Beruflich Dokumente
Kultur Dokumente
of ICU ventilators
Dr. Ananya
Contents
Classification
History
Introduction
Indications
Key terms- compliance , ventilatory work
Components
Control mechanism
Variables
Triggering
Factors to consider in mechanical ventilation
Wave-forms
Classification
According to Robert chatburn
Respiratory Failure
Cardiac Insufficiency
Neurologic dysfunction
LOAD-
The pressure required to deliver the tidal volume is
referred to as the load that the muscles or
ventilator must work against.
load elastic ( volume & inv. Prop t0
compliance)
resistance ( Raw & inspiratory flow)
Equation of motion for respiratory system
Muscle pressure + ventilator pressure =
(volume / compliance)+
(resistance x flow)
It includes pistons
bellows
reducing valves
pneumatic circuits
Piston mechanism
Bellows mechanism
Pneumatic mechanism
Pneumatic circuits- uses pressurized gas as
power source.
these are microprocessor controlled with
solenoid valves.
use programmed algorithms in
microprocessor to open and close solenoid
valves to mimic any flow or pressure wave
pattern.
Control circuit
Its the system that governs the ventilator
drive mechanism or output control valve.
Classified as-
Open circuits- desired output is selected and
venti. achieves it without any further input
from clinician.
Closed circuits- desired output is selected
and venti. Measures a specific parameter
(flow/vol/press) continuously and input is
constantly adjusted to match desired output.
a.k.a SERVO controlled.
Control parameters
Pressure
Volume
Flow
Time
Ventilators deliver gas to the lungs using
positive pressure at a certain rate. The
amount of gas delivered can be limited by
time, pressure or volume. The duration can
be cycled by time, pressure or flow.
If volume is set, pressure varies..if
pressure is set, volume varies..
.according to the compliance...
Mechanical- employs levers or pulleys to
control drive mechanism.
Pneumatic
Fluidic- applies gas flows and pressure to
control direction of other gas flows and to
perform logic functions based on the COANDA
effect.
Electronic- uses resistors and diodes and
integrated circuits to provide control over the
drive mechanism.
Pressure controller
Ventilator controls the trans-respiratory
system pressure .
This trans-respiratory system gradient
determines the depth or volume of
respiration.
Based on this a ventilator can be positive or
negative pressure ventilator.
Volume controller
Volume cycled ventilation delivers a:
set volume;
with a variable Pressure - determined by
resistance, compliance and inspiratory effort
Flow controller
Allows pressure to vary with changes in patient s
compliance and resistance while controlling flow.
This flow is measured by vortex sensors or venturi
pnemotachometers.
Time controller
measures and controls inspiratory and expiratory time.
These ventilators are used in newborns and infants
Inspiratory time is a combination of the inspiratory
flow period and time taken for inspiratory pause. The
following diagram depicts how the addition of an
inspiratory pause extends total inspiratory time.
Normal inspiratory time of a spontaneously breathing healthy adult is approximately 0.
8- 1.2 seconds, with an inspiratory expiratory (I: E) ratio of 1:1.5 to 1:2 2.
Its advantageous to extend the inspiratory time in order to:
improve oxygenation - through the addition of an inspiratory pause; or to
increase tidal volume - in pressure controlled ventilation
Adverse effects of excessively long inspiratory times are haemodynamic compromise,
patient ventilator dysynchrony, and the development of autoPEEP.
Phase variables
A. Trigger .
What causes the breath to begin?
B C
B. Limit
What regulates gas flow during the breath?
A
C. Cycle .
What causes the breath to end?
Phases of ventilator supported breath
inspiration
change from inspiration to expiration
expiration
change from expiration to
inspiration
Types of ventilator breaths-
Mandatory breath
Assisted breath
Spontaneous breath
Trigger variable
Its the variable that determines start of inspiration
Triggering refers to the mechanism through which the
ventilator senses inspiratory effort and delivers gas flow or
a machine breath in concert with the patients inspiratory
effort.
Can use pressure or volume or time or flow as a trigger.
In modern ventilators the demand valve is triggered by
either a fall in pressure (pressure triggered) or a change in
flow (flow triggered).
With pressure triggered a preset pressure sensitivity has to
be achieved before the ventilator delivers fresh gas into the
inspiratory circuit. With flow triggered a preset flow
sensitivity is employed as the trigger mechanism.
