Sie sind auf Seite 1von 30

Operating Manual

Alarmbox AB 01

BINDER GmbH
Address Post office box 102
D-78502 Tuttlingen
Tel. +49 7462 2005 0
Fax +49 7462 2005 93 0
Internet http://www.binder-world.com
E-mail info@binder-world.com
Service Hotline +49 7462 2005 555
Service Fax +49 7462 2005 93 555
Service E-Mail service@binder-world.com
Service Hotline USA +1 866 816 8191
Service Hotline Asia Pacific +603 6204 2855

Issue 04/2006 Art. No. 7001-0038


EG - KONFORMITÄTSERKLÄRUNG
EC - DECLARATION OF CONFORMITY
CE - DECLARATION DE CONFORMITE

Anbieter / Supplier / Fournisseur: BINDER GmbH


Anschrift / Address / Adresse: Im Mittleren Ösch 5, D-78532 Tuttlingen
Produkt / Product / Produit: Alarmbox - Alarmgerät
Alarm Box – alarm device
Boîte d’alarme – dispositif d’alarme
Typenbezeichnung / Type / Type: AB 01

Die oben beschriebenen Produkte sind konform mit folgenden harmonisierten Normen:
The products described above are in conformity with the following harmonized standards:
Les produits décrits ci-dessus sont conformes aux normes harmonisées suivantes:

Sicherheit / safety / sécurité:


IEC/CEI 61010-1:2001 Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel-
und Laborgeräte – Teil 1: Allgemeine Anforderungen
Safety requirements for electrical equipment for measurement,
control, and laboratory use – Part 1: General requirements
Règles de sécurité pour appareils électriques de mesurage, de
régulation et de laboratoire – Partie 1 : Prescriptions générales
IEC/CEI 61010-2-010:2003 Sicherheitsbestimmungen für elektrische Meß-, Steuer-, Regel-
und Laborgeräte – Teil 2-010: Besondere Anforderungen an
Laborgeräte für das Erhitzen von Stoffen
Safety requirements for electrical equipment for measurement,
control, and laboratory use – Part 2-010: Particular requirements
for laboratory equipment for the heating of materials
Règles de sécurité pour appareils électriques de mesurage, de
régulation et de laboratoire. Partie 2-010 : Prescriptions
particulières pour appareils de laboratoire utilisés pour
l’échauffement des matières
EMV / EMC / CEM:
IEC/CEI 61326:1997 + A1:1998 + Elektrische Betriebsmittel für Leittechnik und Laboreinsatz – EMV-
A2:2000 Anforderungen
Electrical equipment for measurement, control and laboratory use
– EMC requirements
Matériels électriques de mesure, de commande et de laboratoire
– Prescriptions relatives à la CEM

1/2

AB 01 04/2006 page 2/30


Die oben beschriebenen Produkte sind konform mit folgenden EG-Richtlinien:
The products described above are in conformity with the following EC guidelines:
Les produits décrits ci-dessus sont conformes aux directives CE suivantes:

Niederspannungsrichtlinie Richtlinie 73/23/EWG des Rates vom 19. Februar 1973 zur
73/23/EWG, Änderung 93/68/EWG Angleichung der Rechtsvorschriften der Mitgliedstaaten betreffend
Low voltage directive elektrische Betriebsmittel zur Verwendung innerhalb bestimmter
73/23/EEC, amended 93/68/EEC Spannungsgrenzen
Directive basse tension Council Directive of 19 February 1973 on the harmonization of the
73/23/CEE, modifiée 93/68/CEE laws of Member States relating to electrical equipment designed
for use within certain voltage limits (73/23/EEC)
Directive 73/23/CEE du Conseil, du 19 février 1973, concernant le
rapprochement des législations des États membres relatives au
matériel électrique destiné à être employé dans certaines limites
de tension
EMV-Richtlinie Richtlinie 89/336/EWG des Rates vom 3. Mai 1989 zur
89/336/EWG, Änderung 93/68/EWG Angleichung der Rechtsvorschriften der Mitgliedstaaten über die
EMC Directive elektromagnetische Verträglichkeit
89/336/EEC, amended 93/68/EEC Council Directive 89/336/EEC of 3 May 1989 on the
Directive CEM approximation of the laws of the Member States relating to
89/336/CEE, modifiée 93/68/CEE electromagnetic compatibility
Directive 89/336/CEE du Conseil du 3 mai 1989 concernant le
rapprochement des législations des États membres relatives à la
compatibilité électromagnétique

Die oben beschriebenen Produkte tragen entsprechend die Kennzeichnung CE.


The products described above, corresponding to this, bear the CE-mark
Les produits décrits ci-dessus, en correspondance, portent l’indication CE.

D-78532 Tuttlingen, 10.04.2006

BINDER GmbH

P. M. Binder Dr.-Ing. V. Kek


Geschäftsführender Gesellschafter Leiter F & E
Managing Director Head of R & D
Directeur général Chef de service R&D

2/2

AB 01 04/2006 page 3/30


Content
1. SAFETY.................................................................................................................. 5
1.1 Legal considerations ...........................................................................................................................5
1.2 Structure of the safety instructions......................................................................................................5
1.3 Type plate ...........................................................................................................................................7
1.4 General safety instructions on installing and operating the Alarm Box AB 01 ...................................8
1.5 Intended use .......................................................................................................................................8
2. WARRANTY ........................................................................................................... 9
3. UNIT DESCRIPTION ............................................................................................ 10
3.1 Equipment overview..........................................................................................................................10
3.2 Alarm output......................................................................................................................................11
3.3 Interface RS 422 ...............................................................................................................................12
4. SCOPE OF DELIVERY, TRANSPORTATION, STORAGE, AND INSTALLATION12
4.1 Unpacking, and checking equipment and scope of delivery.............................................................12
4.2 Storage..............................................................................................................................................12
4.3 Location of installation and ambient conditions ................................................................................12
5. INSTALLATION OF THE EQUIPMENT ............................................................... 13
5.1 Electrical connection .........................................................................................................................13
5.2 Connection to the communication software APT-COM® 2.0 ............................................................14
5.3 Connection to the communication software APT-COM® 3 ...............................................................15
6. START UP ............................................................................................................ 16
6.1 Settings at the Alarm Box AB 01.......................................................................................................16
6.2 Settings in APT-COM® 2.01 ..............................................................................................................16
6.2.1 Settings for remote alarm..................................................................................................................16
6.3 Settings in APT-COM® 3 DataControlSystem ..................................................................................17
6.3.1 Connection of the Alarm Box AB 01 when operating APT-COM® 3 and the Watch Tool on the
same computer .................................................................................................................................17
6.3.2 Connection of the Alarm Box AB 01 when operating APT-COM® 3 and the Watch Tool on
different computers ...........................................................................................................................18
6.3.3 Testing the Alarm Box AB 01: Function “Alarmbox manual switch” .................................................19
6.3.4 Checking the connection to the Alarm Box AB 01 ............................................................................19
6.3.5 Setting the tolerance limits in the APT-COM® 3 measurement window ...........................................20
7. ALARM MESSAGE .............................................................................................. 20
7.1 Buzzer ...............................................................................................................................................21
7.2 Alarm course in APT-COM® 2.01......................................................................................................21
7.3 Alarm course in the Watch Tool (APT-COM® 3)...............................................................................21
8. MANUAL ALARM RESET.................................................................................... 22
8.1 Alarm reset in APT-COM® 2.01 ........................................................................................................22
8.2 Alarm reset in APT-COM® 3..............................................................................................................22
9. CLEANING, AND SERVICE................................................................................. 23
9.1 Cleaning ............................................................................................................................................23
9.2 Service ..............................................................................................................................................23
10. DISPOSAL............................................................................................................ 24
10.1 Disposal of the transport packing .....................................................................................................24
10.2 Decommissioning..............................................................................................................................24
10.3 Disposal of the unit in the Federal Republic of Germany .................................................................25
10.4 Disposal of the unit in the member states of the EC except for the Federal Republic of Germany .26
10.5 Disposal of the unit in non-member states of the EC .......................................................................27
11. TECHNICAL DATA .............................................................................................. 27
12. CONTAMINATION CLEARANCE CERTIFICATE ............................................... 28

