Sie sind auf Seite 1von 68

Druckschalter

Pressure Switch
Manocontacteur électronique

EDS 3400 / 3300

mit frontbündiger Membran


with flush membrane
avec membrane affleurante

Mat –Nr. 669873 / Stand: 28.06.2011


2 EDS 3400 / 3300 mit frontbündiger Membran

Stand: 28.06.2011 HYDAC ELECTRONIC GMBH Mat. Nr.: 669873


EDS 3400 / 3300 mit frontbündiger Membran 3

Inhalt
1 Allgemeines _________________________________________________ 5
2 Sicherheitshinweise _________________________________________ 5
3
4
Funktionen des EDS 3000____________________________________ 5
Montage ____________________________________________________ 6
D
5 Bedienung __________________________________________________ 7
6 Digitalanzeige _______________________________________________ 7
7 Ausgangsverhalten __________________________________________ 8
7.1 Schaltausgänge ______________________________________________________________________8
7.1.1 Einstellung auf Schaltpunkt (SP) _________________________________________________8
7.1.2 Einstellung auf Fensterfunktion (WIN) ___________________________________________8
7.2 Analogausgang ______________________________________________________________________9

7.3 Einstellen der Schaltpunkte und Hysteresen bzw. Schaltwerte für die
Fensterfunktion ___________________________________________________________________________9
7.4 Einstellbereiche für die Schaltausgänge _______________________________________10
7.4.1 Schaltpunkt-Funktion ____________________________________________________________10
7.4.2 Fenster-Funktion _________________________________________________________________10
8 Grundeinstellungen ________________________________________ 11
8.1 Ändern der Grundeinstellungen __________________________________________________11
8.2 Übersicht der Grundeinstellungen _______________________________________________12
9 Programmierfreigaben ______________________________________ 15
9.1 Ändern der Betriebs-Programmierfreigabe_____________________________________15
9.2 Ändern der Haupt-Programmierfreigabe ________________________________________15
10 Fehlermeldungen ___________________________________________ 16
11 Anschlussbelegung ________________________________________ 17
12 Technische Daten __________________________________________ 18
13 Typenschlüssel ____________________________________________ 19
13.1 Standard-Ausführung ____________________________________________________________19
13.2 DESINA®-konforme Ausführung bzw. an DESINA® anschließbar _________20
14 Zubehör____________________________________________________ 21
14.1 Für den elektrischen Anschluss _______________________________________________21
14.1.1 Für die Ausgangsvarianten „1“, „2“ und „3“ _________________________________21
14.1.2 Für alle Ausgangsvarianten und DESINA®-Ausführung _____________________21
14.2 Für den mechanischen Anschluss_____________________________________________22
15 Geräteabmessungen _______________________________________ 23

Stand: 28.06.2011 HYDAC ELECTRONIC GMBH Mat. Nr.: 669873


4 EDS 3400 / 3300 mit frontbündiger Membran

Vorwort

D
Für Sie, den Benutzer unseres Produktes, haben wir in dieser
Dokumentation die wichtigsten Hinweise zum Bedienen und
Warten zusammengestellt.

Sie dient Ihnen dazu, das Produkt kennen zu lernen und seine
bestimmungsgemäßen Einsatzmöglichkeiten optimal zu nutzen.

Diese Dokumentation muss ständig am Einsatzort verfügbar


sein.
Bitte beachten Sie, dass die in dieser Dokumentation
gemachten Angaben der Gerätetechnik zu dem Zeitpunkt der
Literaturerstellung entsprechen. Abweichungen bei technischen
Angaben, Abbildungen und Maßen sind deshalb möglich.

Entdecken Sie beim Lesen dieser Dokumentation Fehler oder


haben weitere Anregungen und Hinweise, so wenden Sie sich
bitte an:

HYDAC ELECTRONIC GMBH


Technische Dokumentation
Hauptstraße 27
66128 Saarbrücken
-Deutschland-
Tel: +49(0)6897 / 509-01
Fax: +49(0)6897 / 509-1726
Email: electronic@hydac.com

Die Redaktion freut sich über Ihre Mitarbeit.

„Aus der Praxis für die Praxis“

Stand: 28.06.2011 HYDAC ELECTRONIC GMBH Mat. Nr.: 669873


EDS 3400 / 3300 mit frontbündiger Membran 5

1 Allgemeines
Die Druckschalter EDS 3000 mit frontbündiger Membran wurden speziell für Applikationen

D
entwickelt, in denen die verwendeten Medien zu einem Verstopfen, Verkleben oder
Einfrieren eines Standard-Druckanschlusses führen könnten.
Weitere Anwendungsfelder finden sich in Prozessen mit wechselnden Medien, in denen
Rückstände zu Vermischung oder Verunreinigung der Medien führen könnte.

Der Druckanschluss ist frontseitig durch eine voll verschweißte Edelstahlmembran bündig
abgeschlossen und intern mit einer Druckmittlerflüssigkeit gefüllt, welche den Prozessdruck
hydrostatisch zur Messzelle überträgt.

Falls Sie Fragen bezüglich der technischen Daten oder Eignung für Ihre Anwendungen
haben, wenden Sie sich bitte an unseren technischen Vertrieb.
Die Druckschalter der Serie EDS 3000 werden einzeln auf einem rechnergesteuerten
Prüfplatz abgeglichen und einem Endtest unterzogen. Sie sind wartungsfrei und sollten
beim Einsatz innerhalb der Spezifikationen (siehe Technische Daten) einwandfrei arbeiten.
Falls trotzdem Fehler auftreten sollten, wenden Sie sich bitte an den HYDAC–Service.
Fremdeingriffe in das Gerät führen zum Erlöschen jeglicher Gewährleistungsansprüche.

2 Sicherheitshinweise
Überprüfen Sie vor der Inbetriebnahme den Zustand des Gerätes sowie des mitgelieferten
Zubehörs. Lesen Sie vor der Inbetriebnahme des Gerätes die Bedienanleitung und stellen
Sie sicher, dass das Gerät für Ihre Anwendung geeignet ist.

Falsche Handhabung bzw. die Nichteinhaltung von Gebrauchshinweisen oder technischen


Angaben kann zu Sach- und / oder Personenschäden führen.

Es ist unbedingt auf die Verträglichkeit der Messmedien mit den verwendeten Werkstoffen
des Druckschalters zu achten; ebenso sind die Überlast- und Berstdrücke einzuhalten

3 Funktionen des EDS 3000


Je nach Ausführung bietet das Gerät folgende Funktionen:
• Messwertanzeige des aktuellen Druckes in PSI, MPa, bar oder einem frei skalierbaren
Bereich.
• Anzeige des Maximalwertes oder eines eingestellten Schaltpunktes.
• Schalten der Schaltausgänge entsprechend dem Druck und den eingestellten
Schaltparametern.
• Analogausgang
• Menü zur Grundeinstellung (Anpassen des EDS 3000 an die jeweilige Applikation).
• Programmierfreigaben

Stand: 28.06.2011 HYDAC ELECTRONIC GMBH Mat. Nr.: 669873


6 EDS 3400 / 3300 mit frontbündiger Membran

4 Montage

D
Der EDS 3000 kann über den Druckanschluss direkt an einen Hydraulikblock montiert
werden.
Hierbei ist zu beachten, dass die frontbündige (außenliegende) Membran vor
mechanischer Beschädigung geschützt wird.
Zur optimalen Ausrichtung ist eine Verdrehung um 340° in der Längsachse, sowie um 270°
im Display inkl. der Bedientasten möglich.
Das Anzugsdrehmoment beträgt für beide G 1/2 Gewindeanschlüsse 45 Nm.
Der elektrische Anschluss sollte von einem Fachmann nach den jeweiligen
Landesvorschriften durchgeführt werden (VDE 0100 in Deutschland). Das
Druckschaltergehäuse ist dabei ordnungsgemäß zu erden.

Die Druckschalter der Serie EDS 3000 tragen das – Zeichen. Eine Konformitäts-
erklärung ist auf Anfrage erhältlich.
Es gelten die EMV–Normen: EN 61000-6-1, -2, -3 und -4.
Die Forderungen der Normen werden nur bei ordnungsgemäßer und fachmännischer
Erdung des Druckschaltergehäuses erreicht. Beim Einschrauben in einen Hydraulikblock
ist es ausreichend, wenn der Block über das Hydrauliksystem geerdet ist.

ACHTUNG:
Das Einschrauben des EDS 3000 muss mit einem passenden Maulschlüssel
(Schlüsselweite 27) am Sechskant des Druckanschlusses erfolgen.
Eine unsachgemäße Montage, wie z. B. durch manuelles Eindrehen über das
Gehäuse, kann aufgrund der Verdrehbarkeit des EDS 3000 zu Beschädigungen
am Gehäuse, bis hin zum vollständigen Ausfall des Gerätes führen.

Zusätzliche Montagehinweise, die erfahrungsgemäß den Einfluss elektromagnetischer


Störungen reduzieren:
• Möglichst kurze Leitungsverbindungen herstellen
• Leitungen mit Schirm verwenden (z.B. LIYCY 4 x 0,5 mm²)
• Der Kabelschirm ist in Abhängigkeit der Umgebungsbedingungen fachmännisch und zum
Zweck der Störunterdrückung einzusetzen
• Direkte Nähe zu Verbindungsleitungen von Leistungsverbrauchern oder störenden
Elektro- oder Elektronikgeräten ist möglichst zu vermeiden

Stand: 28.06.2011 HYDAC ELECTRONIC GMBH Mat. Nr.: 669873


EDS 3400 / 3300 mit frontbündiger Membran 7

5 Bedienung

6 Digitalanzeige
Nach Einschalten der Versorgungsspannung zeigt das Gerät kurz "EdS" an und
beginnt mit der Anzeige des aktuellen Druckes.

