Titrant Aqueous acid/bases Specifics Electrode description* Cat. No./ Order No. Reference Electrolyte c(KCl) = 3 mol/L 020096101 (6.2308.020) alkaline samples Combined LL pH glass electrode, U glass 020912162 (6.0222.100) c(KCl) = 3 mol/L 020096101 (6.2308.020) titrations performed at high temperatures Unitrode 020910674 (6.0259.100) c(KCl) = 3 mol/L 020096101 (6.2308.020) titrations with small sample volumes Combined LL micro pH glass electrode OR Combined flat membrane pH glass electrode 020912235 (6.0234.100) 020915617 (6.0227.100) c(KCl) = 3 mol/L 020096101 (6.2308.020) Shaft diameter 6.4 mm, minimal immersion depth; 10 mm Shaft diameter 12 mm, minimal immersion depth; 12 mm c(KCl) = 3 mol/L 020096101 (6.2308.020) Perchloric acid See AB 80 and Monograph"Nonaqueous titration of acids and bases with potentiometric endpoint indication" See AB 80 and Monograph"Nonaqueous titration of acids and bases with potentiometric endpoint indication" Solvotrode - combined LL pH glass electrode, ground-joint diaphragm (PCTFE) 020910551 (6.0229.100) Typical application: water analysis - p&m value etc. Remarks
Aquatrode plus
020910631 (6.0252.300)
c(LiCl) = 2 mol/L in easy to clean sleeve Ethanol diaphragm 020094435 (6.2312.010) IMPORTANT: When not in daily use, store in LiCI sat. in ethanol and soak membrane in distilled water before use. However, when in daily use, store in distilled water, so that only the membrane is immersed, not the diaphragm. Before the measurement lift the diaphragm sleeve slightly and allow a small amount of electrolyte to flow out. Rinse the electrode with distilled water, replenish the reference electrolyte and start the titration (see bottom of chart). IMPORTANT: Never fill the Solvotrode with LiCIO4 in glacial acetic acid! Electrode is also available with a plug-in head S: 20 94 852-3 (6.0430.500)3
TBAOH
020910551 (6.0229.100)
Potassium methylate
020910551 (6.0229.100)
Cyclohexylamine
020910551 (6.0229.100)
Silver nitrate
AB 25
arsenite, cerium sulphate, iron (III), iodine, potassium bromate, sodium nitrite, sodium thiosulphate, oxalic acid, permanganate, titanium
For titrations where the Pt-Titrode pH value during the titration doesn't change. For titrations during which the pH value changes. Combined LL Pt-ring electrode
6.0431.100
Electrode is also available with a plug-in head S: 20 94 852-3 (6.0431.500)3 c(KCl) = 3 mol/L 020096101 (6.2308.020)
6.0451.100
020928841 (6.0338.100) 020948604 (6.0435.110) L = 160 mm, Loger version available: (6.9904.023), L= 250 mm c(KCl) = 3 mol/L 020096101 (6.2308.020) 020959002 (6.0502.140) c(KCl) = 3 mol/L 020096101 (6.2308.020) a separate reference electrode is needed. Order no: 20 94 411-1 (6.0726.107) a separate reference electrode is needed. Order no: 20 94 411-1 (6.0726.107)
EDTA
020955856 (6.0508.110)
Determination of Al, Ion selective electrode Cu2+ Ba, Ca, Cd, Bi, Co, Fe, Mg, Ni, Pb, Zn. See AB 125
Determination of Ca, Mg, See AB 125 Surfactants in aqueous media Tego trant A 100,2 see AB 233
020955856 (6.0508.110) 020959321 (6.0507.120) c(KCl) = 3 mol/L 020096101 (6.2308.020) a separate reference electrode is needed. Order Nr: 6.0726.100
020959321 (6.0507.120)
020959304 (6.0507.010)
Inner chamber of a separate reference the reference electrode is needed. electrode: Order Nr: 6.0726.107 c(KCI) = 3 mol/L 020096101 (6.2308.020) Outer chamber: c(NaCL) = 1 mol/L c(KCl) = 3 mol/L 020096101 (6.2308.020) a separate reference electrode is needed. Order no: 20 94 411-1 (6.0726.107) electrode not suitable for high salt concentrations and pH > 10
Surfactrode resistant
020959312 (6.0507.130)
Surfactrode refill
020959339 (6.0507.140)
a separate reference electrode is needed. Order no: 20 94 411-1 (6.0726.107) electrode not suitable for Chloroform
AB = Application Bulletin * All these electodes are equipped with a Metrohm plub-in head G3
1
TEABr 0.4 mol/L tetra ethyl ammonium bromide in ethylene glycol (020094451) is an electrolyte solution for non-aqueous titrations. It is very useful when titrating low acid contents with TBAOH. In such cases the 2 mol/L LiCI in ethanol (020094435) would cause an alkali error during the measurement. The Solvotrode is a combined pH glass electrode specially designed for non-aqueous titrations. This electrode is normally filled with LiCI sat. in ethanol. However, when titrating with TBAOH we strongly recommend replacing the CiCI electrolyte with 0.4 mol/L TEABr in ethylene glycol. This electrolyte can of course also be used for all other non-aqueous titrations with various titrants.
2
TEGO trant A 100 020959274 (6.2317.000) (6 g) 020959282 (6.2317.010) (60 g) Available on request 6.2317.020 (500 g)
3
To connect an electrode with plug-in head G to a Titrino or 726/796 Titroprocessor, use one of the following cables: 020977345 (1 m) or 020977353 (3 m). See Metrosensor Electrode Catalog for additional information. Storage of the Solvotrode when not in daily use: Store the Solvotrode in 2 mol/L LiCI in ethanol. Use a normal Electrode holder.