UV ABSORPTION METHOD
Interference's
Water vapor may also interfere with the UV method when water vapor concentrations are high and variable. These interference's appear to be due to the condensation of water vapor on imperfect absorption cell windows and Ozone scrubber.
In tropical area we recommend the installation of heated Ozone Scrubber to reduce interference's from water vapour.
ECRD Publications
EC9810 Principle
Sample Inlet Particulate Filter Microprocessor O3 Output
Detector
UV source
EC9810 Operation
The EC9810 ozone monitor is a US EPA equivalent method (U.V.) photometer. This measurement principal is based on the ozones characteristics of absorbing U.V.light at a wavelength of 253.7 nm. Two measurements of a U.V. lamps light intensities are taken in a measurement cycle. In the first, an air sample is passed directly to the sample cell, the second is directed through the ozone scrubber. The scrubber function is to selectively remove the ozone from the sample for the reference portion of the measurement cycle,leaving all other potential interferences intact. These interferences are then compensated by the differential process of subtracting the measurement cycle from the reference cycle to produce an ozone measurement.
ECRD Publications
ECRD Publications
ECRD Publications
EC9810 Specifications
Ozone Range: Resolution: Precision : Zero drift 30 day: Span drift 30 day: Temperature range: EPA designated range: Sample flow: 0 - 20 ppm 0.001 ppm 0.001 ppm 0.001 ppm 0.002 ppm/C0 or 0.5% 5 - 40 C0 0 - 50 to 0 - 1000 ppb 0 - 0.5 to 0 - 0.75 slpm