Beruflich Dokumente
Kultur Dokumente
October 2010
Agenda
Why nucleonic measurement? The nucleonic measuring principle Typical industries and markets Application examples Don't be afraid of gamma radiation PROTRAC: typically VEGA plics the idea with a future Configuration and parameter adjustment Source container shipping
Level switching (1) Continuous level (2) Density and concentration as well as flow measurement of liquids (3) Interface measurement (4)
plics Electronics
4 20 mA HART Features
Profibus PA
Foundation Fieldbus
Relay output Analogue and digital inputs Serial multi-gauge communication bus Qualified up to SIL2 Simple electronics exchange
plics Housing
Aluminium
Stainless steel
PLICSCOM
VEGACONNECT
Simple navigation
Clear menu structures Tree structures Unambiguous terms
NEW
PVT scintillator
NaI crystal
Figure: VEGASOURCE 35
Radiation-based process instrumentation Page 28 31 August 2011 Roland Bonath
Figure: VEGASOURCE 31
Equivalent dose
The equivalent dose defines the energy dose taken in by the body through ionising radiation Measurement unit for the dose rate is the microsievert/hour (Sv/h) The sievert (Sv) unit is a measure of the biological effects of radiation The largest exposure results from a medical x-ray examination The dose rate a radiation source generates depends on the activity and the distance
Mount Everest
Scintillation detector
1.0
Zugspitze
Venice
Mounting instructions
Mounting instructions
Mounting position for cont. level
The source container must be mounted at the height of, or slightly above, the max. level The radiation must be aligned exactly with the detector The source container and detector should be mounted as close as possible to the vessel to avoid control zones An gap between source container and vessel must be blocked off and marked. In general two or more source containers are used for large measuring ranges
Mounting instructions
Mounting position for cont. level
The source container must be mounted at the height of, or slightly above, the max. level The radiation must be aligned exactly with the detector The source container and detector should be mounted as close as possible to the vessel to avoid control zones An gap between source container and vessel must be blocked off and marked. In general two or more source containers are used for large measuring ranges
Mounting instructions
Mounting position for cont. level
The source container must be mounted at the height of, or slightly above, the max. level The radiation must be aligned exactly with the detector The source container and detector should be mounted as close as possible to the vessel to avoid control zones An gap between source container and vessel must be blocked off and marked. In general two or more source containers are used for large measuring ranges
Mounting instructions
Mounting position for point level
The source container version with an collimation angle = 5 is recommended for point level If larger collimation angle (20 or 40) are used, ensure that the ray is horizontal For this purpose, mount the radiation source container in such a way that the eyelet is positioned horizontally.
Mounting instructions
Mounting position for density
The most constant conditions for density measurement in pipes are achieved if the unit is mounted on vertical pipes If only horizontal pipes are accessible, the path of the ray should also be arranged horizontally to reduce the influence of air bubbles and build-up. In order to achieve a longer path of the radiation through the medium and thus a larger measuring effect
Mounting accessoires
Setup / Adjustment
Standard communication protocols are available for adjustment and connection in process automation
4 20 mA /HART Profibus PA and Foundation Fieldbus
VEGAMET
All settings can be made with one adjustment software (DTM or EDD) Simple setup is also possible via the indicating and adjustment unit PLICSCOM
EDD
EDD
Function control
Standardized status messages acc. to NE 107 or VDI/VDE 2650 Detailed diagnosis in the instrument with cleartext display Digital transmission via HART, Profibus and Foundation Fieldbus
Out of specification
Preventive maintenance
Early recognition of malfunctions Diagnosis evaluation with PACTware 4.0 Simplified servicing
The background calibration is necessary in order to register the natural background radiation The pulse rate of this background radiation is automatically subtracted from any other measured pulse rate
The radiation is switched on Most of the Gammaphotones will be blocked by the liquid The pulse rate is low Background pulse rate of the background radiation is automatically subtracted
The radiation is switched on Within the measuring range, the vessel is emptied as far as possible Ideally: 0%, maximum 40% Background pulse rate of the background radiation is automatically subtracted
The amount of pulse rate depends on the filling level An low number of pulses means an high filling level of the vessel The measured pulse rate is inverse proportional to the filling level An detailed relation between filling level and pulse rates is defined by the Linearization curve
The absorption coefficient of the measured material The reference pulse rate Io
sample 1
For many applications an one-point calibration is sufficient This calibration point should be located as near as possible to the operating point If high accuracy is required over the complete measuring range two or more calibration points should be applied.
sample 1
DTM
Radiation-based process instrumentation Page 48 31 August 2011 Roland Bonath
1491
sample 2
sample 2
DTM
Radiation-based process instrumentation Page 49 31 August 2011 Roland Bonath
1491
sample 1
sample 2
I = 10 * e - * roh * d
Radiation-based process instrumentation Page 50 31 August 2011 Roland Bonath
LIVE Demo
Power supply Power supply 4 20 mA active output (only explosion protected devices) 4 20 mA passive output (only explosion protected devices) 4 20 mA input Switching input open collector Switching input relay contact Switching output (NPN transistor Multisensor communication bus 2x address switches for MGC
6. 7. 8. 9. 10.
4 20 mA output option (IS safe devices) PLICSCOM connection Connections for external indication (VEGADIS) Ground connection
4.
Lesson:
Setup the gauge Build an linearization curve
1. Serial number VEGASOURCE container 2. Order code 3. Document ID 4. Radiation angle 5. Local dose rate at a defined distance from the surface
6. Wording "Hochradioaktive Strahlenquelle" (highly radioactive source), depending on activity 7. Isotope (Cs-137 or Co-60) 8. Serial number of the source capsule 9. Source activity in Bq and mCi 10. Manufacturing date of the source capsule
Editorial
Field service training radiation-based process instrumentation Version: EN 18 October 2010
VEGA Grieshaber KG Am Hohenstein 113 77761 Schiltach Germany Phone +49 7836 50-0 Fax +49 7836 50-201 E-mail info@de.vega.com www.vega.com