AB 01 04/2006 page 4/30


Dear customer,

For the proper operation of the Alarm Box AB 01, it is necessary to read this operating manual completely
and carefully and to observe the given instructions.

1. Safety
This operating manual is part of the scope of delivery. Always keep it at hand.
To avoid injuries and damage observe the safety instructions of the operating manual.

WARNING
Failure to observe the safety instructions.
Serious injuries and unit damage.
¾ Observe the safety instructions in this operating manual
¾ Carefully read the complete operating instructions of the Alarm Box AB 01.

1.1 Legal considerations


This operating manual contains information necessary for the intended use, correct installation, start-up
and operation, and for the maintenance of the unit.
Understanding and observing the instructions in this operating manual are prerequisites for hazard-free
use and safety during operation and maintenance.
This operating manual cannot cover all conceivable applications. If you would like additional information,
or if special problems arise that you feel are not sufficiently addressed in this manual, please ask your
dealer or contact us directly.
Furthermore, we note that the contents of this operating manual are not part of an earlier or existing
agreement, promise, or legal relationship, nor do they modify such a relationship. All obligations on the
part of BINDER derive from the respective purchase contract, which also contains the entire and
exclusively valid statement of warranty administration. The statements in this manual neither augment nor
restrict the contractual warranty provisions.

1.2 Structure of the safety instructions


In this operating manual, the following harmonized denominations and symbols indicate dangerous
situations following the harmonization of ISO 3864-2 and ANSI Z535.4.
Signal word panel
Depending on seriousness and probability of the consequences, dangers are identified with a signal
word, the corresponding safety color, and if appropriate, the safety alert symbol.

DANGER
Indicates an imminently hazardous situation that, if not avoided, will result in death or serious
(irreversible) injury.

WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious
(irreversible) injury

AB 01 04/2006 page 5/30


CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in moderate or minor
(reversible) injury

CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in damage of the product
and/or its functions or of a property in its ambiance.

Safety alert symbol


Use of the safety alert symbol indicates risk of injury.
Observe all measures that are marked with the safety alert symbol in order to avoid death or
injury.

Pictograms
Warning signs

Electrical hazard Explosive substances Pollution Hazard Harmful substances

Biohazard
Mandatory action signs

Mandatory regulation Read operating Pull the power plug Environment protection
instructions
Prohibition signs

Do NOT touch Do NOT spray with


water

Information to be observed in order to ensure optimum function of the product.

AB 01 04/2006 page 6/30


Word message panel structure
Type / cause of hazard.
Possible consequences.
∅ Instruction how to avoid the hazard: prohibition.
¾ Instruction how to avoid the hazard: mandatory action.

Observe the other notes and information not specially emphasized in the same way, in order to avoid
disturbances which could result in direct or indirect injuries or property damage.

1.3 Type plate

Nenntemperatur --- °C 0,02 kW SCHALTER GRÜN: GERÄT 0-1


--- °F 230 V 1 N ~ SCHALTER SCHWARZ: SUMMER AUS
Schutzart IP 20 0,1 A MELDELEUCHTE ROT: ALARM
Temp. Schutz DIN 12880 50/60 Hz GERÄTESICHERUNG 10 A MITTELTRÄGE
Klasse --- DIN-BUCHSE MAX. 24 V AC/DC - 2,5 A

Artikel-Nr. 9052-0004 US PATS 4585923 / 5222612 / 5309981 SCHNITTSTELLE RS 422


Projekt-Nr. 5405194 / 5601143 / 5773287 / 6079403
D 78532 Tuttlingen / Germany
Tel. + 49 (0) 7462 / 2005-0 AB 01 # 00-00000
Made in Germany
Internet: www.binder-world.com

Figure 1: Type plate AB 01

Indications of the type plate Information


Nenntemperatur --- °C Nominal temperature
--- °F
Schutzart IP 20 IP type of protection 20 acc. to EN 60529
Temp. Schutz DIN 12880 Temperature safety device acc. to standard DIN 12880
Klasse --- Temperature safety device
Artikel-Nr. 9052-0004 Art. No. 9052-0004
Projekt-Nr. --- (Special application acc. to project no.)
0,02 kW Nominal power 0.02 kW
230 V 1 N ~ Nominal voltage 230 V ± 10%, single-phase unit
0,1 A Nominal current 0.1 A
50/60 Hz Mains frequency 50/60 Hz
AB 01 Model AB 01
# 00-00000 Serial No. 00-00000

Symbol on the type plate Information


CE conformity marking

Electrical and electronic equipment manufactured / placed on the


market in the EC after 13 August 2005 and be disposed of in
separate collection according to the directive 2002/96/EC on waste
electrical and electronic equipment (WEEE).

AB 01 04/2006 page 7/30


1.4 General safety instructions on installing and operating the Alarm Box
AB 01
With regard to operating the Alarm Box AB 01and to the installation location, please observe the
regulations BGR 120 of the German professional association of the chemical industry (formerly ZH 1/119
laboratory guidelines of the employers’ liability insurance association) (for Germany).
The BINDER GmbH is responsible for safety-related unit properties only if skilled electricians or qualified
personnel authorized by BINDER perform all maintenance and repair, and if components relating to
chamber safety are replaced in case of failure with original spare parts.

CAUTION
Danger of overheating.
Damage of the unit.
∅ Do NOT install the unit in unventilated recesses.
¾ Ensure sufficient ventilation for carrying-off the heat.
The Alarm Box AB 01must NOT be operated in hazardous locations.

DANGER
Explosion hazard.
Danger of life.
∅ Do NOT operate the unit in potentially explosive areas.
∅ NO explosive dust or air-solvent mixture in the ambiance.

DANGER
Electrical hazard.
Danger of life.
∅ The unit must NOT become wet during operation or maintenance.

The Alarm Box AB 01has been produced in accordance to the VDE regulations and were routinely tested
in accordance to VDE 0411.

