Beruflich Dokumente
Kultur Dokumente
Publications
Direction 2152916100
Revision 3
GE Medical Systems
HiSpeed CT/I Preinstallation
Operating Documentation
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
i
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
ii
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GEs
field engineers, personnel of thirdparty service companies with equivalent
training, or licensed electricians) to perform electrical servicing on the
equipment.
IMPORTANT...XRAY PROTECTION
Xray equipment if not properly used may cause injury. Accordingly, the
instructions herein contained should be thoroughly read and understood by
everyone who will use the equipment before you attempt to place this equipment
in operation. The General Electric Company, Medical Systems Group, will be glad
to assist and cooperate in placing this equipment in use.
The equipment is sold with the understanding that the General Electric Company,
Medical Systems Group, its agents, and representatives have no responsibility
for injury or damage which may result from improper use of the equipment.
Various protective materials and devices are available. It is urged that such
materials or devices be used.
iii
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is
apparent write Damage In Shipment on ALL copies of the freight or express bill
BEFORE delivery is accepted or signed for by a GE representative or hospital
receiving agent. Whether noted or concealed, damage MUST be reported to the
carrier immediately upon discovery, or in any event, within 14 days after receipt,
and the contents and containers held for inspection by the carrier. A
transportation company will not pay a claim for damage if an inspection is not
requested within this 14 day period.
Call Traffic and Transportation, Milwaukee, WI (414) 785 5052 or 8*323 5052
immediately after damage is found. At this time be ready to supply name of
carrier, delivery date, consignee name, freight or express bill number, item
damaged and extent of damage.
iv
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Chapter 1
Preinstallation
Table Of Contents
Section 1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
1.1 Responsibility of Purchaser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
1.2 Site Preparation Before Equipment Delivery . . . . . . . . . . . . . . . . . . . . . 11
1.3 Other Preinstallation Directions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Section 2
System Catalog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
2.1 Base Scanner System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Section 3
Room Planning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.1 Strength of Floors and Walls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.2 Radiation Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3.3 Mounting Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 310
Section 4
Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
4.1 Temperature and Humidity Specifications . . . . . . . . . . . . . . . . . . . . . . . 41
4.2 Temperature and Humidity Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . 41
4.3 Cooling Requirement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
4.4 Altitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
4.5 ElectroMagnetic Interference (EMI) . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Section 5
Floor Loading and Weights . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
5.1 Floor Loading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
5.2 Seismic Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
i.i
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Section 6
Delivery Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
6.1 For Van Delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
6.2 For Crated Delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
6.3 Storage Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
6.4 Gantry Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
6.5 Shipping Dimensions and Weights (Van Delivery) . . . . . . . . . . . . . . . . 62
6.6 Shipping Dimensions and Weights (Crated Delivery) . . . . . . . . . . . . . . 63
Section 7
Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
7.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
7.2 System Input Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
7.2.1 Facility Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
7.2.2 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
7.2.3 Rating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
7.2.4 Regulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
7.2.5 Phase Imbalance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
7.2.6 Sags, Surges & Transients . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
7.2.7 Microcuts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
7.2.8 Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
7.3 Recommended Power Distribution System . . . . . . . . . . . . . . . . . . . . . . 72
7.4 Power Source Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
7.5 Ground System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Section 8
Interconnection data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
8.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Section 9
Preinstallation Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
i.ii
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
List Of Illustrations
Illustration Page
Illustration 11
Base Scanner System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Illustration 12
Recommended Suite Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Illustration 13
Minimum Room Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Illustration 14
HSA Tube Typical Scatter Survey (Body Filter) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Illustration 15
HSA Tube Typical Scatter Survey (Head Filter) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Illustration 16
Optional Performix Tube Typical Scatter Survey (Body Filter) . . . . . . . . . . . . . . . . . . . . . . . . 38
Illustration 17
Optional Performix Tube Typical Scatter Survey (Head Filter) . . . . . . . . . . . . . . . . . . . . . . . . . 39
Illustration 18
Table and Gantry (Top View) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 311
Illustration 19
Table and Gantry (Side View) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 312
Illustration 110
Power Distribution Unit (PM) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 313
Illustration 111
Compact Power Distribution Unit (PM) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 314
Illustration 112
Compact Power Distribution Unit (PM) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 315
Illustration 113
Operators Console / INDIGO Computer (oc1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 316
Illustration 114
Operators Console / OCTANE Computer (OC1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 317
Illustration 115
Gantry (CT2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Illustration 116
Typical Floor Anchor, Gantry and Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Illustration 117
Patient Table (CT1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Illustration 118
Power Distribution Unit (PM) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Illustration 119
Power Distribution Unit (PM) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Illustration 120
Operators Console / INDIGO Computer (OC1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
i.iii
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Illustration Page
Illustration 121
Operators Console / OCTANE Computer (OC1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Illustration 122
Gantry / Table Mounting Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Illustration 123
Gantry Shipping COnfiguration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Illustration 124
System Ground Map . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Illustration 125
Interconnection Runs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Illustration 126
Contractor Components Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Illustration 127
Primary Power Disconnect (A1)
Fusible Disconnect and Magnetic Contactor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Illustration 128
Typical Customer Supplied Scan Room Warning Light Connection To PM . . . . . . . . . . . . . . . 89
Illustration 129
Typical CuStomer Supplied Scan Room Door Interlock Switch Connection To PM . . . . . . . . 89
i.iv
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
List Of Tables
Table Page
Table 11
Dimensions of Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Table 12
Shielding Requirements Scaling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Table 13
Cooling Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Table 14
HiSpeed Advantage System Floor Loading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Table 15
Shipping Dimensions and Weights
(Van Delivery Typical Inside North America)
catalog: B7500JK, B7500JB, B7500JA, B7500JG, B7500JD, B7500JE . . . . . . . . . . . . . . . . . . 62
Table 16
Shipping Dimensions and Weights
(Crated Delivery Outside North America)
Catalogs: B7500JK, B7500JB, B7500JA, B7500JG, B7500JD, B7500JE . . . . . . . . . . . . . . . . 63
Table 17
Gantry Delivery Size Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Table 18
Facility Power Requirement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Table 19
Component Designators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Table 110
Runs, Wiring and Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Table 111
ContractorSupplied Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
i.v
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
NOTES
i.vi
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
REVISION HISTORY
i.vii
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
NOTES
i.viii
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
SECTION 1
INTRODUCTION
This direction contains physical and electrical data necessary for planning and preparing a site.