In den Grundeinstellungen kann die Anzeige wie folgt geändert werden:


• Anzeige des Maximalwertes "TOP"
Der höchste im System gemessene Druck seit dem Einschalten des Gerätes bzw. dem
letzten Rücksetzvorgang wird permanent angezeigt.
• Anzeige des eingestellten Schaltpunktes "S.P. 1", "S.P. 2"
Je nach Variante kann der Schaltpunkt 1 bzw. Schaltpunkt 2 permanent angezeigt
werden.
• Anzeige dunkel "OFF"
Das Display ist ausgeschaltet. Je nach Einstellung erscheint nach der Einschaltmeldung
kurz "TOP", "S.P. 1", "S.P. 2" oder "OFF" in der Anzeige. Der aktuelle Druck kann
kurzzeitig zur Anzeige gebracht werden, indem die Taste ◄ oder ► betätigt wird.
• Bei der Einstellung "Anzeige des aktuellen Drucks" besteht die Möglichkeit, durch
Betätigen der Taste ◄ oder ► den Maximalwert anzuzeigen. Dieser wird dann für 3s
dargestellt. Der Maximalwert kann in jedem Anzeigemodus durch kurzes, gleichzeitiges
Drücken der ◄ oder ► Taste zurückgesetzt werden. Dieser Vorgang wird durch die
Meldung "rES" im Display quittiert.

• Übersteigt der aktuelle Druck den Nenndruck des Gerätes, so kann er nicht mehr
angezeigt werden und die Anzeige beginnt zu blinken.
• Liegt der aktuelle Druck unterhalb 0,75 % des Nennbereiches, so wird 0 bar
angezeigt.

Stand: 28.06.2011 HYDAC ELECTRONIC GMBH Mat. Nr.: 669873


8 EDS 3400 / 3300 mit frontbündiger Membran

7 Ausgangsverhalten
7.1 Schaltausgänge

D
Der EDS 3000 verfügt über 1 bzw. 2 Schaltausgänge. In den Grundeinstellungen kann
folgendes Schaltverhalten eingestellt werden:

7.1.1 Einstellung auf Schaltpunkt (SP)


Zu jedem Schaltausgang kann ein Schaltpunkt und eine Hysterese eingestellt werden. Der
jeweilige Ausgang schaltet, wenn der eingestellte Schaltpunkt erreicht wurde und schaltet
zurück, wenn der Rückschaltpunkt unterschritten wurde. Der Rückschaltpunkt wird durch
die eingestellte Hysterese bestimmt (Rückschaltpunkt = Schaltpunkt minus Hysterese).

Abkürzungen: "S.P.1", "S.P.2" = Schaltpunkt 1 bzw. 2


"H.Y.1", "H.Y.2" = Hysterese 1 bzw. 2

7.1.2 Einstellung auf Fensterfunktion (WIN)


Die Fensterfunktion ermöglicht es, einen Bereich zu überwachen. Zu jedem Schaltausgang
können jeweils ein oberer und ein unterer Schaltwert eingegeben werden, die den Bereich
bestimmen.
Der jeweilige Ausgang schaltet, wenn der Druck in diesen Bereich eintritt. Bei Verlassen
des Bereiches, d.h. wenn der Rückschaltwert erreicht ist, schaltet der Ausgang zurück.
Der untere Rückschaltwert liegt knapp unter dem unteren Schaltwert (unterer Schaltwert
minus 3-fache Schrittweite, siehe Kapitel 7.4).
Der obere Rückschaltwert liegt knapp über dem oberen Schaltwert (oberer Schaltwert plus
3-fache Schrittweite, siehe Kapitel 7.4).
Der Bereich zwischen Schalt- und Rückschaltwert bildet eine Sicherheitszone, die
verhindert, dass unerwünschte Schaltvorgänge erfolgen (z.B. ausgelöst durch Pulsationen
einer Pumpe).

Beispiel für Schaltausgang 1 (Schließerfunktion):

Abkürzungen: "hi.1", "hi.2" = High level 1 bzw. 2 = oberer Schaltwert 1 bzw. 2


"Lo.1", "Lo.2" = Low level 1 bzw. 2 = unterer Schaltwert 1 bzw. 2
“FS“ (Full Scale) = bezogen auf den vollen Messbereich

Die Fensterfunktion arbeitet nur dann ordnungsgemäß (Ein- und Ausschalten),


wenn alle Schaltwerte (inklusive Sicherheitszone) größer als 0 bar und kleiner als
der Nenndruckbereich sind.

Stand: 28.06.2011 HYDAC ELECTRONIC GMBH Mat. Nr.: 669873


EDS 3400 / 3300 mit frontbündiger Membran 9

7.2 Analogausgang
Analogausgangssignal: 4 .. 20 mA oder 0 .. 10 V
(Einstellbar im Menü "Grundeinstellungen").

7.3 Einstellen der Schaltpunkte und Hysteresen bzw. D


Schaltwerte für die Fensterfunktion
• Taste "mode" betätigen.
• In der Anzeige erscheint "S.P.1" bzw. "hi.1"
• Durch weiteres Betätigen der Taste "mode" den gewünschten Parameter anwählen
(bei eingestellter Schaltpunktfunktion erscheint: "S.P.1", "hYS.1", "S.P.2", "hYS.2" ; bei
eingestellter Fensterfunktion (WIN ) erscheint "Hi.1", "Lo.1", "Hi.2" oder "Lo.2").
• Nach 2 Sekunden blinkt die aktuelle Einstellung.
• Mit den T asten ◄ oder ► die Einstellung ändern.
• Eventuell mit der Taste "mode" weitere Parameter anwählen und mit den
Tasten ◄ oder ► die Einstellung ändern.
• Nach 3 Sekunden ohne Tastenbetätigung schaltet die Anzeige zurück, die Einstellungen
werden gespeichert.

Hinweise:
• Erscheint beim Einstellversuch "LOC" in der Anzeige, ist die Programmierung
gesperrt.
Abhilfe: Programmierfreigabe(n) auf "FREE" setzen.
(siehe Kapitel 9 "Programmierfreigaben")
• Wird beim Ändern die Taste ◄ oder ► festgehalten, wird der Wert automatisch
weitergezählt.
• Wenn eine Einstellung geändert wurde, erscheint beim Umschalten der Anzeige
kurz "PROG" in der Anzeige. Die neue Einstellung wurde dann im Gerät
gespeichert.

Stand: 28.06.2011 HYDAC ELECTRONIC GMBH Mat. Nr.: 669873


10 EDS 3400 / 3300 mit frontbündiger Membran

7.4 Einstellbereiche für die Schaltausgänge


7.4.1 Schaltpunkt-Funktion

7.4.2 Fenster-Funktion

* Alle in den Tabellen angegebenen Bereiche sind im Raster der Schrittweite einstellbar.

Stand: 28.06.2011 HYDAC ELECTRONIC GMBH Mat. Nr.: 669873


EDS 3400 / 3300 mit frontbündiger Membran 11

8 Grundeinstellungen
Zur Anpassung an die jeweilige Applikation kann das Verhalten des EDS 3000 über

D
mehrere Grundeinstellungen verändert werden.
Diese sind zu einem Menü zusammengefasst.

8.1 Ändern der Grundeinstellungen

Wichtiger Hinweis:
Bei aktiviertem Menü werden keine Schaltfunktionen ausgeführt!

Aktivieren des Grundeinstellungsmenüs:

Beenden des Grundeinstellungsmenüs:

Den Menüpunkt "END" anwählen, die Einstellung auf "YES" stellen, der EDS 3000 kehrt
nach 2 s in den normalen Anzeigemodus zurück.

Hinweis:
Erfolgt ca. 25 Sekunden lang keine Tastenbetätigung, wird das Menü
automatisch beendet, ohne dass eventuelle Änderungen wirksam werden.

Stand: 28.06.2011 HYDAC ELECTRONIC GMBH Mat. Nr.: 669873


12 EDS 3400 / 3300 mit frontbündiger Membran

8.2 Übersicht der Grundeinstellungen

Stand: 28.06.2011 HYDAC ELECTRONIC GMBH Mat. Nr.: 669873


EDS 3400 / 3300 mit frontbündiger Membran 13

Stand: 28.06.2011 HYDAC ELECTRONIC GMBH Mat. Nr.: 669873


14 EDS 3400 / 3300 mit frontbündiger Membran

Achtung:
Nach einem Nullpunktabgleich wird z. B. bei einem 600 bar Gerät ein Druck von
bis zu 18 bar als 0 bar angezeigt. Vor Arbeiten an der Hydraulikanlage muss
sichergestellt werden, dass diese drucklos ist.

Sind die Grundeinstellungen verändert worden, so erscheint beim Verlassen des


Grundeinstellungsmenüs für einen kurzen Moment die Meldung "ProG" im Display und
danach wird der in der Primäranzeige gewählte Anzeigewert dargestellt.

Stand: 28.06.2011 HYDAC ELECTRONIC GMBH Mat. Nr.: 669873


EDS 3400 / 3300 mit frontbündiger Membran 15

9 Programmierfreigaben
Das Gerät verfügt über 2 Programmierfreigaben, die beide erteilt sein müssen, um

D
Einstellungen zu ändern. Die Betriebs-Programmierfreigabe kann während des Betriebes
gesetzt bzw. aufgehoben werden. Sie bietet Schutz vor unbeabsichtigten Änderungen.
Ein Sperren der Programmierung über die Haupt-Programmierfreigabe bewirkt, dass
während des Betriebes keine Änderung der Einstellungen vorgenommen werden kann.
Dies dient z.B. als Sicherheitsfunktion oder als Schutz vor unerlaubten Änderungen.

