Hemodynamic Monitoring

Riding the Waves in California


Renee Smith DNP, EdD(c), ACNP

1
Purposes
Aid diagnosis of various cardiovascular
disorders

Guide therapies to minimize cardiovascular


dysfunction
Evaluate the patient’s response to the
therapy

2
A & P of the Heart

3
Hemodynamic Monitoring System
• Basic components of the
monitoring system
• Noncompliant pressure tubing
• Transducer
• Amplifier
• Recorder

• Flush system
• Flush solution
• Infusion system
• Pressure bag

4
Swan Ganz in Heart

5
Hemodynamic Monitoring System (cont.)
• Zeroing and Leveling
• To avoid erroneous readings
• Position of the transducer
• The Phlebostatic axis
• If above---falsely low reading
• If below---erroneous elevated readings
• 2-mmHg change in pressure reading
• Zeroing the system
• 3-way stopcock off patient and open to air

6
Hemodynamic Monitoring System (cont.)

7
Swan Ganz Setup

8
Swan in Patient

9
Pulmonary Artery (PA)
Pressure Monitoring

• Pulmonary artery catheter (Swan Ganz is a brand


name) 1970- Dr. Swan and Dr. Ganz
• Early detection, Identification & treatment of life
threatening conditions, e.g Heart Failure, Cardiac
Tamponade
• Evaluate response to treatment e.g drugs,
mechanical support
• Allows assessment of left ventricular function
• Evaluates CO & CI
10
PA Monitoring Indications

• Any deficit or loss of cardiac function, AMI, CHF, Cardiomyopathy


• All types of shock
• Decreased UOP from dehydration, hemorrhage, GI bleeding, burns or
surgery

11
Pulmonary Artery (PA) Pressure Monitoring (cont.)

• Types of Catheters
• Thermodilution Paceport
for pacing probe (no ballon).
• VIP– additional proximal
lumen and port
• Thermodilution
7-7.1/2 French lumen

12
Swan Ports

• Normally has 4 -5 ports each port are color coded


• Proximal (Blue) used to measure central venous pressure/RAP ( &
injection port for cardiac output
• Distal (Yellow) used to measure PAP (pulmonary Artery Pressure
• Balloon (Red) used to measure PAWP (Pulmonary Artery wedge
Pressure.
• Infusion (White) used for fluid infusion

13
Swan Ganz Waveforms

14
Pulmonary Artery (PA) Pressure Monitoring (cont.)

• Right Atrial Pressure (2-6mmHg)


• Right Ventricular Pressure (20-30/0-8)
• Pulmonary Artery Pressure PA Sysolic pressure =Right VSP
• PA systolic pressure, (20-30 mmHg), equal to right ventricular systolic pressure
• PA diastolic pressure, (8-15 mmHg)
• Reflects the resistance of the pulmonary vascular bed and left ventricular end-diastolic
pressure

Important ****

15
Pulmonary Artery (PA) Pressure Monitoring (cont.)

• Pulmonary Artery Wedge Pressure (PAWP)


• Inflating balloon at the catheter tip (1ml of air)
• Occludes forward flow in the branch of the PA
• Reflects LV end-diastolic pressure
• Normal 8-12 mmHg
• If meet resistance do not force the balloon inflation, it may
have ruptured or it is out of place and could cause infarct.
• Use the PAD reading & notify MD during rounds
***Pulmonary Infract due to inflated balloon don’t force to
pump the ballon so take out the pressure form diastolic
pressure.

16
Pulmonary Artery (PA) Pressure Monitoring (cont.)

17
Pulmonary Artery (PA) Pressure Monitoring (cont.)

• Complications
• Pneumothorax
• Infection
• Ventricular dysrhythmias- if catheter slip out of its place an fall to far
back in the R ventricles cause ventricular Dysthymias.

• Pulmonary artery rupture or perforation-

• Pulmonary artery infarct-

• Embolism -

18
Cardiac Output
Review concepts related to Cardiac Function
• Cardiac output (CO) 48 L per Min if no exercise if exercise
• The amount of blood, in liters, ejected from the left ventricle each minute
• CO = HR x SV ( Heart rate x stroke volume)
• Normal CO for an adult: 4-8 L/min.

• Cardiac Index (CI) **more accurate for the pt than CO


• The amount of blood, in liters, ejected each minute from the left
ventricle per square meter of body surface area
• 2.8 to 4.2 L/min/m2, average 3.0 ± 0.2L/min/m2
• CI = CO (liters/min) ÷ body surface area (m2)

19
Cardiac Output
• Stroke Volume
• Amount of blood ejected by the ventricle during systole, 60-100 ml/beat
•Three factors
• 1. Contractility-
• 2. Preload Fill
• Starling’s Law of the heart
• 3. Afterload push
• systemic vascular resistance
• pulmonary vascular resistance

20
Cardiac Output cont.) Not important
• 10 ml of normal saline or D5W
• Room temperature (or Iced solution-old)
• A closed system: 10-ml syringe attached to the RA port by a stopcock
• Quick and smoothly inject in less than 4 seconds
• Three or more consecutive measurements
• See schematic representation (slope)

21
Central Venous Pressure (CVP)

• Reflects the pressure of blood in the right atrium or vena cava


• Provides information about intravascular blood volume, right
ventricular end-diastolic pressure, and right ventricular
function
• Water manometer -
• Pressure Transducer (hemodynamic monitoring system)
• Acceptable pressure: 5-15 cm H2O

22
Arterial Pressure Monitoring
• Uses
• Continuous monitoring of blood pressure
• Provides access for obtaining blood samples, relatively
closed setup, needleless
• Common sites
• Radial artery --- Allen’s test before putting radial arterial catheter.
Releasing and looking for circulation in that hand 7 second (ulnar
and radial arterial hold at the same time for 7 second and release
then check hand for flush of blood to make sure it’s blood
circulation).
Allen’s test: to test collateral blood flow, compress radial &
ulnar artery, while patient has hand elevated
opening/closing fist, then ulnar is released, should see
good color return to hand w/in 10 secs.
• Femoral artery
• Brachial artery

23
Arterial Line Monitoring

24
Arterial Line Setup

25
26
Arterial Monitor Reading

27
Arterial Pressure Monitoring (cont.)

• Waveform
• Rapid upstroke
• Ejection from the ventricle into the aorta
• Peak = systolic pressure
• Clear dicrotic notch
• Sudden closure of the aortic valve
• The end of ventricular systole or the beginning of diastole
• Definite end-diastole
• The lowest point

28
Arterial Pressure Monitoring (cont.)

• Mean Arterial Pressure (MAP) use opposite arm check MAP and
compare with arterial line.
• Evaluate perfusion of vital body organs
• The average pressure in the peripheral arterial system during the entire
cardiac cycle
• Normal: 70 to 105 mmHg
• Need to maintain more than 60 mmHg to perfuse vital organs automatically
calculated & continuously displayed
• [Systolic BP + (2 x Diastolic BP)] ÷ 3

29
Arterial Pressure Monitoring (cont.)
• Complications
• Infection
• Accidental blood loss
• Impaired circulation to extremity
• Embolism
• Nursing Considerations
• Ensure reading obtained are accurate (AL & SG)
• Level transducer and zeroing system
• Catheter patency (blood clot in tip, kinking of tubing, air bubbles)
• Long tubing (>4ft)
• Too many stopcocks, loose or open stopcocks
• Compare to manual cuff pressure, proper size cuff, proper placement
• Dampened waveform? Check stopcocks, tighten system, flush line, aspirate for air
or clots, change patient position, pt cough
• Monitor complications

30