Sie sind auf Seite 1von 55

MRI

Pulse Sequences
Jerry Allison Ph.D.
Quick Time a nd a
TIFF (U ncompress ed) de compressor
are neede d to se e this pic ture.

1017 pages 2004 2


Outline
I. Spin Echo Imaging
Multiplanar
Multislice
Oblique
II. Inversion Recovery (IR)
III. Gradient Recalled Echo
IV. Three Dimensional (Volume)
Techniques
V. Fast Imaging Techniques
VI. Echoplanar Imaging
Image Contrast

Image contrast in radiography and CT is based upon a


few properties of the tissues or contrast agents
involved:
- physical density (g/cc)
- electron density (electrons/cc)
- atomic number

4
Image Contrast
Contrast in MRI is more complex and depends
on many properties/parameters, which can be
classified into intrinsic properties and
extrinsic parameters. Intrinsic properties
relate directly to the tissue. Extrinsic
parameters relate to the characteristics of the
MR imager and the details of the MRI Pulse
sequence used for imaging.

5
Intrinsic Properties
Proton density
T1 relaxation
T2 relaxation
T2* relaxation
- magnetic susceptibility
Diffusion
Magnetization transfer
-cross relaxation

6
Intrinsic Properties
Chemical Shift
Temperature
Perfusion
Changes in tissue composition (e.g. age)
Viscosity
Physiologic motion
Bulk flow
Blood
CSF

7
Extrinsic Parameters
Magnetic field strength
-static field
-gradient field
Magnetic field homogeneity
Hardware and software parameters
-coil selection
-number of slices acquired
-slice thickness and gap

8
Extrinsic Parameters
Hardware and software parameters
-slice location
-slice orientation
-number of averages or excitations
-RF pulse shape (#sinc lobes)
-RF transmitter bandwidth
-RF receive bandwidth
-pixel size
-matrix size
-field of view
9
Extrinsic Parameters
-acquisition mode ( 2D / 3D )
-artifact suppression
-physiologic triggering / gating
-orientation of phase and frequency
encode gradients

10
Extrinsic Parameters
RF pulse sequences
-inversion recovery
-spin echo
-gradient recalled echo
-fast scan sequences
-echoplanar (single shot techniques)

11
Extrinsic Parameters
Pulse sequence parameters
-repetition time (TR)
-echo time (TE)
-inversion time (TI)
-flip angle ()
-echo train length
Contrast enhancing agents

12
MRI Pulse Sequences
An MRI pulse sequence dramatically
impacts the appearance of an MRI image.

13
Spin Echo Pulse Sequences
T1 weighted PD weighted T2 weighted

TR 510 TR 4500 TR 4500


TE 14 TE 15eff (ETL7) TE 105eff (ETL7)
2min 7sec for 17 slices 2min 39sec for 24 slices 2min 39sec for 24 slices
14
Inversion Recovery Gradient Echo Pulse Sequence

TR 12.1
TE 5.4
15
3min 11sec for 160 slices
MRI Pulse Sequences
More specifically, an MRI pulse sequence is
a sequence of temporal waveforms:
Radiofrequency (RF) pulses
Gradient (magnetic field) pulses
Data acquisiton intervals

16
Here is a pulse-
sequence diagram.
This shows a
timeline for: 1) RF
pulses; 2) gradient
amplitudes for Gx,
Gy, Gz; 3) the
readout (i.e., A/D),
and 4) the signal of
the excited nuclei.

17
Multiplanar Imaging
Axial, sagittal, and coronal images can be
acquired as follows:

Notice that for each plane, the choice of axis for


phase and frequency encoding can vary.
18
MRI Image Weighting
Many MRI images are described as:
Proton density weighted
T1 weighted
T2 weighted
(and T2* weighted)

19
Spin Echo Images
T1 weighted PD weighted T2 weighted

TR 510 TR 4500 TR 4500


TE 14 TE 15eff (ETL7) TE 105eff (ETL7)
2min 7sec for 17 slices 2min 39sec for 24 slices 2min 39sec for 24 slices
20
Proton Density Weighting
Images are (largely) weighted by the mobile
hydrogen content of the tissues (water and
fat).

PD: FAT < WM < GM < CSF


PD images: CSF > GM > WM > Fat

21
Proton Density Weighting
Proton Density
-The nucleus of most hydrogen atoms is a single particle: the proton
-The number of mobile hydrogen nuclei per voxel directly affects the
intensity of the voxel in an MRI image (for all image weightings).
-Proton Density Weighting emphasizes proton density (as opposed to t1, t2 or
T2*)

-Total proton densities


-CSF 0.112 g H/cc
-Grey Matter 0.1058 g H/cc
-White Matter 0.1056 g H/cc
-Fat 0.1 g H/cc
- Protons in lung tissue volume ~ 0.01 g H/cc

So, one of many problems with lung imaging is the low proton density per
volume, leading to very low SNR. 22
Proton Density Weighting

-Although white matter and grey matter have


very similar proton density; they are
differentiated in MRI by their lipid and water
content.
Lipid (g H / cc) Water (g H / cc)
Grey Matter 0.0072 0.0910
White Matter 0.0178 0.0796

23
T1 Weighting
Images demonstrate good contrast between
soft tissue types (because different tissues
have different T1 values).

24
T2 Weighting
Images demonstrate good contrast between
normal tissue and pathology (because many
pathologies have elevated T2 values due
to increased free water content).

