CT/e, ProSpeed AI/FI Series


CT/e Dual, ProSpeed AII/FII Series
User Guide of SSP Phantom

Copyright © 2003 by General Electric Company

Operating Documentation
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE MEDICAL SYSTEMS User Guide of SSP Phantom
REV 1 2404266

LIST OF EFFECTIVE PAGES

Tab - Section Page REV (Table of Contents) i 1

Title page 1 1 (Introduction) 1-1 0


-
Title page rear blank
2 (Material List) 2-1 0
(Warning and other
important information)
a to d 0 3 (Installation Procedure) 3-1 to 3-14 1

(Revision Information) A 1
- Blank/Rear cover -

REVISION HISTORY
REV Date Primary Reason for Change
0 02/19/04 Initial release.

1 06/02/04 Added thickness measurement. (Sec. 3)


Added contents of Linux system in 3-2-4 Measure HMW value, Calculate Thickness and Record. (Sec. 3)

A
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE MEDICAL SYSTEMS User Guide of SSP Phantom
blank

B
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE MEDICAL SYSTEMS User Guide of SSP Phantom
REV 0 2404266

• THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.


• IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER
THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE
WARNING TRANSLATION SERVICES.
• DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE
MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
• FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE
PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK,
MECHANICAL OR OTHER HAZARDS.

• CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS.


• SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRE
LANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’IL INCOMBE DE LE FAIRE
AVERTISSEMENT TRADUIRE.
• NE PAS TENTER D’INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LE
MANUEL SERVICE N’A PAS ÉTÉ CONSULTÉ ET COMPRIS.
• LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE
TECHNICIEN, L’OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES
DANGERS ÉLECTRIQUES, MÉCANIQUES OU AUTRES.

• DIESES KUNDENDIENST–HANDBUCH EXISTIERT NUR IN ENGLISCHER


SPRACHE.
• FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENÖTIGT,
WARNUNG IST ES AUFGABE DES KUNDEN FÜR EINE ENTSPRECHENDE ÜBERSETZUNG
ZU SORGEN.
• VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR DIESES
KUNDENDIENST–HANDBUCH ZU RATE GEZOGEN UND VERSTANDEN
WURDE.
• WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN
DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN
DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE
GEFAHREN KOMMEN.

• ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS.


• SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA
QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN
AVISO SERVICIO DE TRADUCCIÓN.
• NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO, SIN HABER
CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
• LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL
PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN
LESIONES PROVOCADAS POR CAUSAS ELÉCTRICAS, MECÁNICAS O DE
OTRA NATURALEZA.

a
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE MEDICAL SYSTEMS User Guide of SSP Phantom
REV 0 2404266

• ESTE MANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRA DISPONÍVEL


EM INGLÊS.
• SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A
ATENÇÃO GEMS, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇ O.
• NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E
COMPREENDIDO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA.
• O NÃO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANÇA
DO TÉCNICO, OPERADOR OU PACIENTE DEVIDO A‘ CHOQUES ELÉTRICOS,
MECÂNICOS OU OUTROS.

• IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE SOLTANTO IN


INGLESE.
• SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEMS RICHIEDE IL
AVVERTENZA MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE È TENUTO A PROVVEDERE
DIRETTAMENTE ALLA TRADUZIONE.
• SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO DOPO
AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL
CONTENUTO.
• NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR
COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO ALLA
MANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PER
FOLGORAZIONE ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.

b
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE MEDICAL SYSTEMS User Guide of SSP Phantom
REV 0 2404266

IMPORTANT! . . . X-RAY PROTECTION


X–ray equipment if not properly used may cause injury. Accordingly, the instructions herein contained
should be thoroughly read and understood by everyone who will use the equipment before you attempt to
place this equipment in operation. The General Electric Company, Medical Systems Group, will be glad to
assist and cooperate in placing this equipment in use.
Although this apparatus incorporates a high degree of protection against x–radiation other than the useful beam,
no practical design of equipment can provide complete protection. Nor can any practical design compel the opera-
tor to take adequate precautions to prevent the possibility of any persons carelessly exposing themselves or others
to radiation.
It is important that everyone having anything to do with x–radiation be properly trained and fully acquainted with the
recommendations of the National Council on Radiation Protection and Measurements as published in NCRP
Reports available from NCRP Publications, 7910 Woodmont Avenue, Room 1016, Bethesda, Maryland 20814, and
of the International Commission on Radiation Protection, and take adequate steps to protect against injury.