1.5 Intended use


The Alarm Box AB 01 has been developed for the transmission of alarm messages released by the
communication software APT-COM from BINDER, to an already existing alarm facility of the customer, or
by use of the telephone dialing device TWG 01 from BINDER via telephone line.

Respecting the instructions in this operating manual is part of the intended use.

AB 01 04/2006 page 8/30


2. Warranty
BINDER products are manufactured with great care and checked carefully prior to delivery (VDE-piece
checked).
1. Should your BINDER product have a material defect or manufacturing fault, we shall repair it – as long
as it is a new BINDER product – free of charge if the defect or fault occurs within 12 months after
delivery. This period will be extended to 24 months, starting from delivery, provided that all
maintenance and potential repair work within the first 12 months is carried out by us or authorized
service stations.
2. Should you discover a defect or fault in your BINDER product within the aforementioned period,
please inform the company from which you purchased the BINDER product or BINDER directly.
3. All above-mentioned periods or deadlines shall begin with the delivery of a new BINDER product to
the end buyer. Producing the invoice or delivery note shall bring evidence of deadline keeping.
4. In case a BINDER product has a defect or fault, we shall either replace the defective product with a
new one or repair the defective product at our discretion (so called supplementary performance –
“Nacherfüllung”). All parts replaced in the course of such repair shall become our property. Deadlines
mentioned under above number 1 shall not be prolonged by repair work or substitution. In case
supplementary performance should fail, usual warranty rights according to German Law (withdrawal
from contract or price reduction) shall apply, provided the defect or fault occurs within the first 12
months. Should the 24 month-period be relevant and supplementary performance fail, you shall only
be entitled to a price reduction.
5. Should the defect occur within 12 months after delivery, we shall take over all costs necessary for the
repair of the defect, especially transportation, travel and labor costs, in so far as there is no cost
increase due to the product being brought to another location than the place of delivery. However, we
shall decide whether the defect will be repaired at the location where the BINDER product is operating
or whether the BINDER product has to be sent to us. In case the 24 month-period should be relevant,
please send the defective BINDER product at your expense to BINDER or the company from which
you purchased the BINDER product. We shall repair the BINDER product at your premises only if
transportation costs are excessively high. The customer shall be liable for any risk of sudden
destruction or deterioration.
6. Should you have to send back a BINDER product, the BINDER product will have to be supplied in its
original packaging or any similar suitable packaging.
7. All warranty claims as laid down in this document basically become invalid:
• if the defect or fault is caused by repair or intervention by non-authorized persons. With authorized
persons, we mean only Binder service engineers and service engineers of our service partners.
• the same applies if you have attached non-suitable parts or accessories to the BINDER products,
which are not original BINDER spare parts, and if those parts or accessories have caused the
defect or fault.
• parts that become used due to natural wear and tear, e.g. door seals, lamps, lights (enumeration
not exhaustive)
• defects caused by external intervention like e.g. fall or blow
• defects arising from non-compliance with operating instructions, improper use, abnormal ambient
conditions, overload or absence of care/maintenance
• in the case of devices that were converted or redesigned by non-authorized third parties
• in case of minor deviations from the set quality that are irrelevant to the value and usability of the
device
8. All rights or claims beyond the ones described above, especially concerning the right of compensation
for potential damages which might be caused by BINDER products shall be excluded provided that
such exclusion is not prohibited by governing law.
9. Should you send a BINDER product to us for repair or any other reasons, we shall only accept the
BINDER product upon presentation of a so-called authorization number that has previously been
issued to you. We shall issue the authorization number after receiving your complaint in writing or via
telephone prior to your sending (back) the BINDER product to us. The authorization number will be
issued following the receipt of the information mentioned below:

AB 01 04/2006 page 9/30


• BINDER product type and serial number
• Date of purchase
• Name and address of the dealer from which you bought the BINDER product
• Exact description of defect or fault
• Your full address; if possible contact person and availability of that person
• Exact location of the BINDER product
• Contamination clearance certificate via fax in advance!
The authorization number needs to be applied to the packaging in such a way that it can be easily
recognized or be recorded clearly in the delivery documents. For security reasons we cannot accept
your delivery if it does not carry an authorization number!

IMPORTANT NOTICE:
Any warranty repair maintenance or service work may only be carried out by persons or companies being
properly authorized by us. If you do not know an authorized service station, please contact us and we will
name you our service partner closest to you. Otherwise, we will carry out the necessary work ourselves.

Date: November 2003

3. Unit description
The Alarm Box AB 01 has been developed for the transmission of alarm messages released by the
communication software APT-COM from BINDER, to an already existing alarm facility of the customer, or
by use of the telephone dialing device TWG 01 from BINDER via telephone line.
The Alarm BOX AB 01 is equipped with a serial interface RS 422 for connection to the communication
software APT-COM® from BINDER.

3.1 Equipment overview

(3) (2) (1)

Figure 2: Front view of Alarm Box AB 01

(1) Main switch ON-OFF (lit = ON)


(2) Switch alarm buzzer (0 = mute)
(3) Alarm pilot lamp red (lit = alarm state)

AB 01 04/2006 page 10/30


(4) (5) (6) (7)

Figure 3: Rear view of Alarm Box AB 01

(4) Mains cable


(5) Fuse holder
(6) DIN socket for zero-voltage relay alarm output
(7) Interface RS 422

3.2 Alarm output


The Alarm Box AB 01 is equipped at its rear with a zero-voltage relay alarm output via DIN socket (6) The
alarm output is switched by APT-COM® as soon as one or more supervised temperature chambers are
running with their actual value(s) out of the entered tolerance bandwidth.

2
3 1 Pin 1: Pin
2 4 3 Pin 2: Break
6
Pin 3: Make

1 5 If there is no alarm message, Pin 1 and 2 are closed.


The zero-voltage relay alarm output is switched with delay.

Output of Alarm Box AB 01


Pin allocation at the DIN socket
Logic NC

Wiring allocation of the delivered connection cable with DIN plug:


Pin 1 of DIN-plug ---- brown wire
Pin 2 of DIN-plug ---- white wire
Pin 3 of DIN-plug ---- not used

Maximum loading capacity of the switching contact: 24V AC/DC - 2,5A

DANGER
Electrical hazard.
Danger of life.
Damage of switching contact and connection socket.
∅ Do NOT exceed the maximum switching load of 24V AC/DC – 2.5A.
∅ Do NOT connect any devices with a higher loading capacity.

AB 01 04/2006 page 11/30


For connection to the software APT-COM® 2.0 see chap. 5.2 and the description of the
“Check“ window in the APT-COM® 2.0 GLP operating manual.

For connection to the software APT-COM® 3 or to the Watch Tool see chap. 5.3 and the
APT-COM® 3 operating manual., chap. 9 Watch-Tool.

3.3 Interface RS 422


The 25-pin SUB-D interface socket (7) of the Alarm Box AB 01 serves to receive the alarm message
released by APT-COM. The interface settings are fixed inside the unit and can not be changed.

