Sie sind auf Seite 1von 42

X-ray Tube Ratings

Prof. J.K Tonui, PhD

School of Medicine,
Department of Radiology & Imaging
Learning Outcomes
 At the end of this lecture, the student should be able to:

 Describe the factors affecting heat rating of an x-ray tube.

 Calculate the heat rating for 1 and 3 x-ray tubes both in HU

and SI Units.
 Interpret the x-ray tube rating or heat loading charts.

 Understand KW ratings of x-ray tubes.

 Understand various causes of x-ray tube damage and

precautionary measures.
2 RIB 810, 811 & 812 Imaging Physics
Introduction
 X-ray tube produces two forms of energy:

 X-ray (1%) and Heat (99%).

 Heat produced is undesired, hence

 Needs to be disposed off so as not to damage the x-ray tube, and

 Every tube has max. heat dissipation (called heat loading) that

should not be exceeded, because


 Exceeding it will definitely damage it and renders it useless.

3 RIB 810, 811 & 812 Imaging Physics


Tube-Rating
 Heat load on x-ray tube

 Is heat produced during exposure, and

 Total heat produced depends on 3-factos:


1. Kilovoltage (kVp)
2. Milliamperes (mA)
3. Exposure time (s)

 Note that

 The max. temp. a tungsten filament can be raised without dangers of


evaporation is 3000 oC.

4 RIB 810, 811 & 812 Imaging Physics


Heat
 The rate (power) of heat production in an x-ray tube is given
by:
1 W  1 V 1 A

 Hence,
For x-rays, the rate of heat produced is still given by

1 W  1000 V 
1
A OR 1 W  kV  mA
1000

5 RIB 810, 811 & 812 Imaging Physics


Power supply to x-ray
 X-ray tube can be operated using either:

 Single-phase ac source, or

 Three-phase ac source.

 The single-phase ac source

 Uses rms (root mean square) values

 Three-phase ac source

 Uses peak values.

6 RIB 810, 811 & 812 Imaging Physics


Measurement of Heat
 Two units are used to measure heat in x-ray tubes:

 Artificial unit called Heat Unit (HU) and

 SI Units – either Watt-second (Ws) or Joules (J).

 In both cases, the heat produced is the product of:

 Kilovoltage (kV),

 Milliamperes (mA) and

 Exposure time (s).

7 RIB 810, 811 & 812 Imaging Physics


Heat Unit
 Heat Unit (HU)

 A unit of energy used to specify heat loading in x-rays tubes.

 Definitions:

 For Single Phase:


o (Kilovoltage) x (tube current) x (exposure time)

o kVp x mA x s

 For Three Phase (constant potential/high frequency)

o (Kilovoltage) x (tube current) x (exposure time)

o 1.35 x kVp x mA x sec

8 RIB 810, 811 & 812 Imaging Physics


Heat Units:

Technique: Single
Calculation of heat:
Phase Exposure
HU

 70 kVp • 70 X 200 X 0.1


 200 mA
 0.1 second units
?
= 1400 heat

9 RIB 810, 811 & 812 Imaging Physics


Heat Units
Technique: Calculation Heat
Three Phase Exposure In HU

 70 kVp = 70 x 200 x 0.1 x


1.35 = 1890 heat
 200 mA
 0.1 second
units
?
10 RIB 810, 811 & 812 Imaging Physics
Heat in Joules
 The Joule, J,

 Is the SI units of energy and

 Is equal to 1 W x 1 s.

 To calculate heat in Joules, we use

 The rms value for single phase and

 The peak values for three-phase.

12 RIB 810, 811 & 812 Imaging Physics


X-ray tube rating
 The rating of an x-ray tube is given

In terms of kilowatt (kW), and

Is a measure of the ability of the x-ray tube to make a

single exposure in a reasonable exposure duration, where


The reasonable exposure time is usually 0.1 s.

13 RIB 810, 811 & 812 Imaging Physics


Tube Rating Chart
 The tube rating chart

 Indicates the load limit that the tube can safely accept, i.e.

the maximum heat.

 The safe limit is based upon

 Tube construction and

 The kVp waveform (rectifier)

14 RIB 810, 811 & 812 Imaging Physics


Tube, Target, & Ratings
 Surface area bombarded by electrons depends on:

 The actual focal spot size;

 The target angle, and

 The anode diameter.

 The amount of heat produced in the anode depends on:

 Melting point of anode;

 Heat transfer of anode;

 Anode rotation speed and

 Length of exposure.

