Beruflich Dokumente
Kultur Dokumente
PERFORMANCE
LIQUID
CHROMATOGRAPHY
[HPLC]
Preparation of solvents
•Correct solvent preparation is very important.
• It can save vast amounts of time spent troubleshooting such spurious peaks,
baseline noise, etc…
Quality
•All reagents and solvents should be of the highest quality.
• HPLC grade reagents may cost slightly more than lower grade reagents
•HPLC grade reagents contain no impurities to produce spurious peaks in a
chromatogram baseline.
Buffers
•All buffers should be prepared freshly on the day required.
-to ensures that the buffer pH is unaffected by prolonged storage and
that there is no microbial growth present.
INTRODUCTION
Filtration
•All HPLC solvents should be filtered through a 0.45 µm filter before use.
•This removes any particulate matter that may cause blockages.
•After filtration, the solvents should be stored in a covered reservoir to
prevent contamination with dust, etc…
•Pump plungers, seals and check valves will perform better and lifetimes will
be maximized.
Degassing
•Before the freshly prepare mobile phase is pumped around the HPLC
system, it should be thoroughly degassed to remove all dissolved gasses.
•Dissolved gas can be removed from solution by:
• Bubbling with helium
• Sonication
• Vacuum filtration
INTRODUCTION
•If the mobile phase is not degassed, air bubbles can form in the high-
pressure system resulting in problems with system instability, spurious
baseline peaks, etc.
The most efficient form of degassing is bubbling with helium or another low
solubility gas.
It is recommended that the mobile phase is continually degassed at very low
levels throughout the analysis.
-This will inhibit the re-adsorption of gases over the analysis time.
HPLC INSTRUMENT
HPLC INSTRUMENT
Reservoir Pump
Injector
Column system
Detector
HPLC INSTRUMENT
Degasser - to remove bubbles in the column that will cause band spreading
Isocratic elution system – An elution with a single solvent
Gradient elution system – Two and sometime more solvents system that differ significantly in polarity.
Pump
The role of the pump is to force a liquid (called the mobile phase) through the liquid chromatograph at a
specific flow rate, expressed in milliliters per min (mL /min)
During the chromatographic experiment, a pump can deliver a constant mobile phase composition (isocratic)
or an increasing mobile phase composition (gradient).
The type of pump; the screw driven syringe type, the reciprocating pump, and the penumatic pump
Injector system
The injector serves to introduce the liquid sample into the flow stream of the mobile phase.
Typical sample volumes are 5-to 20-microliters (μL).
The injector must also be able to withstand the high pressures of the liquid system.
An auto sampler is the automatic version for when the user has many samples to analyze or when manual
injection is not practical .
HPLC INSTRUMENT
Pump Module Types
Isocratic pump
•Delivers constant mobile phase composition
•solvent must be pre-mixed;
•lowest cost pump
Gradient pump
•Delivers variable mobile phase composition;
•can be used to mix and deliver an isocratic
mobile phase or a gradient mobile phase
Detector
The most widely detector are based on absorption of ultraviolet or visible radiation
PRINCIPLE
Principle of separati on
•When a mixture of components are introduced into the column ,
various chemical and/or physical interactions take place between
the sample molecules and the particles of the column packing .
NORMAL PHASE
REVERSE PHASE
MECHANISM
●
● SP – polar
●
● MP – non polar or less polar
●
● The least polar compound elutes first
●
● SP – Non polar or less polar
●
● MP - polar
●
● The most polar elutes first
PST351-POLYMER CHARACTERISATION 16
NORMAL PHASE
Silica or alumina possess polar sites that interact with polar molecules.
SP = Polar
(hydrophilic)
(ex: Silica gel or
alumina backbone)
MP = Non-polar
(hydrophobic)
(ex:hexane mixed with
polar solvent
Most polar
elute last
REVERSE PHASE
In this column the packing material is relatively non polar and the solvent
is polar with respect to the sample.
MP = Polar (hydophilic)
(Ex: water, methanol,
acetonitrile etc)
PST351-POLYMER CHARACTERISATION 21
MECHANISM
PST351-POLYMER CHARACTERISATION 22
POLARITY ORDER
Less polar
Most separation are achieved by
matching the polarity of the
Hydrocarbons Olefins
analyte to that ofAromatic
the stationary
hydrocarbons
phase; a mobile phase of
considerably different polarity
is then used.
