CONDUCTIVITY ANALYZER
DEFINATION
Conductance Is Reciprocal of Resistance Electric Conduction in a Solid Charge Carrier Electrons Electric Conduction in a Liquid Charge CarrierIons
CONDUCTIVITY ANALYZER
When a liquid contains more ions per unit volume, the number of charge that are transported in a unit time, is greater and
Units of Conductance
As it is defined that conductance is a reciprocal of Resistance G=1/R where G Conductance R Resistance As the unit of Resistance is Ohms then the unit of Conductance or G = ohm-1 But ohm-1 is defined as = mho
mho is the old and formal unit. The new unit is Siemens which is = mho
Units of Conductance
Siemens is wide range and usually this range is seldom used in industries. Smaller unit is siemens. But the unit will be defined as siemens / cm The conductivity of Boiler feed water is 10 siemens / cm
Working Principle
To measure the conductance of a liquid: Two electrodes / or two plates in a single housing: With the help of a electronic circuit The resistance or other words the conductance of the liquid is measured.
Material of Electrode
Construction
Titanium Palladium Electrode Chloro-trifluororthelene insulation
Graphite
Connection:
1 / 2 or 3 / 4 NPT
Cell Constant
Suppose there are two electrodes a & b Both are in equal size A = 1 cm Distance between two electrodes is denoted as L and it is also 1 cm
Cell Constant
Putting the values
cm 1 1 cm cmxcm cm
So; cell constant = cm 1
Cell Constant
Very Important Factor
Cell Constant
Cell Constant
Suppose the Readings are L = 1 meter, A 1 meter2 Then the Cell Constant will be m1 But practically, the area of cell (Electrode) 1 m2 and distance between two electrodes L = 1 meter length is not possible.
Cell Constant
So the cell constant will depends upon at the nature of process. Suppose, the conductivity of pure water is being measured, then the electrodes should be close to each other or cell cell constant should be 0.01 cm-1 or 0.1 cm-1. Where as the conductivity of boiler feed water is more than water and cell constant can be 1 cm1.
Cell Constant
Practically and usually the shape of electrodes are round, so the size of electrodes or the distance between two electrodes can vary. The area of electrode can be 2 cm2 where as distance between them may be 0.5 cm.
Cell constant
of solution varies.
The change in conductivity due to temperature can be rectified with the help of Auto temperature compensator.
Construction of Cell
The cell consists of two electrodes plus temperature compensator. Usually temperature compensator is a thermo-resistance. The temperature of the sample is tried to kept at ambient temperature with the help of coolers, but a thermo-resistance is must in the cell to correct any variation in temperature of sample.
Standard Solutions
Conductivity Probe
Conductivity Probe
Conductivity Probe