Preinstallation work is site preparation for installation of the HSA CT/i product. It is the
responsibility of the purchaser to arrange and pay for this work. Preinstallation work includes:
Installation of electrical conduit, junction boxes, ducts, outlets, and line safety switches.
Installation of interconnection wiring that is AWG stranded copper. The electrical
contractor shall ring out and tag all wires at both ends. Colorcoded wires are
recommended for easier identification. Wires shall be continuous without splices.
Ground wires must conform to local codes.
Any site renovation.
Alterations and modifications to products not specifically included in the sales contract.
All work will conform to local building and safety codes.
Unless specifically mentioned, GE Medical Systems does not provide or install wires, conduits,
junction boxes, and ducts illustrated in this direction.
All CT site plans, preliminary concepts and final working drawings must be reviewed by
General Electric Headquarters Architectural Planning prior to construction or approval.
1.1
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Refer to M959 Laser Imager XL Site Planning Guide for preinstallation information for the 3M
M959 laser camera (catalog B7920SA) and its options. Information needed for preinstallation
of VEIBs, KEIBs and film processors is also in this book.
Refer to M969 Laser Imager HQ Site Planning Guide for preinstallation information for the 3M
969 laser camera (catalog B7920T, MultiUser and B7920TC, SingleUser Imagers) and their
options. Information needed for preinstallation of VEIBs, KEIBs and film processors is also
in this book.
Refer to Kodak Ektascan 2180 Laser Printer, Site Specifications and Requirements document
for site specification information. Available from Kodak.
Refer to Kodak Ektascan 1180 Laser Printer, Site Specifications and Requirements document
for site specification information. Available from Kodak.
1.2
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
SECTION 2
SYSTEM CATALOG
The following is a complete list of system options requiring site planning work for the HSA CT/i
scanner system, catalogs: B7500JK, B7500JB, B7500JA, B7500JG, B7500JD, B7500JE.
Contact your local GE Medical Systems Sales force for a complete list of all system options.
CATALOG NO. OPTION
B7540PB VCR INTERFACE / SCAN CONVERTER
E7014LA COLOR PRINTER
B7999PS EXIDE UNINTERRUPTIBLE POWER SUPPLY
B7999PT LIEBERT UNINTERRUPTIBLE POWER SUPPLY
B7530RC CT HSA CT/I REMOTE COLOR MONITOR OPTION
B7530RB CT HSA CT/I REMOTE BLACK/WHITE MONITOR OPTION
2.1
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Illustration 11
BASE SCANNER SYSTEM
CT3501A PDU
GANTRY
TABLE
OPERATORS
CONSOLE
CHAIR
2.2
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
SECTION 3
ROOM PLANNING
Illustration 12 is a suggested room layout for the CT components. Overall suite size is 20.3
ft (6.19 m) x 23.2 ft (7.07 m) deep, with a 9 ft (2.75 m) ceiling. The area consists of the following
three rooms:
Scan Room: 19.4 ft x 12.9 ft (3.81 m x 6.10 m)
Control Area: 12.3 ft x 9.0 ft (3.81 m x 2.67 m)
Equipment Room: 6.7 ft x 9.0 ft (8.38 m x 3.20 m)
For equipment sizes refer to Illustration 18 through Illustration 113.
Note 1 Other room arrangements are possible. Use cable lengths shown in Section 8 to plan
maximum distance between components. For example:
The control and equipment rooms may be interchanged with the scanner room.
The preferred location of options such as Advantage Windows, laser camera or line
printer is in a separate room.
See Illustration 13 for a room plan with a minimum space requirement.
3.1
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Illustration 12
RECOMMENDED SUITE LAYOUT
EQUIPMENT ROOM
6.7 X 9.0
(8.38 X 3.20)
LASER CAMERA
OPTION
WALL CONTROL AREA
DUCT 12.3 X 9.0 SLAVE
(3.81 X 2.67) MONITOR
OPTION
INJECTOR
CONTROL OPTION
POWER (ELECTRONICS
MODULE OPERATORS ON SHELF
CONSOLE BELOW )
TV
MONITOR
WALL (NOT BY GE)
DUCT
FLUSH
FLOOR OX/VAC/INJECTORS
DUCT IN CEILING
PATIENT
TABLE
SCAN ROOM
19.4 X 12.9
(3.81 X 6.10)
CT
GANTRY
3.2
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Table 11
Dimensions of Components
3.3
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Illustration 13
MINIMUM ROOM LAYOUT
TV
CAMERA
(NOT BY GE)
POWER
MODULE
CT
GANTRY
OX/VAC/INJECTORS
IN CEILING
SINK
TV PATIENT
MONITOR TABLE
(NOT BY GE)
OPERATORS
CONSOLE
SCAN ROOM
16.0 X 20.0
(4.88 X 6.10)
03 6 9 1 2 3 4 CT3508A
3.4
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Table 12
Shielding Requirements Scaling
CHANGED MULTIPLICATION
PARAMETER FACTOR
mA new mA/340
80 kV 0.29
120 kV 0.71
1 second 0.50
4 seconds 2.00
5 mm 0.80
1 mm 0.40
3.5
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Illustration 14
HSA TUBE TYPICAL SCATTER SURVEY (BODY FILTER)
4 7 9.5 7 4 2 1 .50
.50
ELEVATION VIEW
.35
0 24 in.
(610 mm)
mR PER SCAN
SCALE
.45
.35
.85
.60
1.5
3 1
5 2
5
7
TECHNIQUE:
140 kV
340 mA
10 mm APERATURE
2 SECOND ROTATION
BOWTIE FILTER
THROUGH ISOCENTER
PLAN VIEW
3.6
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Illustration 15
HSA TUBE TYPICAL SCATTER SURVEY (HEAD FILTER)
1.5
.35
ELEVATION VIEW
.20
0 24 in.
(610 mm)
mR PER SCAN
SCALE
.30
.12
.60
1 .25
2
.45
3
.80
5
2
3
TECHNIQUE:
140 kV
340 mA
10 mm APERATURE
2 SECOND ROTATION
BOWTIE FILTER
THROUGH ISOCENTER
PLAN VIEW
3.7
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Illustration 16
OPTIONAL PERFORMIX TUBE TYPICAL SCATTER SURVEY (BODY FILTER)
2.2
.56
ELEVATION VIEW
.39
0 24 in.