15 RIB 810, 811 & 812 Imaging Physics


Tube Rating Charts
 The x-ray tube rating charts are

available for:
 Single exposure and

 Multiple rapid exposure

(angiographic) capability

16 RIB 810, 811 & 812 Imaging Physics


Single Exposure Rating Charts
 Single exposure rating charts are

 Designed for a specific x-ray tube

 Every x-ray generator has its own chart:

 That shows allowed single exposure.

 Better ratings

 Means more heat allowed and

 Are associated with large focal spot and high

speed anode rotation.

17 RIB 810, 811 & 812 Imaging Physics


Typical Single-Exposure Tube Rating
Chart
• Fig. shows maximum exposure time for single exposure
at a given kV & mA

18 RIB 810, 811 & 812 Imaging Physics


Example
• What is the maximum exposure time at 90 kVp & 300
mA?

19 RIB 810, 811 & 812 Imaging Physics


Example
• What is the maximum exposure time at 120 kVp &
400 mA?

Can’t do 120 kVp at


400 mA for any
exposure time.
?

20 RIB 810, 811 & 812 Imaging Physics


Single Exposure Rating Charts
 Actually 8 charts combining: 1F 3F
 generator SFS SFS
3400 RPM 3400 RPM
o single phase (1F)
o three phase (3F) 1F 3F
LFS LFS
 focal spot 3400 RPM 3400 RPM
o small
o large 1F 3F
SFS SFS
 anode speed 9600 RPM 9600 RPM
o standard (3400 rpm)
o high (9600 rpm) 1F 3F
LFS LFS
9600 RPM 9600 RPM
21 RIB 810, 811 & 812 Imaging Physics
On-Board Tube Rating Charts
 Checks to see if legal exposure at low-speed rotation and
 Automatically switches to high speed anode rotation as needed
or
 Locks out illegal exposures

Safe at No Safe at No
3400
No Exposure
9600
rpm? rpm?
Allowed

Yes Yes
Allow 3400 Allow 9600
rpm rpm
Exposure Exposure

22 RIB 810, 811 & 812 Imaging Physics


Kilowatt Rating
 Kilowatt rating

 Gives ability of x-ray tube to make single exposure of reasonable

duration (usually 0.1 s)

 The rating can be determined from the tube rating chart using

the following standard assumptions:


 Use 0.1 s exposure time;

 Three phase chart, and

 High speed rotor rotation.


23 RIB 810, 811 & 812 Imaging Physics
Kilowatt Rating (cont…)
 Units
 1 watt = 1 volt x 1 amp

 1 watt = 1 kilovolt x 1 mA

 1 kilowatt (kW) = 1 kilovolt x 1 mA / 1000

 kW rating for a standard 0.1 sec exposure

 kW rating = kVp X mA / 1000

 Use max. mA at given kVp @ .1 sec

24 RIB 810, 811 & 812 Imaging Physics


Kilowatt Rating (cont…)
 To find kW rating,

 Use max. mA at 100

kVp for 0.1 s, hence


X
 kW rating = mA / 10

 Note that
~32 kW
 Each focal spot has its

own kW rating

25 RIB 810, 811 & 812 Imaging Physics


Angiographic Rating Chart
 Provides max. heat units per exposure for

given # of
 Exposures per second and

 The total exposures


Total # of Exposures

2 5 10 20 30
Exposures 1 37,000 24,000 16,000 10,000 7400
per 2 25,000 17,000 12,200 8,000 6,200
second
3 19,000 13,600 10,000 7,000 5,300
4 15,500 11,400 8,600 6,000 4,500
Maximum Load in Peak kV X mA X sec.

26 RIB 810, 811 & 812 Imaging Physics


Example
 How many total exposures can be done at
 90 kVp;
 100 mAs and 90 X 100 = 9000 (Maximum Load)

 3 frames / sec.

Total # of Exposures

2 5 10 13 20 30
Exposures 1 37,000 24,000 16,000 10,000 7400
per 2 25,000 17,000 12,200 8,000 6,200
second
3 19,000 13,600 10,000 7,000 5,300
4 15,500 11,400 8,600 6,000 4,500
Maximum Load in Peak kV X mA X sec.

27 RIB 810, 811 & 812 Imaging Physics


Tube Rating Considerations
 The ability of an x-ray tube to withstand multiple

exposures during several hours of heavy use depends upon


 anode storage / cooling curves

 housing storage / cooling curves

 The housing cooling can be improved with

 Use of fans or

 oil / water circulators

28 RIB 810, 811 & 812 Imaging Physics


Processes of X-Ray Generation
 X-rays are produced when high energy electrons are

stopped by a metal target, and


 The x-rays energy is given by:

E  eV
 Where

o e = 1.6 x 10 -19 C and for 1V

o For 1 V, E = 1.6 x 10 -19 J

29 RIB 810, 811 & 812 Imaging Physics


Difference btw kVp & keV
 Note that

 An “eV” is a unit of energy, where

 1 eV = 1.6 x 10 -19 J of energy and

 1 keV = 1000 eV

 Kilo-peak-voltage (kVp)

 Is the peak or maximum voltage applied across the x-ray tube

(i.e. between anode and cathode) and 1 kVp = 1000 V.