The interface address for correct configuration of the Alarm Box on APT-COM window
“Configure“ (version 2.0 or higher) is fixed as No. 30.

4. Scope of delivery, transportation, storage, and installation


4.1 Unpacking, and checking equipment and scope of delivery
After having unpacked, please check the unit and its optional accessories, if any, based on the delivery
note for completeness and for transportation damage. If transportation damage has occurred,
immediately inform the carrier or post office/parcel service.
Scope of delivery:
• Alarm Box 210 mm x 85 mm x 260 mm (W x H x D)
• Connection cable for alarm output (2-wire) 2 m with DIN-plug
(connection between Alarm Box AB 01 and telephone dialing device TWG 01)
• Connection cable RS422 15m
(connection between plug distributor and Alarm Box AB 01)
• Operating Manual Alarm Box AB 01
For disposal of the transport packing, see chap. 10.1.

4.2 Storage
Intermediate storage of the unit is possible in a closed and dry room. Respect the advice for temporal
decommissioning (chap. 10.2).
• Permissible ambient temperature range: -10°C to +60°C.
• Permissible ambient humidity: max. 70 % r.H., non-condensing
If following storage in a cold location the unit is transferred to the installation site for start-up,
condensation is possible. Wait at least one hour until the oven has attained ambient temperature and is
completely dry.

4.3 Location of installation and ambient conditions


Select a convenient place for the Alarm Box AB 01. The place of the installation should:
• be close to the APT-COM® computer and the plug distributor
• be rather concealed or inaccessible

AB 01 04/2006 page 12/30


Set up the Alarm Box AB 01on a plane surface, free from vibration at a well-ventilated, dry location.

CAUTION
Danger of overheating.
Damage of the unit.
∅ Do NOT set up units in non-ventilated recesses.
¾ Ensure sufficient ventilation for carrying-off the heat.

• Permissible ambient temperature range: +18°C up to +32°C.

The ambient temperature should not be substantially higher than the indicated ambient
temperature of +25°C to which the specified technical data relate. In the case of different
ambient conditions, deviations from the indicated data are possible.

• Permissible ambient humidity: 70 % r.H. max., non-condensing.


• Installation height: max. 2000 m above sea level.

Note that unauthorized persons might switch off the device or the acoustic alarm signal. We
recommend installation e.g., inside a distribution box.

The unit must NOT be installed and operated in hazardous locations.

DANGER
Explosion hazard.
Danger of life.
∅ Do NOT operate the unit in potentially explosive areas.
∅ NO explosive dust or air-solvent mixture in the ambiance.

5. Installation of the equipment


5.1 Electrical connection
The Alarm Box AB 01 has a fixed connection cable, 1800 mm in length (4) and a shockproof plug:
• Mains fuse min. 10 Amp
• Power supply 230 Volt 1N +/- 10 %, 50-60 Hz
• Power consumption <20W
• Unit fuse: Fine fuse 5x20mm 10A (M) (5)
• Protection typeIP20
• Prior to connection and start-up, check the mains voltage. Compare the values to the data specified
on the type plate of the unit (chap. 1.3).
• When connecting, please observe the regulations specified by the local electricity supply company
and as well as the VDE directives (for Germany)
• Pollution degree (acc. to IEC 1010-1): 2
• Over-voltage category (acc. to IEC 1010-1): II

AB 01 04/2006 page 13/30


CAUTION
Danger of incorrect mains voltage.
Damage of the equipment.
¾ Check the mains voltage before connection and start-up.
¾ Compare the mains voltage to the data indicated on the type plate.

We recommend taking the electric power from a central alarm facility (as far as existing) with a backed-up
power feeding 230V AC (e.g. battery buffer for PCs). Only then the alarm transmission during a power
failure is guaranteed.
The power consumption of the unit is less than 20 W. Additionally for alarming during a power failure, it
is necessary that the computer on which APT-COM® 2.0 or the Watch Tool software is running is also
connected to a backed-up power feeding (e.g. battery buffer for PCs).

5.2 Connection to the communication software APT-COM® 2.0


The communication software APT-COM (version 2.0 or higher) generates the alarm message. The alarm
message is transmitted to the Alarm Box AB 01 via their RS422-interface.
The output side of the Alarm Box AB 01 is equipped with a switching contact which is switched in case of
an alarm event. This switching procedure can be used for reporting facilities or central alarm facilities as
the required input signal.
The telephone dialling unit TWG 01 from BINDER can be connected to the Alarm Box for remote
alarming via telephone line.
Also a signal buzzer or an alarm flasher can be switched directly by via this reporting contact.

29
MK

3
Connection cable 15 m KBF
Art. No. 5023-0036

BD
ED 2
FD

APT-Line (E2)

Cable set
Serial computer acc. to type of
interface 1
interface converter
COM1 or COM 2 CB

RS-232
...
RS-422 Telephone
Alarm
Box 30 dialing device
Computer Interface converter Plug distributor expansible by
APT-COM 2.0 Phoenix or W&T modules, Art. No. 8012-0295 Alarmbox AB 01

Figure 4: Sample system APT-COM® 2.01 GLP Edition

AB 01 04/2006 page 14/30


5.3 Connection to the communication software APT-COM® 3

Sample system APT-COM® 3 DataControlSystem GLP Edition with monitoring software “WatchTool"
optionally installed on a second computer:

Ethernet
RS-422
(optional)
Alarm Box AB 01

Serial computer Cable set with installation of monitoring software


interface "Watch Tool" on a second computer 29
acc. to type of
COM1 or COM 2 interface converter (monitoring computer) MK

RS-232 Alarm Telephone


RS-422 Box dialing device
30 3
Monitoring KBF
Interface converter
computer
Phoenix or W&T
Watch Tool
Connection cable 15 m
Art. No. 5023-0036
BD
2
Network connection

ED
FD

APT-Line (E2)

Cable set
Serial computer acc. to type of
interface 1
interface converter
COM1 or COM 2 CB

RS-232
...
RS-422 Telephone
Alarm
Box 30 dialing device
Measuring Interface converter Plug distributor expansible by
computer Phoenix or W&T modules, Art. No. 8012-0295 Alarmbox AB 01
APT-COM 3
with installation of monitoring
Ethernet software "Watch Tool" on the
same computer as APT-COM 3
RS-422
(optional)

Figure 5: Sample system APT-COM® 3 DataControlSystem GLP Edition

AB 01 04/2006 page 15/30


6. Start up
Establish all connections described in chap. 5 prior to commissioning.

6.1 Settings at the Alarm Box AB 01


Following steps have to be carried out to make the Alarm Box AB 01 ready for operation:
• Switch the Alarm Box ON at the green main switch (1)
• Set switch (2) to position I to activate the buzzer for acoustic alarm
The unit is now ready for operation and notification.

6.2 Settings in APT-COM® 2.01


• Announce the Alarm Box in the APT-COM configuration window „Configure“ with the COM-Port
number used and address No. 30 (fix setting of AB 01) and load the corresponding online window.
• Following loading the corresponding online-window in APT-COM go to menu „Function“ and select
„Active“.