(610 mm)
mR PER SCAN
SCALE
.50
.39
.95
.67
1.7
3.4 1.1
5.6 2.2
7.8 5.6
7.8
TECHNIQUE:
140 kV
380 mA
10 mm APERATURE
2 SECOND ROTATION
BOWTIE FILTER
THROUGH ISOCENTER
PLAN VIEW
3.8
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Illustration 17
OPTIONAL PERFORMIX TUBE TYPICAL SCATTER SURVEY (HEAD FILTER)
1.7
.40
ELEVATION VIEW
.22
0 24 in.
(610 mm)
mR PER SCAN
SCALE
.34
.13
.67
1.1 .28
2.2
.50
3.4
.90
5.6
2.2
3.4
TECHNIQUE:
140 kV
380 mA
10 mm APERATURE
2 SECOND ROTATION
BOWTIE FILTER
THROUGH ISOCENTER
PLAN VIEW
3.9
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
3.10
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Illustration 18
TABLE AND GANTRY (TOP VIEW)
CL
SCAN NOTE:
FULL HINGE 25.00 DIMENSIONS ARE IN INCHES
EXTENSION (635.0) (MILLIMETERS)
24.50 23.60
(622.3) (599.4)
FRONT ACCESS NOT TO SCALE
COVER OPENED
ALLOW FOR ADEQUATE
SERVICE CLEARANCE
80.00
(2032.0)
SERVICE CLEARANCE
88.00
(2235.2)
C
L 24.25
(616.0)
TABLE
74.11
(1882.4)
EXTENDER
24.50
REAR GANTRY (622.3)
4.09
COVER (103.9)
80.00 144.00
(2032.0) (3657.6) CT3503A
SERVICE CLEARANCE
3.11
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Illustration 19
TABLE AND GANTRY (SIDE VIEW)
TOP COVER
3.28 OPENING NOTE:
19.65 * 18.26 (83.3) ARC DIMENSIONS ARE IN INCHES
(499.1) (463.8)
GANTRY TILT (MILLIMETERS)
0
0
30 NOT TO SCALE
30 MAX
MAX
INDICATES AIR FLOW
EXTENDER
78.55
(1995.2)
43.78
1112
22.03
(559.6)
17.75
(450.8)
UNDER 34.55
GANTRY CL (877.6)
CABLE SCAN
ACCESS IS 24.56
(623.8)
AVAILABLE
42.75 16.80 57.44 34.47
(1085.8) (426.7) (1459.0) (875.5)
16.00 182.96
(406.4) (4647.2) CT3504A
7 6 IS ACCEPTABLE FOR GANTRY INSTALLATION. THIS LOW CEILING HEIGHT WILL REQUIRE THE REAR GANTRY
COVER TO BE REMOVED FOR A TUBE CHANGE.
3.12
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Illustration 110
POWER DISTRIBUTION UNIT (PM)
NOTE:
2.2
(57) DIMENSIONS ARE IN INCHES
(MILLIMETERS)
INDICATES MINIMUM
2.0
(51) CLEARANCE
9.0 x 12.0 INDICATES AIR FLOW
(229 x 305)
CABLE ACCESS CENTER OF GRAVITY
(CABINET 36.0
FLOOR) (914)
SERVICE AREA
23.25
(591)
33.0 1.0 17.5
VENT (838) VENT (25) (444)
DO NOT 32.0 DO NOT
(813) 6.0
OBSTRUCT OBSTRUCT (152)
78.0
(1,981)
16.5 8.3
(419) (212)
CABLE
CABLE
ACCESS 1.0
ACCESS
(25)
3.0 27.4
10.0 (76) (695)
(254) 4.5
(114)
4.5 5.5
(114) (140)
CT3513A
10.5 12.0 6.0 2.8
(267) (305) (152) (70)
ONE OF TWO POSSIBLE POWER DISTRIBUTION UNITS. REFER TO ILLUSTRATION 111 AND ILLUSTRATION 112 FOR
POWER DISTRIBUTION UNITS SHIPPED AFTER 1 MARCH 1995.
3.13
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Illustration 111
COMPACT POWER DISTRIBUTION UNIT (PM)
NOTE:
DIMENSIONS ARE IN INCHES (MILLIMETERS).
7.3
(185)
50
(1270)
22
(565)
14.5
(370)
4.5
(115)
4.7 3.94 30
(120) (100) (30)
7.3
(185)
POWER DISTRIBUTION UNIT SHIPPED WITH HISPEED CT/I AFTER 11 MARCH 1996.
3.14
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Illustration 112
COMPACT POWER DISTRIBUTION UNIT (PM)
NOTES:
DIMENSIONS ARE IN INCHES (MILLIMETERS).
INDICATES AIR FLOW:
INDICATES CENTER OF GRAVITY
22
14.75 (559)
14.5
(375)
(368)
36.0
(914)
SERVICE AREA
SERVICE AREA
30 24.0
(762) (610)
72
(1829)
50
(1270)
14.75
(375)
14.5 21.75 21.75
(368) (552) (552)
1.34 FRAME THRU
(34) HOLES 4
LOCATIONS
0.3125 (8)
4.85 (123)
19.3
(490)
3.15
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Illustration 113
OPERATORS CONSOLE / INDIGO COMPUTER (OC1)
NOTES:
DIMENSIONS ARE IN INCHES (MILLIMETERS).
INDICATES AIR FLOW:
SEISMIC BRACKET
INDICATES CENTER OF GRAVITY
SUPPLIED WITH CONSOLE
14
0.5x1.0
(355.6)
(12.7x25.4)
Cables and 2 10
(254) Elongated Hole
power Cables in this (50.8) 4 Places
cord in this area.
area. 2.55
(64.9)
1.836
(46.6)
1.75
(44.5)
TOP VEIW
26.1
(663)
ATTACH SEISMIC BRACKETS TO SIDES OF CONSOLE
10.5 FRAME USING (2) 3/8x16 BOLTS,
(267) 2 SPLIT LOCK WASHERS AND 2 FLAT WASHERS
39.00
ATTACH ASSEMBLED UNIT TO FLOOR USING HILTI
KWIKBOLT II 1/2x8 (12.7 x 203).
1.25 1.00
48.00 TYP 16.00
24.00 1.25
33.50
26.1 22.25 10.5
29.50
(663)
25.25 (267)
Cables and
17.1 power cord in
(434) this area.
