Sie sind auf Seite 1von 25

Service Additional :: Selectra / Flexor XL

1
Unpacking the instrument :: XL

2
Unpacking the instrument :: XL

3
Unpacking the instrument :: XL

4
Installing the instrument :: XL
Monitor Table
Working Table

Glass cover

Utilities Cabinet

5
Installing the instrument :: XL
Reagent Rotor L Sample Needle R
Reagent Needle L Cuvette R
Cuvette L Reagent Needle R
Sample Needle L Reagent Rotor R

Sample Rotor
6
Connection plate XL

7
Connecting the cooling unit E and XL
The cooling unit
of the Selectra E
is placed outside
the instrument.

The XL
Cooling unit
Is placed
Inside the
Instrument.

8
Cooling unit E and XL
• Connect unit, using special tubes
• Remove filling lid
• Fill with 2 liters of cooling liquid (Glycol)
• Fill-up with distilled water until the top is reached
• Close the lid
• OPEN AIR TAP IN FILLING LID
• Never switch on the unit without liquid
• Switch on by means of switches Cooling and Pump
• Temperature adjustment is “password” protected and only
accessable by service engineer.
• The procedure for this adjustment is in the service manual.
9
Cooling Unit E and XL

A=Cooling Unit
B=Cooling Agent Reservoir.
C=Cooling Pump
D=Compressor
E=Selectra
F =Heat Exchanger
G=Fan
H=Reagent rotor

10
Fluid and vacuum system Selectra XL

11
Electronics XL

12
Motion Control Board
Old MCB New MCB

13
Motion Control Board :: settings

14
Adjusting the Level Detection XL
• Capacitive Phase Locked Loop (Sample Selectra E)
– See service manual:
• Turn P1 (CCW) to the extreme left position (needle very insensitive).
• Place an empty and dry sample tube in position 2 of the sample rotor.
• Place empty paediatric adapters on positions 1, 3, E3, 4, W, B and S1 of the sample
rotor.
• Make sure that service menu ‘Sample arm’ is selected.
• Press ‘F1’ to reset the sample arm.
• Press three times ‘’, then ‘’ to turn the sample arm into the sample tube on
position 2.
• Check on the screen that ‘Liquid sense = No’.
• Turn P1 (CW) to the right, until ‘Liquid sense = Yes’.
• Turn P1 (CCW) to the left, until ‘Liquid sense = No’.
• Turn P1 360 (one complete turn) to the left (CCW).
• Check also for some other positions if the liquid sensing works well.

15
Level Detection XL

6001-661: PLL Board (left) - PCB layout

6001-662: PLL Board (right) - PCB layout

Note that the PLL boards used in the Junior, E and XL have different part numbers and are not interchangeable.
16
Daily maintenance (customer)
• Fill water container
– 5ml system liquid on 2 liter distilled water
– 25ml system liquid in 10 liter distilled water
• Empty waste container
• Check cuvet rotor blank
– replace cuvet rotor when needed
• Check printer paper
• Fill HCl-bottle (0.1 mol/l)
• Fill tube with hypochlorite solution (W)
• Fill tube with distilled water (B)
• Check wash arm, mixers and cuvet rotor visually
• Check cooling unit

17
Weekly maintenance (customer)
• Perform needle rinse procedure
– clean the sample and regaent needle with hypochlorite solution
• Check syringes for air bubbles and leakage
– clean or replace syringes if necessary

18
Monthly maintenance (customer)
• Clean water and waste container(s)
– 0.1 mol/l NaOH, afterwards rinse several times with water

19
Half-yearly maintenance (customer)
• Run the clean system procedure
– concentrated hypochlorite solution
• Replace mixer belts
• Replace water filter
• Replace drying block on wash arm.

All necessary parts are in the yearly maintenance kit.

20
Yearly maintenance
Replace the following parts:

21
Two-yearly maintenance

• Replace the following parts:

22
Periodic Maintenance Activities
Step Maintenance activity
1 System cleaning procedure
2 Replacement of reagent and sample probe tubing
3 Replacement of ISE sample tubing (only for optional ISE unit)
4 Revision of vacuum pump
5 Revision of water pump
6 Replacement of mixer belts
7 Pipettor check
8 Replacement of water filter
9 Replacement of drying block
10 Cuvette rotor check
11 Lamp check
12 Photometer check

23
Periodic Maintenance Activities
Step Maintenance activity
13 Wash arm check
14 Bellows pump check
15 Reagent and sample arm check
16 ISE-arm & ISE-unit
17 Reagent rotor check
18 Sample rotor check
19 Opto encoder check
20 Vacuum sensor check
21 Water system check
22 Electronic systems check
23 Dichromate endtest execution

24
Required tooling
• cross screw drivers
• flat screw drivers
• hexagon screw drivers
• vernier caliper
• adjustable wrench
• combination spanner
– vertical movement opto-board = 5.5mm
– sprocket for discs = 7 mm
– XL crate = 15mm and 17mm
– external cooling unit = 28mm
• 2Abs test box
• mano meter (vacuum meter)

25

Das könnte Ihnen auch gefallen