9.1 Ändern der Betriebs-Programmierfreigabe

9.2 Ändern der Haupt-Programmierfreigabe


Versorgungsspannung abschalten oder Gerät von der Versorgungsspannung trennen.

Hinweis:
Wenn eine Einstellung geändert wurde, erscheint beim Umschalten der
Anzeigekurz "PRoG" in der Anzeige. Die neue Einstellung wurde dann im Gerät
gespeichert.

Stand: 28.06.2011 HYDAC ELECTRONIC GMBH Mat. Nr.: 669873


16 EDS 3400 / 3300 mit frontbündiger Membran

10 Fehlermeldungen
Wird ein Fehler erkannt, so erscheint eine entsprechende Fehlermeldung, die mit einem

D
beliebigen Tastendruck quittiert werden muss. Mögliche Fehlermeldungen sind:

Stand: 28.06.2011 HYDAC ELECTRONIC GMBH Mat. Nr.: 669873


EDS 3400 / 3300 mit frontbündiger Membran 17

11 Anschlussbelegung

Stand: 28.06.2011 HYDAC ELECTRONIC GMBH Mat. Nr.: 669873


18 EDS 3400 / 3300 mit frontbündiger Membran

12 Technische Daten
Eingangsgrößen EDS 3300 EDS 3400

D
Messbereiche in bar -1..1; 1; 2,5; 6; 10; 16 40; 100; 250; 400; 600
Überlastbereiche in bar 3; 3; 8; 18; 30; 48 80; 200; 500; 800; 900
Berstdruck 1) in bar 5; 5; 15; 30; 50; 80 200; 500; 1000; 2000; 2000
Mechanischer Anschluss G1/2A DIN 3852
G1/2 mit zus. frontseitiger O-Ring-Dichtung
Druckmittlerflüssigkeit Silikon freies Öl
Anzugsdrehmoment 45 Nm
Medienberührende Teile 2) Anschlussstück: Edelstahl Anschlussstück: Edelstahl
(Messzelle: Keramik) (Messzelle: Edelstahl)
Dichtung: FPM Dichtung: FPM
O-Ring: NBR O-Ring: NBR
Ausgangsgrößen
Genauigkeit nach DIN 16086, ≤ ± 0,5 % FS typ.
Grenzpunkteinstellung ≤ ± 1,0 % FS max
(Anzeige, Analogausgang)
Reproduzierbarkeit ≤ ± 0,25 % FS max.
Temperaturdrift ≤ ± 0,025 % FS / °C max. Nullpunkt
≤ ± 0,025 % FS / °C max. Spanne
Analogausgang (optional)
Ausgangssignal (wählbar) 4 .. 20 mA Bürde max. 500 Ω
0 .. 10 V Bürde min. 1 kΩ
Schaltausgänge
Ausführung PNP Transistorschaltausgang
Schaltstrom max. 1,2 A je Ausgang
Schaltzyklen > 100 Millionen
Reaktionszeit ~ 10 ms
Langzeitdrift ≤ ± 0,3 % FS / Jahr typ.
DESINA®-Diagnosesignal (Pin 2)
Funktion I.O.: HIGH Pegel / N.I.O.: LOW Pegel
Pegel HIGH: ~ +UB / LOW: < +0,3 V
Umgebungsbedingungen
Kompensierter Temperaturbereich -10 .. +70 °C
Betriebstemperaturbereich 3) -25 .. +80 °C
Lagertemperaturbereich -40 .. +80 °C
Mediumstemperaturbereich -40 .. +80 °C
- Zeichen EN 61000-6-1 / 2 / 3 / 4
Vibrationsfestigkeit nach ≤ 10 g
DIN EN 60068-2-6 bei 10 .. 500 Hz
Schockfestigkeit nach ≤ 50 g
DIN EN 60068-2-29 (11 ms)
Schutzart nach DIN 40050 IP 67
Sonstige Größen
Versorgungsspannung 9 .. 35 V DC ohne Analogausgang
18 .. 35 V DC mit Analogausgang
Restwelligkeit der ≤5%
Versorgungsspannung
Stromaufnahme max. 35 mA (inaktiver Schaltausgang)
Anzeige 4-stellig, LED, 7-Segment, rot, Zeichenhöhe 7 mm
Gewicht ~ 120 g
1)
Bei G1/2 mit zusätzlicher frontseitiger O-Ring-Dichtung max. 1500 bar
2)
Andere Dichtungsmaterialien auf Anfrage
3)
Bei G1/2 mit zusätzlicher frontseitiger O-Ring-Dichtung -10 .. +80 °C

Stand: 28.06.2011 HYDAC ELECTRONIC GMBH Mat. Nr.: 669873


EDS 3400 / 3300 mit frontbündiger Membran 19

13 Typenschlüssel

D
13.1 Standard-Ausführung
EDS 3 X Z X – X – XXXX – XXX – 000

Genauigkeit
3 = 1 % FS max., Keramik (Dickschicht), relativ
4 = 1 % FS max., Edelstahl (Dünnfilm), relativ

Prozessanschluss mechanisch
Z = Frontbündig

Anschlussart elektrisch
6 = Gerätestecker M12x1, 4-pol.
Nur für Ausgangsvariante „1“, „2“ und „3“
8 = Gerätestecker M12x1, 5-pol.
Nur für Ausgangsvariante „5“

Ausgang
1 = 1 Schaltausgang
Nur in Verbindung mit elektrischer Anschlussart „6“
2 = 2 Schaltausgänge
Nur in Verbindung mit elektrischer Anschlussart „6“
3 = 1 Schaltausgang und 1 Analogausgang
Nur in Verbindung mit elektrischer Anschlussart „6“
5 = 2 Schaltausgänge und 1 Analogausgang
Nur in Verbindung mit elektrischer Anschlussart „8“

Druckbereiche in bar
0001 (-1..1)
01,0; 02,5; 06,0; 0010; 0016
Nur für Genauigkeit „3“

0040; 0100; 0250; 0400; 0600


Nur für Genauigkeit „4“

Anschlussart mechanisch
G01 = G1/2 A, DIN 3852
G02 = G1/2 mit zusätzlicher frontseitiger O-Ring-Dichtung

Modifikationsnummer
000 = Standard

Anmerkungen:
Sonderausführungen auf Anfrage.
Bei Geräten mit anderer Modifikationsnummer ist das Typenschild bzw. die mitgelieferte technische
Änderungsbeschreibung zu beachten.

Stand: 28.06.2011 HYDAC ELECTRONIC GMBH Mat. Nr.: 669873


20 EDS 3400 / 3300 mit frontbündiger Membran

D
13.2 DESINA®-konforme Ausführung bzw. an DESINA® anschließbar
EDS 3 X Z X – X – XXXX – XXX – D00

Genauigkeit
3 = 1 % FS max., Keramik (Dickschicht), relativ
4 = 1 % FS max., Edelstahl (Dünnfilm), relativ

Prozessanschluss mechanisch
Z = Frontbündig

Anschlussart elektrisch
8 = Gerätestecker M12x1, 5-pol.

Ausgang
1 = 1 Schaltausgang
Nur in Verbindung mit elektrischer Anschlussart „6“
3 = 1 Schaltausgang und 1 Analogausgang
Nur in Verbindung mit elektrischer Anschlussart „6“

Druckbereiche in bar
0001 (-1..1)
01,0; 02,5; 06,0; 0010; 0016
Nur für Genauigkeit „3“

0040; 0100; 0250; 0400; 0600


Nur für Genauigkeit „4“

Anschlussart mechanisch
G01 = G1/2 A, DIN 3852
G02 = G1/2 mit zusätzlicher frontseitiger O-Ring-Dichtung

Modifikationsnummer
D00 = DESINA®-konforme Anschlussbelegung mit Eigendiagnose

Anmerkungen:
Sonderausführungen auf Anfrage.
Bei Geräten mit anderer Modifikationsnummer ist das Typenschild bzw. die mitgelieferte technische
Änderungsbeschreibung zu beachten.

Stand: 28.06.2011 HYDAC ELECTRONIC GMBH Mat. Nr.: 669873


EDS 3400 / 3300 mit frontbündiger Membran 21

14 Zubehör
14.1 Für den elektrischen Anschluss
14.1.1 Für die Ausgangsvarianten „1“, „2“ und „3“
D

14.1.2 Für alle Ausgangsvarianten und DESINA®-Ausführung

Stand: 28.06.2011 HYDAC ELECTRONIC GMBH Mat. Nr.: 669873


22 EDS 3400 / 3300 mit frontbündiger Membran

14.2 Für den mechanischen Anschluss

Stand: 28.06.2011 HYDAC ELECTRONIC GMBH Mat. Nr.: 669873


EDS 3400 / 3300 mit frontbündiger Membran 23

15 Geräteabmessungen

Stand: 28.06.2011 HYDAC ELECTRONIC GMBH Mat. Nr.: 669873


24 EDS 3400 / 3300 mit frontbündiger Membran

HYDAC ELECTRONIC GMBH

D
Hauptstr. 27
D-66128 Saarbrücken
Germany

Web: www.hydac.com
E-Mail: electronic@hydac.com
Tel.: +49 (0)6897 509-01
Fax.: +49 (0)6897 509-1726

HYDAC Service
Für Fragen zu Reparaturen steht Ihnen der HYDAC Service zur Verfügung.