3.00
3.00
23.06 1.61
47.00 (40)
12.00
32.06
5.2 (132)
10 0.425 (11)
(25.4)
3.16
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Illustration 114
OPERATORS CONSOLE / OCTANE COMPUTER (OC1)
NOTES:
DIMENSIONS ARE IN INCHES (MILLIMETERS).
INDICATES AIR FLOW:
SEISMIC BRACKET INDICATES CENTER OF GRAVITY
SUPPLIED WITH CONSOLE 14 0.5x1.0
(355.6) (12.7x25.4)
CABLES AND 2 10
POWER (254) Elongated
CABLES IN (50.8)
CORD IN THIS AREA. Hole
THIS AREA. 2.55 4 Places
1.836 (64.9)
(46.6)
1.75
TOP VEIW (44.5)
24.12
(614) ATTACH SEISMIC BRACKETS TO SIDES OF
9.57 CONSOLE FRAME USING (2) 3/8X16 BOLTS, 2
(243) 39.00 SPLIT LOCK WASHERS AND 2 FLAT WASHERS
1.25 1.00
48.00 TYP 16.00
24.00 1.25
33.50
22.25 29.50
24.12 25.25 9.57
(614) (243) CABLES AND
15.63 POWER
CORD IN
(397)
3.00
47.00 23.06 1.61
12.00 (40)
32.06
5.2 (132)
0.425 (11)
10
(25.4)
FRONT VIEW SIDE VIEW
3.17
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
NOTES
3.18
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
SECTION 4
ENVIRONMENTAL CONDITIONS
Ratings and duty cycles of CT subsystems apply if site environment meets the standards of this
section.
Maintain environmental conditions listed below at all times Including, for example,
overnight, weekends and holidays. Shut down the CT system if air conditioning is not working.
When system is shut down for major repair, air conditioning may be shut down also.
Temporarily install a temperature and humidity recorder close to gantry in scan room. Use
recorder before installation and again at installation to verify true temperature and humidity
conditions.
You may wish to consider an air conditioner with two compressor units. Such an air conditioner
permits CT system operation during an extended repair of the primary air conditioner.
Over half the cooling requirement is for gantry. For patient comfort, locate a wall
air-conditioning vent at floor level beside and behind gantry to meet gantry cooling
requirement.
4.1
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Table 13
Cooling Requirements
SYSTEM
COMPONENT BTU/HR WATT BTU/HR WATT BTU/HR WATT BTU/HR WATT BTU/HR WATT
SCANNER SYSTEM
5. TV camera 34 10
6. TV monitor 300 88
SELECTED OPTIONS
Note 3 With 75 slices per patient, minimum gantry cooling is enough for up to three patients per
hour. Recommended cooling is enough for up to four patients per hour. Use growth BTU/hr
to size cooling for future, more powerful X-ray tubes.
Note 4 Cooling requirements do not include cooling for room lighting, personnel or nonCT
equipment.
4.4 Altitude
System operating altitude is from mean sea level to 10,000 ft (3050 m).
4.2
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
4.3
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
NOTES
4.4
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
SECTION 5
FLOOR LOADING AND WEIGHTS
Table 14 shows CT components with size and weight, floor loading and normal mounting
requirements.
Table 14
HiSpeed Advantage System Floor Loading
Gantry 3750 88 X 43 153 Rectangular base plate 24 X 81 (610 X 2057) Hilti KwikBolt II 1/2in (12.7mm)
(1701) (2235 X 1092) (736) with four round pads, each 2.5 (63.5) in diameter by 8in (203mm) long per
contact with floor and supporting 1850 lbs P/N 2106573 at four leveling pads
(840 kg) center to center. into concrete floor.
Patient table 1185 (539) 24.3 X 198.3 182 Rectangular base 18 X 57 (457 X 1448) with Hilti KwikBolt II 1/2in (12.7mm)
Includes (617 X 5037) (897) five round pads, each 2.5 (63.5) in contact diameter by 8in (203mm) long per
450 (204) with floor. Worst-case cantilever pull on any P/N 2106573 at three leveling pads
Patient bolt is 1235 lb (562 kg). into concrete floor.
Power 800 30 X 22 180 Four Casters or support area of 30 X 22 (838 Casters are for positioning and
distribution unit (363) (762 X 559) (82) X 457). service. Set on floor. May be
anchored to floor using angle
brackets in seismic zones.
Operators 662 48 X 39 50 Four Casters or Leveling Feet support area of Casters are for positioning. Set on
console/ (300) (1219x100) (23) 46 X 19 (1168 X 483). floor. Four leveling pads. May be
computer anchored to floor using angle
brackets in seismic zones.
5.1
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Illustration 115
GANTRY (CT2)
SCAN
CENTERLINE
7.1 (180.3) to CG
82.00 78.55
(2082.8) (1995.2)
CG
43.00
(1092.2)
FLOOR CT3502A
Illustration 116
TYPICAL FLOOR ANCHOR, GANTRY AND TABLE
63.5mm DIAMETER
LEVELING PAD
195mm
110mm
5.2
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Illustration 117
PATIENT TABLE (CT1)
These two locations, leveling only No Anchors
NOTE:
CENTER OF GRAVITY IS
WITH 450 LB (205 KG)
PATIENT FULLY EXTENDED
(Y = 32.8 (833.1)) 96.21
(2443.7)
198.96
(5053.6)
90.73
(2304.5)
TRAVEL
16.00
(406.4)
CRADLE
EXTEN- 42.00
SION (1066.8) Z
43.78 32.8
(1112) (833.1)
20.26
(514.6)
91.90
This location leveling only (2334.3)
CT3511A
5.3
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Illustration 118
POWER DISTRIBUTION UNIT (PM)
3.0
(76) NOTE:
29.0 2.0
(737) (51) DIMENSIONS ARE IN
INCHES (MILLIMETERS)
VIEW AA
33.0
(838)
32.0
(813)
27.0 22.3
(686) (566)
13.5 13.5 3.0 17.5
(686) (686) (76) (444)
TA
0.38 (9.5) BOLT IN STEEL
STUD WALL OR IN
CONCRETE ANCHORS
(CONCRETE WALL)
0.19 X 6 (4.8 X 152)
STEEL PLATE WELD TO
78.0 3 STEEL STUDS AT A
(1981)
76.5
STEEL STUDS (1943)
OR
CONCRETE 8.3
16.5 WALL (212)
(419)
A
A
27.4
(695)
CT3513A1
0.38 (9.5) BOLT IN
CONCRETE ANCHORS
5.4
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Illustration 119
POWER DISTRIBUTION UNIT (PM)
NOTES:
DIMENSIONS ARE IN INCHES (MILLIMETERS).