The equipment is sold with the understanding that the General Electric Company, Medical Systems Group, its
agents, and representatives have no responsibility for injury or damage which may result from improper use of the
equipment.

Various protective material and devices are available. It is urged that such materials or devices be used.

CERTIFIED ELECTRICAL CONTRACTOR STATEMENT

All electrical installations that are preliminary to positioning of the equipment at the site prepared for the equipment
shall be performed by licensed electrical contractors. In addition, electrical feeds into the Power Distribution Unit
shall be performed by licensed electrical contractors. Other connections between pieces of electrical equipment,
calibrations, and testing shall be performed by qualified GE Medical personnel. The products involved (and the
accompanying electrical installations) are highly sophisticated, and special engineering competence is required.

In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s
electrical work on these products will comply with the requirements of the applicable electrical codes.

The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third–
party service companies with equivalent training, or licensed electricians) to perform electrical servicing on the
equipment.

c
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE MEDICAL SYSTEMS User Guide of SSP Phantom
REV 0 2404266

DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent, have notation “damage in
shipment” written on all copies of the freight or express bill before delivery is accepted or “signed for” by a Gen-
eral Electric representative or a hospital receiving agent. Whether noted or concealed, damage MUST be reported
to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and con-
tainers held for inspection by the carrier. A transportation company will not pay a claim for damage if an inspection
is not requested within this 14 day period.

Call Traffic and Transportation, Milwaukee, WI (414) 827-3449 / 8*285-3449 immediately after damage is found.
At this time be ready to supply name of carrier, delivery date, consignee name, freight or express bill number, item
damaged and extent of damage.

Complete instructions regarding claim procedure are found in Section “S” of the Policy & Procedure Bulletins.

OMISSIONS & ERRORS


GE personnel, please use the GEMS CQA Process to report all omissions, errors, and defects in this documenta-
tion. Customers, please contact your GE Sales or Service represenatives.

CAUTION
Do not use the following devices near this equipment. Use of these devices near this equipment could cause
this equipment to malfunction.

Devices not to be used near this equipment:

Devices which intrinsically transmit radio waves such as; cellular phone, radio transceiver, mobile radio transmit-
ter, radio-controlled toy, etc.

Keep power to these devices turned off when near this equipment.

Medical staff in charge of this equipment is required to instruct technicians, patients and other people who may be
around this equipment to fully comply with the above regulation.

d
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE MEDICAL SYSTEMS User Guide of SSP Phantom
REV 1 2404266

TABLE OF CONTENTS

SECTION PAGE

SECTION 1 - INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1

SECTION 2 - MATERIAL LIST. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1

SECTION 3 - INSTALLATION PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1


3-1 PREPARATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3-2 INSTALLATION PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3-2-1 SSP phantom centering (Refer to Illustration 3-2) . . . . . . . . . . . . . . . . . . . . . 3-2
3-2-2 Get The Scout and Axial Images . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3-2-3 Calculate coefficient value “K” of your SSP Phantom . . . . . . . . . . . . . . . . . . 3-6
3-2-4 Measure HMW value, Calculate Thickness and Record . . . . . . . . . . . . . . . . 3-7
3-2-4-1 For CT of IRIX system (O2 Computer). . . . . . . . . . . . . . . . . . . . 3-7
3-2-4-2 For CT using Linux system (PC Computer), SSP profile and HWM can
be saved in ImageWorks and printed on the film. . . . . . . . . . . . . . . . . . . . . . 3-10

i
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE MEDICAL SYSTEMS User Guide of SSP Phantom

ii
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE MEDICAL SYSTEMS User Guide of SSP Phantom
REV 0 2404266

SECTION 1 - INTRODUCTION

This option enables CT/e, ProSpeed AI/FI Series and CT/e Dual, ProSpeed AII/FII Series systems to use SSP
Phantom. This manual describes direction.

CAUSE
Calculate slice trim thickness with right-angled triangle with the tungsten wire image which got by scan under defined
scan condition with required thickness.

1-1
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE MEDICAL SYSTEMS User Guide of SSP Phantom
blank

1-2
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE MEDICAL SYSTEMS User Guide of SSP Phantom
REV 0 2404266

SECTION 2 - MATERIAL LIST

Remote Tilt Option Kit (for Linux System)


2402714

ITEM PART NO. DESCRIPTION QTY.


