INTRODUCTION
The Model 108 is a 3 or 4 gas analyzer Dual beam IR sensor for CO2 (0-100%) Dual beam IR sensor for CH4 (0-100%) Electrochemical sensor for O2 (acidic electrolyte) (0-100%) Balance Gas or Electrochemical sensor for H2S (0 2,000 ppm) It has internal data logging capability (>7000 pts) with downloading SW in ASCII format Two point cal for each sensor Internal pump Easy to use/easy to train Simple sampling system for water removal (optional)
IR SENSOR SCHEMATIC
IR SENSOR S C H E M AT I C P H O TO O F C O M PA C T L F G A N A LY Z E R
ELECTROCHEMICAL SENSORS
O X Y G E N ( O 2) HYDROGEN SULFIDE ( H 2S ) O P T I O N
Acidic electrolyte Cathode: O2 + 4H+ = 4e- + 2H2O Anode: 2Pb 2Pb+2 + 4eOverall: O2 + 4H+ + 2Pb = 2H2O + 2Pb+2 Lifetime: 4 years
Oxidation Rxn.:
H2S + 4 H2O =
H2SO4 + 8H+ +8e-
BIOGAS
Chem CH4 CO2 N2 H2 % 5075 2550 010 01
H2 S
O2
01
0.1-1
H2 S
O2
03
02
BALANCE GAS
The Model 108 has software that measures & calculates the balance gas on a continuous basis: Balance gas = 100% - ( % CH4+% CO2 + %O2) OR The fourth position can be used for an electrochemical sensor such as H2S
SAMPLING CONSIDERATIONS
Gas Temperature-Samples must be cooled down to near ambient temperatures for optimum measurements from the sensors
Particulate- a filter is necessary to remove particulate to prevent clogging of lines, fouling the sensors & pump
Moisture- must be removed to prevent loss of sample gases and fouling of optical sensors
10
60
50
40 CH4 CO2 O2 20
30
Bal
10
0 1 2 3 4 5 Sample # 6 7 8 9 10
11
12
GC312
13
SUMMARY
The 108 is the ideal analyzer for checking landfill gas Calibrate the analyzer before and after the landfill gas measurements Be sure to use sampling probe to remove water and particulate
Download the data to the PC using the software provided with the analyzer
14