INDICATES AIR FLOW:
INDICATES CENTER OF GRAVITY
22
14.75 (559)
14.5
(375)
(368)
36.0
(914)
SERVICE AREA
SERVICE AREA
30 24.0
(762) (610)
72
(1829)
50
(1270)
14.75
(375)
14.5 21.75 21.75
(368) (552) (552)
1.34 FRAME THRU
(34) HOLES 4
LOCATIONS
0.3125 (8)
4.85 (123)
19.3
(490)
5.5
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Illustration 120
OPERATORS CONSOLE / INDIGO COMPUTER (OC1)
NOTES:
DIMENSIONS ARE IN INCHES (MILLIMETERS).
INDICATES AIR FLOW:
SEISMIC BRACKET INDICATES CENTER OF GRAVITY
SUPPLIED WITH CONSOLE 14 0.5x1.0
(355.6) (12.7x25.4)
CABLES AND 2 10
POWER (254) Elongated
CABLES IN (50.8)
CORD IN THIS AREA. Hole
THIS AREA. 2.55 4 Places
1.836 (64.9)
(46.6)
1.75
TOP VEIW (44.5)
26.1
(663) ATTACH SEISMIC BRACKETS TO SIDES OF
10.5 CONSOLE FRAME USING (2) 3/8X16 BOLTS, 2
(267) 39.00 SPLIT LOCK WASHERS AND 2 FLAT WASHERS
1.25 1.00
48.00 TYP 16.00
24.00 1.25
33.50
22.25 29.50
26.1 25.25 10.5
(663) (267) CABLES AND
17.1 POWER
CORD IN
(434)
3.00
47.00 23.06 1.61
12.00 (40)
32.06
5.2 (132)
0.425 (11)
10
(25.4)
FRONT VIEW SIDE VIEW
5.6
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Illustration 121
OPERATORS CONSOLE / OCTANE COMPUTER (OC1)
NOTES:
DIMENSIONS ARE IN INCHES (MILLIMETERS).
INDICATES AIR FLOW:
SEISMIC BRACKET INDICATES CENTER OF GRAVITY
SUPPLIED WITH CONSOLE 14 0.5x1.0
(355.6) (12.7x25.4)
CABLES AND 2 10
POWER (254) Elongated
CABLES IN (50.8)
CORD IN THIS AREA. Hole
THIS AREA. 2.55 4 Places
1.836 (64.9)
(46.6)
1.75
TOP VEIW (44.5)
24.12
(614) ATTACH SEISMIC BRACKETS TO SIDES OF
9.57 CONSOLE FRAME USING (2) 3/8X16 BOLTS, 2
(243) 39.00 SPLIT LOCK WASHERS AND 2 FLAT WASHERS
1.25 1.00
48.00 TYP 16.00
24.00 1.25
33.50
22.25 29.50
24.12 25.25 9.57
(614) (243) CABLES AND
15.63 POWER
CORD IN
(397)
3.00
47.00 23.06 1.61
12.00 (40)
32.06
5.2 (132)
0.425 (11)
10
(25.4)
FRONT VIEW SIDE VIEW
5.7
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Illustration 122
GANTRY / TABLE MOUNTING DIMENSIONS
5.8
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
SECTION 6
DELIVERY DATA
6.1
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Table 15
SHIPPING DIMENSIONS AND WEIGHTS
(VAN DELIVERY TYPICAL INSIDE NORTH AMERICA)
CATALOG: B7500JK, B7500JB, B7500JA, B7500JG,
B7500JD, B7500JE
DIMENSIONS
2 Gantry (dollies) Refer to Illustration 123 and 109 2769 54 1372 82 2083 3992 1812
Table 17 for Gantry delivery options
13 Parts and accessories software/bay collector (2 48 1219 40 1016 48 1219 450 205
boxes on skid)
6.2
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Table 16
SHIPPING DIMENSIONS AND WEIGHTS
(CRATED DELIVERY OUTSIDE NORTH AMERICA)
CATALOGS: B7500JK, B7500JB, B7500JA, B7500JG,
B7500JD, B7500JE
DIMENSIONS1
1 Gantry (Dolly in raised position extends beyond 120 3048 56 1423 86 2185 4573 2075
end of skid.) (Slat Crate)
Illustration 123
GANTRY SHIPPING CONFIGURATION
NOTES:
DIMENSIONS ARE IN INCHES
(MILLIMETERS).
1 DRAWING NOT TO SCALE
2 EACH DOLLY IS EQUIPPED WITH A
THIRD DEMOUNTABLE CASTER
TO STABILIZE IT DURING ASSEMBLY
TO GANTRY.
3 DIMENSIONS SHOWN ARE WITH DOLLY
CASTERS RETRACTED AND GANTRY
RESTING ON ITS BASE PLATE.
4 SIDE RAILS MAY BE REMOVED
FOR HALLWAY OR DOOR CLEARANCE
89
(2260)
6.3
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Table 17
GANTRY DELIVERY SIZE OPTIONS
Remove Protective 3712 109 43 (1092) 83 (2108) 2 0.5 With rails removed, be very careful to
Side Rails (1683) (2769) prevent damage to covers.
Remove Side Rails 3476 99 (2515) 43 (1092) 83 (2108) 2 1.5 This configuration is used to get gantry
& one dolly (1576) into a 9ft. elevator.
Gantry set on Aux 3400 88 (2235) 43 (1092) 84.3 2 2.5 Do NOT push on gantry covers. Push
Dollies (1542) (2141) only on frame members. Aux. dollies pro-
vide 2.3 in. (58 mm) clearance from
floor.
Remove Covers 3392 110 40 77 4 1.5 Use extreme care in removing, handling
Use Shipping (1538) (2794) (1016) (1956) and installing covers. See procedure in
Dollies Direction 2152926100, HiSpeed CT/i
General System Installation.
6.4
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
SECTION 7
POWER REQUIREMENTS
7.1 Introduction
The power distribution unit (PDU) supplied with the HSA CT/i system transforms and
distributes power to all system components. The PDU is the only power entry point required.
