Beruflich Dokumente
Kultur Dokumente
04/25/2014
OMP40 Setup
OMP40 Setup
OMP40 Setup
OMP40 Setup
OMP40 Setup
MP700 Setup
OMI Setup
SW1-1 OFF
SW1-2 OFF
SW1-3 ON
SW1-4 OFF
Stylus Run-out
The part probe stylus should be indicated in
to be sure that it is centered on the probe body
(center line of rotation of the machine spindle).
This needs to be done in order to probe a part
with repeated accuracy. When rotated, the
stylus should have minimal run out, no greater
than .0003TIR. This run out is adjusted in the same
manner on the MP700 as well as the OMP40 and
the OMP400.
The closer to .0000TIR, the more accuracy can
be expected when probing
Tools required for this procedure are:
.0001 dial indicator
Magnetic base height gage
Two - 2mm hex wrenches
Integrator Parameters
Integrator Parameters
Stylus Run-out
When the stylus is indicated in for the first time, you
may want to use a .0005 dial indicator to get the
stylus close to center and then switch to the .0001
indicator for fine adjustment.
Be sure to pick up center of stylus ball in both X and
Z axis before making adjustments. Move X axis in
the + and direction to pick up the high point, and
then repeat for Z axis.
Stylus Run-out
Using two hex wrenches to adjust centering gives you a better feel for how
much adjustment you are making. Turn both screws clockwise or counter clockwise,
simultaneously, to move the probe body from side to side. All setscrews should be kept
slightly snug to control movement during this process.
Never back out the setscrews completely because this could allow the probe to fall
from the tool holder.
Stylus Run-out
There are two sets of setscrews.
Two upper screws spaced at
180 degrees apart and four
spaced at 90 degrees apart.
The four bottom screws are used for centering the stylus and the top screws are used for
final tightening of the probe. All screws should be tight when the process is completed.
Renishaw recommends a maximum .049 lbf for tightening.
Once the part probe is indicated in, you are ready to perform part probe operations.
Typical part probe parameters access via : Input / Part Setup / Part Probing / Part
Probe Parameters
Select the following: Input console button, Part Setup, Part Probing, Part Probe
Deflection offsets, Ring Gage Method.
0.0006
Note that the part probe will only need to be
recalibrated if the stylus is changed.
When you probe the Z location of the part you will be prompted to retain that
value as the Part Z Offset value. Say Yes to this prompt.
Input Console Button / Part Setup / Part Probing / Part Zero Probe Cycles / Hole
or Circle Pocket.
Input Console Button / Part Setup / Part Probing / Part Zero Probe Cycles / Edge
RETAIN PROBED PART SETUP has 2 selections that specify how run time probing
data is to be handled. The run time probing data always takes effect in the part
program. Retain all will show the new run-time values in part setup and tool lengths
(part zero X, Y, probe Z, skew and zero calibration values) when the program exits. Do
not retain will not retain any new values after the part program is finished.
START COORDINATE REFERENCE lets you specify if the start positions for the
conversational part probing data blocks are either machine reference or part reference
locations. This gives you more flexibility to program automatic part probing cycles.
Summary