30 RIB 810, 811 & 812 Imaging Physics


Difference btw kVp & keV
 Note that

 The max. energy that an x-ray

acquires is proportional to the


kVp used, but
 Few of such photons are produce

with majority having lower


energy.

31 RIB 810, 811 & 812 Imaging Physics


Tube Damage

Warning

32 RIB 810, 811 & 812 Imaging Physics


Anode Damage
 When heat capacity of a tube is exceeded, then

 It produces melted spots or potholes on the anode surface, and

 Is usually caused by thermal shock (i.e. high mA on cold anode)

can cause cracks in anode (tube death)

33 RIB 810, 811 & 812 Imaging Physics


Protecting the Anode
 To protect the anode,

 Requires warming up the tube slowly, which

 Eliminates thermal shock from high mA exposures on

cold anode.

 Note that

 Warm-up is needed whenever the tube is cold and

 If done once in the morning is not sufficient if tube

not used for several hours.


34 RIB 810, 811 & 812 Imaging Physics
High Voltage Arcs
 Occurs when electrons move from filament to
tube housing instead of to anode
 Can be caused by filament evaporation
 deposition of filament on glass envelope as result of
o high filament currents
o long filament boost time
• reduce by not holding first trigger longer than needed
 Very short exposure with instantaneously very
high mA
 Generator often drops off line +

35 RIB 810, 811 & 812 Imaging Physics arcing


Tube Insert Damage
 Bearing Damage
 prevents proper rotation of anode
 anode can run too slow
 anode can stop
 results in thermal damage to anode (melted
spots)
 Filament break
 renders one focal spot
completely inoperative

36 RIB 810, 811 & 812 Imaging Physics


Reducing Tube Wear: Lower mA

80 kVp 80 kVp
or 100 mA, 0.5 sec
500 mA, 0.1 sec

 Both exposure are 50 mAs


 Same radiation to image receptor
 Same dose to patient

Don’t
smoke
that tube

37 RIB 810, 811 & 812 Imaging Physics


Reducing Tube Wear: Lower mA
80 kVp 80 kVp
or
500 mA, 0.1 sec 100 mA, 0.5 sec

 Low mA reduces tube wear Don’t


smoke
 filament temperature is lower that tube
 reduces filament evaporation

38 RIB 810, 811 & 812 Imaging Physics


Reducing Tube Wear: Lower mA

80 kVp 80 kVp
or
500 mA, 0.1 sec 100 mA, 0.5 sec

 Use lowest mA (and largest focal spot)


consistent with patient motion
considerations
 Large focal spot allows higher mA to be
Don’t
used smoke
that tube

39 RIB 810, 811 & 812 Imaging Physics


Reducing Tube Wear: Raise kVp
70 kVp 90 kVp
or
100 mAs 40 mAs
 High kVp exposures require less heat units for same film
density
 higher kVp more penetrating
 High kVp also reduces patient exposure
 More penetrating beam
 BUT higher kVp reduces contrast Don’t
smoke
 Use highest kVp consistent with required contrast that tube

40 RIB 810, 811 & 812 Imaging Physics


Reducing Tube Wear
 Reduce use of high speed anode rotation
 use longer times instead of higher kV and/or mA

 High speed rotation greatly increases bearing wear


 generators automatically select high speed for high
combinations of kV & mA
 BUT longer exposure times
o increase exposure time & patient motion
 use lowest mA consistent with patient motion
considerations Don’t
smoke
that tube
41 RIB 810, 811 & 812 Imaging Physics
Reducing Tube Wear
 Reduce first trigger holding time

 Reduces bearing wear

o Reduces tube rotation time

 Reduces filament evaporation

o filament evaporation can lead to tube arcing

 Holding first trigger sometimes necessary

 synchronizing breathing for children


Don’t
RIB 810, 811 & 812 Imaging Physics
smoke
42
that tube
Oil Leaks

 May be accompanied by air bubble


in housing
 Eventually causes high voltage
arcing
 Requires immediate service
attention

43 RIB 810, 811 & 812 Imaging Physics

Das könnte Ihnen auch gefallen