Power wiring between the facility main distribution panel and the PDU should be kept as short
as possible. This minimizes voltage regulation effects.
Carefully consider advantages and disadvantages of conduits, floor ducts and surface raceways
for running cables. Make cable passageways large enough to install any cable with all other
cables already installed.
When routing power cables, all three phase wires and ground must be run in the same conduit
or raceway duct. Power cables should be routed separately from system control cables. for
example, use a separate trough in duct.
7.2.2 Configuration
The HSA CT/i system is designed to operate on threephase, fourwire delta or wye power.
A ground referenced wye source produces the lowest leakage currents and is preferred.
However, the neutral wire does not need to be run to the system, ie., fourwire connection. (A
dummy terminal is provided for parking the neutral wire in the event a fivewire service is
already installed at the site.)
7.2.3 Rating
Voltage 380 to 480 VAC1
Capacity 90 KVA @ 85% P.F.
Frequency 50 or 60 Hz +/0.5 Hz
1 The absolute range of line voltage at the input to the PDU must remain within one of the ranges
shown in Table 18 at all times.
7.2.4 Regulation
The size of the facility transformer and feeder wires determine load regulation presented to the
system. Total load regulation as measured at the PDU input terminals must not exceed 6%.
7.1
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
7.2.7 Microcuts
The HSA CT/i System is generally unaffected by microcuts.
7.2.8 Grounding
A dedicated 1/0 (55mm2) or larger insulated copper ground wire must be run with the phase
wires from the main distribution panel to the PDU. The ground wire should be bonded to any
intermediate distribution panels through which it passes.
The resistance between the PDU ground and the facility earth ground must not exceed 0.5 ohm.
In addition, the total resistance between the PDU ground and earth must not exceed 2 ohms.
7.2
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Table 18
Facility Power Requirement
NOMINAL LINE VOLTAGE (VAC) 380 400 420 440 460 480
MOMENTARY LINE CURRENT (A) 137 130 124 118 113 108
@ NOMINAL LINE VOLTAGE
MAXIMUM LINE CURRENT (A) 149 141 135 128 123 117
@ LOW LINE VOLTAGE
251300 ft (7791 m) 2/0 (70) 1/0 (55) 1/0 (55) 1 (45) 1 (45) 2 (35)3
Note 7 Note 7 Note 7 Note 7 Note 7 Note 7
301350 ft (92106 m) 3/0 (85) 2/0 (70) 2/0 (70) 1/0 (55) 1/0 (55) 1 (45)
Note 7 Note 7 Note 7 Note 7 Note 7 Note 7
351400 ft (107122 m) 3/0 (85) 3/0 (85) 2/0 (70) 2/0 (70) 1/0 (55) 1/0 (55)
Note 7 Note 7 Note 7 Note 7 Note 7 Note 7
Note 5 The feeder table above is based on the use of copper wire, rated 75C and run in steel
conduit. Wire size is in AWG (mm 2). Ampacity is determined in accordance with the
National Electric Code (NFPA 70), Table 31016 (1996).
Note 6 The wire size shown is the minimum allowable for the specified overcurrent protection
rating.
Note 7 The feeder is sized to contribute 2.4% maximum regulation.
Note 8 A minimum 1/0 (55 mm 2) ground wire is required.
Note 9 A neutral wire is not required and, if present, is not used.
Note 10 The Primary Disconnect (A1) must have provision for tagout / lockout.. Illustration 127
shows the primary power disconnect (A1) using a combination fusible disconnect and
magnetic contactor. Alternatively, use a GE Spectra RMS molded case circuit breaker or
equivalent.
If the HSA CT/i system must be powered from an existing distribution transformer and
secondary feeder, such as the equipment distribution panel of an Xray department, installation
with other XRay equipment which use rapid film changers should be avoided. These changers
use a large number of high powered, closely spaced exposures which may coincide with a CT
scan and produce image artifacts.
In all cases, qualified personnel must verify that the transformer and feeder, at point of takeoff,
plus the run to HSA CT/i meets all the requirements of Section 7.2.
Power Systems for XRay Rooms, Direction 46013833, is available for additional information
on power requirements for Xray systems.
7.3
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
7.4
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Illustration 124
SYSTEM GROUND MAP
A1
POWER GANTRY
DISCONNECT (CT2)
ROTATING
ASSEMBLY
TABLE/GANTRY FRAME
G JUNCTION
#1/0 RACEWAY TILT
MECH
POWER #1/0 #2 Part of Gantry
DISTRIBUTION T FRAME
UNIT
(PM)
2 Ground Braids (Part of Gantry)
TABLE
(CT1)
CAMERA
#2 POWER FROM
NONSYSTEM
SOURCE
CAMERA OR
#2 CAMERA IFACE
(OPTION)
OPERATORS
CONSOLE/
COMPUTER
LINE PRINTER
(OC1) (LP)
(OPTION)
#10 TO CONSOLE/COMPUTER
TO PM (OC1)
VCR
(OPTION)
TO CONSOLE/COMPUTER
(OC1)
SLAVE MONITOR
(SM)
(OPTION)
TO CONSOLE/COMPUTER
(OC1)
NOTES: VIDEO PRINTER
SHIELD / SIGNAL GROUNDS ARE NOT SHOWN.
(OPTION)
= GROUND WIRE IN SUPPLIED CABLE. TO CONSOLE/COMPUTER
DISK UNITS ARE LOCATED IN CC1 AND SCU. (OC1)
7.5
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
NOTES
7.6
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
SECTION 8
INTERCONNECTION DATA
8.1 Introduction
Illustration 125 shows interconnection runs for a 50/60 Hz system.
Table 19 shows component designators for supplied equipment and options, and wall power
outlets.
Table 110 lists customerinstalled wiring and supplied cables for runs shown in
Illustration 125. Actual length of each run is less than the length of supplied cables to allow
for routing inside equipment. Cable diameters and sizes of connectors are provided to aid in
sizing conduit and access plates.
Table 111 lists details for connection to HSA CT/i equipment. Details are listed for the
following types of runs as appropriate;
flushfloor duct
surface floor duct
computer floor
throughfloor duct
throughwall bushing
wall duct
junction box
conduit.
Illustration 126 shows recommended locations of contractorsupplied components
corresponding to equipment locations suggested in Illustration 12.