6.2.1 Settings for remote alarm

Set tolerance bandwidth for the alarm


activation for each individual chamber

Activate limits for each individual


chamber
(position ON)

AB 01 04/2006 page 16/30


Setting of the “Update mode” must
remain “Online“.

After passing the delay time entered


under “Alarm delay” since notification of
the alarm relevant event in the check
window, the alarm is released via the
alarm output of the Alarm Box
AB 01.
(This setting is valid for all supervised
chambers)

As soon as the measured temperature value has been situated for the first time within the tolerance
limits, all alarm functions of APT-COM 2.01 are available for the monitored chamber.
For further instructions please see APT-COM V 2.01 operating manual.

6.3 Settings in APT-COM® 3 DataControlSystem

6.3.1 Connection of the Alarm Box AB 01 when operating APT-COM® 3 and the Watch
Tool on the same computer
Connecting the Alarm Box to APT-COM® 3
If APT-COM 3 and the Watch Tool are operated on a common computer, the Alarm Box AB 01 will be
connected to this computer. Then the Alarm Box AB 01 must be configured like one of the connected
temperature chambers in the configuration menu “Chamber“ of APT-COM 3. The Alarm Box is managed
and directly controlled by APT-COM 3.

AB 01 04/2006 page 17/30


Activating the Alarm Box in the Watch Tool
If the connection between the Watch Tool and APT-COM 3 has been established and both programs
have been called up (see APT-COM 3 operating manual), the Alarm Box AB 01 can be activated in the
Watch Tool configuration menu “Options”. The setting “APT-COM 3 PC (one PC)“ must have been
selected in the Watch Tool.
First the submenu “Config Binder Alarmbox (AB01)“ must be activated by switching on “Enable
Alarmbox“. Prior to this the program has no information about connection of the Alarm Box.

A few menu points can not be addressed even following activation of the submenu, e.g., „COM Port“ and
„Address“. This is because COM port and address are managed by APT-COM 3. Also the „Read test
check for the connection to the Alarm Box AB 01can not be used because the COM connection is
continually monitored by APT-COM 3.

6.3.2 Connection of the Alarm Box AB 01 when operating APT-COM® 3 and the Watch
Tool on different computers
Connecting the Alarm Box to the monitoring computer
The Alarm Box AB 01 is connected to the monitoring computer on which the monitoring software Watch
Tool is running. The Alarm Box can be placed in a distance of up to 50 m from the monitoring computer.

Activation and connection of the Alarm Box in the Watch Tool


If the connection between the Watch Tool and APT-COM 3 has been established and both programs
have been called up (see APT-COM 3 operating manual), the Alarm Box AB 01 can be activated and
managed in the Watch Tool configuration menu “Options”. The setting “Watch PC (two PCs)“ must have
been selected in the Watch Tool.
At first the submenu “Config Binder Alarmbox (AB01)“
must be activated by switching on “Enable Alarmbox“.
Prior to this, the program has no information about the
Alarm Box connection.

Under “COM Port” select the COM port to which the Alarm Box is connected, under “Address” enter 30,
the default Alarm Box address:

AB 01 04/2006 page 18/30


The button “Read Test” serves to check the connection between the Alarm Box and the Watch Tool.

If the signs “COM CRC Error” and “Time out“ remain green, the connection to the Alarm Box is OK and
functional. Under “Software version“ the answer-back code of the Alarm Box controller is displayed.

If one of the signs “COM CRC Error“ or “Time out“ or both of them light up red, the Alarm Box AB 01 is
not or not correctly connected or for another reason communication is not possible.
Only if this test has been successfully passed, the submenu “Alarmbox manual switch“ (switching on and
off the Alarm Box AB 01 for test purposes) is enabled (chap. 6.3.3).

6.3.3 Testing the Alarm Box AB 01: Function “Alarmbox manual switch”
This function serves to switch on and off the Alarm Box AB 01 for test purpose.

Hitting buttons “Switch On/Off” allows manually switching on and off the alarm pilot lamp of the Alarm
Box AB 01.

The state of the sign “Alarm pilot lamp“ corresponds to the state of the Alarm Box AB 01
alarm pilot lamp. This allows to control the switching state of the relay even if the Alarm Box
is not directly visible.

• Sign “Alarm pilot lamp“ red: alarm pilot lamp of the Alarm Box on.
• Sign “Alarm pilot lamp“ green: alarm pilot lamp of the Alarm Box off.
With setting “2 PCs“ the submenu “Alarmbox manual switch“ is enabled only if “Read test“ has
indicated that the connection to the Alarm Box AB 01 is functional (signs “COM CRC Error“ and “Time
out“ remaining green).

6.3.4 Checking the connection to the Alarm Box AB 01


On condition the options to access the Alarm Box AB 01 have been enabled with “Enable Alarmbox“ in
the „Options“ menu, Watch Tool sends a message every 60 sec. to the Alarm Box AB 01 in order to
check the connection. The result is indicated by the sign “Alarmbox status“:

AB 01 04/2006 page 19/30


Green sign: connection OK, Alarm Box functional:

Red sign: no connection, Alarm Box not functional:

Test: Following unplugging the Alarm Box the sign turns red after rundown of max. 2 min (max. 60 sec
plus a possible delay time due to reading access of the Watch Tools to the measuring data).
The interval for checking the Alarm Box is fix.
In case of the very first program start, following activating the option “Enable Alarmbox“ he indication is
green. After max. 2 minutes the real connection state to the Alarm Box AB 01 is displayed.
When disabling “Enable Alarmbox“ the indication “Alarmbox status“ disappears.

6.3.5 Setting the tolerance limits in the APT-COM® 3 measurement window


After measurement start the measurement window for the according measurement opens up.
The tolerance limit for the alarm can here be set individually for each parameter of the corresponding
measurement.

7. Alarm message
The alarm function of APT-COM is activated only if the supervised parameter previously has been within
the tolerance limits for at least one data sampling.

Any leaving of the tolerance bandwidth of one parameter of one of the supervised BINDER temperature
chambers leads to switching of the alarm output (chap. 3.2) to transmit the alarm message to the
telephone dialling device TWG 01 from BINDER or to another alarm facility of the customer.

The red alarm pilot lamp (3) indicates the alarm state. A buzzer sounds simultaneously which
can be muted with switch (2).

In APT-COM® 2.01 the alarm state is also shown on window “Check“ by red coloring of the
wide bar down on window “Check“.

In APT-COM® 3 the alarm state is indicated together with the further alarm specifications in
the Watch Tool window, as described below (chap. 7.3).

AB 01 04/2006 page 20/30


7.1 Buzzer
The alarm buzzer of the Alarm Box AB 01 can be activated or muted by switch (2) (0 = off).

7.2 Alarm course in APT-COM® 2.01


The latest measured value in window “Table“ is checked against the set tolerance
bandwidth, and any deviation is marked as LOW or HIGH. The “Check“ window
takes over the latest measured values with a frequency of actualization that
equals the longest registration interval of all of the temperature chambers from
which actually measured values are read out and stored. Directly following each
actualization of the “Check” window, the values are compared to the tolerance
bandwidth data individually set for each chamber. In case of leaving a tolerance
bandwidth, the alarm rectangle in the corresponding chamber window is colored
red. Now the set „Alarm delay“ time begins to run off. After run-off the wide bar
down on window “Check“ will be colored red as well. Now the moment has come
in which the remote alarm (collective alarm) is transmitted via the Alarm Box.