1 2390281 SSP Phantom 1
2 2111577 Rating Plate Film 1
3 2404266 User Guide of SSP Phantom 1

2-1
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE MEDICAL SYSTEMS User Guide of SSP Phantom
blank

2-2
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE MEDICAL SYSTEMS User Guide of SSP Phantom
REV 1 2404266

SECTION 3 - INSTALLATION PROCEDURE

3-1 PREPARATION

3-2 INSTALLATION PROCEDURE

NOTICE
This SSP (Slice Sensitive Profile) phantom is specialized for thickness measurement.

Illustration 3-1 Thickness Measurement

3-1
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE MEDICAL SYSTEMS User Guide of SSP Phantom
REV 1 2404266

3-2-1 SSP phantom centering (Refer to Illustration 3-2)


1. Put SSP phantom on the phantom holder, open positioning light. Move SSP phantom to left or right to set sag-
ittal marker line at the center of SSP.
2. Use “In/Out/Up/Down” buttons on control panel of Gantry to set coronal marker line on the wire hole.

Illustration 3-2 SSP Phantom Centering

coronal marker line


sagittal marker line

3-2
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE MEDICAL SYSTEMS User Guide of SSP Phantom
REV 1 2404266

3-2-2 Get The Scout and Axial Images


1. Click “New Patient”, input Patient ID and “SSP” in Patient Name. Select head protocol to finish a scout scan
and set up six axial scan groups with different thickness for 1, 2, 3, 5, 7, 10mm. On scout image, move axial
localizer to set DFOV center on superposition of the center of Tungsten wire. (Refer to Illustration 3-3 and Il-
lustration 3-4 for the setup of axial groups)

Note: Set WW=30, WL=-80 on scout image to get a clear Tungsten wire display.
For axial groups, select mA=100, DFOV=5cm, 1 image per group and Interval=0.
That means scan the same position of Tungsten wire with 6 different thickness without cradle movement.

Illustration 3-3 Scout and Axial Images (I)

3-3
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE MEDICAL SYSTEMS User Guide of SSP Phantom
REV 1 2404266

3-2-2 Get the scout and axial images (Continued)

Illustration 3-4 Scout and axial images (II)

3-4
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE MEDICAL SYSTEMS User Guide of SSP Phantom
REV 1 2404266

3-2-2 Get the scout and axial image (Continued)


2. Click “Confirm” to finish the six group scan. Enter “ImageWorks” you will get 6 axial images like the following
Illustration 3-5.
Note: WW=300, WL=-800

Illustration 3-5 6 axial images

3-5
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE MEDICAL SYSTEMS User Guide of SSP Phantom
REV 1 2404266

3-2-3 Calculate coefficient value “K” of your SSP Phantom

IMPORTANT NOTE: - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
This step is only needed for your first time to use SSP phantom in order to get the K value.
For afterwards regular measurement every month, you can skip this step and use the K val-
ue directly in thickness calculation, because one SSP phantom will have a fix K value for
same system.
1. In ImageWorks, view scout image on the screen with one image format.
Note: WW=30, WL=-80

2. Click On button, grid scale will overlay on the scout image.


3. Put mouse on grid, press down left key and hold, draw the grid to set +10 scale of horizontal axis just on the
Tungsten wire. Then you get a right-angle triangle ABC and AC=10. (Refer to Illustration 3-6)

Illustration 3-6 Right-angle triangle ABC and AC=10

10 scale point on
Tungsten wire

4. Count the scales of BC. BC=5 in this example.


Note: BC may be some other number in your system, such as BC=4.8, or 5.2, etc.
5. Calculate K value of your SSP phantom by formula: K=BC/AC
In this example, K=5/10=0.5

3-6
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE MEDICAL SYSTEMS User Guide of SSP Phantom
REV 1 2404266

3-2-4 Measure HMW value, Calculate Thickness and Record


Note
For CT of Linux system, skip Step 3-2-4-1 and go directly to Step 3-2-4-2.

3-2-4-1 For CT of IRIX system (O2 Computer).


1. Open Image Profile function:
Click mouse left key by sequence: Service → Image Quality → Image Profile, Image Profile window occurs
on the screen. Click “Browse” button.

Illustration 3-7 window occurs on the screen

3-7
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE MEDICAL SYSTEMS User Guide of SSP Phantom
REV 1 2404266

3-2-4 Measure HMW value, Calculate Thickness and Record (Continued)


2. Select 1mm SSP axial image, click “Accept”.

Illustration 3-8 1mm SSP axial image

3. In Window Settings region, input WW=300, press “Enter” key; Input WL=-800, press “Enter” key. The image
displays on the window.