Need for additional junction boxes is minimized by use of either a cable raceway system or a
raised computer floor. HSA CT/i uses prefabricated cables with large plugs. Therefore, conduit
or pipe is not recommended for cable runs.
8.1
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Illustration 125
INTERCONNECTION RUNS
TO POWER
SOURCE
SEO A1 WL DS
3
2 8 9
PM
(PDU)
TO
5 TELEPHONE
SYSTEM*
To Laser Imager
Option **
7
KEY:
TO
GANTRY ETHERNET OPTION
(CT2)
CT1791A3
8.2
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Table 19
Component Designators
COMPONENT
DESIGNATOR APPLIES TO SOURCE
8.3
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Table 110
Runs, Wiring and Cables
RUN NO. 1 FROM PRIMARY POWER SOURCE TO FACILITY DISCONNECT (POWER SOURCE A1)
MAXIMUM RUN LENGTH *
RUN NO. 2 FROM FACILITY DISCONNECT TO POWER MODULE (A1 PM) MAXIMUM RUN LENGTH *
RUN N0. 3 FROM FACILITY DISCONNECT TO SYSTEM EMERGENCY OFF (A1 SEO)
RUN NO. 9 POWER MODULE TO SCAN ROOM DOOR INTERLOCK (PM DOOR SWITCH)
8.4
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
RUN NO. 4 TABLE/GANTRY TO POWER MODULE (CT1/CT2 TO PM) MAXIMUM RUN LENGTH 58 FT (17.6m)
B7500J and B7500JE are CE Compliant Systems and use Gantry Model No. 21197342 for selected country shipments only.
GROUND WIRE 1/0 (50) 46327217P1 65 (20) 0.6 0.6 0.6 3.5 (1.1) 3.5 (1.1)
(14) (14) (14)
PDU GANTRY SIGNAL 2118146 60 (18.3) 0.6 2.0X0.8 2.0X0.8 4 (1.2) 2.5 (0.76)
(14) (50X19) (50X19)
TABLE POWER 2118141 60 (18.3) 0.6 1.8x1.3 1.8x1.3 5 (1.5) 2.5 (0.76)
(14) (45X34) (45X34)
Used with 550VDC POWER 46297080G1 60 (18.3) 0.76 NA NA 5 (1.5) 7.5 (2.3)
Gantry (19)
2119732
Used with 550VDC POWER 2118143 60 (18.3) 0.76 NA NA 5 (1.5) 7.5 (2.3)
Gantry (19)
21197342
Used with 120VAC POWER 2119603 60 (18.3) 0.76 NA 2 (51) 5 (1.5) 2.5 (0.76)
Gantry (19)
2119732
Used with 120VAC POWER 2118142 60 (18.3) 0.76 NA 2 (51) 5 (1.5) 2.5 (0.76)
Gantry (19)
21197342
Used with AXIAL MOTOR POWER 46327155G1 60 (18.3) 0.76 NA NA 5 (1.5) 7 (2.1)
Gantry (19)
2118213
Used with AXIAL MOTOR POWER 2118145 60 (18.3) 0.76 NA NA 5 (1.5) 7 (2.1)
Gantry (19)
21197342
RUN NO. 5 OPERATOR CONSOLE TO POWER MODULE (OC1 PM) MAXIMUM RUN LENGTH 76 FT (23m)
OPERATOR CONSOLE POWER 2118139 80 (24.4) 1 (26) NA 2 (51) 1.5 (0.4) 2.5 (0.76)
8.5
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
RUN NO. 7 OPERATOR CONSOLE TO GANTRY (OC1 TO CT2) MAXIMUM RUN LENGTH 74 FT (22.5m)
CONSOLEGANTRY SIGNAL 2139577 80 (24.4) 1.5 2.5X0.8 2.5X0.8 2.0 (0.6) 4 (1.2)
(38) (64X19) (64X19)
CONTROL LAN 2118150 80 0.42 0.6 0.6 3.5 (1.1) 3.5 (1.1)
(24.4) (11) (14) (14)
DAS DATA (FIBER OPTIC) 21178482 80 0.1 0.25 0.25 3.5 (1.1) 3.5 (1.1)
(24.4) (2.5) (6.3) (6.3)
EXPOSURE INTERLOCK 2118148 80 0.16 0.6X0.8 0.6X0.8 3.5 (1.1) 3.5 (1.1)
(24.4) (5) (15X19) (15X19)
Table 111
ContractorSupplied Components
A1 Main 480VAC, Surface or Flush Mount, On/Off Control R4503AK Main Disconnect
(60 Hz) Disconnect Control Control, 480VAC, Surface Mount
(MDC) with Flush Mount Kit included, two
remote Push Button Switches.
A1 Fusible Disconnect 3 Pole, 380V, Magnetic Combination Contactor. With control trans- Source Locally
(50 Hz) and Magnetic former for 220V 50 Hz contactor coil, momentary contact start
(380V) Contactor stop pushbutton station, indicator light, multiple neutral block, mul-
tiple ground block and 160 amp, dual element fuses.
ITL Insuite Telephone Supply 2 voicegrade telephone lines. One line must be a direct num-
Lines ber from outside the facility do not route this line through a tele-
phone switchboard. Telephone line operating charges are paid by
customer.
MFC GE LaserCam HQ Refer to M969 Laser Imager HQ Site Planning Guide, for preinstallation
information for the 3M M969 laser camera (catalog B7920T, multi
user and B7920TC, single user imager) and their options. Informa-
tion needed for preinstallation of VEIBs, KEIBs, and Film Proces-
sors is also in this book.