HYDAC SERVICE GMBH


Hauptstr. 27
D-66128 Saarbrücken
Germany

Tel.: +49 (0)6897 509-1936


Fax.: +49 (0)6897 509-1933

Anmerkung

Die Angaben in dieser Bedienungsanleitung beziehen sich auf die beschriebenen


Betriebsbedingungen und Einsatzfälle. Bei abweichenden Einsatzfällen und/oder
Betriebsbedingungen wenden Sie sich bitte an die entsprechende Fachabteilung.

Bei technischen Fragen, Hinweisen oder Störungen nehmen Sie bitte Kontakt mit Ihrer
HYDAC-Vertretung auf.

Technische Änderungen sind vorbehalten.

Stand: 28.06.2011 HYDAC ELECTRONIC GMBH Mat. Nr.: 669873


EDS 3400 / 3300 with flush membrane 3

Contents
1 General ____________________________________________________ 5
2 Safety Instructions__________________________________________ 5
3 Functions of the EDS 3000 __________________________________ 5
4 Installation _________________________________________________ 6
5 Operation __________________________________________________ 7
6 Digital Display ______________________________________________ 7
7 Output Signals _____________________________________________ 8
7.1 Switching Outputs _________________________________________________________________ 8
7.1.1 Switch Point Setting (SP) _______________________________________________________ 8
E
7.1.2 Window Mode (WIN) _____________________________________________________________ 8

7.2 Analogue Output ___________________________________________________________________ 9


7.3 Setting the Switching Points and Hysteresis or Switching Values for
Window Mode ____________________________________________________________________________ 9
7.4 Setting Ranges for the Switch Outputs _______________________________________ 10
7.4.1 Switching Point Function ______________________________________________________ 10
7.4.2 Window Function _______________________________________________________________ 10

8 Basic Settings _____________________________________________ 11


8.1 Changing the Basic Settings____________________________________________________ 11
8.2 Overview of the Basic Settings _________________________________________________ 12
9 Programming Enables _____________________________________ 15
9.1 Changing the Main Programming Enable_____________________________________ 15
9.2 Changing the Main Programming Enable_____________________________________ 15
10 Error Messages ___________________________________________ 16
11 Pin Assignment ___________________________________________ 17
12 Technical Data:____________________________________________ 18
13 Model Code _______________________________________________ 19
13.1 Standard Version _______________________________________________________________ 19

13.2 DESINA® Compliant Type or Compatible with DESINA®: ________________ 20


14 Accessoires _______________________________________________ 21
14.1 For Electrical Connection _____________________________________________________ 21
14.1.1 For Use with Output Options "1", "2" and "3" ______________________________ 21
14.1.2 For all Output Models and DESINA® Version ______________________________ 21

14.2 For Mechanical Connection ___________________________________________________ 22


15 Dimensions _______________________________________________ 23

Status 28.06.2011 HYDAC ELECTRONIC GMBH Part no.: 669873


4 EDS 3400 / 3300 with flush membrane

Preface

This manual provides you, as user of our product, with key


information on the operation and maintenance of the
equipment.

It will acquaint you with the product and assist you in obtaining
maximum benefit in the applications for which it is designed.

E Always keep the manual with the device for immediate


reference.
Please note: the specifications outlined in this documentation
for the instrument technology are correct at the time of
publishing. Deviations in technical specifications, illustrations
and dimensions are therefore possible.

If you discover errors while reading the documentation or have


additional suggestions or notes, contact us at:

HYDAC ELECTRONIC GMBH


Technical Documentation
Hauptstraße 27
66128 Saarbrücken
-Germany-
Phone: +49(0)6897 / 509-01
Fax.: +49(0)6897 / 509-1726
Email: electronic@hydac.com

The editorial team looks forward to hearing from you.

„Putting Experience into Practice“

Status 28.06.2011 HYDAC ELECTRONIC GMBH Part no.: 669873


EDS 3400 / 3300 with flush membrane 5

1 General
Pressure transmitter EDS 3000 with a flush membrane was designed specifically for
applications in which a standard pressure connection could become blocked, clogged or
frozen by the particular medium used.
Additional fields of application can be found in processes with changing media, in which
residues could lead to mixing or contamination of the media.

The pressure connection is achieved with a fully-sealed stainless steel front membrane
filled internally with a pressure transfer fluid. The process pressure is transmitted

E
hydrostatically to the measurement cell via the pressure transfer fluid.

If you have any queries regarding technical details or the suitability of the unit for your
application, please contact our Technical Sales department.
EDS 3000 pressure switches are individually tested and calibrated at a computer operated
test station. The pressure switches are maintenance-free and operate perfectly when used
according to the specifications (see Technical Specifications). However, if there is a cause
for complaint, please contact HYDAC Service.
Interference by anybody other than HYDAC personnel will invalidate all warranty claims.

2 Safety Instructions
Before commissioning, check the instrument and any accessories supplied. Before
commissioning, please read the operating instructions. Ensure that the instrument is
suitable for your application.

Incorrect use or the non-compliance with the operating instructions and specifications can
lead to damage to property and/or personal injury.

Ensure that the measured fluids in contact with the material are compatible and that the
overload pressure and burst pressure will not be exceeded.

3 Functions of the EDS 3000


Depending on which model you have, the instrument offers the following functions:
• Display of the actual pressure in PSI, MPa, bar or in a freely scalable range.
• Display of the maximum value or of a pre-set switching point.
• Switching of the switch outputs in accordance with the pressure and the pre-set
switching parameters.
• Analogue output
• Menu for the basic settings (EDS 3000 is adjustable to the particular application).
• Programming Enables

Status 28.06.2011 HYDAC ELECTRONIC GMBH Part no.: 669873


6 EDS 3400 / 3300 with flush membrane

4 Installation
Via its pressure port, EDS 3000 can be mounted directly to a hydraulic block.
Please ensure that the flush mounted (external) diaphragm is protected from mechanical
damage.
To ensure optimum positioning, the unit can be rotated 340° about its long axis, and the
display and keypad can be rotated through 270°.
The tightening torque for both G 1/2 thread connectors is 45 Nm.
The electrical connection must be carried out by a qualified electrician according to the
relevant regulations of the country concerned (VDE 0100 in Germany). The pressure switch

E
housing must be earthed correctly at the same time.

EDS 3000 pressure switches are marked. A conformity declaration is available on


request.
The EMV standards: EN 61000-6-1, -2, -3 und -4 apply.
However, the stipulations of those standards will be met only if the sensor's housing has
been correctly earthed by a qualified electrician. When fitted into a hydraulic block, earthing
the block via the hydraulic system is sufficient.

CAUTION:
When installing the EDS 3000 with an integral sensor probe, a suitable spanner
(AF 27) must be used on the hexagonal nut of the pressure port.
Inappropriate installation methods, such as fitting manually above the housing,
can, due to the rotatability of the EDS 3000, can damage the housing or even
cause the device to fail completely.

Additional installation suggestions which, from experience, reduce the effect of


electromagnetic interference:
• Make line connections as short as possible
• Use shielded lines (e.g. LIYCY 4 x 0.5 mm²)
• The cable screening must be fitted by qualified personnel subject to the ambient
conditions and with the aim of suppressing interference
• Keep the unit well away from the electrical supply lines of power equipment, as well as
from any electrical or electronic equipment causing interference

Status 28.06.2011 HYDAC ELECTRONIC GMBH Part no.: 669873


EDS 3400 / 3300 with flush membrane 7

5 Operation

6 Digital Display
After switching on the supply voltage, "EdS" appears briefly in the display
and stards displaying the actual pressure.

In the basic settings, the display can be changed as follows:


• Display of the maximum value "TOP"
The highest pressure measured in the system since the unit was last switched on or since
last reset is displayed permanently.
• Display of the set switchpoint "S.P. 1", "S.P. 2"
Depending on the variant, switchpoint 1 or switchpoint 2 can be displayed permanently.
• Display dark "OFF".
The display is switched off. Depending on the setting, "TOP", "S.P. 1", "S.P. 2" or "OFF"
briefly appears after the switch-on message. The actual pressure can be displayed briefly
pressing ◄ or ►.
• With the setting "Display current pressure" it is possible to display the maximum value by
pressing the button ◄ or ►.
It is then displayed for 3s. The maximum value can be reset in every display mode by
briefly pressing ◄ and ► simultaneously. This process is acknowledged by the message
"rES" on the display.

• If the current pressure exceeds the device's nominal pressure it can no longer be
displayed, and the display begins to flash.
• If the actual pressure is less than 0.75 % of the nominal range, 0 bar is displayed.

Status 28.06.2011 HYDAC ELECTRONIC GMBH Part no.: 669873


8 EDS 3400 / 3300 with flush membrane

7 Output Signals
7.1 Switching Outputs
The EDS 3000 has either 1 or 2 switch outputs. The following settings can be made under
the basic settings:

7.1.1 Switch Point Setting (SP)


One switchpoint and one hysteresis can be set for every switching output. The respective
output switches when the set switchpoint has been reached and it switches back when the
switch-back point has been fallen short of. The switch-back point is determined by the pre-

E
set hysteresis
(switch-back point = switching point minus hysteresis).