25
Approximate T1 and T2 Values for Human Tissue
(37 oC)
T1 at 1.5 T T1 at 0.5 T T2
Tissue
(msec) (msec) (msec)
Skeletal Muscle 870 600 47
Liver 490 323 43
Kidney 650 449 58
Spleen 780 554 62
Fat 260 215 84
Gray Matter 920 656 101
White Matter 790 539 92
Cerebrospinal Fluid >4,000 >4,000 >2,000
T1, T2 Weighting
In images of the head

T1: FAT < WM < GM < CSF


T1 images: FAT > WM > GM > CSF

T2: FAT < WM < GM < CSF


T2 images: CSF > GM > WM > FAT

Careful: CSF or Fat can be suppressed


27
Pulse Sequence Families
Spin Echo: SE
Gradient Echo:
GE
Gradient Recalled Echo (GRE)
Field Echo (FE)
Inversion Recovery: IR
STIR: short tau inversion recovery
Fat suppression
FLAIR: fluid attenuated inversion recovery
Fluid (CSF) suppression

28
Spin Echo Imaging

Easy to control image weighting with SE


T1 weighted
T2 weighted
PD weighted

29
Spin Echo Imaging
The Spin Echo imaging technique has the
advantage that it is not as sensitive to static
inhomogeneity of the magnet and
inhomogeneity caused by magnetic
susceptibility of patient tissue.

30
Spin Echo Imaging

31
Spin Echo Imaging
The pulse sequence must be repeated many times to
produce an MRI image. The time interval between each
execution of the pulse sequence is termed the
Repetition Time (TR).

32
Spin Echo Imaging
The value of the repetition time (TR) and the echo
time (TE) can be varied to control contrast in spin echo
imaging. For example:

TR = 2000 msec TE = 20 msec Proton Density Weighting


TR = 2000 msec TE = 80 msec T2 Weighting
TR = 600 msec TE = 20 msec T1 Weighting

33
Fast Spin Echo Pulse Sequence (FSE)
Turbo Spin Echo (TSE)

Careful: Fat can be excessively 34


bright on FSE images (j-coupling)
Gradient Recalled Echo
Gradient recalled echo techniques have great
versatility. A variety of contrasts can be
produced while imaging rapidly.

GRE techniques include:


GRASS, SPGR, FLASH, FISP, PSIF and many,
many others.

35
36
Gradient Recalled Echo Images

2D-FLASH
TR 25msec MIP
TE 9msec
a = 35o (Maximum Intensity Projection)
37
5.7sec per slice
Gradient Recalled Echo Image

Multi Planar GRASS


mixed T1/T2 weighting
TR 500msec TE 13msec 2NEX a=60o
38
3min 14 sec for 15 slices
Gradient Recalled Echo
Exceptions are:
1. The creation of the echo is accomplished solely by
gradient magnetic fields (no 180o RF pulse).
2. Deposition of RF energy in the patient is lower since
the 180o RF pulses are not used (less heating of patient
tissues).
3. Static inhomogeneity of the magnet and
inhomogeneity caused by magnetic susceptibility of
patient tissue are NOT corrected by gradient recalled
echo techniques.
4. T2 contrast becomes T2* contrast.
39
Gradient Recalled Echo
4. The initial flip angle is frequently chosen to
be less than 90o . The flip angle in gradient
recalled echo techniques is called .

The optimum value of for a particular TR and


tissue having spin lattice relaxation T1 is called
the Ernst angle.

-TR
cos(e) = e T1

40
Gradient Recalled Echo

5. 3D or volume imaging can be accomplished


(resulting in thinner slices).

Vancouver, BC
courtesy of Dr. Rawson

41
Three Dimensional Volume Techniques

3D voxels are isotropic (or nearly isotropic).


The voxels are the same size in all 3 dimensions.
The dimensions of a typical 3D voxel are
1 mm x 1 mm x 1 mm. The acquisition of
isotropic voxels enables the data set to be
reformatted into any oblique plane without
significant loss of resolution using Post
Processing Techniques.

42
Three Dimensional Volume Image

MPRAGE: Magnetization Prepared Rapid Gradient Echo


TR 11.4msec TE 4.2msec a=12o 1.4mm
6min 55sec for 120 slices (168mm slab) 43
Uses Inversion Recovery
Inversion Recovery (IR)
Inversion recovery pulse sequences are useful for:

Suppression of selected tissues (e.g. orbital fat, liver


screening, fatty tumors, CSF)
Creation of heavily T1 weighted images without a
dominant contribution from fat (e.g. brain, liver and
musculoskeletal imaging).

44
Inversion Recovery (IR)
A basic IR spin echo pulse sequence consists of
a 180o inversion pulse, followed by an inversion
time TI, then a 90o RF pulse.

45
Consider two voxels, one of fat and one of H2O

This method of fat suppression is sometimes called short TI


46
inversion recovery or STIR imaging.
Inversion Recovery (IR)

In spin echo inversion recovery imaging


sequences, the 90o pulse is followed by a 180o
pulse in order to produce a spin echo at time TE
following the 90o pulse

47
48
IR Image

STD T2 weighting

vs

FLAIR: fluid attenuated IR (T2 weighted spin echo)


Inversion time: 2.5sec (CSF is suppressed)
TR 10sec TE 119msec (ETL7) 49
3min 49sec for 19 slices
GE MRI Image Annotation

50
GE MRI Image Annotation

51
GE MRI Image Annotation

52
GE MRI Image Annotation

53
GE MRI Image Annotation

54
GE MRI Image Annotation

55