Time triggering
Pressure Triggering
Breath is delivered when ventilator senses patients
spontaneous inspiratory effort.
sensitivity refers to the amount of negative pressure the patient
must generate to receive a breath/gas flow.
If the sensitivity is set at 1 cm then the patient must generate 1
cm H2O of negative pressure for the machine to sense the
patient's effort and deliver a breath.
Acceptable range - -1 to -5 cm H2O below patient s baseline
pressure
If the sensitivity is too high the patient's work of breathing will
be unnecessarily increased. It is not a reasonable course of
action to increase the sensitivity to reduce the patient's
respiratory rate as it only increases their work of breathing.
Flow Triggering
The flow triggered system has two preset variables
for triggering, the base flow and flow sensitivity.
The base flow consists of fresh gas that flows
continuously through the circuit. The patients
earliest demand for flow is satisfied by the base flow.
The flow sensitivity is computed as the difference
between the base flow and the exhaled flow
Here delivered flow= base flow- returned flow
Hence the flow sensitivity is the magnitude of the
flow diverted from the exhalation circuit into the
patients lungs. As the subject inhales and the set
flow sensitivity is reached the flow pressure control
algorithm is activated, the proportional valve opens,
and fresh gas is delivered.
Flow trigger
Advantages -
-The time taken for the onset of inspiratory effort to the onset of
inspiratory flow is considerably less.
-decreases the work involved in initiating a breath.
Limit variable
Cycle variable
Defined as the length of one complete breathing
cycle.
Inspiration ends when a specific cycle variable is
reached.
This variable is used as a feedback signal to end
inspiratory flow delivery which then allows
exhalation to start.
Most new ventilators measure flow and use it as a
feedback signal.
So volume becomes a function of flow and time
Volume= flow x inspiratory time
Baseline variable
The variable controlled during expiration
phase.
Mostly its pressure
Basic definitions
Airway Pressures
Peak Inspiratory Pressure (PIP)
Plateau pressures
Positive End Expiratory Pressure (PEEP)
Continuous Positive Airway Pressure (CPAP)
Plateau Pressure-
The plateau pressure is defined as the end inspiratory
pressure during a period of no gas flow. The plateau
pressure reflects lung and chest wall compliance.
As the plateau pressure is the pressure when
there is no flow within the circuit and patient
airways it most closely represents the alveolar
pressure and thus is of considerable significance
as it desirable to limit the pressure that the
alveoli are subjected to.
Excessive pressure may result in extrapulmonary
air (eg pneumothorax) and acute lung injury.
An increase in airways resistance (including ETT
resistance) will result in an increase in PIP.
An increase in resistance will result in a
widening of the difference between PIP and
plateau pressure.
A fall in compliance will elevate both PIP and
plateau pressure.
It is generally believed that end inspiratory
occlusion pressure (ie plateau pressure) is the
best clinically applicable estimate of average
peak alveolar pressure. Although controversial
it has been generally recommended that the
plateau pressure should be limited to 35 cms
H2O.
Mean Airway Pressure-
The mean airway pressure is an average of the
system pressure over the entire ventilatory
period.
Improves ventilation
Inspiratory Time
In most volume cycled ventilators used in the intensive care
environment it is not possible to set the inspiratory time.
The inspiratory time is determined by the peak inspiratory
flow rate, flow waveform and inspiratory pause. Where
inspiratory time is able to be set, flow becomes dependent
on inspiratory time and tidal volume.
The following example illustrate how these parameters effect
inspiratory time.
Ventilator settings
Tidal volume 1000mls
Peak Flow 60 lpm
Flow Waveform square / constant
Insp. Pause 0 secs
The inspiratory time for this patient would be 1 second because
gas is constantly being delivered at a flow rate of 60 lpm, which
equals 1 litre per second. If an inspiratory pause of 0.5 seconds
were applied then the inspiratory time would be increased to 1.5
seconds.
Changing the patients flow waveform from a square to a
decelerating flow waveform, without changing the flow rate, will
result in an increase in inspiratory time, because the flow of gas
is only initially set at 60 lpm and decreases throughout
inspiration
Output waveforms
Graphical representation of the control or
phase variables in relation to time.
presented as pressure
flow waveforms
volume