Abbreviations: "S.P.1", "S.P.2" = switching point 1 or 2


"H.Y.1", "H.Y.2" = Hysteresis 1 or 2

7.1.2 Window Mode (WIN)


The window function allows you to monitor a range. An upper and a lower switch value
defining the range can be entered for each switch output.
The relevant output will then switch when the pressure enters this range. When leaving this
range, i.e. when the switch-back value has been reached, the output switches back.
The lower switch-back value is just below the lower switching value (lower switching value
minus 3x the increment, see section 7.4).
The upper switch-back value is just above the upper switching value (upper switching value
plus 3x the increment, see section 7.4).
The range between the switch value and the switch-back value forms a safety margin
which prevents unwanted switching operations from being triggered (such as those
triggered by the pulsations of a pump).

Example for switching output 1 (normally open):

Abbreviations: "hi.1", "hi.2" = High level 1 or 2 = top switching value 1 or 2


"Lo 1", "Lo 2" = Low level 1 or 2 = lower switching value 1 or 2
"FS" (Full Scale) = relative to the full measuring range

The window function only works properly (switching on and off) if all switch values
(including the safety margin) are above 0 bar and below the nominal pressure
range.

Status 28.06.2011 HYDAC ELECTRONIC GMBH Part no.: 669873


EDS 3400 / 3300 with flush membrane 9

7.2 Analogue Output


Analogue output signal: 4 .. 20 mA or 0 .. 10 V
(can be set in the Basic Settings menu)

7.3 Setting the Switching Points and Hysteresis or Switching Values for
Window Mode
• Press the "mode" key.
• Either "S.P.1" or "Hi.1" will appear in the display.

E
• Actuate the "mode" button repeatedly to select the required parameter (when the
switchpoint function is switched on, "S.P.1", "hYS.1", "S.P.2", "hYS.2" appears, when
window mode is switched on (WIN) "Hi.1", "Lo.1", "Hi.2" oder "Lo.2" appears).
• After 2 seconds have elapsed, the actual setting will flash.
• Use the ◄ and ► keys to alter the settings.
• If necessary, use the "mode" key to select further parameters and the ◄ and ► keys to
alter the settings.
• If no key has been pressed for 3 seconds, the display will revert to displaying the primary
display and the settings will be saved to the unit.

Note:
• If "LOC" appears in the display when attempting to change a setting, programming
is disabled.
Action: Set the programming enable(s) to "FREE".
(see section 9 Programming Enables)
• If the ; key ◄ or key ► is held down while changing a setting, the value will
advance automatically.
• Whenever any settings have been altered, "PROG" appears briefly in the
display when the display switches over. The new setting is then saved in the
unit.

Status 28.06.2011 HYDAC ELECTRONIC GMBH Part no.: 669873


10 EDS 3400 / 3300 with flush membrane

7.4 Setting Ranges for the Switch Outputs


7.4.1 Switching Point Function

7.4.2 Window Function

* All ranges given in the table can be adjusted by the increments shown.

Status 28.06.2011 HYDAC ELECTRONIC GMBH Part no.: 669873


EDS 3400 / 3300 with flush membrane 11

8 Basic Settings
The function of the EDS 3000 can be adapted to suit the particular application by altering
several basic settings.
These settings are combined in a single menu.

8.1 Changing the Basic Settings

Important Note:
When the menu is activated, no switching operations are carried out!

Activating the Basic Settings menu: E

To exit the Basic Settings menu:

Select the "END" menu item, set to "YES": the EDS 3000 returns to normal display mode
after 2 s.

Note:
If no key is pressed for approx. 25 seconds, the menu closes automatically, and
any changes that may have been made will not be saved.

Status 28.06.2011 HYDAC ELECTRONIC GMBH Part no.: 669873


12 EDS 3400 / 3300 with flush membrane

8.2 Overview of the Basic Settings

Status 28.06.2011 HYDAC ELECTRONIC GMBH Part no.: 669873


EDS 3400 / 3300 with flush membrane 13

Status 28.06.2011 HYDAC ELECTRONIC GMBH Part no.: 669873


14 EDS 3400 / 3300 with flush membrane

Caution:
Following a zero point adjustment, for example on a 600 bar instrument, a
pressure of up to 18 bar will be displayed as 0 bar. Before any work is carried
out on the hydraulic system, ensure that the system is depressurised.

If any basic settings have been altered, "ProG" appears briefly in the display as soon as the
basic settings menu is closed and then the value selected for primary display is shown.

Status 28.06.2011 HYDAC ELECTRONIC GMBH Part no.: 669873


EDS 3400 / 3300 with flush membrane 15

9 Programming Enables
The unit has 2 program enables which must both be enabled in order to modify the
settings. The operating program enable can be set or cancelled during operation. It
provides protection against unintentional alterations of settings.
A programming disable via the main programming enable prevents any changes being
made to the settings during operation. This serves, for example, as a safety function or as
protection against unauthorised alterations.

9.1 Changing the Main Programming Enable

9.2 Changing the Main Programming Enable


Switch off the supply voltage or disconnect the instrument from the supply voltage.

Note:
Whenever any settings have been altered, "PROG" appears briefly in the
display when the display switches over. The new setting is then saved in the
unit.

Status 28.06.2011 HYDAC ELECTRONIC GMBH Part no.: 669873


16 EDS 3400 / 3300 with flush membrane

10 Error Messages
If an error is detected, a corresponding error message appears which must be
acknowledged by pressing any button. Possible error messages are as follows:

Status 28.06.2011 HYDAC ELECTRONIC GMBH Part no.: 669873


EDS 3400 / 3300 with flush membrane 17

11 Pin Assignment

Status 28.06.2011 HYDAC ELECTRONIC GMBH Part no.: 669873


18 EDS 3400 / 3300 with flush membrane

12 Technical Data:
Input data EDS 3300 EDS 3400
Measuring ranges in bar -1..1; 1; 2.5; 6; 10; 16 40; 100; 250; 400; 600
Overload pressures in bar 3; 3; 8; 18; 30; 48 80; 200; 500; 800; 900
Burst pressures 1) in bar 5; 5; 15; 30; 50; 80 200; 500; 1000; 2000; 2000
Mechanical connection G1/2A DIN 3852
G1/2 with additional front O-ring seal
Pressure transfer fluid Silicon-free oil
Torque rating 45 Nm
Parts in contact with fluid 2) Mech. connection: Stainless steel Mech. connection: Stainless steel
(measuring cell ceramics) (measuring cell stainless steel)

E
Seal: FPM Seal: FPM
O-Ring: NBR Seal: FPM
O-Ring: NBR
Output data
Accuracy in accordance with DIN ≤ ± 0.5 % FS typ.
16086, ≤ ± 1,0 % FS max
Limit point setting
(display, analogue output)
Repeatability ≤ ± 0.25 % FS max.
Temperature drift ≤ ± 0.025 % FS / °C max. zero point
≤ ± 0.025 % FS / °C max. range
Analogue output (optional)
Output signal (selectable) 4 .. 20 mA load max. 500 Ω
0 .. 10 V load min. 1 kΩ
Switching outputs
Type PNP transistor output
Switch current Max. 1.2 A per output
Switch cycles > 100 million
Reaction time ~ 10 ms
Long term drift ≤ ± 0.3 % FS / year typ.
DESINA® diagnostic signal (Pin 2)
Functioning OK: HIGH level / not OK: LOW level
Level HIGH: approx. +UB / LOW: < +0.3 V
Environmental Conditions
Compensated temperature range -10 .. +70 °C
0 ..
Operating temperature range 3) -25 .. +80°C
Storage temperature range -40 .. +80°C
Fluid temperature range: -40 .. +80 °C
- mark EN 61000-6-1 / 2 / 3 / 4
Vibration resistance according to ≤ 10 g
DIN EN 60068-2-6 at 10 .. 500 Hz
Shock resistance to ≤ 50 g
DIN EN 60068-2-29 (11 ms)
Protect. class to DIN 40050 IP 67
Other data
Supply voltage 9 .. 35 V DC without analogue output
18 .. 35 V DC with analogue output
Residual ripple of supply voltage ≤5%
Current consumption max. 35 mA (inactive switching output)
Display 4-digit, LED, 7 segment, red,
height of digits 7 mm
Weight ~ 120 g
1)
G1/2 with additional front O-ring seal max. 1500 bar
2)
Other seal materials available on request
3)
For G1/2 with additional front O-ring seal -10 .. +80 °C

Status 28.06.2011 HYDAC ELECTRONIC GMBH Part no.: 669873


EDS 3400 / 3300 with flush membrane 19

13 Model Code

13.1 Standard Version


EDS 3 X Z X – X – XXXX – XXX – 000

Accuracy
3 = 1 % FS max., ceramics (thick film), relative
4 = 1 % FS max., stainless steel (thin film), relative

Mechanical Process Connection


Z = Flush E
Electrical Connection
6 = Male connector M12x1, 4-pole
For output options "1", "2" and "3" only
8 = Male connector M12x1, 5-pole
For output option "5" only

Output
1 = 1 switching output
only in conjunction with electrical connection type "6"
2 = 2 switching outputs
only in conjunction with electrical connection type "6"
3 = 1 switching output and 1 analogue output
only in conjunction with electrical connection type "6"
5 = 2 switching outputs and 1 analogue output
only in conjunction with electrical connection type "8"

Pressure Ranges in bar


0001 (-1..1)
01.0; 02.5; 06.0; 0010; 0016
For accuracy "3" only

0040; 0100; 0250; 0400; 0600


For accuracy "4" only

Mechanical Connection
G01 = G1/2 A, DIN 3852
G02 = G1/2 with additional front O-ring seal

Modification Number
000 = Standard

Notes:
Special models on request.
On instruments with a different modification number, please read the label or the
technical amendment details supplied with the instrument.