The red alarm pilot lamp (3) at the Alarm Box AB 01 indicates the alarm state, and the buzzer sounds.
The zero-voltage relay alarm output (chap. 3.2) is switched. The buzzer can be muted with switch (2).
The alarm state of the Alarm Box is finished automatically during the next update of window “Check” if
there is no parameter running outside its tolerance bandwidth any more.
The red alarm rectangle inside the supervising window of the corresponding chamber must be reset by
hand via double click (chap. 8.1). Thus the message that an error has occurred is conserved, even if the
chamber has meanwhile been able to regulate its values inside the tolerance bandwidth.

7.3 Alarm course in the Watch Tool (APT-COM® 3)


An alarm is indicated by the following signs located in the lower part of the Watch Tool window:

Figure 6: Indications of measurement alarm

The red collective alarm sign „Alarm“ indicates that at least at one of the chambers an error has
occurred.
In addition two yellow signs provide detailed information in addition to the collective red alarm sign
• The sign “Com alarm“ lighting up indicates that during the monitoring interval no new data have
been read.
• The sign “Limit alarm“ lighting up indicates that limit overrun or underrun at least at one of the
chambers has occurred.

For limit alarm detailed information is given in the Watch Tool window “Overview. The round pointer
displays indicate (in percent) whether the limits are respected by the monitored parameters. They
show that the last values read from the database are located inside the limit range set at APT-COM
3. A red or orange pointer color indicated that the limit for the corresponding parameter , has been
exceeded.
Limit Alarm is triggered beginning with a deviation of more than 100% from the set limit (red pointer
color), i.e. the yellow sign for limit alarm and the red collective alarm sign light up, and the Alarm Box
AB 01 responds.

AB 01 04/2006 page 21/30


Red: Deviation of the actual value from the set-point value is more than
100% of the set limit for this parameter in this measurement: limit overrun or
underrun. In this case, following rundown of the alarm delay time Limit
Alarm is triggered
Orange: Deviation of the actual value from the set-point value between
75% and 100% of the set limit for this parameter in this measurement.
Green: Deviation of the actual value from the set-point value up to 75% of
the set limit for this parameter in this measurement.

In case of an alarm, the Alarm Box AB 01 responds (on condition that it has been activated and correctly
connected), and an e-mail message is sent to the address indicated in the “Options“ menu (see for
details the APT-COM 3 operating manual, chap. 9).

8. Manual alarm reset


The alarm reset can must be manually carried out in APT-COM 2.0 or, when using APT-COM 3, in the
Watch Tool.

The alarm can be reset on condition that the alarm cause is no more persistent.

Note for users who have connected the telephone dialing device TWG 01 from BINDER:

The end of the alarm state at the APT-COM software does not influence telephone dialling
unit TWG 01. The alarm has to be reset there additionally by hand by pressing the reset key
. .

8.1 Alarm reset in APT-COM® 2.01


The alarm reset can only be carried out from APT-COM 2.0 window “Check“:
Hereto all red alarm rectangles have to be confirmed individually by a left click. Hereunto a automatic
alarm reset of the collective alarm (the red coloring of the wide bar down on window “Check“
automatically disappears) is performed – But only if all temperature chambers have been returned inside
their limits in the meantime. Otherwise all confirmed alarm rectangles get red again and the alarm state
remains.
8.2 Alarm reset in APT-COM® 3
To reset the alarm indications press button “Alarm reset“ twice:

The first pressing resets the cause indication signs “Com alarm“
or “Limit alarm“. c
The second pressing resets the red collective alarm sign.
d
A condition for alarm reset is that the alarm cause is no more persistent. Otherwise the alarm state is
conserved.

AB 01 04/2006 page 22/30


9. Cleaning, and service
9.1 Cleaning

DANGER
Electrical hazard.
Danger of life.
∅ Do NOT spill water or cleaning agents over the device.
¾ Put off-circuit the unit before cleaning. Pull the power plug.
¾ Completely dry the appliance before switching it on again.

Disconnect the chamber from the mains before cleaning. Pull the power plug!
Wipe the surfaces with a moistened towel. In addition, you can use the following cleaning agents:
• Standard commercial cleaning detergents free from acid or Halogenide and without alcohol.

• We recommend using the neutral cleaning agent Art. No. 1002-0016.

For surface protection, perform cleaning as fast as possible.


After cleaning completely remove cleaning agents from the surfaces with a moistened towel.

Never try to open the device in order to repair it. In case of problems please contact the BINDER service
at the indicated address (chap. 9.2).

CAUTION
Danger of corrosion.
Damage of the unit.
∅ Do NOT use acidic or chlorine cleaning detergents.

We recommend using the neutral cleaning agent Art. No. Art. Nr. 1002-0016 for a thorough
and mild cleaning.
Any corrosive damage that might arise following use of other cleaning agents is excluded from
liability by the BINDER GmbH.

9.2 Service

DANGER
Electrical hazard.
Danger of life.
∅ The unit must NOT become wet during operation.
∅ NEVER open the device for repair purpose.
¾ Have all maintenance work conducted by professional electricians or experts
authorized by BINDER.

AB 01 04/2006 page 23/30


Guarantee / repair and maintenance tasks must only be performed by persons or workshops
that have been authorized by us. Otherwise, BINDER will perform the tasks itself. If you do
not know whom to contact in the event of a fault, contact us and we will advise you of the
nearest authorized workshop

Never try to open the device in order to repair it. In case of problems please contact the BINDER service
at one of the indicated addresses:
BINDER telephone hotline: +49 (0) 7462 2005 555
BINDER fax hotline: +49 (0) 7462 2005 93555
BINDER e-mail hotline: service@binder-world.com
BINDER service hotline USA: +1 866 816 8191 (toll-free in the USA)
BINDER Asia Pacific: + 603 6204 2855
BINDER Internet homepage http://www.binder-world.com
BINDER address BINDER GmbH, post office box 102, D-78502 Tuttlingen
International customers, please contact your local BINDER distributor.

When calling us please provide following information:


• Unit type and serial number
• Date of purchase (see invoice / purchase slip)
• Kind of the fault / description of the defect
• Your full address and the installation location (building, department)
• Times at which we can contact you

10. Disposal
10.1 Disposal of the transport packing

Packing element Material Disposal


Transport box Cardboard Paper recycling
Foamed plastic stuffing PE foam Plastic recycling
Bag for operating manual PE foil Plastic recycling
Insulating air cushion foil PE foil Plastic recycling
If recycling is impossible, all packing parts can also be disposed of in the household waste.

