Display
Display size Resolution Number of traces Display layout and colors 10.4 inch SVGA resolution, 800 x 600 Up to 6 User-configurable
Monitoring: 0.5 to 35 Hz ST: 0.05 to 40Hz Diagnostic: 0.05 to 150 Hz Monitoring: 0.5 to 40 Hz ST: 0.05 to 35Hz Diagnostic: 0.05 to 150 Hz Range 2 to 700 mV Pulse width 0.5 to 2 ms Asystole, Bradycardia, Tachycardia, Ventricular fibrillation, Ventricular tachycardia ST numeric range -9 to +9 mm (-0.9 to +0.9 mV) ST accuracy -8 mm to +8 mm 0.2 mm or 10%, whichever is greater ST numeric resolution 0.1 mm (0.01 mV) ST Trends Up to 72 hours
60 Hz power supply
Controls
TrimKnob* hard keys control and Standard
Alarms
Priority Notification Setting High, Medium, Low and Message Audible and visual Default and individual Audio silence message General alarm message Audio pause Adjustment Trending 2 minutes Central alarm display and adjustment page 10 minutes graphical trends referenced to set alarm limits
ST segment analysis
Impedance respiration Range Accuracy Gain range 4 to 120 resp/min 5% or 5 resp/min, whichever is greater 0.1 to 5 cm/Ohm
Trends
Graphical Numerical All parameters, selectable time scales from 20 minutes to 72 hours All parameters, every 5 minutes sampling or after NIBP determination In both graphical and numerical trends 5 or 30 minute minitrends can be displayed for a continuous historical view
Note: Impedance respiration is intended for patients over three years old.
GE SpO2
Measurement range Pulse oximetry Pulse rate Measurement accuracy Saturation Pulse Rate 0 to 100% 30 to 250 bpm 70 to 100% 2 digits, 0 to 69% unspecified 5% or 5 bpm, whichever is greater
Note: Nellcor OxiMax SpO2 option is not available in all the countries. Please check the availability with your local representatives.
Temperature
Numerical display Scale Measurement range Display Resolution Probe T1, T2, T2-T1 Fahrenheit (F) Celsius (C) 10 to 45C (50 to 113F) 0.1C at 25 to 45C with reusable probes GE approved probes or defibrillator-proof YSI 400 series probes Measurement accuracy 0.1C without probe
NIBP
Measurement technique Oscillometric with step deflation Modes Manual, automatic and stat NIBP Measurement ranges Systolic Adult/Pediatric: 30 to 290 mmHg Neonate: 30 to 140 mmHg MAP Diastolic Accuracy Default initial inflation pressure Maximum determination time Over pressure monitor Adult/Pediatric: 20 to 260 mmHg Neonate: 20 to 125 mmHg Adult/Pediatric: 10 to 220 mmHg Neonate: 10 to 110 mmHg Meets AAMI SP10 Adult/Pediatric: 135 15 mmHg Neonate: 100 15 mmHg Adult/Pediatric 2 minutes Neonate: 85 s Adult/Pediatric: 300 to 330 mmHg Neonate: 150 to 165 mmHg
Thermal Recorder
Operating Principle Waveforms Numerics Tabular trend printout Thermal array Selectable 1, 2, or 3 waveforms HR, SpO2, NIBP, IBP1, IBP2, ETCO2, T1, T2, Resp HR, NIBP, IBP1, IBP2, T1, T2, Et/FiCO2, RR
Graphical trend printout HR, ST, IBP1, IBP2, NIBP, SpO2, Pleth, CO2, Resp, T1+T2 Resolution Vertical Horizontal Paper width Paper speed1 8 dots/mm (200 dots/inch) 24 dots/mm (600 dots/inch) 50 mm, printing width 48 mm 1, 6.25, 12.5, 25 mm/s
I/O connectors
RS-232 computer serial output, Defibrillation synch, Nurse call
Physical specifications
Dimensions (HxWxD) Without extension rack (5 mm) With extension rack (5 mm) Weight Ingress protection 312 mm x 312 mm x 158 mm 312 mm x 352 mm x 178 mm 6Kg IP21
Networking
Work in CARESCAPE Network and S/5 Network environments
Printing
Network laser printer supported in S/5 network
Mounting
GCX compatible Integrated transport handle
Warranty
One year
Electrical specifications
AC input Protection Battery Charging time Run time 100 to 240V 10%, 50/60 Hz, 150VA Class I Exchangeable lithium-ion, 2 pcs max 2 hours per battery pack Up to 4.5 hours
Certifications
IEC 60601-1 passed CE marking according to Directive 93/42/ EEC
GE Healthcare P.O. Box 900, FIN-00031 GE, Finland Tel. +358 10 394 11 Fax +358 9 146 3310 www.gehealthcare.com
2011 General Electric Company All rights reserved. General Electric Company reserves the right to make changes in specifications and features shown herein, or discontinue the product described at any time without notice or obligation. Contact your GE representative for the most current information. GE and the GE Monogram are trademarks of the General Electric Company. *CARESCAPE, DINAMAP, EK-Pro, Ohmeda, PROCARE, SuperSTAT and TrimKnob are trademarks of GE Medical Systems Information Technologies, Inc. The PROCARE trademark is not applicable in China. *Masimo and SET are trademarks of Masimo Corporation. *Nellcor and OxiMax are trademarks of Nellcor Puritan Bennett, Inc. GE Healthcare, a division of General Electric Company.