Status 28.06.2011 HYDAC ELECTRONIC GMBH Part no.: 669873


20 EDS 3400 / 3300 with flush membrane

13.2 DESINA® Compliant Type or Compatible


with DESINA®:
EDS 3 X Z X – X – XXXX – XXX – 000

Accuracy
3 = 1 % FS max., ceramics (thick film), relative

E
4 = 1 % FS max., stainless steel (thin film), relative

Mechanical Process Connection


Z = Flush

Electrical Connection
8 = Male connector M12x1, 5-pole

Output
1 = 1 switching output
only in conjunction with electrical connection type "6"
3 = 1 switching output and 1 analogue output
only in conjunction with electrical connection type "6"

Pressure Ranges in bar


0001 (-1..1)
01.0; 02.5; 06.0; 0010; 0016
For accuracy "3" only

0040; 0100; 0250; 0400; 0600


For accuracy "4" only

Mechanical Connection
G01 = G1/2 A, DIN 3852
G02 = G1/2 with additional front O-ring seal

Modification Number
D00 = pin configuration conforms to DESINA® standard for self-diagnostics

Notes:
Special models on request.
On instruments with a different modification number, please read the label or the
technical amendment details supplied with the instrument.

Status 28.06.2011 HYDAC ELECTRONIC GMBH Part no.: 669873


EDS 3400 / 3300 with flush membrane 21

14 Accessoires
14.1 For Electrical Connection
14.1.1 For Use with Output Options "1", "2" and "3"

14.1.2 For all Output Models and DESINA® Version

Status 28.06.2011 HYDAC ELECTRONIC GMBH Part no.: 669873


22 EDS 3400 / 3300 with flush membrane

14.2 For Mechanical Connection

Status 28.06.2011 HYDAC ELECTRONIC GMBH Part no.: 669873


EDS 3400 / 3300 with flush membrane 23

15 Dimensions

Status 28.06.2011 HYDAC ELECTRONIC GMBH Part no.: 669873


24 EDS 3400 / 3300 with flush membrane

HYDAC ELECTRONIC GMBH


Hauptstr. 27
D-66128 Saarbrücken
Germany

Web: www.hydac.com
Email: electronic@hydac.com
Phone: +49 (0)6897 509-01

E Fax.: +49 (0) 6897 509-1726

HYDAC Service
If you have any questions concerning repairwork, please don’t hesitate to contact
HYDAC Service:

HYDAC SERVICE GMBH


Hauptstr. 27
D-66128 Saarbrücken
Germany

Phone: +49 (0)6897 509-1936


Fax.: +49 (0)6897 509-1933

NOTE

The information and particulars provided in this manual apply to the operating
conditions and applications described herein. For applications and/or operating
conditions not described here, please contact the relevant technical department.

If you have any questions, suggestions, or encounter any problems of a technical


nature, please contact your Hydac representative.

Subject to technical modifications.

Status 28.06.2011 HYDAC ELECTRONIC GMBH Part no.: 669873


EDS 3400 / 3300 avec membrane affleurante 3

Sommaire
1 Généralités __________________________________________________ 5
2 Consignes de sécurité _______________________________________ 5
3 Fonctions de l'EDS 3000 _____________________________________ 5
4 Montage _____________________________________________________ 6
5 Commande __________________________________________________ 7
6 Affichage digital _____________________________________________ 7
7 Signaux de sortie ____________________________________________ 8
7.1 Sorties de commutation ____________________________________________________________8
7.1.1 Fonction seuil (SP) ________________________________________________________________8
7.1.2 Fonction fenêtre (WIN) ____________________________________________________________8
7.2 Sortie analogique ____________________________________________________________________9

F
7.3 Réglage des points d'enclenchement et d'hystérésis ou des valeurs de
commutation pour la fonction fenêtre _________________________________________________9
7.4 Plages de réglage pour les sorties de commutation __________________________10
7.4.1 Fonction point de commutation _________________________________________________10
7.4.2 Fonction fenêtre __________________________________________________________________10
8 Paramètres de base _________________________________________ 11
8.1 Modifier les réglages de base_____________________________________________________11
8.2 Aperçu des réglages de base _____________________________________________________12
9 Autorisations de programmation ____________________________ 15
9.1 Modification de l'autorisation de programmation basse _____________________15
9.2 Modification de l'autorisation de programmation haute _____________________15

10 Messages d'erreur __________________________________________ 16


11 Raccordement électrique ___________________________________ 17
12 Caractéristiques techniques ________________________________ 18
13 Code de commande ________________________________________ 19
13.1 Exécution standard _______________________________________________________________19
13.2 Exécution conforme à DESINA® ou raccordable à DESINA® ______________20

14 Accessoires ________________________________________________ 21
14.1 Pour le raccordement électrique _______________________________________________21
14.1.1 Pour les variantes de base "1" "2" et "3" _____________________________________21
14.1.2 Pour tout signal de sortie et execution DESINA® ____________________________21
14.2 Pour le raccordement mécanique ______________________________________________22
15 Dimensions ________________________________________________ 23

Edition 28.06.2011 HYDAC ELECTRONIC GMBH Code article : 669873


4 EDS 3400 / 3300 avec membrane affleurante

Avant-propos

À l'attention des acquéreurs d'un produit que nous avons


fabriqué. Nous avons rassemblé dans cette documentation les
recommandations essentielles pour l'utilisation et la
maintenance de ce produit.

Cette notice a pour objectif de simplifier la prise de


connaissance du produit et de permettre une exploitation
optimale de ses possibilités d'utilisation, conformément à
l'usage prévu.

Ce document doit toujours être disponible sur le lieu


d'utilisation.

F
Veuillez noter que les informations fournies dans cette
documentation correspondent à la technique de l'appareil au
moment de l'élaboration de ce document. Pour cette raison, les
différentes données techniques, illustrations et mesures sont
susceptibles de diverger.

Si, lors de la lecture de ce manuel, vous deviez détecter des


erreurs ou encore si vous aviez des suggestions ou des
remarques, veuillez vous adresser à :

HYDAC ELECTRONIC GMBH


Documentation technique :
Hauptstrasse 27
D-66128 Saarbrücken
Allemagne
Tél. : +49(0)6897 / 509–01
Fax: +49(0)6897 / 509-1726
Email: electronic@hydac.com

La rédaction vous est reconnaissante de votre participation.

"De la pratique vers la pratique".

Edition 28.06.2011 HYDAC ELECTRONIC GMBH Code article : 669873


EDS 3400 / 3300 avec membrane affleurante 5

1 Généralités
Le transmetteur de pression EDS 3000 avec membrane affleurante a été spécialement
conçu pour les applications dans lesquelles les fluides utilisés peuvent colmater, coller ou
geler un raccord de pression standard.
Les processus avec fluides alternants, dans lesquels des résidus peuvent provoquer un
mélange ou des impuretés, constituent d'autres champs d'application de l'EDS 3000.

Le raccordement pression est raccordé de manière frontale grâce à une membrane en inox
soudée et il est rempli en interne avec un liquide de remplissage du séparateur qui
transmet hydrostatiquement la pression process à la cellule de mesure.

En cas de questions concernant les données techniques et l'aptitude d'utilisation de


l'appareil, veuillez vous adresser à notre service commercial.
Chaque transmetteur de pression basé sur la série EDS 3000 est aligné et soumis à un
test final sur un poste d'essai assisté par ordinateur. Les appareils ne nécessitent aucun
entretien et fonctionnent parfaitement dans les conditions d’utilisation (voir données

F
techniques) spécifiées. Si malgré tout un dysfonctionnement devait survenir, veuillez vous
mettre en relation avec HYDAC-Service.
Toute intervention extérieure dans l’appareil entraînent l’annulation de la garantie.

2 Consignes de sécurité
Avant la première mise en service, merci de vérifier le bon état du matériel et de ses
accessoires éventuels. Lisez la notice d’utilisation avant la mise en service, et assurez-
vous que l’appareil convienne à votre utilisation.

Une mauvaise manipulation comme par exemple le non respect des caractéristiques
techniques ou une mauvaise mise en œuvre peut causer des dégâts matériels et/ou
humains.

Vérifier la compatibilité des fluides de mesure avec les matériaux utilisés par le
manocontacteur. Tenir compte également de la pression de surcharge et la pression
d'éclatement.

3 Fonctions de l'EDS 3000


Selon les versions choisies, l'appareil propose les fonctions suivantes:
• Affichage de la valeur de mesure de la pression actuelle en PSI, MPa, bar ou bien d'un
plage de mesure de choix libre.
• Affichage de la valeur maximum ou d'un point de commutation figé.
• Mode de commutation des sorties selon la pression et paramètres de commutation
réglés
• Sortie analogique
• Menu de base pour la programmation (adaptation de l'ETS 3000 à chaque application)
• Autorisations de programmation

Edition 28.06.2011 HYDAC ELECTRONIC GMBH Code article : 669873


6 EDS 3400 / 3300 avec membrane affleurante

4 Montage
L'EDS 3000 peut être monté directement sur un bloc hydraulique au moyen d'un raccord
de pression.
Il faut ici veiller à ce que la membrane affleurante (externe) soit protégée de dommages
mécaniques.
Pour une orientation optimale, il est possible d'effectuer une rotation de 340° de l'axe
longitudinal et de 270° de l'affichage, touches de réglage comprises.
Le couple de serrage pour chaque raccord taraudé G 1/2 est à 45 Nm.
Le raccordement doit être réalisé par du personnel qualifié selon chaque prescription
nationale (VDE 0100 en Allemagne). Le corps du manocontacteur doit, dans ce cadre, être
relié à la terre.