10.2 Decommissioning
Switch off unit at the main switch (1). Disconnect the unit from the mains.
• Temporal decommissioning: See indications for appropriate storage, chap. 4.2.
• Final decommissioning: Dispose of the unit as described in chap. 10.3 to 10.5.

AB 01 04/2006 page 24/30


10.3 Disposal of the unit in the Federal Republic of Germany
According to directive 2002/96/EC of the European Parliament and of the Council on waste electrical and
electronic equipment (WEEE), BINDER devices are classified as “monitoring and control instruments”
(category 9) only intended for professional use“. They must not be disposed of at public collecting points.
The Alarm Box AB 01 bears the symbol for the marking of electrical and electronic
equipment manufactured / placed on the market in the EC after 13 August 2005 and be
disposed of in separate collection according to the directive 2002/96/EC on waste electrical
and electronic equipment (WEEE) and German national law for electrical and electronic
equipment (Elektro- und Elektronikgerätegesetz, ElektroG). WEEE marking: crossed-out
wheeled bin with solid bar under. An important part of the materials must be recycled in
order to protect the environment.
After the end of utilization have the device disposed of according to the German national
law for electrical and electronic equipment (Elektro- und Elektronikgerätegesetz, ElektroG)
from 23 March 2005, BGBl. I p. 762 or contact the BINDER service who will organize taking
back and disposal of the unit according to the German national law for electrical and
electronic equipment (Elektro- und Elektronikgerätegesetz, ElektroG) from 23 March 2005,
BGBl. I p. 762.

CAUTION
Violation against existing law.
∅ Do NOT dispose of BINDER devices at public collecting points.
¾ Have the device disposed of professionally at a recycling company which is certified
according to the German national law for electrical and electronic equipment (Elektro-
und Elektronikgerätegesetz, ElektroG) from 23 March 2005, BGBl. I p. 762.
or
¾ Instruct the BINDER Service to dispose of the device. The general terms of payment
and delivery of the BINDER GmbH apply, which were valid at the time of purchasing
the unit.

Certified companies disassemble waste BINDER equipment in primary substances for recycling
according to directive 2002/96/EC by. In order to exclude any health hazard for the employees of the
recycling companies, the devices must be free from toxic, infectious or radioactive substances.

It is the user’s responsibility that the unit is free from toxic, infectious or radioactive
substances prior to handing it over to a recycling company.
− Prior to disposal clean the unit from all introduced or sticking toxic substances.
− Prior to disposal disinfect the unit from all sources of infection. Be aware of the fact that
sources of infection might be located as well outside the inner chamber.
− If you cannot safely free the unit from toxic substances and sources of infection, dispose of
it as special waste according to national law.
− Fill out the contamination clearance certificate (chap. 12) and enclose it with the unit.

WARNING
Contamination of the device with toxic, infectious or radioactive substances.
Danger of intoxication.
Danger of infection.
∅ NEVER lead a unit with sticking toxic substances or sources of infection to recycling
according to directive 2002/96/EC.
¾ Prior to disposal, free the unit from sticking toxic substances or sources of infection.
¾ Dispose of a unit which you cannot safely free from all toxic substances or sources of
infection as special waste according to national law.

AB 01 04/2006 page 25/30


10.4 Disposal of the unit in the member states of the EC except for the Federal
Republic of Germany
According to directive 2002/96/EC of the European Parliament and of the Council on waste electrical and
electronic equipment (WEEE), BINDER devices are classified as “monitoring and control instruments”
(category 9) only intended for professional use“. They must not be disposed of at public collecting points.
The Alarm Box AB 01 bears the symbol for the marking of electrical and electronic
equipment manufactured / placed on the market in the EC after 13 August 2005 and be
disposed of in separate collection according to the directive 2002/96/EC on waste electrical
and electronic equipment (WEEE). WEEE marking: crossed-out wheeled bin with solid bar
under.
After the end of utilization, notify the distributor who sold you the device, who will take back
and dispose of the unit according to the directive 2002/96/EC of 27 January 2003 on waste
electrical and electronic equipment (WEEE).

CAUTION
Violation against existing law.
∅ Do NOT dispose of BINDER devices at public collecting points.
¾ Have the device disposed of professionally at a recycling company which is certified
according to conversion of the directive 2002/96/EC into national law.
or
¾ Instruct the distributor who sold you the device to dispose of it. The agreements apply
that were reached with the distributor when purchasing the unit (e.g. his general terms
of payment and delivery).
¾ If your distributor is not able to take back and dispose of the unit, please contact the
BINDER service.

Certified companies disassemble waste BINDER equipment in primary substances for recycling
according to directive 2002/96/EC by. In order to exclude any health hazard for the employees of the
recycling companies, the devices must be free from toxic, infectious or radioactive substances.

It is the user’s responsibility that the unit is free from toxic, infectious or radioactive
substances prior to handing it over to a recycling company.
− Prior to disposal clean the unit from all introduced or sticking toxic substances.
− Prior to disposal disinfect the unit from all sources of infection. Be aware of the fact that
sources of infection might be located as well outside the inner chamber.
− If you cannot safely free the unit from toxic substances and sources of infection, dispose of
it as special waste according to national law.
− Fill out the contamination clearance certificate (chap. 12) and enclose it with the unit.

WARNING
Contamination of the device with toxic, infectious or radioactive substances.
Danger of intoxication.
Danger of infection.
∅ NEVER lead a unit with sticking toxic substances or sources of infection to recycling
according to directive 2002/96/EC.
¾ Prior to disposal, free the unit from sticking toxic substances or sources of infection.
¾ Dispose of a unit which you cannot safely free from all toxic substances or sources of
infection as special waste according to national law.

AB 01 04/2006 page 26/30


10.5 Disposal of the unit in non-member states of the EC

CAUTION
Alteration of the environment.
¾ For final decommissioning and disposal of the oven, please contact the BINDER
service.
¾ Observe the regulations under public law for appropriate disposal protecting the
environment.

The main board of the oven includes a lithium cell. Please dispose of it according to the national
regulations.

11. Technical data

1 Electrical connection 230V ±10% / 1N / 50-60 Hz


2 Enclosure protection IP20 (acc. to EN 60529)
3 Protection against electric shock Class I
4 Installation category II
acc. to IEC 1010-1
5 Pollution degree 2
acc. to IEC 1010-1
6 Max. altitude of the installation site max. 2000 m above sea level
7 Unit fuse 5 x 20 mm 230V/10A semi time lag (M)
8 Zero-voltage relay alarm output 0.02 kW
9 Interface RS422 for connection of APT-COM 24 V, 2.5 A max.
software
10 External dimensions (w x h x d)
11 Electrical connection 205 mm x 87 mm x 270 mm

AB 01 04/2006 page 27/30


12. Contamination clearance certificate
Unbedenklichkeitsbescheinigung

Declaration of harmlessness with regard to safety and health


Erklärung zur Sicherheit and gesundheitlichen Unbedenklichkeit

The German Ordinance on Hazardous Substances (GefStofV), and the regulations regarding safety at
the workplace, require that this form be filled out for all products that are returned to us, so that the safety
and health of our employees can be warranted. In the absence of a completely filled out form, a repair is
not possible.
Die Sicherheit und Gesundheit unserer Mitarbeiter, die Gefahrstoffverordnung GefStofV und die Vorschriften zur
Sicherheit am Arbeitsplatz machen es erforderlich, dass dieses Formblatt für alle Produkte, die an uns
zurückgeschickt wird. Ohne Vorliegen des vollständig ausgefüllten Formblattes ist eine Reparatur nicht möglich.