Sulfides Halides
Most polar
Esters
Nitro
Ethers (aldehyde,
compound
ketone)
Amines alcohols
Carboxylic
Sulfones Amides
acids
How can We Analyze the
Sample?
Carbohydrates
1. fructose
2. Glucose
3. Saccharose
4. Palatinose
5. Trehalulose
6. isomaltose 5
2
3
Zorbax NH2 (4.6 x 250 mm) mAU
4
70/30 Acetonitrile/Water 1
6
1 mL/min Detect=Refractive Index
time
PST351-POLYMER CHARACTERISATION 24
Separations
Separation in based upon
Injector
differential migration between
the stationary and mobile
phases.
Mixer Stationary Phase - the phase
which remains fixed in the
column, e.g. C18, Silica
Detector
Waste
Solvents
Mixer
Pumps
Start Injectiontime
Column
Detector
Solvents
Mixer
Pumps
Start Injectiontime
Column
Detector
Solvents
PST351-POLYMER CHARACTERISATION 27
Injector Chromatogram
Mixer
Pumps
Start Injectiontime
Column
Detector
Solvents
PST351-POLYMER CHARACTERISATION 28
Injector Chromatogram
Mixer
Pumps
Start Injectiontime
Column
Detector
Solvents
PST351-POLYMER CHARACTERISATION 29
Injector Chromatogram
Mixer
Pumps
Start Injectiontime
Column
Detector
Solvents
PST351-POLYMER CHARACTERISATION 30
Injector Chromatogram
Mixer
Pumps
Start Injectiontime
Column
Detector
Solvents
PST351-POLYMER CHARACTERISATION 31
Injector Chromatogram
Mixer
Pumps
Start Injectiontime
Column
Detector
Solvents
PST351-POLYMER CHARACTERISATION 32
Injector Chromatogram
Mixer
Pumps
Start Injectiontime
Column
Detector
Solvents
PST351-POLYMER CHARACTERISATION 33
Injector Chromatogram
Mixer
Pumps
Start Injectiontime
Column
Detector
Solvents
PST351-POLYMER CHARACTERISATION 34
Injector Chromatogram
Mixer
Pumps
Start Injectiontime
Column
Detector
Solvents
PST351-POLYMER CHARACTERISATION 35
Injector Chromatogram
Mixer
Pumps
Start Injectiontime
Column
Detector
Solvents
PST351-POLYMER CHARACTERISATION 36
Injector Chromatogram
Mixer
Pumps
Start Injectiontime
Column
Detector
Solvents
PST351-POLYMER CHARACTERISATION 37
Injector Chromatogram
Mixer
Pumps
Start Injectiontime
Column
Detector
Solvents
PST351-POLYMER CHARACTERISATION 38
Injector Chromatogram
Mixer
Pumps
Start Injectiontime
Column
Detector
Solvents
PST351-POLYMER CHARACTERISATION 39
Injector Chromatogram
Mixer
Pumps
Start Injectiontime
Column
Detector
Solvents
PST351-POLYMER CHARACTERISATION 40
Injector Chromatogram
Mixer
Pumps
Start Injectiontime
Column
Detector
Solvents
PST351-POLYMER CHARACTERISATION 41
CHROMATOGR
AM
tR
mAU Area or height is proportional
to the quantity of analyte.
to
Injection time
PST351-POLYMER CHARACTERISATION 42
Modes of High Performance
Liquid Chromatography
PST351-POLYMER CHARACTERISATION 43
APPLICATIONS
Bioscience
Chemical proteins
peptides
polystyre nucleotides
nes
dyes
phthalate
s
tetracyclines
corticosteroids
Pharmaceuticals
antidepressants
barbiturates Consumer Products
lipids
antioxidants
sugars
Environmental
polyaromatic hydrocarbons Clinical
Inorganic ions amino acids
herbicides vitamins
homocysteine
PST351-POLYMER CHARACTERISATION 44
THE CHROMATOGRAM
tR
tR
mAU Area or height is
proportional
to the quantity of analyte.
to
Injection time
APPLICATIONS