Les manocontacteurs EDS 3000 portent le sigle . Un certificat de conformité est


disponible sur demande.
Les normes EMV applicables: EN 61000–6–1,–-2,–-3 und –-4.
Les éxigences de ces normes ne seront atteintes que par une mise à la terre en bon et du

F
forme du corps de l’EDS. En cas de vissage dans un bloc hydraulique, le fait que le bloc
est relié à la terre via le système hydraulique est suffisant.

ATTENTION !
Le vissage de l'EDS 3000 doit être effectué à l'aide d'une clé adaptée (27) sur
l'écrou hexagonal de l’embase de raccordement.
Un mauvais montage, comme par exemple le vissage par le corps du capteur,
en raison de l'orientation de l'EDS 3000, peut endommager le corps de l'appareil
et peut entrainer la destruction du capteur.

Remarques complémentaires, pour diminuer l'influence des perturbations


électromagnétiques:
• Les raccords électriques doivent être les plus courts possibles.
• Utiliser des câbles blindés (p.ex. LIYCY 4 x 0,5 mm²)
• Le câble blindé est à mettre en œuvre en fonction des conditions environnantes par des
spécialistes afin de diminuer les perturbations électromagnétiques.
• Eviter si possible de placer l'appareil à proximité d'appareils électriques ou électroniques
générateurs de perturbations électromagnétiques.

Edition 28.06.2011 HYDAC ELECTRONIC GMBH Code article : 669873


EDS 3400 / 3300 avec membrane affleurante 7

5 Commande

6 Affichage digital
F
Après mise sous tension, l'appareil affiche brièvement "EdS" et commence avec l'affichage de
la pression actuelle.

Dans les réglages de base, il est possible de modifier l'affichage comme suit :
• Affiche la valeur maximale "TOP"
Affiche en permanence la valeur max de la pression mesurée depuis la mise en service
de l'appareil ou depuis le dernier rafraîchissement.
• Affiche la valeur du seuil d'enclenchement réglé "S.P. 1", "S.P. 2"
Selon la version, le seuil d'enclenchement 1 ou seuil d'enclenchement 2 peut être affiché
en permanence.

• Afficheur éteint "OFF"


L'affichage est hors tension.
Selon les réglages, "TOP", "S.P 1", "SP 2" ou "OFF" est brièvement affiché après
l'information de mise en service. La pression actuelle peut être affichée brièvement en
utilisant la touche ◄ ou ►.
• Lors du réglage de l' "affichage de la pression actuelle", il est possible, en appuyant sur la
touche ◄ ou ►, d'afficher la valeur maximale.
Celle-ci apparaît pendant 3s. La valeur maximale peut être réinitialisée dans chaque
mode d'affichage par une courte pression simultanée des touches ◄ et ►. Cette
procédure est confirmée par le message “rES“ à l'écran.

• Si la pression actuelle dépasse la pression nominale de l'appareil, alors celle-ci ne


peut plus être affichée et l'affichage clignote.
• Si la pression actuelle se trouve en dessous de 0,75% de la zone nominale,
l'appareil affiche 0 bar.

Edition 28.06.2011 HYDAC ELECTRONIC GMBH Code article : 669873


8 EDS 3400 / 3300 avec membrane affleurante

7 Signaux de sortie
7.1 Sorties de commutation
L'EDS 3000 dispose d'1 ou 2 sorties de commmutation. Dans le menu de base, plusieurs
configurations sont possibles :

7.1.1 Fonction seuil (SP)


Pour chaque sortie de commutation, il est possible de régler un seuil d'enclenchement et
une hystérésis. La sortie concernée commute quand la valeur programmée est atteinte;
elle revient à l'état initial quand la pression passe sous le seuil de déclenchement en
descendant. Le seuil de déclenchement est déterminé par l'hystérésis réglée (seuil de
déclenchement = seuil d'enclenchement moins l'hystérésis)

Abréviations : "S.P.1", "S.P.2" = point d'enclenchement 1 ou 2


"H.Y.1", "H.Y.2" = Hysterésis 1 ou 2

7.1.2 Fonction fenêtre (WIN)

F
La fonction fenêtre permet de surveiller un seuil. Pour chaque sortie de commutation il est
possible d'entrer une valeur de commutation supérieure et inférieure qui définissent le
seuil.
La sortie commute lorsque la pression entre dans cette plage. Lorsqu'elle quitte la plage,
c'est à dire que la valeur de déclenchement est atteinte, la sortie revient à l'état initial.
La valeur de déclenchement inférieure se situe tout juste au dessous de la valeur
d'enclenchement inférieure (valeur d'enclenchement inférieur moins le triple incrément, voir
chapître 7.4).
La valeur de déclenchement supérieure se situe tout juste au-dessus de la valeur
d'enclenchement supérieure (valeur d'enclenchement supérieur plus le triple incrément,
voir chapître 7.4).
La plage entre le point d'enclenchement et le point de déclenchement constitue une zone
de sécurité qui permet d'éviter les déclenchement intempestifs (par ex. par les pulsations
d'une pompe).

Exemple pour sortie de commutation 1 (contacts NO) :

Abréviations: "hi.1", "hi.2" = High level 1 ou 2 = point de commutation bas 1 ou 2


"Lo.1", "Lo.2" = Low level 1 ou 2 = valeur d'enclenchement inférieure 1 ou 2
"PE" (Pleine Echelle) = par rapport à la totalité de la plage de mesure

La fonction fenêtre ne fonctionne que si tous les paramètres (y compris la zone de


sécurité) sont supérieurs à 0 bar et inférieurs à la plage de pression nominale.
Edition 28.06.2011 HYDAC ELECTRONIC GMBH Code article : 669873
EDS 3400 / 3300 avec membrane affleurante 9

7.2 Sortie analogique


Signal de sortie analogique: 4 .. 20 mA ou 0 .. 10 V
(paramétrable dans le menu de base)

7.3 Réglage des points d'enclenchement et d'hystérésis ou des valeurs


de commutation pour la fonction fenêtre
• Appuyer sur mode "mode" .
• "S.P.1" ou "hi.1" apparaît dans l'affichage
• En ré-appuyant sur la touche "mode", il est possible de sélectionner le paramètre voulu
("S.P.1", "H.Y.1" à "S.P.2", "H.Y.2") ou alors ("Hi.1", "Lo.1" à "Hi.2" ou "Lo.2") quand on
travaille en fonction fenêtre.
• Après 2 secondes, le paramètre actif clignote.
• Modifier le paramétrage avec les touches ◄ ou ►.
• Sélectionner éventuellement d'autres paramètres avec la touche "mode" et modifier les
réglages avec les touches ◄ et ►.

F
• Après 3 secondes sans appuyer sur une des touches, les valeurs réglées sont
sauvegardées et l'affichage initial est alors redonné.

Remarques :
• S'il apparaît "LOC" à l'affichage lors de votre procédure de réglage, cela signifie
que la programmation n'est pas possible.
Solution : Pour modifier les valeurs, l'autorisation de programmation, doit être
réglée sur "FREE". Voir paragraphe 9 "autorisation de programmation".
• Lors de la modification; un appui maintenu sur la touche ◄ oder ►, incrémente ou
décrémente la valeur.
• Quand une valeur est modifiée, il apparaît brièvement lors de la commutation
l'indication "PROG". La nouvelle valeur est sauvegardée.

Edition 28.06.2011 HYDAC ELECTRONIC GMBH Code article : 669873


10 EDS 3400 / 3300 avec membrane affleurante

7.4 Plages de réglage pour les sorties de commutation


7.4.1 Fonction point de commutation

F
7.4.2 Fonction fenêtre

* Toutes les plages indiquées dans le tableau sont réglables en fonction du pas indiqué.

Edition 28.06.2011 HYDAC ELECTRONIC GMBH Code article : 669873


EDS 3400 / 3300 avec membrane affleurante 11

8 Paramètres de base
Pour s'adapter à chaque application, le fonctionnement de l'EDS 3000 peut être paramétré
grâce à plusieurs réglages de base.
Ceux-ci sont rassemblés dans un menu.

8.1 Modifier les réglages de base

Remarque :
Quand le menu de base est activé, les fonctions de commutations sont
désactivées.

Accéder aux réglages de base :

Sortir du menu réglages de base :

Sélectionner l’option "END", choisir "YES", appuyer sur la touche MODE, l'EDS 3000
repasse en mode normal après 2 s.

Remarque :
Si au bout de 25 secondes, aucune touche n'a été appuyée, l'affichage retourne
à son état de fonctionnement normal sans sauvegarde des modifications.

Edition 28.06.2011 HYDAC ELECTRONIC GMBH Code article : 669873


12 EDS 3400 / 3300 avec membrane affleurante

8.2 Aperçu des réglages de base

Edition 28.06.2011 HYDAC ELECTRONIC GMBH Code article : 669873


EDS 3400 / 3300 avec membrane affleurante 13

Edition 28.06.2011 HYDAC ELECTRONIC GMBH Code article : 669873


14 EDS 3400 / 3300 avec membrane affleurante

Attention :
Après un alignement du point zéro une pression de 18 bar à 0 bar est affiché
pour un appareil de 600 bar. Avant tous travaux sur l'installation hydraulique, il
faut s'assurer que celle-ci est exempte de pression.

Quand les paramètres du menu de base ont été modifiés, "Prog" apparaît un court instant
afin de valider les modifications. Puis l'appareil affiche la valeur de l'affichage primaire
selectionné.

