source
J.Paul Robinson - Purdue University Cytometry Laboratories Slide 27 t:/PowerPoint/confoc/524lect1.ppt

Conventional Finite Optics with Telan system

Modified from Pawley Handbook of Confocal Microscopy, Plenum Press

Ocular Intermediate Image


195 mm 160 mm

Telan Optics Other optics Objective


45 mm

Sample being imaged


J.Paul Robinson - Purdue University Cytometry Laboratories Slide 28 t:/PowerPoint/confoc/524lect1.ppt

Infinity Optics
Ocular Primary Image Plane
The main advantage of infinity corrected lens systems is the relative insensitivity to additional optics within the tube length. Secondly one can focus by moving the objective and not the specimen (stage)

Tube Lens
Infinite Image Distance

Other optics Other optics Objective

Modified from Pawley Handbook of Confocal Microscopy, Plenum Press

Sample being imaged


J.Paul Robinson - Purdue University Cytometry Laboratories Slide 29 t:/PowerPoint/confoc/524lect1.ppt

Summary Lecture 1
Simple versus compound microscopes Achromatic aberration Spherical aberration Khler illumination Refraction, Absorption, dispersion, diffraction Magnification Upright and inverted microscopes Optical Designs - 160 mm and Infinity optics

J.Paul Robinson - Purdue University Cytometry Laboratories

Slide 30 t:/PowerPoint/confoc/524lect1.ppt