• A completely filled out form should be transmitted by Fax (+49 (0) 7462 2005 93555) or by letter in
advance to us, so that this information is available before the equipment/component part arrives. A
second copy of this form should accompany the equipment/component part. Eventually the carrier
should be informed.
Eine vollständig ausgefüllte Kopie dieses Formblattes soll per Telefax (Nr. +49 (0) 7462 2005 93555) oder Brief
vorab an uns gesandt werden, so dass die Information vorliegt, bevor das Gerät/Bauteil eintrifft. Eine weitere
Kopie soll dem Gerät/Bauteil beigefügt sein. Ggf. ist auch die Spedition zu informieren.

• Incomplete information or non-conformity with this procedure will inevitably lead to substantial delays
in processing. We hope you will have understanding for this measure, which lies outside of our area
of influence, and that you will help us to speed up this procedure.
Unvollständige Angaben oder Nichteinhalten dieses Ablaufs führen zwangsläufig zu beträchtlichen
Verzögerungen in der Abwicklung. Bitte haben Sie Verständnis für Maßnahmen, die außerhalb unserer
Einflussmöglichkeiten liegen und helfen Sie mit, den Ablauf beschleunigen.

• Please fill out this form completely.


Bitte unbedingt vollständig ausfüllen!

1. Unit/ component part / type: / Gerät / Bauteil / Typ:

2. Serial No./ Serien-Nr.:

3. Details about utilized substances / biological substances / Einzelheiten über die eingesetzten
Substanzen/biologische Materialien:

3.1 Designations / Bezeichnungen:


a) ____________________________________________________________________________

b) ____________________________________________________________________________

c) ____________________________________________________________________________

3.2 Safety measures required for handling these substances / Vorsichtsmaßnahmen beim Umgang
mit diesen Stoffen:

a) ____________________________________________________________________________

b) ____________________________________________________________________________

c) ____________________________________________________________________________

AB 01 04/2006 page 28/30


3.3 Measures to be taken in case of skin contact or release into the atmosphere / Maßnahmen
bei Personenkontakt oder Freisetzung:

a) ____________________________________________________________________________

b) ____________________________________________________________________________

c) ____________________________________________________________________________

d) ____________________________________________________________________________

3.4 Other important information that must be taken into account / Weitere zu beachtende und
wichtige Informationen:

a) ____________________________________________________________________________

b) ____________________________________________________________________________

c) ____________________________________________________________________________

4. Declaration on the risk of these substances (please checkmark the applicable items) /
Erklärung zur Gefährlichkeit der Stoffe (bitte Zutreffendes ankreuzen) :
‰ 4.1 For non toxic, non radioactive, biologically harmless materials / für nicht giftige, nicht
radioaktive, biologisch ungefährliche Stoffe:

We herewith guarantee that the above-mentioned unit / component part… / Wir versichern, dass o.g.
Gerät/Bauteil...

‰ Has not been exposed to or contains any toxic or otherwise hazardous substances / weder giftige noch
sonstige gefährliche Stoffe enthält oder solche anhaften.

‰ That eventually generated reaction products are non-toxic and also do not represent a hazard / auch
evtl. entstandene Reaktionsprodukte weder giftig sind noch sonst eine Gefährdung darstellen.

‰ Eventual residues of hazardous substances have been removed / evtl. Rückstände von Gefahrstoffen entfernt
wurden.

‰ 4.2 For toxic, radioactive, biologically harmful or hazardous substances, or any other
hazardous materials / für giftige, radioaktive, biologisch bedenkliche bzw. gefährliche Stoffe oder
anderweitig gefährliche Stoffe.

We herewith guarantee that … / Wir versichern, dass …


‰ The hazardous substances, which have come into contact with the above-mentioned
equipment/component part, have been completely listed under item 3.1 and that all information in this
regard is complete / die gefährlichen Stoffe, die mit dem o.g. Gerät/Bauteil in Kontakt kamen, in 3.1 aufgelistet sind und
alle Angaben vollständig sind.

‰ That the unit /component part has not been in contact with radioactivity / das Gerät/Bauteil nicht mit
Radioaktivität in Berührung kam

5. Kind of transport / transporter / Transportweg/Spediteur:

Transport by (means and name of transport company, etc.) Versendung durch (Name Spediteur o.ä.)

___________________________________________________________________________________

Date of dispatch to BINDER GmbH / Tag der Absendung an BINDER GmbH:

___________________________________________________________________________________

AB 01 04/2006 page 29/30


We herewith declare that the following measures have been taken / Wir erklären, dass folgende
Maßnahmen getroffen wurden:

‰ Hazardous substances were removed from the unit / component part, so that no hazard exists for
corresponding persons in the handling or repair of these items / das Gerät/Bauteil wurde von Gefahrstoffen
befreit, so dass bei Handhabung/Reparaturen für die betreffenden Person keinerlei Gefährdung besteht

‰ The unit was securely packaged and properly identified / das Gerät wurde sicher verpackt und vollständig
gekennzeichnet.

‰ Information about the hazardousness of the shipment (if required) has been provided to the
transporter / der Spediteur wurde (falls vorgeschrieben) über die Gefährlichkeit der Sendung informiert.
We herewith commit ourselves and guarantee that we will indemnify BINDER GmbH for all damages
that are a consequence of incomplete or incorrect information provided by us, and that we will exempt
BINDER GmbH from eventual damage claims by third parties./ Wir versichern, dass wir gegenüber BINDER für
jeden Schaden, der durch unvollständige und unrichtige Angaben entsteht, haften und BINDER gegen eventuell entstehende
Schadenansprüche Dritter freistellen.

We are aware that, in accordance with Article 823 of the German Civil Code (BGB), we are directly liable
with regard to third parties, in this instance especially the employees of BINDER GmbH, who have been
entrusted with the handling / repair of the unit / component. / Es ist uns bekannt, dass wir gegenüber Dritten – hier
insbesondere mit der Handhabung/Reparatur des Geräts/des Bauteils betraute Mitarbeiter der Firma BINDER - gemäß §823 BGB
direkt haften

Name: _______________________________________________________________________

Position: _______________________________________________________________________

Date / Datum: _______________________________________________________________________

Signature / Unterschrift: _______________________________________________________________

Company stamp / Firmenstempel:

Equipment that is returned to the factory for repair must be accompanied by a completely filled
out contamination clearance certificate. For service and maintenance works on site, such a
contamination clearance certificate must be submitted to the service technician before the
start of the works. No repair or maintenance of the equipment is possible, without a properly
filled out contamination clearance certificate.

AB 01 04/2006 page 30/30