3-8
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE MEDICAL SYSTEMS User Guide of SSP Phantom
REV 1 2404266

3-2-4 Measure HMW value, Calculate Thickness and Record (Continued)


4. Click “Horizontal” button, then click “FWHM” button, press and hold mouse left key to draw a line overlay on
the Tungsten wire but longer than it at two ends.

Illustration 3-9 Window Settings region_1mm

End
Start

5. A profile occur and a HMW value displays on the bottom-right corner of the image.
For this example, the HMW=2.64mm.
6. Multiply HMW value by K value to get measured thickness, we have got K=0.5 in 3-2-3, Step 5.
2.64x0.5=1.32(mm)
7. Compare measured thickness 1.32mm with selected scan thickness 1mm to get the tolerance.
(1.32-1)/1=0.32=3.2%
8. Repeat steps 2 to 7 to measure other thickness.
For thickness<=2mm, the tolerance standard is <±50%
For thickness>2mm, the tolerance standard is <±1mm

3-9
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE MEDICAL SYSTEMS User Guide of SSP Phantom
REV 1 2404266

3-2-4 Measure HMW value, Calculate Thickness and Record (Continued)


9. Record the measurement value in the following form manually.

Table 3-1 The measurement value

Thickness K value HMW Measured Thickness Tolerance Tolerance Standard (±)


1 0.5 2.64 1.32 32% 50%
2 0.5 4.00 2.00 0% 50%
3 0.5 6.15 3.08 0.08mm <1mm
5 0.5 10.35 5.18 0.18mm <1mm
7 0.5 14.16 7.08 0.08mm <1mm
10 0.5 20.21 10.11 0.11mm <1mm

3-2-4-2 For CT using Linux system (PC Computer), SSP profile and HWM can be saved in ImageWorks
and printed on the film.

NOTICE
For CT/e Linux, software version should be 2.31 or latter.
For CT/e Dual Linux, software version should be 3.61 or latter.
1. Open Image Profile function:
Click mouse left key by sequence: Service → Image Quality → Image Profile, Image Profile window occurs
on the screen.

3-10
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE MEDICAL SYSTEMS User Guide of SSP Phantom
REV 1 2404266

3-2-4 Measure HMW value, Calculate Thickness and Record (Continued)


2. Input K value which you get from 3-2-3, for example, K=0.48, press “Enter” key. Then click "Browse" button.

Illustration 3-10 Input K value

Input K value

3. For example, select 5mm SSP axial image, click “Accept”.

Illustration 3-11 5mm SSP axial image

3-11
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE MEDICAL SYSTEMS User Guide of SSP Phantom
REV 1 2404266

3-2-4 Measure HMW value, Calculate Thickness and Record (Continued)


4. In Window Settings region, input WW=300, press “Enter” key. Input WL=-800, press “Enter” key. The image
displays on the window.
5. Click “Horizontal” button, then click “FWHM” button, press and hold mouse left key to draw a line overlay on
the Tungsten wire but longer than it at two ends.

Illustration 3-12 Window Setting region_5mm

Start End

6. A profile occur and three value displays on the bottom-right corner of the image.
K val: the K value of SSP phantom which you get in 3-2-3 and input in 3-2-4-2 Step 2.
THK: the thickness value, which is got by K val multiply HMW value.
HMW: FWHM value (Full Width Half Maximum).
7. Click “Save Screen” button. Wait for the following prompt window. This SSP profile image will be saved under
the same Examination number with series name “SSAVE”.

8. Repeat Steps 3 to 7 to measure SSP profile of all the other thickness.

3-12
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE MEDICAL SYSTEMS User Guide of SSP Phantom
REV 1 2404266

3-2-4 Measure HMW value, Calculate Thickness and Record (Continued)


9. Open Browser window on ImageWorks, you will find SSP profile image saved under the same Examination
number with series name “SSAVE”, select it .

Illustration 3-13 Save SSP profile image

10. Click “Viewer”, the SSP profile image is displayed on the screen.
Send all of the SSP profile images to “Film Composer” and print them on a film.

Illustration 3-14 SSP profile image

3-13
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE MEDICAL SYSTEMS User Guide of SSP Phantom
REV 1 2404266

3-2-4 Measure HMW value, Calculate Thickness and Record (Continued)

Illustration 3-15 Examples of SSP profile image

3-14