Edition 28.06.2011 HYDAC ELECTRONIC GMBH Code article : 669873


EDS 3400 / 3300 avec membrane affleurante 15

9 Autorisations de programmation
Cet appareil dispose de deux autorisations de programmation. Chacune doit être accordée
pour modifier les réglages. L'autorisation de programmation basse peut être accordée ou
annulée pendant le fonctionnement. Ceci sert comme protection contre des modifications
non volontaires.
Un verrouillage de la programmation via l'autorisation principale de programmation a pour
effet qu'aucune modification des paramètres ne peut être effectuée lors du fonctionnement.
Ceci sert p.ex. comme fonction de sécurité ou comme protection contre des modifications
non autorisées.

9.1 Modification de l'autorisation de programmation basse

9.2 Modification de l'autorisation de programmation haute


Couper la tension d'alimentation ou démonter l'unité de la tension d'alimentation

Remarque :
Quand une valeur est modifiée, "PROG" apparaît brièvement dans l'affichage
lors de la commutation. La nouvelle valeur est sauvegardée.

Edition 28.06.2011 HYDAC ELECTRONIC GMBH Code article : 669873


16 EDS 3400 / 3300 avec membrane affleurante

10 Messages d'erreur
Dès que l'appareil détecte une erreur, un message d'erreur est affiché : celui-ci doit être
acquitté avec n'importe quelle touche. Les différents messages d'erreurs sont :

Edition 28.06.2011 HYDAC ELECTRONIC GMBH Code article : 669873


EDS 3400 / 3300 avec membrane affleurante 17

11 Raccordement électrique

Edition 28.06.2011 HYDAC ELECTRONIC GMBH Code article : 669873


18 EDS 3400 / 3300 avec membrane affleurante

12 Caractéristiques techniques
Valeurs d'entrée EDS 3300 EDS 3400
Plages de mesure en bar -1..1; 1; 2,5; 6; 10; 16 40; 100; 250; 400; 600
Surpression en bar 3; 3; 8; 18; 30; 48 80; 200; 500; 800; 900
Pression d‘éclatement 1) en bar 5; 5; 15; 30; 50; 80 200; 500; 1000; 2000; 2000
Raccordement mécanique G1/2A DIN 3852
G1/2 avec joint torique frontal supplémentaire
Liquide de remplissage du séparateur Huile exempte de silicone
Couple de serrage 45 Nm
Materiaux en contact avec le fluide 2) Raccord : acier inoxydable Raccord : acier inoxydable
(cellule de mesure en céramique) (cellule de mesure en acier inoxydable)
Joint : FPM Joint : FPM
Joint torique : NBR Joint torique : NBR
Valeurs de sortie
Précision selon DIN 16086, ≤ ± 0,5 % PE typ
Réglage du seuil ≤ ± 1,0 % FS max
(affichage, sortie analogique)
Reproductibilité ≤ ± 0,25 % PE max.
Dérive en température ≤ ± 0,025 % PE / °C max. point zéro
≤ ± 0,025 % FS / °C max. étendue

F
Sortie analogique (en option)
Signal de sortie (sélectionnable) 4 .. 20 mA charge max. 500 Ω
0 .. 10 V charge min. 1 kΩ
Sorties de commutation
Exécution PNP sortie de commutation transistor
Courant de commutation max. 1,2 A par sortie
Cycles de commutation > 100 millions
Temps de réaction ~ 10 ms
Dérive dans le temps ≤ ± 0,3 % PE / an typ.
Signal de diagnostic DESINA® (broche 2)
Fonction I.O.: niveau HIGH / N.I.O.: niveau LOW
Niveau HIGH: ~ +UB / LOW: < +0,3 V
Conditions environnementales
Plage de température compensée -10 .. +70 °C
Plage de température d'utilisation 3) -25 .. +80 °C
Plage de température de stockage -40 .. +80 °C
Plage de température du fluide -40 .. +80 °C
Sigle EN 61000-6-1 / 2/ 3/ 4
Résistance aux vibrations ≤ 10 g
DIN EN 60068-2-6 à 10 .. 500 Hz
Résistance aux chocs selon ≤ 50 g
DIN EN 60068-2-29 (11 ms)
Indice de protection selon DIN 40050 lP 67
Autres valeurs
Tension d'alimentation 9 .. 35 V DC sans sortie analogique
18 .. 35 V DC avec sortie analogique
Oscillation résiduelle de la tension ≤5%
d'alimentation
Consommation de courant max. 35 mA (sortie de commutation inactive)
Affichage 4 caractères, LED, 7 segments rouge, hauteur 7 mm
Masse ~ 120 g
1)
G1/2 avec joint torique frontal supplémentaire max. 1500 bar
2)
Autres matériaux de joints sur demande
3)
G1/2 avec joint torique frontal supplémentaire -10 .. +80 C

Edition 28.06.2011 HYDAC ELECTRONIC GMBH Code article : 669873


EDS 3400 / 3300 avec membrane affleurante 19

13 Code de commande

13.1 Exécution standard


EDS 3 X Z X – X – XXXX – XXX – 000

Précision
3 = 1 % FS max., céramique (couche épaisse), relative
4 = 1 % FS max., acier inox (couche mince), relative

Raccordement procès mécanique


Z = Affleurant

Raccord électrique
6 = Embase M12x1, 4pôles
pour type de signaux de sortie "1“, "2“ et "3“ uniquement
8 = Embase M12x1, 5pôles

F
pour type de signaux de sortie "5" uniquement

Sortie
1 = 1 sortie de commutation
avec type de raccordement électrique "6" uniquement
2 = 2 Sorties de commutation
avec type de raccordement électrique "6" uniquement
3 = 1 sortie de commutation et 1 sortie analogique
avec type de raccordement électrique "6" uniquement
5 = 2 contacts + 1 sortie analogique
avec type de raccordement électrique "8" uniquement

Plages de pression en bar


0001 (-1..1)
01,0; 02,5; 06,0; 0010; 0016
(Uniquement pour précision "3")

0040; 0100; 0250; 0400; 0600


(Uniquement pour précision "4")

Raccord mécanique
G01 = G1/2 A, DIN 3852
G02 = G1/2 avec joint torique frontal supplémentaire

Numéro de modification
000 = standard

Remarques :
Exécutions spéciales sur demande.
Pour les appareils ayant un autre indice de modification, veuillez respecter la plaque signalétique
ou la description des modifications techniques jointe à la livraison.

Edition 28.06.2011 HYDAC ELECTRONIC GMBH Code article : 669873


20 EDS 3400 / 3300 avec membrane affleurante

13.2 Exécution conforme à DESINA® ou raccordable à DESINA®


EDS 3 X Z X – X – XXXX – XXX – D00

Précision
3 = 1 % FS max., céramique (couche épaisse), relative
4 = 1 % FS max., acier inox (couche mince), relative

Raccordement procès mécanique


Z = Affleurant

Raccord électrique
8 = Embase M12x1, 5pôles

Sortie

F
1 = 1 sortie de commutation
avec type de raccordement électrique "6" uniquement
3 = 1 sortie de commutation et 1 sortie analogique
avec type de raccordement électrique "6" uniquement

Plages de pression en bar


0001 (-1..1)
01,0; 02,5; 06,0; 0010; 0016
(Uniquement pour précision "3")

0040; 0100; 0250; 0400; 0600


(Uniquement pour précision "4")

Raccord mécanique
G01 = G1/2 A, DIN 3852
G02 = G1/2 avec joint torique frontal supplémentaire

Numéro de modification
D00 = Raccordement avec auto-diagnostic conforme à DESINA®

Remarques :
Exécutions spéciales sur demande.
Pour les appareils ayant un autre indice de modification, veuillez respecter la plaque signalétique
ou la description des modifications techniques jointe à la livraison.

Edition 28.06.2011 HYDAC ELECTRONIC GMBH Code article : 669873


EDS 3400 / 3300 avec membrane affleurante 21

14 Accessoires
14.1 Pour le raccordement électrique
14.1.1 Pour les variantes de base "1" "2" et "3"

14.1.2 Pour tout signal de sortie et execution DESINA®

Edition 28.06.2011 HYDAC ELECTRONIC GMBH Code article : 669873


22 EDS 3400 / 3300 avec membrane affleurante

14.2 Pour le raccordement mécanique

Edition 28.06.2011 HYDAC ELECTRONIC GMBH Code article : 669873


EDS 3400 / 3300 avec membrane affleurante 23

15 Dimensions
Diamètre

Embase

Orientation
afficheur 270° F
Orientation
corps 340°

Joint
elastomère

Ètanchéité DIN3869

Edition 28.06.2011 HYDAC ELECTRONIC GMBH Code article : 669873


24 EDS 3400 / 3300 avec membrane affleurante

HYDAC ELECTRONIC GMBH


Hauptstr. 27
D-66128 Saarbrücken
Allemagne

Web: www.hydac.com
Email: electronic@hydac.com
Tél. : +49(0)6897 / 509-01
Fax: +49(0)6897 / 509-1726

F
HYDAC Service
HYDAC Service se tient à votre disposition pour toute question concernant les
réparations.

HYDAC SERVICE GMBH


Hauptstr. 27
D-66128 Saarbrücken
Allemagne

Tél. : +49(0)6897/509-1936
Fax : +49(0)6897/509-1933

Remarques

Les données de ce prospectus se réfèrent aux conditions de fonctionnement et


d’utilisation décrites. Pour des cas d'utilisation et/ou conditions de fonctionnement
différents, veuillez vous adresser au service technique compétent.

Pour toute question technique, demande de conseils ou en cas de panne, veuillez vous
mettre en relation avec votre représentation HYDAC.

Sous réserve de modifications techniques.