Outline of Instructions
Purpose of Wipe Tests UTHSC-H Wipe Test Requirements Wipe Test Materials Appropriate Lab Attire UTHSC-H Radiation Safety Forms Instructions for Wipe Tests How to Clean a Contaminated Area Safety Resources on the Web Summary
Additional requirements may occur for researchers working with greater than 5 mCi
Urine bioassay and more frequent surveys
Contamination Survey
For the protection of the researchers and all workers in the lab, a contamination survey should be performed at the end of each experiment or project. Not documented Can be performed with a Geiger-Mueller detector Simple check to make sure there is no contamination in the lab that may have resulted from the radiation work
www.uth.tmc.edu safety
Radiation Safety
RS-08
Gather equipment
Place samples in separate vial or envelope Label each vial or envelope noting the location of the sample Fill out Laboratory Wipe Test Report
Take samples using a S motion while rotating the cotton swab. Also shown, collecting a sample using a chem wipe.
Cut stick of cotton swab down to fit inside vial and close vial with lid.
Take cut cotton swab and place in tube Fill tube with approximately 1mL of cocktail Close tube
When the tab is in the opposite position it signals to the user and LSC that the sample has been read.
Various LSCs
Laboratory Wipe Test Report for Radioactive Materials for Authorized Users
Building: ________ Room Number: ________ Wipe Tested by: __________________________ Radioisotopes used in this Laboratory: Date: ____________________
Note: If Storage Only - Wipe test must at least be taken of storage area If No Use or Storage - In comment section indicate No Use or Storage This Month
Authorized User: _________________________ Counting Equipment . Type: LSC or Other (specify) _____________
Standard Isotope Example: H-3 _____ Example: P-32 ____ _________________ _________________ _________________ Efficiency (%) * 30 % ___________ 75 % ___________ _______________ _______________ _______________
Laboratory Diagram
dpm
cpm background
efficiency
Number Count Rate (dpm) 1 ________ 2 ________ 3 ________ 4 ________ 5 ________ 6 ________ 7 ________ 8 ________ 9 ________ 10 ________ 11 ________ 12 ________ 13 ________ 14 ________ 15 ________ 16 ________ 17 ________ 18 ________ Rewipe of # ___ ________ Rewipe of # ___ ________
Comments:___________________________________________________________________________________________________ ____________________________________________________________________________________________________________ ____________________________________________________________________________________________________________ ____________________________________________________________________________________________________________ ____________________________________________________________________________________________________________
Laboratory Wipe Test Report for Radioactive Materials for Authorized Users
Building: MSB____ Room Number: 5.432___ Wipe Tested by: John Smith ________________ Date: July 11, 2006_________
Authorized User: R. Emery ________________ Counting Equipment . _ Type: LSC or Other (specify) LSC_________
Standard Isotope Example: H-3_____ Example: P-32 ____ _________________ _________________ _________________ Efficiency (%) * 30 % ___________ 75 % ___________ _______________ _______________ _______________
Laboratory Diagram
6 7
dpm
cpm background
efficiency
2 8 1 5
Number Count Rate (dpm) 1 20 ______ 2 23 ______ 3 19 ______ 4 22 ______ 5 26 ______ 6 250 _____ 7 18 ______ 8 21 ______ 9 ________ 10 ________ 11 ________ 12 ________ 13 ________ 14 ________ 15 ________ 16 ________ 17 ________ 18 ________ Rewipe of # __6_ 22______ Rewipe of # ___ ________
3. Fume hood
4. Sink 5. Door 6. Lab bench 7. Lab bench 8. Sink
Comments: Wipe test for area 6 showed slight contamination. After cleaning the area it was re-wiped and tested to provide proof of decontamination. ____________________________________________________________________________________________________________ ____________________________________________________________________________________________________________ ____________________________________________________________________________________________________________
If Area is Contaminated
Take proper precautions IF MINOR Spill or Contaminated Area
Clean the affected area per proper lab cleanup procedures
IF MAJOR Spill or Contaminated Area Secure the area, prevent re-entry Contact Radiation Safety for Assistance
0485-005-317 During Business Hours 7534-005-317 After Hours (500-HELP) UT Police
Spray area with RADCON or Scrubbing Bubbles (DOW) to lift contamination from the surface
Review
Documented Monthly Wipe Tests for Radioactive Material Contamination are Required A Radioactive Spill or Contaminated Area Should be Decontaminated (Cleaned) and Wipe Tested Again For Assistance, You can Contact Radiation Safety at 713-500-5840