8.6
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Illustration 126
CONTRACTOR COMPONENTS LOCATION
EQUIPMENT ROOM
68 x 90
LASER CAMERA
OPTION
CONTROL AREA
12.3 X 9.0
(3.81 X 2.67)
INJECTOR
A1 CONTROL OPTION
(ELECTRONICS
OPERATORS ON SHELF
CONSOLE BELOW )
PDU
TV
PM MONITOR SEO
WALL (NOT BY GE)
DUCT
FLUSH
FLOOR OX/VAC/INJECTORS
DUCT IN CEILING
PATIENT
TABLE
SCAN ROOM
195 x 1211 WL
CT
CT1 GANTRY
AND
CT2 DS
03 6 9 1 2 3 4 CT3509A
8.7
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Illustration 127
PRIMARY POWER DISCONNECT (A1)
FUSIBLE DISCONNECT AND MAGNETIC CONTACTOR
X1 X2
NC
M
STOP STOP START
SYSTEM EMERGENCY
OFF
M TO
IN FRONT COVER CONTROL
MECHANICAL MAINS TRANSFORMER
SWITCH FUSES
125 AMPS
A A
MAINS B B TO
INPUTS PDU
C C
M
GND GND
NEUTRAL NEUTRAL
CONTACTOR ENCLOSURE
CT3306A
8.8
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
Illustration 128
TYPICAL CUSTOMER SUPPLIED SCAN ROOM WARNING LIGHT CONNECTION TO PM
A3J2 FUSE
A3K3 Hi Line
1 24 V Secondary
Neutral
Lo
Hi
Lo
4 Room Warning
Lamp
FACILITY SUPPLIED
Illustration 129
TYPICAL CUSTOMER SUPPLIED SCAN ROOM DOOR INTERLOCK SWITCH CONNECTION TO PM
A3 J6 1,2
TO RELAY
CONTROL N.O. Switch
CIRCUIT
BOARD
J3 46,50
Customer Supplied
8.9
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
NOTES
8.10
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
SECTION 9
PREINSTALLATION CHECKLIST
Scheduled
Delivery Date _______________________ Salesperson _____________________
Customer ____________________________ FDO # __________ Room # __________
Equipment
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
YES NO NA
____ ____ ____ 1. Are detailed signed plans available for the installation crew?
____ ____ ____ 2. Has all preliminary work been done per plan?
____ ____ ____ 3. Site conditions:
____ ____ ____ a Have doors, elevators, corridors, etc. been checked to make sure
equipment can reach room destination?
____ ____ ____ b Are all boxes and conduits installed?
____ ____ ____ c Are wires pulled and 6 ft (1.83 m) wire length available at all
boxes?
____ ____ ____ d Is flexible stranded copper wire used?
____ ____ ____ e Is the finish floor down?
____ ____ ____ f Is the primer paint on walls?
____ ____ ____ g Are lights in the room?
____ ____ ____ h Is power available to operate power tools?
____ ____ ____ i Is correct power available?
____ ____ ____ j Are line and remote disconnects available?
____ ____ ____ k Is there a lock on door or a locked room available?
9.1
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
YES NO NA
____ ____ ____ l Are junction box covers and fittings on job site?
____ ____ ____ m Are Unistrut support structures installed and leveled?
____ ____ ____ o Are ceiling, walls and floor clear of all obstructions?
Ceiling Height __________ Type _______________
____ ____ ____ p Has any work which would create dust in adjacent areas been
completed?
____ ____ ____ s Have preparations been made for an equipotential grounding
system?
____ ____ ____ t Do conduit lengths for General Electricsupplied cables allow
for required cable outside of conduit?
____ ____ ____ u Is the floor compatible for equipment mounting? Has concrete
strength been determined if anchors are to be used instead of
throughbolting?
____ ____ ____ v Have arrangements been made for removal of old equipment in
the room?
____ ____ ____ y Is adequate interface information available for old and new
equipment?
____ ____ ____ aa Are center positions clearly marked to eliminate interference
with ceiling, rear wall and any equipment located at head, foot
or rear of table?
____ ____ ____ ad Have system options been determined? (Refer to Section 2.)
____ ____ ____ af Are any required nonelectrical lines (air, oxygen, vacuum,
water) provided for?
____ ____ ____ ag Is the InSite phone Line Installed and Tested?
____ ____ ____ ah Is the InSite Modem on order for the installation?
____ ____ ____ ai Are the Internet Addresses for the System/InSite Modem available?
9.2
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
YES NO NA
____ ____ ____ 4. Inspection Dates
Comments:
9.3
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
NOTES
9.4
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
0("
(++. * %+./ &"/)& /)(( ,+3". !&/0.&10&+* 1*&0
/" / **". /4/0") &((1/0.0&+*
(++. ,(0"/ $*0.4
"*0".+#$.2&04
$*0.4
+,".0+./ +*/+(" +),10".
1)&!&04 &. ++(&*$ ."-1&.")"*0/ *! /," &#& 5
,+3". !&/0.&10&+* 1*&0
0&+*/
/)(( ,+3". !&/0.&10&+* 1*&0
0("
%" ' (&/0 ,."&*/0((0&+* 6
+),+*"*0 *0". +**" 0&+* .1*/
+*0. 0+. /1,,(&"! +*0. 0+. /1,,(&"!
(+ 0&+* &* /),(" /1&0"
!&$.)
++(&*$ ."-1&.")"*0/ &. ++(&*$ ."-1&.")"*0/ *! 0(" +# ("/ /1,,(&"! *! !!&0&+*( 3&.&*$ ."-1&."!
/," &#& 0&+*/
+1*0&*$ ."-1&.")"*0/
(++. ,(0" &*/0((0&+*
"(&2".4 !0 !"/ .&,0&+*
$*0.4
+3".
&*,10 +*#&$1.0&+*
(" 0.+)$*"0& &*0".#"."* "
."-1&.")"*0/ 4 +),+*"*0
+3". !&/ +**" 0 !&$.) #1/&(" !&/ +**" 0 *! )$5
*"0& +*0 0+.
+3". ."-1&.")"*0/
!&/0.&10&+* 0.*/#+.)".
(++. $.+1*!&*$
(+!&*$ !&$.)
4 +),+*"*0 (&*" )+*&0+.
/1)).4 ."$1(0&+*
/0."*$0% +# ."&*/0((0&+* %" ' (&/0 6
Index.1
GE MEDICAL SYSTEMS HISPEED CT/I PREINSTALLATION
REV 3 January, 1998 DIRECTION 2152916100
-/,1#!1',+ 1 )#
1,/%#
,,* ,,/ +1#/),!( 6-'!) 4'1!& +" ,++#!1',+
*#"'
0601#* -/',/ 1, '+01))1',+
,,* )6,21
2'1# )6,21
601#* !,*-,+#+10
,,* /+'+% '%&1 6-'!) ,+1/,) '/!2'1 +" ,+7
+#!1',+ 601#* ,-1',+0
Index.2
CT SERVICE ENGINEERING, NB918
P.O. BOX 414
MILWAUKEE WISCONSIN 532010414 U.S.A.
Fax: (414) 7855874