Beruflich Dokumente
Kultur Dokumente
INTRODUCTION
The goal of the task is to analyze elements (Fe, Mg, Cu, Ca and Zn) in a sample of water with ICP-OES and determine which elements are contained in the sample and amount of these elements in water. Inductively coupled plasma optical emission spectrometry (ICP-OES), is an analytical technique used for the detection of trace metals. The sample is usually transported into the instrument as a stream of liquid sample. Inside the instrument, the liquid is converted into an aerosol through a process known as nebulization. The sample aerosol is then transported to the plasma where it is desolvated, vaporized, atomized, and excited and/or ionized by the plasma. The excited atoms and ions emit their characteristic radiation which is collected by a device that sorts the radiation by wavelength. The radiation is detected and turned into electronic signals that are converted into concentration information for the analyst. The intensity of this emission is indicative of the concentration of the element within the sample.
PLASMA
Cross section of an ICP torch and load coil depicting an ignition sequence. A - Argon gas is swirled through the torch. B - RF power is applied to the load coil. C - A spark produces some free electrons in the argon. D The free electrons are accelerated by the RF fields causing further ionization and forming plasma. E - The sample aerosol-carrying nebulizer flow punches a hole in the plasma.
The first function of the high temperature plasma is to remove the solvent from the aerosol, usually leaving the sample as microscopic salt particles. The next steps involve decomposing the salt particles into a gas of individual molecules (vaporization) that are dissociated into atoms (atomization).
The temperature varies in different areas of the plasma, being two or three times higher than those achieved in the flames of burning in the flame methods. Due to the high temperatures reached, the atomization is more complete and less chemical interference problem.
In ICP-OES, the light emitted by the excited atoms and ions in the plasma is measured to obtain information about the sample. The emission from the plasma is polychromatic, because the excited species in the plasma emit light at several different wavelengths. This polychromatic radiation must be separated into individual wavelengths so the emission from each excited species can be identified and its intensity can be measured without interference from emission at other wavelengths. The separation of light according to wavelength is generally done using a monochromator, which is used to measure light at one wavelength at a time, or a polychromator, which can be used to measure light at several different wavelengths at once.
At first, we had to prepare different solutions to calibrate the instrument (then, the wavelength will be correct and we will eliminate background in the measurement). To determine limit of detection, we analyzed the blank 10 times. We prepared the calibration solution with the metals (Zn, Fe, Mg, Cu and Ca) to identificate these elements in the sample after.
Zn poly
C (mg/L) 0 0,1 1 Zn poly 19,6133 42,5477 274,300
250
200 Intensity
150
100
50
The next graph shows the confidence bands of the calibration curve.
We can determine the limit of detection with the confidence bands (interpolating in the graph) or we can do it with this equation:
LOD (3)
0,0042
The results are different because with the graphical method, we determine the limit of detection without background. With the 3 method limit of detection is lower than in graphical method, because in this method we determine the limit of detection with
noise; the standard deviation is the mean value of the noise, and when we multiply by 3, we are supposing that we are measuring the intensity of the sample.
Zn mono
C (mg/L) 0 0,1 1
200
Intensity
150
100
50
0,1439 0,0029
The sensitivity of a method is related to the standard deviation. So that a steeper slope of the calibration curve implies a greater sensitivity. In this case the measure is more sensitive with the polychromator, because the slope is greater.
Fe polychromator
C (mg/L) 0 0,1 1
Intensity
0,0715 0,0030
Fe monochromator
C (mg/L) 0 0,1 1
200
0.0719 0,0018
In the case of iron, the measure is more sensitive with the monochromator.
Mg polychromator
C (mg/L) 0 0,1 1 Mg poly 2,8645 12,3747 111,3567
120 100 80 Intensity 60 40 20 0 0 0.2 0.4 0.6 C [mg/L] 0.8 1 1.2 y = 109.08x + 2.2023 R = 0.9999
INTENSITY 872,8367
0.0425 0,0004
Mg monochromator
C (mg/L) 0 0,1 1 Mg mono 79,9727 406,1767 3773,5
4000 3500 3000 Intensity 2500 2000 1500 1000 500 0 0 0.2 0.4 0.6 C [mg/L] 0.8 1 1.2 y = 3712.5x + 58.635 R = 0.9999
0,2391 0,0008
Cu polychromator
C (mg/L) 0 0,1 1
900 800 700 Intensity 600 500 400 300 200 100 0 0 0.2 0.4 0.6 C [mg/L] 0.8 1 1.2 y = 711.04x + 92.933 R = 0.9999
INTENSITY 123,3
0,1439 0,0019
Cu monochromator
C (mg/L) 0 0,1 1 Cu mono 1,5270 2,5176 13,5373
16 14 12 Intensity 10 8 6 4 2 0 0 0.2 0.4 0.6 C [mg/L] 0.8 1 1.2 y = 12.103x + 1.423 R = 0.9997
INTENSITY 1,5367
0,1449 0,0071
The measure with polychromator has a greater slope, then this measure is more sensitive.
Ca polychromator
C (mg/L) 0 0,1 1
70 60 50 Intensity 40 30 20 10 0 0 0.2 0.4 0.6 C [mg/L] 0.8 1 1.2 y = 52.498x + 4.8407 R = 0.9989
INTENSITY 297
0,1792 0,0013
Ca monochromator
C (mg/L) 0 0,1 1
700 600 Intensity 500 400 300 200 100 0 0 0.2 0.4 0.6 C [mg/L] 0.8 1 1.2 y = 559.52x + 49.904 R = 0.9987
INTENSITY 2291,031
0,3649 0,0034
The slope with monochromator is greater than with polychromator. Then, the measure is more sensitive with monochromator.
CONCLUSION
Water sample "Dobr voda": Ca 5.31 mg/l Mg 7.65 mg/l
The results are close to zero, then there isnt Zn in the water sample.
Fe POLY Sample
INTENSITY 3,1545
Fe MONO Sample
INTENSITY 12,8703
The results are close to zero, then there isnt Fe in the water sample.
INTENSITY 872,8367
The concentration isnt close to zero, is aproximately 8,05 mg/L and the producer declared that in Dobr Voda there was 7,65 mg/L. The result is correct and the difference could be due to possible errors in calibration. We had a problem with the calibration curve, because we pressed the start button to measure the sample before to put it in the pump.
Cu POLY Sample
INTENSITY 123,3
Cu MONO Sample
INTENSITY 1,5367
The results are close to zero, then there isnt Cu in the water sample.
The concentration isnt close to zero, is aproximately 5,41 mg/L and the producer declared that in Dobr Voda there was 5,31 mg/L. So, the result is correct.
With the values obtain with ICP, we can affirm that in the sample there are Ca and Mg with concentration 5,41 mg/L and 8,05 mg/L, respectively.